HG.115/116.v2 Endoscopes Inspection Report Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for other items at the end of the Scopes Report) These reports are compiled by our Medical Engineering Team & are for guidance only; all lots are sold as seen, as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase. Lot Description Condition 110 OLYMPUS CF-1T200L Video Scope 1600x15mm Insertion tube Good condition Bending mechanism & Angulation control Good movements and locks Light guide cable & Light transmissionBoth light channels are functional Objective lens and focusing Functional Remote control switches Functional Distal end Good Image Quality Good image Working Channel (s) Clear Leak test Pass Over all condition Overall physical condition and image quality are good 111 OLYMPUS CYF-4A Direct View 350x5.54mm Insertion tube Good condition Bending mechanism & Angulation control Good movements and locks Light guide cable & Light transmissionBoth light channels are functional Objective lens and focusing Functional Remote control switches N/A Distal end Good condition Image Quality Good image Working Channel (s) Clear Leak test Pass Over all condition Overall physical condition and image quality are good 112 FUJINON UG1-FP2 Direct View 1000x10.30mm Insertion tube Fair condition Bending mechanism & Angulation control Good movements and locks Light guide cable & Light transmissionBoth light channels are functional Objective lens and focusing Functional Remote control switches N/A Distal end Average condition as some physical damage is present Image Quality Fair image Working Channel (s) Clear Leak test Leaking Over all condition Overall physical condition and image quality are in fair condition but scope is leaking 113 OLYMPUS BF TYPE P30 Insertion tube Fair condition but has kinks on the insertion tube near the control box Bending mechanism & Angulation control Good movements Light guide cable & Light transmissionBoth light channels are functional Objective lens and focusing Functional Remote control switches N/A Distal end Good condition

Upload: others

Post on 24-Feb-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014

(Condition Reports for other items at the end of the Scopes Report)

These reports are compiled by our Medical Engineering Team & are for guidance only; all lots are sold as seen,

as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase.

Lot Description Condition

110 OLYMPUS

CF-1T200L

Video Scope

1600x15mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – Functional

Distal end – Good

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition and image quality are good

111 OLYMPUS

CYF-4A

Direct View

350x5.54mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition and image quality are good

112 FUJINON

UG1-FP2

Direct View

1000x10.30mm

Insertion tube – Fair condition

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Average condition as some physical damage is present

Image Quality – Fair image

Working Channel (s) – Clear

Leak test – Leaking

Over all condition – Overall physical condition and image quality are in

fair condition but scope is leaking

113 OLYMPUS

BF TYPE P30

Insertion tube – Fair condition but has kinks on the insertion tube near the

control box

Bending mechanism & Angulation control – Good movements

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Good condition

Page 2: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Direct View

540x5.58mm

Image Quality – Poor image with 50+ fibres missing and water damage

Working Channel (s) – Clear

Leak test – Fail - Leaking

Over all condition – Overall physical condition is fair but the image is poor,

water damage is visible and has 50+ fibres missing

114 OLYMPUS

CYF-4A

Direct View

350x5.65mm

Insertion tube – Fair condition but has a kink on the insertion tube near the

control box and there is a dent near the distal end

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition is fair but the image is good

115 OLYMPUS

CYF-4

Direct View

350x5.78mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition is good and the image

quality is good

116 OLYMPUS

BF TYPE 240

Video Scope

500x6.20mm

Insertion tube – Fair condition but dent along the insertion tube

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – Functional

Distal end – Good condition

Image Quality – Image is average but the light does not focus and therefore

shows bright regions on the output image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition is fair but the image quality

is poor as the light is not focused and shows bright patches

117 OLYMPUS

CYF TYPE 240

Insertion tube – Fair condition but has a kink near the control box

Bending mechanism & Angulation control – Movements and locks are

functional but is stiff

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – Not functional

Distal end – Good condition

Page 3: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Video Scope

350x5.75mm

Image Quality – Image is poor as the image is blurred

Working Channel (s) – Clear

Leak test – Fail- Leaking

Over all condition – Overall physical condition is fair but the image quality

is poor as the image is blurred and the scope is leaking

118 OLYMPUS

CYF TYPE 240

Video Scope

350x5.65mm

Insertion tube – Fair condition but has a kink near the control box

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – Functional

Distal end – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition is fair and the image quality

is good.

119 OLYMPUS

CYF TYPE 240

Video Scope

350x5.48mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – Functional

Distal end – Good condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Overall physical condition is fair and the image quality

is good.

120 OLYMPUS

CF-200S

Videoscope

600x14.80mm

Insertion tube – Satisfactory condition

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– Good

Objective lens and focusing – Good optics

Remote control switches – Fully functional

Distal end – Satisfactory condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Satisfactory overall condition

121 OLYMPUS

CF-240AL

Videoscope

1600x12.90mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– Good

Objective lens and focusing – Good optics

Remote control switches – Fully functional

Distal end – Satisfactory condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Page 4: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Over all condition – Good overall condition

122 OLYMPUS

BF-P30

Direct view

500x5.30mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movement

Light guide cable & Light transmission– Good

Objective lens and focusing – Satisfactory

Remote control switches – N/A

Distal end – Good condition

Image Quality – Below satisfactory, bad water damage to whole image and

25+ fibres missing

Working Channel (s) – Clear

Leak test – Leaking

Over all condition – Below satisfactory overall condition as scope has

water damage and 25+ fibres missing to image, also the scope is leaking

123 OLYMPUS

SIF-230

Videoscope

2000x12.00mm

Insertion tube – Satisfactory condition

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– Good

Objective lens and focusing – Good optics

Remote control switches – Fully functional

Distal end – Satisfactory condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Leaking when movement controls are moved to maximum

Over all condition – Good overall condition, however there is a leak when

movement controls are moved to maximum

124 OLYMPUS

CF-240AL

Videoscope

1600x13.0mm

Insertion tube – Very good condition

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– Good

Objective lens and focusing – Good optics

Remote control switches – Fully functional

Distal end – Satisfactory condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Pass

Over all condition – Very good overall condition

125 OLYMPUS

CF-H260AZL

Videoscope

Insertion tube – Very good condition

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– Good

Objective lens and focusing – Unable to test as do not have the correct

pigtail to test with

Remote control switches – Unable to test as do not have the correct pigtail

to test with

Distal end – Satisfactory condition

Image Quality – Unable to test as do not have the correct pigtail to test

with

Working Channel (s) – Clear

Leak test – Leaking

Page 5: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

1600x14.10mm Over all condition – Good physical condition, unable to test scopes

working condition as do not have the correct pigtail to test with. The scope

however is leaking

126 OLYMPUS

GIF-PQ20

Direct view

1000x9.90mm

Insertion tube – Satisfactory physical condition slightly stained in colour

and slight dent near bottom of insertion tube

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– One of the light outputs is not

functional

Objective lens and focusing – Good optics, however there are 15 fibres

missing from image

Remote control switches – N//A

Distal end – Satisfactory physical condition

Image Quality – Satisfactory image as there are 15 fibres missing from

image

Working Channel (s) – Clear

Leak test – Leaking

Over all condition – Below satisfactory physical condition, 15 fibres

missing from image, one of the light outputs is not functional and scope is

leaking

127 OLYMPUS

PCF-230

Videoscope

1300x12.0mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movement and locks

Light guide cable & Light transmission– Good

Objective lens and focusing – Good optics

Remote control switches – Fully functional

Distal end – Satisfactory condition

Image Quality – Good image

Working Channel (s) – Clear

Leak test – Leaking

Over all condition – Good condition, scope is leaking

176 VISION

SCOPE

-

Direct View

280x3.22mm

Insertion tube – Poor condition where there is a dent along the tube, kinks

near the control box and the outer insulating sheath near the distal end is

missing

Bending mechanism & Angulation control – Good movements

Light guide cable & Light transmission– Both light channels are functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Fair condition

Image Quality – Fair image but there is slight water damage visible on the

image near the edge

Leak test – Unable to test as there is no connector at time of testing

Over all condition – Overall physical condition is poor due to the condition

as the insertion tube but the image is fair but slight water damage is visible

177 VISION

SCIENCES

ENT – 2000

Insertion tube – Poor condition where there is a dent along the tube and

kinks near the control box

Bending mechanism & Angulation control – Good movements

Light guide cable & Light transmission– Light channels are functional

Page 6: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Direct View

300x3.49mm

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Good condition

Image Quality – Good image

Leak test – Pass

Over all condition – Overall physical condition is poor but the image

quality is good

181 OLYMPUS

CYF-4A

Direct View

350x5.56mm

Insertion tube – Good condition

Bending mechanism & Angulation control – Good movements and locks

Light guide cable & Light transmission– Both the light channels are

functional

Objective lens and focusing – Functional

Remote control switches – N/A

Distal end – Good condition

Image Quality – Good image

Working Channel - Clear

Leak test – Pass

Over all condition – Overall physical condition is good and the image

quality is good

Page 7: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Condition Report – Sale: 1406 Date: 13 / 05 / 2014

These reports are compiled by our Medical Engineering Team & are for guidance only; all lots are sold as seen,

as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase.

Lot

Number

Description Report

1 Datascope System

98XT Balloon Pump

with 1x Box of

Acessories

Powers up – display screen and control panel are in good working

order and the printer is also functional. Doppler is included and the gas

cylinder and adult condensate removal/safety disk are in place. Overall

the unit is in good physical condition. Not fully tested. Includes

manual, ECG lead set and a spare gas cylinder. Operating hours: 2198.

2 Zeiss Operating

Microscope

The operating microscope powers up, with unit able to produce a good

light output. The mechanical movements and locks are in working

order. The operating microscopes optical system shows to produce a

clear image, with unit showing to have good focusing with the

magnification functional. The unit is in fair physical condition.

3 Haag-Streit 900 Slit

Lamp with Table

and Accessories Case

The slit lamp is able to power up, with it able to produce a light output.

The fixation light is in place and shows to function. Units joystick

movements shows to be freely movable, with its lock functional. Units

chin rest is in place and its chin adjustment is in working order. The slit

lamps tables mechanical height adjustment is in working order. The slit

lamp is missing its eyepieces, however a set of eyepieces were used for

tested purposes and showed the rest of the optical system to be able to

produce a clear image. Overall the slit lamp is in satisfactory physical

condition, with the side compartment on the table missing its door.

4 Drager Fabius

Anaesthetic Machine

The anaesthetic machine was tested using a medical compressor and

the anaesthetic machine showed to have its gauges and flow meters in

working order. The anti-hypoxic safety feature is in working order and

the O2 flush is functional. The gas cut off alarm is functional. The

anaesthetic machine includes O2, air and N2O hoses. The unit is in

good physical condition.

Manufacture date: 2001

6 Zeiss OPMI-1

Operating

Microscope

The power supply comes on and all the selected outputs power is

working. The lamp is working fine. The optical system needs cleaning

up but still with fair magnification and focus. All mechanical

movements and locks are functional. Magnification range – 0.4 to 2,5

and eye pieces 12,5x. Objective lens f=200. System is in fair overall

physical condition.

Page 8: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

7 Medtronic Lifepak

20 Defib With ECG

Leads x1

The defibrillator powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

defibrillator is testes using a defibrillator analyser and the output

energies are within the limits. The AED and manual mode are in good

working order. The ECG signal is tested using an ECG simulator and,

the defibrillator appears to give good strong traces of ECG signals.

Pacer option is included, which has not been fully tested but appears to

be functional. Printer is functional. The service indicator stays on.

Includes III lead ECG, AED therapy cable and two defibrillator pads

(expire 03/2014 and 04/2014). The overall physical condition is

reasonable. The defibrillator is a Biphasic unit and its Y.OM is 2004.

8 Aloka Echo Camera

SSD-500 Ultrasound

with Probe Aloka

7.5MHz

The ultrasound powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

ultrasound is tested using a phantom and set measurements taken are

within the limits. The resolution is average. Ultrasound includes modes

B and M, which are in good working order. The probe (model: UST-

5512U-7.5MHz) is in good physical condition and is intact. Overall

physical condition is fair.

10 Pfizer Valleylab

Force 30 Diathermy

(No Foot Switch)

The diathermy powers up well and passed its initial self-test. The

display and function buttons are in good working order. Some of the

function buttons has some wear and tear. The diathermy has not been

fully tested but the patient monitoring system is in good working order.

No footswitches are included. Overall physical condition is reasonable.

16 Zoll M Series Defibs

x 2

Both the defibrillators powers up well, its display, functions and

function buttons are in good working order. The defibrillators have

been tested using a defibrillator analyser and the output energies are

within the limits. The ECG is tested using an ECG simulator and the

defibrillators give good strong traces of ECG signals. The pacer option

is included, which has not been fully tested but appear to be functional.

The printer is functional. The defibrillators include two batteries which

appear to be flat and are in need of charge. Two AED therapy cables

are included but no ECG leads are included. Overall physical condition

is reasonable.

24 Wolf Stack Trolley

With Panasonic

M2080 Monitor,

Wolf 5512 Processor,

Wolf 5124 Light

Source, Wolf

Camera Head and

Wolf 2232 Laparo

CO2 Insufflator

All units power up. The processor and camera head are able to produce

a satisfactory image on the Panasonic monitor, with the camera head

also showing to have satisfactory focusing. The light source is able to

produce a good light output, with its brightness adjustment in working

order. The insufflator has not been tested but is able to power up ok.

Overall lot is in satisfactory physical condition.

Page 9: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

25 Smith and Nephew

Stack System Sony

Trinitron Monitor

PVM-20L2MD, 640

Image System 300

XL Xenon Light

Source, 350 3-CCD

Camera Operator

and Camera Head,

Sony Colour Video

Printer UP-21 MD

(037097)

All units are able to power up. The camera head and processor are able

to produce a satisfactory image on the Sony monitor, with the camera

head also showing to have satisfactory focusing. The light source is

able to produce a good light output with its brightness control

functional. The 640 image management system is able to power up and

works on the National Display Systems flat screen, but in the boot up

unit displays message ‘CMOS checksum error’ which is able to be

passed. Unit is able to power up successfully and shows to be

functional. The color video printer is functional. Overall lot is in

satisfactory physical condition.

Image management system Version: 2.3

28 Philips Intellivue

MP70 Patient

Monitors x2

The two patient monitor’s powers up well and passed its initial self-

test. Its display and LCD touchscreen are in good working order. The

monitors do no detect the multi-parameter and recorder modules.

Therefore unable to carry out any further checks. No leads are

included. Overall physical condition is reasonable. Y.O.M is 2004 and

2005. Software Versions are: REV B.05.67 and REV F.01.42

29 Datex Capnomac

Ultima Patient

Monitors x2

Both monitors power up ok and the display screens and control panels

are in good working order. CO2 functions appear to working but gas

monitoring functions have not been tested. Overall the units are in

reasonable physical condition. The alarm silence buttons on both

monitors have been worn out.

30 Penlon AV900

Ventilator

The ventilator powers up well and passed its initial self-test. The

display, LCD touchscreen and function buttons are in good working

order. The unit has not been fully tested but appears to be functional.

Oxygen hose is included. Overall physical condition is as there are

scratches on the top cover. S/W is V3.11

31 Penlon AV900

Ventilator

The ventilator powers up well and passed its initial self-test. The

display, LCD touchscreen and function buttons are in good working

order. The unit has not been fully tested but appears to be functional.

No hoses are included. Overall physical condition is reasonable. S/W is

V3.30

34 Graseby 3400

Syringe Pump, 3200

Syringe Pump and

Ivac P6000 Syringe

Pump x2, Dinamap

Pro 100V2 Patient

Monitor and Nellcor

NBP-4000 Patient

One of the IVAC syringe drivers is not powering up neither is the

Dinamap. The Nellcor power s up but the display is not displaying

anything. The rest of the units included in lot are powering up with

their function buttons and displays looking to be in working order.

None of the units have been tested. The lot is in fair physical condition,

with the Dinamap’s printer door missing, and part of the Nellocrs

handle missing. No accessories included with lot.

Page 10: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Monitor

41 Huntleigh Flowtron

Pumps x5

Unable to check to units as their plugs are missing, the other three units

power up , with pumps looking to function however they have not been

fully test. Overall the lot is in reasonable physical condition.

43 Storz Stack Trolley

with Storz 20090231

Monitor. NDS

Radiance Monitor,

Storz Xenon Nova

20131520 Light

Source, Storz Image

22200020 Camera

Control Unit, Storz

CD Re-Writer and

Keeler Headset

All units are able to power up. Unit does not include a camera head so

unable to test the processor. The light source is missing its

interchangeable light guide adaptor; however the light source is able to

produce a good light output. The imaging unit is able to power up, and

then is able to open the AIDA compact programme but it fails to fully

load as message ‘No dongle found’ is displayed which is unable to be

passed. The head gear is able to power up and is able to produce a light

output with its brightness adjustment functional. Overall lot is in

satisfactory physical condition.

44 Valley Lab Force 2

Diathermy - No

Pedals

The diathermy has its switch missing, therefore unable to power up and

carry out any further checks. No footswitches are included. Overall

physical condition is fair.

45 ATL HDI 5000

Ultrasound with

Mitsubishi

CP200,C5-2,L12-5

and C8-4V Probes

The ultrasound powers up but does not go into its main menu screen,

therefore unable to carry out any further checks. The lot includes four

probes (models: L7-4, C5-2, L12-5 and C8-4v), which are physically

intact but have not been tested. Printer is included, which powers up

but has not been tested. Overall physical condition is reasonable.

50 GE Dash 4000

Patient Monitor,

Datex Ohmeda S/5

Patient Monitors x2

(Damage to One)

The three patient monitors power up well, and passed its initial self-

test. The Dash 4000 powers on for a few minutes but turns itself off;

also its display monitor has lines running through the LCD screen. The

Datex Ohmeda (S/N: 4293223) has many pixels missing and therefore

unable to clearly read the display monitor. The monitor appears to be

functional, where it’s ECG and NIBP are tested but has not been

verified, unable to test the SPO2 as display set up is needed to be

accessed but difficult to read the display monitor. This monitor has a

number of cracks along the casing and its rotary knob is missing.

Recorder is missing but parameters IBP, ECG, SPO2, Temp and NiBP

are included.

The other monitor (S/N: 4749116) is in good working order where its

ECG is fully tested using an ECG simulator and is functional. The

NiBP and SPO2 appear to be functional but have not been fully tested.

The printer is included and functional. Included parameters NiBP,

ECG, SPO2, IBP, CO2 and Temp. Two batteries are included but are

flat and in need of a charge. Lot includes x1 AC adaptor but no leads

are included. Overall physical condition is average due to usage.

Page 11: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

65 Datex Ohmeda

Aestiva 3000

Anaesthetic Trolley

with Ventilator

The anaesthetic machine powers up ok and the pressure gauges, flow

meters, O2 flush; Basel flow and anti-hypoxic safety feature are all in

good working order. The 7900 Smart Vent powers up ok and

completes the system self-test successfully. The function buttons are

working fine but the display has several pixels out, however the

displayed values are not affected. The ventilator was tested with test

lung and medical air and the set parameters and alarm limits were

working fine. The flow and oxygen sensors are working fine. The

absorber, bellows and rebreathing assembly are in good condition with

no leak detected during test run. The backup battery for the Smart Vent

is functional. Overall, the system is in good physical condition. O2,

Air, N2O hoses and AGSS receiver are included

Smart Vent Software Version: 3.2

66 SLE 2000 Infant

Ventilator with SLE

2100 Respiration

Monitor

The ventilator powers up, with its function buttons and numerical

displays in working order. Unit is missing its expiratory block, but one

was used for test purposes. The ventilator showed to be functional,

when it was tested using a medical compressor. The ventilator showed

to be functional in all ventilation modes. Units output has not been

verified to see if it is within set limits. Unit oxygen is able to calibrate

to 21%. Unit includes O2 and air hoses. The SLE 2100 is able to power

up, with it then passing through its initial self-test successfully. Units

display is functional and the function buttons look to be responding.

The SLE2100 has not been tested. The humidifier is also powering up

but again has not been test. The lot is in satisfactory physical condition.

Vent modes: SIMV, PTV, CMV and CPAP

Operation hours: 12835.53h

68 Eschmann TD850

Diathermy with

Yellow and Blue

Foot Switch on ST80

Suction Trolley and

Lina Smoke

Evacuator

The diathermy powers up well, its display, functions and function

buttons are in good working order. The diathermy is tested using an

ESU analyser and the monopolar (cut and coag) output results are out

of limits. Bipolar is not tested as there are no footswitches. Includes

monopolar footswitches (cut and coag). The ST 80 suction has its

mains cable cut off and therefore unable to fully test. The Lina suction

powers up and appears to be functional but has not been fully tested.

Overall physical condition is reasonable.

69 Olympus CLE-4U

Light Source and

Atlantech Xenon

Fibre-Optic Light

Both units are able to power up, however neither unit was able to

produce a light output. The Olympus CLE-4U has its air feeding pump

functional. Units are in below satisfactory physical condition.

Page 12: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Source

70 Medtronic Lifepak

20 x1, Physio

Control Lifepak 9

with Shock Advisory

Adapter x1

Lifepak 20

The defibrillator is able to power up and is able to pass through its

initial self-test successfully. Units function buttons, rotary knob and

display are in working order. Unit does not include any paddles but

does include (1x) AED cable. Units output was tested using a

defibrillator analyser and all readings taken showed to be within set

limits. Unit was functional in both manual and AED modes. Unit was

able to charge to 360J in manual mode. The defibrillator is biphasic.

The audio voice prompts are functional. Unit includes 3 wire ECG

leads and units ECG was verified as functional. Units printer is

functional but the battery is not. Unit is in good physical condition.

Lifepak 9

The defibrillator is able to power up, with its function buttons and

display in working order. Unit does not include any paddles but

includes a shock advisory adaptor and AED cable. Unit has not been

fully tested. Unit was able to give a test output of 360J, but the output

has not been verified. Only manual mode has been checked. The shock

advisory adaptor has not been checked. Unit includes (1x) 3 wire ECG

and units ECG has been verified as functional. Units printer is

functional and the battery is low. Unit is in fair physical condition.

Lifepak 9 manufacture date: 1999

Lifepak 12 manufacture date: 2006

74 Drager Oxylog 2000

Portable ventilator

Unit power up and system self-test completed successfully. The lithium

battery and mains charger is in good working order. LCD and controls

are functional. Airways pressure indicator and alarm setting are

working fine. The ventilator modes include: IPPV, SIMV, CPAP and

SIPPV and they all appears to be functional. Test run with medical air

compressor and test lung, indicates that all parameter and alarm

settings were in good working order. The minute Volume (MV) was

verified with the set valve of Tidal volume and Frequency (BPM) and

found to be within limit. The system is in good physical and cosmetic

condition. Oxygen hose with schrader connector is included.

76 Graseby 3200

Syringe Pumps x5,

Graseby 3100

Syringe Pumps x3

and Alaris GH

Syringe Pump and

Abbot Flexiflow

The Alaris Plus syringe driver is not power up and one of the Graseby

3100 displays ‘fault 86’ which is not able to be passed. The rest of the

units are able to power up, with their function buttons and displays in

working order. Units have not been tested. The lot is in fair physical

condition.

Page 13: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Enteral Pump

88 Olympus PSD 20

Diathermy with Foot

Switch

Unit is able to power up, with it numerical displays in working order

and its function buttons looking to respond. Unit has not been tested

with its output not checked. Units patient plate alarm is functional.

Unit includes (1x) footswitch for monopolar cut/coag, which looks to

be responding. Unit is in good physical condition.

92 Cardiac Science

AED Trainer x2,

Lifepak CRT

Trainer x1

The lifepak CR-T does not include a battery so unable to determine if

the unit is able to power up.

Both the Cardiac trainers are able to power up, however only one of the

units includes batteries. Both units’ displays are functional. Both units

voice prompts are functional. Both units were tested on a defibrillator

analyser, one unit was unable to find a signal the other unit was able to

find a signal but then no actual output (0J) was recorded when a

shock was delivered.

Overall lot is in satisfactory physical condition.

103 Omega Excel Laser

System with 1 x

Probe, Manual and

Key

The laser unit powers up well and passed its initialization phase. The

display, functions and function buttons are in good working order. The

beam test was conducted and appears to be in good working order.

Includes probe which is in good working order (200mW and lambda

820nm). Key is included. Overall physical condition is fair. Y.O.M

2001

142 Smith and Nephew

30º Autoclavable

Rigid Scopes x 4

The four rigid scopes appear to be physically intact and are inclined at

30 degrees. Four of the rigid scopes do not produce a good output

image. Overall physical condition is reasonable.

143 Smith and Nephew

30º Autoclavable

Rigid Scopes x 4

The four rigid scopes appear to be physically intact and are inclined at

30 degrees. Four of the rigid scopes do not produce a good output

image. Overall physical condition is reasonable.

144 Smith and Nephew

30º Autoclavable

Rigid Scopes x 4

The four rigid scopes appear to be physically intact and are inclined at

30 degrees. Four of the rigid scopes do not produce a good output

image. Overall physical condition is reasonable.

145 Smith and Nephew

30º Autoclavable

Rigid Scopes x 4

The four rigid scopes appear to be physically intact and are inclined at

30 degrees. Four of the rigid scopes do not produce a good output

image. Overall physical condition is reasonable.

175 Olympus MAJ-554

Camera Head

Unable to test the camera head as there is no processor to test with. The

Camera head looks to be intact and in satisfactory physical condition.

Page 14: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

178 Downs 0° Rigid

Scope with Light

Lead

The rigid scope is physically intact and gives a good output image. The

light source cable is included which appears to be intact. Overall

physical condition is fair.

183 GE 4C-RS Curved

Array Convex Probe

GE Curved Array Convex Probe, 4C-RS (year 2007) -The probe

connection, cable and head are intact and in good physical condition;

but not tested due to unavailability of suitable unit. The probe is

compatible with model GE Logiq e, Vivid e & Vivid i ultrasound

system.

184 GE RNAS-9-RS 3D

Micro-Convex Probe

GE 4D Micro-convex Probe, RNA5-9-RS (year 2009), Frequency

range: 3.3 to 9.1 MHz. The probe connection, cable and head are intact

and in good physical condition; but not tested due to unavailability of

suitable unit. The probe is compatible with model GE Voluson I

ultrasound system.

185 GE 11L-D Linear

Probe

Unable to test the probe as there is no ultrasound to test with. The

probe looks to be intact and in satisfactory physical condition

187 Philips C5-1

Ultrasound Probe -

BOFN912

The probe has not been tested on a compatible ultrasound. Therefore

unable to carry out any further tests. Overall physical condition is

reasonable but near the scan head it has been taped previously. (Note

the probe has been tested on iU22 and HD11XE and was not detected)

188 Qty of Urology

Instruments to Inc.

Storz Resectoscopes

with Obturator,

Comeg Cystoscope,

Ellick Connectors

Lot is in satisfactory physical condition, lot is incomplete consisting of

various pieces.

195 Aloka UST-9123

Convex Ultrasound

Probe

The probe is in good physical and cosmetic condition and the scanhead

is physically intact. A compatible ultrasound is not available therefore

the probe cannot be tested.

196 Storz Hopkins 7200

30º Rigid Scope

The rigid scope is physically intact but does not give a good output

image. The rigid scope is inclined at 30 degrees. Overall physical

condition is fair.

200 Olympus OTV-S6

Camera Head

Unable to test camera head. The camera head is missing its lens

adaptor. The camera head is missing its rubber grommet at the bottom

of the camera head causing a large section of the inner insulation to be

exposed. The camera head is in below satisfactory physical condition.

235 Pentax LH 150FP

Light Source

The light source powers up, with its function knobs in working order.

The light source is able to produce a light output with its brightness

Page 15: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

adjustment functioning. The air pump appears to be functioning. The

light source is in fair physical condition for unit’s age.

239 Bayer Clinitek 50

Analysers x2 (1

Spare only)

Unable to power up units as lot does not include correct power cables.

Units are in reasonable physical condition with one unit missing its

printer cover.

246 Blease Frontline 690

Anaesthetic Machine

with 8200s

Ventilator

The anaesthetic machine has not been tested but shows to have its flow

meters and gauges intact. The ventilator also has not been tested but

showed to be able to power up with its numerical displays functioning.

Unit’s bellows and absorber are in place. Unit includes an AGSS

receiving system and hose and a high flow high vacuum regulator and

hose. The unit is also complete with O2, air and N2O hoses. Overall

the unit is in fair physical condition.

248 Smith and Nephew

Stack Trolley with

Sony LCD Monitor,

Smith and Nephew

660 HD Image

Management

System, Dyonics

300XL Xenon Light

Source, Smith and

Nephew 460P 3-CCD

Digital Camera

Control Unit, HP

Printer and Camera

All units power up, with the function buttons functional. The camera

head and processor produce a satisfactory image on the Sony flat

screen monitor, with the camera head also showing to have satisfactory

focusing. The light source is able to produce a good light output with

its brightness adjustment functional. The image management system

showed to have its touch screen display functional with the unit

showing to be in working order. The printer powers up. Overall lot is

in good physical condition.

Image management system version: 1.5

252 Olympus TK-E1

Trolley with

Olympus CV240

Control Unit with

Pigtail and Olympus

CLV-U20 Light

Source

All units on the Olympus trolley power up ok. The light source xenon

lamp is working fine and light output model selector (OES, EVIS 100

& EVIS 200) are fully functional. Brightness control in auto and

manual mode are working fine. The video processor with the video

cable MH977 are in good working order when tested with a

videoscope. All the function keys and video controls are working.

Good quality image was displayed when connected to a colour

monitor. The system is in good overall physical condition.

259 Sensor Medics

3100A Oscillatory

Ventilator with

Fisher and Paykel

MR700 Humidifier

The ventilator is able to power up, with its function buttons/dials and

the numerical displays showing to function. The ventilator has not been

tested. The oxygen blender monitor is in place but unable to verify if it

is powering up. The humidifier is powering up with its function

buttons/dial and numerical display functioning, however again unit has

not been tested. The ventilator includes O2 and air hoses. Overall the

lot is in fair physical condition.

Operation hours: 11278.7

Page 16: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

260 Sensor Medics

3100A Oscillatory

Ventilator with

Fisher and Paykel

MR700 Humidifier

The ventilator is able to power up, with its numerical displays in

working order. The silencing alarm function button is not functioning

unable to determine if the rest of the function buttons are responding.

The ventilator has not been tested. The Air-oxygen blender is in place.

The humidifier is powering up but does not appear to be functioning

with unit not responding and only displaying ‘EOo’ on the numerical

display. The ventilator includes O2 and air hoses. Overall the lot is in

fair physical condition.

Operation hours: 10897.3

261 Blease Frontline 440

Anaesthetic Machine

with 8200s

Ventilator and

Marquette Solar

8000 Patient

Monitor

The anaesthetic machine has not been tested but shows to have its flow

meters and gauges intact. The oxygen monitor shows to be powering

up with its function buttons and display in working order. The

ventilator is also powering up with its numerical displays functioning.

Again the ventilator has not been tested. The units bellows and

absorber are in place. Unit includes an AGSS receiving system and

hose and a high vacuum regulator and hose. The unit is also complete

with O2, air and N2O hoses. The patient monitor is able to power up

but however does not include any modules, unable to test unit. Overall

the lot is in reasonable physical condition.

265 Nihon Kohden

Cardiofax Q ECG

Recorder with ECG

Lead

The ECG machine powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

ECG machine has not been fully tested but appears to be functional.

The printer and battery is in good working order. Includes 12 lead ECG

wire. Overall physical condition is fair.

266 Nihon Kohden

Cardiofax GEM

ECG Recorder with

ECG Lead

The ECG machine powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

ECG machine has not been fully tested but appears to be functional.

The printer and battery is in good working order. Includes 12 lead ECG

wire. Overall physical condition is fair.

278 Smith and Nephew

Stack Trolley with

Sony LCD Monitor,

Smith and Nephew

660 HD Image

Management

System, Dyonics

Xenon XL Light

Source, Smith and

Nephew 460P 3-CCD

Digital Camera

All units power up, with function buttons functional. The camera head

and the processor are able to produce a good image on the Sony flat

screen monitor, with the camera head also showing to have good

focusing. The light source is able to produce a good light output with

its brightness adjustment functional. However the light sources power

switch is loose. The image management system showed to have its

touchscreen display in working order with the unit showing to be in

working order. The printer powers up. Overall the lot is in satisfactory

physical condition.

Image management system version: 1.5

Page 17: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Control Unit,

Camera Head and

HP Printer

279 Conmed Excalibur

Plus Diathermy on

Suction Trolley with

Blue, Yellow and

Black Foot Switches

The diathermy powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

diathermy is tested using an ESU analyser and the monopolar (cut and

coag) and bipolar output readings are within the limits. However the

pure cut mode on the higher output appear to be out of range otherwise

all the other modes are within the range. Includes monopolar (cut and

coag) and bipolar footswitches. The suction unit powers up but the

suction is not functional, therefore unable to carry out any further tests.

The overall physical condition is reasonable.

280 Conmed Excalibur

Plus Diathermy on

Suction Trolley with

Blue, Yellow and

Black Foot Switches

The diathermy powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

diathermy is tested using an ESU analyser and the monopolar (cut and

coag) and bipolar output readings are within the limits. Includes

monopolar (cut and coag) and bipolar footswitches. The suction is in

good working order but output has not been verified. The overall

physical condition is reasonable.

286 Kodak Dryview

8150 Laser Imager

The unit powers up however the display panel is not functional, so

unable to do anything with unit. Unit is in satisfactory physical

condition.

287 Fisher and Paykel

Cosy Cot

Resusitator

Unit powers up, with its indication lights and display in working order.

Units function buttons are functional however the manual mode

function button is missing its button cover. Unit showed to heat up well

in ‘Prewarm’ and ‘manual’ modes. unit does not include a skin

temperature probe. Units overhead light and timer both showed to be

functional. The neopuff section of the unit has not been tested. Units

electrical height adjustment is in working order. Unit includes O2 and

air hoses and a mattress. Overall unit is in satisfactory physical

condition.

Manufacture date: 2007-01

288 Fisher and Paykel

Cosy Cot

Resuscitator

Unit powers up but immediately alarms and gives an error message ‘Er

017’. The error message is unable to be passed. No further checks are

able to be carried out on the unit. The neopuff section of the unit has

not been tested. Unit include O2 and air hoses and a mattress. Unit is in

satisfactory physical condition

Manufacture date: 2004-01

294 Stryker Resus

Trolley -30109

The patient trolley is physically intact; its upholstery is in good

condition. All articulating movements are in good working order. Its

Page 18: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

height adjustment is functional. The side rails are intact. The brakes are

in good working order. Overall physical condition is reasonable.

302 Fresenius 4008s

Hemofiltration

Machine

Unit is not powering up with unit only able to give a constant alarm.

Unit includes modules, double blood pumps, heparin pump, air

detector and BiBag. The Bibag module panel cover is completely

missing. Unit includes water and drainage hoses. Overall the unit is in

fair physical condition.

Manufacture date: 2010

Operation hours: 8972.03

303 Welch Allyn Patient

Monitors on Stands

with SPO2 Leads

The three vital signs monitor power up well, its display, functions and

function buttons are in good working order. The NiBP pump and SPO2

is functional but has not been fully tested. Lot includes two NiBP cuffs.

Battery is in good working order. Overall physical condition is fair.

308 Medelac Saphire 2M

and DP Medical

Light Source

Medelec Sapphire II. Powers up ok - display screen and controls are in

good working order and the printer is also functional. No further

checks carried out. Lot includes a DP Medical 202032 light source

which is in good working order. Overall the unit is in reasonable

physical condition.

311 Braun Dialog

Dialysis Machine x 2

Unable to power up units as both of the mains cables has been

removed. Both of the units include (1x) single blood pump, (1x)

heparin pump and (1x) air detector. Both of the units have their water

and drainage hoses cut off from the bottom. Overall lot is in reasonable

physical condition.

312 Braun Dialog

Dialysis Machine x 2

Unable to power up units as both of the mains cables have been

removed. Both of the units include (1x) single blood pump, (1x)

heparin pump and (1x) air detector. One of the units includes water and

drainage hoses the other unit has its water and drainage hoses missing.

Overall lot is in reasonable physical.

315 Conmed Excalibur

Plus Diathermy on

Suction Trolley with

Blue, Yellow and

Black Foot Switches

The diathermy powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

diathermy is tested using an ESU analyser and the monopolar (cut and

coag) and bipolar output readings are within the limits. Includes

monopolar (cut and coag) and bipolar footswitches. The suction and its

regulator are functional but the output has not been verified. The

overall physical condition is reasonable.

316 Conmed Excalibur

Plus Diathermy on

Suction Trolley with

Blue, Yellow and

Black Foot Switches

The diathermy powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

diathermy is tested using an ESU analyser and the monopolar (cut and

coag) and bipolar output readings are within the limits. Includes

monopolar (cut and coag) and bipolar footswitches. The suction is

Page 19: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

functional but the regulator is not functional an unable to change the

suction output. The overall physical condition is reasonable.

317 Drager 5400

Transport Incubator

The Transport incubator 5400 power up ok on mains; battery power

has not been tested. The incubator set temperature is maintained

accordingly. The BabyMix 2 ventilator is in good physical condition

but not fully tested. The Oxygen monitor Oxydig has no battery, not

tested. The lot is in fair overall ohysical condition.

322 Marquette Mac8

ECG Recorder with

ECG Lead

Unit is able to power up after a long period of time. Then the unit

shows to have its function buttons in working order. Units display

functions however the display flickers which makes it hard to read in

places, with some lines of pixels missing. Unit makes a loud constant

humming noise. Unit has not been tested. Unit includes 12 wire ECG

however all of the leads are missing their end connection pins. Unit is

in reasonable physical condition.

325 Conmed Excalibur

Plus Diathermy on

Trolley with Yellow,

Blue and Black Foot

Switches

The diathermy powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

diathermy is tested using an ESU analyser and the monopolar (cut and

coag) and bipolar output readings are within the limits. Includes

monopolar (cut and coag) and bipolar footswitches. The overall

physical condition is reasonable.

326 Drager D-Vapour

Desflurane

Vapouriser

The vaporiser is in good physical condition, with its volume control

dial and lock intact. There are some minor marks to paintwork.

328 Drager Oxylog

Portable Ventilator

Unit power up when connected to medical gas. Airways pressure

indicator working fine. Test run with medical air compressor and test

lung, indicates that all parameter and alarm settings were in good

working order. The minute Volume (MV) was verified with the set

valve of Tidal volume and Frequency (BPM) and found to be within

limit. The system is in fair physical condition. Oxygen hose with

schrader connector is included.

361 The ultrasound is unable to power up. Unable to test the unit. Units

(1x) probe appears to be physically intact. Unit is in reasonable

physical condition, with its covers colour being worn away in large

patches.

369 Drager 8000SC

Incubator

Unit is able to power up, with units function buttons and displays in

working order. In ‘Air’ mode unit showed to be functional, with unit

able to reach and maintain the set temperature. Temperature output has

not been verified. Unit does not include a skin temperature probe so

unable to test ‘skin’ mode. ‘Humidifier’ mode has not been checked.

Units electrical height adjustment and mechanical cot tilt are in

working order. Unit includes a mattress. Unit is in satisfactory physical

Page 20: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

condition.

374 Edwards Lifescience

Aquarius

Hemofiltration

Machine

Unit is able to power up, units function buttons, rotary knob and

display show to be functioning. Unit has not been tested. The blood

pump and filtrate cover panel is missing, also the prefillter sensor is

bent and unable to close. Overall unit is in fair physical condition.

Manufacture date: 2009

Software version: 6

379 Sonosite 180 Plus

with Probe

The ultrasound powers on, but is not functional as the display gives a

badly flickering display with no actual picture showing. Unable to

check unit. Units (1x) probe appears to be physically intact. The

ultrasound is in satisfactory physical condition.

Manufacture date: 2004-08

382 Zeiss IOL Master

with Printer and

Table

The IOL Master powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

IOL Master is tested using a test eye and test results taken where within

the limits. The functions included are Axial Length Measurement,

Corneal Curvature Measurement and Anterior Chamber Depth

Measurement, which are in good working order. Printer is functional.

The height adjustment is in good working order. Overall physical

condition is reasonable. IOL Master Version: 3.02.0304

387 Haag-Streit Slit

Lamp with Table

The slit lamp is able to power up, with it able to produce a light output,

the reflector mirror for the light output is in place but has a large chip

at one corner. The fixation light is in place but however it is not

functional and has its outer cover missing. The joystick movements

show to be functioning but however is slightly stiff, the movement lock

is functional. The chin rest is in place and its chin adjustment is

functional. The slit lamps tables mechanical height adjustment is in

working order. The slit lamps optical system is able to produce a clear

image, with it also showing to have good focusing. Overall the slit

lamp is in satisfactory physical condition, with the side compartment

on the table missing its bottom panel.

(1x) set of eyepieces (10x)

395 Quantel Medical

Virdis Twin Laser

with Key and Foot

Switch, BMBQ Slit

Lamp Adaptor

Quantel Medical Viridis Twin, software version: M:10.02 B:10.00.

Unit power up ok, the LCD screen and function keys are working fine.

The aiming beam is working and the 532 nm laser function screen

comes when the key is turned on. The parameters were adjustable for

power, pulse etc. Test firing of the laser indicate green laser at the out

but the power was not verified. When the key is turned to select 810

Page 21: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

nm laser an error message is displayed – E100A:0 ‘fibre disconnected.

Only one fibre cable with laser aperture is included with slit lamp

adapter model: BMBQ. One YAG safety goggle is included. System is

in fair physical condition.

396 Hill-Rom Air Shields

Resusitaire - 001297

Unit powers up, with its function buttons, indication lights and numeric

displays in working order. Unit showed to function with unit able to

heat up in ‘pre-warm’ and ‘manual’ modes. In ‘manual’ mode the heat

adjustment is functional. Unable to test ‘baby’ mode as unit does not

include a skin temperature probe. Units exam light and timer are both

functional. Units electrical height adjustment is functional as is the

mechanical cot tilt. Units Suction and patient supply were checked

using a medical compressor, and the suction and patient supply showed

to be functioning with their gauges in working order. Unit includes O2,

air and N2O hoses, unit also includes a mattress and suction jar.

Overall unit is in satisfactory physical condition.

399 Stryker Endoscopy

Trolley with

Trinitron Monitor

Q5000 Light Source,

X6000 Light Source,

2 x Sony Colour

Video Printer, 2 x

988i 3 Chip Camera

Controller and 3 x

988i Camera Heads

All units are able to power up, with the function buttons functional.

Both of the processors and two of the camera heads are able to produce

a good image on the Sony monitor, with both camera heads showing to

have good focusing. However one of the two camera heads does show

some slight interference from the connection end if moved. The third

camera head is able to produce a reasonable image however it is

constantly affected with bad interference when the camera head is

moved, the camera heads focusing is functional. Both of the light

sources are able to produce a good light output with their brightness

adjustments functional. Both of the color video printers are powering

up. Overall the lot is in satisfactory physical condition.

Bulb hours for Q5000: 205.2

Bulb hours for x6000: 183

400 Alcon Surgical

Ultrasonic Imaging

System with

Monitor and Probe

10MHz and Foot

Switch on Trolley

Alcon Ultra Scan imaging System, software version: 3.00. Unit power

up ok and the CRT display and keyboard are working fine. The

ultrasound probe 10MHz is appears to be working with a scan image

display on the screen. Unit is not fully tested. In fair physical

condition.

402 Eschmann TD411-S

Diathermy with Foot

Switches on ST80

Suction Trolley

The diathermy powers up well and passed its initial self-test. The

display, rotary knobs and functions are in good working order. The

diathermy is tested using an ESU analyser and cut outputs are out of

range whereas the coagulation modes are within the range. The bipolar

mode has not been tested as there are no footswitches. Monopolar cut

and coagulation footswitches are included. Overall physical condition

Page 22: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

is fair.

405 GE Mac 5000

Pagewriter ECG

With Leads

Unit is able to power up, with its display and function buttons in

working order. Unit has not been tested. Unit includes 12 wire ECG.

Unit does not include a battery, unable to test printer as there is no

paper. Unit is in satisfactory physical condition.

409 Zeiss OPMI-1FC

Colposcope with

Panasonic Camera

Unit powers up, unit is able to give a light output however the light

guide cable has a lot of its fibers damaged causing unit to have a poor

light transmission. Units mechanical movements are in working order.

Units optical system is able to produce a good image, with unit

showing to have its focusing an magnification in working order. The

Panasonic WV-CL110AE has no power supply so unable to test.

Overall the lot is in satisfactory physical condition.

(1x) eyepieces (10x/22B)

(1x) objective lens

414 Mortara ELI 250

with ECG

Pagewriter Leads

The ECG machine powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

ECG machine is tested using an ECG simulator and gives out good

strong traces of ECG signals. The printer is in good working order.

Includes 12 lead ECG. Overall physical condition is fair.

415 Zeiss IOL Master

with Printer and

Table

The IOL Master powers up well and passed its initial self-test. At time

of testing the monitor switched off and the display screen stayed blank,

but afterwards the screen went into its normal menu screen. The

display, functions and function buttons are in good working order. The

IOL Master is tested using a test eye and test results taken where within

the limits. The functions included are Axial Length Measurement,

Corneal Curvature Measurement and Anterior Chamber Depth

Measurement, which are in good working order. Printer is functional

but the ink cartridge needs replacing. The height adjustment is in good

working order. Overall physical condition is reasonable. IOL Master

Version: 3.01.0294

418 GE Mac 5000

Pagewriter ECG

With Leads

Unit powers up, with its function buttons and display in working order.

Unit has not been fully tested. Unit includes 12 wire ECG. Units

Printer is functional, however the battery is not functional. Unit is in

satisfactory physical condition.

440 Takagi S10-15 Slit

Lamp Microscope

The slit lamp is able to power up, with the unit showing to be able to

give out a light output. The fixation light is in place and is functional,

but is missing its fixation light cover. The joystick movements are

freely movable with the movement lock functional. The chin rest is in

place with its chin adjustment functional. The mechanical table height

Page 23: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

adjustment is in working order. The slit lamps optical system is able to

produce a clear image, with the magnification in working order. The

CV-S3200 camera head powers up and was testing using a monitor

(not included). The camera head showed to be able to produce a clear

image. Lot is in satisfactory physical condition.

(1x) set of eyepieces (16x)

443 Penlon AV900

Ventilator

The ventilator powers up well and passed its initial self-test. The

display, LCD touchscreen and function buttons are in good working

order. The unit has not been fully tested but appears to be functional.

No hoses are included. Overall physical condition is reasonable. S/W is

V3.33

450 3M Espe Eliper

Freelight 2 Curing

Light, Shrewsbury

Medical Ultrasound

Therapy System

The shrewsbury unit powers up, with units displays and function knobs

in working order. Unit showed to be functioning with unit giving an

output and with its timer functional. unit includes a (1x) transducer

(1.1, 3.2 MHZ). The unit is in fair physical condition

The Freelight 2 has its power indication lights on, but unable to

determine if unit is functional. Not tested. Lot includes its docking

station. Unit is in fair physical condition.

453 Olympus KV-5

Suction Pumps x2

Both of the suction pumps are able to power up. Both units show to be

giving suction, the suction output has not been verified. Both of their

vacuum regulators are in working order. However both of the units are

missing their suction jars. Overall lot is in good physical condition.

454 Olympus KV-5

Suction Pumps x2

Both of the suction pumps are able to power up. Both units show to be

giving suction, however one of the units is unable to reach the units

maximum output of 85KPa showing to be around 5+ KPa short.

Neither units output has been verified. Both of the units vacuum

regulators are in working order. Both units are missing their suction

jars. Overall lot is in good physical condition.

462 Nellcor Butitan

Bennett 840

Ventilator System

Con# D4133,

Ass#G7940

Unit is able to power up, with its touchscreen display functional and its

function buttons and rotary knob are looking to respond. Unit is

missing its expiratory filter, so unable to test unit. The ventilation

modes are selectable but have not been tested. Unit includes O2 and air

hoses. Overall the ventilator is in good physical condition.

Vent modes: A/C, SIMV, SPONT and BILEVEL

Software: 4-070000-85-AB

Operation time: 53614Hours

Manufacture date: 2000-03

Page 24: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

463 Nellcor Butitan

Bennett 840

Ventilator System

Con# D4133,

Ass#G9127

Unit is able to power up, with its touchscreen display functional and its

function buttons and rotary knob are looking to respond. Unit is

missing its expiratory filter, so unable to test unit. The ventilation

modes are selectable but have not been tested. Unit includes O2 and air

hoses. Overall the ventilator is in good physical condition.

Vent modes: A/C, SIMV, SPONT and BILEVEL

Software: 4-070000-85-AB

Operation time: 48823Hours

481 Valleylab Force FX

Electrosurgical Unit

Unit power up and self-test completed successfully. All displays,

output power and modes selector keys are in good working order. The

patient return plate monitoring circuit is functioning. The unit’s RF

output was verified using ESU analyser and the selected outputs were

all within limits for both bipolar and monopolar. Unit is in good

physical condition and mounted on a trolley. Footswitches and

cables/accessories are not included

482 Valleylab Force FX

Electrosurgical Unit

Unit power up and self-test completed successfully. All displays,

output power and modes selector keys are in good working order. The

patient return plate monitoring circuit is functioning. The unit’s RF

output was verified using ESU analyser and the selected outputs were

all within limits for both bipolar and monopolar. Unit is in good

physical condition and mounted on a trolley. Footswitches and

cables/accessories are not included

485 Stryker X6000 Light

Source

The light source is able to power up. Unit showed to be able to produce

a good light output, with its brightness adjustment in working order.

Bulb hours: 200. The light source is in good physical condition.

487 Wolf Arthro Power

Unit with Cable and

9 Rotating Cutting

Tools

Unit powers up, unable to verify if unit is functional as unit does not

include a footswitch. Unit does however include a flexible cable and

(9x) various hand pieces. Unit is in reasonable physical condition.

490 Penlon Enflurane

Vapouriser x 2

Both of the vaporisers are in satisfactory physical condition. With both

of the vaporisers volume control dials being intact.

491 Penlon Enflurane

Vapouriser x 2

Both of the vaporisers are in satisfactory physical condition. With both

of the vaporisers volume control dials being intact.

493 3M Unitek Ortholux

XT Curing Light,

3M XL3000 Curing

Light

Both units are able to power up, with both of the units able to give a

light output. Both units timers showed to be functional. The lot is in

satisfactory physical condition.

Page 25: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

497 Penlon Sigma

Sevoflurane

Vapouriser (Back

Lock Missing)

The vaporiser is in good physical condition, with its volume control

dial looking to be intact.

498 Penlon Sigma

Sevoflurane

Vapouriser

The vaporiser is in good physical condition, with its volume control

dial looking to be intact.

507 Respironics BIPAP x

2 and Resmed VPAP

II x 2

Bipap Synchrony

Both units power up, however one unit displays a blank screen and

gives a constant audible alarm, unit is not functional. The other unit has

its function buttons and displays in working order. With unit showing

to give an output but units have not been fully tested. Units are in fair

physical condition.

VPAP

Both units are able to power up, one units function buttons do not look

to be responding correctly with some of its display hard to read. The

second units function buttons and display are functional. Both units

showed to be giving an output but units have not been fully test. Units

are in fair physical condition.

521 Alcon Surgical

Ultrascan Imaging

System with Foot

Switch and Sony

Video Graphic

Printer

Alcon Ultra Scan imaging System, software version: 3.00. Unit power

up ok and the CRT display and keyboard are working fine. The

ultrasound probes included are damaged. Unit was tested with the

probe of lot 400 but still could not obtain a scan image. Unable to carry

out further test on this unit. The attached printer is working fine.

System is in fair physical condition.

522 Stack Trolley Inc

Trinitron Monitor

Storz Image Hub,

Xenon Nova 300

Light Source Aida

Control

All units are able to power up. Lot does not include a camera head so

the processor is unable to be tested but its power indication light

showed to come on. The light source showed to be able to produce a

good light output with its brightness adjustment functional. The AIDA

control had its power indication light on but the unit has not been

tested. The monitor showed to be functional. Overall the lot is in fair

physical condition.

523 SLT CL100 Yag

Laser with Box of

Accessories and

Consumables

The unit power up ok and the displays and function keys are working

fine. No fully tested. System is in fair physical condition. Year of

manufacture January 1992, Serial No. 91015501. The following are

included:

IV Pole, Footswitch, User & Maintenance Manual & ON/OFF Key

Page 26: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Accessories:

Laser bare fibre x5

Laser goggles x3

Black Laser glasses x3

Grey laser glasses x4

Chisel Probe x2

Round Probe x3

Hook Probe x1

Frosted Scalpel x1

Metal cones x5

CO2 Regulator & hose

SLT laparoscopy fibre x2

524 Conmed Excalibur

Plus Diathermy on

Suction Trolley with

Blue, Yellow and

Black Foot Switches

Unit is able to power up, with its displays in working order and its

function buttons looking to respond. The unit has not been fully tested.

The patient plate alarm is functional. Lot includes mono-polar and bi-

polar footswitches, the footswitches look to be responding. The suction

pump is able to power up, with unit able to produce suction, however

the regulator is not functional and is stuck on maximum suction. Unit

incudes (1x) suction jar. Lot is in reasonable physical condition.

525 Conmed Excalibur

Plus Diathermy on

Suction Trolley with

Blue, Yellow and

Black Foot Switches

Unit is able to power up, with its displays in working order and its

function buttons looking to respond. The unit has not been fully tested.

The patient plate alarm is functional. Lot includes mono-polar and bi-

polar footswitches, the foot switches look to be responding. The

suction pump is able to power up, with the unit able to produce suction,

its regulator is in working order. Unit includes (2x) suction jars with

one of the jars missing its lid. Lot is in reasonable physical condition.

526 BiPAP Vision

Respironics

Ventilatory System

on Stand - 046721

Unit powers up, before then powering back down again immediately

and then giving a continuous alarm for ‘Vent Inop’. Unit includes O2

hose. Unit is in satisfactory physical condition.

528 Keymed Trolley

With Olympus CV-

200 Camera Control

Unit, Olympus CLV-

U20 Light Source,

Olympus MH-977

'Pig Tail' Connector

Con# D4029,

7366097,7254862

Both the processor and light source are able to power up. The processor

and pigtail were tested using a monitor and a videoscope (both not

included). The processor and the pigtail both showed to be functional

and both helped in producing a good image. The processors function

buttons showed to be in working order. The light source showed to be

functional also with its function buttons in working order. All of its

light output modes (OES, Evis 100 and Evis 200) showed to be

functional. The light sources brightness adjustment worked in both

manual and automatic modes. The light sources air pump showed to be

temperamental in functioning. Overall the lot is in satisfactory physical

condition.

Page 27: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

531 Zeiss IOL Master

Eye Scanner

The IOL Master powers up well and passed its initial self-test. The

display, control panel and functions are in good working order. The

IOL Master camera is faulty and unable to take an image and therefore

unable to carry out any further tests. The test modes included are Axial

Length Measurement, Corneal Curvature Measurement and Anterior

Chamber Depth Measurement. Printer is included, which is functional.

The table height adjustment is in good working order. Overall physical

condition is reasonable. IOL Master Version 3.01.0294.

543 Cinescan Quantel

Medical with HP /

Mitsubishi Printers

and Foot Switch

The scanner power up ok, display and keyboard are functional. Both A

& B scan appears to be working fine. Good image displayed from the

US probe. All seem to be working and system is in good physical

condition. The B&W printer is also functional.

548 Smith and Nephew

Stack Trolley with

Sony LCD Monitor,

Smith and Nephew

660 HD Image

Management

System, Smith and

Nephew 460P 3-CCD

Digital Camera

Control Unit,

Camera and HP

Printer

All units power up, with function buttons in working order. The

processor and camera head are able to produce a good image on the

Sony flat screen monitor, the camera head also shows to have good

focusing. The image management system has its touch screen display

functional with unit showing to be in working order. The printer

powers up. Overall the lot is in good physical condition.

Image management system version: 1.5

551 Olympus Stack

Trolley with Sony

PVM20M2MDE

Monitor, Olympus

UHI-3 Insufflator,

Olympus CLV-S20

Light Source, Sony

SVD-3500MDP

VCR, Olympus MH-

317 Foot Switches x2

- Con# 4018, Ass# -

7939,7941,7942,7943,

5905

All units are powering up. The light source is able to produce a good

light output with its brightness adjustment functional, the spare lamp

also shows to be functional. The insufflator has not been tested but its

displays show to be functioning and its function buttons look to

respond. The monitor showed to be functional. The VCR is powering

up. Overall lot is in satisfactory physical condition.

552 Respironics BiPAP

Focus

Unit powers up, but unit is unable to complete its system start up, with

unit giving a constant audible alarm and displays system error ‘343

battery faulty or missing’ which is unable to be passed. Unable to carry

out any further tests on unit. The unit is in satisfactory physical

Page 28: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

condition.

Boot block version: 1.03.000

553 E-Vent Inspiration

LS Mobile

Ventilator

The ventilator is able to power up, with it showing to be able to pass

through its initial self-test ok. Units function buttons and display are in

working order. The ventilator showed to be functional when it was

tested using a medical compressor. The BPM, Vte and Ve are able to

match up accordingly. Units output has not been verified with a flow

analyser. Unit was only able to be tested with a pediatric flow sensor in

units pediatric setting. Unit also was only tested in V-SIMV mode,

with the rest of the vent modes not being verified. The battery is not

functional. Unit includes O2 and air hoses, no flow senor included in

lot. The ventilator is in satisfactory physical condition.

Vent modes: V-CMV, V-SIMV, PRVC-CMV, PRVC-SIMV, P-CMV,

P-SIMV, VS, SPONT and SPAP

Software version: 5.0.1

554 E-Vent Inspiration

LS Mobile

Ventilator

The ventilator is able to power up, with it able to complete its initial

self-test successfully. Units display is functional however the rotary

selection knob is temperamental, the rest of the function buttons are in

working order. The ventilator showed not to be functioning/ventilating

correctly when tested using a medical compressor. With units BPM,

Vte and Ve not able to match up and with them showing to be out of

limits from the values set. Output has not been tested with a flow

analyser. Unit was only able to be tested with a pediatric flow sensor in

units pediatric setting. Unit also was only tested in V-SIMV mode, rest

of vent modes not verified. The battery is functional. Unit includes O2

and air hoses and also a pediatric flow sensor. The ventilator is in

satisfactory physical condition.

Vent modes: V-CMV, V-SIMV, PRVC-CMV, PRVC-SIMV, P-CMV,

P-SIMV, VS, SPONT and SPAP

Software version: 5.0.0

555 E-Vent Inspiration

LS Mobile

Ventilator

The ventilator is able to power up, with it showing to be able to pass

through its initial self-test ok. Units function buttons are in working

order, with the display functioning however it can flicker and the

display picture can flip and appear upside down on the display. The

ventilator showed to be functional when it was tested using a medical

compressor. The BPM, Vte and Ve are able to match up accordingly.

Units output has not been verified with a flow analyser. Unit was only

able to be tested with a paediatric flow sensor in units paediatric

setting. Unit also was only tested in V-SIMV mode, with the rest of the

Page 29: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

vent modes not being verified. The battery is functional. Unit includes

O2 and air hoses and also a pediatric flow sensor. The ventilator is in

satisfactory physical condition.

Vent modes: V-CMV, V-SIMV, PRVC-CMV, PRVC-SIMV, P-CMV,

P-SIMV, VS, SPONT and SPAP

Software version: 5.0.1

556 Alcon Surgical

Series 20000 Legacy

Phaco Machine and

Foot Switch

Unit is able to power up, with its function buttons and touchscreen

display in working order. Unit does not include any hand pieces and

unit has not been fully tested. Modes look to be selectable but have not

been tested. Units audio voice prompts are functional. The IV pole

height adjustment is in working order. The foot control looks to be

responding and the remote is functional. Overall the unit is in good

physical condition.

V3.00

Manufacture date: 04/00

561 Nellcor Butitan

Bennett 840

Ventilator System

Con# D4133,

Ass#G9128

Unit is able to power up, with its touchscreen display functional and its

function buttons and rotary knob looking to respond. Unit is missing its

expiratory filter, so unable to test unit. The ventilation modes are

selectable but have not been tested. Unit includes O2 and air hoses.

Overall the ventilator is in good physical condition.

Vent modes: A/C, SIMV, SPONT and BILEVEL

Software: 4-070000-85-AB

Operation time: 44328Hours

564 Toshiba ECC CEE

Ultrasound with

Probes x3 -PLF

805ST x2, 375MT

The ultrasound powers up well and passed its initialization. The

display, functions and function buttons are in good working order. The

ultrasound is tested using a phantom and appears to be functional. The

set measurements taken are within the limits. The greyscale and

resolution is average. The ultrasound modes included are: B, M Colour

and PW. The three probes included are intact but all three probes

shows interference and lines through the image. One of the PLF-805ST

probes has discoloration on the probe membrane and has a large crystal

damage, which affects the image. The PVF-375MT probe has also

crystal damage but does not affect the image. Overall physical

condition is average.

565 Olympus TI-1900

Imaging Trolley with

Sony PVM-

20M2MDE Monitor,

All units are able to power up. Unable to test the two processors as lot

does not include any camera heads but their indication lights show to

be functional, unable to verify if function buttons are functional. The

light source is able to produce a good light output with its brightness

Page 30: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Olympus UHI-2

Insufflator, Olympus

OTV-S6 Camera

Control System (No

Camera), Olympus

CLV-S30 Light

Source, Olympus

OTV-SX2 camera

Control System (No

Camera), Stryker

Endoscopy SE5

Arthroscopy

adjustment in working order, its spare lamp is functional. The

insufflator has not been tested, but its displays look to be functional.

The SE5 powers up but again has not been tested, but has its displays

functional. The Sony monitor is functional. Overall lot is in satisfactory

physical condition.

566 Stockert Heart/Lung

Bypass Machine

Stockert S 3 Roller Pump. The system powers up and the control

console is in good working order. All 5 pumps are functional and the

display panels and speed controls are also in good working order. The

backup battery is also working and overall the unit is reasonable

physical condition. Year of manufacture: 2003

579 Steris Surgimax

Electric Operating

Table (No Pads or

Hand Control)

The operating table is powering up. With its function buttons on the

base of the unit functional. All of the electrical articulating movements

showed to be in working order. As did the mechanical movements.

Units brake lock is functional with the wheels freely movable. The

operating table is missing its head rest panel, unit is also missing all of

its cushions. Unit does not include a remote. The battery is functional.

The operating table is in satisfactory physical condition.

581 Erbe ICC 200 and

Erbe APC 300 On

Trolley with Foot

Switches

The diathermy powers up well and passed its initialization. The

display, functions and function buttons are in good working order. The

diathermy is tested using an ESU analyser and the cut output are within

the range however the coag modes are out of range. The bipolar

footswitches are not included and therefore unable to test. Includes

monopolar cut and coag footswitches. The Argon Plasma Coagulation

powers up but as there is no Argon gas at ti142me of tested therefore

unable to fully test. Overall physical condition is reasonable. Software

Version is V2.20 for the Argon Plasma coagulation.

587 Drager 8000 Resus

Unit

The resuscitaire powers up well, its display, functions and function

buttons are in good working order. The gauges and flowmeters are

intact and functional. The resuscitaire heater is in good working order.

The heater manual mode is tested, which appears to be functional

however as there are no skin probes included therefore unable to verify

the skin mode. The suction is in good working order along with its

regulator but the output has not been verified. The Resus unit has not

been fully tested but has been tested for leaks and no leak is present.

Page 31: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

The motorised height adjustment is in good working order. Includes

gel mattress, medical gas hoses (air and oxygen). The overall physical

condition is reasonable. Y.O.M is 1998

611 Olympus PSD-60

Diathermy with Foot

Switch

Unit is able to power up and pass through its initial self-check

successfully. Units displays look to be functional and its function

buttons look to be responding. The unit has not been tested, with its

output not verified. Units patient plate alarm appeared to be functional.

The (1x) foot switch for monopolar cut and coag looked to be

responding. Unit is in satisfactory physical condition.

Version: V1.0.1

649 Philips Pagewriter

Trim III ECG

Recorder and Leads

- 18822 Serial

Number

US20817891

The ECG machine is able to power up, with units function buttons,

rotary knob and display in working order. However the display has

been badly scratched, especially in one corner. Unit includes (1x) 12

wire ECG and units ECG was tested using an ECG analyser. Unit

showed to have good traces of ECG. Units printer is functional,

however units battery is not. Unit is in reasonable physical condition,

with bad scratches to the display.

652 Getinge 1345

Washer on Cabinet

Unit is able to power up, Units function buttons and display are in

working order. Unable to test unit. Units main door is able to open with

its lock in working order, inside unit includes its sterilizing rack. Units

cabinet door lock is locked with the door open, resulting it the cabinet

door not being able to close. Unit includes (1x) Martel MCP8810X-

184 printer which is not powering up and (1x) star TSP650 printer

which is missing its correct power adaptor. Lot is in good physical

condition, with however a dent to its back panel.

Version: 5.06(320)

669 Datascope Accutor

Plus SPO2 Monitor

with Cuff

Unit is able to power up, with its function buttons and displays in

working order. Unit has not been fully tested. Units NiBP pump shows

to be functional. Unable to test units SPO2 parameter as there is no

SPO2 probe included in lot. Lot includes (1x) NiBP cuff and hose.

Battery not functional. Unit is in fair physical condition.

671 Marquette 1/K

Operating Table

Remotes- 03115

Maquet cordless control 1133.9/A0 – in good physical condition but

not tested.

687 Fisher and Paykel

Cosy Cot Resusitator

Unit powers up, with units function buttons, indication lights and

numerical display in working order. Unit shows to be functional, with

unit able to heat up in ‘prewarm’ and ‘manual’ modes. With the

manual modes heat adjustment functional. Unable to test ‘baby’ mode

as lot does not include a skin temperature probe. The overhead light is

Page 32: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

functional as is the timer. The electrical height adjustment is in

working order. Unit includes a mattress and O2 and air hoses. The

‘Neopuff’ part of unit has not been tested. Lot is in satisfactory

physical condition.

Manufacture date: 2009-03

688 Fisher and Paykel

Cosy Cot Resusitator

Unit powers up, with units function buttons, indication lights and

numerical display in working order. Unit shows to be functional with

unit able to heat up in ‘prewarm’ and ‘manual’ modes. With the

manual modes heat adjustment functional. There is no skin temperature

probe included with lot. Units overhead light is functional as is the

timer. The electrical height adjustment is in working order. Unit

includes a mattress and O2 and air hoses. The ‘Neopuff’ part of unit

has not been tested. Lot is in satisfactory physical condition.

Manufacture date: 2007-10

689 Fisher and Paykel

Cosy Cot Resusitator

Unit powers up, with units function buttons, indication lights and

numerical display in working order. Unit shows to be functional with

unit able to heat up in ‘prewarm’ and ‘manual’ modes. With the

manual modes heat adjustment functional. There is no skin temperature

probe included with lot. Units overhead light is functional as is the

timer. The electrical height adjustment is in working order. Unit

includes a mattress and O2 and air hoses. The ‘Neopuff’ part of unit

has not been tested. Lot is in satisfactory physical condition.

Manufacture date: 2007-10

711 Marquette Betastar

Electric Operating

Table with Remote

Control (End Section

Missing Incomplete

Pads)

The operating table is able to power up. All electrical articulating

movements except for the leg movement are functional from the

function buttons on the main unit. The leg extension movement has not

been tested as table is missing its leg extension part of the table. Lot

includes (1x) remote which is not fully functional as only trendlenburg

movement works from it. The tables wheels are freely movable and the

brakes are functional. Unit does not include a complete set of cushions

with lot only including a cushion for the main body section, which

looks to be in satisfactory physical condition. Units battery is not

functional. The operating table is in satisfactory physical condition.

Page 33: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

713 Optiflex Leg

Extender

Unit is able to power up. Unit includes (1x) remote which shows to be

in working order. Unit shows to be functional, with its angle

adjustment functional as is its manual speed selection. The external

power supply makes a continuous humming noise. Overall the unit is

in reasonable physical condition.

725 PneuPAC VentiPAC

200D & paraPAC

200D, portable

ventilators (x2)

Both units power up when connected to medical gas. Airways pressure

indicator working fine and the battery powered alarm are working

fine.. Test run with medical air compressor and test lung, indicates that

all parameter and alarm settings were in good working order. The

minute Volume (MV) was verified with the set valve of Tidal volume

and Frequency (BPM) and found to be within limit. The system is in

fair physical condition. Oxygen hose with schrader connector is

included.

727 Gynecare Morcelltor Unit is able to power up, its indication lights are in working order. Unit

has not been tested as there is no hand piece for the lot. Unit is in fair

physical condition.

730 Dyonics 300XL

Xenon Light Source

- 10084

The light source is able to power up, with its function buttons and

numerical display in working order. Unit was able to produce a good

light output, with its brightness adjustment functional. Unit is in fair

physical condition.

Bulb hours: 101

778 Zeiss Colposcope The colposcope is able to power up. Unit is able to produce a good

light output with its brightness adjustment functional. Units mechanical

movements are functional, the lock on the main arm does not

completely tighten. Units optical system is able produce a clear image

with units magnification in working order. Overall the colposcope is in

satisfactory physical condition, with the cap cover for the brightness

adjustment knob missing.

(1x) set of eye pieces (12.5x)

(1x) objective lens (f250)

779 Alcon Surgical

Series 20000 Legacy

Phaco System

Unit powers up, with units function buttons and touch screen display in

working order. The unit does not include any hand pieces and unit has

not been fully tested. Modes are selectable but have not been verified.

Units IV pole height adjustment is functional as are the audio voice

prompts. The foot control shows to respond however the remote

control is not functional. Lot is in good physical condition.

Version: V3.12

Manufacture date: 01/97

Page 34: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

781 Laserscope

Greenlight HPS 2070

with Foot Switch (No

Key)

The laser unit has not got a key and therefore unable to power up.

Includes footswitch and the mains cable is an industrial plug. Overall

physical condition is fair. Single phase, Y.O.M 2006

795 Carl Zeiss

Microscope Parts x

3

Zeiss operating microscope binocular assembly x2. Both with good

optics. One with eyepieces 12,5x/18B, f170. The second one with eye

pieces 10x/22B, f170 with extension bridge.

808 Alcon Surgical

Series 20000 Legacy

Phaco System

Unit powers up, with units function buttons and touch screen display in

working order. The unit does not include any hand pieces and unit has

not been fully tested. Modes are selectable but have not been verified.

Units IV pole height adjustment is functional as are the audio voice

prompts. The remote control is functional. Unit does not include a foot

control. Lot is in good physical condition.

Version: V3.12

Manufacture date: 11/01

809 Kodak Dryview

8900 Laser Imager

Unit power up and the drive motors inside con be heard initializing but

the LCD screen remain blank.

809 Kodak Dryview

8900 Laser Imager

Unit powers up, however units power indication light above the display

is not coming on, also units touch screen display is not coming on or

functional. Unit shows to be not functional and unable to carry out any

further checks. Unit is in satisfactory physical condition.

810 Kodak Dryview

8900 Laser Imager

Unit not powering up and cover panel at the back is missing. Fair

physical condition.

810 Kodak Dryview

8900 Laser Imager

Unit sounds to power up for a brief second, before cutting out.

However the power indication light above the display does not come

on and neither does the touchscreen display. Unit shows to be not

functional and unable to carry out any further checks. Unit is in

satisfactory physical condition.

861 Philips Heartstart

XL (No Batteries)

The defibrillator powers up, with its display functional and its function

buttons looking to respond. The unit has not been tested. Unit includes

pacer function, which also has not been tested. Unit does not include

any accessories. The printer in functional, however the printer doors

hinge is broken making it difficult to close. Units battery is missing.

Unit is in satisfactory physical condition.

Manufacture date: 2004-03

Page 35: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

866 Bair Hugger Patient

Warmers x 2

Both patient warmers power up ok and the heaters are in good working

order. The units are in reasonable physical condition. Hoses are not

included.

875 Alcon Surgical

Series 20000 Legacy

Phaco System

Unit powers up, with units function buttons and touch screen display in

working order. The unit does not include any hand pieces and unit has

not been fully tested. Modes are selectable but have not been verified.

Units IV pole height adjustment is functional as are the audio voice

prompts. The foot control shows to be responding and the remote

control is functional. Lot is in good physical condition.

Version: V3.12

Manufacture date: 11/01

879 Olympus PSD 20

Diathermy with Foot

Switch

The diathermy is able to power up, with its displays functional and its

function buttons looking to respond. The unit has not been tested with

its output not being verified. Units patient plate alarm is functional.

Unit includes (1x) monopolar cut/coag footswitch which appears to be

responding. Unit is in good physical condition.

882 GE Dash 3000

Patient Monitor

The patient monitor powers up well and passed its initial self-test. The

display, functions and function buttons are in good working order. The

patient monitor has not been fully tested but appears to be functional.

The NiBP pump is functional but has not been fully tested. Includes

incomplete ECG lead, SpO2 trunk and NiBP hose. Battery has not

been fully tested. Overall physical condition is fair.

891 LTE Falcon 30

Autoclave

The autoclave power but with error displayed. Printer error and

instruction to switch unit off. Unable to test any of the cycle and the

door id disabled. Unit is in fair physical condition. But not fully tested.

896 Kodak Dryview 8900

Laser Imager

The laser imager power up ok and touch screen is fully functional. MS

Windows 2000 Professional start-up completed. The system self-test

which include warming up and calibration are completed successfully

and ‘Ready’ is displayed as the unit status. Test patterns were

successfully printed from all three film drawers. The system appears to

be working fine but not tested fully. Unit is in good physical condition.

897 Alcon Series 20000

Legacy Phaco

machine, 8 x Phaco

Hand Pieces with

Foot Switch, Remote

Control and

Manuals

Unit powers up, with its function buttons and touchscreen display in

working order. Unit has not been fully tested, however unit does

include (ref: TURBOSONIC-375) hand pieces (x8). Units IV pole

height adjustment is in working order. Units audio voice prompts are

functional. Unit includes (1x) foot control which looks to be

responding. Unit includes (1x) remote. Unit is in satisfactory physical

condition.

Page 36: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Version: V3.00

Manufacture: 03/97

899 Olympus WM-30

Mobile Workstation

with Olympus CV-

230 Camera Control

Unit. Olympus CLV-

U40 Light Source

and Sony PVM-

20M2MDE Monitor

All units are able to power up, lot does not include a pig tail or a

videoscope but some were used for test purposes. The processor

showed it could produce a clear image, however the image was far too

bright as the light sources auto brightness focusing is not functional.

The light source showed to give a light output in all light output modes

(OES, EVIS 100 and EVIS 200). However the auto brightness focusing

was not functional and in manual mode the manual light output

adjustment also did not function. The air pump part of the light source

showed to function. The monitor is functional. Lot is in satisfactory

physical condition.

900 Spembly Medical

140 Cryo Unit

The Cryo Unit’s control, footswitch and regulator appear to be ok. The

probe is not included. Lot is in fair physical condition.

902 Storz Hamou Micro

Hysteroflator

The insufflator power up ok the flow and pressure display and

adjustment keys are working fine. System is in good physical

condition.

904 Olympus PSD-20

Diathermy and MH-

551 Foot Switch

The diathermy is able to power up, with its displays functional and its

function buttons looking to respond. The unit has not been tested with

its output not being verified. Units patient plate alarm is functional.

Unit includes (1x) monopolar cut/coag footswitch which appears to be

responding. Unit is in good physical condition.

906 Vivident Astralis 7

Dental Curing Light

Unit

Dental curing light power supply is working ok, the lamp comes on

and the timer is working. The unit is still in good overall condition.

925 Alcon Ocuscan with

Foot Switch and

Probe Con# D4126

Ass#G21727

The Ocuscan power up ok, display and function keys are working. The

probe test could not be checked at the start up. Although the scan

graphs were displayed but the automatic reading and value were not

coming. Unit is in good physical condition but not fully tested. Two

ultrasound probes is included. The printer is working fine.

932 Sonacel Multiphon

Plus Ultrasonic

Therapy System

The treatment ultrasound unit power up ok The output intensity control

and moving coil meter is in good working order. The timer & US head

selector are functional. The output was tested and it is working fine.

Lot comes with a 1.5MHz and 3MHz head. Fair overall condition.

938 Gettinge 1345

Washer/Disinfector

and Pump

Unit power up ok, display and function keys are working. The unit is

stuck in a disinfection phase and unable to proceed. The door is locked

and the condition of the interior is unknown. The system is in good

physical condition but the process was not tested fully. Year: 2002,

Page 37: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

PAC 5,06 (320), Panel Version: 3.4. counter at 734 cycle. Grundfros X

water pump included

942 Kodak Dryview

8150 Laser Imager

Unit powers up, with units touch screen display in working order. Unit

has not been tested, but the menus look to be accessible. Unit is in

satisfactory physical condition.

Version: 1.1.2

969 Olympus UHI-3

Insufflator and MH-

317 Foot Switch

The insufflator powers up, with its function buttons looking to respond

and its numerical displays in working order. Unit has not been tested.

Unit includes (1x) footswitch. The insufflator is in good physical

condition.

973 Diomed 15 Plus

Laser and Ethicon

Endo Surgery Unit

Unit has no On/Off key – unable to check function – No hand piece or

laser probe included.

986 EMS Swiss

Lithoclast Master

Ultrasound System

and Microaire

Both unit powers up ok, the display and function keys are working.

The Lithoclast system has two of its push button missing. There are no

hand pieces to check the unit. Average physical condition.

1006 Philips Series 50XM

CTG Fetal Monitor

with US and Toco

and Pressure Cuff

Unit powers up, with its function buttons and displays in working

order. The units toco and cardio 2 parameters showed to be

functioning, but the outputs have not been fully tested. The

cardio1/combi parameter has not been checked as the port is damaged.

Units SPO2 shows to be functional as does the NiBP pump, neither of

the parameters outputs have been fully tested or verified. The printer

shows to be functional. The event marker is functional. Unit is in fair

physical condition.

Unit includes: (1x) toco transducer, (1x) US transducer, (1x) event

marker, (1x) SPO2 probe and (1x) NiBP cuff and hose.

1007 Philips Series 50XM

CTG Fetal Monitor

with US and Toco

Leads

Unit powers up, with its function buttons and display in working order.

The units toco and cardio1/combi parameters showed to be

functioning, but the outputs have not been fully tested. The cardio2

parameter is not functional giving ‘ERR’ message displayed. Units

NiBP pump is functional, but has not been fully tested. Unble to check

units SPO2 as unit does not include an SPO2 probe. The printer is

functional. Unit does not include an event marker. Unit is in fair

physical condition.

Unit includes: (1x) Toco transducer, (1x) US transducer and (1x) NiBP

Page 38: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

cuff

1008 Hoskins Cot and

Mattress x 5

The cots are all in good condition.

1019 Kodak Dryview 8150

Laser Imager

Unit is able to power up, with units touchscreen display showing to be

in working order. Unit has not been tested. Unit is in good physical

condition.

Version: 1.1.b04

1026 Kodak Dryview 8900

Laser Imager

Powers up but unable to complete the system boot-up (message on

display panel reads: invalid system disk). No further checks could be

carried out. The front corner of the outer casing of the unit has been

damaged otherwise it is in reasonable physical condition.

1030 Cots with Pentaflex

Paediatric

Mattresses x2

Fair physical condition

1031 Cots with Pentaflex

Paediatric

Mattresses x2

Fair physical condition

1032 Cots with Pentaflex

Paediatric

Mattresses x2

Fair physical condition

1033 Cots with Pentaflex

Paediatric

Mattresses x2

Fair physical condition

1035 Arjo Grey Electric

Tilt Table

Unit is able to power up, however units footswitch is damaged causing

the up and down movement to only be able to go down with its button

getting stuck in place. Units motorised forward/backward tilt showed

to be functional. The mechanical head tilt is functional. Unit is in fair

physical condition.

1040 Thermo Klenz

Washers - 1015547

The unit is able to power up, with its display in working order and its

function buttons looking to respond. Unit has not been tested. Unit is in

satisfactory physical condition.

1041 Thermo Klenz

Washers - 1015546

The unit is able to power up, with its display in working order and its

function buttons looking to respond. Unit is displaying message for

‘maintenance’. Unit has not been tested. Unit is in satisfactory physical

Page 39: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

condition.

1042 Thermo Klenz

Washers - 1015548

The unit is able to power up, with its display in working order and its

function buttons looking to respond. Unit is displaying message for

‘maintenance’. Unit has not been tested. Unit is in satisfactory physical

condition

1043 Thermo Klenz

Washers - 1015558

The unit is able to power up, with its display in working order and its

function buttons looking to respond. Unit has not been tested. Unit is in

satisfactory physical condition.

1046 Kodak Carestream

Dryview 8150 Laser

Imager

Powers up and completes the initialisation process successfully.

Display panel and touch screen interface are in good working order and

the system menus are accessible. No further checks carried out. The

unit is in reasonable physical condition. Software version: 1.1.B04

1052 Kodak Dryview 5800

Laser Imager

Powers up and completes the initialisation process successfully.

Display panel and function keys are in good working order and the

system menus are accessible. Includes 2x film trays (8x10-20x25 and

14x17-35x43). No further checks carried out. The unit is in good

physical condition. Software version: 2.4.B01P1.

1053 Kodak Dry View

5800 Laser Imager

Powers up and completes the initialisation process successfully.

Display panel and function keys are in good working order and the

system menus are accessible. Includes 3x film trays (8x10-20x25,

10x12-25x30 and 14x17-35x43). No further checks carried out. The

unit is in good physical condition. Software version: 2.3.B02.

1054 Lisa 500 Autoclave

and Printer - 090592

W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch

screen are fully functional. A vacuum tested cycle was carried out and

passed with no problem. Printer is working but need ribbon and paper.

The interior of the chamber and door gasket are in good condition. The

memory card MMC- SD and trays are not included. Unit is in good

physical condition with some scuff and marks on the covers. Counter at

1556. Type B autoclave, Year 2009

1055 Lisa 500 Autoclave

and Printer

W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch

screen are fully functional. A vacuum tested cycle was carried out and

passed with no problem. Printer is working fine. The interior of the

chamber and door gasket are in good condition. The memory card

MMC- SD and trays are not included. Unit is in good physical

condition with some scuff and marks on the covers. Counter at 1578.

Type B autoclave, Year 2009

Page 40: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

1056 Lisa 500 Autoclave

and Printer

W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch

screen are fully functional. A vacuum tested cycle was carried out and

unit failed the test – error A390 was displayed. Printer is working fine.

The interior of the chamber and door gasket are in fair condition. The

memory card MMC- SD and trays are not included. Unit is in fair

physical condition with some scuff and marks on the covers. Counter at

1674. Type B autoclave, Year 2009

1057 Lisa 500 Autoclave

and Printer

W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch

screen are fully functional. A vacuum tested cycle was carried out and

passed with no problem. However, a ‘Clock Battery Low’ message is

displayed on the screen. Printer is working fine. The interior of the

chamber and door gasket are in good condition. The memory card

MMC- SD and trays are not included. Unit is in good physical

condition with some scuff and marks on the covers. Counter at 1641.

Type B autoclave, Year 2009

1061 Agfa Drystar 3000

X-Ray Printer

Unit is able to power up, with it able to pass through its initial

initialisation successfully. Units display is functional, with its function

buttons looking to respond. Unit has not been test. Unit is in

satisfactory physical condition.

Software version: R1.95

1062 Marquette Mac 5000

ECG Recorder with

Leads - 16673

Units power indication light illuminates however units display does not

show to be coming on. unable to carry out any further checks. Unit

includes 12wire ECG. Unit is in fair physical condition.

1064 Stryker Transfer

Trolley

Units mechanical height adjustment is in working order, with the its

mechanical tilt also in working order. Units wheels are freely movable

with the brake lock functional. The cot sides are in place. The mattress

is in place and in fair physical condition. Unit is in satisfactory

physical condition.

1075 Maquet Electric

Operating Table

with Qty of Extra

Accessories and

Extensions

The operating table is in reasonable physical condition with all

cushions in place and the transport cart is also included. Remote

control is included but only the up/down motorised movement of the

table is functional, other functions appear to be not working (the motor

starts when the keys are activated from the remote control but no

movement is detected). The battery is in working condition.

Accessories included with the table consists of extensions, head rests

and leg rests.

Page 41: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

1078 Huntleigh Akron

Electric Patient

Couch - Green

Unit looks to be powering up, with its electrical height adjustment in

working order. The mechanical movements also show to be in working

order. The wheels are freely movable and the brake locks are

functional. Unit includes (1x) footswitch which is functional. Unit is

in fair physical condition

1080 Luxo, Daray x 2 and

Heine HL5000

Examination Lights-

100587, 6321, 126511

The two Daray examination lights are not powering up, the Luxo

examination light and the Heine wall mounted examination light are

both able to power up and produce a light output. All of the units

mechanical movements are work however there locks are not. All four

of the examination lights have bad damage such as cracks and

scratches to casing etc, bad dents to casing etc, loose casing etc. Lot is

in below satisfactory physical condition.

1081 Green Electric

Dialysis Chair

Unit powers up, with its electrical height adjustment in working order.

Units mechanical movements are also in working order. Units wheels

are freely movable and its brakes are functional. Unit includes (1x)

remote which is functional. Unit is in satisfactory physical condition.

1082 HNE Akron Electric

Couch - Red

Unit powers up, with its electrical movements in working order. Units

mechanical movements are also in working order. Units wheels are

movable and the brakes function. Unit includes (1x) remote which is

functional. Unit is in reasonable physical condition.

1085 Medtronic Lifepak

20 Defib and Leads -

13251 Serial

Number 30493063

The defibrillator powers up and is able to pass through its initial self-

test successfully. Units function buttons, rotary knob and display are in

working order. Unit does not include any paddles but does include (1x)

AED cable. Units output was tested using a defibrillator analyser and

all readings taken showed to be within set limits. Unit showed to be

functional in both manual and AED modes. Unit was able to charge to

360J. The defibrillator is biphasic. The audio voice prompts are

functional. Unit includes pacer function which showed to be functional

but however out of limits. Unit includes (1x) 3 wire ECG and units

ECG was verified as functional. Units SPO2 parameter has not been

tested but unit includes (1x) SPO2 trunk Units printer is functional but

the battery is not. Unit is in good physical condition.

Manufacture date: 2002

1088 Pentax EPM-3500

Camera

Control/Lightsource

- 055808 Serial

Number E0011304

The processor powers up, but unable to determine if units function

buttons are functional, none of the indication lights are on. The unit has

not been tested. Unit is in satisfactory physical condition.

Manufacture date: 2004

Page 42: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Lamp hours: 0.0

Total hours: 1693.5

1095 SLE 2000 Infant

Ventilator

The ventilator is able to power up, with its numerical displays in

working order. Unit is missing its reset function button but the rest of

units function buttons are in working order, also units CPAP/PEEP

regulator is damaged and not functional. The ventilator showed to be

functional when it was tested on a medical compressor. However the

PEEP cannot be adjusted and CPAP mode is not able to function as the

CPAP/PEEP regulator is not working. The rest of the vent modes

(SIMV, PTV and CMV) showed to be in working order. Units output

has not been verified. Unit is able to be calibrated to 21%. Unit does

not come with any O2 or air hoses, also unit is missing its stand. The

ventilator is in reasonable physical condition.

Vent modes: SIMV, PTV, CMV and CPAP

Operation hours: 4563.25h

1111 Welch Allyn CP10

ECG Recorder with

ECG Lead

Unit is able to power up, units function buttons and display are in

working order. Unit has not been fully tested. Unit includes 12 wire

ECG. No paper to check the printer. The battery is functional. There is

a rattling sound with something loose inside of the unit. Unit is not on

a trolley. Unit is in fair physical condition.

Software version:1.06

1115 Ascena Syringe

Pumps x 18 - 7525,

103094, 054391,

101273, 118485,

7044, 101577, 101621

The syringe drivers are in satisfactory physical condition. 4 of the units

have damaged/broken cases and 2 of the units have parts removed from

inside. Not tested.

1124 Endoscopy Stack

Trolley with Sony

Trinitron Monitor

and Olympus

Surgical CO2

Insufflator

Unable to test lot as the processor does not include a camera head. The

processor does however power up, with its indication lights in working

order and its function buttons looking to respond but not verified. The

Sony monitor also powers up but again has not been tested. The lot is

in satisfactory physical condition.

1125 Fisher and Paykel

Neopuff Infant

Resuscitator

Unit powers up, with its function buttons and numerical displays in

working order. Unit showed to heat up well in ‘prewarm’, ‘Manual’

and ‘Baby’ modes. With the heat adjustment in ‘Manual’ and ‘Baby’

modes functional. ‘Baby’ mode showed to be in working order. Unit

includes (1x) skin temperature probe which is functional, however its

reading has not been verified. Units examination light and timer are

both functional. Units electrical height adjustment is functional but

does jolt when raising up the unit. The Neopuff part of the unit has not

Page 43: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

been checked. Unit includes an O2 hose and a mattress. Unit is in

satisfactory physical condition.

1126 Fisher and Paykel

Neopuff Infant

Resuscitator

Unit powers up, with its function buttons and numerical displays in

working order. Unit showed to be able to heat up well in ‘prewarm’,

‘manual’ and ‘baby’ modes. With the heat adjustment in ‘manual’ and

‘baby’ modes functional. However after a brief period of time the unit

displays alarm message ‘sensor disconnected’ with the numerical

display stopping, this happens when the skin temperature probe is

actually in place. ‘Baby’ mode is not fully functional. Unit includes

(1x) skin temperature probe. Units examination light and timer are both

functional. Units electrical height adjustment is functional. The

Neopuff part of the unit has not been checked. Unit includes an O2

hose and a mattress. Unit is in satisfactory physical condition.

Manufacture date: 2004-11

1129 Siemens Servo

Ventilator 300 and

Parts

The ventilator powers up, with units function buttons, indication lights

and numerical displays in working order. The ventilator showed to be

functional when it was tested using a medical compressor. With the

BPM, tidal volume and minute volume all matching up accordingly.

Units output has not been verified. Unit showed to be functional in

both volume and pressure modes. Units O2 was able to calibrate to

21%. Units battery functions but alarms for being low. Unit includes

O2 and air hoses. The ventilator is in satisfactory physical condition

with t however not including a stand and there are also chips to

paintwork around edges of unit.

Operation hours: 30950h

1130 Siemens Servo

Ventilator 300 and

Parts

The ventilator powers up, with units function buttons, indication lights

and numerical displays in working order. The ventilator showed to be

functional when it was tested using a medical compressor. With the

BPM, tidal volume and minute volume all matching up accordingly.

Units output has not been verified. Unit showed to be functional in

both volume and pressure modes. Units O2 was able to calibrate to

21%. Units battery functions but alarms for being low. Unit includes

O2 and air hoses. The ventilator is in satisfactory physical condition

with it however not including a stand and there are also chips to

paintwork around edges of unit.

Operation hours: 31147h

1131 Drager Photo

Therapy 4000 and

8000 Therapy

Drager photo therapy lamp 4000 & 8000. The 4000 has both lamp type

coming on and the counter is at 125831hrs. The 8000 lamp is working

fine with the fan and alarm tested ok. Counter at 296639hrs. Both stand

Page 44: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

Lamps and castors are in fair condition.

1143 SLE 2000 HFO

Ventilator

The ventilator powers up, with units main display and numerical

displays in working order. Units function buttons look to be

responding. Unable to test the ventilator as the unit is missing its

expiratory valve. Unit includes O2 and air hoses. Units blended gas

output connection is badly damaged. The ventilator is in satisfactory

physical condition.

Vent modes: SIMV, PTV, CMV and CPAP

Operation hours: 25618.91h

1144 SLE 2000 Infant

Ventilator

The ventilator is able to power up, with its function buttons, numerical

displays and gauges in working order. The ventilator showed to be

functional when tested with a medical compressor. All ventilation

modes showed to be in working order. Units output has not been

verified. Units oxygen was able to calibrate to 21%. Unit includes O2

and air hoses. The ventilator is in satisfactory physical condition.

Vent modes: SIMV, PTV, CMV and CPAP

Operation hours: 6839.96h

1145 Dantec Dynamic

Keypoint EMG

Stimulator

The unit is able to power up, with units function buttons and monitor

functional. The unit has not been tested and unit does not include any

accessories or leads. Unit is able to enter its main menu. The unit is in

reasonable physical condition.

V2.12

1151 Engell

Fridge/Freezer

The unit is able to power up, the unit shows signs it is able to start to be

able to cool down, however the unit has not been fully tested. Unit is

fair physical condition.

1153 Kodak Dryview 8900

Laser Imager

Unit is able to power up, with its touchscreen display in working order.

The unit has not been tested. Unit is in satisfactory physical condition.

Version: DV8900_2.14

1154 Dometic Pharmacy

Refrigerator

The unit is able to power up, with the unit not showing any signs it is

able to cool down, however the unit has not been fully tested. Unit

includes its key for its door lock. Unit is in satisfactory physical

condition.

1156 Fixed Height Patient

Couch - Black

The couch is in satisfactory physical condition, with some rips at

corners of the cushions. The back movement is freely movable.

Page 45: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

1158 Grey Electric

Dialysis Chair

The unit is able to power up, with units electrical movements showing

to be in working order. Unit includes (1x) remote. Units wheels show

to be freely movable. Units cushions have some rips to the corners.

Unit is in reasonable physical condition.

1169 HP Pagewriter 200I

ECG Recorder with

ECG Lead

Unit powers up, with units function buttons looking to respond and its

display functional however there are lots of pixels missing from it. Unit

has not been tested. Unit include 12 wire ECG. Unit does not include

any paper to test the printer. The battery is functional. Unit is not on a

trolley. Unit is in fair physical condition.

1175 Flat Screen Monitors

x 3

The two flat screen monitors both power up, however the they have

large cracks/damage to the screens and the two monitors are not

functional. The Dell monitor is able to power up, not fully tested. Lot is

in poor physical condition as the screens are badly damaged.

1180 Braun SCT 2

Tourniquets x 2 -

14822, 18867

Both units appear to have their gauges and regulators intact. Lot has

not been tested. Both units are complete with their correct hoses. Lot is

in reasonable physical condition with chips to paint work.

1190 Trolley with Siemens

Unity Audiology

Test Equipment

The unity PC probe Mic is able to power up, with it not being tested

and neither has the PC hearing instruments analyser. There are no

complete leads or accessories with lot so unable to test unit. The Hi-pro

powers up. Lot is in fair physical condition.

1192 FMS Duo with

Arthroscopy Fluid

Management System

with Integrated

Shower and Foot

Switch

Unit powers up, with its displays functional and its function buttons

looking to respond. Unit has not been tested. Unit is in complete and

missing its shaver. Unit includes (1x) foot control but unable to verify

if it is responding. Unit is in fair physical condition, with the suction

pumps cover missing.

1207 Quayle Deroter and

Audio/Hearing Aid

Fitting Tool with

Dust Extractor, Foot

Pedal and Drill Head

- MS3031,

CP0027975

The polisher is able to power up, with the polisher and fans functional.

Some of the vacuum pumps wires have been cut so unable to power up

unit. Unit includes a micro drill, with its hand piece included, the micro

drill has not been tested. Unit includes (1x) footswitch which has

exposed cables wires where the mains cable is coming away from the

footswitch. Unit includes a stand for the polisher. Unit includes a box

of hearing accessories (not checked). Lot is in reasonable physical

condition.

1218 Medilink System

Treatment Unit -

Cased

Unit is able to power up, with its function buttons and displays in

working order. Unit has not been fully tested, but module modes are

selectable (ultrasound and shortwave). Unit includes (2x) ultrasound

Page 46: Endoscopes Inspection Report – Sale ------ date ... Report Sale 1406.pdf · HG.115/116.v2 Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014 (Condition Reports for

HG.115/116.v2

transducers, both 1.1 and 3.3 MHz with crystal areas (7.0cm2 and

0.79cm2). Unit also includes shortwave (1x) disk electrode. Unit is in

fair physical condition.

Version: v1.20

1227 Braun Tourniquet x

2

Both tourniquets are in fair condition with the regulators. Manometer

and pneumatic switch appears to be functional. Units were not tested

with medical air.

1229 Datex Normocap

200 x 2

Only one of the gas monitor are working. The working ones has its

display and function keys working fine. Co2 function is working but

battery failure and air leak detected is displayed. There is a piece of the

cover chipped off near the water trap assembly. The second unit is not

powering up at all.

1231 Aesculap Drill Unit

Type: GA 244

The main drill power up but there is no foot switch or hand piece to

test functionality. The irrigation pump is not powering up. Lot in

average physical condition.

1248 Mobile Eye Chart Unit powers up, with snellen back light is illuminating. Unit is missing

its plastic caster for the stand. Unit is in poor physical with damage to

the plastic graphs.

1249 Lisa 500 Autoclave

and Printer

W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch

screen are fully functional. A vacuum tested cycle was carried out and

passed with no problem. ‘Dust filter needs replacement. Printer is

working fine. The interior of the chamber and door gasket are in good

condition. The memory card MMC- SD and trays are not included.

Unit is in good physical condition with some scuff and marks on the

covers. Counter at 1560. Type B autoclave, Year 2009

1250 Lisa 500 Autoclave

and Printer

W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch

screen are fully functional. A vacuum tested cycle was carried out and

passed with no problem. Printer is working fine. The interior of the

chamber and door gasket are in good condition. The memory card

MMC- SD and trays are not included. Unit is in good physical

condition with some scuff and marks on the covers. Counter at 1578.

Type B autoclave, Year 2009