•EquipmentClassification- Class II equipment: - Type B shock protection: - Recommended sterilization: See 7. Sterilization-Donotuseinthepresenceofflammableliquidsorgases.
Cautions for safe handling and operation
Readthesesafetyprecautionsfirstbeforeusingthisproduct.These instructions will show you how to operate this product safely and prevent injury to yourself orothers.Theyareclassifiedbythedegreeand/orseverityofdanger.Allcautionsregardingsafetyshouldbeobserved.
Classification Degree and severity of danger
Warning Provides an instruction where personal injury may occur.
Caution Provides an instruction where minor to medium injury may occur.
Notice Providesaninstructionthatshouldbeobservedforsafetyreasons.
Warning
-ToPreventElectricShock:Donotpluginthepowercordwithwethands.-ToPreventElectricShock:Preventwaterfromgettingontothecontrolbox.-ToPreventElectricShock:Pluginandremovethepowerbyholdingontothebaseoftheplugand
not the cord.-ToPreventElectricShock:Useanelectricaloutletwhichisgrounded.-Ifyoufeelanyabnormalitysuchasexcessivevibration,heat,ornoise,stopusingtheunit immediately.
Caution
-Whenoperatingthissystem,alwaysconsiderthesafetyofthepatient.-Theproductisdesignedforclinicaldentalusebyqualifiedpersonnel-Everyconnectionmustbefirm.Ifitisnotfirmlyconnected,itmaycauseanairorwaterleak,ortheLEDmightnotturnon.-Donotdroptheproductorcauseittobeimpactedinanyway.Thiscouldcauseamalfunctionanddamage.-ThissystemistobeusedwithISO9168,typeBortypeChoseconnector.-Thissystemismeanttofunctionintemperaturesof0-40°C(32-104°F),humidityat10-85%RH,atmosphericpressure700-1060hPaandnomoistureorcondensationintheControlUnit.Operatingoutside of these parameters can cause a malfunction.-WheninstallingtheControlBoxorMotor,donotbendortwistthetubingorthecord.Noexcessive
force is need to do the installation.-Donotusecontaminatedair(bydustormoisture).Iftheaircontainswaterordust,itmightcauseamalfunctionoroverheating.
3
Caution
Unit-Graspcordbytheplugtoremovefromtheoutlet.Donotpulloryankthecorditself.Itmaycause
the wire to malfunction. -CareshouldbetakennottoplacetheACcordnearagasburner.Neverattempttorepairaburnedmotorcord.Alwaysreplaceitwithanewcord.-Priortouse,alwayscheckforvibration,noiseandoverheating.Ifanyabnormalitiesaredetected,stopusingimmediatelyandcontactyourdealer.-Becarefulnottospillwaterontotheunitasitmayresultinafireoranelectricshockduetoashort
circuit.
Motor, Hand piece -Donotusethisproductunderstrongstressforextendedperiodsoftime.Itcancauseoverheating.- Do not connect or disconnect the cord until the drive motor has completely stopped.-Donotconnect/disconnectthehandpieceduringoperation.-VerifythattheSpeedControlSwitchisadjustedwithintheallowablespeedbeforeuse.-Connectonly4-hole,5-holeor6-pintubing.-AirRequirements:Dry,freefromcontaminationandoil.Useacompressorwithadryairsystem/Installanairfilterifnecessary/Blowoutthelinesbeforeinstallation.-Donotautoclave(oranyotherhightemperaturesterilization)ControlBox,ACAdaptor,MotorCode.-Theusershallberesponsibleforoperationandmaintenanceandinspection.
NoticeDuringrotation,themotorandmotorcordmayaffectcomputerandLANcables.Noisecouldbeheardduringoperationnearradioreceivers.Aftertreatment,immediatelyturnthepowerandairsupplyoff.Removethepowercordiftheunitisnottobeusedforalongtime.
1. Specifications
Model Maxso E600P
Speed 2,000~40,000rpm
Torque 3.4Ncm
Rated Input AC24V/DC36
NoiseLevel <60dB
Dimensions W127xD149xH54(mm)
4
2. Component Names
1. ControlBox 2. DigitalSpeedDisplay 3. Gear Ratio Lamp 4. Gear Ration Select Key 5. FWD/REVSelectKey 6. Program1Button 7. Program2Button 8. SpeedAdjusters
19
102
3
4
6
5
7
8
12
11 13 14
16
15
9. Motor10. MotorCord11. PowerSwitch12. ACAdapterConnector13. MotorCordConnector14. TurbineHoseConnector15. ACCord16. ACAdaptor
5
3. Control Box Operation
Caution
Makesurethereisnoresidualairorwaterintheturbinehosewhenconnectingittothecontrolbox.Handtightenthenutproperlywithoutexcessiveforcewhenconnectingtheturbinehoseandthemotor cord connector.
3-1 Connect the hand piece tubing from your delivery unit. Fittheturbinehoseplugfromthedeliveryunittotheturbinehoseconnectorandtightencompletely.
Caution
- Do not use any other hose than the ones provided with the unit- Do not use contaminated air-Blowoutthelinesbeforeinstallation
Caution
NeveruseanyACadapterotherthantheoneprovidedwiththisproduct.
3-4 Connecting the Power CordInsertthepowercordintotheinletoftheACadapter.
3-2 Connecting the Motor.Connectthemotorbyaligningthemotorcordplugandinsertfirmly.
3-3 Connecting the AC adapterInserttheACadapterplugintotheACadapterconnectorbeforepluggingitintothewalloutlet.
6
4. Handling the Motor
4-1 Connecting/Disconnecting the motor and motor cord.Alignandinsertthemotorpinfirmlyintothepinholesofthemotor cord connector and fasten the motor cord nut securely.
Toremovethemotorcordfromthemotor,unscrewanddetachthemotorcordnut,andgentlypulloutthemotorcordconnector.
4-2 Connecting/ Disconnecting the motor and hand pieceToinserttheE-Typehandpieceintothemotor,alignthehandpieceandmotor,thenturnuntilyouhearaclick(positioningpinsarealigned)
Toremovethehandpiece,simplypullthehandpiecefromthemotor.
Caution
-Careshouldbetakenwhenusinganyhandpieceattachmentwhichrequireslubrication(E-Typemicromotor).-Followinglubrication,standthehandpieceattachmentverticallyuntiltheexcesssolutiondrains.
Wipe the hand piece then attach it to the motor.- Do not connect or disconnect the handpiece until the motor has completely stopped.
5. Operation
5-1InsertthepowercordplugintotheACoutletandturnthepowerswitchon.Whenthepowerswitchison,thelampsandindicator will come on.
5-2Beforeusingthissystem,makesuretooperatetheFootAirCalibrationfunction.
5-3PushtheGearRatioSelectKeytoindicatethespeedoftheattachment you will use.
*CheckTableBelowforSpeeds
HandPieceGearRatio Indicator Speed (min-1(rpm))
16:1 1-25 100–2,500
1:1 2–40 2,000–40,000
1:5 1–20 10,000–200,000
MotorMotor CordConnector
Motor CordNut
Motor Cord
Handpiece(Option)Positioning
Gear Ratio Lamp
Motor Insert
Push
7
5-4SetthemaximumspeedusingtheRotationAdjustmentKey(+/-).
5-5SelectFWD/REVrotationusingtheFWD/REVSelectkey.
5-6Themicromotoroperationiscontrolledbytheairswitch/foot pedal of the delivery unit.
5-7ProgramFunction.Youcanmemorizethesetting:1)setthevaluesforgearratio,speed2)PressP1orP2keyformorethan3secondsuntilitbeeping,thesettingiscompleted.
Push
DigitalSpeedDisplay
Increase
Decrease
FWD/REVLamp
6. Sterilization
Sterilize the motor only.Forsterilization,werecommendtheautoclavesterilizationmethod.Sterilizationisrequiredthefirsttimeyouusetheunitaswellasaftereachpatientasnotedbelow.
Autoclaving1)Turnoffthepower.2)Detachthemotorfromthemotorcord.3)Cleanthesurfaceofthemotorwithabrush(Donotuseametalbrush),andwipeitwithcottonanddisinfectantalcohol.4)ScrewtheMotorCaptotheMotor.PuttheAutoclavePlugonto the motor insert.5)Insertintoautoclaveouch.Sealthepouch.6)Autoclavableuptoamaximumof135°C.7)KeeptheMotorintheautoclavablepouchuntilyouareready to use it to keep it clean.
Caution
-Donotautoclave(oranyotherhightemperaturesterilization)ControlBox,ACAdaptor,orMotorCord.-Donotlubricatethemotor.- Do not wipe or immerse the system in acidic water or acidic solutions.-Donotsterilizewithdirtonthesurface,itmaycauserust.
AutoclavePlug
Motor Cap
Tighten
8
8. Error Codes
Error Code Issue Cause Solution
E0 Overcurrenterror(InCircuit)
Detected an overcurrent in the circuit
Contact your dealer
E1 Overcurrenterror.(InSoftware)
Detected an overcurrent in the software
Contact your dealer
E2 Retry Error Motor was overloaded and stopped and can not restart.
Contact your dealer
E3 ITRIP Error Faultymotororinternalcircuit Contact your Dealer
E5 OvervoltageInputError
Overvoltagetothecontrolbox Contact your dealer
E6 LEDvoltageerror Overvoltageintheinternalcir-cuit.(morethan3.7V)
Contact your dealer
E8 Overheatingerror Hightemperaturesafetysystemactivated due to extended use .
Wait for a while and letitcooldownbeforeusingagain.
7. Maintenance
IftheO-Ringsbecomeworn,themotormightnotconnectwelltothehandpiece,orwatermayleak.ReplacetheO-Rings.
Changing the O-RingRemovetheO-Ringatthemotorinsertwithapointedtool,andmountthenewO-ringsintothegroove.
Caution
Thereare4O-RingsontheMotorInsert.Theblueoneisthinnerthantheother3.Makesuretoinstall them in the proper spot.
Caution
IftheO-Ringsbecomeworn,itmaycause:-Air/Waterleak-Difficultyattaching/disconnectingthehandpiece-ExcessVibration
9
9. Troubleshooting
Whenhavinganissuewiththeunit,checkthisguidebeforecontactingyourdealer.Thesearethemost common issues which may occur.
Issue Cause Solution
Pilot Lamp does not come on Power switch is off-ACAdaptorisnotconnectedproperly-Internalfuseisblown
Turn on the switch-Check the connection-Contact your dealer
Motor doesn’t not run Tubing,MotorCord,ACadapt-er is not connected properly-Airpressureisnotsufficient
Check the connection-Check the air pressure from the delivery unit-refertoErrorcode9
The rotation speed is too slow -Airpressurefromdeliveryunitis too low-Airpressureonthedeliveryunit is lower than the “Lower limit of air pressure”.
Check the air pressure on the delivery system.-Do“Foot/Aircalibration
Beepswhenturningontheswitch
Youaresteppingonthefootpedalwhenturningonthepower switch
Do not step on the foot pedal whileturningontheswitch
E9 Motor start error Motor didn’t reach the preset speedintheprescribedamountof time.Break down of the motor cord wire,orinternalcircuitmalfunc-tion
Contact your dealer
EA LEDvoltageerror VoltagefortheLEDisn’thighenough
Contact your dealer
Eb Airoffseterror Lower than preset ‘Lower limit of air pressure”
Displayedbyfoot/aircalibrationfunction.NotaMalfunction
Ec Airfullscaleerror Higherthanpreset“Upperlimitofair pressure”Airpressurehigherthan0.4MPa
Displayedbyfoot/aircalibrationfunction.Notamalfunction
EE EEPROMError Malfunctionofthesettingsmemory
Contact your dealer
10
10.1 Scope of Warranty
BEYESDentalCanadaInc.(‘BEYES’)warrantstotheoriginalretailpurchaserthatitwillbeatBEYESoptiontorepairorreplacecomponentsofthedentalproductsmanufacturedbyBEYES(exceptforcomponentsnotwarrantedunder‘Exclusions’)thataredefectiveinmaterialorworkmanshipundernormaluseandservice.BEYES’obligationunderthislimitedwarrantyislimitedtotherepairorreplacementoftheapplicablecomponents.ThislimitedwarrantyshallonlyapplytodefectsthatarereportedtoBEYESwithintheapplicablewarrantyperiodandwhich,uponexaminationbyBeyes,provetobedefective.Thiswarrantyextendsonlytothefirstretailpurchaserofaproductandisnottransferableorassignable.Replacementcomponentsorproductsmaybeusedand/orrefurbishedcomponentsorproducts,providedtheyareoflikequalityandspecificationsasnewcomponentsorproducts.
10.2 Applicable Warranty PeriodTheapplicablewarrantyperiod,measuredfromthedateofinvoicetotheoriginaluser,shallbewarrantedforaperiodof25months.
10.3 ExclusionsThislimitedwarrantydoesnotcoverandBEYESshallnotbeliableforthefollowing;1)Defects,damageorotherconditionscaused,inwholeorinpart,bymisuse,abuse,negligence,alteration,accident,freightdamage,negligentstorage,tamperingorfailuretoseekandobtainrepairorreplacementinatimelymanner;2)Productswhicharenotinstalled,used,andproperlycleanedandmaintainedasrequiredorrecommendedintheBEYES‘Installation’and/or‘Installation/User’sManual’fortheapplicableproduct,includingthespecifiedstructuralandoperationalenvironmentconditionsandelectricalpowerrequirements;3)Productsconsideredtobeofaconsumableorsterilenature;4)AccessoriesorpartsnotmanufacturedbyBEYES;5)Chargesbyanyoneforadjustments,repairs,replacementparts,installationorotherworkperformeduponorinconnectionwithsuchproductswhicharenotexpresslyauthorizedinwritinginadvancebyBEYES;
LEDdoesnotlightup Reached the end of life expec-tancy
Contact your dealer
Motorheatsupabnormallyduringrotation
Coolantairisnotcomingoutofthe delivery unit
Check the air pressure of the delivery system
WaterLeakage Turbine,Motorcordisnotconnected correctly
Check the connection
Whenturnedon,thesettingvalues are different from what they were set to.
Turned off the motor while it wasrotating
Do not turn off the unit until afterthemotorstopsspinning
10. Beyes Limited Warranty Statement
11
6)Costsandexpensesofroutinemaintenanceandcleaning;7)RepresentationsandwarrantiesmadebyanypersonorentityotherthanBEYES;8)Matchingofcolor,grainortextureexcepttocommerciallyacceptablestandards;9)Changesincolorcausedbynaturalorartificiallight;10)Custommanufacturedproducts;11)AlterationsormodificationstotheproductbyanypersonorentityotherthanBEYES;12)ProductsthatwouldotherwisebecoveredunderSections1and2ofthislimitedwarranty,butareacquired:(i)fromapersonorentitythatisnotBEYESoroneofitsauthorizeddealers;or(ii)fromaBEYESdealerthatisnotauthorizedtoselltheproductatissueinthegeographicterritorywherethepurchaserislocated,orisnotauthorizedtoselltheproductatissuewithinthemedical,animalhealthordentalmarket,asthecasemaybe,inwhichpurchaserintendstousetheproduct.
10.4 Exclusive Remedy; Consequential Damages DisclaimerBEYES’OBLIGATIONUNDERTHISLIMITEDWARRANTYISTHEREPAIRORREPLACEMENTOFDEFECTIVEPARTS.BEYESSHALLNOTBELIABLEFORANDHEREBYDISCLAIMSANYDIRECT,SPECIAL,INDIRECT,INCIDENTAL,EXEMPLARYORCONSEQUENTIALDAMAGESORDELAYS,INCLUDING,BUTNOTLIMITEDTO,DAMAGESFORLOSSOFPROFITSORINCOME,LOSSOFUSE,DOWNTIME,COVERANDEMPLOYEEORINDEPENDENTCONTRACTORWAGES,PAYMENTSANDBENEFITS.
10.5 Warranty DisclaimerTHISLIMITEDWARRANTYISBEYES'ONLYWARRANTYANDISINLIEUOFALLOTHERWARRANTIES,EXPRESSORIMPLIED.BEYESMAKESNOIMPLIEDWARRANTIESOFANYKINDINCLUDINGANYIMPLIEDWARRANTIESOFMERCHANTABILITYORFITNESSFORAPARTICULARPURPOSE.THISWARRANTYISLIMITEDTOTHEREPAIRORREPLACEMENTOFDEFECTIVEPARTS.
Beyes Dental Canada Inc.23-595MiddlefieldRoadToronto,Ontario,P1V3S2Canada Tel:1-855-603-1888Fax: 1-855-720-1228Email: [email protected]:www.beyes.ca
Lotus NL B.V.KoninginJulianaplein10,IeVerd,2595AA,TheHague,NetherlandsTel: +31645171879(English) +31626669008(Dutch)Email: [email protected]:www.beyes.ca 0197
Printed in CanadaFile#:ENI019Rev.3/06.01.21
Federallawrestrictsthisdevicetosalebyorontheorderofadentist,physician,oranyotherpractitionerlicensedbythe law of the states in which he or she practices to use or order the use of this device. Beyes Dental Canada Inc. is not responsibleforanytypographicalerrors.Subjectstotechnicalchangeswithoutnotice.*Allbrandsareholdersoftheirrespective trademarks.