dycor system 200 software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · viewing devices...

79
PN 90422VE Rev C COMMITTED TO TOTAL QUALITY Q M T Dycor System 200 Software User Manual Process Instruments 150 Freeport Road Pittsburgh, PA 15238

Upload: others

Post on 10-Nov-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

PN 90422VE Rev C

COMMITTED TO TOTAL QUALITY

QMT

Dycor System 200 Software

User Manual

Process Instruments150 Freeport Road

Pittsburgh, PA 15238

Page 2: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

ii | Dycor System 200 Software

AMETEK Process Instruments150 Freeport RoadPittsburgh, PA 15238, USAPhone: 412-828-9040Fax: 412-826-0399

AMETEK Process Instruments455 Corporate BoulevardNewark, Delaware 19702, USAPhone: 302-456-4400 (Main)

800-537-6044 (Service)800-222-6789 (Ordering)

Fax: 302-456-4444

AMETEK Process InstrumentsAMETEK-APIE OfficePostfach 2165D-40644 Meerbusch

ORRudolf-Diesel Strasse 16D-40670 MeerbuschGermanyPhone: 49-21-59-9136-0Fax: 49-21-59-9136-39

AMETEK Process Instruments7950 Whithorn DriveHouston, Texas 77095, USAPhone: 281-463-2820Fax: 281-463-2701

AMETEK Singapore PVT. Ltd.10 Ang Mo Kio Street 65#05-12 TechpointSingapore 569059Republic of SingaporePhone: 65-484-2388Fax: 65-481-6588

AMETEK Process Instruments2876 Sunridge Way N.E.Calgary, AlbertaCanada T1Y 7H9Phone: 403-235-8400Fax: 403-248-3550

Offices

© 2001 AMETEK

This manual is a guide for the use of the Dycor System 200 Software. Data herein has been verified and validated and is believed adequate for theintended use of this software. If the software or procedures are used for purposes over and above the capabilities specified herein, confirmation of theirvalidity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability. This publication is not alicense to operate under, or a recommendation to infringe upon, any process patents.

Page 3: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

| iii

Table of Contents

CHAPTER 1 Introduction to System 200 SoftwareMinimum PC Requirements for System 200 Software ......................... 1-1Dycor System 200 Software Installation .................................................. 1-2

CHAPTER 2 User Interface / Setting Up your DeviceThe View Menu ........................................................................................... 2-1Setting Up Devices...................................................................................... 2-3

Changing a Device Property .............................................................. 2-3Tune parameters tab ......................................................................... 2-4

Viewing Devices .......................................................................................... 2-8

CHAPTER 3 Modes of OperationAnalog Mode Overview ............................................................................ 3-3Bar Mode Overview ................................................................................... 3-4

Placing Labels ........................................................................................ 3-4Meter Mode Overview .............................................................................. 3-6Annunciator Mode Overview .................................................................. 3-7Trend Mode Overview ............................................................................... 3-8Tabular Mode Feature................................................................................. 3-9Leak Check Mode Overview .................................................................. 3-10Custom Modes .......................................................................................... 3-11

Creating a Custom Mode .................................................................. 3-11Adding a display window in custom mode ................................... 3-11Moving or resizing a window in custom mode ............................. 3-12Deleting a window in custom mode .............................................. 3-12Saving custom mode files .............................................................. 3-12Opening custom mode files ........................................................... 3-13Adding custom mode files to the mode menu ................................ 3-13

CHAPTER 4 PreferencesSelect Edit/Preferences ............................................................................... 4-1

CHAPTER 5 Scanning DataDefining Scan Settings ............................................................................... 5-1Turning the Filament On/Off .................................................................... 5-4Turning the Multiplier On/Off ................................................................. 5-4Scanning ....................................................................................................... 5-5

Starting a Scan ...................................................................................... 5-5Stopping a Scan .................................................................................... 5-5Scheduling Scans .................................................................................. 5-5

Page 4: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

iv | Dycor System 200 Software

CHAPTER 6 CalculationsPerform a Calculation .............................................................................. 6-1

CHAPTER 7 Setting Up the DisplayDIsplay Screen .......................................................................................... 7-1

Title ...................................................................................................... 7-2Enable Total Pressure ......................................................................... 7-2Display Mass Range........................................................................... 7-2

Low mass ........................................................................................ 7-2High mass....................................................................................... 7-2

Autoscale ............................................................................................. 7-2Logarithmic ......................................................................................... 7-3Upper Limit ........................................................................................ 7-3Lower Limit ........................................................................................ 7-3Units .................................................................................................... 7-3Mode .................................................................................................... 7-3Channel Color .................................................................................... 7-4Minutes ............................................................................................... 7-4Enable Leak Tone .............................................................................. 7-4Hide ..................................................................................................... 7-4

CHAPTER 8 Setting AlarmsAccessing the Set Alarms Screen ............................................................ 8-1

Enable All Alarms .............................................................................. 8-2High Limit ........................................................................................... 8-2High Warning ..................................................................................... 8-2Low Limit ............................................................................................ 8-3Low Warning ...................................................................................... 8-3AutoSet Button ................................................................................... 8-3

Action Button ............................................................................................ 8-4

CHAPTER 9 Analog OutputsSetting the Analog Outputs .................................................................... 9-1

CHAPTER 10 Degassing a Quadrupole HeadHow to Degas the Quadrupole Head ..................................................10-1

CHAPTER 11 Automatic CalibrationAuto Calibration Dialog Box ..................................................................11-1Performing an Automatic Calibration ...................................................11-4Resolving an RF Tune Error ...................................................................11-7Coarse Adjustment of Mass Position ....................................................11-8

Page 5: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

| v

CHAPTER 12 Working with Data FilesSaving Data Files .....................................................................................12-1Opening Data Files .................................................................................12-3Moving Through Data Records .............................................................12-4Converting Data File ...............................................................................12-5

CHAPTER 13 Printing Data

CHAPTER 14 Dynamic Data Exchange (DDE)Copying Magnitudes ..............................................................................14-1Copying Channels ...................................................................................14-1Direct DDE Control .................................................................................14-2

Requesting Values in the Devices & Log Window. ......................14-3Microsoft Excel 7.0 Example ...................................................................14-5

CHAPTER 15 Setting Security Levels

CHAPTER 16 System Error Messages

Page 6: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Introduction | 1-1

INTRODUCTION TO SYSTEM 200 SOFTWARE

The Dycor System 200 software provides a graphical user interface tocontrol your RGA data. Running under Windows 95 or Windows NT, itprovides powerful features and flexible operating modes.

Minimum PC Requirements for System 200 Software

System 200 Software requires the following:

� Pentium 100 or better

� 16 MB RAM for Windows 95/98, 32 MB RAM for Windows NT

� Sound Card (for Leak mode tone and alarms). Recommended, notrequired.

� Microsoft Windows 95/98 or Windows NT 4.0

� 640x480, 256-color display.

For optimum display quality we recommend running at 800x600 or1024x768, High Color (16-bit).

To change your display resolution or color palette:

� Open Display Properties from the Windows Control Panel.

� Select the Settings tab.

� In Color Palette option, set High Color (16-bit).

� In Desktop Area option, set 800 x 600 pixels or more.

Page 7: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

1-2 | System 200 Software

Dycor System 200 Software Installation

1. Insert the Setup Disk 1 in your floppy disk drive.

2. Choose RUN... from the Windows Start Menu.

3. Type the drive letter, followed by a colon ( : ) and a backslash (\) andthe word �setup.�

a:\setup

4. Follow the instructions on your screen. Click Next to continue theSetup process.

Instructions For Using Dycor System 200 Software

Follow the instructions below to configure the Dycor System 200 Softwareon your system to obtain RGA and other input data. It is presumed thatyou have already correctly set up the appropriate hardware. See yourdevice hardware manual for information on hardware setup.

1. Start the Dycor System 200 Software.• Choose the Windows Start button, select Programs and Dycor

System 200. If you did not choose the default options duringinstallation, the paths or names could have changed. Anotheroption is to use the file manager program to look for the Dycor 200subdirectory and to activate the Dycor.exe executable program.

2. Set Device Settings.Communications settings, tune and filament protection parametersand device type.

3. Select a Display Mode.Standard RGA-type modes available through Mode menu option.

AnalogBarTrendTabularMeterAnnunciatorLeak

Page 8: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Introduction | 1-3

4. Define Scan Settings.Move the mouse to the display window. Click the right mouse button.Click Scan and define the following:

• Device to be scanned,• Mass or range of masses to be scanned,• Number of samples for each scanned value, and• Dwell time.

5. Set-Up Display Parameters.You can create your own custom displays through Custom modes.These can be saved for retrieval later. Three default custom modes areprovided to make it easier to create and save custom scan, display,alarm and analog output settings. Different options are providedwithin each mode. You also have the option of linking data andincluding calculations based on linked data.

6. Set-Up Alarms. (optional)Includes the ability to have a master alarm that triggers based on avariety of other alarm conditions using Boolean logic.

7. Set-Up Analog Outputs. (optional)

8. Schedule Scans.Scan continuously or schedule scans to be performed at timed inter-vals.

9. Start Scanning.

10. Save Scanned Data for Later Viewing or Printing. (optional)

Page 9: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

1-4 | System 200 Software

This page intentionally left blank.

Page 10: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

User Interface and Setting Up Your Device | 2-1

USER INTERFACE / SETTING UP YOUR DEVICE

The Dycor System 200 screen allows you to customize how data is viewedand how otherwise to set up your display.

The View Menu

From the View Menu select the menu option for the viewing area youwant displayed. If the option has a checkmark beside it, the viewing areais active. To toggle the on/off checkmark, select the menu option again.Toolbars displayed on the Main Window (Figure 2-1) include:

• Standard ToolbarIcons used most often including the main display modes.

• Custom Mode ToolbarIcons that are used for the Custom mode.

• Data File ToolbarIcons relating to saving data files.

Figure 2-1.Dycor System 200screen

Standard Toolbar

Custom Mode Toolbar

Data File Toolbar

Main Window

Status Bar

Page 11: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

2-2 | System 200 Software

• Status BarBar at the bottom of the screen that displays system messages includ-ing whether or not the system is currently scanning.

• Devices & LogDisplays a window showing all devices and their status. Also showsthe system event log.

• LibraryDisplays the mass spectra library window. Use the arrow buttons onthe main toolbar to step through the library spectra.

Page 12: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

User Interface and Setting Up Your Device | 2-3

Setting Up Devices

Changing a Device Property

• To change device properties, select Edit/Device Properties.

General tab

Figure 2-2.Edit Device Properties,General tab.

The following are defined using the General tab (Figure 2-2).

• Communications Section

Port Select the COM port on the computer used toconnect to your device.

• Device Section

Name Provides a name for the device. It is used by the sys-tem when printing or displaying information for thedevice.

Description

Page 13: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

2-4 | System 200 Software

• Device Enabled

Not Checked System temporarily ignores the device.Checked Device is available for use by the system.

• Live Data

Not Checked System uses demonstration data.Checked System uses live data from quadrupole head.

Tune parameters tab

Tune FileThe tune parameters (Figure 2-3) are saved in the file named in theTune File text box. Each display mode or configuration can have adifferent tune file.

Figure 2-3.Device Propertiesdialog box - TuneParameters tab.

Browse Button Retrieves a previously saved tune file. Navigate tothe location of the file and select it.

Save Button Saves current tune parameters to a tune file (de-fault extension .tun). These settings can be im-ported later using the Tune File text box or usingthe Browse button.

Make Default Sets the current tune file as the one to use in allmodes that use this quadrupole.

Page 14: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

User Interface and Setting Up Your Device | 2-5

Emission CurrentElectron emission current used to ionize molecules in the ion source.The filament power will be adjusted until the actual electron currentequals this value. 1E-3 amps (1 ma.) is the best choice for mediumpressure work. If this value is changed, the sensitivity of the analyzerwill change and it should be recalibrated.Range: 1E-4 to 1E-2 amps (.1 to 10 ma.)

Electron EnergyEnergy of the ionizing electrons in the ion source. It is measured inelectron-volts or the difference in potential between the filament andthe source grid. Most data available on the ionization efficiencies andcracking patterns of various molecules are taken at -70 volts and rarelywill this parameter need to be changed. During a degas operation thisvoltage will increase to -250 volts until the degas is complete.Range: -150 to -30 volts

Focus VoltageNormally set to -20 volts. A slight increase in sensitivity can be ob-tained by adjusting this voltage but severe mass discrimination canoccur at extreme limits.Range: -200 to +10 volts

Repeller Voltage (Non-CIS Head)This parameter sets the voltage on the repeller grid that surrounds theionizer. It is normally set to a few volts above the Electron EnergyParameter so the electrons are contained within the ionizer andshould never be set positive with respect to the filament. This param-eter is normally set to minus 72 volts when the Electron EnergyParameter is set to -70 volts.Range: -200 to +10 volts

FrequencyRF frequency applied to the mass filter. Should be adjusted so the RFTune Value is minimized within its typical operating range. Frequencyparameter should be adjusted up or down until the RF Tune Value isminimized. Normally use the Auto Calibration.Range: 1.000 to 3.000 MHz

Cal Low Mass*Normally keep this value set to 1. At the low mass, the low position,low sensitivity and low resolution calibration values will be used.Range: 1 to 100

Cal High Mass*Set to 100 or 200 depending on whether the device has a 100- or 200-AMU range. At the high mass, the high position, high sensitivity andhigh resolution calibration values will be used.Range: 1 to 200

Page 15: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

2-6 | System 200 Software

Low PositionUsed to adjust the peak positions at the low end of the scan range sothe peaks correspond to their associated AMU values. Increasing thisvalue shifts the peaks to the right. Decreasing this value shifts peaksto the left.Range: -10 to 10

High PositionUsed to adjust the peak positions at the high end of the scan range sothe peaks correspond to their associated AMU values. Increasing thisvalue shifts the peaks to the right. Decreasing this value shifts peaksto the left.Range: -10 to 10

Low ResolutionUsed to adjust resolution of the peaks at the low end of the scanrange. Normally adjusted for a 10% valley between the peaks ofadjacent masses.Range: 50 to 4100

High ResolutionUsed to adjust resolution of the peaks at the intermediate and higherend of the scan range. Normally adjusted for a 10% valley betweenthe peaks of adjacent masses.Range: 50 to 4100

Low SensitivityThe ion energy in electron-volts that the ions entering the mass filterwill have for the low masses. If the energy is increased, the sensitivitywill increase but the resolution may decrease. Standard setting (6.00)provides a partial compensation for the decrease in sensitivity at highmasses while keeping the best possible resolution.

Range: 3.00 to 10.00

High SensitivityThe ion energy in electron-volts that the ions entering the mass filterwill have for the high masses. If the energy is increased, the sensitivitywill increase but the resolution may decrease. Standard setting (7.00for 100-AMU units and 8.00 for 200-AMU units) provides a partialcompensation for the decrease in sensitivity at high masses whilekeeping the best possible resolution.Range: 3.00 to 10.00

Page 16: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

User Interface and Setting Up Your Device | 2-7

Amplifier CalibrationWith Emission Current set at 1.0E-3, an AMP CAL of 5.0E-3 allows thepartial pressure readings to be approximated in torr. This is a gainparameter by which each current reading is multiplied. When Ampli-fier Calibration is set to one, the partial pressure readings are in amps.Range: 1.0E-10 to 1.0E10

Total SensitivityThis calibration factor can be set between 0.1 and 999.9. It can beadjusted so that the total pressure reading agrees with an ion gaugereading. The number is a simple multiple of the measured totalcurrent to the Faraday Cup when reading total pressure.Range: .1 to 999.9

Multiplier VoltageApplies only if you have a multiplier-type analyzer (not a FaradayCup). This is the multiplier voltage placed on the multiplier. A typicalvalue is -1000 volts.Range: 0 to -2000 volts

Page 17: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

2-8 | System 200 Software

Viewing Devices

1. Select View/Devices & Log or click on the icon. A log is dis-played showing the active device and its status.

2. Here it is possible to view a log of all related activity for the activedevices, including when the filament or multiplier was turned on oroff and when scans were started or completed as well as any alarmconditions. It also provides information on the data source for thereading in the Source field. The Status field indicates whether live ordemo data is being used. Demo data can be used to simulate real datafor testing purposes.

3. To remove the log display select View Devices & Log again or click onthe icon. The system log is saved in a text file in the �log� folder.

4. To toggle the on/off state of any of the viewing areas, select the menuoption again.

The system log is saved in a text file in the �Log� folder.!NOTE

Page 18: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-1

MODES OF OPERATION

Any setting you defined in a mode will be retained by the systemwhen you change the mode or exit the program. Any display or scanparameters that you defined for a mode will not be transferred to anew mode when you switch modes. However, the system will con-tinue to scan when you change modes.

• Analog ModeScan over a range of masses to view both peak shapes and heights.

• Bar ModeScan over a range of masses to view mass peaks and heights.

• Meter Mode View the partial pressure of a specific mass or digital/analog-compat-ible module input in a meter-type format. Alarms can be set up anddisplayed to the right of the meter to view the current reading inrelation to the high- and low-alarm limits defined. In Meter mode,multiple meters can be set up with, for example, each meter corre-sponding to a different mass from a different quadrupole head.

• Annunciator Mode Track whether a measured parameter is within acceptable levels usinga simple color/text indication. If the level is

within the alarm limits defined = GREEN OK

above the high-alarm warning = YELLOW HIGH

above the high-alarm limit = RED HIGH

below the low-alarm warning level = YELLOW LOW

below the low-alarm limit = RED LOW

• Trend Mode View trends of selected masses over time.

!NOTE

Page 19: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-2 | System 200 Software

• Tabular Mode View information is a textual, table-type format.

• Leak Check ModeCheck for leaks in the vacuum system by monitoring the level of aselected tracer gas over time.

• Custom ModeCreate custom setups including all scanning, display, alarm andanalog output options. Custom mode setups can include either asingle mode display window or multiple mode display windows. Forexample, you can customize the screen to show one Analog modedisplay window, one Trend mode window and two Meter windows.Or, you could set up two Analog mode display windows or two Barmode windows.

Page 20: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-3

Analog Mode Overview

Analog mode allows scanning across a range of masses to observe theirpartial pressure. Both height and shape of peaks can be viewed.

1. Select Analog from the Mode menu or select the icon from thetoolbar.

2. Specific values for different mass units can be viewed.

• Move the cursor to the main Dycor screen area and hold down theleft mouse button.

• The title bar will change to allow you to view the specific partialpressure and possible gas identification for the selected mass.

• You can move the cursor to show partial pressure for differentmasses.

The gas identification list is stored in a text file called �mass.txt.� Itcan be customized for your application.

!NOTE

Page 21: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-4 | System 200 Software

Bar Mode Overview

Bar mode allows scanning across a range of masses to observe their partialpressures.

1. Select Bar from the mode menu or select the icon from thetoolbar.

2. Specific values for different mass units can be viewed.

• Move the cursor to the main Dycor screen area and hold down theleft mouse button.

• The title bar will change to allow you to view the specific partialpressure and possible gas identification for the selected mass.

• You can move the cursor to show partial pressure for differentmasses.

The gas identification list is stored in a text file called �mass.txt.� Itcan be customized for your application.

Placing Labels

A label can be placed on the screen only from an Analog or Bar modewindow.

1. Right mouse click anywhere from within the Analog or Bar modemain screen area.

2. Select Label. The Label dialog box appears.

!NOTE

Figure 3-1.Label dialog box forAnalog and Bar modes.

Page 22: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-5

3. Type the label text into the text box and then click OK.

4. A pen icon appears. Move the pen icon to the location where the labelis to appear and click the left mouse button.

5. To move the location of the label after it is placed, press and holddown the left mouse button on the label and move the label to thedesired location.

6. To delete the label, move the label off the main screen.

Page 23: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-6 | System 200 Software

Meter Mode Overview

Meter mode allows use of up to 12 meters. Each meter can be indepen-dently set to monitor one of the following:

• Specific partial pressure from a quadrupole head. For example, youcould set up the meter to look for the partial pressure of Mass 28.

• Calculation or link from another data source.

• In addition, you can define unique scanning, display, alarm, andanalog output settings.

1. Select Meter from the Mode menu or select the icon fromthetoolbar.

2. Each meter has a bar representing a measured value. Text under themeter displays the numeric value of the measured parameter.

3. Each meter can be set to display alarm limits. As long as the measuredvalue is within the high and low alarm limits, the meter bar appears ingreen.

4. When an alarm condition occurs, the meter bar appears in yellow orred.

5. Any other actions associated with alarms are activated if the limit isexceeded (e.g., running an information file, setting off a digital oranalog output).

6. Before changing meter properties, it is necessary to select the meter byclicking on it. The meter will be highlighted with a yellow rectangle.

Page 24: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-7

Annunciator Mode Overview

Annunciator mode allows viewing of up to 12 Annunciator displaywindows, each of which can be set to monitor the following:

• Specific partial pressure from a quadrupole head. For example, youcould set up the meter to look for the partial pressure of Mass 28.

• Calculation or link from another data source.

• In addition, you can define unique scanning, display, alarm, andanalog output settings.

1. Select Annunciator from the Mode menu or select the iconfromthe toolbar.

2. Annunciator is normally used to define alarm limits so that one canview whether the selected value is within the alarm limits using visualindicators.

3. For each channel, a green light indicates the value is within acceptablelimits.

4. A yellow light indicates an alarm warning.

5. A red light indicates an alarm condition.

Page 25: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-8 | System 200 Software

Trend Mode Overview

Trend mode allows monitoring of up to eight different channels versustime. A channel can correspond to any of the following:

• Specific partial pressure from a quadrupole head. For example, youcould set up the meter to look for the partial pressure of Mass 28.

• Calculation or link from another data source.

• In addition, you can define unique scanning, display, alarm, andanalog output settings.

1. Select Trend from the Mode menu or select the icon from thetoolbar.

2. The main Trend mode screen shows each channel as a colored line.The Y axis shows partial pressure or measured value; the X axis showstime.

3. You can view the exact value of a trend line by holding down the leftmouse button in the trend area. Display the Data File Toolbar to seethe time of each data point.

Page 26: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-9

Tabular Mode Feature

Tabular Mode allows use of up to eight channels per display window toview the following information in table-type format:

• Name of channel

• Value for selected channel

• Existence of an alarm condition

• Specific partial pressure from the quadrupole head

1. Select Tabular from the Mode menu or select the icon from thetoolbar.

2. If an alarm was defined and the alarm has been exceeded, an alarmindicator will appear to the right of the main table.

Page 27: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-10 | System 200 Software

Leak Check Mode Overview

Leak Check mode checks the vacuum system for leaks.

By applying a tracer gas of known mass to the outside walls of thevacuum system, you can monitor the internal vacuum system partialpressure of the tracer gas. An increase in partial pressure indicates aleak. You can further isolate the leak by observing the increase in themeasured tracer gas in response to different injection points.

1. Select Leak Check from the Mode menu or click the icon onthe toolbar.

2. Leak Check display window shows the level of the selected mass onthe Y axis and time on the X axis.

3. Up and down arrows (on the right of the leak check screen) indicatewhether the current measured value of the selected tracer gas isincreasing or decreasing.

4. Current value of the gas is also shown in text format between thearrow indicators (e.g., 10E-11).

5. Recommended sample dwell time is 120 milliseconds (ms).

6. If you have a sound card and speakers, you can enable a leak tone inthe Display Properties dialog box.

Page 28: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-11

Custom Modes

Custom modes allow the mixing and matching of the best features ofvarious modes to create a unique mode setup. Three default custom modesetups are included with the Dycor software and more can be added asneeded.

For each display window added in Custom mode, the scanning, display,alarm and analog output options are set as if in that mode. For example, ifadding an Analog mode window within the Custom Mode, all settings forthat window are defined as if in Analog mode.

Creating a Custom Mode

1. Select File/New or select the icon on the toolbar.

2. Select Edit/Add Display. Select the type of display you want to addfrom the list of standard mode types.

3. Specify the data source. Select the applicable source of data from theData Source selection.

4. Select File/Save As to save the Custom mode. Or select one of

icons in the toolbar to save the Custom mode to one ofthe predefined Custom mode file names (Dycor1.dyc, Dycor2.dyc,Dycor3.dyc). Custom mode files are given a .dyc file extension.

5. To retrieve the custom mode file at a later time, select File/OpenConfiguration. If you used one of the predefined icons, just click onone of the icons to retrieve the file.

Adding a display window in custom mode

1. Select Edit/Add Display. Choose the type of display window to add.

2. Select the data source for the display window.

3. The system adds a display window to the Custom mode screen.

Page 29: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-12 | System 200 Software

!NOTE

Moving or resizing a window in custom mode

1. Select the display window to move or resize.

2. The window is highlighted in yellow.

3. Move the window.

4. Press and hold down the mouse button (just an outline of the windowis visible) while moving the display window, then release the mousebutton.

5. Resize the window.

6. Use standard Windows 95 methods.

Deleting a window in custom mode

1. Click on the window to delete.

2. The window is highlighted in yellow.

3. Select Edit/Delete Display, press the Del button or select theicon from the Custom Mode toolbar.

4. The display window is deleted.

Saving custom mode files

1. Select File/Save As. You can save the Custom mode setup under a newname or use one of the predefined icons.

2. By default, the system places these files in the folder where the DycorSystem 2000 Software is installed and gives the file a .dyc extension.

3. Select the Save button to save the file.

If you save a custom mode file in the install folder with the name�startup.dyc� it will automatically be opened when the programstarts.

Page 30: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Modes of Operation | 3-13

Opening custom mode files

1. Select File/Open Configuration or select the icon from theCustom Mode toolbar.

2. Navigate to the location of the Custom mode file to open.

3. Select the Custom mode file.

4. Click Open.

Adding custom mode files to the mode menu

1. Select the Set Menu option from the Mode menu.

2. Set Custom Menu dialog box appears. You can associate a Custommode file to appear at the bottom of the mode menu.

3. You can also provide a menu/button name to appear instead of thefilname at the bottom of the menu.

4. Browse Button - navigates to the location of the Custom mode file.

5. Menu/Button Name - this is the name that appears at the bottom ofthe mode menu.

These files are recalled when pressing the icons.

Figure 3-2.Set Custom Modemenu screen.

Page 31: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

3-14 | System 200 Software

This page intentionally left blank.

Page 32: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Preferences | 4-1

PREFERENCES

The Automatic Filament Enable feature allows you to automatically turnon the filament for the quadrupole head.

Select Edit/Preferences

• Automatic Filament Enable Check BoxWhen checked: turns the filament on for the connected quadrupolehead automatically when scan is started.

• Pressure UnitsSelect the pressure units to use for the quadrupole device.

• DDE Item SeparatorSelect the character that will separate multiple data items in a DynamicData Exchange (DDE) message.

• Trend ColorsSet the default trend channel colors.

• Data File SizeSet the maximum size for a data file. This can be an integer between 1 and 16.The normal value is 1 megabyte. Setting the data file size to a higher valuewill increase the size of your data files as well as the size of the trend graphbuffer. As a result of this increase, it will require more disk space and memory.

Figure 4-1.Preferences DialogBox, AutomaticFilament Enable checkbox.

Page 33: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

4 - 2 | System 200 Software

This page intentionally left blank.

Page 34: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Scanning Data | 5-1

SCANNING DATA

This chapter contains information on how to scan data, including how toperform calculations or link data from a different display.

Defining Scan Settings

Scan settings allow you to define the type of information you want to gatherfrom the selected device. For Trend and Tabular modes you must first se-lect a channel before you can define scan settings.

Right click on the applicable display window and then click Scan (Figure5-1).

• Data Source SectionSelect the device from which to scan and temporarily disable the scan-ning device.

DisabledAll other options from Scan tab are disabled. Use this option todisable the display.

DeviceUse this list box to select the device from which to scan.

Figure 5-1.Properties Screen -Scan tab.

Page 35: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

5-2 | System 200 Software

FunctionDetermines the function of the device. Options are Ion Current orTotal Pressure. In addition, the Dymaxion has Analog and DigitalInput functions.

Total Pressure - options for specific mass unit or dwell time aredisabled. Only total pressure is displayed.

Linking Data in a DisplayData can be linked from one window into another. This can savescanning time since the system only needs to scan the data onetime.

1. Select the display window to receive the linked data from an-other data source.

2. Click the right mouse button from a display window and thenclick Scan.

3. From the Data Source section, select the Link From button.

4. Select the data source to link from using the list box to the rightof the Link From button.

You cannot link data into an Analog or Bar mode window.

CalculationDisplay the results of a calculation.

• Scan MassRange (for Analog or Bar mode)Set low and high mass numbers to define the range to scan.

• Mass (all modes other than Analog and Bar)Define the specific mass number to track. You can also select a gas fromthe list to the right of the mass number.

The gas list is stored in a text file called �gas.txt.� It can be custom-ized to your application.

!NOTE

!NOTE

Page 36: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Scanning Data | 5-3

• Samples per AMU (Analog Mode only)Number of samples the system will take for each selected AMU. Thehigher the samples-per-AMU, the better the resolution. In Analog mode,more samples per AMU allow you to resolve separate peaks more eas-ily. You can set this value between 4 and 15.

• Dwell TimeTime the system allots for each sample. Dwell time determines scan-ning speed and background noise. The more time allotted for eachsample, the higher the signal-to-noise ratio. Very low pressures requirea slower scan speed.

• Auto ZeroChecked - Allows subtraction of the amplifier drift effects from all read-ings.

There will be a speed penalty if Autozero is turned on because the systemtakes two readings for each reading displayed.

Page 37: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

5-4 | System 200 Software

Turning the Filament On/Off

1. Select Control/Filament or click the icon on the toolbar.

Turning the Multiplier On/Off

1. Select Control/Multiplier or select the icon from the toolbar.

4. To turn the multiplier on for all quadrupole heads, select the All Onbutton. To turn off the multiplier for all quadrupole heads, select theAll Off button.

Page 38: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Scanning Data | 5-5

Scanning

Starting a Scan

1. Select the icon from the toolbar or select the Control/Scan menuoption.

2. Scanning will appear in the status bar at the bottom of the screen.

3. If you have scheduled scans at timed intervals and the system has fin-ished the previous scan and is waiting to start the next scan, the timeremaining before the next scan begins is displayed.

4. Idle is displayed when the system is not scanning.

Stopping a Scan

Select the icon from the toolbar or select the Control/Scan menu op-tion to toggle the scan off.

Scheduling Scans

1. Select Control/Schedule or click the icon in the toolbar.

2. The Schedule dialog box appears (Figure 5-2).

� Single Scan Radio ButtonTells the system to run a single scan when the icon is selected.The system will stop scanning after one scan.

Figure 5-2.Schedule dialog box.

Page 39: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

5-6 | System 200 Software

� Continuous Radio ButtonTells the system to run scans at timed intervals once the icon isselected.

� Scan Interval Text BoxType a number to specify the time delay between scans. The unitsthis number represents depend on whether the scans were sched-uled using minutes or seconds as the time interval.

MinutesUse minutes as the interval between successive scans. Uses the num-ber specified in the Scan Interval text box to determine the schedul-ing.

SecondsUse seconds as the interval between successive scans. Uses the num-ber specified in the Scan Interval text box to determine the schedul-ing.

Page 40: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Calculations | 6-1

CALCULATIONS

Calculations are available for Analog and Bar modes only.

Figure 6-1.Properties screen,Calculation tab.

Perform a Calculation

1. Right click from a display window.

2. Click Calculation.

OffsetAdds the offset value to the measured reading. Set value to 0 for nooffset.

GainMultiplies the reading by thenumber entered in the gain text box.Set value to 1 for no gain.

Offset 2Adds the entered value after the first offset is applied and the gainis applied. Set value to 0 for no offset value.

3. Click OK to apply these values to the readings.

Page 41: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

6-2 | System 200 Software

This page intentionally left blank.

Page 42: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Setting Up the Display | 7-1

SETTING UP THE DISPLAY

DIsplay Screen

To set the display for a mode, click the right mouse button on a displaywindow and then click Display. The Display tab is highlighted. For modesusing channels, Trend and Tabular Modes, you must first select a channelbefore you can define the scan settings.

Figure 7-1.Properties Screen -Display tab. Analogmode.

Figure 7-2.Properties Screen -Display tab. Trendmode.

Page 43: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

7-2 | System 200 Software

Title

Type the title for the display window. It will then appear in the title bar forthe window and in other areas of the software to identify the display win-dow.

Enable Total Pressure

Check Box. Applies only for Analog/Bar modes. When checked - total pres-sure for the quadrupole head is displayed in the top right of the title bar.

Display Mass Range

Applies only for Analog/Bar modes. Determine which mass range to dis-play. Note that the instrument continues to scan over the range selectedunder the scan properties but will only display the selected range in thedisplay properties. This enables the user to zoom in on a selected rangewhile the instrument continues to collect data over the entire scan range.

Low mass

Tells the system the first mass unit to display on the far left of the Xaxis. If the mass scanning range is changed, the low mass display set-ting will automatically change to match the new mass scanning range.If you want the display to show a different range than the scanningrange, you can return to this Display button option and change therange of AMUs to display on the screen without affecting the scanningrange.

High mass

Tells the system the last mass to display on the X axis. If the mass scan-ning range is changed, the high mass display setting will automaticallychange to match the new mass scanning range. If you want the displayto show a different range than the scanning range, you can return tothis Display button option and change the range of AMUs to displayon the screen without affecting the scanning range.

Autoscale

When checked the system automatically sets the Y axis based on the high-est value from the previous scan across the mass range. If you turn on theAutoscale feature, the system overrides the Y axis. As the highest peak

Page 44: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Setting Up the Display | 7-3

increases or decreases, the display scale is changed to accommodate thechanging highest peak.

Logarithmic

When checked the software uses a logarithmic scale for the display. If notchecked it uses a linear scale to display readings.

Upper Limit

This value is used only when the Autoscale check box is not checked. Youenter the upper limit to show on the Y axis.

Lower Limit

This value is used only when the Autoscale check box is not checked. If theLogarithmic box is checked, use this list box to select the number of loga-rithmic decades to display on the Y axis using the upper limit as the top ofthe display range. If the Logarithmic is not checked, enter a value into thistext box for the lower limit.

Units

Click the radio button for the type of units on which you want the scannedinformation displayed.

CustomType a custom unit name.

PrecisionApplies to all except Analog/Bar modes. Defines how precise the read-ing should display by specifying how many decimal places to use. Forexample, if the precision is set to 2, a reading might be 5.22.

Mode

Applies only to Analog/Bar modes. Enables subtract or compare modes.

NormalNormal operations. Does not include any data files with the currentdata.

SubtractSubtracts the saved data from the current data.

Page 45: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

7-4 | System 200 Software

CompareShows the saved data file side-by-side with the current data.

Channel Name

Applies for Trend and Tabular modes. Names the displayed channel.

Channel Color

Applies for Trend Mode. Defines a different color for the displayed chan-nel.

Minutes

Applies for Trend, Leak modes. Defines the time interval to show on thedisplay.

Display Values

If checked, displays channel values in Trend Mode.

Enable Leak Tone *

If checked, an audio tone will track the measured ion current.

Hide**

If checked, the data channel will not be displayed but will keep trackingthe data. This is useful when viewing a scan that overlaps another scan.

* Leak Mode only**Trend Mode only

Page 46: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Setting Alarms | 8-1

SETTING ALARMS

Accessing the Set Alarms Screen

Setting alarms includes defining alarm limits and defining the actions thatshould be taken when an alarm limit is reached. Alarm limits also appearon the main display for the current mode. High and low limit alarms ap-pear in red, high and low warning alarms appear in yellow.

To set alarms, click on the right mouse button from the displaywindow and then click Alarms.

For Trend and Tabular modes, first select the channel as the sourceof the alarms before you define the alarm values.

!NOTE

Figure 8-1.Set Alarms screen.

Page 47: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

8-2 | System 200 Software

Enable All Alarms

When checked - enables all high and low alarms. Deactivated when box isnot checked. Alarms can be deactivated when, for example, your system isnot operating.

High Limit

When checked - enables the high limit alarm.

NameProvide a name for the alarm. Used to identify the alarm to the system.

ValueType in the high limit value for the alarm. For analog mode, any massvalues that exceed the alarm limit are counted even if they are from thesame scan.

CountDefines the number of consecutive readings that must be above highalarm limit before the alarm is triggered. Higher count values decreasethe possibility that a transient or noise signal will set off a false alarm.

High Warning

When checked - enables high warning alarm system.

NameProvide a name for the alarm. Used to identify the alarm to the system.

ValueType in the high limit value for the alarm. Must be lower than the highalarm value. For analog mode, any mass values that exceed the alarmlimit are counted even if they are from the same scan.

CountDefines the number of consecutive readings that must be above thislimit before the alarm is triggered. Higher count values decrease thepossibility that a transient or noise signal will set off a false alarm.

Page 48: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Setting Alarms | 8-3

Low Limit

When checked - enables the low limit alarm.

NameProvide a name for the alarm. Used to identify the alarm to the system.

ValueType in the low limit value for the alarm. For analog mode, any massvalues that fall below this limit are counted even if they are from thesame scan.

CountDefines the number of consecutive readings that must be below thislimit before the alarm is triggered. Higher count values decrease thepossibility that a transient or noise signal will set off a false alarm.

Low Warning

When checked - enables the low warning alarm.

NameProvide a name for the alarm. Used to identify the alarm to the system.

ValueType in the low limit value for the alarm. Must be higher than the lowalarm value. For analog mode, any mass values that fall below this limitare counted even if they are from the same scan.

CountDefines the number of consecutive readings that must be below thislimit before the alarm is triggered. Higher count values decrease thepossibility that a transient or noise signal will set off a false alarm.

AutoSet Button

When selected, this button automatically sets the alarm levels as follows:

High Limit 10 times the current scan valueHigh Warning 5 times the current scan valueLow Warning 5 times less than the current scan valueLow Limit 10 times less than the current scan value

Page 49: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

8-4 | System 200 Software

Action Button

For all four alarms. The Action button can be used to specify what action the systemshould take when the specified alarm is triggered.

Dialog Box SectionSpecify a message to display when the alarm is reached.

Info File Select a file by typing in the name.

Browse Navigate to the location of the file and select it.

This file will be launched when you click on the Information button inan alarm dialog box. The file could, for example, provide help to theoperator on what action to take when the alarm is triggered.

Sound SectionUse the Browse button and navigate to the location of the sound fileand select the file. This allows you to provide a sound file with instruc-tions or a warning sound to the operator when the alarm is triggered.

Digital OutputNot active.

Test Dialog ButtonClick to test the alarm.

Figure 8-2.Alarm Action screen.

Page 50: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Analog Outputs | 9-1

ANALOG OUTPUTS

Process readings can be directed to devices capable of receiving analogoutput signals.

Setting analog outputs does not apply to Analog, Bar or Leak modes.

Setting the Analog Outputs

� Right click from a display window and then click Analog.

!NOTE

For Trend and Tabular modes, first select the channel as the source ofthe analog input.

DeviceIndicates to which output device to send the analog output signal.For outputdevices the system sends the reading to the module speci-fied for the output module for that device.

Figure 9-1.Properties screen -Analog tab.

Page 51: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

9-2 | System 200 Software

ModuleIndicates the output module to which the system will direct theanalog input.

Full ScaleUser-defined value as the top of the analog output range.

ZeroUser-defined value as the bottom of the analog output range.

Page 52: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Degassing a Quadrupole Head | 10-1

DEGASSING A QUADRUPOLE HEAD

Degassing a quadrupole head involves increasing the filament current toburn off any contaminants that may have collected on the filament.

How to Degas the Quadrupole Head

1. Select Control/Degas.

2. Click on the quadrupole head. The light bulb will turn yellow.

3. Type in the number of minutes for the system to degas in the Minutestext box. The limit is 10 minutes.

5. To start the degassing process, click the Start button. A dialog box mes-sage appears indicating that the system is degassing.

6. To cancel the degas operation, select Cancel.

While in the degas operation, the filament cannot be turned off.!NOTE

Figure 10-1.Control/Degas screen.

Page 53: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

10-2 | System 200 Software

This page intentionally left blank.

Page 54: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Automatic Calibration | 11-1

AUTOMATIC CALIBRATION

The Automatic Calibration function adjusts and optimizes RF frequency,mass resolution, and mass position. You can also check to see if the currenttune parameters are within tolerance. Ideally, you should use a calibrationgas with peaks at the low and high ends of the mass range. However, themass spectrometer can be calibrated using residual air.

Auto Calibration Dialog Box

Click on Control on the menu bar and select Auto Calibration from thedrop-down menu to bring up the Automatic Calibration dialog box.

Device List Box

Allows you to select the quadrupole head on which to perform theautomatic calibration.

Figure 11-1.Automatic Calibrationdialog box.

Page 55: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

11-2 | System 200 Software

Properties Button

Takes you to the Device Properties dialog box where you can view orchange properties for the device selected from the Device list box.

Mass Text Box

Defines the low and high mass of the calibration gases. The left text boxcorresponds to the low mass. Ideally a calibration gas as close as pos-sible to 1 should be used. For the high mass, use a calibration gas that isclose to the top of the operating range. For example, if operating up to100 AMU, use a calibration gas in the 90s AMU range.

WidthText Box

Allows you to set the resolution by specifying the width in AMU thateach peak should have for both the low and high mass calibration gas.Width is measured at 25% of the total peak height. If, for example, youset the peak width at .8, you are telling the system that at 25% peakheight it should be .8 AMU wide. Decreasing the width improves theresolution but lowers the sensitivity. To get better sensitivity, increasethe width. If the peak width is set too wide, you may start to lose theability to distinguish between adjacent peaks.

RF Tune Box

Displays the RF Tune value in the box to the right of the label. Thesquare to the left appears green if the value is within specifications. Itappears red if out of specification.

Resolution Box

The square to the left appears green if the value is within specificationsand red if out of specification. For example, if the width is set to .8, theindicator will appear green if the AMU width is sufficiently close to .8AMU at 25% peak height.

Page 56: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Automatic Calibration | 11-3

Position Box

Tells you whether selected masses (from high and low calibration gases)are properly positioned. In other words, you want to ensure that thepeak for 18 appears at 18 and not at 18.5 or 17.5. It allows you to shiftthe peaks left or right. The square to the left appears green if the valueis within specification or red if out of specification.

Check Button

Checks the RF Tune, resolution and position but does not attempt tochange them. The settings appear on the display window above thePosition and Width text boxes.

Start Button

Starts the automatic calibration, optimizing RF Tune, resolution andposition. When complete, the display windows should show data thatis properly positioned, resolved and with proper sensitivity.

Stop Button

Stops the calibration or calibration check process.

Cancel Button

Cancels the automatic calibration and discards all calibration changes.

OK Button

Exits the dialog box and saves all calibration changes made while in thethis dialog box.

Status Box

Indicates whether or not the system is scanning, the filament is warm-ing up or the system is finished doing the calibration or calibrationcheck.

Page 57: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

11-4 | System 200 Software

Performing an Automatic Calibration

Before performing an automatic calibration, check to make sure you havecorrectly entered the Cal Low Mass and Cal High Mass on the Tune Pa-rameters tab of the Device Properties screen.

� Click on the Edit button on the menu bar and select Device Propertiesfrom the drop-down menu.

� Select your device. The Device Properties screen appears.

� Select the Tune Parameters tab. Check to make sure you have enteredthe correct values for Cal Low Mass and Cal High Mass. Correct thevalues, if necessary.

To perform an automatic calibration, access the automatic calibration screenby clicking on the Control button on the menu bar and then on AutomaticCalibration from the drop-down menu. The Automatic Calibration screenappears.

Figure 11-3.Automatic Calibrationscreen.

Figure 11-2.Tune Parameters tabon Device Propertiesscreen.

Page 58: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Automatic Calibration | 11-5

� Select your device from the drop-down box.

� Adjust the mass and width values or use the defaults. Check to makesure that you have peaks present on your analog screen before con-tinuing.

� Click on the Start button.

� RF TuneTurns Yellow.Turns Green.

RF Tune Value appears in the box next to it.

� Resolution

Turns YellowPeaks appear in the Low and High boxes.Turns Green.

Figure 11-4.RF Tune turns greenand value appears inRF Tune box.

Figure 11-4.Resolution turns greenand peaks appear.

Page 59: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

11-6 | System 200 Software

� Position

Turns Yellow.Turns Green.

� Click on the OK button after automatic calibration to save all changesmade while in this dialog box.

Figure 11-5.Position turns greenand auto calibration iscomplete.

Page 60: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Automatic Calibration | 11-7

Resolving an RF Tune Error

Analyzer electronics must be RF-tuned to match the quadrupole analyzerhead. New analyzers are tuned at the factory. If you see the message “RFTune Failure” in the Devices & Log status box, it may be caused by one ofthe following.

• The electronics were removed from onehead and installed on another.

• Circuit boards were replaced during a repair.

• There is an electronics circuit failure.

• The analyzer head is damaged.

Setting the RF Tune

The RF Tune is set using the Auto Calibration dialog box in the System 2000software.

1. On the Control menu, click Auto Calibration.

If you have more than one device, make sure you have selected thecorrect device to initialize.

2. On the Automatic Calibration dialog box, click the Start button. Whenthe RF Tune box turns green, click the Stop button.

If the RF Tune box turns red, there is a circuit failure, or the analyzerhead is damaged. Pleas contact AMETEK for assistance.

3. Click OK to close the Auto Calibration dialog box.

4. On the Mode menu, click Analog. Right-click on the screen and clickon Display. In the Display mass range boxes, enter “1” for Low and “50for High. Allow the analyzer to warm up for approximately an hour.

5. At the end of an hour, observe the positions of the peaks on the analogscreen to determine how much coarse tuning is needed before you canperform an auto calibration. For example, if you are sampling air, lookat peaks at mass/charge 14 and 40.

!NOTE

!NOTE

Page 61: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

11-8 | System 200 Software

Click on the left mouse button and hold it down to move the cursoranywhere on the mass scale to see the location on the mass scale.

If the peaks of interest are off by more than 0.5 AMU, you must manuallyadjust the mass position. If the peaks are off by less than 0.5 AMU, go toStep 2 of “Coarse Adjustment of Mass Position.” For example, if the maxi-mum of peak 14 appears at 13.6 and peak 40 appears at 39.8, no adjustmentis necessary. However, if peak 14 reads 13.1 and peak 40 reads 39.03, youneed to perform coarse adjustment.

Coarse Adjustment of Mass Position

On the Edit menu, click Device Properties and then click the Tune Param-eters tab.

1. While looking at the scan, note the positions of the peaks to determinein which direction you need to adjust the mass position.

If the peaks shift to the right (higher)Decrease the numbers in the tune file at “low position” and/or “highposition.”

If the peaks shift to the left (lower)Increase the numbers in the tune file at “low position” and/or “highposition.”

Start out by changing the numbers by approximately 1 unit (e.g., from3.20 to 2.20, or from -1.50 to -2.50). After you change the numbers, clickOK and observe the next analog spectrum to see how the peaks haveshifted. If the resulting peak positions are acceptable, go to Step 2. Ifthe peak positions are still off by more than 0.5 AMU, repeat Step 1.

2. On the Control menu, click Auto Calibration. Make sure you have thecorrect device selected.

Click the Start button and observe the boxes for RF Tune, Resolutionand Position. When all three status lights have turned green, click OK.

Your analyzer should now be properly calibrated.

!NOTE

Page 62: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Working with Data Files | 12-1

WORKING WITH DATA FILES

Saving Data Files

Displayed information can be saved for later viewing or printing. Eachdata record within a data file contains scan data from the display window(s).Data records are created in serial format as the data becomes available fromthe system. For a Custom mode, for example, each display will be saved asa data record.

1. Select the display window or windows for which data will be saved.

• Hold down the shift key to select multiple display windows.

• Selecting a highlighted display window deselects the display win-dow.

• If no display window is selected, the system saves data for all dis-play windows in the mode.

2. Select File/Save Data or the icon on the toolbar. To activate this bar,select View/Data File Toolbar.

3. You are prompted with the Save As dialog box. You can specify data fileattributes.

4. Once the attributes are set, name the data file and place it in a folderusing standard Windows 95/98 techniques.

Figure 12-1.File/Save Data Menu -Save As dialog box.

Page 63: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

12-2 | System 200 Software

� Start Saving Radio ButtonPrompts the system to start saving data with the next scan and tocontinue saving data files.

� Save Now Radio ButtonPrompts the system to save the most recent scanned data and thenstop scanning. If there are multiple Trend modes within a Custommode setup, only one can be selected for which to save data.

� When File Full SectionChooses how to handle cases in which the data file becomes greaterthan 1 megabyte.

· Start New FileSave the data to a new file. The system automatically appendsconsecutively numbered digits at the end of each new file itcreates so it is easier to follow the order in which the files werecreated. Each file will be one megabyte in size so the data can becopied to disk.

· Stop Radio ButtonStops scanning when the file is full.

5. To stop saving data, select File/Stop Saving.

Page 64: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Working with Data Files | 12-3

Opening Data Files

1. Select File/Open Data or select the icon in the toolbar.

2. Navigate to the location of the data file. Click on the file.

3. Select Open. The data file appears on the screen. From here it can ei-ther be viewed or printed.

You can open System 1000 data files. The file is converted to theSystem 200 format and saved. If the new file was named �test.dat� itwill be saved as �new test.dat.�

!NOTE

Page 65: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

12-4 | System 200 Software

Moving Through Data Records

1. Select either File/Data File Position or use the data toolbar to move backand forth through the data file using the standard recording symbolssuch as rewind or fast forward.

2. To activate the data toolbar, select View/Data File Toolbar.

FORWARD

Go forward one record in the data file.

BACKWARD

Go back one record in the data file.

GO TO END

Go to the last data record of the data file.

GO TO START

Go to the first record in the data file.

Page 66: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Working with Data Files | 12-5

Converting Data File

Use this utility to convert System 2000 data files into ASCII text.

1. Select File/Convert Data File.

2. Select one or more data files.

3. Click Convert.

4. The file will be converted and saved with a CSV extension.

Page 67: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

12-6 | System 200 Software

This page intentionally left blank.

Page 68: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Printing Data | 13-1

PRINTING DATA

The Dycor System 2000 Software will print scanning information in a man-ner very similar to how it is displayed on the screen using the displayparameters created for that mode. You can view a preview of what will beprinted by selecting File/Print Preview.

In Custom Mode, the system prints the currently selected display win-dow. If you do not select a display window, the system prints informationfor each display window using the required number of pages.

1. Select the display window to be printed. If a window is not selected,the entire screen will be printed. To select multiple windows, hold downthe shift key while selecting.

2. Select File/Print or click on the icon.

3. Type a printer header comment. This information appears in the titleof the printed pages. Click OK.

4. If you selected File/Print, a standard Windows 95/98 Print Dialog Boxappears. Set the additional print information.

5. Click OK to print the scanning information.

Page 69: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

13-2 | System 200 Software

This page intentionally left blank.

Page 70: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Dynamic Data Exchange (DDE) | 14-1

DYNAMIC DATA EXCHANGE (DDE)

You can copy any of the columns from a Tabular display window to theclipboard.

Copying Magnitudes

1. Select Tabular mode.

2. Right click on the mouse and select Copy Magnitudes. This places theDDE data into the clipboard.

3. Use Edit/Paste Special (or Edit/Paste Link) in another Windows 95 ap-plication to create the DDE link and continuously update the data inreal time.

Copying Channels

1. Select Tabular mode.

2. Right click on the mouse and select Copy Channels. This places thename of the channel onto the clipboard.

3. Use Edit/Paste Special (or Edit/Paste Link) in another Windows 95 ap-plication to create the DDE link and continuously update the channelnames in real time.

If you are importing data into Microsoft Excel, go to the Edit/Prefer-ences menu option in the Dycor System 200 Software and select DDEItem Separator. From this menu select Return. This places the num-bers into separate columns. Select tab to have the number placed inrows.

!NOTE

Page 71: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

14-2 | System 200 Software

Direct DDE Control

The DDE service name is �Sys200.� The permanent topics available are:

• Command

• Status

• Other topics are created with the name of each Tabular display win-dow.

Sending a Command

1. Initiate a DDE conversation using the �Command� topic.

2. The following DDE Execute commands are available:

[OPEN filename]

Opens the Dycor System 2000 configuration file filename.

[FILAMENT(ON)]

[FILAMENT(OFF)]

[FILAMENT(ON, �device�)]

[FILAMENT(OFF, �device�)]

Sets the filament state. If the device isn�t specified, turns all filamentson or off.

[MULTIPLIER(ON)]

[MULTIPLIER(OFF)]

[MULTIPLIER(ON, �device�)]

[MULTIPLIER(OFF, �device�)]

Sets the multiplier state. if the device isn�t specified, turns all multipli-ers on or off.

Page 72: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Dynamic Data Exchange (DDE) | 14-3

[SCAN(ON)]

[SCAN(OFF)]

Enables scanning.

[DEGAS(ON, �device�)]

[DEGAS(OFF, �device�)]

Enables degas for a specified QuadLink device.

[SCHEDULE(SINGLE)]

[SCHEDULE(CONTINUOUS, interval)]

Sets to either single or continuous scan mode. If set to continuous, setthe interval in seconds.

3. The square brackets and quotes are required. Multiple commands canbe sent in the same DDE Execute command line.

Requesting Values in the Devices & Log Window.

1. Use the �Status� topic to request the values. The DDE items availableare:

ITEM RETURNS

Devices List of devices

Filaments List of ON or OFF state of the filaments

Multiplier List of ON or OFF state of the electron multipliers

Status List of device statuses

ScanState �Idle� or �Scanning�

Page 73: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

14-4 | System 200 Software

2. Use the name of the Tabular display window as the topic name to re-quest those values.

3. The DDE items available for a data window are:

ITEM RETURNS

Channels List of channels

Magnitudes List of current readings

Page 74: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Dynamic Data Exchange (DDE) | 14-5

!NOTE

Microsoft Excel 7.0 Example

Use the Edit/Preferences menu option in the Dycor System 200 Soft-ware to set the DDE Item separator to tab when using this type ofexample to create a DDE link with Microsoft Excel.

Sub Main ()

DDEcmd = Application.DDEInitiate(�Sys2000�, �Command�) �Initiate the DDE conversation.

Application.DDEExecute DDEcmd, �[open(��custom1.dyc��)]� �Open a configuration file.Application.DDEExecute DDEcmd, �[filament(on)][scan(on)]� �Turn on filament and start scan.

Application.DDETerminate DDEcmd �Wait 5 seconds for file to open

and take some data.Application.WaitTimeSerial(Hour(Now()), Minute (Now()), Second(Now()) + 5)

DDEdata = Application.DDEInitiate(�Sys2000�, �Tabular�) �Initiate the DDE conversation.

returnList = Application.DDERequest(DDEdata, �Channels�) �Get a list of channels.For i = LBound(returnList) To UBound(returnList)

Worksheets(�Sheet1�).Cells(i, 1).Value = returnList(i)

Next i

returnList = Application.DDERequest(DDEdata, �Magnitudes�) �Get the data

For i = LBound(returnList) To UBound(returnList) Worksheets(�Sheet1�).Cells(i, 2).Value = returnList(i)

Next i

Application.DDETerminate DDEdata

End Sub

Page 75: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

14-6 | System 200 Software

This page intentionally left blank.

Page 76: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Setting Security Levels | 15-1

SETTING SECURITY LEVELS

Allows you to set the operator privilege level. There are two levels: Super-visor and Operator. In Supervisor level, all security is off. When in Opera-tor, the user cannot save configuration or device parameters. In addition,there are optional restrictions that can be set in the Option menu.

Figure 15-1.Set Security Levels andOperator Optionsscreens.

To set Operator level:

1. Select Edit/Security Level.

2. Use the Options button to set Operator level privileges.

3. Use the Set Password button to set an optional password. The pass-word is required to go from Operator to Supervisor level.

4. Set the level to Operator.

Page 77: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

15-2 | System 200 Software

This page intentionally left blank.

Page 78: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

Error Messages | 16-1

!NOTE

ERROR MESSAGES

SYSTEM 200 ERROR MESSAGES

Here are all the error messages that can appear in the Devices & Log box.There are many other self-explanatory informational and warning mes-sages —these are only the errors.

“Quad1” is used as an example device. The actual device name will beshown in the Source column.

Source Message Comment

Quad1 Protect pressure exceeded The filament has been turned offbecause the pressure is too high.The filament protectpressure isset in the Protection tab of theDevice Properties dialog box.

Quad1 Filament open failure The filament has burned out andmust be replaced. The device hasbeen set to disabled in the DeviceProperties dialog box.

Quad1 Filament half open One half of the two-part filamenthas burned out. The system is stillusable, but the filament should bereplaced as soon as possible.

Quad1 RF tune failure The RF electronics frequency andthe quadrupole analyzer head aremismatched. Follow the proce-dure in the Auto Calibrationchapter to resolve this error.

Page 79: Dycor System 200 Software - irem senuhv.cheme.cmu.edu/manuals/dycorsoftware.pdf · Viewing Devices ... device hardware manual for information on hardware setup. 1. ... fault extension

16-2 | System 200 Software

Quad1 Filament tripped The filament has been turned offbecause the pressure is too highfor safe operation. The device hasbeen set to disabled in the DeviceProperties dialog box.

Quad1 Filament short failure The filament resistance is too low.The filament may be touchinganother ionizer element or anelectronics failure. The device hasbeen set to disabled in the DeviceProperties dialog box.

Quad1 No response The device is not responding atall to communications.

Quad1 Receive TIME-OUT error.

Quad1 Receive OVERRUN error

Quad1 Receive FRAMING error

Quad1 XMT CHECKSUM error

Quad1 CMD error

Quad1 ACK error

Quad1 Receive CHECKSUM error

Quad1 Transmit error

Quad1 Communications error These errors all indicate that mes-sages between the device and thePC are being corrupted.

For all communications problems, check the communications settings inthe Device Properties dialog box. Errors could also be caused by a badcable, RS-232 to RS-485 converter, a powered-off analyzer, or electronicsfailure.