27.10.11 University of Rostock | Faculty of Medicine | Department of Orthopaedics Electro-Stimulating Implants for Bone Regeneration: Parameter Analysis on Design and Implant Position 1 Y.Su* 1 , R.Souffrant 1 , D.Klüß 1 , M.Ellenrieder 1 , H.Ewald 2 , W.Mittelmeier 1 , R.Bader 1 1 Department of Orthopaedics, University of Rostock, Rostock, Germany 2 Department of General Electrics, University of Rostock, Rostock, Germany *Doberaner Strasse 142, D-18057 Rostock, Germany, [email protected] www.forbiomit.med.uni-rostock.de

Upload: others

Post on 01-Jun-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

27.10.11 University of Rostock | Faculty of Medicine | Department of Orthopaedics

Electro-Stimulating Implants for Bone Regeneration:

Parameter Analysis on Design

and Implant Position

1

Y.Su*1, R.Souffrant1, D.Klüß1, M.Ellenrieder1,

H.Ewald2, W.Mittelmeier1, R.Bader1

1Department of Orthopaedics, University of Rostock, Rostock, Germany

2Department of General Electrics, University of Rostock, Rostock, Germany *Doberaner Strasse 142, D-18057 Rostock, Germany,

[email protected] www.forbiomit.med.uni-rostock.de

Outline

• Background • Method and Materials • Results of Numerical Simulation • Discussion and Outlook

2 University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

3 University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

Bone defects

• Tumor

• Infection

• Fracture

• Implant

lossenning

Why electro-stimulation of bone tissue?

• Bone growth

• Bone regeneration

• Fracture healing

Theory: Reciprocal piezoelectric effect

Piezoelectric effect Yasuda et al. 1957

4 University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

Habel 2004

Clinical Application of Electro-stimulation Treatment of avascular femoral head necrosis by ASNIS-S screw

Stryker Trauma GmbH, ASNIS III S-Series operative technique manual

5 University of Rostock | Faculty of Medicine | Department of Orthopaedics

stimulation parameter: sinusoidal, 20 Hz, 700mV

27.10.11

6

Methods

Comsol Multiphysics Real

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

Artificial femur (Sawbone)

COMSOL Settings

7 University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

• Geometry shape order: Linear

• Module: AC/DC module-Electric currents-Frequency domain

• Bondary conditions

ASNIS screw: Electric potential (700mV) and electric isolation

• Mesh

Screw boundary: free triangular, max element size: 35 µm

Bone: free tetrahedral, max element size: 1.5 mm

Surrounding material: free tetrahedral, max element size: 2 cm

Total approx. 1,28 million elements, approx. 1,7 million DOF

• Solver: Iterative solver- Conjugate gradients

Material Parameters

8

- Sinusoidal external magnetic field 20Hz - Voltage in the implant 700mV

[1] Gabriel C, Gabriel S, Corthout E. The dielectric properties of biological tissues: I. Literature survey. Phys Med Biol 1996; 41: 2231-49.

[2] Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys Med Biol 1996; 41(11): 2251-69.

[3] Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues. Phys Med Biol 1996; 41: 2271-93

Blood

Cortical bone

Cancellous bone

ASNIS screw

Material Conductivity σ [S/m]

Relative Permitivity

εr

Human Blood

0.7 5 260

Cortical bone

0.020045 25 119

Cancellous bone

0.078902 4 020 200

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

Parameter variation - Tip shape

9

Contour: Electric field norm (V/m)

• Optimized electric filed for avascular femoral head necrosis: 5-70V/m

5 W. Kraus Magnetfeldtherapie und magnetisch induzierte Elektrostimulation in der Orthopädie. Orthopäde, V 13, PP 78-92, 1984

University of Rostock | Faculty of Medicine | Department of Orthopaedics

69

27.10.11

10

Isosurface: Electric field norm (V/m)

5

35 Parameter variation - Tip shape

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

11

Femoral head

Parameter variation - Tip shape

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

12

Contour: Electric field norm (V/m)

Parameter variation – Screw positioning backward

University of Rostock | Faculty of Medicine | Department of Orthopaedics

5

69

27.10.11

13

B0 B1 B2

B3 B4 B5 Isosurface: Electric field norm (V/m)

5

35

Parameter variation – Screw positioning backward

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

14

Femoral head

Parameter variation – Screw positioning backward

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

15

Contour: Electric field norm (V/m)

Parameter variation – Screw positioning forward

University of Rostock | Faculty of Medicine | Department of Orthopaedics

5

69

27.10.11

16

F0 F1 F2

F3 F4 F5

Isosurface: Electric field norm (V/m)

Parameter variation – Screw positioning forward

5

35

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

17

Femoral head

Parameter variation – Screw positioning forward

University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

Discussion

18 University of Rostock | Faculty of Medicine | Department of Orthopaedics 27.10.11

• Simulation with bone material parameter data of our working group • Simulation and validation with a real human femoral head • Classification of patients bone defects according to own MRI data • Simulation with different bone defects

Outlook

• Numerical simulation is not yet validated

• Artifical bone compares to real human femoral head geometry

• Bone defects are not considered

14.07.

11 University of Rostock | Faculty of Medicine

Kind thanks www.forbiomit.med.uni-rostock.de