February 2016The 46th Annual Meeting of the Japanese Society for Replacement Arthroplasty
SpecialReport
No.81 (2017.3)
1. Introduction
T-smart, being developed as the tomosynthesis providing tomographic images with few metal artifacts, is considered useful for the postoperative evaluation of total knee arthroplasty (TKA) due to its capacity for detecting minute changes in periprosthetic bone. We compared radiography and tomosynthesis images obtained after TKA, and investigated the usefulness of tomosynthesis for the early stage biological fixation.
2. Subjects and Methods
Subjects included 47 knees among 30 patients (bilateral simultaneous TKA in 17 patients and unilateral TKA in 13 patients) who underwent cemented poster ior-stabi l ized TKA wi th the Persona (Zimmer Biomet) for the medial knee osteoarthritis between April 2014 and March 2015. The mean age at time of surgery was 77.7 years (66 to 86 years).
(1) Radiography images and tomosynthesis images were obtained 6 months after surgery, and the rate of emergence of periprosthetic radiolucent lines (RLLs) were determined. The emergence of a gap of 0.5 mm or more was defined as positive.
(2) The proportion of appropriate and evaluable frontal radiography of the tibia and lateral radiography of the femur was determined.
The segmented zones of each radiography are shown in Fig. 1. The X-ray system used was a Shimadzu SONIALVISION series R/F system.
The Detection of Radiolucent Line in Total Knee Arthroplasty Using of TomosynthesisDepartment of Orthopaedic Surgery, Faculty of Medicine, Saga UniversitySatomi Nagamine, Shuya Ide, Tomonori Tajima, Shinsuke Someya, Motoki Sonohata, and Masaaki Mawatari
Fig.2
Detection Rate for Gap Between Femoral Component and Bone (n=47) Detection Rate for Gap Between Tibial Component and Bone (n=47); Frontal View
Detection Rate for Gap Between Tibial Component and Bone (n=47); Lateral View
0
5
10
15
20
25
30
Zone 1 Zone 4
7 8
14
22
14.9% 17.0% 29.8%
46.8%
0
5
10
15
20
25
30
Zone 1 Zone 4, medial Zone 4, lateral
7
0 0
14
5
19
14.9% 10.6%
40.4%
0
5
10
15
20
25
30
Zone 1 Zone 2 Zone 3 Zone 4 Zone 5
15
1
6
0
6
15
5
18
3
9 31.9% 31.9%
10.6% 2.1%
38.3%
12.8% 6.4%
19.1% 12.8%
0
5
10
15
20
25
30
Zone 1 Zone 2 Zone 3 Zone 4 Zone 5
5 7
10
1 4
17 19
26
7 4
10.6%
36.2% 40.4%
14.9%
55.3%
21.3%
2.1% 8.5% 8.5%
14.9%
1
2 3
4
Tomosynthesis
Radiography
Medial posterior condyle
Lateral posterior condyle
Tomosynthesis
Radiography
Tomosynthesis
Radiography
29.8%
Fig.1
Lateral radiograph of the femur
Frontal radiograph of the tibia
Lateral radiograph of the tibia
1
2 3
4 1
2
3
4
5 1
2
3
4
5
This article describes the poster presentation made about Shimadzu tomosynthesis.
February 2016The 46th Annual Meeting of the Japanese Society for Replacement Arthroplasty
SpecialReport
No.81 (2017.3)
3. Results
A result of investigation (1) is shown in Fig. 2. The red numbers indicate radiolucent zones and the red circles show areas of note.As a result of investigation (2), the appropriate and evaluable frontal images of the tibial component, and the lateral images of the femoral component were obtained by radiography in 57.4 %, and 46.8 % respectively (Fig. 3).
4. Discussion
RLLs are evaluated accurately by fluoroscopy guided radiography 1 week after TKA at our hospital. The radiological technologists are instructed that an X-ray irradiation angle should be accurately parallel to the component/bone boundary plane. Frontal radiographs require accurate alignment with the boundary plane between the tibial component
and bone, and lateral radiographs require accurate alignment with the boundary plane between the femoral component and bone, though there are limitations in conventional radiography. While a variety of refinements are implemented when performing tomosynthesis at our hospital, the two practices shown in Fig. 5 are considered particularly important.
5. Conclusions
• Tomosynthesis was able to visualize periprosthetic bone accurately that was difficult to evaluate with radiography. Furthermore, with T-smart processing that reduces metal artifacts, tomosynthesis could be used for early detection of periprosthetic minute changes of bone.
• Lateral tomosynthesis could be used to evaluate the medial and lateral posterior condyle of the femur separately.
• Metal artifacts in tomosynthesis images altered based on factors such as X-ray parameters, materials of prostheses, patient postures, and scanning directions. And tomosynthesis required refinement of the imaging methods and imaging condition settings.
• The scanning along the transverse axis were effective for both the periprosthetic tibia and femur tomosynthesis.
The Detection of Radiolucent Line in Total Knee Arthroplasty Using of Tomosynthesis
Fig.4
【Characteristics of Tomosynthesis】
【Issues of Tomosynthesis】
All the above phenomena are evaluated as RLL positive.
• Tomosynthesis images are evaluable and accurate since a large number of cross-sectional images parallel to the R/F table top are obtained in a single acquisition.
• Changes at the periprosthetic bone can be observed in the earlier stage by tomosynthesis because the effects of metal artifacts are minimal.
• Although tomosynthesis reduces metal artifacts, they are not prevented entirely, and image quality will alter depending on such as imaging conditions.
• The RLL detection rate with tomosynthesis is higher than the actual RLL emergence rate.
Radiolucent lineReactive lineInitial gap (anteromedial gap in grand-piano sign, initial gap of patella groove, etc.)Artifact gap (alters depending on X-ray parameters, materials of prostheses, patient postures, scanning directions, and component designs)
✔
✔
✔
✔
Fig.5
Accurately measured values should be entered for the height of the central imaging position and the thickness of the subject.
Practice 1: Accurate Measurements
Practice 2: Scanning Along Transverse Axis
Frontal view Lateral view Frontal view Lateral view Scanning along transverse axis
Metal artifacts are minimized when the scanning is performed along the transverse axis compared to the longitudinal axis.
Scanning along longitudinal axis
Fig.3
Frontal radiograph of tibial component
inappropriate and unevaluable appropriate and evaluable
An X-ray irradiation angle should be parallel to the boundary surface between the implant and bone.
Lateral radiograph of femoral component
inappropriate and unevaluable appropriate and evaluableThe medial and lateral pegs are not superimposed.
The medial and lateral posterior condyles are not superimposed.