electronic instrumentation & control systems

90
ELECTRONIC INSTRUMENTATION & CONTROL SYSTEMS (WLE-306) Presented by: Mr. Shahnawaz Uddin 1

Upload: shahnawazuddin

Post on 18-Oct-2015

17 views

Category:

Documents


0 download

TRANSCRIPT

  • 5/28/2018 Electronic Instrumentation & Control Systems

    1/90

    ELECTRONIC INSTRUMENTATION &CONTROL SYSTEMS

    (WLE-306)

    Presented by:

    Mr. Shahnawaz Uddin

    1

  • 5/28/2018 Electronic Instrumentation & Control Systems

    2/90

    Unit-4

    MISCELLANEOUS INSTRUMENTS

    2

  • 5/28/2018 Electronic Instrumentation & Control Systems

    3/90

    Amplitude Distortion Distortionis the alteration of the original shape (or other characteristic)

    of a signal, waveform, or other form of information

    Distortion is usually unwanted and in practice, many methods are

    employed to minimize it

    In signal processing, a noise-free system can be characterized by a

    transfer function, such that the output y(t) can be written as a function of

    the input x(t) as: y(t) = F(x(t))

    When the transfer function comprises only a gain (A) and delay (T), thenthe output is undistorted

    Distortion occurs when the transfer function F is more complicated than

    this, e.g., if F is a linear function of frequency (for instance a filter whose

    gain and/or delay varies with frequency), then the signal will experiencelinear distortion

    The linear distortion will not change the shape of a single sinuosoid, but

    will usually change the shape of a multi-tone signal

    3

  • 5/28/2018 Electronic Instrumentation & Control Systems

    4/90

    Amplitude distortionis distortion occurring in a system, subsystem,

    or device when the output amplitude is not a linear function of the

    input amplitude

    For example, in case of a transistor, output is a linear function of

    input only for a fixed portion of the transfer characteristic, i.e., Ic = Ib

    When output is not in this portion, two forms of amplitude distortion

    might arise:

    (i) Harmonic Distortion, & (ii) Intermodulation Distortion(i) Harmonic distortion:

    The creation of harmonics of the fundamental frequency of a

    sinusoidal wave to a system

    (ii) Intermodulation distortion: This form of distortion occurs when two sinusoidal waves of

    frequencies f1and f2are present at the input, resulting in the creation

    of several other frequency components, whose frequencies include

    (f1+ f2), (f1- f2), (2f1- f2), (2f2f1), and in general (mf1 nf2) for

    integer m and n

    Amplitude Distortion (-contd.)

    4

  • 5/28/2018 Electronic Instrumentation & Control Systems

    5/90

    Generally the strength of the unwanted output falls rapidly as m and n

    increase

    Amplitude distortion is measured with the system operating understeady-state conditions with a sinusoidal input signal

    When other frequencies are present, the term "ampl i tude"refers to

    the amplitude of fundamental frequency component only

    It can be shown mathematically (Fourier Series Analysis) that any

    complex waveform is made up of a fundamental frequency (f0)component and its harmonics (2f0, 3f0, 4f0, )

    It is often desired to measure the amplitude of fundamental or each

    harmonic individually, and can be performed by instruments called

    wave analyzers Wave analyzers are also referred to as frequency selective

    voltmeters, carrier frequency voltmeters, orselective level

    voltmeters

    Some wave analyzers have the facility of automatic frequency

    control, in which the tuning automatically locks to the signal

    Amplitude Distortion (-contd.)

    5

  • 5/28/2018 Electronic Instrumentation & Control Systems

    6/90

    This makes it possible to measure the amplitude of signals that are

    drifting in frequency by amounts that would carry them outside the

    widest pass-band available

    Harmonic distortion analyzers measure the total harmonic content in

    the waveforms

    Harmonic distortion can be quantitatively measured very accurately with

    harmonic distortion analyzer, generally called a distortion analyzer

    The total harmonic distortion (THD) is given by

    where, D2, D3, D4, represent 2nd, 3rd, 4th, harmonics

    The harmonic distortion analyzer measures the total harmonic distortion

    without individually the amplitude & frequency of each component

    These analyzers can be used along with a frequency generator or a source

    of white (or pseudo-random) noise to measure the frequency response of

    amplifiers, filters, etc.

    Amplitude Distortion (-contd.)

    ...DDDD 242

    3

    2

    2

    6

  • 5/28/2018 Electronic Instrumentation & Control Systems

    7/90

    Fig. (4.1) Graph of a Waveform and the distorted versions of the same waveform

    Amplitude Distortion (-contd.)

    7

  • 5/28/2018 Electronic Instrumentation & Control Systems

    8/90

    Basic Wave Analyzer

    A basic wave analyzer is shown in fig. (9.1a), and consists of a

    primary detector (a simple LC circuit)

    This LC circuit is adjusted for resonance at the frequency of the

    particular harmonic component to be measured

    The intermediate stage is a full wave rectifier, to obtain the

    average value of the input signal

    The indicating device is a simple dc voltmeter that is calibrated

    to read the peak value of the sinusoidal input voltage

    Since, the LC circuit is tuned to a single frequency, it passes

    only the frequency to which it is tuned and rejects all other

    frequencies

    A number of tuned filters, connected to the indicating device

    through a selector switch, would be required for a Wave

    Analyzer

    8

  • 5/28/2018 Electronic Instrumentation & Control Systems

    9/90

    Basic Wave Analyzer (-contd.)

    Fig. (9.1a) Basic Wave Analyzer

    9

  • 5/28/2018 Electronic Instrumentation & Control Systems

    10/90

    Basic Wave Analyzer (-contd.)

    10

  • 5/28/2018 Electronic Instrumentation & Control Systems

    11/90

    Frequency Selective Wave Analyzer

    Wave analyzer (fig. 9.1b) consists of a very narrow pass-band

    filter section which can be tuned to a particular frequency within

    the audible frequency range (20 Hz -20 kHz) The complex wave to be analyzed is passed through an adjustable

    attenuator, which serves as a range multiplier and permits a large

    range of signal amplitudes to be analyzed without loading the

    amplifier

    The driver amplifier applies the attenuated input signal to a high-Q

    active filter (a low pass filter, which allows the selected frequency

    to pass and reject all others)

    The magnitude of this selected frequency is indicated by the meter

    and the filter section identifies the frequency of the component

    The filter circuit consists of a cascaded RC resonant circuits and

    amplifiers

    11

  • 5/28/2018 Electronic Instrumentation & Control Systems

    12/90

    The capacitors are varied for range changing (i.e., coarse tuning)

    & the potentiometer is used to change the frequency within the

    selected pass-band (i.e., fine tuning), hence, this wave analyzer is

    also called a frequency selective voltmeter

    The selected signal output from the final amplifier stage is applied

    to the meter circuit & to an un-tuned buffer amplifier

    The main function of the buffer amplifier is to drive output devices,such as recorders or electronics counters

    The meter has several voltage ranges as well as decibel scales

    marked on it

    It is driven by an average reading rectifier type detector The bandwidth of the instrument is very narrow, typically about 1%

    of the selective band given in response characteristics (fig. 9.2)

    Frequency Selective Wave Analyzer (-contd.)

    12

  • 5/28/2018 Electronic Instrumentation & Control Systems

    13/90

    Frequency Selective Wave Analyzer (-contd.)

    13

  • 5/28/2018 Electronic Instrumentation & Control Systems

    14/90

    Frequency Selective Wave Analyzer (-contd.)

    14

    H t d W A l

  • 5/28/2018 Electronic Instrumentation & Control Systems

    15/90

    Heterodyne Wave Analyzer

    The wave analyzers are useful for measurement in the audio

    frequency range only, i.e., for measurements in the RF range and

    above (MHz range), an ordinary wave analyzer cant be used

    Hence, special types of wave analyzers working on the principle of

    heterodyning (mixing) are used, which are known as Heterodyne

    wave analyzers

    In Heterodyne wave analyzer, the input signal to be analyzed is

    heterodyned with the signal from the internal tunable local oscillatorin the mixer stage to produce a higher IF frequency

    By tuning the local oscillator frequency, various signal frequency

    components can be shifted within the pass-band of the IF amplifier

    The output of the IF amplifier is rectified and applied to the metercircuit

    An instrument that involves the principle of heterodyning is the

    Heterodyning tuned voltmeter (shown in fig. 9.3)

    The input signal is heterodyned to the known IF by means of a

    tunable local oscillator 15

    H t d W A l ( td )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    16/90

    The amplitude of the unknown component is indicated by the

    VTVM (Vacuum Tube Voltmeter) or output meter

    The frequency of the component is identified by the local oscillatorfrequency, i.e., the local oscillator frequency is varied so that all

    the components can be identified

    The fixed frequency amplifier is a multistage amplifier, which can

    be designed conveniently because of its frequency characteristics With the use of a suitable attenuator, a wide range of voltage

    amplitudes can be covered

    Their disadvantage is the occurrence of spurious cross-modulation

    products, setting a lower limit to the amplitude that can bemeasured

    Heterodyne Wave Analyzer (-contd.)

    16

    H t d W A l ( td )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    17/90

    Two types of frequency-selective amplifiers find use in Heterodyne

    wave analyzers

    The first type employs a crystal filter (band-pass arrangement),having a center frequency of 50 kHz; another type uses a resonant

    circuit in which the effective Q has been made high and is controlled

    by negative feedback

    When a knowledge of the individual amplitudes of the component

    frequency is desired, a heterodyne wave analyzer is used

    A modified heterodyne wave analyzer is shown in fig. 9.4

    In this analyzer, the attenuator provides the required input signal for

    heterodyning in the first mixer stage, with the signal from a local

    oscillator having a frequency of 30-48 MHz The first mixer stage produces an output which is the difference of

    the local oscillator frequency and the input signal, to produce an IF

    signal of 30 MHz

    Heterodyne Wave Analyzer (-contd.)

    17

    H t d W A l ( td )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    18/90

    This IF frequency is uniformly amplified by the IF amplifier

    This amplified IF signal is fed to the second mixer stage, where it

    is again heterodyned to produce a difference frequency or IF ofzero frequency

    The selected component is then passed to the meter amplifier and

    detector circuit through an active filter having a controlled band-

    width The meter detector output can then be read off on a db-calibrated

    scale, or may be applied to a secondary device such as a recorder

    This wave analyzer is operated in the RF range of 10 kHz -18 MHz

    with 18 overlapping bands selected by the frequency range controlof the local oscillator

    The bandwidth, which is controlled by the active filter, can be

    selected at 200 Hz, 1 kHz, and 3 kHz

    Heterodyne Wave Analyzer (-contd.)

    18

    H t d W A l ( td )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    19/90

    Heterodyne Wave Analyzer (-contd.)

    19

    Heterod ne Wa e Anal er ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    20/90

    Heterodyne Wave Analyzer (-contd.)

    20

    H i Di t ti A l

  • 5/28/2018 Electronic Instrumentation & Control Systems

    21/90

    Harmonic Distortion AnalyzerFundamental Suppression Type:

    Distortion analyzer measures the total harmonic power present in the

    test wave rather than the distortion caused by each component The simplest method to suppress the fundamental frequency by

    means of a high pass filter whose cut-off frequency is a little above

    the fundamental frequency

    Thus, the high pass filter allows only the harmonics to pass and the

    total harmonic distortion (THD) can then be measured

    The most commonly used harmonic distortion analyzers based on

    fundamental suppression are as follow:

    (i) Employing a Resonance Bridge, (ii) Wien's Bridge Method

    (iii) Bridged T -Network Method

    (i) Employing a Resonance Bridge:

    The bridge, shown in fig. (9.5), is balanced for the fundamental

    frequency, i.e., L & C are tuned to the fundamental frequency21

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    22/90

    The bridge is unbalanced for theharmonics, i.e., only harmonic

    power will be available at the output terminal and can be measured

    If the fundamental frequency is changed, the bridge must bebalanced again by varying L & C

    If L & C are fixed components, then this method is suitable only when

    the test wave has a fixed frequency

    Indicators can be thermocouples or square law VTVMs (VacuumTube Volte Meters), which indicate the rms value of all harmonics

    When a continuous adjustment of the fundamental frequency is

    desired, a Wien bridge arrangement is used (shown in fig. 9.6)

    (ii) Wien's Bridge Method:

    The bridge is balanced for the fundamental frequency, therefore,

    fundamental energy is dissipated in the bridge circuit elements

    Only the harmonic components reach the output terminals

    Harmonic Distortion Analyzer (-contd.)

    22

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    23/90

    The harmonic distortion output can then be measured witha meter

    For balance at the fundamental frequency:

    C1= C

    2= C, R

    1= R

    2= R, R

    3= 2R

    4

    (iii) Bridged T -Network Method:

    As shown in fig. (9.7), L & C's are tuned to the fundamental

    frequency, and Ris adjusted to bypass fundamental frequency

    The tank circuit being tuned to the fundamental frequency, thefundamental energy will circulate in the tank and is bypassed by the

    resistance

    Only harmonic components will reach the output terminals and the

    distorted output can be measured by the meter

    The Q of the resonant circuit must be at least 3-5

    One method of using a bridge T-network is given in fig. (9.8)

    The switch S is first connected to point A so that the attenuator is

    excluded and the bridge T-network is adjusted for full suppression of

    the fundamental frequency, i.e., minimum output

    Harmonic Distortion Analyzer (-contd.)

    23

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    24/90

    Minimum output indicates that the bridged T-network is tuned to the

    fundamental frequency & fundamental frequency is fully suppressed

    The switch is next connected to terminal B, i.e. the bridged T-network

    is excluded

    Attenuation is adjusted until the same reading is obtained on the

    meter

    The attenuator reading indicates the total rms distortion

    Note:

    Distortion measurement can also be obtained by means of a wave

    analyzer; knowing the amplitude & frequency of each component; the

    harmonic distortion can be calculated

    However, distortion meters based on fundamental suppression aresimpler to design and less expensive than wave analyzers

    The disadvantage with the harmonic distortion analyzers is that they

    give only the total distortion and not the amplitude of individual

    distortion components

    Harmonic Distortion Analyzer (-contd.)

    24

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    25/90

    Harmonic Distortion Analyzer (-contd.)

    Fig. (9.5) Resonance Bridge

    25

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    26/90

    Harmonic Distortion Analyzer (-contd.)

    Fig. (9.6) Wiens Bridge Method 26

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    27/90

    Harmonic Distortion Analyzer (-contd.)

    27

    Harmonic Distortion Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    28/90

    Harmonic Distortion Analyzer (-contd.)

    28

    Spectrum Analyzer

  • 5/28/2018 Electronic Instrumentation & Control Systems

    29/90

    Spectrum Analyzer The most common way of observing signals is to display them on an

    oscilloscope, with time on the x-axis (i.e., amplitude of the signal

    versus time)

    It is also useful to display signals in the frequency domain; theinstrument providing this frequency domain view is the spectrum

    analyzer

    A spectrum analyzer provides a calibrated graphical display on its

    CRT, with frequency on the horizontal axis and amplitude (voltage)on the vertical axis

    Displayed as vertical lines against these coordinates are sinusoidal

    components of which the input signal is composed

    The height represents the absolute magnitude, and the horizontal

    location represents the frequency

    These instruments provide a display of the frequency spectrum over

    given frequency band

    Spectrum analyzers use either (i) a parallel filter bank, or(ii) a

    swept frequency technique 29

    Spectrum Analyzer ( contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    30/90

    (i) Spectrum Analyzer using Parallel Filter Bank:

    In a parallel filter bank analyzer, the frequency range is covered by a

    series of filters whose central frequencies and bandwidths are so

    selected that they overlap each other (as shown in Fig. 9.9a)

    Typically, an audio analyzer will have 32 of these filters, each covering

    one third of an octave

    For wide band narrow resolution analysis, particularly at RF or

    microwave signals, the swept technique is preferred

    (ii) Spectrum Analyzer using Swept Receiver Design:

    As shown in fig. (9.9b), the sawtooth generator provides the sawtooth

    voltage which drives the horizontal axis element of the scope and this

    sawtooth voltage is frequency controlled element of the voltage tunedoscillator

    As the oscillator sweeps from fminto fmaxof its frequency band at a linear

    recurring rate, it beats with the frequency component of the input signal

    & produces an IF, whenever a frequency component is met during its

    sweep

    Spectrum Analyzer (-contd.)

    30

    Spectrum Analyzer (-contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    31/90

    The IF corresponding to the frequency component is amplified and

    detected if necessary, and then applied to the vertical plates of the

    CRO, producing a display of amplitude versus frequency

    One of the principal applications of spectrum analyzers has been in

    the study of the RF spectrum produced in microwave instruments

    In a microwave instrument, the horizontal axis can display a wide

    range (2-3 GHz) for a broad survey and a narrow range (30 kHz) as

    well for a highly magnified view of any small portion of the spectrum

    Signals at microwave frequency separated by only a few kHz can be

    seen individually

    The basic block diagram of an RF spectrum analyzer (fig. 9.13)

    covers the range 500 kHz to 1 GHz, which is representative of asuper-heterodyne type

    The input signal is fed into a mixer which is driven by a local oscillator

    (which is linearly tunable electrically over the range 2-3 GHz)

    Spectrum Analyzer (-contd.)

    31

    Spectrum Analyzer (-contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    32/90

    The mixer provides two signals at its output that are proportional in

    amplitude to the input signal but of frequencies which are the sum

    and difference of the input signal & local oscillator frequency

    The IF amplifier is tuned to a narrow band around 2 GHz, since the

    local oscillator is tuned over the range of 2-3 GHz, only the inputs

    that are separated from the local oscillator frequency by 2 GHz will be

    converted to IF frequency band, pass through the IF frequency

    amplifier, get rectified & produce a vertical deflection on the CRT From this, it is observed that as the sawtooth signal sweeps, the local

    oscillator also sweeps linearly from 2-3 GHz

    The tuning of the spectrum analyzer is a swept receiver, which

    sweeps linearly from 0 to 1 GHz

    The sawtooth scanning signal is also applied to the horizontal plates

    of the CRT to form the frequency axis

    Spectrum analyzers are widely used in radars, oceanography, and

    bio-medical fields

    Spectrum Analyzer (-contd.)

    32

    Spectrum Analyzer (-contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    33/90

    Spectrum Analyzer (-contd.)

    33

    Basic Spectrum Analyzer Using Swept Receiver Design

  • 5/28/2018 Electronic Instrumentation & Control Systems

    34/90

    Basic Spectrum Analyzer Using Swept Receiver Design

    34

    Basic Spectrum Analyzer Using Swept Receiver Design

  • 5/28/2018 Electronic Instrumentation & Control Systems

    35/90

    Basic Spectrum Analyzer Using Swept Receiver Design

    35

    Basic Spectrum Analyzer Using Swept Receiver Design

  • 5/28/2018 Electronic Instrumentation & Control Systems

    36/90

    Basic Spectrum Analyzer Using Swept Receiver Design

    Fig. (9.12) Test Waveform as seen on X-axis (time) & Z-axis (frequency)

    Fig. (9.13) RF Spectrum Analyzer 36

    Q-METER

  • 5/28/2018 Electronic Instrumentation & Control Systems

    37/90

    Q METER

    The overall efficiency of coils and capacitors intended for RF

    applications is best evaluated using the Q-value

    The Q-meter is an instrument designed to measure some electrical

    properties of coils and capacitors

    The principle of Q-meter is based on series resonance; the voltage

    drop across the coil or capacitor is Q-times the applied voltage

    (where Q is the ratio of reactance to resistance, XL/R)

    If a fixed voltage is applied to the circuit, a voltmeter across thecapacitor can be calibrated to read Q directly

    At resonance XL= XCand EL= I XL, EC= I XC, E = IR

    Therefore,

    From the above equation, if E is kept constant, the voltage across the

    capacitor can be measured by a voltmeter calibrated to read directly

    in terms of Q

    E

    E

    R

    X

    R

    XQ CCL

    37

    Q-METER (-contd )

  • 5/28/2018 Electronic Instrumentation & Control Systems

    38/90

    A practical Q-meter circuit is shown in fig.(10.7)

    The wide range oscillator, with frequency range from 50 kHz to 50 MHz,

    delivers a current to the shunt resistance (Rsh) having a value of 0.02

    Rsh introduces almost no resistance into the tank circuit and therefore,

    represents a voltage source of magnitude e with a small internal

    resistance

    The voltage across the capacitor is measured by an electronic voltmeter

    corresponding to ECand calibrated directly to read Q The circuit is tuned to resonance by varying C until the electronic

    voltmeter reads the maximum value

    The resonance output voltage E, corresponding to EC , is E = Q x e

    That is, Q = E/e

    Since, e is known, the electronic voltmeter can be calibrated to read Q

    directly

    The inductance of the coil can be determined by connecting it to the test

    terminals of the instrument

    Q METER ( contd.)

    38

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    39/90

    The circuit is tuned to resonance by varying either the capacitance or the

    oscillator frequency

    If the capacitance is varied, the oscillator frequency is set to a given

    frequency & resonance is obtained If the capacitance is preset to a desired value, the oscillator frequency is

    varied until resonance occurs

    The inductance of the coil can be calculated from known values of the

    resonant frequency & resonating capacitor (C)

    The Q indicated is not the actual Q, because the losses of the resonating

    capacitor, voltmeter and inserted resistance are all included in the

    measuring circuit The actual Q of the measured coil is somewhat greater than the

    indicated Q

    This difference is negligible except where the resistance of the coil is

    relatively small compared to the inserted resistance Rsh

    Q METER ( contd.)

    C)f2(

    1Lor,

    LC2

    1f,XX

    2CL

    39

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    40/90

    Q METER ( contd.)

    Fig. (10.7) Circuit Diagram of a Q-meter

    40

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    41/90

    Factors Causing Error during Q-measurement:

    (1)At high frequencies the electronic voltmeter may suffer from losses

    due to the transit time effect

    The effect of Rshis to introduce an additional resistance in the tank

    circuit, as shown in fig. (10.8)

    To make the Qobsvalue as close as possible to Qact, Rshshould bemade as small as possible (Rshvalue of 0.02-0.04 introduces

    negligible error)

    (2)Another source of error, and probably the most important one, is the

    distributed capacitance or self capacitance of the measuring circuit

    Q METER ( contd.)

    )R

    R1(QQ,Hence

    RR1

    RRR

    QQ

    RR

    LQand

    R

    LQ

    shobsact

    shsh

    obs

    act

    sh

    obsact

    41

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    42/90

    Q METER ( contd.)

    42

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    43/90

    The presence of distributed or stray capacitances modifies the actual

    Q and the inductance of the coil

    At the resonant frequency, at which the self capacitance and inductance

    of the coil are equal, the circuit impedance is purely resistive; thischaracteristic can be used to measure the distributed capacitance

    One of the simplest methods of determining the distributed capacitance

    (Cs) of a coil involves the plotting of a graph of 1/f2against C (in pF) as

    shown in fig. (10.9a)

    The frequency of the oscillator in the Q meter is varied and the

    corresponding value of C for resonance is noted

    The straight line produced to intercept the x-axis gives the value of Cs

    Q METER ( contd.)

    s2

    s

    2

    2

    s

    2

    2

    CCthen,0f

    1

    If

    )CC(L4f

    1or,

    )CC(L2

    1fand

    L4Slope,therefore,

    4

    SlopeL

    43

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    44/90

    The value of unknown can also be determined from the above

    equation

    Another method of determining the stray or distributed capacitance

    (Cs) of a coil involves making two measurements at different

    frequencies

    The capacitor C of the Q-meter is calibrated to indicate the

    capacitance value

    The test coil is connected to the Q-meter terminals as shown infig.(10.9b)

    The tuning capacitor is set to a high value position (to its maximum)

    and the circuit is resonated by varying the oscillator frequency

    Suppose the meter indicates resonance & the oscillator frequency isfound to be f1& the capacitance value to be C1

    The oscillator frequency of the Q-meter is now increased to twice the

    original frequency, i.e., f2= 2f1, and the capacitor is varied until

    resonance occurs at C2

    Q METER ( contd.)

    44

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    45/90

    The resonant frequency of an LC circuit is given by

    Therefore, for the initial resonance condition, the total capacitance of the

    circuit is (C1+ Cs)and the resonant frequency is given by

    After the oscillator and the tuning capacitor are varied for the new value

    of resonance, the capacitance is (C2+ Cs), therefore,

    But f2= 2f1, therefore,

    Hence, C1+ Cs= 4 (C2+ Cs)

    The distributed capacitance can be calculated using the above equation

    Q ( co td )

    LC2

    1f

    )CC(L2

    1f

    s1

    1

    )CC(L2

    1f

    s2

    2

    )CC(L2

    12

    )CC(L2

    1

    s1s2

    3

    C4CC 21s

    45

    Q-METER (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    46/90

    Q METER ( contd.)

    46

    Examples

  • 5/28/2018 Electronic Instrumentation & Control Systems

    47/90

    ExamplesEx. 10.1: The self capacitance of a coil is measured by using the

    outlined in the previous section. The first measurement is at f1=1 MHz

    & C1=500 pF. The second measurement is at f2=2 MHz & C2=110 pF.

    Find the distributed capacitance. Also calculate the value L.(Ans. 20 pF, 48.712 H)

    Ex. 10.2: Calculate the value of the self capacitance when the following

    measurements are performed: f1=2 MHz & C1=500 pF

    f2=6 MHz & C2=50 pF

    (Ans. 6.25 pF)

    Problem-1:The distributed capacitance was found to be 20 pF by use

    of a Q-meter. The first resonance occurred at C1=300 pF & f1was

    half the second resonance frequency. Determine the value of f2at the

    second resonance (given L=40 H) (Ans. 2.8 MHz)47

    Electroencephalogram (EEG)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    48/90

    Electroencephalogram (EEG)

    An electroencephalogram (EEG) is a test that measures and records

    the electrical activity of the brain

    Special sensors (electrodes) are attached to your head and hookedby wires to a computer

    The computer records your brain's electrical activity on the screen or

    on paper as wavy lines

    Certain conditions, such as seizures, can be seen by the changes in

    the normal pattern of the brain's electrical activity

    EEG may be done to:

    Diagnose epilepsy and see what type of seizures are occurring

    Check for problems with loss of consciousness or dementia

    Find out if a person who is in a coma is brain-dead

    Study sleep disorders, such as narcolepsy

    Watch brain activity while a person is receiving general

    anesthesia during brain surgery48

    EEG (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    49/90

    Help find out if a person has a physical problem (problems in the

    brain, spinal cord, or nervous system) or a mental health problem

    How EEG is Done? The EEG record is read by a doctor who is specially trained to

    diagnose and treat disorders affecting the nervous system

    (neurologist)

    You will be asked to lie on your back on a bed or table or relax in achair with your eyes closed

    The EEG technologist will attach 10 to 20 flat metal discs (electrodes)

    to different places on your head, using a sticky electrolyte paste or

    jelly to hold the electrodes in place (A cap with fixed electrodes may

    be placed on your head instead of individual electrodes) The electrodes are hooked by wires to an EEG machine that records

    the brain activity drawn by a row of pens on a moving piece of paper

    or as an image on the computer screen

    EEG ( contd.)

    49

    EEG (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    50/90

    You may be asked to breathe deeply and rapidly (hyperventilate), usually

    20 breaths a minute for 3 minutes

    You may be asked to look at a bright, flashing light called a strobe

    (photic or stroboscopic stimulation)

    Results: There are several types of brain waves:

    Alpha Waves have a frequency of 8 to 12 cycles per second. Alpha

    waves are present only in the waking state when your eyes are closed

    but you are mentally alert. Alpha waves go away when your eyes areopen or you are concentrating.

    Beta Waves have a frequency of 13 to 30 cycles per second. These

    waves are normally found when you are alert or have taken high doses

    of certain medicines, such as benzodiazepines.

    Delta Waves have a frequency of less than 3 cycles per second. Thesewaves are normally found only when you are asleep or in young children.

    Theta Waves have a frequency of 4 to 7 cycles per second. These

    waves are seen in drowsiness or arousal in older children and adults; it

    can also be seen in meditation

    ( )

    50

  • 5/28/2018 Electronic Instrumentation & Control Systems

    51/90

    Fig. (1)The cerebrum contains the frontal, parietal, temporal and occipital lobes51

  • 5/28/2018 Electronic Instrumentation & Control Systems

    52/90

    Fig. (2)The 1020 electrode system for measuring the EEG

    52

  • 5/28/2018 Electronic Instrumentation & Control Systems

    53/90

    Fig. (3)A man undergoing an EEG, wearing a cap equipped with electrodes

    53

  • 5/28/2018 Electronic Instrumentation & Control Systems

    54/90

    Fig. 4(a) Four types of EEG waves

    Fig. 4(b) When the eyes are

    opened, alpha waves disappear

    54

  • 5/28/2018 Electronic Instrumentation & Control Systems

    55/90

    Electroencephalogram (EEG)

    Normal In adults who are awake, the EEG shows mostly alpha waves and betawaves.The two sides of the brain show similar patterns of electrical activity.There are no abnormal bursts of electrical activity and no slow brain

    waves on the EEG tracing.If flashing lights (photic stimulation) are used during the test, one area

    of the brain (the occipital region) may have a brief response after eachflash of light, but the brain waves are normal.

    Abnormal The two sides of the brain show different patterns of electricalactivity. This may mean a problem in one area or side of the brain is

    present.The EEG shows sudden bursts of electrical activity (spikes) or sudden

    slowing of brain waves in the brain. These changes may be caused by

    a brain tumor, infection, injury, stroke, or epilepsy.55

    El h l (EEG)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    56/90

    Electroencephalogram (EEG)

    Abnormal The EEG records changes in the brain waves that may not be injust one area of the brain. A problem affecting the entire brain-

    such as drug intoxication, infections (encephalitis), or metabolic

    disorders (such as diabetic ketoacidosis) that change the chemical

    balance in the body, including the brain-may cause these kinds of

    changes.The EEG shows delta waves or too many theta waves in adultswho are awake. These results may mean brain injury or a brain

    illness is present. Some medicines can also cause this.The EEG shows no electrical activity in the brain (a "flat" or

    "straight-line" EEG). This means that brain function has stopped,

    which is usually caused by lack of oxygen or blood flow inside

    the brain. This may happen when a person has been in a coma. In

    some cases, severe drug-induced sedation can cause a flat EEG.56

    EEG (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    57/90

    What factors may affect the EEG Test?

    Reasons why the results may not be helpful include:

    (i) Moving too much

    (ii) Taking some medicines, such as those used to treat seizures

    (antiepileptic medicines) or sedatives, tranquilizers, and barbiturates

    (iii) Being unconscious from severe drug poisoning or a very low body

    temperature (hypothermia)(iv) Having hair that is dirty, oily, or covered with hairspray or other hair

    preparations. This can cause a problem with the placement of the

    electrodes.

    ( )

    57

    Electrocardiography

  • 5/28/2018 Electronic Instrumentation & Control Systems

    58/90

    Electrocardiography

    An electrocardiogram (ECG or EKG) is an electrical recording of the

    heart activity over time and is used in the investigation of heart

    diseaseBritish physiologist Augustus D. Waller was the pioneer of

    electrocardiography and in 1887 published the first human

    electrocardiogram

    In 1903 Dutch physiologist, Willem Einthoven, transformed thiscurious physiologic phenomenon into an indispensable clinical

    recording device that is still used today

    ECG is a surface measurement of the electrical potential generated

    by electrical activity in cardiac tissue

    The human heart can be considered as a large muscle whose

    beating is simply a muscular contraction which develops a potential

    to be measured in the form of ECG58

  • 5/28/2018 Electronic Instrumentation & Control Systems

    59/90

    Fig. (1)59

    Electrocardiography (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    60/90

    Three Leeds of ECG:

    The differential potential is

    measured between the right and left

    arm, between the right arm and the

    left leg and between left arm and left

    leg

    These three measurements are

    referred to as leads I, II, IIIrespectively

    The signal from the body is being

    amplified because the signals from

    the body are small and weak,

    ranging from 0.5 mV to 5.0 mV

    Signals are filtered to remove the

    noise, then after digital conversion

    through ADC the digital signal is

    sent to computer

    g p y ( )

    Fig. (2)

    60

    Electrocardiography (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    61/90

    Fig. (3)Block diagram of an electrocardiograph. The normal locations for

    surface electrodes are right arm (RA), right leg (RL), left arm (LA), and left

    leg (LL). Physicians usually attach several electrodes on the chest of the

    patients as well.

    Resistors

    and switch

    Amp ADC

    Signal

    processorMonitor

    PrinterStorage

    LA

    LL

    RA

    RL

    Electrocardiography ( contd.)

    61

    Electrocardiography (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    62/90

    Fig. (4) Schematic representation of normal ECG

    Electrocardiography ( contd.)

    62

    Types of ECG Recordings

  • 5/28/2018 Electronic Instrumentation & Control Systems

    63/90

    yp g

    Bipolar Leads recordvoltage between electrodesplaced on wrists & legs(right leg is grounded)

    Lead Irecords between

    right arm & left arm

    Lead II: right arm & left leg

    Lead III: left arm & left leg

    Fig. (5)63

  • 5/28/2018 Electronic Instrumentation & Control Systems

    64/90

    Fig. (6)Einthovens triangle. Lead I is from RA to LA, lead II is from RA to

    LL, and lead III is from LA to LL.

    0IIIIII

    64

    Causes of Cardiac

  • 5/28/2018 Electronic Instrumentation & Control Systems

    65/90

    Cycle

    3 distinct waves are

    produced during cardiac

    cycle

    P wave caused by atrial

    depolarization

    QRS complex caused by

    ventricular depolarization T wave results from

    ventricular repolarization

    Fig. (7)65

    Elements of the ECG

  • 5/28/2018 Electronic Instrumentation & Control Systems

    66/90

    P wave: (Depolarization of both atria)

    Relationship between P and QRS helps distinguish various cardiac

    arrhythmiasShape and duration of P may indicate atrial enlargement

    PR interval:(from onset of P wave to onset of QRS)

    Normal duration = 0.120.2 sec

    Represents atria to ventricular conduction time (through His

    bundle)

    Prolonged PR interval may indicate a 1st degree heart block

    QRS complex:(Ventricular depolarization)Larger than P wave because of greater muscle mass of ventricles

    Normal duration = 0.08 - 0.12 sec

    66

    Elements of the ECG (-contd.)

  • 5/28/2018 Electronic Instrumentation & Control Systems

    67/90

    Its duration, amplitude, and morphology are useful in diagnosing

    cardiac arrhythmia, ventricular hypertrophy, Myocardial Infarction

    (MI), electrolyte derangement, etc.

    Q wave greater than 1/3 the height of the R wave, greater than

    0.04 sec are abnormal and may represent MI

    ST segment:

    Connects the QRS complex and T wave

    Duration of 0.08-0.12 secT wave:

    Represents repolarization or recovery of ventricles

    Interval from beginning of QRS to apex of T is referred to as the

    absolute refractory period

    QT Interval:

    Measured from beginning of QRS to the end of the T wave

    Normal QT is usually about 0.40 sec

    QT interval varies based on heart rate67

  • 5/28/2018 Electronic Instrumentation & Control Systems

    68/90

    Fig. (8)

    68

    Ultrasound System

  • 5/28/2018 Electronic Instrumentation & Control Systems

    69/90

    Ultrasound is one of the most widely used modalities in medical imaging,

    which is regularly used in cardiology, obstetrics, gynaecology, abdominal

    imaging, etc.

    Mostly, it is used in non-invasive techniques, although an invasive

    technique like intra-vascular imaging is also possible

    Ultrasound systems are signal processing intensive with various imaging

    modalities and different processing requirements in each modality, digital

    signal processors (DSP) are finding increasing use in such systems The advent of low power system-on-chip (SoC) with DSP and RISC

    processors is providing portable and low cost systems without

    compromising the image quality necessary for clinical applications

    The term ultrasound refers to frequencies that are greater than 20 kHz,

    which is commonly accepted to be the upper frequency limit the humanear can hear

    Typically, ultrasound systems operate in the 2 MHz to 20 MHz frequency

    range, although some systems are approaching 40 MHz for harmonic

    imaging 69

    Ultrasound System: Basic Funct ional i ty

  • 5/28/2018 Electronic Instrumentation & Control Systems

    70/90

    70

    Ultrasound System: Basic Funct ional i ty

  • 5/28/2018 Electronic Instrumentation & Control Systems

    71/90

    Fig.(1 ) shows the basic functionality of an ultrasound system, which

    demonstrates how transducers focus sound waves along scan lines

    in the region of interest

    In principle, the ultrasound system focuses sound waves along a

    given scan line so that the waves constructively add together at the

    desired focal point

    As the sound waves propagate towards the focal point, they reflect

    off on any object they encounter along their propagation path

    Once all of the sound waves along the given scan line have been

    measured, the ultrasound system focuses along a new scan line until

    all of the scan lines in the desired region of interest have been

    measured

    To focus the sound waves towards a particular focal point, a set of

    transducer elements are energized with a set of time-delayed pulses

    to produce a set of sound waves that propagate through the region of

    interest, which is typically the desired organ and the surrounding

    tissue 71

    Ultrasound System: Basic Funct ional i ty

  • 5/28/2018 Electronic Instrumentation & Control Systems

    72/90

    This process of using multiple sound waves to steer and focus a

    beam of sound is commonly referred to as beam-forming

    Once the transducers have generated their respective soundwaves, they become sensors that detect any reflected sound

    waves that are created when the transmitted sound waves

    encounter a change in tissue density within the region of interest

    By properly time delaying the pulses to each active transducer, the

    resulting time-delayed sound waves meet at the desired focal

    point that resides at a pre-computed depth along a known scan

    line

    The amplitude of the reflected sound waves forms the basis for the

    ultrasound image at this focal point location Envelope detection is used to detect the peaks in the received

    signal and then log compression is used to reduce the dynamic

    range of the received signals for efficient display and can be

    analysed by the doctor or technician72

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    73/90

    73

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    74/90

    The beam-former control unit, as shown in Fig. (2), is responsible for

    synchronizing the generation of the sound waves and the reflected

    wave measurements The controller knows the region of interest in terms of width and

    depth and gets translated into a desired number of scan lines and a

    desired number of focal points per scan line

    The beam-former controller begins with the first scan line and excites

    an array of piezo-electric transducers with a sequence of high-voltage

    pulses (of the order 100 V & 2 A) via transmit amplifiers

    The pulses go through a Tx/Rx switch, which prevents the high-

    voltage pulses from damaging the receive electronics

    Note that these high-voltage pulses have been properly time delayedso that the resulting sound waves can be focused along the desired

    scan line to produce a narrowly focused beam at the desired focal

    point

    74

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    75/90

    The beam-former controller determines which transducer elements to

    energize at a given time and the proper time delay value for each

    element to properly steer the sound waves towards the desired focalpoint

    As the sound waves propagate toward the desired focal point, they

    migrate through materials with different densities; with each change

    in density, the sound wave has a slight change in direction &

    produces a reflected sound wave

    Some of the reflected sound waves propagate back to the transducer

    & form the input to the piezo-electric elements in the transducer

    The resulting low voltage signals are scaled using a variable

    controlled amplifier (VCA) before being sampled by ADCs The VCA is configured so that the gain profile being applied to the

    received signal is a function of the sample time since the signal

    strength decreases with time (e.g., it has travelled through more

    tissue)75

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    76/90

    The number of VCA and ADC combinations determines the number

    of active channels used for beam-forming

    It is usual to run the ADC sampling rate 4 times or higher than thetransducer centre frequency

    Once the received signals reach the Rx beam-former, the signals are

    scaled and appropriately delayed to permit a coherent summation of

    the signals

    This new signal represents the beam-formed signal for one or more

    focal points along a particular specific scan line

    Once the data is beam-formed, depending on the imaging modes,

    various processings are carried out, e.g., it is common to run the

    beam-formed data through various filtering operation to reduce outband noise

    In B (Brightness) mode, demodulation followed by envelope detection

    and log compression is the most common practice

    76

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    77/90

    Several 2D noise reduction and image enhancement functions are

    also performed in this mode

    In spectral mode, a windowed Fast Fourier Transform (FFT) isperformed on the demodulated signal & displayed separately

    It is also common to present the data on a speaker after

    separation of forward and reverse flow

    In these systems, a repeated set of pulse is sent through thetransducer

    In between the pulses, the received signal is recorded

    There is an alternate mode where a continuous pulse sets are

    transmitted, which are known as continuous wave (CW) systems

    These systems are used where a more accurate measurement of

    velocity information is desired using Doppler techniques

    The disadvantage of this system is that it loses the ability to

    localize the velocity information 77

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    78/90

    In these systems, a separate set of transducers are used for

    transmission and reception

    Due to large immediate reflection from the surface of thetransducer, the dynamic range requirement becomes very high to

    use ADC to digitize the reflected ultrasound signal and maintain

    enough signal to noise (SNR) for estimating the velocity

    information

    Therefore, an analog beam-forming is usually used for CW

    systems followed by analog demodulation

    Such systems can then use lower sampling rate (usually in kHz

    range) ADCs with higher dynamic range

    78

    Ultrasound System: System Components

  • 5/28/2018 Electronic Instrumentation & Control Systems

    79/90

    79

    Ultrasound System: Imaging Modes

  • 5/28/2018 Electronic Instrumentation & Control Systems

    80/90

    A-mode (Amplitude) Imaging:

    It displays the amplitude of a sampled voltage signal for a single

    sound wave as a function of time This mode is considered 1D and used to measure the distance

    between two objects by dividing the speed of sound by half of the

    measured time between the peaks in the A-mode plot, which

    represents the two objects in question

    This mode is no longer used in ultrasound systems

    B-mode (Brightness) Imaging:

    It is the same as A-mode, except that brightness is used to represent

    the amplitude of the sampled signal

    B-mode imaging is performed by sweeping the transmitted soundwave over the plane to produce a 2D image

    Typically, multiple sets of pulses are generated to produce sound

    waves for each scan line, each set of pulses are intended for a

    unique focal point along the scan line 80

    CW (C ti W ) D l

    Ultrasound System: Imaging Modes

  • 5/28/2018 Electronic Instrumentation & Control Systems

    81/90

    CW (Continuous Wave) Doppler:

    In this mode, a sound wave at a single frequency is continuously

    transmitted from one piezo-electric element and a second piezo-

    electric element is used to continuously record the reflected soundwave

    By continuously recording the received signal, there is no aliasing in

    the received signal

    Using this signal, the blood flow in veins can be estimated using theDoppler frequency

    However, since the sensor is continuously receiving data from

    various depths, the velocity location cannot be determined

    PW (Pulse Wave) Doppler:

    For this several pulses are transmitted along each scan line and the

    Doppler frequency is estimated from the relative time between the

    received signals

    Since pulses are used for the signaling, the velocity location can also

    be determined81

    Ultrasound System: Imaging Modes

  • 5/28/2018 Electronic Instrumentation & Control Systems

    82/90

    Color Doppler:

    For this, the PW Doppler is used to create a color image that is super-

    imposed on top of B-mode image

    A color code is used to denote the direction and magnitude of the flow,

    e.g., red typically denotes flow towards the transducer and blue denotes

    flow away from it

    A darker color usually denotes a larger magnitude while a lighter color

    denotes a smaller magnitudePower Doppler:

    In this, instead of estimating the actual velocity of the motion, the

    strength or the power of the motion is estimated and displayed

    It is useful to display small motion and there is no directional information

    in this measurement

    Spectral Doppler:

    Itshows the spectrum of the measured velocity in a time varying manner

    Both PW & CW Doppler systems are capable of showing spectral

    Doppler

    82

    Ultrasound System: Imaging Modes

  • 5/28/2018 Electronic Instrumentation & Control Systems

    83/90

    M-mode:

    This display refers to scanning a single line in the object and then

    displaying the resulting amplitudes successively, which shows themovement of a structure such as a heart

    Because of its high pulse frequency (up to 1000 pulses per second),

    this is useful in assessing rates and motion and is still used

    extensively in cardiac and fetal cardiac imaging

    Harmonic Imaging:

    It is a new modality where the B-mode imaging is performed on the

    second (or possibly other) harmonics of the imaging

    Due to the usual high frequency of the harmonic, these images have

    higher resolution than conventional imaging, however, due to higherloss, the depth of imaging is limited

    Some modern ultrasound systems switch between harmonic and

    conventional imaging based on depth of scanning83

    Ultrasound System: Imaging Modes

  • 5/28/2018 Electronic Instrumentation & Control Systems

    84/90

    This system imposes stringent linearity requirements on the signal

    chain components

    Elasticity/Strain Imaging: It is a new modality where some measures of elasticity (like Youngs

    modulus) of the tissue (usually under compression) is estimated and

    displayed as an image

    These types of imaging have been shown to be able to distinguish

    between normal and malignant tissues

    This is currently a very active area of research both on clinical

    applications and in real-time system implementation

    84

    Basic Ultrasound Machine

  • 5/28/2018 Electronic Instrumentation & Control Systems

    85/90

    Basic Ultrasound Machine Components:

    Central Processing Unit (CPU)

    Transducer probe

    Transducer Pulse Controls

    Display

    Keyboard/Cursor

    Disk Storage

    Printers

    85

    Wh t i EEG?

  • 5/28/2018 Electronic Instrumentation & Control Systems

    86/90

    What is an EEG? An electroencephalogram is a measure of the brain's

    voltage fluctuations as detected from the electrodes.

    It is an approximation of the cumulative electrical

    activity of neurons.

    Background 1875 - Richard Caton discovered electrical

    properties of exposed cerebralhemispheres of rabbits and monkeys.

    1924 - German Psychiatrist Hans Bergerdiscovered alpha waves in humans and

    coined the term electroencephalogram 1950s - Walter Grey Walter developed

    EEG topography - mapping electricalactivity of the brain.

    H B i

  • 5/28/2018 Electronic Instrumentation & Control Systems

    87/90

    Human Brain

    Frontal LobesPersonality, emotions, problem solving.

    Parietal lobesCognition, spatial relationships andmathematical abilities, nonverbal

    memory.

    Occipital lobesVision, color, shape and movement.

    Temporal lobes

    Speech and auditory processing,language comprehension, long-termmemory.

    Diff t i EEG

  • 5/28/2018 Electronic Instrumentation & Control Systems

    88/90

    Different waves in EEGSlowest but highest

    amplitude waves,deepest stages of sleep

    it tends to appear during

    drowsy, meditative, or

    sleeping states.

    Predominantly originates

    From occipital lobe during

    wakeful relaxation with

    closed eyes.

    associated with active, busy,

    or anxious thinking andactive concentration.

    relate to neural consciousnes

    via the mechanism for

    conscious attention

    P bl ith EEG

  • 5/28/2018 Electronic Instrumentation & Control Systems

    89/90

    Problems with EEG

    Electrical activity generated by complex systemof billions of neurons.

    Difficult to register electrode location.

    Artifacts from motion, eye blinks, swallows, heartbeat, sweating

    Food, age, time of day, fatigue, motivation of

    subject.

    Advantages of EEG Many EEG studies have reported reproducible

    changes in brain dynamics that are task dependent!

    People are able to control their brainwaves via

    biofeedback!

  • 5/28/2018 Electronic Instrumentation & Control Systems

    90/90

    Fig.Basic structure of the heart. RA is the right atrium, RV is the right

    ventricle; LA is the left atrium, and LV is the left ventricle. Basic pacing rates