Transcript
Page 1: Pipet Calibration - Wisconsin DNR

1

Calibration of Calibration of Mechanical Pipettes in Mechanical Pipettes in

the Modern the Modern Wastewater LaboratoryWastewater Laboratory

DeWayne KennedyDeWayne Kennedy--Parker and Parker and Roger SchultzRoger Schultz

Copyright 2007, University of Wisconsin Board of Regents

Page 2: Pipet Calibration - Wisconsin DNR

2

Any reference to product or Any reference to product or company names does not constitute company names does not constitute

endorsement by the Wisconsin endorsement by the Wisconsin State Laboratory of Hygiene, or State Laboratory of Hygiene, or

the University of Wisconsin.the University of Wisconsin.

DisclaimerDisclaimer

Page 3: Pipet Calibration - Wisconsin DNR

3

Class A volumetric pipettes

• Do not require periodic calibration• Are very accurate and precise

• Require practice to master the technique of use

• Have defined volumes• Are usually broken if dropped

with a bulb

Page 4: Pipet Calibration - Wisconsin DNR

4

These are fine.

But, maybe it’s time to go “Mechanical”.

One adjustable One adjustable volume mechanical volume mechanical pipette can replace pipette can replace a number of single a number of single volume pipettes.volume pipettes.

Page 5: Pipet Calibration - Wisconsin DNR

5

Mechanical airMechanical air--displacement displacement pipettes are becoming more pipettes are becoming more commonly used in the modern commonly used in the modern wastewater laboratory due to their wastewater laboratory due to their many advantages over traditional many advantages over traditional glass pipettes.glass pipettes.

Page 6: Pipet Calibration - Wisconsin DNR

65 - 27

Page 7: Pipet Calibration - Wisconsin DNR

7

Where would mechanical Where would mechanical pipettes be used in a wastewater pipettes be used in a wastewater

laboratory?laboratory?

•• Chlorine TestingChlorine Testing•• Phosphorous Testing, essential if using Phosphorous Testing, essential if using

TestTest--nn--tube® methodtube® method•• Standards PreparationStandards Preparation•• Any pipetting operation up to 10 mL that is Any pipetting operation up to 10 mL that is

repeated often in the laboratoryrepeated often in the laboratory

Page 8: Pipet Calibration - Wisconsin DNR

8

Mechanical air displacement pipette Mechanical air displacement pipette operation.operation.

Operating button is Operating button is pressed to the first pressed to the first stop, moving the stop, moving the piston and expelling piston and expelling a specified volume of a specified volume of air from the tipair from the tip

Page 9: Pipet Calibration - Wisconsin DNR

9

After immersing the tip into the liquid, the operating button is released. This creates a partial vacuum and the specified volume of liquid is aspirated into the tip.

Page 10: Pipet Calibration - Wisconsin DNR

10

When the operating button is pressed to the first stop again, the air dispenses the liquid. To empty the tip completely, the operating button is pressed to the second stop (blow out).

Page 11: Pipet Calibration - Wisconsin DNR

11

Costs are higher, initiallyCosts are higher, initially

•• Most class A glass volumetric pipettes Most class A glass volumetric pipettes $10 to $40 ( depending on the size)$10 to $40 ( depending on the size)

•• Mechanical (nonMechanical (non--electronic) electronic) $200 to $400 $200 to $400

•• Electronic ( multiElectronic ( multi--function ) function ) $400 to $600$400 to $600

Page 12: Pipet Calibration - Wisconsin DNR

12

AdvantagesAdvantages•• Highly precise and accurateHighly precise and accurate•• Adjustable volumes. Adjustable volumes. •• Disposable tips. No need for washing, Disposable tips. No need for washing,

saving analyst time. Eliminates saving analyst time. Eliminates possibility of carryover between possibility of carryover between samples.samples.

•• Possible laboratory savings. Despite Possible laboratory savings. Despite higher initial investment, cost may be higher initial investment, cost may be lower over time due to saved analyst lower over time due to saved analyst time and lower replacement rate than time and lower replacement rate than easily broken glass pipettes.easily broken glass pipettes.

Page 13: Pipet Calibration - Wisconsin DNR

13

DisadvantagesDisadvantages•• Higher initial costHigher initial cost

•• Failure to maintain and calibrate may lead Failure to maintain and calibrate may lead to gross errors and poor test results being to gross errors and poor test results being reported on discharge monitoring reports.reported on discharge monitoring reports.

•• Failures may be “unpredictable in timing Failures may be “unpredictable in timing and undetectable by the user.” Periodic and undetectable by the user.” Periodic calibration and preventive maintenance calibration and preventive maintenance are, therefore, essential to ensure the are, therefore, essential to ensure the integrity of laboratory results. integrity of laboratory results.

*George Rodrigues, PhD, ARTEL*George Rodrigues, PhD, ARTEL

Page 14: Pipet Calibration - Wisconsin DNR

14

Calibration ApproachesCalibration Approaches•• Gravimetric Gravimetric –– most commonmost common

•• All you need is a calibrated analytical All you need is a calibrated analytical balance capable of measuring down to 0.1 balance capable of measuring down to 0.1 mg (same as for total suspended solids)mg (same as for total suspended solids)

•• ThermometerThermometer•• Room temperature distilled waterRoom temperature distilled water•• A light weighing vessel for weighing A light weighing vessel for weighing

sample replicatessample replicates•• Clean, unused pipette tipsClean, unused pipette tips•• Spreadsheet or printed form for recording Spreadsheet or printed form for recording

and calculating pipette precision and and calculating pipette precision and accuracyaccuracy

Page 15: Pipet Calibration - Wisconsin DNR

15

Calibration ApproachesCalibration Approaches

•• Photometric Photometric –– not as commonnot as common•• Requires an expensive instrument Requires an expensive instrument

($5,000)($5,000)•• Expensive reagentsExpensive reagents

Page 16: Pipet Calibration - Wisconsin DNR

16

Annual Pipette Calibration (gravimetric)

Date Analyst BalanceTemp. (C) Z-Factor #N/A Serial #

Pipette # 117205 Pipette # Pipette #VOLUM E WEIGHT VOLUME WEIGHT VOLUME WEIGHT

(ml) (g) (ml) (g) (ml) (g)

MEAN #DIV/0! MEAN #DIV/0! MEAN #DIV/0!CORR. MEAN #DIV/0! CORR. MEAN #DIV/0! CORR. MEAN #DIV/0!STD DEV #DIV/0! STD DEV #DIV/0! STD DEV #DIV/0!% CV #DIV/0! % CV #DIV/0! % CV #DIV/0!% INACC. #DIV/0! % INACC. #DIV/0! % INACC. #DIV/0!PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0!

Pipette # Pipette # Pipette #VOLUM E WEIGHT VOLUME WEIGHT VOLUME WEIGHT

(ml) (g) (ml) (g) (ml) (g)

MEAN #DIV/0! MEAN #DIV/0! MEAN #DIV/0!CORR. MEAN #DIV/0! CORR. MEAN #DIV/0! CORR. MEAN #DIV/0!STD DEV #DIV/0! STD DEV #DIV/0! STD DEV #DIV/0!% CV #DIV/0! % CV #DIV/0! % CV #DIV/0!% INACC. #DIV/0! % INACC. #DIV/0! % INACC. #DIV/0!PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0!

Corrective Action Due To Failures?

Acceptance Criteria% Inaccuracy must be less than 2.00No replicate may be greater than 2% from true volume. %CV must be less than 1.00

Spreadsheet created by: Donna Johnsen, WSLH

Page 17: Pipet Calibration - Wisconsin DNR

17

Quarterly Pipette Calibration (gravimetric)

Date Analyst Balance Ae 240Temp. (C) Z-Factor #N/A Serial # J52053

Pipette # Pipette # Pipette #VOLUME WEIGHT VOLUME WEIGHT VOLUME WEIGHT

(ml) (g) (ml) (g) (ml) (g)

MEAN #DIV/0! MEAN #DIV/0! MEAN #DIV/0!CORR. MEAN #DIV/0! CORR. MEAN #DIV/0! CORR. MEAN #DIV/0!STD DEV #DIV/0! STD DEV #DIV/0! STD DEV #DIV/0!% CV #DIV/0! % CV #DIV/0! % CV #DIV/0!% INACC. #DIV/0! % INACC. #DIV/0! % INACC. #DIV/0!PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0!

Pipette # Pipette # Pipette #VOLUME WEIGHT VOLUME WEIGHT VOLUME WEIGHT

(ml) (g) (ml) (g) (ml) (g)

MEAN #DIV/0! MEAN #DIV/0! MEAN #DIV/0!CORR. MEAN #DIV/0! CORR. MEAN #DIV/0! CORR. MEAN #DIV/0!STD DEV #DIV/0! STD DEV #DIV/0! STD DEV #DIV/0!% CV #DIV/0! % CV #DIV/0! % CV #DIV/0!% INACC. #DIV/0! % INACC. #DIV/0! % INACC. #DIV/0!PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0! PASS / FAIL ? #DIV/0!

Corrective Action Due To Failures?

Acceptance Criteria% Inaccuracy must be less than 2.00No replicate may be greater than 2% from true volume. %CV must be less than 1.00

Spreadsheet created by: Donna Johnsen, WSLH

Page 18: Pipet Calibration - Wisconsin DNR

18

Temperature Z-Factor15 1.002

15.5 1.00216 1.0021

16.5 1.002217 1.0023

17.5 1.002418 1.0025

18.5 1.002619 1.0027

19.5 1.002820 1.0029

20.5 1.00321 1.0031

21.5 1.003222 1.0033

22.5 1.003423 1.0035

23.5 1.003624 1.0037

24.5 1.003825 1.0039

Page 19: Pipet Calibration - Wisconsin DNR

19

Date 10/25/07 Analyst DKP Balance Ae 240Temp. (C) 22 Z-Factor 1.0033 Serial # J52053

Page 20: Pipet Calibration - Wisconsin DNR

20

Pipette # 117205VOLUME WEIGHT

(ml) (g)2.50

MEAN #DIV/0!CORR. MEAN #DIV/0!STD DEV #DIV/0!% CV #DIV/0!% INACC. #DIV/0!PASS / FAIL ? #DIV/0!

Page 21: Pipet Calibration - Wisconsin DNR

21

Page 22: Pipet Calibration - Wisconsin DNR

22

Page 23: Pipet Calibration - Wisconsin DNR

23

Page 24: Pipet Calibration - Wisconsin DNR

24

Page 25: Pipet Calibration - Wisconsin DNR

25

Page 26: Pipet Calibration - Wisconsin DNR

26

Pipette # 117205VOLUME WEIGHT

(ml) (g)2.50 2.511

2.5082.5142.5172.5252.5142.5212.5032.4962.517

MEAN 2.51257CORR. MEAN 2.52086STD DEV 0.00858% CV 0.340383825% INACC. 0.83445924PASS / FAIL ? PASS

Page 27: Pipet Calibration - Wisconsin DNR

27

MEAN 2.51257CORR. MEAN 2.52086STD DEV 0.00858% CV 0.340383825% INACC. 0.83445924PASS / FAIL ? PASS

MEAN 2.51257CORR. MEAN 2.52086STD DEV 0.00858% CV 0.340383825% INACC. 0.83445924PASS / FAIL ? PASS

Acceptance Criteria% Inaccuracy must be less than 2.00

%CV must be less than 1.00No replicate may be greater than 2% from true volume.

Page 28: Pipet Calibration - Wisconsin DNR

28

CalculationsCalculations

Mean Weight = sample replicate 1 + sample replicate 2 + (etc.)Number of replicates

Mean Volume (Corrected Mean) = mean weight × Z-Factor

% Inaccuracy = [(Corrected Mean – true value) ÷ true value] × 100

% CV (Coefficient of Variation) = (Standard Deviation ÷ Corrected Mean) × 100

Page 29: Pipet Calibration - Wisconsin DNR

29

Temperature: 20° C - 25° C, constant to ± 0.5° C

# of measurements: 10

Courtesy of Eppendorf

≤ 0.2%± 0.6%100 µl - 200 µl

≤ 0.3%± 0.7%50 µl - 90 µl≤ 0.3%± 0.8%20 µl - 40 µl≤ 0.5%± 1.0%10 µl - 15 µl≤ 0.8%± 1.5%5 µl≤ 1.2%± 2.0%2 µl≤ 1.8%± 2.5%1 µl

ImprecisionInaccuracyVolume

Page 30: Pipet Calibration - Wisconsin DNR

30

A pipette is only as good as it’s A pipette is only as good as it’s operator.operator.

Courtesy: ARTEL

Page 31: Pipet Calibration - Wisconsin DNR

31

A little training can go a long way.A little training can go a long way.

Courtesy: ARTEL

Page 32: Pipet Calibration - Wisconsin DNR

32Courtesy: ARTEL

Page 33: Pipet Calibration - Wisconsin DNR

33

Pipette MaintenancePipette Maintenance•• Mechanical pipettes must be Mechanical pipettes must be

inspected, cleaned and inspected, cleaned and maintained regularly to assure maintained regularly to assure proper operation.proper operation.

•• Refer to the pipette’s instruction Refer to the pipette’s instruction manual for guidance on cleaning manual for guidance on cleaning and maintenance.and maintenance.

Page 34: Pipet Calibration - Wisconsin DNR

34

•• Care must be taken to not aspirate any Care must be taken to not aspirate any liquid into the shaft of the pipette. Even liquid into the shaft of the pipette. Even nonnon--corrosive liquids can affect the corrosive liquids can affect the performance of the pipette.performance of the pipette.

•• Do not lay the pipette down when a filled Do not lay the pipette down when a filled pipette tip is attached as liquid will enter pipette tip is attached as liquid will enter the shaft.the shaft.

•• If liquid accidentally enters the shaft, If liquid accidentally enters the shaft, disassemble the pipette and clean and disassemble the pipette and clean and dry the shaft and piston.dry the shaft and piston.

Page 35: Pipet Calibration - Wisconsin DNR

35

•• Inspect all seals and the piston for Inspect all seals and the piston for wear or damage. Replace if wear or damage. Replace if necessarynecessary

Page 36: Pipet Calibration - Wisconsin DNR

36

•• After cleaning (or replacement), the After cleaning (or replacement), the piston should be lubricated lightly.piston should be lubricated lightly.

Page 37: Pipet Calibration - Wisconsin DNR

37

In Summary……..• Air-displacement pipettes are precision

instruments that should not be taken for granted.

• Pipettes tend to fail silently and randomly, impacting sample and reagent delivery.

• Periodic calibration and preventive maintenance are, therefore, essential to ensure the integrity of laboratory results.

Page 38: Pipet Calibration - Wisconsin DNR

38

Summary continued

• Operator technique is also a significant source of pipetting error.

• Pipetting-technique training, especially if it offers the opportunity to obtain immediate feedback, is easy to do and has a significant positive effect on performance.

Page 39: Pipet Calibration - Wisconsin DNR

39

ReferencesArtel, Incorporated. Ten Tips to Improve Your Pipetting Technique. Artel, Incorporated, Westbrook, Maine, 1998.

ASTM International. Standard Specification for Piston-or Plunger-Operated Volumetric Apparatus. ASTM E1154. ASTM International, West Conshohoken, Pennsylvania, 1997.

Conners, M. and Curtis, R. H. Pipetting error: A real problem with a simple solution. Part I, American Laboratory News (1999) 31(13):16-19.

Curtis, R. H. Controlling pipette performance in the real world, Calibration Laboratory (2000) 7(1):32-36.

Page 40: Pipet Calibration - Wisconsin DNR

40

References- Continued

Curtis, R. H. Performance verification of manual action pipettes. Part 1, American Clinical Laboratory(1994) 12(7):8-10.

Curtis, R. H. Performance verification of manual action pipettes. Part 2, American Clinical Laboratory(1994) 12(9):16-17.

ISO. Piston-Operated Volumetric Apparatus: Piston Pipettes. Part 2, ISO 8655-2. ISO, Geneva, Switzerland, 2002.

Page 41: Pipet Calibration - Wisconsin DNR

41

References- Continued

NCCLS. Determining Performance of Volumetric Equipment: Proposed Guideline. NCCLS, Wayne, Pennsylvania, 1984.

Pentheny, G. Effects of common pipettingtechnique errors on accuracy and precision, American Clinical Laboratory (1997) 16(2):10-12.

Page 42: Pipet Calibration - Wisconsin DNR

42

•• The WSLH will have an The WSLH will have an instructional CD companion instructional CD companion to this presentation available to this presentation available soon. A sign up sheet will be soon. A sign up sheet will be available if you are interested available if you are interested in receiving a copy when it is in receiving a copy when it is available.available.

Page 43: Pipet Calibration - Wisconsin DNR

43

Thanks for having us!Thanks for having us!

DeWayneDeWayne KennedyKennedy--ParkerParker

Chemist Supervisor

WI State Lab of HygieneEnvironmental Health Division 2601 Agriculture DriveMadison, WI 53718

(608) 224-6282

[email protected]

Roger SchultzRoger Schultz

Senior Chemist

WI State Lab of HygieneEnvironmental Health Division 2601 Agriculture DriveMadison, WI 53718

(608) 224-6282

[email protected]

For More InformationFor More Information


Top Related