_~.;;:,A~ t"v$'~.'OPERATING MANUAL
Minute Volumeter 3000
From Dräger:Minute Volumeter 3000OPERATING INSTRUCTIONS
Important NoticeFor correct and effective use of the appliance, and to avoid hazards, we wouldpoint out the following:
Any use of the appliance requiresprecise knowledge and observationof these operating instructions.
2 The appliance is intended only for thepurposes specified in the OperatingManual or for purposes confirmed inwriting by Drägerwerk AG.
3 The appliance should be inspected byexperts at regular time intervals. Anofficial report of the inspectionsshould be drawn up.
4 Only original Dräger spare partsr' ,uld be used for maintenance and'_,Jairs. Repairs and maintenance,and the replacement of spare partsshould only be carried out by experts.
5 We recommend having inspectionsand repair work carried out by the
2
Technical Customer Service of yourDräger Branch or Agent.Regular inspection is best ensured byentering into an Inspection ServiceContract with the Technical Customer Service ofyour Dräger Branch orAgent.
6 Responsibility for the reliable function of the appliance passes to theowner or operator in all cases wherethe appliance has been inexpertlymaintained or repaired by personsnot employed by the Dräger Organisation or where it has been used in amanner which does not conform tothe normal conditions of use.
We would also point out that the nationar recommendations, regulations andlaws governing the use of technicalequipment should be observed.
DRÄGERWERK AG LÜBECK
ContentsPage
Important Notice 2
Intended Use 3
Approvals 3
Technical Data 3
Initial Preparation 4
Operational Use 4
Cleaning and Disinfection 6
Maintenance and Care 7
Trouble Shooting 7
Order List.. 8
Intended UseThe Dräger minute Volumeter 3000shows the patient's respiratory minutevolume and tidal volume independentot the gas composition, and it therebyappreciably simplities the determination ot respiratory values needed tortherecord.
Application ot Volumeter 3000• In assisted or controlled ventilation;• During anaesthesia;
• In the determination ot vita I capacity;• To check spontaneous respiration
(e. g. during postoperative recovery).
No heating is required tor the Volumeter 3000.
It is resistant to corrosion in respect tothe usual inhalation anaesthetics (e. g.Halothane, Entlurane, or Methoxytlurane).
Technical Data
ApprovalsThe Volumeter 3000 is exclusivelymechanical in operation and it is theretore explosion proot.
Fig.1 Minute Volumeter 300036270
Measurement range1 turn ot large hand1 turn ot small hand
Necessary initial ti ow
Max. permissible peak tlow
Flow resistance
Heat stability
Connections
Air inlet (above)Air outlet (below)
Weight (without heating)
Dimensions
Diameter/Length
1000 cm315 I
3.5 I/min
120 I/min
1 mbar at 80 I/min
up to 120°C
Male thread M 33x 1Union nut M 33x 1
0.9 kg
80 0/90 mm
3
Initial PreparationThe Volumeter is connected to the expiratory side of a ventilator or to a circlesystem of an anaesthetic machine. Theexpiratory air flows from top to bottom.
The Volumeter can be directly screwedonto the Dräger ventilators- andanaesthetic machines (Fig. 2).
Operational UseMeasurement of Minute Volume
When the left button is depressed (Fig.3), a timer begins to run, which after oneminute wil! stop the hand ofthe volumeter. In this manner, the minute volume isdetermined. When the left button ispushed, a black spot wil! also sim ultaneously appear in the left aperture,and after one minute, a black disc wil! fil!the rectangular space, located belowthe hand of the indicator.
Measurement of Tidal Volume
When the right button is depressed (Fig.4) a black spot wil! appear in the rightaperture and the tidal volume wil! bemeasured.
To stop the indicator hand after acertainnumber of respiratory cycles, the rightbutton is depressed half way. To restartthe Volumeter, the same button iscompletely depressed.
38271 38272
Fig.2 Minute Volumeter 3000 in circle Fig. 3 Release of indicator hand for minutesystem of anaesthetic machine volume
4
38213
Fig. 4 Release of indicator hand for tidalvolume
Measurement of Vital Capacity
(Special equipment 2M 18476 for theminute Volumeter 3000 in a kit with accessories).
The accessory kit (Fig. 5) enables measurement of vital capacity using a facemask (Fig. 6) or a moufh piece (Fig. 7).
During postoperative recovery, the minute volume of a patient with an indwelling endotracheal tube can be determined using the same special equipment without further expenditure.
38625
Fig. 6 Measuring the vita I capacity using aface piece
38626
Fig. 7 Measuring the vital capacity using amouth piece
38621
Fig.5 Dräger Minute Volumeter 3000 with special equipment kit
1 Minute Volumeter 30002 Face mask size 1 (large)3 Face mask size 2 (smalI)4 Head st rap5 Valve6 Mouth piece7 Nose clamp8 Corrugated hose with
connecting nozzleconsisting of:GasketThreaded connectorCorrugated hoseUnion collarConnecting nozzle
9 Hose for endotracheal tubes10 Bag with 5 cardboard
mouth pieces11 Cone shaped adapter
2 M 18250M 8983M 9126M 8135
2 M 21032 M 2353
R 5868
2 M 2346
M 92572 M 2328
M 12558M 9226
2 M 2091M 10071
2 M 23558400851
5
Cleaning and DisinfectionCleaning
The Volumeter is to be cleaned aftereach day's use. Run hot water from thetap through the Volumeter measurement chamber, the water flowing framtop to bottom, so that none of the watercan enter the bleed vents of the contralmechanism (Fig. 8). Do not immerse theVolumeter in water.
Shake out the Volumeter manually (Fig.9) and then dry it using the DrägerVolumeter Dryer 2M 14850 (Fig. 10) or
the Dräger Dryer 2M 8220. The connecting nozzle 2M 13921 is to be used to attach the Volumeterto the 2M 8220Dryer.
Disinfection in the Dräger Aseptor
Clean and dry the Volumeter as described before. Then attach it to theholding screw in the Aseptor and it willbe disinfected with formaldehyde.
Autoclave Sterilization
Clean the Volumeter as described be-
fore. It does not need to be dried.Sterilize in an autoclave set to a glovepragramme at 120°C.
..Q
38628
Fig. 8 Flushing out Volumeter
6
38629
Fig. 9 Shaking out remaining water38274
Fig.10 Dräger Volumeter-Dryer 2 M 14850
Maintenance and CareThe bearings of the Volumeter's rotatingelements become accessible through 4lateral bores; these bores are marked byprinted arrows, This makes it possible tolubricate the bearings with a speciallysuited lubricant »Oxigënoex S 4« (approved for fittings carrying oxygen).A lubricating set 2M 18180 is suppliedwith each Volumeter.
Lubrication must be carried out afterabout 30 steam sterilisations or at least
every 2 months (Fig. 11).Note!
Only use pipette supplied with the kit toavoid damage to the bearings.
Each bearing must be lubricated as follows:
• Remove protective cover frompipette and open lubricant bottle.
• Dip tip of pipette into lubricant andsuck one drop of »Oxigenoex S 4«into pipette by pulling the pin upwards.
• Insert pipette into one of the fourbores as far as it will go (bores indicated by arrows on the side of theVolumeter). Push the pin forward tothe stop whereby lubricant is pressedinto the hole.
Fig. 11 Lubrication of the bearings38275
Trouble ShootingIf the timer mechanism of the Volumeter 3000 does not start followingsterilization using superheated steam,th is is due to the small shafts sticking intheir bearings. This can be corrected bygiving the Volumeter a sharp turnaround its longitudinal axis.
When a new piece of tempered plateglass hast to be put in, remove the retaining ring over the glass using apointed instrument (such as forceps) bylevering it out to the front. When theVolumeter is tipped. the glass will fallout.
When putting in a new glass:
• Clean the dial face;
• Clean the new glass before putting itin;
• Snap in the retaining ring.
7
Order List
DescriptionDesignationOrder-No.
Minute Volumeter 3000
Volumeter 30002 M 18250
Minute Volumeter 3000
Volumeter 3000 in
in carrying case for measurementcarrying case2 M 18476
of vital capacity; complete with 2 face masks, valve, mouth pieceand connecting hosesSpare and wearing partsTempered plate glass
GlassE9285
Retaining ring
Retaining ringE 9284
Gasket
GasketsM 22154
(on upper connecting piece)(pack of 10)
Bag with cardboard mouth pieces;
Bag mouthpiece2M2355
bag contains five pieces (belongs to Volumeter in carrying case)Lubrication-Set for
Lubrication-Set2 M 18180
Volumeter 3000
Subject to alterations!
8