ECG Measurement System
By Nick Rose Under guidance of Dr. Jack Ou
Summer 2012
Built off of Chia-Hung Chen, Shi-Gun Pan, and Peter Kinget’s design for an ECG Measurement System from the Columbia Integrated Systems Lab
Full Circuit
R=10 M
Potentiometer
9 V
Gain= 200
Bandpass Filter Transfer Function
0.01 0.1 1 10 100 1000 10000
-35
-30
-25
-20
-15
-10
-5
0
5
Freq (Hz)
H(
measured) (dB)
Op-Amp
• Purpose: To bias voltage at output of AD624
-10 -8 -6 -4 -2 0 2 4 6 8 10
-8
-6
-4
-2
0
2
4
6
8
10
V3 (V)
V6 (V)
Opto-Coupler
V(1) versus V(2) V(1) versus V(3)
-10 -8 -6 -4 -2 0 2 4 6 8 10
-10
-8
-6
-4
-2
0
2
4
6
8
10
V1 (V)
V3 (V)
V2 (V)
Instrumentation Amplifier
Gain of 200
Gain of 1000
-0.008 -0.006 -0.004 -0.002 0 0.002 0.004 0.006 0.008 0.01
-8
-6
-4
-2
0
2
4
6
8
10
Vin (V)
V9 (V
)
V(3) of op-amp versus:2nd Harmonic Distortion 3rd Harmonic Distortion
-6.5 -6 -5.5 -5 -4.5 -4 -3.5 -3 -2.5 -2 -1.5
-70
-60
-50
-40
-30
-20
-10
0
V3 (V)
H2 dist (dB)
-6.5 -6 -5.5 -5 -4.5 -4 -3.5 -3 -2.5 -2 -1.5
-70
-60
-50
-40
-30
-20
-10
0
V3 (V)
H3 dist (dB)
Gain of 200, V(3) of 741 op-amp = -4.4 V
Adjusted V(3) to -4.4 V to minimize 2nd and 3rd harmonic distortion
Magnitude of input sine wave versus gain (gain includes gain of opto coupler)
0 0.002 0.004 0.006 0.008 0.01 0.012570
580
590
600
610
620
630
640
Input mag (V)
Gain
Power Dissipation
• The entire circuit (AD624, 741, optocoupler)..• Current draw: 114 mA• Voltage: dual power supply (+-9V)
Shielded Versus Unshielded Cables
• Untwisted cable, shielded by aluminum foilAluminum shield is grounded
Unshielded AD624 Output
Shielded AD624 Output
Unshielded ECG Output
Shielded ECG Output
Shield Input of ECG
Unshield Input of ECG
Shielded AD624 Output at DC
Shielded AD624 Output (0 -12.5 Hz)
Shielded AD624 Output at 60 Hz
Unshielded AD624 Output at 60 Hz
Unshielded ECG Output
Shielded ECG Output
Extra Components for stabilizing the Output Voltage of the AD624
• 10 Mohm resistors were used at pin 1 and 2 of AD624 in order to provide bias current transistors in the pre-amplifier stage
• An input offset potentiometer with the middle pin attached to +9 V was used to reduce offset voltage. The voltage at pin 9 was used.
Different Design Iterations
• Breadboard• PCB• Equipment
Future Work: Active Grounding Technique
• Problems observed:– Common mode sensing resistors attached to node
16 and 3 of AD624 can affect of node 9 of AD624.• 25 Kohms, DC voltage at node 9 is disturbed. • 1 Mohm, DC voltage at node 9 is stabilized.
– Large values of Rf makes it difficult to for negative feedback to pin node 2 of 741 op-amp to ground.
– What is the right ratio of Rf/Ra?