![Page 1: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/1.jpg)
Validation of new systems in an ISO 9001 certified environment
Emma Davis CIGMR
University of Manchester
![Page 2: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/2.jpg)
Overview
• What is BioBanking Solutions
• What is ISO 9001
• Validation of Chemagic STAR
![Page 3: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/3.jpg)
What is Biobanking Solutions?
• BioBanking Solutions (BBS) is:o a research project based at Centre for Integrated Genomic Medical
Research (CIGMR), UoMo a 15 strong team of academics, project managers, IT specialists and
technical staffo an ISO 9001:2008 certified biobanking solution provider for projects
with sample handling requirementso A provider of services and advice for banking and distributing
biological samples.
http://www.biobankingsolutions.ac.uk/
![Page 4: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/4.jpg)
We provide
• Project management • Labelling solutions• Sample tracking for multiple sample types• DNA extraction (manual and robotic)• DNA measurement (Nanodrop and PicoGreen)• Sample storage (room temperature, -80C)• Sample distribution • Sample replenishment
![Page 5: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/5.jpg)
Biobanking process
Biobank•Study Design•Storage•Processing•Aggregation•Quality control
Collectors Collaborators
![Page 6: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/6.jpg)
What is ISO 9001
• Current version is ISO 9001:2008• Quality management system standard• Based upon mapped processes• Requires a customer focus, management
commitment and continual improvement of your systems
• We have been ISO 9001 certified since 2005• We are audited by an external auditor
![Page 7: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/7.jpg)
Mapped process
• In order to ensure that all of our services are run in a controlled and standardised way we have:– Mapped core and support processes– Validated core machinery– Trained to SOPs– Developed a Quality Management System (QMS)
which is certified to ISO 9001:2008
![Page 8: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/8.jpg)
What is a Quality Management System (QMS)?
• The web definition is:“A quality management system is a set of documented, interrelated or interacting elements that organizations use to direct and control how quality policies are implemented and quality objectives are achieved.”
![Page 9: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/9.jpg)
What does this mean?
• Meeting the quality expectations of upstream and downstream users through monitoring of your procedures and continual improvement.
OR
• Doing all aspects of your work in the best way you can.
![Page 10: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/10.jpg)
Mapped BBS processes
![Page 11: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/11.jpg)
Example of a mapped process
• Our processes detail basic steps• Supported by detailed flowcharts and SOPs• Controlled and reviewed by senior staff
![Page 12: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/12.jpg)
Introducing a new system into the Biobank
• We implement a 2 stage approach when introducing a new system or procedure into the biobank:
– Method development
– Validation
• These of both fall under the R&D process
![Page 13: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/13.jpg)
Method development in the R&D process
• This section is performed by a senior member of laboratory staff :– Scoping the system– Initial tests of the system– Changes are made to the system to fulfil
requirements– Drafting SOP– Develop validation test plan– Hand-over to validation staff
![Page 14: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/14.jpg)
Method Validation
• Validation is a procedure that is used to check that a product, service or system meets requirements and fulfils its intended purpose.
• Validation is required when working to ISO 9001 and is important when you are providing a service. (section 7.3 – design and development)
![Page 15: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/15.jpg)
Method Validation in the R&D process
• Performed by specifically trained individuals– Thorough testing of every function of the system– Completion of the test plan– Additional testing where necessary– Assessment of the SOP– Activation of the system– Probationary period
![Page 16: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/16.jpg)
Example section of a test planRestart the Method. Select the matrix to matrix files for both user prompts.
User prompt: Please choose the quantification technique used. Absorbance. Fluorescence. Click Abort.
Method Aborts. Restart the Method. Select the matrix to matrix files for both user prompts for the worklists.
User prompt: Please choose the quantification technique used. Absorbance. Fluorescence. Select Absorbance, Click OK.
The ‘Destination plate type’ prompt appears. Click Abort.
Method Aborts.
![Page 17: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/17.jpg)
Why we validate
• Ensure traceability and sample identification• Maintain the integrity of the samples• Provide accurate data• Prevent loss of sample• Prove the system does what is supposed to do• Peace of mind for collectors and collaborators
![Page 18: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/18.jpg)
Frequency of validation
• Core systems should be validated after any change to the system
• The level of validation can range from:– A single test run after minor changes– A full system check after upgrades
![Page 20: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/20.jpg)
Chemagic overview
• The Chemagic STAR, with the 24 rod head, can run 2 sets of 24 samples in one run
• Each set takes approximately 1.5hrs to run• It can process up to 4mls of blood or saliva• It uses magnetic bead extraction• It is integrated with our LIMS and includes
barcode reading
![Page 21: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/21.jpg)
Chemagic – critical procedure
• DNA extraction is a service BBS provides• High quality DNA extraction is important for
our collectors• Failure of DNA extraction results in loss of the
sample• Repeat samples are not always available• Poor quality DNA extraction may cause DNA
to degrade over time
![Page 22: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/22.jpg)
Plan for validation of the Chemagic 24-rod head
Run various numbers of samples in simulation for both blood and saliva◦E.g. <24, 24, 24+1, 2x24, 2x24-1 (testing the
programming)Run the same tests with water
◦Testing the robotics, labware definitions and liquid handling
Run at least 24 saliva kits and bloods◦Samples should ideally be from one donor for blood
and a mixture of donors for salivaCheck the user interface for errors and check
volume calculations
![Page 23: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/23.jpg)
Plan for validation continued
Validate connection to LIMS and transfer of data
Validate chemistry◦Assess DNA quality and quantify by Nanodrop,
picogreen, Gel, rtPCR/PCRUse extracted DNA to assess stabilityRepeat process for 12 rod head.
![Page 24: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/24.jpg)
Example of simulation test data
• Purpose of the test was to see which methods could be run with different combinations of labware
protocol Medium VolumeSample Type Kit
Output 1ml tips 5ml tips 1ml tips 5ml tips 1ml tips 5ml tips 1ml tips 5ml tips 1ml tips 5ml tipsBlood Kit 1 (medium) Abgene x x x No x x x No x x
Greinerbioone x x x No x x x No x xMatrix x x x x x x x No x x
Blood Kit 2 (medium) Abgene x x x No x x x No x xGreinerbioone x x x No x x x No x xMatrix x x x x x x x No x x
Blood Kit 3 (medium) Abgene x x x No x x x No x xGreinerbioone x x x No x x x No x xMatrix x x x x x x x No x x
Saliva Kit 1(medium) Abgene x x x No x x x No x xGreinerbioone x x x No x x x No x xMatrix x x x x x x x No x x
Input LabwareSaliva TekLab 1ml Teklab 5ml Vacuette 1ml Vacuette 5ml
x = compatibleChecked in simulation mode
![Page 25: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/25.jpg)
Looking for patterns154.3 171.365 8.32 180.81
184.335 227.03 165.655 26.39
202.875 15.22 231.64 174.465
229.705 197.31 230.495 10.825
214.055 217.72 221.76 227.335
167.15 215.78 170.52 9.61
Data from before head was changed(16/1)average conc ng/ul 199.1
Average yeild per ml (ug) 24.8Average 260/280 1.932368421Average 260/230 2.008421053Highest value cell Cell 9 231.64Lowest value cell Cell 7 8.32Standard deviation 80.39
164.77 299.435 213.17 191.1
226.98 245.18 227.805 225.21
242.665 243.26 238.915 225.805
231.875 234.605 241.675 223.51
254.595 234.91 233.885 228.735
204.18 244.72 237.545 225.14
Date from after head was changed (09/3)average conc ng/ul 230
Average yeild per ml (ug) 28.8Average 260/280 1.917291667Average 260/230 2.0575Highest value cell Cell 13 299.435Lowest value cell Cell 19 164.77Standard deviation 24.11015
![Page 26: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/26.jpg)
Post validation
• Generate a report detailing all issues and fixes• SOPs undergo a probationary process• Samples are being run routinely• Data is being collected to monitor the
extraction technique
![Page 27: Validation of new systems in an ISO 9001 certified environment Emma Davis CIGMR University of Manchester](https://reader035.vdocuments.net/reader035/viewer/2022062511/5517e571550346d0568b46cc/html5/thumbnails/27.jpg)
Thanks Bill Ollier Martin Yuille Kate Dixon Andrew Platt Debbie Payne Biobanking team
All CIGMR Staff
Collectors & Collaborators
Hamilton and Chemagen Support