ELISA Technical Guide Corporate Headquarters IDEXX Laboratories, Inc. One IDEXX Drive, Westbrook, Maine 04092 USA Tel: 1-207-556-4890 or 1-800-548-9997 Fax: 1-207-556-4826 or 1-800-328-5461 European Headquarters IDEXX Europe B.V. Koolhovenlaan 20, 1119 NE Schiphol-Rijk The Netherlands Tel: 31-20-655-23-00 or 00-800-727-43399 Fax: 31-20-655-23-33 HerdChek, FlockChek and xChek are trademarks or registered trademarks of IDEXX Laboratories, Inc. in the United States and/or other countries. © 2007 IDEXX Laboratories, Inc. All rights reserved. • 09-65846-00 (2) idexx.com

Upload: mueberra-ayazoglu

Post on 28-Nov-2014

187 views

Category:

Documents


6 download

TRANSCRIPT

Page 1: Eliza Testi

ELISA Technical Guide

Corporate Headquarters IDEXX Laboratories, Inc.One IDEXX Drive, Westbrook, Maine 04092 USATel: 1-207-556-4890 or 1-800-548-9997Fax: 1-207-556-4826 or 1-800-328-5461

European Headquarters IDEXX Europe B.V. Koolhovenlaan 20, 1119 NE Schiphol-Rijk The NetherlandsTel: 31-20-655-23-00 or 00-800-727-43399Fax: 31-20-655-23-33

HerdChek, FlockChek and xChek are trademarks or registered trademarks of IDEXX Laboratories, Inc. in the United States and/or other countries.

© 2007 IDEXX Laboratories, Inc. All rights reserved. • 09-65846-00 (2)i d e x x . c o m

Page 2: Eliza Testi

Introduction 1

ELISATechnology 2

ELISAKits 3

ELISAEquipment 5

EquipmentMaintenanceandCalibration 6

ReagentHandlingandPreparation 7

HandlingandPreparingKitComponents 8

QualityControl 9

SampleHandling 10

PipettingMethods 12

ELISAPlateTiming 15

ELISAPlateWashing 16

ReadingPlatesandDataManagement 18

ELISATroubleshooting 19

Appendixes

A:GravimetricPipetteCalibrationProcedure 23

B:FlockChek*andHerdChek*InventoryControlTrackingChart 24

C:LaboratoryTrackingChart 25

D:MaintenanceandCalibrationSchedule 26

E:QualityControlQuickCheck 27

ELIS

A Technical G

uide

Page 3: Eliza Testi

ELI

SA

Tec

hnic

al G

uid

e

Page 4: Eliza Testi

ELIS

A Technical G

uide

ELISAIDEXXLaboratoriesmanufacturesdiagnostictestkitsforthedetectionofdiseasesinruminants,horses,swineandpoultry.

Theenzyme-linkedimmunosorbentassay(ELISA)isoneofthemostsensitiveandreproducibletechnologiesavailable.Theseassaysarerapid,simpletoperformandeasilyautomated.IDEXXintroducedthefirstcommercialpoultryELISAforinfectiousbursaldisease(IBD)in1985andthefirstcommerciallivestockELISAforAujeszky’sdisease/pseudorabiesin1986,enhancingthewaylaboratoriestestproductionanimalstoday.

Aswithanyassay,thereproducibilityandreliabilityofELISAsaredependentuponpropertechniqueandattentiontodetail.ThisELISAtechnicalguidewillincreaseyourawarenessofELISAtechniquesandhelpyoumaintainproficiencywiththismethodology.

Checkyourpackageinsertforspecificinstructionsforeachassayyouperform.Periodically,improvementsandrevisionsaremadetokitandpackageinserts.Therefore,itisimportanttoreviewtheprotocolonaregularbasis.Ifyouhavequestionsconcerninganyofthefollowinginformation,callIDEXXTechnicalServicesforassistance.WithintheUnitedStates,call1-800-548-9997or1-800-943-3999.OutsideoftheUnitedStates,call1-207-556-0895andselecttheProductionAnimalServicesTechnicalSupportoption.OrvisitourWebsiteatwww.idexx.com/production/contact/contactpas.jsp.

Introduction

Page 5: Eliza Testi

ELI

SA

Tec

hnic

al G

uid

e

ELISATheELISAisarapidtestusedfordetectingandquantifyingantibodiesorantigensagainstviruses,bacteriaandothermaterials.Thismethodcanbeusedtodetectmanyinfectiousagentsaffectingpoultryandlivestock.

InELISAtechnology,thesolidphaseconsistsofa96-wellpolystyreneplate,althoughothermaterialscanbeused.Thefunctionofthesolidphaseistoimmobilizeeitherantigensorantibodiesinthesample,astheybindtothesolidphase.Afterincubation,theplatesarewashedtoremoveanyunboundmaterial.Conjugateisthenaddedtotheplateandallowedtoincubate.

Theconjugateconsistsofeitheranantigenorantibodythathasbeenlabeledwithanenzyme.Dependingupontheassayformat,theimmunologicallyreactiveportionoftheconjugatebindswitheitherthesolidphaseorthesample.Theenzymeportionoftheconjugateenablesdetection.

Theplatesarewashedagainandanenzymesubstrate(hydrogenperoxideandachromogen)isaddedandallowedtoincubate.ColordevelopsinthepresenceofboundenzymeandtheopticaldensityisreadwithanELISAplatereader.

ELISA FormatsELISAsaredividedintothreemainformats—indirect,blocking(competitive)andantigen-capture(direct).

Indirect FormatIntheindirectformat,thesampleantibodyissandwichedbetweentheantigencoatedontheplateandanenzyme-labeled,anti-speciesglobulinconjugate.Theadditionofanenzymesubstrate-chromogenreagentcausescolortodevelop.Thiscolorisdirectlyproportionaltotheamountofboundsampleantibody.Themoreantibodypresentinthesample,thestrongerthecolordevelopmentinthetestwells.Thisformatissuitablefordeterminingtotalantibodylevelinsamples(Newcastlediseasevirus,B. abortus,etc.).

ELISA Technology

ELISA formats provide the ability to: • Test a large number of

samples at the same time

• Automate the procedure using robotics or other types of automated equipment

• Computerize the calculation and reporting of results

Indirect ELISA

Steps in an Indirect ELISA

Page 6: Eliza Testi

ELIS

A Technical G

uide

ELISABlocking (Competitive) FormatInthisformat,thespecificsampleantibodiescompetewith,orblock,theenzyme-labeled,specificantibodyintheconjugate.Theadditionofanenzymesubstrate-chromogenreagentcausescolortodevelop.Thiscolorisinverselyproportionaltotheamountofboundsampleantibody.Themoreantibodiespresentinthesample,thelesscolordevelopmentinthetestwells(CAV,CSFVAb,etc.).

Antigen-Capture (Direct) FormatIntheantigen-captureformat,theantigeninthesampleissandwichedbetweenantibodiescoatedontheplateandanenzyme-labeledconjugate.Theantibodyconjugatecanbeeithermonoclonalorpolyclonal.Theadditionofanenzymesubstrate-chromogenreagentcausescolortodevelop.Thiscolorisdirectlyproportionaltotheamountofthetargetantigenpresentinthesample(CSFVAg,LLAg,etc.).

Steps in Blocking ELISA

Blocking ELISA

ELISA Kits

AnELISAkitisasetofstandardizedreagentsandmicrowellplatesmanufacturedforaspecifictest.IDEXXELISAkitsmaycontainsomeorallofthefollowingcomponents:coatedplates(solidand/orstripplates),samplediluent,controls,washconcentrate,conjugate,substrateandstopsolution.Thekitsaremanufacturedinbatchesorlots.Eachcomponentofeachkitlotisoptimizedandmanufacturedtoworkasaunit.Thekitspassmanyquality-controltestsconductedbyIDEXX,numerousworldwidereferencelaboratories,and/ortheUSDAbeforetheyareapprovedandreleasedforsale.

NOTE:Donotmixorusecomponentsfromdifferentkitlotnumbers.

Page 7: Eliza Testi

ELI

SA

Tec

hnic

al G

uid

e

ELISACoated Plates The96-wellplatesaremadeofpolystyreneandcoatedwitheitherinactivatedantigenorantibody.Thiscoatingisthebindingsitefortheantibodiesorantigensinthesample.Unboundantibodiesorantigensinthesamplearewashedawayafterincubation.

Forsomeassays,normalhostcell(NHC)antigensandpathogen-specificantigensarecoatedinalternatingrowsorcolumns.TheNHCantigensareusedtoassesswhetherantibodiesagainsttissueculturecomponentspresentinvaccinesarecontributingtosamplereactions.Whenusingtheseplates,eachsampleisaddedtobothasamplewellandanNHCwell.

Sample DiluentMostassaysrequireaspecificdilutionofthesample.Samplesareaddedtothesamplediluentandmixedpriortoputtingthemontothecoatedplates.

ControlsThepositivecontrolisasolutionthatcontainsantibodyorantigen.Thenegativecontrolisasolutionwithoutantibodyorantigen.Thecontrolshelptonormalizeorstandardizeeachplate.Controlsarealsousedtovalidatetheassayandtocalculatesampleresults.Inmostkits,thecontrolsarepredilutedandreadytouse.Besuretofollowtheinstructionsinthepackageinsert.

ConjugateELISAconjugatesareenzyme-labeledantibodiesorantigensthatreactspecificallytoplate-boundsampleanalytes.Unboundconjugateiswashedawayafterincubationandbeforetheadditionofsubstrate.Theopticaldensityofthecolorimetricsubstrateisdirectlyproportionaltothequantityofboundenzymepresent.

SubstrateForperoxidaseconjugates,thesubstrateisamixtureofhydrogenperoxideandachromogenthatreactswiththeenzymeportionoftheconjugatetoproducecolor.

Wash ConcentrateThewashconcentrateisabufferedsolutioncontainingdetergentusedtowashawayunboundmaterialsfromtheplates.

Stop SolutionThestopsolutionstopstheenzyme-substratereactionand,thereby,thecolordevelopment.

ELISA Kits continued

IDEXX kits are manufactured in batches or lots according to strict quality standards. Each component or reagent in a kit lot is optimized to work with the other reagents contained in the kit. This includes measurements of sensitivity, specificity and repeatability. Therefore, it is very important not to mix reagents from different kit lots.

All kit components have an expiration date.

Page 8: Eliza Testi

ELIS

A Technical G

uide

ELISAEquipmentforELISAtestingiswidelyavailable.Readers,washersandpipettesareavailableasmanualorautomatedsystems.Someofthefactorsaffectingequipmentselectionarethenumberandtypesoftestsandsamples,technicaltrainingofstaffandfinancialconsiderations.BelowisabriefoutlineofsomeequipmentavailableforperformingELISAtesting.

Pipettes• Single-channel,fixed-volumeandadjustable-volume(1–20µL,

10–100µL,20–200µL,etc.)

• Multichannel,8-and12-channels

• Semi-automateddispensingunits

• Fullyautomatedsystemsthatcanprocessmultipleplates

Dilutors• Single-channel

• Multichannel

• Automateddispensingunits

Washer Systems• Manualsystemsthatwashoneroworcolumnatatime

• Semi-automatedsystemsthathandleonestriporplateatatime

• Fullyautomatedsystemsthatcanprocessmultipleplates

ELISA Plate Readers• Manualreadersthatreadoneroworwellatatime

• Semi-automatedreadersthatreadoneplateatatime

• Fullyautomatedsystemsthatcanprocessmultipleplatessimultaneously

Other• Humiditychamber(notrequiredforallELISAtests)

• Platesealersforassaysthathavelongincubationtimes(toavoidevaporation)

ELISA Equipment

There is a large selection of equipment available. When purchasing a plate reader, call IDEXX Technical Services to make sure the xChek* software has the proper interface.

Multichannel pipette and single-channel pipette

Semi-automated wash system

Manual wash system

Plate reader

Page 9: Eliza Testi

ELI

SA

Tec

hnic

al G

uid

e

ELISAThemaintenanceandcalibrationofyourlaboratoryequipmentisextremelyimportantinobtainingaccurateandreproducibleresults.

TheMaintenanceandCalibrationSchedule(AppendixD)shouldbeusedasaguideline.Adjustmaintenanceroutinesaccordingtotheamountofdailytestingperformedinyourlaboratory.Alwaysrefertoyourequipmentmanufacturer’sguidefortheirrecommendations.

Calibration ProtocolsEquipmentalwaysneedstobeinpropercalibration.Equipmentthatisoutofcalibrationcanproducefalseorinaccurateresults.RefertotheMaintenanceandCalibrationSchedule(AppendixD)andyourmanufacturer’sinstructionsforthepropercalibrationprotocolandrequiredfrequency.

Options for Calibrating Pipettes• PerformthegravimetricmethodoutlinedinAppendixA.

• UseacommercialautomatedcalibrationsystemlikethePCS®‚producedbyArtel.SeeAppendixA,GravimetricPipetteCalibrationProcedure,formoreinformation.

• Sendthepipettetothemanufacturer;seeyourowner’smanualforinstructions.

• Sendthepipettetoapipettecalibrationservice.

Sendingpipettesoutforserviceisbeneficialwhenrepairormaintenanceisnecessary.However,thispracticeprovidesonlyalimitedlevelofqualitycontrol,whichcanbeincreasedwithin-housecalibration.

Operatortechniqueandlaboratoryenvironmentaretwocriticalvariablesthatdeterminehowapipettewillperformwhenusedonyourbenchtop.Athoroughquality-controlprogrammustincludeaquantitativeaccountoftheseeffects.Itisbeneficialtohaveamethodinplacethatallowsyoutoperformregular,routineperformanceverificationsonyourownpipettes.Bydoingso,youwillbeabletotrackpipettesthataredriftingoutoftolerance.Whenthishappens,thefailingpipetteshouldbesentoutforcorrectivemaintenanceorrepairbyaqualifiedservicebeforeitcompromisesyourlaboratorydataandproductivity.

Equipment Maintenance and Calibration

Be sure to label your pipettes with the calibration date and keep a log for the calibration and maintenance of all your equipment.

Pipette with calibration label

Page 10: Eliza Testi

ELIS

A Technical G

uide

ELISAReceiving KitsWhenyoureceiveyourELISAkits,recordthedateonyourInventoryControlTrackingChart(AppendixB)andonthekitboxes.Inspectthemfordamageandstorethemat2°–7°C.Whenusingkitsfromyourinventory,usethefirst-in-first-out(FIFO)method.Inotherwords,usethekitsthataretheoldest(orwillexpire)first.Individualkitcomponentsmayhavelongerexpirationdatesthantheactualdateontheouterkitbox.However,youshouldgobytheexpirationdateontheouterlabelofthekit.Ifyoudonotuseanentirekit,markthedateitisopenedandeachtimeitisusedthereafter.Thisway,youcankeeptrackofhowmanytimesitmovesorcyclesfromtherefrigeratortoroomtemperature.Keepthenumberofcyclestoaminimumbybatching(oraccumulating)samplesintolargergroupswheneverpossible.

General Reagent HandlingBesuretocheckyourpackageinsertforguidelinesonhandlingandpreparingreagents.Sometestkitsrecommendthatallreagentsandplatesbebroughttoroomtemperature(18°–25°C)priortouse;othersindicatethatonlyspecificreagentsbebroughttoroomtemperature.Whenyouneedtobringakittoroomtemperature(18°–25°C),takeitoutoftherefrigeratorandtakethekitcomponentsoutofthekitboxatleast2–3hoursbeforebeginningtheassay.

Measureallreagentsusingsterileorcleanvessels.Becarefultomeasureonlywhatisneededforthenumberofplatesbeingrun.Thiswillhelptomaintaintheintegrityofthereagents.Donotreturnreagentstotheoriginalstockbottles.Westronglyrecommendusingdisposablepipettesandreservoirswhenhandlingreagentstominimizetheriskofcontamination.However,ifyouchoosetoreuseanydisposabledevice,useaseparatereservoirforeachreagentandbesuretolabelthem.Also,washandthoroughlyrinsethewellswithdeionizedordistilledwateraftereachuse.Changeanddiscardthedisposablereservoirsasfrequentlyaspossible.Neverusethesamereservoirforconjugateandsubstrate,evenifithasbeenwashed.

Reagent Handling and Preparation

The contamination of reagents may compromise your test results. Labeling your reagent reservoirs and using a separate one for each reagent will help minimize the risk.

Label your kit with the date it was received.

Labeled reservoirs

Page 11: Eliza Testi

ELI

SA

Tec

hnic

al G

uid

e

ELISAPlatesMostassayplatesareprovidedinresealablebagsthatincludeadesiccant.Insomekits,theplatesarepackagedindividually.Whenopeningabagthatcontainsseveralplates,removeonlythenumberofplatesyouneed,resealthebagandreturnittotherefrigeratorforstorage.Ifpossible,theplatesthatweretakenfromtheoriginalbagshouldbestoredinanotherbagwithdesiccantsinit.Ifapartialplateisused,aspiratealltheliquidfromtheusedwellsandcoverthemwithsealingtape.Placeseveraldesiccantsinthebag.Besuretomarktheplateaccordinglyandstoreitseparatelyfromunusedassayplates.Somekitshavetheplatesformattedintostrips.Werecommendlabelingeachstripbeforeusetopreventconfusion.

Sample Diluent and Wash ConcentrateMakesurethesamplediluentandwashconcentratehavecometoroomtemperature(18°–25°C)beforeuse.Theseareusuallythelargestbottlesinakitandrequirethemosttimetoequilibrate.Ifthewashconcentratestillshowscrystalformationafterreachingroomtemperature,mixitbyinvertingitseveraltimes.

ControlsMostkitsareformulatedwithpredilutedcontrols.However,somerequirethatyoudilutetheminthesamemannerasyoursample.Controlsshouldbeaddedtotheplateinthesamemethodandatthesametimeasthesamples.

ConjugateIfthekitrequiresyoutopreparea“working”conjugatesolution,besuretofollowtheinstructionsclosely.Prepareonlywhatyouimmediatelyneed,anddonotsaveleftoversolutionforfutureuse.Ifconjugatesarecontaminatedorimproperlystored,theymayloseenzymaticactivityormayhaveanapparentincreaseinbackground.Mostkitssupplyaready-to-useconjugate.

SubstrateOurELISAtestkitsincludeaready-to-usesubstrate.Thechemicalactivityofthesubstratewillbecompromisedifitisexposedtolightorcomesintocontactwithmetal.Protectthissolutionbystoringitinadarkcontaineruntilreadyforuse.

Stop SolutionBesuretousethestopsolutionincludedinthekit.Followanysafetyprecautionsinthepackageinsert.Thestopsolutionshouldbeatroomtemperaturebeforeuse.Ifthestopsolutionshowscrystalformationafterreachingroomtemperature,mixitbyinvertingseveraltimes.Thestopsolutionmaycrystalizeatlowertemperatures.Beforeuse,makesurethatitiscompletelydissolvedandappearsclear.

Handling and Preparing Kit Components

Refer to the package insert for specific details on the kit you are using.

Do not exchange components between kit lot numbers, even if kits are of similar type. Test results may be severely and adversely affected.

Seal, label and store partially used plates in a bag with desiccant.

Page 12: Eliza Testi

ELIS

A Technical G

uide

ELISAIn-House ControlsWerecommendusingin-houseassaycontrolstomonitoryourELISAtechniquesandkitperformanceovertime.

Becauseseraaregenerallyreceivedinsmallquantities,controlswillneedtobemadebypoolingsamples.Collectnegativeandpositivesamplesseparately.Whensufficientquantitiesofeachhavebeencollected,poolsimilarsamplestogether.Mixthepooledsamplesthoroughly.Insmallquantities,performtheserialdilutionofpositiveserainnegativeserum.Assayeachdilutionaccordingtostandardkitprotocol(samesampledilutionasdescribedinthekitpackageinsert).Selectthedilutionthatismostcomparabletothesample-to-positive(S/P)orsample-to-negative(S/N)valuesthatyouwanttomonitor.Makelargequantitiesofthatdilution.Prefilterthepreparedcontrolsusinga0.45-micronfiltermembrane;youmaythenchoosetofilterwitha0.20-micronfiltermembrane(optional).Putasmallamountofthefreshlymadein-housecontrol(volumeenoughfor1–2tests)intoairtightvials,label,dateandstorefrozenat-70°Cifpossible.Keeparecordofthispreparationinanotebookforreference.

Tousethiscontrol,thaw,mixanddiluteitinthesamemannerasaroutinesample.Runitoneveryplatenexttothekitcontrols.Donotrefreezeyourin-housecontrol.Youcanstoreitfor3–5daysat4°C.

RecordyourresultsontheLaboratoryTrackingChart(AppendixC)andgraphthem.Anyvariationsortrendsshouldalertyoutoreviewyourtechniqueandquality-controlmeasures.

Monitoring TemperatureELISAtestsaresensitivetotemperatureextremes.Trytomaintainalaboratorytemperatureof18°–25°C.Avoidrunningassaysunderornearairventsasthismaycauseexcessivecooling,heatingand/orevaporation.Also,donotrunassaysindirectsunlightasthismaycauseexcessiveheatandevaporation.Coldbenchtopsmayaffectyourassayandshouldbeavoidedbyplacingseverallayersofpapertowelsorsomeotherinsulatingmaterialundertheassayplatesduringincubation.

Recordandtrackthetemperatureduringeachassay.Ifyourlaboratory’stemperaturefluctuatesfrommorningtoafternoon,recordthisonyourtrackingchart.Ifyouhaveconditionsthataredifficulttocontrol,itisagoodideatouseatemperaturecontrolchambertoincubateyourplates.UsingELISAplatecoverswillhelpcontrolevaporationandaccidentalspills.

Quality Control CheckUsetheQualityControlQuickCheck(AppendixE)totroubleshootanyproblems.

Quality Control

To help troubleshoot questionable results, record and graph the laboratory temperature.

IBV Control Tracking September

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

0.4

Day

Positive Control Negative Control

Co

ntr

ol/V

alu

es

in O

ptical D

en

sity

1 110 5 7 9 13 15 17 19 21 20 25 27 29

Tracking chart for controls

Laboratory temperature tracking chart

Page 13: Eliza Testi

�0

ELI

SA

Tec

hnic

al G

uid

e

ELISAIncoming Sample QualitySamplequalitycanhaveasignificantimpactonfinalassayresults.Mostlabshavenochoiceregardingthequalityofincomingsamples.Inmanycases,thesamplediluentformulationcompensatesforvariationsinsamplequality.

Grossfungalorbacterialcontaminationcanhaveadverseeffectsontheantibodyorotherproteincomponentsofasampleandmayhaveanundesirableeffectontestresults.Ifsamplequalityishighlyquestionable,obtainingafreshsampleisstronglyadvised,whenpossible.

Serum/Plasma SamplesSerumsampleswithtracehemolysis(light-redcolor)andmoderatelipemia(milkyappearance)mayhavelittleornoeffectonELISAresults.Avoidusingsamplesthatareheavilyhemolyzed(dark-redcolor)orgrosslylipemic.Checkyourpackageinsertforinformation.Whenserumisontheclot,becarefulnottoaspirateanyoftheclottedmaterialorbloodcells.

Meat Juice SamplesMeatjuicesamplesshouldbeascleanaspossible.Removedebrisandlipidsfromthesamplewhenpipetting.

Milk SamplesWholemilksamplescanbeusedaftercentrifugationfor15minutesat2000xg,orleftovernightifrefrigerated(2°–8°C).Thesampleintendedfortheassayshouldbedrawnfrombelowthecreamlayer.

Egg Yolk SamplesCollectsampleswithacleantuberculinsyringeandmixthedilutedsamplesthoroughlybyvortexing.

Other Sample TypesRefertoyourpackageinsertforsamplehandling,preparationandstorageofothersampletypes(e.g.,albumin,cloacalswabs).

Meat juice sample Take sample from the area indicated.

Blood sample Take sample from the area indicated.

Milk sample Take sample from the area indicated.

Sample Handling

Page 14: Eliza Testi

��

ELIS

A Technical G

uide

ELISA

Light hemolysis

Dark hemolysis

Unmixed thawed sample Proteins settled on bottom of tube; mix prior to taking sample

Avoid numerous freeze-thaw cycles, as this may damage the antibodies or antigens in the sample. We recommend no more than 3–5 cycles.

Storing SamplesBesuresamplesareproperlystored.Ingeneral,serumsamplesshouldberefrigeratedat2°–7°Cfornomorethan3–5days.Ifsamplesneedtobestoredforalongerperiod,theyshouldberemovedfromtheclotandfrozentoaminimumof-20°C.Makesureallstoredsamplesareproperlylabeledandsealedtopreventevaporation.Evidenceoflyophilization(concentrationofthesample)canbeseenascrystallizationandiscommoninself-defrostingfreezers.Thisshouldbeavoidedbecausetheintegrityofthesamplewillmostlikelybecompromised.

Using Frozen SamplesFrozensamplescanbethawedatroomtemperatureorinarefrigerator.Allthawedsamplesneedtobethoroughlymixedpriortodilutiontoensurethattheproteinsaredispersedthroughoutthesample.Mixbygentlevortexingorinvertingatleastfivetimes.Frothingorover-mixingofsampleswillcausedenaturationofserumproteins.

Sample Handling continued

Page 15: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISATwopipettingmethodsusedforELISAarestandard(forward)andreverse.Notallpipettesarecapableofreversepipetting.Refertotheinstructionsincludedwithyourpipettefordetails.

Usestandard(forward)pipettingforthepreparationofsampledilutions,andreversepipettingfortheadditionofdilutedsamples,controlsandreagents.

Carefulpipettingiscrucialforaccuratetestresults.Becomefamiliarwiththepipetteandbothmethodsbeforerunningactualtests.Besuretousethecorrectpipetteandtip(volumecapacity)forthevolumebeingtransferred.

Pipetting Technique1. Drawthecalibratedvolumeofsampleintothetip.

2. Touchthesideofthetubewiththetiptoremovetheexcessliquid.

3. Ensurethatyouhavethepropervolumeofsampleinthetip.

Whenusingamultichannelpipette,ifthewellsonyourplateareempty,positionthetipsintothelowercornerofeachwell,makingcontactwiththeplastic.Ifthewellsonyourplatecontainliquid,positionthetipsabovetheliquid,makingcontactwiththeplastic.

Use standard (forward) pipetting for the preparation of sample dilutions, and reverse pipetting for the addition of diluted samples, controls and reagents.

Pipetting a Sample

Proper Pipetting

Draw up the calibrated volume of sample into the tip.

The drop on the tip needs to be removed.

Touch the side of the tube with the tip to remove the excess liquid.

Ensure that you have the proper volume of sample in the tip.

Proper position to dispense reagents into empty wells using a multichannel pipette; in the lower corner of each well

Proper position to dispense reagents into wells containing liquid using a multichannel pipette; above the liquid

Pipetting Methods

Page 16: Eliza Testi

��

ELIS

A Technical G

uide

ELISAStandard (Forward) Pipetting and Sample Preparation1. Putanewtiponasingle-channelpipetteandmakesurethatitisontight.

2. Presstheplungertothefirststop.

3. Somemanufacturersrecommendthatyouprewetthetipbyaspiratingandexpellinganamountofthesample.Checktheinstructionsthatcamewithyourpipette.

4. Drawthecalibratedvolumeofsampleintothetipandpauseforonesecondwiththetipstillinthesample.Becarefulnottoplacethetiptoodeeplyintothesample.

5. Touchthetiptothesideofthesamplecontainertoremoveanyexcessliquidontheoutsideofthetip.

6. Dispensethesampleintothemeasureddiluentbydepressingtheplungerpastthefirststoptothesecondstop.Becarefulnottoplacethetiptoodeeplyintothesamplediluent.

Forsampleslessthanorequalto10µL:Afterdispensingthesampleintothediluent,rinsethepipettetipinthediluentbypushingtheplungerdown2–3timesbeforeejectingthetip.

7. Mixsampleswithamultichannelpipettepriortodispensingsamplesontotheplate.Youcandothisbypushingtheplungerdown3–6times.

8. Ejectthetipintoawastecontainer.

Reverse Pipetting Using a Multichannel Pipette1. Putnewtipsonthepipette.Makesuretheyareontightandstraight.

2. Presstheplungerpastthefirststopandhalfwaytothesecondstop.

3. Drawtheliquidinaslowmotion,beingcarefulnottodrawanyairbubblesintothetips.Checkforconsistencyofvolumeinthetips.

4. Touchthetipstotheedgeofthereagentreservoirtoremoveexcessliquidontheoutsideofthetips.

a. Ifthewellsonyourplateareempty,positionthetipsintothelowercornerofeachwell,makingcontactwiththeplastic.

b. Ifthewellsonyourplatecontainliquid,positionthetipsabovetheliquid,makingcontactwiththeplastic.

5. Slowlydispensetheliquidintothewellsbydepressingtheplungertothefirststop.Becarefulnottosplashliquidoutofthewells,andmakesuretherearenodropsleftonthetips.

6. Torepeat,holdtheplungeratthefirststopandcontinuewithstep3.

7. Ejectthetipsintoawastecontainer.

NOTE:Reversepipettingusesmorereagent/volume(=“deadvolume”).

Bent tips; needs to be replaced

Pipetting Methods continued

Page 17: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISAAutomated Dilution Systems and Competitive AssaysForthosesystemsandassaysusingneatsamplesorlowerdilutionfactors,thesamplecanbeputdirectlyintothewellsofthecoatedplates.

Followthesequencebelow:

1. Addthediluenttotheplate.

2. Addthesampleintothediluent.

3. Mixbytappingtheplateorrepeatingpipetting.

Automated equipment uses more reagent/volume than semiautomated. Check your manufacturer’s recommendations for purging and priming your system.

Pipetting Methods continued

Page 18: Eliza Testi

��

ELIS

A Technical G

uide

ELISAAdding Samples and ControlsIncubationsforassayplatesshouldbetimedaspreciselyaspossible.Usuallytheprocessofaddingsamplestotheplaterequiresthemosttime.Whenyoudispensesamplesontotheplate,itiscriticaltokeepthetimedifferencebetweenthefirstandthelastsampletoaminimum.

Useamultichannelpipettewheneverpossibletominimizethetimeintervalfromthestartoftheplatetotheend.Ifthetimeintervalbecomestoolongoryouareinterruptedwhileaddingsamplestotheplate,putapositivecontroland/oryourin-housecontrolattheendoftheplateandcomparetheseresultswiththecontrolsatthebeginningoftheplate.

Fortightercontroloverthetimedifferentiationfromwhenthecontrolsandsamplesareadded,youcanputyourcontrolsinatubethatisrackedinpositionwithyoursamples.Thenuseamultichannelpipetteandputthecontrolsontotheplateatthesametimeyouareaddingthesamples.

Multiple Plate RunsWhentimingmultipleplates,itisimportanttokeeptrackofthetimeintervalfromthefirstplatetothelastplateintherun.Keepyourbatchsizessmallenoughsoyourprocessesdonotoverlap.Youdonotwanttobewashingaplate(unlessitisdonebyanautomatedwasher)whileanotherneedstohaveconjugateadded.Ifpossible,useatimerforeveryplate.

To minimize incubation time between controls and samples, rack the controls with the samples and add them to your plate using a multichannel pipette.

Use several timers when incubating multiple plates.

ELISA Plate Timing

Page 19: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISAAutomated or Semiautomated SystemsIngeneral,anautomatedorsemiautomatedwashsysteminproperworkingorderwillprovidemoreconsistentwashingthanmanualmethods.Checkthatallthedispensingneedlesaredispensingwithasmooth,steadystreamandthatallaspirationportsaspirateuniformly.

Makesureyourwashsystemisproperlycleanedandmaintained.RefertotheEquipmentMaintenanceandCalibrationsectioninthisguide(page6)andyourowner’smanualforpropermaintenance.Theplate-washingtechniqueshouldbeconsistentfromplatetoplateandfromrowtorowwithinaplate.Avoidprolongedsoaktimesunlessspecificallyrecommendedinthepackageinsert.

Preparethewashsolutionaccordingtothepackageinsert.Useonlythewashsolutionincludedwith,orrecommendedfor,yourkit.

Aspiratereagentsfromtheplatebeforedispensingthewashsolution.

Followthespecificrecommendationsinyourpackageinsertforthenumberofwashestouseateachstepofanassay.Mostassaysrequireapproximately300–350µLperwellperwash.Becarefultofillthewellsabovethelevelofthereagents.Donotallowwellstooverflow.Ifthisoccurs,thetestresultsmaybeinvalid.

Donotallowtheplatetodrybetweenplatewashingsandpriortotheadditionofreagents.

Afterthefinalaspiration,tapoutanyremainingliquidontoseverallayersofpapertowels.

Whentestingmilk,albumin,wholebloodoryolksamples,takeextracaretoinspectthewells.Becauseoftheirproteinorfatcomposition,thesesampletypesaresometimesmoredifficulttowashfromthewellsandmayrequirethemaximumrecommendednumberofwashcycles.

After tapping out plates, check paper towels for any evidence of color. This may indicate that the plates were not washed properly and there are reagents remaining in or around the wells.

Microbial growth in wash system tube; needs to be replaced

ELISA Plate Washing

Page 20: Eliza Testi

��

ELIS

A Technical G

uide

ELISAManual or Semi-Manual SystemsWorkquicklysothetimefromwashingthefirstwell/rowtothelastisminimal.Ifthetimeistoolong,theemptywellsmaydryoutandthelastwellswillhavealongerincubationthanthefirstwells.

Washingbymultichannelpipetteorwashbottlesshouldbeavoidedifatallpossibleasthisgivesinsufficientwashingthatresultsinhighandinconsistentbackground.

Makesuretoaspiratealltheliquidfromthewellsbyplacingtheaspirationneedlesatthebottomandinthecornersofthewells.Donotscrapethesurfaceoftheplateasthiswillremovetheantigen/antibodyboundtothesurfaceandcauseinconsistentorinaccurateresults.Afteraspiration,wellsshouldnotdrybeforetheadditionofthenextreagent.

Aftertappingouttheplates,checkthepapertowelsforanyevidenceofcolor.Thismayindicatethattheplateswerenotwashedproperlyandtherearereagentsremaininginoraroundthewells.

Overflowing plate; this can contaminate other wells

Proper position of manual washer needles for dispensing wash solution

Proper position of manual washer needles for aspirating liquid

ELISA Plate Washing continued

Page 21: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISAReading PlatesThelaststepinanELISAistoreadandinterprettheresults.Formostassays,theopticaldensity(amountofcolor)ofthesolutionontheplateisreadwithaspectrophotometer,commonlyknownasaplatereader.Therearemanymodelsandmanufacturersofplatereaders;refertothemanufacturer’sinstructionsfordetailsofoperation.

Thepackageinsertspecifieswhichwavelengthisrequiredfortheassay.Mostassaysspecifytheabsorbancereadingat450nmor650nm.Mostassaysareoptimizedusingaplatereaderequippedwitha650-nmfilter.Otherfilterscanbeused,butwillresultinloweropticaldensity(OD)values.Theuseof630-nmor620-nmfilterswilllowertheODvaluesofboththecontrolsandsamples,butwilldosoequivalentlyacrosstheentireplate.Theuseofthesealternativefilterswillnotaffectthetestresults.

Platesshouldbereadassoonaspossiblefollowingtheadditionofstopsolution.Absorbancereadingsmaydriftifexcessivetimeelapsesbetweenstoppingthereactionandreadingtheplates.

Data ManagementIDEXXprovidesthexChek*softwaretoassistyouinthecollectionandmanagementofthedatafromyourELISAassays.ThexCheksoftwareinterfaceswithmostcommonplatereaderstoreadtheplate,sendtheopticaldensitiestothecomputerandcalculatetheresults.AnIDEXXTechnicalServicesRepresentativecanassistyouinlearningmoreaboutthissoftware.

The xChek* software interfacing with the ELISA plate reader

Reading Plates and Data Management

Page 22: Eliza Testi

��

ELIS

A Technical G

uide

ELISAThisinformationisintendedtohelpyoutroubleshootyourELISAprocedure.Ifyouneedassistance,contactIDEXXProductionAnimalServicesTechnicalServicesat1-207-556-0895,option2;or1-800-548-9997,option2.

NOTE:TheconditionsdescribedheremaynotpertaintoeveryELISAkitbecauseperformancerequirementsvaryforindividualassays.Besuretocheckyourpackageinsertforspecifications.

ELISA Troubleshooting

Issue:High background or excessive color development (high optical density [OD] readings)

Possible Causes Recommended Actions

Poor-qualitywaterwasusedtowashplatesortoprepare

washsolution.

Checkthewaterquality.Ifitisquestionable,trysubstitutinganalternatewatersource,suchasbottleddistilledwater,towashplatesorpreparethewashsolution.

Substratesolutionhasdeteriorated.

Makesurethesubstrateiscolorlesspriortoadditiontotheplate.

Therewasinsufficientwashingorpoor

washerperformance.

Tryusingthehighestnumberofwashesrecommendedfortheassay.Makesurethatatleast350µLofwashsolutionisdispensedperwellperwash.Verifytheperformanceofthewashersystem.Havethesystemrepairedifanyportsdrip,dispenseoraspiratepoorly.

Washersystemhadmicrobialcontamination.

Cleanoutmicrobialcontaminationbyflushingthesystemwithadilutesolutionofbleach(10%byvolume)followedbyalargeamountofdistilledordeionizedwater.Primethesystemwiththeappropriatewashsolutionbeforeuse.Thetubingmayneedtobechangedifthecontaminationisheavy.

Washsystemcontainedanalternatewashformulation.

Besureeachuniquewashsolutionisproperlylabeled.Primethesystemthoroughlywhenswitchingbetweensolutions.

Readerwasmalfunctioningornotblankedproperly;

thisisapossiblecauseiftheODreadingswerehighand

thecolorwasnotdark.

Verifythereader’sperformanceusingacalibrationplateandcheckthelampalignment.Verifytheblankingprocedure,ifapplicable,andreblank.

Laboratorytemperaturewastoohighortoolow.

Maintaintheroomtemperaturewithin18°–25°C.Avoidrunningassaysnearheatsources,indirectsunlightorunderairvents.

Reagentswereintermixed,contaminatedor

preparedincorrectly.

Ensurethatthecorrectreagentswereused,thatworkingsolutionswerepreparedcorrectlyandthatcontaminationhasnotoccurred.

Page 23: Eliza Testi

�0

ELI

SA

Tec

hnic

al G

uid

e

ELISAIssue:Insufficient color development (low optical density [OD] readings)

Possible Causes Recommended Actions

Laboratorytemperaturewastoolow.

Maintaintheroomtemperaturewithin18°–25°C.Avoidrunningassaysunderairconditioningventsornearcoldwindows.

Washsolutionwaspreparedincorrectlyorthewrongwash

solutionwasused.

Besuretousethewashsolutionrecommendedforthekitandthatitispreparedcorrectly.Labeleachuniquewashsolutiontoavoidusingthewrongone.

Washersystemhadmicrobialcontamination

orcontainedanalternatewashformulation.

Cleanoutmicrobialcontaminationbyflushingthesystemwithadilutesolutionofbleach(10%byvolume)followedbyalargeamountofdistilledordeionizedwater,thenprimethesystemwiththeappropriatewashsolution.Besureeachuniquewashsolutionisproperlylabeled.Primethesystemthoroughlywhenswitchingbetweensolutions.

Toomanywashcycleswereused.

Staywithintherecommendedrangeforthenumberofwashcycles.Trytousethelowestnumberofwashesrecommendedfortheassay.

Incubationperiodsweretooshort.

Followprotocolforincubationtimes.Timeeachplateseparatelytoensureaccurateincubationperiods.

Reagentsandplatesweretoocold.

Makesureplatesandreagentsareatroomtemperaturebytakingthemoutoftherefrigerator,andthekitcomponentsoutofthekitbox,atleast2–3hoursbeforestartingtheassay.

Reagentswereexpiredorintermixedfromadifferent

lotnumber.

Verifytheexpirationdatesandlotnumbersonthereagents.

Wrongconjugatewasused,conjugatewas

preparedincorrectlyorhasdeteriorated.

Besurethattheconjugateusedistheonethatcamewiththekitandallconjugatesarekit-andlot-specific.Ifpreparationofaworkingconjugateisneeded,besurethattheconcentrateanddiluentaremixedinappropriatevolumes.Donotpreparetheworkingsolutiontoofarinadvance,anddonotsaveanyunusedportionforfutureuse.Ifnoconjugatepreparationisnecessary,besuretopouroutonlytheamountrequiredforimmediateuse,anddonotreturnanyunusedportiontothestockbottle.

Assayplatereadwasatwrongwavelength,orreader

wasmalfunctioning.

Verifythecorrectwavelengthfortheassayandreadtheplateagain.Verifyreadercalibrationandlampalignment.

Positivecontrolwasdiluted(indirectformatonly).

Donotdilutecontrolsunlessspecifiedinthepackageinsert.

Excessivekitstresshasoccurred.

Checkrecordstoseehowmanytimesthekithascycledfromtherefrigerator.Checktoseeifthekitwasleftoutonaloadingdockorotherareafortoolongoratextremetemperatures.

Assayplateswerecompromisedorpreviouslyused.

Besuretorefrigerateplatesinsealedbagswithadesiccanttomaintainstability.Preventcondensationfromformingonplatesbyallowingthemtoequilibratetoroomtemperaturewhileinthepackaging.Ifpartialplatesareused,besuretolabelusedwellstopreventreuse;coverthemwithsealingtapeandusetheremainingwellsassoonaspossible.Donotstorepartiallyusedplateswithotherplates.Includeadesiccantinthestoragebag.

ELISA Troubleshooting continued

Page 24: Eliza Testi

��

ELIS

A Technical G

uide

ELISAELISA Troubleshooting continued

Issue:Replicates within a plate show poor reproducibility

Possible Causes Recommended Actions

Excessivetimewastakentoaddsamples,controlsorreagentstotheassayplate.

Besuretohaveallmaterialssetupandreadytousequickly.Useamultichannelpipettetoaddreagentstomultiplewellssimultaneously.Rackcontrolswithsamplesanddispensethemontotheplateatthesametimeasthesamples.

Multichannelpipettewasnotfunctioningproperly.

Verifypipettecalibrationandcheckthattipsareontight.Besureallchannelsofthepipettedrawanddispenseequalvolumes.

Therewasinconsistentwashingorwasher

systemmalfunctioning.

Verifytheperformanceofthewashersystem.Havethesystemrepairedifanyportsdripordispense/aspiratepoorly.

Therewaspoordistributionofantibodyinthesample.

Ifthesamplewasthawedorrefrigerated,makesureitwasmixedpriortodilution.Dilutedsamplesalsoneedtobemixedpriortoaddingthemtotheplate.

Issue:No color development

Possible Causes Recommended Actions

Reagentswereusedinthewrongorderoranassaystep

wasomitted.

Checkthepackageinsertfortheassayprotocolandrepeattheassay.

Sampleswerenotaddedtodiluent(indirectformatonly).

Verifythatthesampleswereaddedtothediluent.

Wrongconjugatewasused,conjugatewas

preparedincorrectlyorhasdeteriorated.

Besurethattheconjugateusedistheonethatcamewiththekit.Allconjugatesarekit-andlot-specific.Ifpreparationofaworkingconjugateisneeded,besurethattheconcentrateanddiluentaremixedincorrectvolumes.Donotpreparetheworkingsolutiontoofarinadvanceanddonotsaveanyunusedportionforfutureuse.Ifnoconjugatepreparationisnecessary,besuretopouroutonlytheamountrequiredforimmediateuseanddonotreturnanyunusedportiontothestockbottle.

Page 25: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISAELISA Troubleshooting continued

Issue:Poor reproducibility plate to plate

Possible Causes Recommended Actions

Inconsistentincubationtimesoccurredfromplatetoplate.

Timeeachplateseparatelytoensurethatplateshaveconsistentincubationperiods.

Inconsistentwashingoccurredfromplatetoplate.

Usethesamenumberofwashesforeachplate.Verifytheperformanceofthewashersystem.Havethesystemrepairedifanyportsdripordispenseoraspiratepoorly.

Pipettewasworkingimproperly.

Checkthepipettecalibration.Verifythatpipettetipsareontightbeforeuseandthatallchannelsdrawanddispenseequalvolumes.

Kitcontrolsandsampleswereatdifferent

temperatures.

Besuretoallowsufficienttimeforsamplediluent,samplesandkitcontrolstocometoroomtemperaturebyremovingthemfromthekitbox.Largervolumeswillrequirelongerequilibrationtime.Ifusingawaterbathtohastenequilibration,makesureitismaintainedatroomtemperature;donotuseawarmwaterbathforcontrols,samplesordiluent.

Reagentswerebeingusedfromdifferentkitlots.

Ifrunningtwodifferentkitlotsatthesametime,makesuretolabelreagenttrays,etc.,soallreagentswithinalotareusedwiththecorrespondingplates.

Page 26: Eliza Testi

��

ELIS

A Technical G

uide

ELISAFor adjustable-volume pipettes, calibrations should be performed at a minimum of two settings—a low volume and a high volume at commonly used settings.

Materials

•Pipette •AnalyticalBalance •GlassBeaker

•DeionizedorDistilledWater •WeighingVessel •Thermometer

Procedure1. Toavoiderroneousresultsduetoevaporation,werecommend

humidifyingtheanalyticalbalancechamberatleast2hourspriortocalibration.Thiscanbeachievedbyplacingasmall,half-filledbeakerofwaterintothechamberwithallitsdoorsclosed.

2. Followthemanufacturer’sdirectionsforcleaningandlubricationofpipettespriortocalibration.

3. Theroomtemperatureshouldremainconstantwithin±0.5°C,preferablybetween18°–25°C.

4. Allowasufficientvolumeofdeionizedordistilledwatertocometoroomtemperatureandtakeatemperaturereading.

5. Recordthebeginningweightoftheweighingvesselorzerothebalance.

6. Usinganewpipettetipwitheachdelivery,pipettewaterintotheweighingvesselandrecordtheweight.Repeatthisstep10times.

7. Calculatethevolumedispensedforeachdelivery.

Calculations1. Calculatetheactualvolumedeliveredasfollows:

Volume= WeightofWater

DensityofWater

Densityofwaterat16°–21°C=0.998mg/µL Densityofwaterat22°–25°C=0.997mg/µL

2. Calculatethemean(M),standarddeviation(SD)andcoefficientofvariance(CV)ofthe10volumestodetermineprecisionofthepipette.

3. Determinetheaccuracyofthepipetteasfollows:(1-[Differencebetweenstatedandactualvolume/statedvolume])X100=%accuracy

Recommended specifications1. Precision:CV≤5.0%

2. Accuracy:≥95%

LabelingLabelthepipettewiththecalibrationdate,thetechnician’sinitials,theprecisionandtheaccuracy.Also,recordthesedatainalaboratorynotebookorlogforlong-termstorage.

Appendix A: Gravimetric Pipette Calibration Procedure

Automated Calibration SystemThe ARTEL PCS® Pipette Calibration System is an automated instrument/reagent system designed to facilitate regular and frequent verification of pipette performance and operator technique. For more information call ARTEL, at 1-888-406-3463 or 1-207-854-0860, or visit their Web site at artel-usa.com.

Page 27: Eliza Testi

Ap

pe

nd

ix B

:

Inve

nto

ry C

on

trol

Tra

ck

ing

Ch

art

Kit

Typ

eLo

t N

umb

erN

umb

er o

f Kits

Rec

eive

dD

ate

Rec

eive

dE

xpir

atio

n D

ate

C

om

men

ts

��

ELI

SA

Tec

hnic

al G

uid

e

ELISA

Page 28: Eliza Testi

��

ELIS

A Technical G

uide

ELISAA

pp

en

dix C

: Flo

ck

Ch

ek

*, He

rdC

he

k* a

nd

CH

EK

IT* L

ab

ora

tory T

rack

ing

Ch

art A

ssayNam

e:

Co

mm

ents

*FlockChek,H

erdChekandC

HE

KITaretradem

arksorregisteredtrademarksofID

EXXLaboratories,Inc.intheU

nitedStatesand/orothercountries.

Date

Technician

Kit Lot Number

Plate Numbers

Negative

Control

Positive Control

Internal Control

Internal Control

Internal Control

Internal Control

Room Temperature

Page 29: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISAAppendix D: Maintenance and Calibration Schedule

Procedure Daily Weekly Monthly Quarterly Yearly

Pipettes

Cleanexterior X

Checkcalibration X

CleaninteriorandO-rings X

Dilutor

FlushsystemwithDIwater X

Purgesystem X

Checkcalibration* X

Soaksyringes X

Check/Changetubings Check Change

Washer

FlushsystemwithDIwaterwhenusingwashsolutionsotherthanDIwaterandwhenchangingbetweenwashsolutiontypes

Checktraps,filtersandfoaming X

Checkaspirationanddispensingneedlesfordripsanddebris

Checktubesandbottlesformicrobialgrowth X

Decontamination—flushsystemwithbleachoralcoholsolution* X

Calibrationcheck—Purgesystem X

Cleanexterior X

Runthrough“cleancycle”withDIwater*

Check/Changetubing Check Change

Reader

Calibrationplate** X

Lampalignment Afterbulbreplacement

Cleanoptics X

Cleanexterior X

*Refertoyourmanufacturer’sguideforspecificinstructionsforyourmakeandmodel.**CallIDEXXTechnicalServicesoryourreadermanufacturerforrecommendationsoncalibrationplates.

X

X

Page 30: Eliza Testi

��

ELIS

A Technical G

uide

ELISAMonitoringandtrackingassayperformanceonquality-controlchartsprovidesinsightastowhenitisnecessarytotroubleshootproblems.BelowisachecklisttoreviewifyouarehavingproblemswithyourELISA.Ifyouhavefollowedallthestepsbelowandarestillhavingproblemswithyourassay,contactIDEXXProductionAnimalServicesTechnicalServicesat1-207-856-0895,option2;or1-800-548-9997,option2.

✔ EquipmentmKeeppreventivemaintenanceup-to-date

mCalibrateandcleanpipettes

mCalibratereader

mSanitizeandmaintainwashsystem

✔ ReagentsmMaintaininventorycontrol—FIFO

mInspectcomponents

mBringreagentstoroomtemperature

mAvoidcontamination

mVerifyproperstorage

✔ TechniquemMonitorsamplequality

mVerifyreagentpreparation

mVerifyappropriatesamplemixing

mVerifyproperpipetting

mChecktiming—multipleplateruns

mCheckwashingofassayplates

mUsein-housecontrols—trackresults

✔ OthermMonitorlaboratorytemperature

mUsesteriledisposablesandreservoirs

mMonitorassayperformance;recordinalog

Appendix E: Quality Control Quick Check

Page 31: Eliza Testi

��

ELI

SA

Tec

hnic

al G

uid

e

ELISANotes

Page 32: Eliza Testi

ELISA Technical Guide