ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an ECG Measurement System from the Columbia Integrated Systems Lab

Upload: darren-lindsey

Post on 05-Jan-2016

216 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

ECG Measurement System

By Nick Rose Under guidance of Dr. Jack Ou

Summer 2012

Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an ECG Measurement System from the Columbia Integrated Systems Lab

Page 2: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Full Circuit

R=10 M

Potentiometer

9 V

Gain= 200

Page 3: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Bandpass Filter Transfer Function

0.01 0.1 1 10 100 1000 10000

-35

-30

-25

-20

-15

-10

-5

0

5

Freq (Hz)

H(

measured) (dB)

Page 4: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Op-Amp

• Purpose: To bias voltage at output of AD624

-10 -8 -6 -4 -2 0 2 4 6 8 10

-8

-6

-4

-2

0

2

4

6

8

10

V3 (V)

V6 (V)

Page 5: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Opto-Coupler

Page 6: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

V(1) versus V(2) V(1) versus V(3)

-10 -8 -6 -4 -2 0 2 4 6 8 10

-10

-8

-6

-4

-2

0

2

4

6

8

10

V1 (V)

V3 (V)

V2 (V)

Page 7: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Instrumentation Amplifier

Gain of 200

Page 8: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Gain of 1000

-0.008 -0.006 -0.004 -0.002 0 0.002 0.004 0.006 0.008 0.01

-8

-6

-4

-2

0

2

4

6

8

10

Vin (V)

V9 (V

)

Page 9: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

V(3) of op-amp versus:2nd Harmonic Distortion 3rd Harmonic Distortion

-6.5 -6 -5.5 -5 -4.5 -4 -3.5 -3 -2.5 -2 -1.5

-70

-60

-50

-40

-30

-20

-10

0

V3 (V)

H2 dist (dB)

-6.5 -6 -5.5 -5 -4.5 -4 -3.5 -3 -2.5 -2 -1.5

-70

-60

-50

-40

-30

-20

-10

0

V3 (V)

H3 dist (dB)

Page 10: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Gain of 200, V(3) of 741 op-amp = -4.4 V

Adjusted V(3) to -4.4 V to minimize 2nd and 3rd harmonic distortion

Magnitude of input sine wave versus gain (gain includes gain of opto coupler)

0 0.002 0.004 0.006 0.008 0.01 0.012570

580

590

600

610

620

630

640

Input mag (V)

Gain

Page 11: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Power Dissipation

• The entire circuit (AD624, 741, optocoupler)..• Current draw: 114 mA• Voltage: dual power supply (+-9V)

Page 12: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded Versus Unshielded Cables

• Untwisted cable, shielded by aluminum foilAluminum shield is grounded

Page 13: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Unshielded AD624 Output

Page 14: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded AD624 Output

Page 15: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Unshielded ECG Output

Page 16: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded ECG Output

Page 17: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shield Input of ECG

Page 18: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Unshield Input of ECG

Page 19: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded AD624 Output at DC

Page 20: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded AD624 Output (0 -12.5 Hz)

Page 21: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded AD624 Output at 60 Hz

Page 22: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Unshielded AD624 Output at 60 Hz

Page 23: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Unshielded ECG Output

Page 24: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Shielded ECG Output

Page 25: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Extra Components for stabilizing the Output Voltage of the AD624

• 10 Mohm resistors were used at pin 1 and 2 of AD624 in order to provide bias current transistors in the pre-amplifier stage

• An input offset potentiometer with the middle pin attached to +9 V was used to reduce offset voltage. The voltage at pin 9 was used.

Page 26: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Different Design Iterations

• Breadboard• PCB• Equipment

Page 27: ECG Measurement System By Nick Rose Under guidance of Dr. Jack Ou Summer 2012 Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an

Future Work: Active Grounding Technique

• Problems observed:– Common mode sensing resistors attached to node

16 and 3 of AD624 can affect of node 9 of AD624.• 25 Kohms, DC voltage at node 9 is disturbed. • 1 Mohm, DC voltage at node 9 is stabilized.

– Large values of Rf makes it difficult to for negative feedback to pin node 2 of 741 op-amp to ground.

– What is the right ratio of Rf/Ra?