echelle grating spectrometers in analytical spectrometry

8

Click here to load reader

Upload: charles-c

Post on 06-Aug-2016

222 views

Category:

Documents


5 download

TRANSCRIPT

Page 1: Echelle grating spectrometers in analytical spectrometry

Peter N. Keliher and Charles C. Wohlers‘ Chemistry Department. Villanova University, Villanova, Pa. 19085

Echelle Grating Spectra ...,.,. in Analytical Spectrometry

Almost since its inception nearly 100 years ago, one of the major prob- lems facing analytical spectrometry has been the limitation of spectrome- ter resolution and dispersion on ana- lytical results. This problem was quite important to spectroscopists decades ago when they tried to perform such analyses as boron in steel and were se- verely limited by spectral interference, or when they were required to do such things as determine single isotopes of uranium. Analyses like these required very high dispersion and resolution; thus, these qualities often came to be the most important factors when de- signing spectrometers. For this reason, gratings replaced prisms, and grating spectrometers themselves began to be made with ever longer focal lengths for higher linear dispersion and with more finelyruled gratings for higher resolution. These solutions began to become exhausted in the 1940’s when focal lengths grew so long that light losses became too great even for such bright sources as the DC arc, and it became too difficult to rule gratings with finer and finer groove spacings and still keep efficiency reasonably high and ghost intensities reasonably low.

In the late 1940’9, Harrison and his group a t MIT decided to take a com- pletely different approach in design- ing a high resolution and dispersion spectrometer, and in so doing devel- oped the echelle grating (I, 2). They did this by applying the same hasic formulas for diffraction gratings that others had, hut instead came up with a different conclusion. The hasic for- mula for resolution, assuming that the angle between the incident and dif- fracted rays is small, is

-= h 2 N d . ~ i n f l = ~ ~ Ah h

where N is the number of grooves in the grating, d is the groove spacing, fl

Present address, Jarrell-Ash Division, 590 Lincoln Street, Waltharn, Mass. 02154.

Figure 1. Echelle grating: W, grating width: d, tion; i. angle of incidence

groove spacing; r, angle of reflec-

is the angle between the diffracted ray and the grating normal, and m is the

quite similar to a normal blazed plane grating except that the “short side”of

order in which-the grating is used. In- stead of using high groove densities (i.e., large nnmher of grooves) to achieve high resolution, the echelle grating increases the blaze angle (and the order) to achieve very high resolu- tion.

The equation for dispersion, again assuming that the angle between the incident and diffracted rays is small, is

dl 2 f . t a n f l rnf d h h d . cos f l

where f is the focal length of the spec- trometer. Here, the echelle grating spectrometer gives high dispersion again by using a high blaze angle and high orders instead of long focal lengths.

ure 1. The echelle grating appears

- = = _ _

The echelle grating is shown in Fig-

the grooves is used, i.e., echelle grat- ings are designed to be used a t blaze angles greater than 45”. In this way, a spectrometer using an echelle grating gives high dispersion without a very long focal length and high resolution without extremely fine groove spac- ings. The effect of this high blaze angle and use of high orders is shown in Figures 2 and 3 which compare the blaze of conventional and echelle spec trometers.

Some advantages of the echelle grating are indicated in Table I which compares two “typical” conventional and echelle spectrometers of 0.5-m focal length. As may be seen, the ech- elle spectrometer has better than one order of magnitude higher resolution and dispersion with no sacrifice of op- tical speed.

Table I also emphasizes an impor- tant property of echelle gratings, the

ANALYTICAL CHEMISTRY, VOL. 48. NO. 3, MARCH 1976 * 333A

Page 2: Echelle grating spectrometers in analytical spectrometry

Figure 2. Typical spectrometer blazed at 3000 A; represents grating blaze function

Figure 3. Echelle spectrometer blazed at all wavelengths; represents grating blaze function. Stacked orders fall with- in optimum blaze region

use of multiple high orders. These or- ders must be separated from each other in some way, which is not gener- ally necessary in conventional spec- trometers. This is accomplished by placing an auxiliary dispersing ele- ment, usually a prism or low disper- sion grating in the spectrometer so tha t it disperses wavelengths a t right angles to the echelle and thus effec- tively separates the orders. Typical echelle spectra are shown in Figures 4 and 5 . In the photograph of the 150-W xenon lamp continuum (Figure 41, one can see the different orders appearing as horizontal lines, with the lowest order (longest wavelength) appearing a t the bottom and the highest order (shortest wavelength) appearing a t the top. Here, the prism-cross dispersion element is, separating the wavelengths in the vertical direction, while the ech- elle grating separates them horizontal- ly. The free spectral range (FSR- wavelength range best covered by one order) varies from 1.8 nm at 200 nm to 11.1 nm at 500 nm for this grating. This type of two-dimensional pattern can be used in either a spectrographic mode or a phlotoelectric mode. When an echelle spectrometer is used as a monochrc;’mator in the photoelectric

Table 1. Comparison of Typical Conventional and Echelle Spectrometers

Conventional Echelle

Focal length Groove density Angle of diffraction Width, W Order (m) a t 300 nm Resolution at 300 nm Linear dispersion a t 300 nm Reciprocal linear dispersion a t 300 nm f-Num ber

0.5 M 1200 grooveslmm 10” 22’ 52 mm 1 62,400 0.61 mrnlnm 16 A/mm fj9.8

0.5 M 79 grooveslmm 63” 26’ 128 mm 75 763,000 6.65 rnrnlnm 1.5 &mm fl8.8

mode, two wavelength dials are re- quired, one to change orders and the other to select the wavelength within an order. Otherwise, the operation and appearance of an echelle monochro- mator are the same as any convention- al monochromator. Early models of the only commercially available ech- elle monochromator (Spectrametrics. Inc.. Andover, Mass.) used arbitrary coordinates for the vertical (order) and horizontal (wavelength) dials. Wavelengths were obtained by using “ Wavelength-Arbitrary Coordinate” tables provided by the manufacturer (3). For example, the Cu 324.7-nm line in the 79th order was “dialed in” by adjusting the vertical counter to 50148 and the horizontal counter to 50587. Newer systems manufactured by Spectrametrics have a more conve- nient direct wavelength control. As with normal monochromators, some “peaking” may be necessary after the wavelength has been set. Sometimes a spectral line will appear in more than one order, but there will be a most de- sired order with respect to relative sig- nal intensity. Several examples of this are given in Table 11.

The two-dimensional pattern of wavelengths found with an echelle monochromator can, however, be a disadvantage in some cases. There is a limitation on slit height (the maxi- mum slit height is generally 1 mm or less) so tha t interference between or- ders is avoided. This can affect the lu- minosity (or light throughput) when photoelectric readout is used. The lu- minosity of a monochromator (for photoelectric readout) is given by

BAtshWH * COS 13 f‘ a =

where Bh is the brightness of the source directed toward the entrance slit in the wavelength interval selected by the monochromator, t is the effi- ciency of the optics, s is the slit width, h is the slit height, W is the grating width, and H is the grating height. This equation assumes that the spec- tral line width is smaller than the spectral slit width used, and that the

slit width is sufficiently larger than the critical slit width so that diffrac- tion effects are negligible.

Therefore, if one wants t o compare the luminosity of echelle and conven- tional spectrometers having similar focal lengths, important factors to consider are the angle of diffraction 6, the slit height h , and the grating effi- ciency. If the echelle grating is physi- cally somewhat larger than a conven- tional grating, as is often t he case, this will make u p for any loss of luminosity due to the higher angle of diffraction. This leaves the slit height as the prin- cipal limitation on luminosity in an echelle spectrometer when the unit is used in the photoelectric mode. How- ever, balanced against this is the fa- vorable grating efficiency of the ech- elle as discussed earlier. For a given spectral resolution, a high resolution echelle spectrometer can have much wider slits than a medium resolution conventional spectrometer, thus in- creasing the relative luminosity of the echelle instrument.

Historical Background

As mentioned previously, the ech- elle grating was first developed by Harrison and coworkers in 1949 ( I , 2) to provide a high resolution and dis- persion spectrometer in a small, high luminosity package. Although there was some early analytical interest ( 4 , 5 ) , this did not continue, probably due to the difficulties of ruling high angle of diffraction, ghost-free, Igratings and also due to the difficulties of using photographic detection with short slits. This first problem was overcome in time by Harrison and his group a t MIT, primaril) through the develop- ment of the interferometrically con- trolled ruling of gratings (6-8), which was designed specifically for ruling echelle gratings. In spite clf these im- provements, echelle gratings found lit- tle use in analytical spectrometry and instead found their first widespread use in astronomy (9-12), <as is often the case with new developments in spectrometry.

In 1969 a commercial echelle spec-

3 3 4 A * ANAL.YTICAL CHEMISTRY, VOL. 48, NO. 3, MARCH 1976

Page 3: Echelle grating spectrometers in analytical spectrometry

lPANIII

SPECTRASPAN 111 is the latest in a line of field proven plasma emission spectrometers from SMI. The most up-to-date technology available has been used in making SPECTRASPAN 111 a high per- formance, high reliability, easily operated system for trace element analysis. The high temperature plasma and high resolution spectrometer which have provided interference free low-level analyses in the past have been further refined for still higher performance. Multi-element capability has been increased to 20 simultaneous channels, and the number of 20 channel programs which may be pre-aligned is unlimited. Easy conversion between qualitative and quantitative operation, sequential and multi-element modes provides unmatched versatility. Modular design allows growth to meet future requirements and guards against obsolescence. Operation has been greatly simplified and produc- tivity increased through the addition of a built-in microcomputer which performs all calibrations,

SPECTRAMETRICS, I NCORPORAl 204 ANDOVER ST, ANDOVER, MA 01810 -

Computer Aided Plasma Spectrometry for Trace Element Measurements.

corrections, and conversions. Operational monitors and diagnostic routines insure reliable data.

SPECTRASPAN 111 - features a high dispersion Echelle Grating spectrometer and a high tempera- ture argon plasma in a compact package.

SPECTRASPAN 111 - Equals or exceeds A. A. performance on routine metals and measures refractories and some non metals - on a routine basis. All with no lamps, no explosive gases, minimal sample treatment and simple operation. SPECTRASPAN 111 - Is an up-to-date system for today’s analytical requirements - at a price you can afford.

TED (617) 475-7015

CIRCLE 201 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 48, NO. 3, MARCH 1976 335A

Page 4: Echelle grating spectrometers in analytical spectrometry

Figure 4. Ec:helle spectrogram of 150-W xenon arc lamp. Note distinct orders

Figure 5. 1,000 ppm iron solution aspirated into premixed ni- trous oxide-acetylene flame. Note sodium impurity in solu- tion; 589.0-589.6-nm doublet, lower right of photograph

L- I

trometer was rntroduced by Spectra- metrics. This :system was referred to as a “SpectraSpan” and was described in some detail by Elliott (13) and later by Matz (14 ) . By the simple adjust- ment of a mirror. the SpectraSpan could be immediately converted from the spectrographic to the photoelec- tric (monochromator) mode. Since 1970 a number of reports have ap- peared concerning the application of the Spectrametrics system in analyti- cal spectrometry (15-22).

Applications

There are four major areas in which echelle spectrometers might be useful in analytical spectrometry: providing lower detection limits and greater se- lectivity in emission spectrometry through the decrease in spectral back- ground; use of a continuum source in atomic absorption spectrometry; mea- surement of spectral line profiles; and multielement analysis systems. These areas are considered in detail below.

Emission Spectrometry

The various factors influencing de- tection limits in emission spectrome- try have been covered theoretically by Winefordner and Vickers (23) and also by Laqua et al. (241, who specifi- cally considered the effect of mono- chromator resolution. The latter workers concluded that. for best de- tection limits, one should use a very high resolution monochromator and have a slit width equal to 1.5 times the critical slit width, assuming tha t source noise is the limiting factor. In emission spectrometry the two major sources of noisie which limit the sensi- tivity are from the source (flame flick- er, etc.) and from the photomultiplier tube (PMT) detector (dark current). Since P M T da.rk current noise will re- main constant, i t is important to de- crease the ‘source noise from the back- ground by decreasing the spectral

Table I I . Relative Intensities of Spectral Lines in Different Ordersa

Rei Line, nm Order intensity

Mercury, 253.7 99 12 100 34 ioi loo 102 40

Mercury, 434.8 57 13 58 100 59 82

Sodium. 489.0 42 5 1 _ - 43 100 44 27

a See refs. 13-1 7 and 21 for descrip- tion of instrumental system used.

bandpass until P M T noise predomi- nates or until the spectral line width is reached. Therefore, it is an advantage to have a high resolution, high lumi- nosity monochromator so that one can easily isolate a spectral line profile from its background without loss o f light throughput (compared to a medi- um resolution system) and consequent decrease of the source noise relative to the P M T noise.

Since it is relatively easy to design an echelle monochromator to have ten times the resolution and dispersion of conventional medium resolution mo- nochromators of the type used in atomic absorption, significant im- provements in detection limits should be possible, particularly when high background “atom reservoirs” such as the nitrous oxide-acetylene flame or various “plasma”-type systems are used. Also. spectral interferences, such as CaOH band emission on barium 553.5-nm atomic emission, or samar- ium 492.41 nm on neodymium 492.43-nm atomic emission, are elimi-

nated when a high resolution echelle monochromator is used (17). This im- portant practical property of an ech- elle spectrometer (increase in speci- ficity) should be of great importance in emission spectrometry as high tem- perature and high background plas- mas (DC, microwave, or R.F) become more and more commonly used.

Atomic Absorption

Another application of echelle grat- ings is in the area of atomic absorption using a continuum source (AAC). This use of a continuum source (xenon lamp) in atomic absorption would be an advantage in that one would need only one source for all elements, in- stead of a different spectral source (hollow cathode lamps or electrodeless discharge lamps) for each element. This would, of course, save consider- able time and expense when a number of different elements are to be ana- lyzed. Also, it would make background correction and multielement analysis in atomic absorption considerably eas- ier. It is difficult to use a conventional medium resolution monochromator in AAC since it cannot completely isolate the spectral line from the continuum background and thus will have a very low signal-to-background ratio. How- ever, with an echelle monochromator, the spectral line can be isolated and sensitive results quite similar to those using spectral sources may be ob- tained. This has been shown by the authors (19. 22) for two versions of a modified commercially available ech- elle monochromator; some typical re- sults are shown in Table 111. Here, sensitivities (defined as the concentra- tion necessary to give 1% absorption of the incident light, i.e., an absorbance of 0.0044) are compared for the AAC system with values from hollow, cath- ode lamp (HCL,) sources both from published tables (25) and also from experimental values obtained on the

3 3 6 A ANALYTICAL CHEMISTRY, VOL. 48, NO. 3, MARCH 1976

Page 5: Echelle grating spectrometers in analytical spectrometry
Page 6: Echelle grating spectrometers in analytical spectrometry

Table 111. Sensitivities Using Continuum and Line Sources

Sensitivity, AALa Sensitivity, AACa

Element and line, nm Ref. 25 Refs. 19, 22b Refs. 19, 22b Silver, 328.1 0.06 0.2 0.15 Ai u m in urn I 309.2 1 .o 2.7 4.1 Calcium, 42;I.J 0.08 0.15 0.20 Chrom i u in, 3 5 7.9 0.1 0.3 5 0.30 Copper, 324.7 0.09 0.065 0.20 Iron, 248.3 0.12 0.15 0.3 Potassium, 766.5 0.04 0.035 0.06 Magnesium, ,285.2 0.007 0.016 0.020 Sodium, 589.0 0.015 0.045 0.025 Lead, 283.3 0.5 1.1 1.5 Vanadiur1,318.5~ 1.7 2.7 13.0

dent l ight (0 0044 absorbance). All values are in ppm. Air-acetylene flame used for all elements ex iep t aluminum and vanadium where nitrous oxide-acetylene flame used. b Villanova results. C Multiplet, only one line o f which i s used in AAC.

QSensitivity i s defined as the concentration necessary to give 1% absorption o f the inci-

same echelle system using wide slits. I t was found possible to determine all elements having resonance lines at wavelengths greater than 225 nm, which includes the great majority of elements riorrnally determined by atomic absorption.

for AAC aire somewhat worse than those for AAL, (atomic absorption using a spectral line source) for most elements, but not to any great extent. The AAC ,results could be improved by using an echelle monochromator specifical1:y designed for AAC. Detec- tion limits, which compared less favor- ably for AAC than did sensitivities be- cause of higher source noise, could be improved by using a higher luminosity monochromator and/or a more intense and stable lamp; this would also ex- tend the lower wavelength limit below 225 nm. Very recently, Zander and co- workers (2’6) have described an ech- elle-AAC system with built in wave- length modulation. Wavelength mod- ulation for AAC has the significant ad- vantage of suppressing most source noise (27) The result is improved sig- nal-to-noise rstios and detection lim- its. Detection limits in the echelle- AAC-wavelength modulation mode are three times better than in normal echelle-AAC. Working curves are lin- ear over about three orders of magni- tude of conceiitration.

Spectral Line Profiles

The scanning of spectral line pro- files does not have a direct application to analytical work, but knowledge of spectral line widths and profiles can provide clues as to means of improv- ing analytical methods and also gives a better understanding of the physical processes taking place. The instru- ment normally used in scanning spec- tral line profiles is the Fabry-Perot in- terferometer. Although i t has been

The sensitivities shown in Table 111

used for many years for this purpose and although much useful data have been obtained with it, i t still has a number of disadvantages. First, to ob- tain high resolution on a Fabry-Perot interferometer, one needs mirrors hav- ing high reflectance dielectric coat- ings. However, these are not easily available for wavelengths below 300 nm. These coatings are also only us- able over a relatively narrow wave- length region of a few hundred ang- stroms; therefore, if one wishes to look a t widely separated regions of the spectrum, the mirrors must be changed. The Fabry-Perot is also com- plicated to operate, and it is frequent- ly very difficult to achieve good stabil- ity. The echelle monochromator pro- vides an alternative system for the measurement of spectral line profiles. As early as 1957, Walsh and coworkers (28) used an echelle monochromator for the direct measurement of the cad- mium 228.8-nm line from a vapor dis- charge lamp to illustrate the unsuita- bility of tha t device in AAL. Self-re- versa1 of the cadmium line was clearly observed.

The echelle monochromator has been used by Stroke and coworkers to study isotopic shift and hyperfine structure of short-lived radioactive isotopes (29-35). This is one applica- tion where a high resolution, high lu- minosity monochromator is obviously important. The authors have used an echelle monochromator to obtain qualitative and semiquantitative re- sults in scanning line profiles (1 7,20). Unfortunately, an echelle system can- not provide the extreme high resolu- tion of which a good interferometer is theoretically capable, and it was not possible to obtain the line profile di- rectly without instrumental broaden- ing, as is also the case with most Fabry-Perots. Attempts to correct for this broadening were unsuccessful.

The authors were able to demonstrate, however, the dependence of atomic absorption response on HCL line width (20). A particular aluminum HCL went through a maximum in the line width as the current was in- creased, and this exactly corresponded to the same current, 15 mA, where a minimum in the atomic absorption re- sponse was observed. Scans of spectral line profiles have also been made ( I 7 ) showing the onset of self-absorption and self-reversal in lines emitted by microwave-powered electrodeless dis- charge lamps. Further progress in the determination of spectral h e profiles using an echelle monochromator would probably require an instrument of higher resolution than that used by the authors.

Multielement Analysis

The last, but perhaps the potential- ly most important application of ech- elle gratings, is as the dispersing de- vice in a multielement spectrometer. TV-type image detectors are becom- ing popular in multielement applica- tions (36-41), and the two-dimension- al nature of these systems lends itself to adaptation with the echelle grating. In fact, what is most important when an image detector system is used is the two-dimensional nature of the echelle pattern, rather than high resolution and dispersion. When an image detec- tor (of the ones currently available) is used with an echelle, some sacrifice in resolution and dispersion is necessary. Use of a high dispersion echelle spec- trometer with such a system would re- quire an image tube detector much larger and with higher resolution than those currently available (39, 40). Current vidicons are generally 25 mm in diameter and have on the order of 105 resolving elements. A square ech- elle pattern 17 mm on a side would fit into this 25-mm circle, and it could be generated covering a wavelength re- gion of 200-600 nm using a 50 X 100- mm, 79 groove/mm echelle grating (angle of diffraction 63’ 26’) in a spec- trometer having a focal length of 160 mm. This spectrometer would have a reciprocal linear dispersion of ap- proximately 3.0 A/mm a t :200 nm and 8.5 A/mm a t 500 nm. Since the resolu- tion capabilities of the vidicon would be approximately 20 line pairs/mm, this means that each tube element would be looking a t about a 0.25 8, window, depending upon wavelength. This type of resolution, however, should be adequate for many applica- tions.

The echelle-image tube combina- tion in analytical spectrometry was first proposed by Margoshes in 1970 (42) for use with a vidicon detector. Some characteristics of the Margoshes

3 3 8 A ANALYTICAL CHEMISTRY, VOL. 48, NO. 3, MARCH 1976

Page 7: Echelle grating spectrometers in analytical spectrometry

Chemistry and Biochemistry of Thiocyanic Acid and its Derivatives edited by A. A. NEWMAN

CONTENTS: M . N. Hughes, General Chemistry. J . L. Burnieister, Coordination Chemistry. D. H . Kerridge. Molten Thiocyanates. J . L. Wood. Biochemistry. H . A . Bec4huis. Technology and Industrial Applications. M . R . F. Ashworth. Analytical Chemistry: 1 . Thiocyanate Ion and Thiocyanic Acid. 2. Organic Thiocyanates, R - S - ('3 ,V. 3. Organic Isothiocyanates. R - N = C = S. 197.5. 366 pp . , $31 .00/EI 2.00

Chemical Analysis of Organometallic Compounds Volume 4

by T. R. CROMPTON Yuniber 4 in The Analysis of Organic Materials/An Interna- tional Series of Monographs/series editors: R. Belcher and D. M. W. Anderson

This is the fourth volume i n this critical review of the organometallic compounds of all the elements, com- pounds which are attracting increasing interest all over the world. There is a main chapter on the analysis of or- ganophosphorus compounds. and minor chapters on or- ganoarsenic. organoantiniony and organobismuth com- pounds. Each chapter provides a complete review of all aspects of the analysis of the organocompounds in ques- tion, together wi th critical comments on the relative im- portance. merits and demerits of the various methods. In particular. the section on organophosphorus compounds contains much information on organophosphorus insecti- cides and pesticides. including a detailed discussion on the application of gas chromatography to their analysis. 197.5. 292 pp.. S24.2.5/€9.40

Determination of Organic Structures by Physical Methods Volume 6

editcd by F. C. NACHOD. J. J . ZUCKERMAN, and EDWARD W. RANDALL

CONTENTS: P. J . Arpino and F. W. McLufferry, Amino Acid Sequencing of Oligopeptides by Mass Spec- trometry. G. M . P e q x a and F. W. McLafferty, Comput- erized Structure Retrieval and Interpretation of Mass Spectra. R. L. Strong. Flash Photolysis and Structure. J . Schraml and J . M. Bellama. 29Si Nuclear Magnetic Resonance. J . K . Saunders and J . W. Euston. The Nu- clear Overhauser Effect. L. Lunazzi. Molecular Struc- tures by NMR in Liquid Crystals. 1976. about 500 pp. , in preparation

N.B.: Postage plus 50c handling charge on all orders not accompanied by payment. Prices are subject io rhange without notice.

ACADEMIC PRESS A Subsidiary of Harcourt Brace Jouanooich. Publishers I l l FIFTH AVENUE:, NEW YORK, N . Y . 10003 24-28 OVAL ROAD, LONDON N W I 7DX

CIRCLE 2 ON READER SERVICE CARD

Highest quality, fully fused cells for UV, visible and infrared in lightpaths (LP) of 2 to 100 mm . . . including flow cells, jacketed cells, and cell washers.

W e 1 Type 9 Type 17 or 18 Standard Semi-Micro Micro

A wide spectrum to choose from. All cells matched at no extra charge.

Immediate Delivery - P h o n e To d a y (51 6)93 5- 8 774

Internal Type-LP Material Width Height Price

1Q-lOmm Quartz 12.5mm 45mm $35.00 1G-lOmm Glass 12.5mm 45mm 12.00 9Q-lOmm Quartz 4.0mm 45mm 36.00 9G-lOmm Glass 4.0mm 45mm 19.00

29Q-1Omm Quartz 4.0mm 48mm* 44.00 29G-10mm Glass 4.0mm 48mm* 22.00 17Q-1 Omm Quartz 2.0mm 25mm 34.00 18Q-1Omm Quartz 2.0mm 45mm 39.00

*Teflon Stoppered cell for high temperature measurements. Q is the highest Quality cell in Quartz. Standard quartz (H) available at lower cost.

Free Cell Rack with each order of 4 or more quartz cuvettes; 8 or more glass cuvettes.

e Savant Instruments, Inc.

221 Park Avenue, Hicksville, N.Y. 1 1801

Write for complete 1976 catalog:

Savant Cells Division CIRCLE 200 ON READER SERVICE CARD

ANALYTICAL CHEMISTRY, VOL. 48, NO. 3, MARCH 1976 3 3 9 A

Page 8: Echelle grating spectrometers in analytical spectrometry

Table IV. Characteristics of Echelle-Vidicon System Proposed by Margosheso

order Central

wavelength, nrn

134 175.2 117 200.7 78 301 .O 59 397.9 39 602.0 29 809.5

a Taken from ref. 42.

Reciprocal linear dispersion, A h r n Resolution, A

2.45 0.080 2.8 1 0.094 4.21 0.14 5.57 0.19 8.43 0.28

11.3 0.38

system are shown in Table IV. The table gives. for several orders, the cen- tral wavelength, the dispersion, the resolution (controlled by the resolu- tion of the vidicon). and the equiva- lent slit width that would have to be used to achieve this resolution on a Conventional spectrometer with 5 &mm dispersion. The dispersion and resolution are particularly good in the ultraviolet region of the spectrum. This is, of course, an advantage since the spectra of atomic lines are general- ly more complex a t shorter wave- lengths. A working instrument was first constructed by Danielsson and coworkers (43, 44) in Sweden. An image dissector tube was used. This type of image detector is basically a photomult,iplier tube in which the photocathode is scanned so that only a small portion is chosen for amplifica- tion by the dynode chain. This detec- tor has the advantage of lower noise and higher rewlution than the vidicon tube. but the readout is inherently se- quential rather than having true si- multaneous multiple-channel detec- tion. The system built by Danielsson and coworkers covers a spectral range of 200-400 nni with a resolution of 0.023 8, a t 200 nm and 0.045 A a t 400 nm. The focal length is 220 nim, and the display is intercepted by a 20-mm diameter image dissector.

if'ood and coworkers ( 4 5 ) have re- cently designed an echelle-image tube combination, but their system uses an SEC vidicon rather than an image dis- sector. The vidicon is preceded by an image converi:er to overcome the lack of response of' the vidicon below ca. 300 nm. The instrument described by iVood and his, colleagues has a resolu- tion of 0.7 280 nm and is able to cover a wave- length range of 230-860 nm. The focal length of the system is 600 mm, and the vidicon used has a diameter of 40 mm. A1thoug.h the Wood spectrometer was designed for the analysis of atom- ic emission spectra, the versatility of the system has suggested some appli- cations in molecular spectrometry. Johnson has very recently reported ( 4 6 ) on the use of the Wood echelle-

a t 810 nm and 0.19 8, a t

Equiv sl i t width, prn

16 19 28 38 56 76

vidicon system for molecular absorp- tion and fluorescence spectrometry. A special high intensity mercury line source was used to generate molecular fluorescence spectra.

An alternative device for multiele- ment spectrometry is the PMT. The PMT is the most commonly used de- tector in analytical optical spectroine- try, both in single-channel systems such as conventional monochroma1;ors and in polychromator direct reader systems. Elliott (13) has, in fact. used a single conventional P M T in conjunc- tion with a n optical encoding device for multielement echelle spectrome- try. A more desirable approach is the use of a P M T "bank", Le.. an array of PMT's arranged in a two-dimensional pattern ( 4 7 ) . Although the PMT's are fixed and not easily moved (as is true with a conventional direct reader), it is relatively easy to arrange certain lines to pass through the system via a deflecting mirror arrangement. This system is described in detail elsewhere (48 ) .

This paper has given a general idea of what advantages echelle gratings have over conventional gratings and in what areas of analytical spectrometry these advantages might best be ex- ploited. Although echelle gratings have been in existence now for over 25 years. it is apparent that their use in analytical spectrometry is only in its infancy, and that future years will see the echelle grating becoming more and more important in this area.

References

(1) G . R. Harrison, J . O p t SrJc.. Am., 39,

( 2 ) G. R. Harrison, J. E. Archer. and .J. 522 (1949).

Camus. ibid. , 42. 706 (1952). ( 3 ) Spectrametrics. Inc.. Andover. Mass.

manufacturer's data. (4) I). Richardson, Spectrochim. A c t a , 6,

61 (1953). (5) LV. G. Kirchgessner and N. A. Finkel-

stein, A n a ) . Chem., 25, 1 ( M (1953). (6) G. R. Harrison and G. IT. Stroke, J

Opt. .To(,. Am., 45, 118 (19%J. (71 G. LV. Stroke. i b id . , 51, 1321 (1961;l. ( 6 ) G. R. Harrison, A p p l . O p t , 12, 2039

(9) D. J. Schroeder. i b id . , 6, 1976 (1967). (10) R. Tousey. .J. D. Purcell. and D. L,.

(1973).

Garret t , ib id . , p 365.

(11) W. Liller, ibtd., 9, 2332 (1970). (12) D. J. Schroeder, Pub. As,!ron. SOC.

Pac., 82, 1253 (1970). (13) W. G. Elliott, Am Lab. , 2 ( 3 ) , 67

(1970). (14) G. J . Matz, ibid., 5 (3 ) . 75 (1973). (15) M. S. Cresser, P. N. Keliher, and C. C.

CVohlers, Spectrosc. Lett . , 21, 179 (1970). 116) F. L. Corcoran, Jr.. P. N. Keliher, and

C. C. Wohlers,Am. Lab., 4 l(3). 51 (1972). (17) hl. S. Cresser, P. N. Keliher, and C. C.

LVohlers. Ana/. Chem.. 35. 1.11 (1973). (18) S. S. Karacki and F. L.'Corcoran, hr.,

(19) P. N. Keliher and C. C . \Vohlers,

(20) P. N. Keliher and C. C. Wohlers.

Appl . Spectrosc., 27,41 (1973).

.4nai. Chem., 46,682 (1974).

Appl . Spectrosc., 29, 198 (1975). 121) M. S. Cresser, P. N. Keliher, and C. C .

iVohlers, Lab. I'ract.. 26, 335 (1975). ( 2 2 ) P . N. Keliher and C. C. \'Vohlers,

Anal. C'hem., 48, 140 (1976;i. (23) .J. D. LVinefordner and T. J . L'ickers,

rhid., 36, 1939 (1964). (241 K. Laqua, LV. D. Hagenah, and H.

LVaechter, Z. Ana/ . Chem. , 225, 142 (1967).

(251 "Analytical Methods for Atomic Ah- sorption Spectrophotometry", Perkin- Elmer Corp., Norwalk, Conn., 1973.

(26) A. T. Zander, T . C. O'Haver, and P. N. Keliher, paper presented a t the 27th Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy. Cleveland. Ohio, March 1, 1976.

( 2 7 ) LV. Snelleman, Spectrochin. Acta, Part R, 23,403 (1968).

128) B. .J. Russell. J . P . Sheltcm. and A. Walsh, Spectrcichim. Acta, 8,~317 (1957).

SOC. Am., 49,1088 (1959). (29) R. J . Hull and H. H. Stroke, J . Opt.

( 3 0 ) R. J . Hull and H. H. Stmke. ibid. . 51. 1203 (1961).

ley, and H. H. Stroke. ihid., 52,501 (1962).

1147 (1963).

(31) C. J . Schuler, M. Ciftan, L. C. Brad-

(32) R. .J. Hull and H. H. Stroke. ibid. , 53,

( 3 3 ) P. C . Maenante and H. H. Stroke. ihid. , 59, 838 ( 1 969).

(34) R. Chuckrow and H. H . Stroke, ibid.,

(35) 0. Redi and H. H. Stroke, ibid., 65, 1

( 3 6 ) D. G. Mitchell. K. LV, Jackson. and K.

61,218 (1971).

(1975).

M. Aldous. Anal C h e m . , 45, 1215A 11973).

( 3 7 ) K. IV. .Jackson, K. M. Aldous, and D. G. Mitchell. i l p p i . Spwtroqc. , 28, 569 (1974).

LV. ,Jackson. Anal . C'hem., .17, 1034 i 19751.

( 3 8 ) K. M. Aldoris, 11. G. Mitchell. and K.

(39) Y. Talmi, ibid., p 658A. 140) Y. Talmi, ii-'id.. p 697.4. (41) K. tV. Rusch, N. G. Howell, and G. H.

Morrison, ibid , 16,575 11974) (421 hl. Margoshes, paper presented a t the

2lst Pittsburgh Conference on Analyti- cal Chemistry and Applied Spectroscopy, Cleveland, Ohio, March 3, 1970.

(43) A. Danielsson and P. Lindhlom, P h ~ s . Scr.. 5 , 227 (1972).

(4.1) 4. Danielsson. P. Lindblom, and E. Soderman, Chem. Scr., 6, !j (1974).

(45) D. L. LVood. A. B. DargEs, and D. L. Nash. Appl . S , ~ e c t r o s c . , 29, 310 1197.5).

146) J . D. Johnst.)n. paper presented at the 10th Middle Atlantic RegiNunal Meeting (MARMi, American Chemical Society, Philadelphia, Pa., February 2 3 , 1976.

(47 ) P. N. Keliher, paper prckented a t the 170th National Meeting of the American Chemical Society, Chicago, Ill., August 25, 1975.

(481 P. N. Keliher, Res. D e i . , in press.

P.N.K. thanks the Villanova University Faculty Research Program for its support.

3 4 0 A ANALYTICAL CHEMISTRY, VOL. 48, NO. 3, MARCH 1976