csRAAD Box™ Smart Sensor Network With the RAAD Box™, conventional strain gauge load cell systems can be transformed into advanced sensor networks. Instead of analyzing the combined signal from the load cells in the system, the RAAD Box embeds intelligence into each individual load cell, enabling you to analyze your system at the lowest level. Failures can be detected immediately, guard-ing you against costly downtime or product waste.
Power and Performance for Weighing Applications
IND110 Weight TransmitterThe IND110 provides an economi-cal way to connect an analog strain gauge scale or weigh mod-ules via 4-20mA to process control equipment.
IND131 Process TransmitterThis newest generation of process transmitters combines ultra-fast A/D conversion and digital filter-ing with a compact footprint and a wide choice of process interfaces. A crisp OLED display makes con-figuration very convenient. Also available in a junction box design for harsh and heavy washdown applications.
IND331 Process TerminalAll you get from an IND131, but additionally with a panel-mountable display. The design allows to panel mount the entire terminal as well as remote installation of the display, separate from the electronic. Also available in a desk/wall-mountable harsh housing.
Model IND110 IND131 / IND 131xx IND331 / IND331xxResolution (max.)Approved
4,000 divisions, (non-approved) 100,000 divisions, OIML Class III 6,000e,NTEP Class III/IIIL 10,000dMID approvals available
100,000 divisions, OIML Class III 6,000eNTEP Class III/IIIL 10,000d MID approvals available
Housing Plastic Plastic (DIN rail)Stainless steel enclosure (J-Box)
Stainless steel front (panel)Stainless steel enclosure (harsh)
Protection IP20, Type 1 IP20, Type 1 (DIN rail)IP69k (J-Box)
IP65, Type 4x and 12 (panel front); IP66, Harsh (not approved for Haz applications)
Interfaces 4 - 20mA 4-20 mA, AB RIO, DeviceNet, Profi bus DP, Ethernet IP, Modbus TCP, ControlNet and serial with support for Modbus RTU
4-20 mA, AB RIO, DeviceNet, Profibus DP, Ethernet IP, Modbus TCP, ControlNet and serial with support for Modbus RTU
Power 20 to 28VDC, 300 mA max. 85-264 VAC or 24 VDC; IND131xx available in 24VDC version only
85-264 VAC or 24 VDC; IND331xx avail-able in 24VDC version only
Applications Signal conversion, tank weighing Tank and simple process weighing, level indication and loss-in-weight
Tank and simple process weighing, level indication and loss-in-weight
Number of attachable scales
Single platform, up to 4 load cells or weigh modules
Single platform, up to 8 350 Ω load cells (AC version), up to 4 350 Ω load cells (DC version) or weigh modules
Single platform, up to 8 350 Ω load cells (AC version), up to 4 350 Ω load cells (DC version)
Page 191 193, 197 195, 197
191
Weig
hing
Elec
troni
cs
IND110Weight Transmitter
Customer benefits • Compact footprint uses minimal amount of valuable
enclosure space• DIN rail mounting for easy, fast installation• Setpoint outputs included• Remote sense inputs provide load cell cable
resistance compensation and allow for use with intrinsically safe barrier
• Digital filtering increases accuracy across entire weight range and in applications with vibration
• Simple and fast calibration, with zero calibration independent of span calibration value• Global approvals provide product acceptance in
local markets• Harsh and 7/9 enclosure optionally available
Direct Interface to PLCsand Other Control Equipment
ConnectEasy integration into process-based network architectures provides valuable plant-floor information that can be used to optimize production and help reduce down-time.
Complywith global regulations and standards. Harsh and 7/9 enclosure optionally available. 7/9 enclosure is certified safe for use in most Division 1 or 2 hazardous classified areas.
192
Parameter Specifi cationModel No. IND110Display none LED Indicators Power ON with external power supplied HI ON when load cell signal > high setpoint LO ON when load cell signal < low setpointTemperature Range Operating -10 ~ +45 °C (14 ~ +113°F)ATEX Approval (1) Number (7/9 enclos. only) DEMCO ATEX 0303070U (certifi cate for fl ameproof enclosure) Rating EExd IIB H2 IP66.Haz Approval (1) Number (7/9 enclos. only) UL number QUZW.E152336 Rating Class I, Div 1, Groups C-D Class II, Div 1, Groups E-G Class III Div 1 Remarks HazLoc approval only in conjunction with ISB15 or ISB15x barrier.Approvals UL, cUL, CE marked (UL fi le E139501 for module, E152336 for 7/9 ecnlosure) Load Cell Input Excitation Voltage 10 VDC, 120 mA max. (4x 350 Ohm cells) Load Cell Range 1.5 to 3 mV/V with full bridge Span Adjustment Range 25% to 110% of cell capacity @ 2 mV/V Zero Adjustment Range 0 to 100% of load cell output @ 2 mV/VOutputs Discrete 1 low & 1 high level output, open collector type, 5 to 25 VDC, 60 mA max sink current Analogue Signal: 4 to 20 mA (1 mA to 24 mA limits) Load resistance 0 to 500 Ohm maxLinearity +/- 0.012% maxTemperature Stabillity <20 ppm/°C (span), 50 ppm/°C (zero)Update Rate 20/second Calibration DIP switch mode selection enables front panel calibration of load cell zero/span and setpoint low/high limits Protection IP65 (optional Type 4x enclosure) Power Supply 20 to 28 VDC, 300 mA max., 8 watts Class 2 power supplyWeight 0.18 kg (0.39 lb)Mounting 35mm DIN rail (module only) or wall mount (optional enclosure)
1) See certifi cate for complete information
IND110 Weight Transmitter Specifi cations
IND110 Weight Transmitter Ordering Information
Description Item No.IND110 Load Cell Signal Converter DIN Rail 71208076Type 4X Enclosure for IND110 64053670IND110 Load Cell Signal Converter incl. 7/9 enclosure 64053672
www.mt.com/IND110
Weig
hing
Elec
troni
cs
193
Weig
hing
Elec
troni
cs
IND131 / IND331 Weighing Terminals
The IND131 and IND331 analog scale terminals deliver precision measurement data in a single, cost-effective package. Each terminal features:
• Ultra-fast A/D conversion rate• Patented TraxDSP® digital filtering• Organic LED display readable in all light conditions• Enclosures designed to match application environment• Modular design, multiple mounting possibilities• Support for a wide variety of options, including DIO,
additional serial port, and PLC interfaces• Choice of AC or 24 VDC power supply• Portable calibration/configuration settings and firmware
upgrades via an SD memory card• Scale calibration through PLC interface• CalFREE® allows scales to be tested without test weights
Max Connectivity, Mini Footprint Easy Integration for Weighing
The IND131/IND331 deliv-ers excellent accuracy, reliability and efficiency for process weighing applications.
Designed with the per-formance and reliability expected of Weights and Measures and product safety compliant terminals, these superior weighing terminals include features and functionality never before offered in such a compact and versatile form factor.
The inherent qualities of process weighing ter-mi nals from METTLER TOLEDO make it easier for OEMs and System Integrators to build best-in-class weighing systems cost-effectively.
Easy integration into pro-cess-based network archi-tectures provides valuable plant-floor information to process owners to opti-mize production and help reduce down-time.
194
IND131 / IND331Specifications
IND131/DIN IND331/Panel IND131/JBox IND331/Harsh
Enclosure typesUnit of
Measure DIN rail-mount, plastic, with operator interface
Panel mount, stainless steel
+ plastic chassis
Stainless steelenclosure including
internal boardto sum load cells
Stainless steel desk-or wall-mountable
enclosure, with operator interface
Dimensions* L × H × D mmin.
68 × 138 × 1112.7 × 5.4 × 4.4
Front panel:168 × 68 × 126.6 × 2.7 × 0.5
251 × 261 × 1239.9 × 10.3 × 4.8
220 × 131 × 1778.7 × 5.2 × 7.0
Shipping weight kg / lb 1 / 2.2 1.5 / 3.3 5.5 / 12.1 3 / 6.5
Environmental protection IP20, Type 1 IP65, Type 4x and 12 IP69k IP66
Ambient environmentOperation °C / °F -10 to 40 / 14 to 104, 10 % to 90 % relative humidity, non-condensing
Storage °C / °F -20 to 60 / -4 to 140, 10 % to 90 % relative humidity, non-condensing
Power requirementsAC version VAC / Hz / mA 85-264 / 49-61 / 27-73
DC version VDC / mA 18-36; nominal 24 / 84-170; nominal 120 (IND131), nominal 130 (IND331)
DisplayType Green OLED including weight display, weight units, gross/net indication and graphic symbols for motion and
center of zero. 10 updates/sec.
Character height mm / in. 5.6 / 0.22 12 / 0.47 5.6 / 0.22 (internal) 12 / 0.47
Weight display Maximum displayed resolution of 100 ,000 divisions
Scale type Analog load cells
Number of cells Up to 8 350 Ω load cells (AC version), up to 4 350 Ω load cells (DC version), 2 or 3 mV/V
Number of scales 1
Analog/Digital update rates Hz Internal analog: 366 / Target comparison: 50 / PLC Interface: 20
Digital Filtering TraxDSP®
Memory Stores two-speed target control values and limit values for three comparators
Applications Basic process weighing for gain-in-weight (filling), loss-in-weight (dosing) and level indication
Load cell excitation voltage VDC 5
μV Build Minimum/Approved microvolts 0.1 / 0.6
Keypad 4 keys (Clear, Zero, Tare, Print); 1.22 mm thick polyester overlay (PET) with polycarbonate display lens
Communications
Serial inter-faces
Standard: One serial port (COM1), RS-232, 300 to 115,200 baudOptional: Serial port (COM2), RS-232/485, 300 to 115,200 baud
Protocol Serial inputs: ASCII commands for CTPZ (Clear, Tare, Print, Zero), SICS (most commands, levels 0 and 1)Serial outputs: Continuous, Extended continuous, or Demand (limited formats)
Approvals
Weights andMeasures
USA: NTEP Class III/IIIL - 10,000d; CoC 09-051Canada: Class III/IIIHD - n max. 10000/20000;
AM-5744Europe: OIML, Class III, 6000e; R76/2006-NL1-09.26
MID OIML R51 (Automatic Catchweighing) T10262; OIML R61 (Automatic Gravimetric Weighing) T10261
Product Safety UL, cUL, CE
* Dimensional drawings are included on following pages
Options and AccessoriesMultiple PLC Options
COM2 / DIO, includes:
– COM2 RS-232/485 serial, 4 discrete outputs, solid state or relay
– Modbus RTU capability, 2 discrete inputs (selectable active or passive)
72229290 Panther Terminal adapter plate for IND331 Panel unit
71209353 Wall mounting bracket (for IND331 Harsh unit)
22015188 Swivel mounting bracket (for IND331 Harsh unit)
72229289 2GB SD Memory Card for configuration and calibration data backup and restore
All IND131 and IND331 terminals can be equipped with any of a wide array of PLC options, including:
• 4-20mA Analog output
• Allen-Bradley RIO
• DeviceNet™
• EtherNet/IP™
• Modbus TCP
• PROFIBUS® DP
• ControlNet™ (24 VDC units only)
• Modbus RTU
www.mt.com/IND131-331
195
Weig
hing
Elec
troni
cs
IND131xx and IND331xxProcess Weighing in Hazardous Areas
IND131xx / IND331xxWeighing Terminals
The IND131xx and IND331xx analog scale terminals deliver precision measurement data in a single, cost-effective package, safe for use in Zone 2/22 and Division 2/Zone 2 hazardous areas. Terminal features include:
• Organic LED display readable in all light conditions• Ultra-fast A/D conversion rate and TraxDSP® digital
filtering• Enclosures designed to match application
environment − modular design enables multiple mounting possibilities
• 24 VDC operation• Support for a wide variety of options, including DIO,
additional serial port, and PLC interfaces• Portable calibration/configuration settings and
firmware update via an SD memory card• Scale calibration through PLC interface. CalFREE®
allows scales to be calibrated without test weights
Zone 2/22, Division 2
With the performance and reliability expected of Weights and Measures and product safety compliant terminals, the inherent properties of these process weighing terminals from METTLER TOLEDO make it easier for OEMs and System Integrators to build best-in-class weighing systems cost-effectively.
The IND131/IND331 deliv-ers excellent diagnostics, reliability and efficiency for process weighing applications. These supe-rior weighing terminals include features and functionality never before offered in such a compact and versatile form factor.
Easy integration into pro-cess-based network archi-tectures provides valuable plant-floor information that can be used to opti-mize production and help reduce down-time.
IECEx
196
IND131xx / IND331xxSpecifications
IND131xxDIN mounted
IND331xxPanel mounted
IND131xxJBox enclosure
Enclosure types Unit ofMeasure
DIN rail-mount, plastic, with operator interface
Panel mount, stainless steel with operator interface + plastic chassis
Stainless steelenclosure including
internal boardto sum load cells
Dimensions L × H × D mmin.
68 × 138 × 1112.7 × 5.4 × 4.4
Front panel:168 × 68 × 126.6 × 2.7 × 0.5
251 × 261 × 1239.9 × 10.3 × 4.8
Shipping weight kg / lb 1 / 2.2 1.5 / 3.3 5.5 / 12.1
Environmental protection IP20, Type 1 IP65, Type 4x and 12 IP69k
Ambient environmentOperation °C / °F -10 to 40 / 14 to 104, 10 % to 90 % relative humidity, non-condensing
Storage °C / °F -20 to 60 / -4 to 140, 10 % to 90 % relative humidity, non-condensing
Power requirements VDC / mA 18-36; nominal 24 / 84-170; nominal 120 (IND131xx), nominal 130 (IND331xx)
DisplayType Green OLED including weight display, weight units, gross/net indication and graphic symbols for motion and center
of zero. 10 updates/sec.
Character height mm / in. 5.6 / 0.22 12 / 0.47 5.6 / 0.22 (internal)
Weight display Maximum displayed resolution of 100 ,000 divisions
Scale type Analog load cells
Number of cells Up to 4 350 Ω load cells, 2 or 3 mV/V
Number of scales 1
Analog/Digital update rates Hz Internal analog: 366 / Target comparison: 50 / PLC Interface: 20
Digital Filtering TraxDSP®
Memory Stores two-speed target control values and limit values for three comparators
Application Basic process weighing for gain-in-weight (filling), loss-in-weight (dosing) and level indication
Load cell excitation voltage VDC 5
μV Build Minimum/Approved microvolts 0.1 / 0.6
Keypad 4 keys (Clear, Zero, Tare, Print); 1.22 mm thick polyester overlay (PET) with polycarbonate display lens
Communications
Serialinterfaces
Standard: One serial port (COM1), RS-232, 300 to 115,200 baudOptional: Serial port (COM2), RS-232/485, 300 to 115,200 baud
ProtocolSerial inputs: ASCII commands for CTPZ (Clear, Tare, Print, Zero),
SICS (most commands, levels 0 and 1)Serial outputs: Continuous, Extended continuous, or Demand (limited formats)
Approvals
Weights andMeasures
USA: NTEP Class III/IIIL - 10,000d; CoC 09-051Canada: Class III/IIIHD - n max. 10,000/20,000; AM-5744
Europe: OIML, Class III, 6000e; R76/2006-NL1-09.26
Hazardous locations
MID OIML R51 (Automatic Catchweighing) T10262OIML R61 (Automatic Gravimetric Weighing) T10261
Product safety, ordinarylocations
UL, cUL, CE
USA - FM Canada - FM ATEX IECEx NEPSI
NI / I, II, III / 2 / ABCDEFG / T5 Ta = 40°C
I / 2 / IIC / T5 Ta = 40°C
NI / I / 2 / ABCD / T5 Ta = 40°C
DIP / II, III / 2 / EFG / T5 Ta = 40°C
II 3 G Ex nA nL [nL] IIC T5II 3 D Ex tD A22 IP 65 T
100°C
Ex nA nL [nL] IIC T5Ex tD A22 IP65 T100°C
Pending
Install per Control Drawing 72238303R Install per Installation Drawing 72246295R
* Dimensional drawings are included on following pages
197
Weig
hing
Elec
troni
cs
IND131 and IND331 Packaged SolutionsEnhanced Terminals in Robust Enclosures
• IND131 and IND331 in enclosure with optional I/O• Each version can be configured with any of the terminals’
multiple interface and connectivity • Optional DC power supply available
Enclosure SpecificationsEnclosures are also available in painted mild steel (shown at left), stainless steel and fiberglass. The stainless steel enclo-sure housing the IND131 DIN and IND331 panel packaged solutions is rated IP66 (IP65, when used with IND331 panel mount terminal). Each type includes integrated mounting back-ets, and if required can be configured with a power supply.
Multiple Terminal Configurations AvailableWhen a single terminal is not enough, multi-terminal designs are available, with power supply, optional I/O, and additional features as required for the application.
Faster basic process controlThe IND131/IND331 combination serial and relay Discrete I/O option allows the terminal or terminals to be configured for efficient process control. User selection of latched or concidence targets in combination with user-configured limit values for up to three comparators create a time-saving and cost-effective process controller.
Customized InterfacesCustomize the packaged IND131 and IND331 for specific uses with convenient operators for I/O, such as annunciators and pushbuttons, making operator interactions even easier, more consistent and safer.
198
IND131xx and IND331xx Packaged Solutions
• Packaged IND131xx and IND331xx in Enclosures Approved for Division 2, Zone 2 and Zone 22
• Each version can be configured with any of the terminals' multiple interface and connectivity options
• Packages include an agency-approved 24 VDC power supply
Enclosure SpecificationsThe stainless steel enclosure housing the IND131xx DIN and IND331xx panel packaged solutions is FM-approved for use in safe areas, and in Division 2, Zone 2 hazardous areas. The enclosure is ATEX/IECEx certified for Zone 22 (dust) hazardous areas. Cable entry points can be either conduit hubs for use in the U.S., or ATEX-approved glands for European applications.• FM-certified enclosure, gas and dust • ATEX and IECEx certified for Zone 2/22 environments• Rated IP66, IP65 when used with panel-mount terminal• T5 temperature class rating
Multi-Terminal ConfigurationsWhen a single terminal is not enough, multi-terminal designs are available, with agency-approved power supply, optional ISB (Intrinsically Safe Barrier) and additional features as required for the application.
Load Cells in Hazardous AreasIn applications with load cells in-stalled in Division 1/Zone 1 or Zone 1/21 areas, an intrinsically safe ana-log load cell barrier (ISB) is required. IND131xx and IND331xx terminals packaged with a METTLER TOLEDO ISB05 provide an integrated solution for such applications.
Packaged Model Configurations
IND131xx and IND331xx packages are highly customizable. Figure 1 shows a system compris-ing an IND131xx with optional ISB05 barrier. Figure 2 shows an IND331xx panel-mount terminal (its front panel interface not shown). Packaged model configurations include an FM/CSA/ATEX approved power supply (24VDC at 1.3 A, 100-240 VAC, 50/60 Hz) and are available with any of the PLC inter-faces (EtherNet/IP, PROFIBUS DP, Modbus TCP, Allen-Bradley RIO, ControlNet, Device-Net), and Solid State DIO.
Figure 1 Figure 2
IND131xx-based• ISB05 (optional), ATEX glands
(for European installations)• FM, ATEX and IECEx- certified
enclosure
IND331xx-based, standard(front panel display not shown)• Conduit hubs (for installation in
US/Canada)• FM, ATEX and IECEx-certified
enclosure
IECEx
Weig
hing
Elec
troni
cs
199
IND331, Harsh enclosure
IND131/IND131xx, DIN-mount enclosure
138 [5.43]
130 [5.12]
ø2.5 [R0.1]
68 [
2.68
]
100
[3.9
6]
110.
5 [4
.35]
220 [8.66]
130.
5 [5
.14]
176.9 [6.96]
145.8 [5.74]
120 [4.72]
95 [
3.74
]
M4 [0.16]
IND131/IND131xx/IND331/IND331xxDimensional Drawings in mm and [inches]
IND331/331xx, Panel-mount
Template for IND331 panel mounting cut-outs
148.2 [5.3]
74.1 [2.92]35 [1.38]
9.9 [0.39]
40 [1.57]
14 [0.55]
6 - ø4.5 [0.17]
3 [0.12] ø30 [1.18]46 [1
.81]
31 [1
.22]
156 [6.14]130 [5.12]
110.
5 [4
.35]
11 [0.43]
40 [1
.57]
00.5
[3.
96]
115.
2 [4
.96]
168 [6.61]
R4 [0.16]
68 [
2.68
]
60 [
2.38
]
68 [
2.68
]
Panel
200
249.
5 [9
.82]
IND131/IND131xx, J-box enclosure
184 [7.24]
200 [7.87]237 [9.33]
164.
1 [6
.46]
144.
8 [5
.70]
125.
5 [4
.94]
223.5 [8.80]250.8 [9.87]
237 [9.33]200 [7.87]
168.
7 [6
.64]
165.
7 [6
.52]
151.
5 [5
.97]
Ø 12.7 [0.5]
R3.6 [0.14]
IND131/IND131xx /IND331/IND331xxModel Identification
www.mt.com/IND131-331