Page 1: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

1

prepFASTThe Next Step in Automation

I C P | I C P M S | A A

• Autodilution • Autocalibration • Intelligent QC Dilution

Elemental Scientific

Page 2: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

2

prepFAST Inline Syringe Dilution

prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions for samples and standards. Capable of up to 400x dilution, the prepFAST is the fastest, simplest way to ensure high quality data in every run.

S400V Syringe Pump

The precise (<± 0.05%) and accurate (<± 0.2%) S400V syringe pump provides smooth and balanced delivery of solutions over a wide range of flow rates (1 to 40,000 μL/min) to ensure rapid reliable inline dilutions.

Four Syringes: 1. Rinse 2. Carrier 3. Diluent 4. Internal Standard

Syringe Dilution• Clean• Chemically Resistant - Organics - Strong Acids• Low Maintenance• Accurate and Precise• Long-Term Stability

prepFAST Automation

Page 3: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

3

Benefits• Fully Automated - Autocalibrate - Autodilute - QC Autodilute• Rapid uptake, stabilization and washout• Inline dilution• Integrated with ICP/ICPMS Software

Valve Injection• FAST Stabilization• FAST Washout• Low Carryover

P7+ Valve

autosamplermovement

autosamplermovement

sampleflush

readdelay

measurement

measurement

wash

wash

overhead

overheadreaddelaysampleuptake

0s 200s100s 300s

Normal Analysis

FAST AnalysisSix Steps in a Standard Analysis1. Autosampler Movement2. Uptake/Dilution3. Stabilization4. Measurement5. Wash6. Overhead

Normal Analysis

prepFAST Analysis

Page 4: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

4

Inline Dilutions

Rinse

Vacuum

3. Diluted sample is injected and sample loop is cleaned

Carrier

To ICPMS

Carrier

S4: InternalStandard

S3: Diluent

S2: Carrier

2. Syringes add internal standard and dilute sample into second loop

Waste

To ICPMS

S1: Rinse

4. Sample and dilution loops are cleaned

Waste

Carrier

To ICPMS

Sample

Vacuum

1. Vaccum load sample into loop

Carrier

To ICPMS

Automated Inline Autodilution

*Patent Pending

Page 5: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

5

Certified Reference Material

Figure 2. Omaha tap water is used as the EPA 200.8 defined, Laboratory Fortified Matrix (LFM) and Quality Control Standard (QCS). Recovery of 95-105% greatly exceeds EPA requirements of 70-130%.

< 1% RSD over 14 hours

Figure 1. prepFAST autodilution stability throughout a run for a wide range of dilution factors.

Accurate Spike Recovery

FAST, Stable and Accurate

Page 6: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

6

Autocalibration

Figure 3. Illustration of conventional calibration vs. prepFAST autocalibration function using a single multi-element standard. The full calibration curve is created from vial 2 in the standards rack. Dilution occurs rapidly using small volumes at high flow rates. Sample is then injected at the user's desired flow rate.

Conventional Calibration vs prepFAST Autocalibration

Figure 4. Predefined dilution factors for a single multi-element standard are used to build calibration curves. Accuracy of dilution results in high linearity (R2 = 0.999997).

Autocalibration

R2 = 0.999997

prepFAST Autocalibration(seven points)

Inline Prep: 1 Blank + 1 Standard

STDPosition

InlineDilutionFactor Std Diluent Total Concentration

10

222

2

2

2

2

2

200x

200x100x

40x

10x

5x

2.5x

1x

50

50

100

250

1000

2000

4000

10000

9950

9950

9900

9750

9000

8000

6000

0

10000

10000

10000

10000

10000

10000

10000

10000

0

2.5

5

12.5

50

100

200

500

prepFAST Calibration: Blank can be analyzed diluted or undiluted.

BLANK

1

2

3

4

5

500 ppb

6

7

8

9

100

2.5

5

12.5

50

100

200

500

10

1

2

3

4

5

6

7

BLANK

1

2

3

4

5

6

7

8

9

10

STDPosition Concentration

Conventional Calibration(seven points)

Offline Prep: 1 Blank + 7 Standards

500 ppb

200 ppb

100 ppb

50 ppb

12.5ppb

5 ppb

2.5 ppb

Dilution Rate

Page 7: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

7

ICH "J" Calibration Coefficients

Figure 6. Multiple calibration curves for the USP elements have good linearity for drugs with a large range of daily doses. (ICH = International Conference on Harmonisation)

Figure 5. USP protocols define a tight calibration range for each drug. In this example multiple drug specific calibration curves are automatically generated from a single stock standard.

Multiple Curves from a Single Stock Standard

1.000

Page 8: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

8

Manual vs. Inline Sample Dilution

Figure 8. Mixing inline eliminates the need to premix/dilute samples in discrete tubes prior to analysis. This cost effective approach improves sample throughput, saves time, and reduces labware and reagent consumption. After in-valve dilution, the sample is injected (400 µL/min) at the user's desired flow rate.

Inline Autodilution

Load Sample

Autodiluteand Analyze

Dilution Correction

Report Data

QCPASS

ICP/ICPMS SOFTWARE

Figure 7. Flowchart illustrates the workflow logic for prescriptive sample dilution.

Page 9: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

9

Sample Stability

0.00  

0.10  

0.20  

0.30  

0.40  

0.50  

0   60   120   180   240   300   360   420   480  

Concen

tra)

on  Fou

nd  (p

pm)  

Time  (Minutes)  

Rh  concentra)on  in  700  samples  over    8  hours.  RSD  ≤  2%  

50% Aqua Regia20x Inline autodilution

88 sample/hr

Figure 9. Fire assay digests for precious metals in Aqua Regia require dilution before analysis. prepFAST inline 20x autodilution eliminates the manual dilution step.

Figure 10. Precise and accurate results for concentrations of metals in a Trace Metals in Water Certified Reference Material (CRM, NIST 1643e) are obtained with prepFAST autocalibration and autodilution.

Precise and Accurate Autodilution

Page 10: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

10

Intelligent Autodilution

QCFAIL

Load Sample

Rerun with Intelligent Dilution

INTELLIGENT DILUTION

Load Sample

Autodiluteand Analyze

Dilution Correction

Report Data

QCPASS

ICP/ICPMS SOFTWARE

Automated QC Dilutions

Figure 11. Flowchart illustrates the workflow logic for intelligent autodilution of samples that fail QC.

Page 11: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

11

Calibration Over Range

Figure 12. prepFAST autodilution in combination with ICPMS QC software can eliminate in-run QC failures. After a QC failure is detected, ICPMS software immediately inserts a new sample line with an intelligent dilution factor. The prepFAST automatically reruns the sample at the new intelligent dilution factor. Advanced prepFAST QC functions are currently available on select instruments.

prepFAST Washout

Figure 13. For a selection of elements, using the prepFAST, washout greater than 20,000x is achieved in 10 seconds. This is more than sufficient to eliminate any carry-over from the high concentration QC over range sample onto the intelligently diluted re-run.

Page 12: Elemental Scientific prepFAST - Separation · 2 prepFAST Inline Syringe Dilution prepFAST is an autodilution system that automatically performs precise and accurate inline dilutions

12

©Elemental Scientific | 7277 World Communications Drive | Omaha, NE 68122p: 402-991-7800 | f: 402-991-7799 | [email protected] | [email protected]

www.prepFAST.com | www.icpms.com

Contact us by phone at 402.991.7800 or by e-mail at [email protected]. Our scientists and engineers are available to answer your questions related to elemental analysis. We are