electric force microscopy (efm) efm is used to map the vertical (z) and near-vertical gradient of...
TRANSCRIPT
Electric Force Microscopy (EFM)
EFM is used to map the vertical (z) and near-vertical gradient of the electric field between the tip and the sample versus the in-plane coordinates x and y. This is done using LiftModeTM. The field due to trapped charges—on or beneath the sample surface—is often sufficiently large to generate contrast in an EFM image. Otherwise, a field can be induced by applying a voltage between the tip and the sample. The voltage may be applied directly from the microscope’s electronics under AFM software control, or from an external power supply with appropriate current-limiting elements in place. EFM is performed in one of three modes:
amplitude detection, phase detection, or frequency modulation (FM).
Review Last Week about EFM
EFM is used for electrical failure analysis, detecting trapped charges, mapping electric polarization, and performing electrical read/write, among other applications.
The application of EFM
Electrostatic Interactions is long range interaction
EFM/MFM
Conductive Domains
Magnetic tip, instead of only a conductive tip
Piezoelectric materials
In this method, the cantilever is vibrated by a small piezoelectric element near its resonant frequency. The cantilever’s resonant frequency changes in response to any additional force gradient. Attractive forces make the cantilever effectively “softer,” reducing the cantilever resonant frequency. Conversely, repulsive forces make the cantilever effectively “stiffer,” increasing the resonant frequency.
Resonance frequency
Phase line
Driving frequency
Normal TM AFM: ver der Waal Force gradientEFM: lift mode: Electrostatic Force gradient
Mechanical Driven Mode
V piezo = Vac sin(t+ 1)
Free cantilever: A1 (d) sin(dt)+ 1)Interacted cantilever: A2(d) sin( d t) + 2)
Constant drive frequency
Normal TM AFM: ver der Waal Force gradientEFM: lift mode: Electrostatic Force gradient
Electric Force Microscopy (EFM)
• Mechanical vibration given to tip
• Electric field will produce a change in phase of tip
• Measure phase lag relative to piezo drive signal.
• In most cases, it is necessary to apply a voltage across the tip or sample to achieve a high-quality image.
• Even if a layer of insulating material covered the conductive domains, they still can be detected.
EFM/MFM
Conductive Domains
Magnetic domains
Magnetic tip, instead of only a conductive tip
Piezoelectric materials
Comparison: Phase imaging and the phase image in EFM
• Phase image in EFM
During phase imaging, tapping mode imaging the tip is tapping the surface, the phase change due to the ver der Waal force gradient of the sample
Noncontact techniques, phase images are obtained in lift mode And the phase changes results from Electrostatic Force gradient
Surface Potential Detectionor Kelvin Probe Microscopy
Surface Potential
(SP) Imaging
SP imaging maps the electrostatic potential on the sample surface with or without a voltage applied to the sample. SP imaging is a nulling technique. As the tip travels above the sample surface in LiftMode, the tip and the cantilever experience a force wherever the potential on the surface is different from the potential of the tip. The force is nullified by varying the voltage of the tip so that the tip is at the same potential as the region of the sample surface underneath it. The voltage applied to the tip in nullifying the force is plotted versus the in-plane coordinates, creating the surface potential image.
Surface Potential Detectionor Kelvin Probe Microscopy
The piezo disengaged to stop mechanically driving the cantilever to vibrate, a AC voltage directly applied to the tip
• Detects the potential of the surface
• AC voltage to the tip
• (no induced mechanical vibration)
Vac= 0 - 12 Vpp
V1 V2
Sample surface
but the conducting regions should not be passivated.
• Let = fo
• Potential on surface will shift phase of the cantilever
• Detect phase change, use potential feedback loop to
keep phase constant
Surface Potential Detection
Vac= 0 - 12 Vpp
V1 V2
Sample surface
Electrostatic Force in a Capacitor
22
it charge to done workthe is capacitor a in Energy
Vz
C
z
CV
z
UF
z
WF
zFW
2Vd
C
d
VqqEF sample
Energy Treatment
Classical Treatment
Surface Potential Detection
First Harmonic Force Second Harmonic Force
• Let = Vdriving
• The lock-in technique allows extraction of the first
harmonic of tip deflection proportional to F. A feedback
loop is employed to keep it equal to zero by adjusting
the Vdc on the tip.
• Vdc=Vsurf
• Therefore the surface potential is directly measurement
by adjusting the potential offest on the tip and keep the
fisrt harmonic response to zero.
Not force gradient
( Deflection Amplitude caused by electrical force)
Comparison:
EFM and Surface Potential Microscopy
• Phase image in EFMNoncontact techniques, phase images are obtained in lift mode And the phase changes results from Electrostatic Force gradient
• The lock-in technique allows extraction of the first harmonic of
tip deflection proportional to F. A feedback loop is employed to
keep it equal to zero by adjusting the Vdc on the tip.
• Vdc=Vsurf
• Therefore the surface potential is directly measurement by
adjusting the potential offest on the tip and keep the first
harmonic response to zero.
• Potential image in Surface potential Microscope
(Amplitude modulation ( AM-KPFM)
EFM and Frequency modulation KPFM
Kelvin Probe Force Microscopy Study on Conjugated Polymer/Fullerene Bulk Heterojunction Organic Solar Cells
Nano Letters 2005, 5, 269
A breakthrough in power conversion efficiency in polymer solar cells was achieved for bulk heterojunctions with the MDMO-PPV/PCBM (poly-[2-(3,7-dimethyloctyloxy)-5-methyloxy]-para-phenylene-vinylene/1-(3-methoxycarbonyl) propyl-1-phenyl [6,6]C61) system.
The drastic increase in the power conversion efficiency was a result of changing the spin casting solvent from toluene to chlorobenzene, which accounted for a change in the film nanomorphology.
Charge separation in bulk heterojunctions is based on the Photoinduced charge transfer, and photoexcited excitons in Organic materials have a very limited diffusion length on the Order of 10 nm.
Not all photoexcitations will dissociate into separated charges carries.Photoluminescence of PCBM was found for toluene cast blends. This means that part of the adsorbed photons is not used for Photocurrent generation.
Another study reported that hole mobility of prinstine MDMO-PPV Films to decrease when cast from toluene as compared to chlorobezene.
In this study, Kelvin probe force microscope was used to reveal that the electronic work functions of films prepared from Different solvent.
sample - tip = q CP
tip=4.28 eV, PtIr coated Si cantilevers
CP: contact potential, q is the elementary charge