HG.115/116.v2
Endoscopes Inspection Report – Sale: 1406 Date: 13 / 05 / 2014
(Condition Reports for other items at the end of the Scopes Report)
These reports are compiled by our Medical Engineering Team & are for guidance only; all lots are sold as seen,
as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase.
Lot Description Condition
110 OLYMPUS
CF-1T200L
Video Scope
1600x15mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – Functional
Distal end – Good
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition and image quality are good
111 OLYMPUS
CYF-4A
Direct View
350x5.54mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition and image quality are good
112 FUJINON
UG1-FP2
Direct View
1000x10.30mm
Insertion tube – Fair condition
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Average condition as some physical damage is present
Image Quality – Fair image
Working Channel (s) – Clear
Leak test – Leaking
Over all condition – Overall physical condition and image quality are in
fair condition but scope is leaking
113 OLYMPUS
BF TYPE P30
Insertion tube – Fair condition but has kinks on the insertion tube near the
control box
Bending mechanism & Angulation control – Good movements
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Good condition
HG.115/116.v2
Direct View
540x5.58mm
Image Quality – Poor image with 50+ fibres missing and water damage
Working Channel (s) – Clear
Leak test – Fail - Leaking
Over all condition – Overall physical condition is fair but the image is poor,
water damage is visible and has 50+ fibres missing
114 OLYMPUS
CYF-4A
Direct View
350x5.65mm
Insertion tube – Fair condition but has a kink on the insertion tube near the
control box and there is a dent near the distal end
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition is fair but the image is good
115 OLYMPUS
CYF-4
Direct View
350x5.78mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition is good and the image
quality is good
116 OLYMPUS
BF TYPE 240
Video Scope
500x6.20mm
Insertion tube – Fair condition but dent along the insertion tube
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – Functional
Distal end – Good condition
Image Quality – Image is average but the light does not focus and therefore
shows bright regions on the output image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition is fair but the image quality
is poor as the light is not focused and shows bright patches
117 OLYMPUS
CYF TYPE 240
Insertion tube – Fair condition but has a kink near the control box
Bending mechanism & Angulation control – Movements and locks are
functional but is stiff
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – Not functional
Distal end – Good condition
HG.115/116.v2
Video Scope
350x5.75mm
Image Quality – Image is poor as the image is blurred
Working Channel (s) – Clear
Leak test – Fail- Leaking
Over all condition – Overall physical condition is fair but the image quality
is poor as the image is blurred and the scope is leaking
118 OLYMPUS
CYF TYPE 240
Video Scope
350x5.65mm
Insertion tube – Fair condition but has a kink near the control box
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – Functional
Distal end – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition is fair and the image quality
is good.
119 OLYMPUS
CYF TYPE 240
Video Scope
350x5.48mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – Functional
Distal end – Good condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Overall physical condition is fair and the image quality
is good.
120 OLYMPUS
CF-200S
Videoscope
600x14.80mm
Insertion tube – Satisfactory condition
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– Good
Objective lens and focusing – Good optics
Remote control switches – Fully functional
Distal end – Satisfactory condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Satisfactory overall condition
121 OLYMPUS
CF-240AL
Videoscope
1600x12.90mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– Good
Objective lens and focusing – Good optics
Remote control switches – Fully functional
Distal end – Satisfactory condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
HG.115/116.v2
Over all condition – Good overall condition
122 OLYMPUS
BF-P30
Direct view
500x5.30mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movement
Light guide cable & Light transmission– Good
Objective lens and focusing – Satisfactory
Remote control switches – N/A
Distal end – Good condition
Image Quality – Below satisfactory, bad water damage to whole image and
25+ fibres missing
Working Channel (s) – Clear
Leak test – Leaking
Over all condition – Below satisfactory overall condition as scope has
water damage and 25+ fibres missing to image, also the scope is leaking
123 OLYMPUS
SIF-230
Videoscope
2000x12.00mm
Insertion tube – Satisfactory condition
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– Good
Objective lens and focusing – Good optics
Remote control switches – Fully functional
Distal end – Satisfactory condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Leaking when movement controls are moved to maximum
Over all condition – Good overall condition, however there is a leak when
movement controls are moved to maximum
124 OLYMPUS
CF-240AL
Videoscope
1600x13.0mm
Insertion tube – Very good condition
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– Good
Objective lens and focusing – Good optics
Remote control switches – Fully functional
Distal end – Satisfactory condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Pass
Over all condition – Very good overall condition
125 OLYMPUS
CF-H260AZL
Videoscope
Insertion tube – Very good condition
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– Good
Objective lens and focusing – Unable to test as do not have the correct
pigtail to test with
Remote control switches – Unable to test as do not have the correct pigtail
to test with
Distal end – Satisfactory condition
Image Quality – Unable to test as do not have the correct pigtail to test
with
Working Channel (s) – Clear
Leak test – Leaking
HG.115/116.v2
1600x14.10mm Over all condition – Good physical condition, unable to test scopes
working condition as do not have the correct pigtail to test with. The scope
however is leaking
126 OLYMPUS
GIF-PQ20
Direct view
1000x9.90mm
Insertion tube – Satisfactory physical condition slightly stained in colour
and slight dent near bottom of insertion tube
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– One of the light outputs is not
functional
Objective lens and focusing – Good optics, however there are 15 fibres
missing from image
Remote control switches – N//A
Distal end – Satisfactory physical condition
Image Quality – Satisfactory image as there are 15 fibres missing from
image
Working Channel (s) – Clear
Leak test – Leaking
Over all condition – Below satisfactory physical condition, 15 fibres
missing from image, one of the light outputs is not functional and scope is
leaking
127 OLYMPUS
PCF-230
Videoscope
1300x12.0mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movement and locks
Light guide cable & Light transmission– Good
Objective lens and focusing – Good optics
Remote control switches – Fully functional
Distal end – Satisfactory condition
Image Quality – Good image
Working Channel (s) – Clear
Leak test – Leaking
Over all condition – Good condition, scope is leaking
176 VISION
SCOPE
-
Direct View
280x3.22mm
Insertion tube – Poor condition where there is a dent along the tube, kinks
near the control box and the outer insulating sheath near the distal end is
missing
Bending mechanism & Angulation control – Good movements
Light guide cable & Light transmission– Both light channels are functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Fair condition
Image Quality – Fair image but there is slight water damage visible on the
image near the edge
Leak test – Unable to test as there is no connector at time of testing
Over all condition – Overall physical condition is poor due to the condition
as the insertion tube but the image is fair but slight water damage is visible
177 VISION
SCIENCES
ENT – 2000
Insertion tube – Poor condition where there is a dent along the tube and
kinks near the control box
Bending mechanism & Angulation control – Good movements
Light guide cable & Light transmission– Light channels are functional
HG.115/116.v2
Direct View
300x3.49mm
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Good condition
Image Quality – Good image
Leak test – Pass
Over all condition – Overall physical condition is poor but the image
quality is good
181 OLYMPUS
CYF-4A
Direct View
350x5.56mm
Insertion tube – Good condition
Bending mechanism & Angulation control – Good movements and locks
Light guide cable & Light transmission– Both the light channels are
functional
Objective lens and focusing – Functional
Remote control switches – N/A
Distal end – Good condition
Image Quality – Good image
Working Channel - Clear
Leak test – Pass
Over all condition – Overall physical condition is good and the image
quality is good
HG.115/116.v2
Condition Report – Sale: 1406 Date: 13 / 05 / 2014
These reports are compiled by our Medical Engineering Team & are for guidance only; all lots are sold as seen,
as per our Conditions of Sale. We recommend that wherever possible you inspect lots prior to purchase.
Lot
Number
Description Report
1 Datascope System
98XT Balloon Pump
with 1x Box of
Acessories
Powers up – display screen and control panel are in good working
order and the printer is also functional. Doppler is included and the gas
cylinder and adult condensate removal/safety disk are in place. Overall
the unit is in good physical condition. Not fully tested. Includes
manual, ECG lead set and a spare gas cylinder. Operating hours: 2198.
2 Zeiss Operating
Microscope
The operating microscope powers up, with unit able to produce a good
light output. The mechanical movements and locks are in working
order. The operating microscopes optical system shows to produce a
clear image, with unit showing to have good focusing with the
magnification functional. The unit is in fair physical condition.
3 Haag-Streit 900 Slit
Lamp with Table
and Accessories Case
The slit lamp is able to power up, with it able to produce a light output.
The fixation light is in place and shows to function. Units joystick
movements shows to be freely movable, with its lock functional. Units
chin rest is in place and its chin adjustment is in working order. The slit
lamps tables mechanical height adjustment is in working order. The slit
lamp is missing its eyepieces, however a set of eyepieces were used for
tested purposes and showed the rest of the optical system to be able to
produce a clear image. Overall the slit lamp is in satisfactory physical
condition, with the side compartment on the table missing its door.
4 Drager Fabius
Anaesthetic Machine
The anaesthetic machine was tested using a medical compressor and
the anaesthetic machine showed to have its gauges and flow meters in
working order. The anti-hypoxic safety feature is in working order and
the O2 flush is functional. The gas cut off alarm is functional. The
anaesthetic machine includes O2, air and N2O hoses. The unit is in
good physical condition.
Manufacture date: 2001
6 Zeiss OPMI-1
Operating
Microscope
The power supply comes on and all the selected outputs power is
working. The lamp is working fine. The optical system needs cleaning
up but still with fair magnification and focus. All mechanical
movements and locks are functional. Magnification range – 0.4 to 2,5
and eye pieces 12,5x. Objective lens f=200. System is in fair overall
physical condition.
HG.115/116.v2
7 Medtronic Lifepak
20 Defib With ECG
Leads x1
The defibrillator powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
defibrillator is testes using a defibrillator analyser and the output
energies are within the limits. The AED and manual mode are in good
working order. The ECG signal is tested using an ECG simulator and,
the defibrillator appears to give good strong traces of ECG signals.
Pacer option is included, which has not been fully tested but appears to
be functional. Printer is functional. The service indicator stays on.
Includes III lead ECG, AED therapy cable and two defibrillator pads
(expire 03/2014 and 04/2014). The overall physical condition is
reasonable. The defibrillator is a Biphasic unit and its Y.OM is 2004.
8 Aloka Echo Camera
SSD-500 Ultrasound
with Probe Aloka
7.5MHz
The ultrasound powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
ultrasound is tested using a phantom and set measurements taken are
within the limits. The resolution is average. Ultrasound includes modes
B and M, which are in good working order. The probe (model: UST-
5512U-7.5MHz) is in good physical condition and is intact. Overall
physical condition is fair.
10 Pfizer Valleylab
Force 30 Diathermy
(No Foot Switch)
The diathermy powers up well and passed its initial self-test. The
display and function buttons are in good working order. Some of the
function buttons has some wear and tear. The diathermy has not been
fully tested but the patient monitoring system is in good working order.
No footswitches are included. Overall physical condition is reasonable.
16 Zoll M Series Defibs
x 2
Both the defibrillators powers up well, its display, functions and
function buttons are in good working order. The defibrillators have
been tested using a defibrillator analyser and the output energies are
within the limits. The ECG is tested using an ECG simulator and the
defibrillators give good strong traces of ECG signals. The pacer option
is included, which has not been fully tested but appear to be functional.
The printer is functional. The defibrillators include two batteries which
appear to be flat and are in need of charge. Two AED therapy cables
are included but no ECG leads are included. Overall physical condition
is reasonable.
24 Wolf Stack Trolley
With Panasonic
M2080 Monitor,
Wolf 5512 Processor,
Wolf 5124 Light
Source, Wolf
Camera Head and
Wolf 2232 Laparo
CO2 Insufflator
All units power up. The processor and camera head are able to produce
a satisfactory image on the Panasonic monitor, with the camera head
also showing to have satisfactory focusing. The light source is able to
produce a good light output, with its brightness adjustment in working
order. The insufflator has not been tested but is able to power up ok.
Overall lot is in satisfactory physical condition.
HG.115/116.v2
25 Smith and Nephew
Stack System Sony
Trinitron Monitor
PVM-20L2MD, 640
Image System 300
XL Xenon Light
Source, 350 3-CCD
Camera Operator
and Camera Head,
Sony Colour Video
Printer UP-21 MD
(037097)
All units are able to power up. The camera head and processor are able
to produce a satisfactory image on the Sony monitor, with the camera
head also showing to have satisfactory focusing. The light source is
able to produce a good light output with its brightness control
functional. The 640 image management system is able to power up and
works on the National Display Systems flat screen, but in the boot up
unit displays message ‘CMOS checksum error’ which is able to be
passed. Unit is able to power up successfully and shows to be
functional. The color video printer is functional. Overall lot is in
satisfactory physical condition.
Image management system Version: 2.3
28 Philips Intellivue
MP70 Patient
Monitors x2
The two patient monitor’s powers up well and passed its initial self-
test. Its display and LCD touchscreen are in good working order. The
monitors do no detect the multi-parameter and recorder modules.
Therefore unable to carry out any further checks. No leads are
included. Overall physical condition is reasonable. Y.O.M is 2004 and
2005. Software Versions are: REV B.05.67 and REV F.01.42
29 Datex Capnomac
Ultima Patient
Monitors x2
Both monitors power up ok and the display screens and control panels
are in good working order. CO2 functions appear to working but gas
monitoring functions have not been tested. Overall the units are in
reasonable physical condition. The alarm silence buttons on both
monitors have been worn out.
30 Penlon AV900
Ventilator
The ventilator powers up well and passed its initial self-test. The
display, LCD touchscreen and function buttons are in good working
order. The unit has not been fully tested but appears to be functional.
Oxygen hose is included. Overall physical condition is as there are
scratches on the top cover. S/W is V3.11
31 Penlon AV900
Ventilator
The ventilator powers up well and passed its initial self-test. The
display, LCD touchscreen and function buttons are in good working
order. The unit has not been fully tested but appears to be functional.
No hoses are included. Overall physical condition is reasonable. S/W is
V3.30
34 Graseby 3400
Syringe Pump, 3200
Syringe Pump and
Ivac P6000 Syringe
Pump x2, Dinamap
Pro 100V2 Patient
Monitor and Nellcor
NBP-4000 Patient
One of the IVAC syringe drivers is not powering up neither is the
Dinamap. The Nellcor power s up but the display is not displaying
anything. The rest of the units included in lot are powering up with
their function buttons and displays looking to be in working order.
None of the units have been tested. The lot is in fair physical condition,
with the Dinamap’s printer door missing, and part of the Nellocrs
handle missing. No accessories included with lot.
HG.115/116.v2
Monitor
41 Huntleigh Flowtron
Pumps x5
Unable to check to units as their plugs are missing, the other three units
power up , with pumps looking to function however they have not been
fully test. Overall the lot is in reasonable physical condition.
43 Storz Stack Trolley
with Storz 20090231
Monitor. NDS
Radiance Monitor,
Storz Xenon Nova
20131520 Light
Source, Storz Image
22200020 Camera
Control Unit, Storz
CD Re-Writer and
Keeler Headset
All units are able to power up. Unit does not include a camera head so
unable to test the processor. The light source is missing its
interchangeable light guide adaptor; however the light source is able to
produce a good light output. The imaging unit is able to power up, and
then is able to open the AIDA compact programme but it fails to fully
load as message ‘No dongle found’ is displayed which is unable to be
passed. The head gear is able to power up and is able to produce a light
output with its brightness adjustment functional. Overall lot is in
satisfactory physical condition.
44 Valley Lab Force 2
Diathermy - No
Pedals
The diathermy has its switch missing, therefore unable to power up and
carry out any further checks. No footswitches are included. Overall
physical condition is fair.
45 ATL HDI 5000
Ultrasound with
Mitsubishi
CP200,C5-2,L12-5
and C8-4V Probes
The ultrasound powers up but does not go into its main menu screen,
therefore unable to carry out any further checks. The lot includes four
probes (models: L7-4, C5-2, L12-5 and C8-4v), which are physically
intact but have not been tested. Printer is included, which powers up
but has not been tested. Overall physical condition is reasonable.
50 GE Dash 4000
Patient Monitor,
Datex Ohmeda S/5
Patient Monitors x2
(Damage to One)
The three patient monitors power up well, and passed its initial self-
test. The Dash 4000 powers on for a few minutes but turns itself off;
also its display monitor has lines running through the LCD screen. The
Datex Ohmeda (S/N: 4293223) has many pixels missing and therefore
unable to clearly read the display monitor. The monitor appears to be
functional, where it’s ECG and NIBP are tested but has not been
verified, unable to test the SPO2 as display set up is needed to be
accessed but difficult to read the display monitor. This monitor has a
number of cracks along the casing and its rotary knob is missing.
Recorder is missing but parameters IBP, ECG, SPO2, Temp and NiBP
are included.
The other monitor (S/N: 4749116) is in good working order where its
ECG is fully tested using an ECG simulator and is functional. The
NiBP and SPO2 appear to be functional but have not been fully tested.
The printer is included and functional. Included parameters NiBP,
ECG, SPO2, IBP, CO2 and Temp. Two batteries are included but are
flat and in need of a charge. Lot includes x1 AC adaptor but no leads
are included. Overall physical condition is average due to usage.
HG.115/116.v2
65 Datex Ohmeda
Aestiva 3000
Anaesthetic Trolley
with Ventilator
The anaesthetic machine powers up ok and the pressure gauges, flow
meters, O2 flush; Basel flow and anti-hypoxic safety feature are all in
good working order. The 7900 Smart Vent powers up ok and
completes the system self-test successfully. The function buttons are
working fine but the display has several pixels out, however the
displayed values are not affected. The ventilator was tested with test
lung and medical air and the set parameters and alarm limits were
working fine. The flow and oxygen sensors are working fine. The
absorber, bellows and rebreathing assembly are in good condition with
no leak detected during test run. The backup battery for the Smart Vent
is functional. Overall, the system is in good physical condition. O2,
Air, N2O hoses and AGSS receiver are included
Smart Vent Software Version: 3.2
66 SLE 2000 Infant
Ventilator with SLE
2100 Respiration
Monitor
The ventilator powers up, with its function buttons and numerical
displays in working order. Unit is missing its expiratory block, but one
was used for test purposes. The ventilator showed to be functional,
when it was tested using a medical compressor. The ventilator showed
to be functional in all ventilation modes. Units output has not been
verified to see if it is within set limits. Unit oxygen is able to calibrate
to 21%. Unit includes O2 and air hoses. The SLE 2100 is able to power
up, with it then passing through its initial self-test successfully. Units
display is functional and the function buttons look to be responding.
The SLE2100 has not been tested. The humidifier is also powering up
but again has not been test. The lot is in satisfactory physical condition.
Vent modes: SIMV, PTV, CMV and CPAP
Operation hours: 12835.53h
68 Eschmann TD850
Diathermy with
Yellow and Blue
Foot Switch on ST80
Suction Trolley and
Lina Smoke
Evacuator
The diathermy powers up well, its display, functions and function
buttons are in good working order. The diathermy is tested using an
ESU analyser and the monopolar (cut and coag) output results are out
of limits. Bipolar is not tested as there are no footswitches. Includes
monopolar footswitches (cut and coag). The ST 80 suction has its
mains cable cut off and therefore unable to fully test. The Lina suction
powers up and appears to be functional but has not been fully tested.
Overall physical condition is reasonable.
69 Olympus CLE-4U
Light Source and
Atlantech Xenon
Fibre-Optic Light
Both units are able to power up, however neither unit was able to
produce a light output. The Olympus CLE-4U has its air feeding pump
functional. Units are in below satisfactory physical condition.
HG.115/116.v2
Source
70 Medtronic Lifepak
20 x1, Physio
Control Lifepak 9
with Shock Advisory
Adapter x1
Lifepak 20
The defibrillator is able to power up and is able to pass through its
initial self-test successfully. Units function buttons, rotary knob and
display are in working order. Unit does not include any paddles but
does include (1x) AED cable. Units output was tested using a
defibrillator analyser and all readings taken showed to be within set
limits. Unit was functional in both manual and AED modes. Unit was
able to charge to 360J in manual mode. The defibrillator is biphasic.
The audio voice prompts are functional. Unit includes 3 wire ECG
leads and units ECG was verified as functional. Units printer is
functional but the battery is not. Unit is in good physical condition.
Lifepak 9
The defibrillator is able to power up, with its function buttons and
display in working order. Unit does not include any paddles but
includes a shock advisory adaptor and AED cable. Unit has not been
fully tested. Unit was able to give a test output of 360J, but the output
has not been verified. Only manual mode has been checked. The shock
advisory adaptor has not been checked. Unit includes (1x) 3 wire ECG
and units ECG has been verified as functional. Units printer is
functional and the battery is low. Unit is in fair physical condition.
Lifepak 9 manufacture date: 1999
Lifepak 12 manufacture date: 2006
74 Drager Oxylog 2000
Portable ventilator
Unit power up and system self-test completed successfully. The lithium
battery and mains charger is in good working order. LCD and controls
are functional. Airways pressure indicator and alarm setting are
working fine. The ventilator modes include: IPPV, SIMV, CPAP and
SIPPV and they all appears to be functional. Test run with medical air
compressor and test lung, indicates that all parameter and alarm
settings were in good working order. The minute Volume (MV) was
verified with the set valve of Tidal volume and Frequency (BPM) and
found to be within limit. The system is in good physical and cosmetic
condition. Oxygen hose with schrader connector is included.
76 Graseby 3200
Syringe Pumps x5,
Graseby 3100
Syringe Pumps x3
and Alaris GH
Syringe Pump and
Abbot Flexiflow
The Alaris Plus syringe driver is not power up and one of the Graseby
3100 displays ‘fault 86’ which is not able to be passed. The rest of the
units are able to power up, with their function buttons and displays in
working order. Units have not been tested. The lot is in fair physical
condition.
HG.115/116.v2
Enteral Pump
88 Olympus PSD 20
Diathermy with Foot
Switch
Unit is able to power up, with it numerical displays in working order
and its function buttons looking to respond. Unit has not been tested
with its output not checked. Units patient plate alarm is functional.
Unit includes (1x) footswitch for monopolar cut/coag, which looks to
be responding. Unit is in good physical condition.
92 Cardiac Science
AED Trainer x2,
Lifepak CRT
Trainer x1
The lifepak CR-T does not include a battery so unable to determine if
the unit is able to power up.
Both the Cardiac trainers are able to power up, however only one of the
units includes batteries. Both units’ displays are functional. Both units
voice prompts are functional. Both units were tested on a defibrillator
analyser, one unit was unable to find a signal the other unit was able to
find a signal but then no actual output (0J) was recorded when a
shock was delivered.
Overall lot is in satisfactory physical condition.
103 Omega Excel Laser
System with 1 x
Probe, Manual and
Key
The laser unit powers up well and passed its initialization phase. The
display, functions and function buttons are in good working order. The
beam test was conducted and appears to be in good working order.
Includes probe which is in good working order (200mW and lambda
820nm). Key is included. Overall physical condition is fair. Y.O.M
2001
142 Smith and Nephew
30º Autoclavable
Rigid Scopes x 4
The four rigid scopes appear to be physically intact and are inclined at
30 degrees. Four of the rigid scopes do not produce a good output
image. Overall physical condition is reasonable.
143 Smith and Nephew
30º Autoclavable
Rigid Scopes x 4
The four rigid scopes appear to be physically intact and are inclined at
30 degrees. Four of the rigid scopes do not produce a good output
image. Overall physical condition is reasonable.
144 Smith and Nephew
30º Autoclavable
Rigid Scopes x 4
The four rigid scopes appear to be physically intact and are inclined at
30 degrees. Four of the rigid scopes do not produce a good output
image. Overall physical condition is reasonable.
145 Smith and Nephew
30º Autoclavable
Rigid Scopes x 4
The four rigid scopes appear to be physically intact and are inclined at
30 degrees. Four of the rigid scopes do not produce a good output
image. Overall physical condition is reasonable.
175 Olympus MAJ-554
Camera Head
Unable to test the camera head as there is no processor to test with. The
Camera head looks to be intact and in satisfactory physical condition.
HG.115/116.v2
178 Downs 0° Rigid
Scope with Light
Lead
The rigid scope is physically intact and gives a good output image. The
light source cable is included which appears to be intact. Overall
physical condition is fair.
183 GE 4C-RS Curved
Array Convex Probe
GE Curved Array Convex Probe, 4C-RS (year 2007) -The probe
connection, cable and head are intact and in good physical condition;
but not tested due to unavailability of suitable unit. The probe is
compatible with model GE Logiq e, Vivid e & Vivid i ultrasound
system.
184 GE RNAS-9-RS 3D
Micro-Convex Probe
GE 4D Micro-convex Probe, RNA5-9-RS (year 2009), Frequency
range: 3.3 to 9.1 MHz. The probe connection, cable and head are intact
and in good physical condition; but not tested due to unavailability of
suitable unit. The probe is compatible with model GE Voluson I
ultrasound system.
185 GE 11L-D Linear
Probe
Unable to test the probe as there is no ultrasound to test with. The
probe looks to be intact and in satisfactory physical condition
187 Philips C5-1
Ultrasound Probe -
BOFN912
The probe has not been tested on a compatible ultrasound. Therefore
unable to carry out any further tests. Overall physical condition is
reasonable but near the scan head it has been taped previously. (Note
the probe has been tested on iU22 and HD11XE and was not detected)
188 Qty of Urology
Instruments to Inc.
Storz Resectoscopes
with Obturator,
Comeg Cystoscope,
Ellick Connectors
Lot is in satisfactory physical condition, lot is incomplete consisting of
various pieces.
195 Aloka UST-9123
Convex Ultrasound
Probe
The probe is in good physical and cosmetic condition and the scanhead
is physically intact. A compatible ultrasound is not available therefore
the probe cannot be tested.
196 Storz Hopkins 7200
30º Rigid Scope
The rigid scope is physically intact but does not give a good output
image. The rigid scope is inclined at 30 degrees. Overall physical
condition is fair.
200 Olympus OTV-S6
Camera Head
Unable to test camera head. The camera head is missing its lens
adaptor. The camera head is missing its rubber grommet at the bottom
of the camera head causing a large section of the inner insulation to be
exposed. The camera head is in below satisfactory physical condition.
235 Pentax LH 150FP
Light Source
The light source powers up, with its function knobs in working order.
The light source is able to produce a light output with its brightness
HG.115/116.v2
adjustment functioning. The air pump appears to be functioning. The
light source is in fair physical condition for unit’s age.
239 Bayer Clinitek 50
Analysers x2 (1
Spare only)
Unable to power up units as lot does not include correct power cables.
Units are in reasonable physical condition with one unit missing its
printer cover.
246 Blease Frontline 690
Anaesthetic Machine
with 8200s
Ventilator
The anaesthetic machine has not been tested but shows to have its flow
meters and gauges intact. The ventilator also has not been tested but
showed to be able to power up with its numerical displays functioning.
Unit’s bellows and absorber are in place. Unit includes an AGSS
receiving system and hose and a high flow high vacuum regulator and
hose. The unit is also complete with O2, air and N2O hoses. Overall
the unit is in fair physical condition.
248 Smith and Nephew
Stack Trolley with
Sony LCD Monitor,
Smith and Nephew
660 HD Image
Management
System, Dyonics
300XL Xenon Light
Source, Smith and
Nephew 460P 3-CCD
Digital Camera
Control Unit, HP
Printer and Camera
All units power up, with the function buttons functional. The camera
head and processor produce a satisfactory image on the Sony flat
screen monitor, with the camera head also showing to have satisfactory
focusing. The light source is able to produce a good light output with
its brightness adjustment functional. The image management system
showed to have its touch screen display functional with the unit
showing to be in working order. The printer powers up. Overall lot is
in good physical condition.
Image management system version: 1.5
252 Olympus TK-E1
Trolley with
Olympus CV240
Control Unit with
Pigtail and Olympus
CLV-U20 Light
Source
All units on the Olympus trolley power up ok. The light source xenon
lamp is working fine and light output model selector (OES, EVIS 100
& EVIS 200) are fully functional. Brightness control in auto and
manual mode are working fine. The video processor with the video
cable MH977 are in good working order when tested with a
videoscope. All the function keys and video controls are working.
Good quality image was displayed when connected to a colour
monitor. The system is in good overall physical condition.
259 Sensor Medics
3100A Oscillatory
Ventilator with
Fisher and Paykel
MR700 Humidifier
The ventilator is able to power up, with its function buttons/dials and
the numerical displays showing to function. The ventilator has not been
tested. The oxygen blender monitor is in place but unable to verify if it
is powering up. The humidifier is powering up with its function
buttons/dial and numerical display functioning, however again unit has
not been tested. The ventilator includes O2 and air hoses. Overall the
lot is in fair physical condition.
Operation hours: 11278.7
HG.115/116.v2
260 Sensor Medics
3100A Oscillatory
Ventilator with
Fisher and Paykel
MR700 Humidifier
The ventilator is able to power up, with its numerical displays in
working order. The silencing alarm function button is not functioning
unable to determine if the rest of the function buttons are responding.
The ventilator has not been tested. The Air-oxygen blender is in place.
The humidifier is powering up but does not appear to be functioning
with unit not responding and only displaying ‘EOo’ on the numerical
display. The ventilator includes O2 and air hoses. Overall the lot is in
fair physical condition.
Operation hours: 10897.3
261 Blease Frontline 440
Anaesthetic Machine
with 8200s
Ventilator and
Marquette Solar
8000 Patient
Monitor
The anaesthetic machine has not been tested but shows to have its flow
meters and gauges intact. The oxygen monitor shows to be powering
up with its function buttons and display in working order. The
ventilator is also powering up with its numerical displays functioning.
Again the ventilator has not been tested. The units bellows and
absorber are in place. Unit includes an AGSS receiving system and
hose and a high vacuum regulator and hose. The unit is also complete
with O2, air and N2O hoses. The patient monitor is able to power up
but however does not include any modules, unable to test unit. Overall
the lot is in reasonable physical condition.
265 Nihon Kohden
Cardiofax Q ECG
Recorder with ECG
Lead
The ECG machine powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
ECG machine has not been fully tested but appears to be functional.
The printer and battery is in good working order. Includes 12 lead ECG
wire. Overall physical condition is fair.
266 Nihon Kohden
Cardiofax GEM
ECG Recorder with
ECG Lead
The ECG machine powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
ECG machine has not been fully tested but appears to be functional.
The printer and battery is in good working order. Includes 12 lead ECG
wire. Overall physical condition is fair.
278 Smith and Nephew
Stack Trolley with
Sony LCD Monitor,
Smith and Nephew
660 HD Image
Management
System, Dyonics
Xenon XL Light
Source, Smith and
Nephew 460P 3-CCD
Digital Camera
All units power up, with function buttons functional. The camera head
and the processor are able to produce a good image on the Sony flat
screen monitor, with the camera head also showing to have good
focusing. The light source is able to produce a good light output with
its brightness adjustment functional. However the light sources power
switch is loose. The image management system showed to have its
touchscreen display in working order with the unit showing to be in
working order. The printer powers up. Overall the lot is in satisfactory
physical condition.
Image management system version: 1.5
HG.115/116.v2
Control Unit,
Camera Head and
HP Printer
279 Conmed Excalibur
Plus Diathermy on
Suction Trolley with
Blue, Yellow and
Black Foot Switches
The diathermy powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
diathermy is tested using an ESU analyser and the monopolar (cut and
coag) and bipolar output readings are within the limits. However the
pure cut mode on the higher output appear to be out of range otherwise
all the other modes are within the range. Includes monopolar (cut and
coag) and bipolar footswitches. The suction unit powers up but the
suction is not functional, therefore unable to carry out any further tests.
The overall physical condition is reasonable.
280 Conmed Excalibur
Plus Diathermy on
Suction Trolley with
Blue, Yellow and
Black Foot Switches
The diathermy powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
diathermy is tested using an ESU analyser and the monopolar (cut and
coag) and bipolar output readings are within the limits. Includes
monopolar (cut and coag) and bipolar footswitches. The suction is in
good working order but output has not been verified. The overall
physical condition is reasonable.
286 Kodak Dryview
8150 Laser Imager
The unit powers up however the display panel is not functional, so
unable to do anything with unit. Unit is in satisfactory physical
condition.
287 Fisher and Paykel
Cosy Cot
Resusitator
Unit powers up, with its indication lights and display in working order.
Units function buttons are functional however the manual mode
function button is missing its button cover. Unit showed to heat up well
in ‘Prewarm’ and ‘manual’ modes. unit does not include a skin
temperature probe. Units overhead light and timer both showed to be
functional. The neopuff section of the unit has not been tested. Units
electrical height adjustment is in working order. Unit includes O2 and
air hoses and a mattress. Overall unit is in satisfactory physical
condition.
Manufacture date: 2007-01
288 Fisher and Paykel
Cosy Cot
Resuscitator
Unit powers up but immediately alarms and gives an error message ‘Er
017’. The error message is unable to be passed. No further checks are
able to be carried out on the unit. The neopuff section of the unit has
not been tested. Unit include O2 and air hoses and a mattress. Unit is in
satisfactory physical condition
Manufacture date: 2004-01
294 Stryker Resus
Trolley -30109
The patient trolley is physically intact; its upholstery is in good
condition. All articulating movements are in good working order. Its
HG.115/116.v2
height adjustment is functional. The side rails are intact. The brakes are
in good working order. Overall physical condition is reasonable.
302 Fresenius 4008s
Hemofiltration
Machine
Unit is not powering up with unit only able to give a constant alarm.
Unit includes modules, double blood pumps, heparin pump, air
detector and BiBag. The Bibag module panel cover is completely
missing. Unit includes water and drainage hoses. Overall the unit is in
fair physical condition.
Manufacture date: 2010
Operation hours: 8972.03
303 Welch Allyn Patient
Monitors on Stands
with SPO2 Leads
The three vital signs monitor power up well, its display, functions and
function buttons are in good working order. The NiBP pump and SPO2
is functional but has not been fully tested. Lot includes two NiBP cuffs.
Battery is in good working order. Overall physical condition is fair.
308 Medelac Saphire 2M
and DP Medical
Light Source
Medelec Sapphire II. Powers up ok - display screen and controls are in
good working order and the printer is also functional. No further
checks carried out. Lot includes a DP Medical 202032 light source
which is in good working order. Overall the unit is in reasonable
physical condition.
311 Braun Dialog
Dialysis Machine x 2
Unable to power up units as both of the mains cables has been
removed. Both of the units include (1x) single blood pump, (1x)
heparin pump and (1x) air detector. Both of the units have their water
and drainage hoses cut off from the bottom. Overall lot is in reasonable
physical condition.
312 Braun Dialog
Dialysis Machine x 2
Unable to power up units as both of the mains cables have been
removed. Both of the units include (1x) single blood pump, (1x)
heparin pump and (1x) air detector. One of the units includes water and
drainage hoses the other unit has its water and drainage hoses missing.
Overall lot is in reasonable physical.
315 Conmed Excalibur
Plus Diathermy on
Suction Trolley with
Blue, Yellow and
Black Foot Switches
The diathermy powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
diathermy is tested using an ESU analyser and the monopolar (cut and
coag) and bipolar output readings are within the limits. Includes
monopolar (cut and coag) and bipolar footswitches. The suction and its
regulator are functional but the output has not been verified. The
overall physical condition is reasonable.
316 Conmed Excalibur
Plus Diathermy on
Suction Trolley with
Blue, Yellow and
Black Foot Switches
The diathermy powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
diathermy is tested using an ESU analyser and the monopolar (cut and
coag) and bipolar output readings are within the limits. Includes
monopolar (cut and coag) and bipolar footswitches. The suction is
HG.115/116.v2
functional but the regulator is not functional an unable to change the
suction output. The overall physical condition is reasonable.
317 Drager 5400
Transport Incubator
The Transport incubator 5400 power up ok on mains; battery power
has not been tested. The incubator set temperature is maintained
accordingly. The BabyMix 2 ventilator is in good physical condition
but not fully tested. The Oxygen monitor Oxydig has no battery, not
tested. The lot is in fair overall ohysical condition.
322 Marquette Mac8
ECG Recorder with
ECG Lead
Unit is able to power up after a long period of time. Then the unit
shows to have its function buttons in working order. Units display
functions however the display flickers which makes it hard to read in
places, with some lines of pixels missing. Unit makes a loud constant
humming noise. Unit has not been tested. Unit includes 12 wire ECG
however all of the leads are missing their end connection pins. Unit is
in reasonable physical condition.
325 Conmed Excalibur
Plus Diathermy on
Trolley with Yellow,
Blue and Black Foot
Switches
The diathermy powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
diathermy is tested using an ESU analyser and the monopolar (cut and
coag) and bipolar output readings are within the limits. Includes
monopolar (cut and coag) and bipolar footswitches. The overall
physical condition is reasonable.
326 Drager D-Vapour
Desflurane
Vapouriser
The vaporiser is in good physical condition, with its volume control
dial and lock intact. There are some minor marks to paintwork.
328 Drager Oxylog
Portable Ventilator
Unit power up when connected to medical gas. Airways pressure
indicator working fine. Test run with medical air compressor and test
lung, indicates that all parameter and alarm settings were in good
working order. The minute Volume (MV) was verified with the set
valve of Tidal volume and Frequency (BPM) and found to be within
limit. The system is in fair physical condition. Oxygen hose with
schrader connector is included.
361 The ultrasound is unable to power up. Unable to test the unit. Units
(1x) probe appears to be physically intact. Unit is in reasonable
physical condition, with its covers colour being worn away in large
patches.
369 Drager 8000SC
Incubator
Unit is able to power up, with units function buttons and displays in
working order. In ‘Air’ mode unit showed to be functional, with unit
able to reach and maintain the set temperature. Temperature output has
not been verified. Unit does not include a skin temperature probe so
unable to test ‘skin’ mode. ‘Humidifier’ mode has not been checked.
Units electrical height adjustment and mechanical cot tilt are in
working order. Unit includes a mattress. Unit is in satisfactory physical
HG.115/116.v2
condition.
374 Edwards Lifescience
Aquarius
Hemofiltration
Machine
Unit is able to power up, units function buttons, rotary knob and
display show to be functioning. Unit has not been tested. The blood
pump and filtrate cover panel is missing, also the prefillter sensor is
bent and unable to close. Overall unit is in fair physical condition.
Manufacture date: 2009
Software version: 6
379 Sonosite 180 Plus
with Probe
The ultrasound powers on, but is not functional as the display gives a
badly flickering display with no actual picture showing. Unable to
check unit. Units (1x) probe appears to be physically intact. The
ultrasound is in satisfactory physical condition.
Manufacture date: 2004-08
382 Zeiss IOL Master
with Printer and
Table
The IOL Master powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
IOL Master is tested using a test eye and test results taken where within
the limits. The functions included are Axial Length Measurement,
Corneal Curvature Measurement and Anterior Chamber Depth
Measurement, which are in good working order. Printer is functional.
The height adjustment is in good working order. Overall physical
condition is reasonable. IOL Master Version: 3.02.0304
387 Haag-Streit Slit
Lamp with Table
The slit lamp is able to power up, with it able to produce a light output,
the reflector mirror for the light output is in place but has a large chip
at one corner. The fixation light is in place but however it is not
functional and has its outer cover missing. The joystick movements
show to be functioning but however is slightly stiff, the movement lock
is functional. The chin rest is in place and its chin adjustment is
functional. The slit lamps tables mechanical height adjustment is in
working order. The slit lamps optical system is able to produce a clear
image, with it also showing to have good focusing. Overall the slit
lamp is in satisfactory physical condition, with the side compartment
on the table missing its bottom panel.
(1x) set of eyepieces (10x)
395 Quantel Medical
Virdis Twin Laser
with Key and Foot
Switch, BMBQ Slit
Lamp Adaptor
Quantel Medical Viridis Twin, software version: M:10.02 B:10.00.
Unit power up ok, the LCD screen and function keys are working fine.
The aiming beam is working and the 532 nm laser function screen
comes when the key is turned on. The parameters were adjustable for
power, pulse etc. Test firing of the laser indicate green laser at the out
but the power was not verified. When the key is turned to select 810
HG.115/116.v2
nm laser an error message is displayed – E100A:0 ‘fibre disconnected.
Only one fibre cable with laser aperture is included with slit lamp
adapter model: BMBQ. One YAG safety goggle is included. System is
in fair physical condition.
396 Hill-Rom Air Shields
Resusitaire - 001297
Unit powers up, with its function buttons, indication lights and numeric
displays in working order. Unit showed to function with unit able to
heat up in ‘pre-warm’ and ‘manual’ modes. In ‘manual’ mode the heat
adjustment is functional. Unable to test ‘baby’ mode as unit does not
include a skin temperature probe. Units exam light and timer are both
functional. Units electrical height adjustment is functional as is the
mechanical cot tilt. Units Suction and patient supply were checked
using a medical compressor, and the suction and patient supply showed
to be functioning with their gauges in working order. Unit includes O2,
air and N2O hoses, unit also includes a mattress and suction jar.
Overall unit is in satisfactory physical condition.
399 Stryker Endoscopy
Trolley with
Trinitron Monitor
Q5000 Light Source,
X6000 Light Source,
2 x Sony Colour
Video Printer, 2 x
988i 3 Chip Camera
Controller and 3 x
988i Camera Heads
All units are able to power up, with the function buttons functional.
Both of the processors and two of the camera heads are able to produce
a good image on the Sony monitor, with both camera heads showing to
have good focusing. However one of the two camera heads does show
some slight interference from the connection end if moved. The third
camera head is able to produce a reasonable image however it is
constantly affected with bad interference when the camera head is
moved, the camera heads focusing is functional. Both of the light
sources are able to produce a good light output with their brightness
adjustments functional. Both of the color video printers are powering
up. Overall the lot is in satisfactory physical condition.
Bulb hours for Q5000: 205.2
Bulb hours for x6000: 183
400 Alcon Surgical
Ultrasonic Imaging
System with
Monitor and Probe
10MHz and Foot
Switch on Trolley
Alcon Ultra Scan imaging System, software version: 3.00. Unit power
up ok and the CRT display and keyboard are working fine. The
ultrasound probe 10MHz is appears to be working with a scan image
display on the screen. Unit is not fully tested. In fair physical
condition.
402 Eschmann TD411-S
Diathermy with Foot
Switches on ST80
Suction Trolley
The diathermy powers up well and passed its initial self-test. The
display, rotary knobs and functions are in good working order. The
diathermy is tested using an ESU analyser and cut outputs are out of
range whereas the coagulation modes are within the range. The bipolar
mode has not been tested as there are no footswitches. Monopolar cut
and coagulation footswitches are included. Overall physical condition
HG.115/116.v2
is fair.
405 GE Mac 5000
Pagewriter ECG
With Leads
Unit is able to power up, with its display and function buttons in
working order. Unit has not been tested. Unit includes 12 wire ECG.
Unit does not include a battery, unable to test printer as there is no
paper. Unit is in satisfactory physical condition.
409 Zeiss OPMI-1FC
Colposcope with
Panasonic Camera
Unit powers up, unit is able to give a light output however the light
guide cable has a lot of its fibers damaged causing unit to have a poor
light transmission. Units mechanical movements are in working order.
Units optical system is able to produce a good image, with unit
showing to have its focusing an magnification in working order. The
Panasonic WV-CL110AE has no power supply so unable to test.
Overall the lot is in satisfactory physical condition.
(1x) eyepieces (10x/22B)
(1x) objective lens
414 Mortara ELI 250
with ECG
Pagewriter Leads
The ECG machine powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
ECG machine is tested using an ECG simulator and gives out good
strong traces of ECG signals. The printer is in good working order.
Includes 12 lead ECG. Overall physical condition is fair.
415 Zeiss IOL Master
with Printer and
Table
The IOL Master powers up well and passed its initial self-test. At time
of testing the monitor switched off and the display screen stayed blank,
but afterwards the screen went into its normal menu screen. The
display, functions and function buttons are in good working order. The
IOL Master is tested using a test eye and test results taken where within
the limits. The functions included are Axial Length Measurement,
Corneal Curvature Measurement and Anterior Chamber Depth
Measurement, which are in good working order. Printer is functional
but the ink cartridge needs replacing. The height adjustment is in good
working order. Overall physical condition is reasonable. IOL Master
Version: 3.01.0294
418 GE Mac 5000
Pagewriter ECG
With Leads
Unit powers up, with its function buttons and display in working order.
Unit has not been fully tested. Unit includes 12 wire ECG. Units
Printer is functional, however the battery is not functional. Unit is in
satisfactory physical condition.
440 Takagi S10-15 Slit
Lamp Microscope
The slit lamp is able to power up, with the unit showing to be able to
give out a light output. The fixation light is in place and is functional,
but is missing its fixation light cover. The joystick movements are
freely movable with the movement lock functional. The chin rest is in
place with its chin adjustment functional. The mechanical table height
HG.115/116.v2
adjustment is in working order. The slit lamps optical system is able to
produce a clear image, with the magnification in working order. The
CV-S3200 camera head powers up and was testing using a monitor
(not included). The camera head showed to be able to produce a clear
image. Lot is in satisfactory physical condition.
(1x) set of eyepieces (16x)
443 Penlon AV900
Ventilator
The ventilator powers up well and passed its initial self-test. The
display, LCD touchscreen and function buttons are in good working
order. The unit has not been fully tested but appears to be functional.
No hoses are included. Overall physical condition is reasonable. S/W is
V3.33
450 3M Espe Eliper
Freelight 2 Curing
Light, Shrewsbury
Medical Ultrasound
Therapy System
The shrewsbury unit powers up, with units displays and function knobs
in working order. Unit showed to be functioning with unit giving an
output and with its timer functional. unit includes a (1x) transducer
(1.1, 3.2 MHZ). The unit is in fair physical condition
The Freelight 2 has its power indication lights on, but unable to
determine if unit is functional. Not tested. Lot includes its docking
station. Unit is in fair physical condition.
453 Olympus KV-5
Suction Pumps x2
Both of the suction pumps are able to power up. Both units show to be
giving suction, the suction output has not been verified. Both of their
vacuum regulators are in working order. However both of the units are
missing their suction jars. Overall lot is in good physical condition.
454 Olympus KV-5
Suction Pumps x2
Both of the suction pumps are able to power up. Both units show to be
giving suction, however one of the units is unable to reach the units
maximum output of 85KPa showing to be around 5+ KPa short.
Neither units output has been verified. Both of the units vacuum
regulators are in working order. Both units are missing their suction
jars. Overall lot is in good physical condition.
462 Nellcor Butitan
Bennett 840
Ventilator System
Con# D4133,
Ass#G7940
Unit is able to power up, with its touchscreen display functional and its
function buttons and rotary knob are looking to respond. Unit is
missing its expiratory filter, so unable to test unit. The ventilation
modes are selectable but have not been tested. Unit includes O2 and air
hoses. Overall the ventilator is in good physical condition.
Vent modes: A/C, SIMV, SPONT and BILEVEL
Software: 4-070000-85-AB
Operation time: 53614Hours
Manufacture date: 2000-03
HG.115/116.v2
463 Nellcor Butitan
Bennett 840
Ventilator System
Con# D4133,
Ass#G9127
Unit is able to power up, with its touchscreen display functional and its
function buttons and rotary knob are looking to respond. Unit is
missing its expiratory filter, so unable to test unit. The ventilation
modes are selectable but have not been tested. Unit includes O2 and air
hoses. Overall the ventilator is in good physical condition.
Vent modes: A/C, SIMV, SPONT and BILEVEL
Software: 4-070000-85-AB
Operation time: 48823Hours
481 Valleylab Force FX
Electrosurgical Unit
Unit power up and self-test completed successfully. All displays,
output power and modes selector keys are in good working order. The
patient return plate monitoring circuit is functioning. The unit’s RF
output was verified using ESU analyser and the selected outputs were
all within limits for both bipolar and monopolar. Unit is in good
physical condition and mounted on a trolley. Footswitches and
cables/accessories are not included
482 Valleylab Force FX
Electrosurgical Unit
Unit power up and self-test completed successfully. All displays,
output power and modes selector keys are in good working order. The
patient return plate monitoring circuit is functioning. The unit’s RF
output was verified using ESU analyser and the selected outputs were
all within limits for both bipolar and monopolar. Unit is in good
physical condition and mounted on a trolley. Footswitches and
cables/accessories are not included
485 Stryker X6000 Light
Source
The light source is able to power up. Unit showed to be able to produce
a good light output, with its brightness adjustment in working order.
Bulb hours: 200. The light source is in good physical condition.
487 Wolf Arthro Power
Unit with Cable and
9 Rotating Cutting
Tools
Unit powers up, unable to verify if unit is functional as unit does not
include a footswitch. Unit does however include a flexible cable and
(9x) various hand pieces. Unit is in reasonable physical condition.
490 Penlon Enflurane
Vapouriser x 2
Both of the vaporisers are in satisfactory physical condition. With both
of the vaporisers volume control dials being intact.
491 Penlon Enflurane
Vapouriser x 2
Both of the vaporisers are in satisfactory physical condition. With both
of the vaporisers volume control dials being intact.
493 3M Unitek Ortholux
XT Curing Light,
3M XL3000 Curing
Light
Both units are able to power up, with both of the units able to give a
light output. Both units timers showed to be functional. The lot is in
satisfactory physical condition.
HG.115/116.v2
497 Penlon Sigma
Sevoflurane
Vapouriser (Back
Lock Missing)
The vaporiser is in good physical condition, with its volume control
dial looking to be intact.
498 Penlon Sigma
Sevoflurane
Vapouriser
The vaporiser is in good physical condition, with its volume control
dial looking to be intact.
507 Respironics BIPAP x
2 and Resmed VPAP
II x 2
Bipap Synchrony
Both units power up, however one unit displays a blank screen and
gives a constant audible alarm, unit is not functional. The other unit has
its function buttons and displays in working order. With unit showing
to give an output but units have not been fully tested. Units are in fair
physical condition.
VPAP
Both units are able to power up, one units function buttons do not look
to be responding correctly with some of its display hard to read. The
second units function buttons and display are functional. Both units
showed to be giving an output but units have not been fully test. Units
are in fair physical condition.
521 Alcon Surgical
Ultrascan Imaging
System with Foot
Switch and Sony
Video Graphic
Printer
Alcon Ultra Scan imaging System, software version: 3.00. Unit power
up ok and the CRT display and keyboard are working fine. The
ultrasound probes included are damaged. Unit was tested with the
probe of lot 400 but still could not obtain a scan image. Unable to carry
out further test on this unit. The attached printer is working fine.
System is in fair physical condition.
522 Stack Trolley Inc
Trinitron Monitor
Storz Image Hub,
Xenon Nova 300
Light Source Aida
Control
All units are able to power up. Lot does not include a camera head so
the processor is unable to be tested but its power indication light
showed to come on. The light source showed to be able to produce a
good light output with its brightness adjustment functional. The AIDA
control had its power indication light on but the unit has not been
tested. The monitor showed to be functional. Overall the lot is in fair
physical condition.
523 SLT CL100 Yag
Laser with Box of
Accessories and
Consumables
The unit power up ok and the displays and function keys are working
fine. No fully tested. System is in fair physical condition. Year of
manufacture January 1992, Serial No. 91015501. The following are
included:
IV Pole, Footswitch, User & Maintenance Manual & ON/OFF Key
HG.115/116.v2
Accessories:
Laser bare fibre x5
Laser goggles x3
Black Laser glasses x3
Grey laser glasses x4
Chisel Probe x2
Round Probe x3
Hook Probe x1
Frosted Scalpel x1
Metal cones x5
CO2 Regulator & hose
SLT laparoscopy fibre x2
524 Conmed Excalibur
Plus Diathermy on
Suction Trolley with
Blue, Yellow and
Black Foot Switches
Unit is able to power up, with its displays in working order and its
function buttons looking to respond. The unit has not been fully tested.
The patient plate alarm is functional. Lot includes mono-polar and bi-
polar footswitches, the footswitches look to be responding. The suction
pump is able to power up, with unit able to produce suction, however
the regulator is not functional and is stuck on maximum suction. Unit
incudes (1x) suction jar. Lot is in reasonable physical condition.
525 Conmed Excalibur
Plus Diathermy on
Suction Trolley with
Blue, Yellow and
Black Foot Switches
Unit is able to power up, with its displays in working order and its
function buttons looking to respond. The unit has not been fully tested.
The patient plate alarm is functional. Lot includes mono-polar and bi-
polar footswitches, the foot switches look to be responding. The
suction pump is able to power up, with the unit able to produce suction,
its regulator is in working order. Unit includes (2x) suction jars with
one of the jars missing its lid. Lot is in reasonable physical condition.
526 BiPAP Vision
Respironics
Ventilatory System
on Stand - 046721
Unit powers up, before then powering back down again immediately
and then giving a continuous alarm for ‘Vent Inop’. Unit includes O2
hose. Unit is in satisfactory physical condition.
528 Keymed Trolley
With Olympus CV-
200 Camera Control
Unit, Olympus CLV-
U20 Light Source,
Olympus MH-977
'Pig Tail' Connector
Con# D4029,
7366097,7254862
Both the processor and light source are able to power up. The processor
and pigtail were tested using a monitor and a videoscope (both not
included). The processor and the pigtail both showed to be functional
and both helped in producing a good image. The processors function
buttons showed to be in working order. The light source showed to be
functional also with its function buttons in working order. All of its
light output modes (OES, Evis 100 and Evis 200) showed to be
functional. The light sources brightness adjustment worked in both
manual and automatic modes. The light sources air pump showed to be
temperamental in functioning. Overall the lot is in satisfactory physical
condition.
HG.115/116.v2
531 Zeiss IOL Master
Eye Scanner
The IOL Master powers up well and passed its initial self-test. The
display, control panel and functions are in good working order. The
IOL Master camera is faulty and unable to take an image and therefore
unable to carry out any further tests. The test modes included are Axial
Length Measurement, Corneal Curvature Measurement and Anterior
Chamber Depth Measurement. Printer is included, which is functional.
The table height adjustment is in good working order. Overall physical
condition is reasonable. IOL Master Version 3.01.0294.
543 Cinescan Quantel
Medical with HP /
Mitsubishi Printers
and Foot Switch
The scanner power up ok, display and keyboard are functional. Both A
& B scan appears to be working fine. Good image displayed from the
US probe. All seem to be working and system is in good physical
condition. The B&W printer is also functional.
548 Smith and Nephew
Stack Trolley with
Sony LCD Monitor,
Smith and Nephew
660 HD Image
Management
System, Smith and
Nephew 460P 3-CCD
Digital Camera
Control Unit,
Camera and HP
Printer
All units power up, with function buttons in working order. The
processor and camera head are able to produce a good image on the
Sony flat screen monitor, the camera head also shows to have good
focusing. The image management system has its touch screen display
functional with unit showing to be in working order. The printer
powers up. Overall the lot is in good physical condition.
Image management system version: 1.5
551 Olympus Stack
Trolley with Sony
PVM20M2MDE
Monitor, Olympus
UHI-3 Insufflator,
Olympus CLV-S20
Light Source, Sony
SVD-3500MDP
VCR, Olympus MH-
317 Foot Switches x2
- Con# 4018, Ass# -
7939,7941,7942,7943,
5905
All units are powering up. The light source is able to produce a good
light output with its brightness adjustment functional, the spare lamp
also shows to be functional. The insufflator has not been tested but its
displays show to be functioning and its function buttons look to
respond. The monitor showed to be functional. The VCR is powering
up. Overall lot is in satisfactory physical condition.
552 Respironics BiPAP
Focus
Unit powers up, but unit is unable to complete its system start up, with
unit giving a constant audible alarm and displays system error ‘343
battery faulty or missing’ which is unable to be passed. Unable to carry
out any further tests on unit. The unit is in satisfactory physical
HG.115/116.v2
condition.
Boot block version: 1.03.000
553 E-Vent Inspiration
LS Mobile
Ventilator
The ventilator is able to power up, with it showing to be able to pass
through its initial self-test ok. Units function buttons and display are in
working order. The ventilator showed to be functional when it was
tested using a medical compressor. The BPM, Vte and Ve are able to
match up accordingly. Units output has not been verified with a flow
analyser. Unit was only able to be tested with a pediatric flow sensor in
units pediatric setting. Unit also was only tested in V-SIMV mode,
with the rest of the vent modes not being verified. The battery is not
functional. Unit includes O2 and air hoses, no flow senor included in
lot. The ventilator is in satisfactory physical condition.
Vent modes: V-CMV, V-SIMV, PRVC-CMV, PRVC-SIMV, P-CMV,
P-SIMV, VS, SPONT and SPAP
Software version: 5.0.1
554 E-Vent Inspiration
LS Mobile
Ventilator
The ventilator is able to power up, with it able to complete its initial
self-test successfully. Units display is functional however the rotary
selection knob is temperamental, the rest of the function buttons are in
working order. The ventilator showed not to be functioning/ventilating
correctly when tested using a medical compressor. With units BPM,
Vte and Ve not able to match up and with them showing to be out of
limits from the values set. Output has not been tested with a flow
analyser. Unit was only able to be tested with a pediatric flow sensor in
units pediatric setting. Unit also was only tested in V-SIMV mode, rest
of vent modes not verified. The battery is functional. Unit includes O2
and air hoses and also a pediatric flow sensor. The ventilator is in
satisfactory physical condition.
Vent modes: V-CMV, V-SIMV, PRVC-CMV, PRVC-SIMV, P-CMV,
P-SIMV, VS, SPONT and SPAP
Software version: 5.0.0
555 E-Vent Inspiration
LS Mobile
Ventilator
The ventilator is able to power up, with it showing to be able to pass
through its initial self-test ok. Units function buttons are in working
order, with the display functioning however it can flicker and the
display picture can flip and appear upside down on the display. The
ventilator showed to be functional when it was tested using a medical
compressor. The BPM, Vte and Ve are able to match up accordingly.
Units output has not been verified with a flow analyser. Unit was only
able to be tested with a paediatric flow sensor in units paediatric
setting. Unit also was only tested in V-SIMV mode, with the rest of the
HG.115/116.v2
vent modes not being verified. The battery is functional. Unit includes
O2 and air hoses and also a pediatric flow sensor. The ventilator is in
satisfactory physical condition.
Vent modes: V-CMV, V-SIMV, PRVC-CMV, PRVC-SIMV, P-CMV,
P-SIMV, VS, SPONT and SPAP
Software version: 5.0.1
556 Alcon Surgical
Series 20000 Legacy
Phaco Machine and
Foot Switch
Unit is able to power up, with its function buttons and touchscreen
display in working order. Unit does not include any hand pieces and
unit has not been fully tested. Modes look to be selectable but have not
been tested. Units audio voice prompts are functional. The IV pole
height adjustment is in working order. The foot control looks to be
responding and the remote is functional. Overall the unit is in good
physical condition.
V3.00
Manufacture date: 04/00
561 Nellcor Butitan
Bennett 840
Ventilator System
Con# D4133,
Ass#G9128
Unit is able to power up, with its touchscreen display functional and its
function buttons and rotary knob looking to respond. Unit is missing its
expiratory filter, so unable to test unit. The ventilation modes are
selectable but have not been tested. Unit includes O2 and air hoses.
Overall the ventilator is in good physical condition.
Vent modes: A/C, SIMV, SPONT and BILEVEL
Software: 4-070000-85-AB
Operation time: 44328Hours
564 Toshiba ECC CEE
Ultrasound with
Probes x3 -PLF
805ST x2, 375MT
The ultrasound powers up well and passed its initialization. The
display, functions and function buttons are in good working order. The
ultrasound is tested using a phantom and appears to be functional. The
set measurements taken are within the limits. The greyscale and
resolution is average. The ultrasound modes included are: B, M Colour
and PW. The three probes included are intact but all three probes
shows interference and lines through the image. One of the PLF-805ST
probes has discoloration on the probe membrane and has a large crystal
damage, which affects the image. The PVF-375MT probe has also
crystal damage but does not affect the image. Overall physical
condition is average.
565 Olympus TI-1900
Imaging Trolley with
Sony PVM-
20M2MDE Monitor,
All units are able to power up. Unable to test the two processors as lot
does not include any camera heads but their indication lights show to
be functional, unable to verify if function buttons are functional. The
light source is able to produce a good light output with its brightness
HG.115/116.v2
Olympus UHI-2
Insufflator, Olympus
OTV-S6 Camera
Control System (No
Camera), Olympus
CLV-S30 Light
Source, Olympus
OTV-SX2 camera
Control System (No
Camera), Stryker
Endoscopy SE5
Arthroscopy
adjustment in working order, its spare lamp is functional. The
insufflator has not been tested, but its displays look to be functional.
The SE5 powers up but again has not been tested, but has its displays
functional. The Sony monitor is functional. Overall lot is in satisfactory
physical condition.
566 Stockert Heart/Lung
Bypass Machine
Stockert S 3 Roller Pump. The system powers up and the control
console is in good working order. All 5 pumps are functional and the
display panels and speed controls are also in good working order. The
backup battery is also working and overall the unit is reasonable
physical condition. Year of manufacture: 2003
579 Steris Surgimax
Electric Operating
Table (No Pads or
Hand Control)
The operating table is powering up. With its function buttons on the
base of the unit functional. All of the electrical articulating movements
showed to be in working order. As did the mechanical movements.
Units brake lock is functional with the wheels freely movable. The
operating table is missing its head rest panel, unit is also missing all of
its cushions. Unit does not include a remote. The battery is functional.
The operating table is in satisfactory physical condition.
581 Erbe ICC 200 and
Erbe APC 300 On
Trolley with Foot
Switches
The diathermy powers up well and passed its initialization. The
display, functions and function buttons are in good working order. The
diathermy is tested using an ESU analyser and the cut output are within
the range however the coag modes are out of range. The bipolar
footswitches are not included and therefore unable to test. Includes
monopolar cut and coag footswitches. The Argon Plasma Coagulation
powers up but as there is no Argon gas at ti142me of tested therefore
unable to fully test. Overall physical condition is reasonable. Software
Version is V2.20 for the Argon Plasma coagulation.
587 Drager 8000 Resus
Unit
The resuscitaire powers up well, its display, functions and function
buttons are in good working order. The gauges and flowmeters are
intact and functional. The resuscitaire heater is in good working order.
The heater manual mode is tested, which appears to be functional
however as there are no skin probes included therefore unable to verify
the skin mode. The suction is in good working order along with its
regulator but the output has not been verified. The Resus unit has not
been fully tested but has been tested for leaks and no leak is present.
HG.115/116.v2
The motorised height adjustment is in good working order. Includes
gel mattress, medical gas hoses (air and oxygen). The overall physical
condition is reasonable. Y.O.M is 1998
611 Olympus PSD-60
Diathermy with Foot
Switch
Unit is able to power up and pass through its initial self-check
successfully. Units displays look to be functional and its function
buttons look to be responding. The unit has not been tested, with its
output not verified. Units patient plate alarm appeared to be functional.
The (1x) foot switch for monopolar cut and coag looked to be
responding. Unit is in satisfactory physical condition.
Version: V1.0.1
649 Philips Pagewriter
Trim III ECG
Recorder and Leads
- 18822 Serial
Number
US20817891
The ECG machine is able to power up, with units function buttons,
rotary knob and display in working order. However the display has
been badly scratched, especially in one corner. Unit includes (1x) 12
wire ECG and units ECG was tested using an ECG analyser. Unit
showed to have good traces of ECG. Units printer is functional,
however units battery is not. Unit is in reasonable physical condition,
with bad scratches to the display.
652 Getinge 1345
Washer on Cabinet
Unit is able to power up, Units function buttons and display are in
working order. Unable to test unit. Units main door is able to open with
its lock in working order, inside unit includes its sterilizing rack. Units
cabinet door lock is locked with the door open, resulting it the cabinet
door not being able to close. Unit includes (1x) Martel MCP8810X-
184 printer which is not powering up and (1x) star TSP650 printer
which is missing its correct power adaptor. Lot is in good physical
condition, with however a dent to its back panel.
Version: 5.06(320)
669 Datascope Accutor
Plus SPO2 Monitor
with Cuff
Unit is able to power up, with its function buttons and displays in
working order. Unit has not been fully tested. Units NiBP pump shows
to be functional. Unable to test units SPO2 parameter as there is no
SPO2 probe included in lot. Lot includes (1x) NiBP cuff and hose.
Battery not functional. Unit is in fair physical condition.
671 Marquette 1/K
Operating Table
Remotes- 03115
Maquet cordless control 1133.9/A0 – in good physical condition but
not tested.
687 Fisher and Paykel
Cosy Cot Resusitator
Unit powers up, with units function buttons, indication lights and
numerical display in working order. Unit shows to be functional, with
unit able to heat up in ‘prewarm’ and ‘manual’ modes. With the
manual modes heat adjustment functional. Unable to test ‘baby’ mode
as lot does not include a skin temperature probe. The overhead light is
HG.115/116.v2
functional as is the timer. The electrical height adjustment is in
working order. Unit includes a mattress and O2 and air hoses. The
‘Neopuff’ part of unit has not been tested. Lot is in satisfactory
physical condition.
Manufacture date: 2009-03
688 Fisher and Paykel
Cosy Cot Resusitator
Unit powers up, with units function buttons, indication lights and
numerical display in working order. Unit shows to be functional with
unit able to heat up in ‘prewarm’ and ‘manual’ modes. With the
manual modes heat adjustment functional. There is no skin temperature
probe included with lot. Units overhead light is functional as is the
timer. The electrical height adjustment is in working order. Unit
includes a mattress and O2 and air hoses. The ‘Neopuff’ part of unit
has not been tested. Lot is in satisfactory physical condition.
Manufacture date: 2007-10
689 Fisher and Paykel
Cosy Cot Resusitator
Unit powers up, with units function buttons, indication lights and
numerical display in working order. Unit shows to be functional with
unit able to heat up in ‘prewarm’ and ‘manual’ modes. With the
manual modes heat adjustment functional. There is no skin temperature
probe included with lot. Units overhead light is functional as is the
timer. The electrical height adjustment is in working order. Unit
includes a mattress and O2 and air hoses. The ‘Neopuff’ part of unit
has not been tested. Lot is in satisfactory physical condition.
Manufacture date: 2007-10
711 Marquette Betastar
Electric Operating
Table with Remote
Control (End Section
Missing Incomplete
Pads)
The operating table is able to power up. All electrical articulating
movements except for the leg movement are functional from the
function buttons on the main unit. The leg extension movement has not
been tested as table is missing its leg extension part of the table. Lot
includes (1x) remote which is not fully functional as only trendlenburg
movement works from it. The tables wheels are freely movable and the
brakes are functional. Unit does not include a complete set of cushions
with lot only including a cushion for the main body section, which
looks to be in satisfactory physical condition. Units battery is not
functional. The operating table is in satisfactory physical condition.
HG.115/116.v2
713 Optiflex Leg
Extender
Unit is able to power up. Unit includes (1x) remote which shows to be
in working order. Unit shows to be functional, with its angle
adjustment functional as is its manual speed selection. The external
power supply makes a continuous humming noise. Overall the unit is
in reasonable physical condition.
725 PneuPAC VentiPAC
200D & paraPAC
200D, portable
ventilators (x2)
Both units power up when connected to medical gas. Airways pressure
indicator working fine and the battery powered alarm are working
fine.. Test run with medical air compressor and test lung, indicates that
all parameter and alarm settings were in good working order. The
minute Volume (MV) was verified with the set valve of Tidal volume
and Frequency (BPM) and found to be within limit. The system is in
fair physical condition. Oxygen hose with schrader connector is
included.
727 Gynecare Morcelltor Unit is able to power up, its indication lights are in working order. Unit
has not been tested as there is no hand piece for the lot. Unit is in fair
physical condition.
730 Dyonics 300XL
Xenon Light Source
- 10084
The light source is able to power up, with its function buttons and
numerical display in working order. Unit was able to produce a good
light output, with its brightness adjustment functional. Unit is in fair
physical condition.
Bulb hours: 101
778 Zeiss Colposcope The colposcope is able to power up. Unit is able to produce a good
light output with its brightness adjustment functional. Units mechanical
movements are functional, the lock on the main arm does not
completely tighten. Units optical system is able produce a clear image
with units magnification in working order. Overall the colposcope is in
satisfactory physical condition, with the cap cover for the brightness
adjustment knob missing.
(1x) set of eye pieces (12.5x)
(1x) objective lens (f250)
779 Alcon Surgical
Series 20000 Legacy
Phaco System
Unit powers up, with units function buttons and touch screen display in
working order. The unit does not include any hand pieces and unit has
not been fully tested. Modes are selectable but have not been verified.
Units IV pole height adjustment is functional as are the audio voice
prompts. The foot control shows to respond however the remote
control is not functional. Lot is in good physical condition.
Version: V3.12
Manufacture date: 01/97
HG.115/116.v2
781 Laserscope
Greenlight HPS 2070
with Foot Switch (No
Key)
The laser unit has not got a key and therefore unable to power up.
Includes footswitch and the mains cable is an industrial plug. Overall
physical condition is fair. Single phase, Y.O.M 2006
795 Carl Zeiss
Microscope Parts x
3
Zeiss operating microscope binocular assembly x2. Both with good
optics. One with eyepieces 12,5x/18B, f170. The second one with eye
pieces 10x/22B, f170 with extension bridge.
808 Alcon Surgical
Series 20000 Legacy
Phaco System
Unit powers up, with units function buttons and touch screen display in
working order. The unit does not include any hand pieces and unit has
not been fully tested. Modes are selectable but have not been verified.
Units IV pole height adjustment is functional as are the audio voice
prompts. The remote control is functional. Unit does not include a foot
control. Lot is in good physical condition.
Version: V3.12
Manufacture date: 11/01
809 Kodak Dryview
8900 Laser Imager
Unit power up and the drive motors inside con be heard initializing but
the LCD screen remain blank.
809 Kodak Dryview
8900 Laser Imager
Unit powers up, however units power indication light above the display
is not coming on, also units touch screen display is not coming on or
functional. Unit shows to be not functional and unable to carry out any
further checks. Unit is in satisfactory physical condition.
810 Kodak Dryview
8900 Laser Imager
Unit not powering up and cover panel at the back is missing. Fair
physical condition.
810 Kodak Dryview
8900 Laser Imager
Unit sounds to power up for a brief second, before cutting out.
However the power indication light above the display does not come
on and neither does the touchscreen display. Unit shows to be not
functional and unable to carry out any further checks. Unit is in
satisfactory physical condition.
861 Philips Heartstart
XL (No Batteries)
The defibrillator powers up, with its display functional and its function
buttons looking to respond. The unit has not been tested. Unit includes
pacer function, which also has not been tested. Unit does not include
any accessories. The printer in functional, however the printer doors
hinge is broken making it difficult to close. Units battery is missing.
Unit is in satisfactory physical condition.
Manufacture date: 2004-03
HG.115/116.v2
866 Bair Hugger Patient
Warmers x 2
Both patient warmers power up ok and the heaters are in good working
order. The units are in reasonable physical condition. Hoses are not
included.
875 Alcon Surgical
Series 20000 Legacy
Phaco System
Unit powers up, with units function buttons and touch screen display in
working order. The unit does not include any hand pieces and unit has
not been fully tested. Modes are selectable but have not been verified.
Units IV pole height adjustment is functional as are the audio voice
prompts. The foot control shows to be responding and the remote
control is functional. Lot is in good physical condition.
Version: V3.12
Manufacture date: 11/01
879 Olympus PSD 20
Diathermy with Foot
Switch
The diathermy is able to power up, with its displays functional and its
function buttons looking to respond. The unit has not been tested with
its output not being verified. Units patient plate alarm is functional.
Unit includes (1x) monopolar cut/coag footswitch which appears to be
responding. Unit is in good physical condition.
882 GE Dash 3000
Patient Monitor
The patient monitor powers up well and passed its initial self-test. The
display, functions and function buttons are in good working order. The
patient monitor has not been fully tested but appears to be functional.
The NiBP pump is functional but has not been fully tested. Includes
incomplete ECG lead, SpO2 trunk and NiBP hose. Battery has not
been fully tested. Overall physical condition is fair.
891 LTE Falcon 30
Autoclave
The autoclave power but with error displayed. Printer error and
instruction to switch unit off. Unable to test any of the cycle and the
door id disabled. Unit is in fair physical condition. But not fully tested.
896 Kodak Dryview 8900
Laser Imager
The laser imager power up ok and touch screen is fully functional. MS
Windows 2000 Professional start-up completed. The system self-test
which include warming up and calibration are completed successfully
and ‘Ready’ is displayed as the unit status. Test patterns were
successfully printed from all three film drawers. The system appears to
be working fine but not tested fully. Unit is in good physical condition.
897 Alcon Series 20000
Legacy Phaco
machine, 8 x Phaco
Hand Pieces with
Foot Switch, Remote
Control and
Manuals
Unit powers up, with its function buttons and touchscreen display in
working order. Unit has not been fully tested, however unit does
include (ref: TURBOSONIC-375) hand pieces (x8). Units IV pole
height adjustment is in working order. Units audio voice prompts are
functional. Unit includes (1x) foot control which looks to be
responding. Unit includes (1x) remote. Unit is in satisfactory physical
condition.
HG.115/116.v2
Version: V3.00
Manufacture: 03/97
899 Olympus WM-30
Mobile Workstation
with Olympus CV-
230 Camera Control
Unit. Olympus CLV-
U40 Light Source
and Sony PVM-
20M2MDE Monitor
All units are able to power up, lot does not include a pig tail or a
videoscope but some were used for test purposes. The processor
showed it could produce a clear image, however the image was far too
bright as the light sources auto brightness focusing is not functional.
The light source showed to give a light output in all light output modes
(OES, EVIS 100 and EVIS 200). However the auto brightness focusing
was not functional and in manual mode the manual light output
adjustment also did not function. The air pump part of the light source
showed to function. The monitor is functional. Lot is in satisfactory
physical condition.
900 Spembly Medical
140 Cryo Unit
The Cryo Unit’s control, footswitch and regulator appear to be ok. The
probe is not included. Lot is in fair physical condition.
902 Storz Hamou Micro
Hysteroflator
The insufflator power up ok the flow and pressure display and
adjustment keys are working fine. System is in good physical
condition.
904 Olympus PSD-20
Diathermy and MH-
551 Foot Switch
The diathermy is able to power up, with its displays functional and its
function buttons looking to respond. The unit has not been tested with
its output not being verified. Units patient plate alarm is functional.
Unit includes (1x) monopolar cut/coag footswitch which appears to be
responding. Unit is in good physical condition.
906 Vivident Astralis 7
Dental Curing Light
Unit
Dental curing light power supply is working ok, the lamp comes on
and the timer is working. The unit is still in good overall condition.
925 Alcon Ocuscan with
Foot Switch and
Probe Con# D4126
Ass#G21727
The Ocuscan power up ok, display and function keys are working. The
probe test could not be checked at the start up. Although the scan
graphs were displayed but the automatic reading and value were not
coming. Unit is in good physical condition but not fully tested. Two
ultrasound probes is included. The printer is working fine.
932 Sonacel Multiphon
Plus Ultrasonic
Therapy System
The treatment ultrasound unit power up ok The output intensity control
and moving coil meter is in good working order. The timer & US head
selector are functional. The output was tested and it is working fine.
Lot comes with a 1.5MHz and 3MHz head. Fair overall condition.
938 Gettinge 1345
Washer/Disinfector
and Pump
Unit power up ok, display and function keys are working. The unit is
stuck in a disinfection phase and unable to proceed. The door is locked
and the condition of the interior is unknown. The system is in good
physical condition but the process was not tested fully. Year: 2002,
HG.115/116.v2
PAC 5,06 (320), Panel Version: 3.4. counter at 734 cycle. Grundfros X
water pump included
942 Kodak Dryview
8150 Laser Imager
Unit powers up, with units touch screen display in working order. Unit
has not been tested, but the menus look to be accessible. Unit is in
satisfactory physical condition.
Version: 1.1.2
969 Olympus UHI-3
Insufflator and MH-
317 Foot Switch
The insufflator powers up, with its function buttons looking to respond
and its numerical displays in working order. Unit has not been tested.
Unit includes (1x) footswitch. The insufflator is in good physical
condition.
973 Diomed 15 Plus
Laser and Ethicon
Endo Surgery Unit
Unit has no On/Off key – unable to check function – No hand piece or
laser probe included.
986 EMS Swiss
Lithoclast Master
Ultrasound System
and Microaire
Both unit powers up ok, the display and function keys are working.
The Lithoclast system has two of its push button missing. There are no
hand pieces to check the unit. Average physical condition.
1006 Philips Series 50XM
CTG Fetal Monitor
with US and Toco
and Pressure Cuff
Unit powers up, with its function buttons and displays in working
order. The units toco and cardio 2 parameters showed to be
functioning, but the outputs have not been fully tested. The
cardio1/combi parameter has not been checked as the port is damaged.
Units SPO2 shows to be functional as does the NiBP pump, neither of
the parameters outputs have been fully tested or verified. The printer
shows to be functional. The event marker is functional. Unit is in fair
physical condition.
Unit includes: (1x) toco transducer, (1x) US transducer, (1x) event
marker, (1x) SPO2 probe and (1x) NiBP cuff and hose.
1007 Philips Series 50XM
CTG Fetal Monitor
with US and Toco
Leads
Unit powers up, with its function buttons and display in working order.
The units toco and cardio1/combi parameters showed to be
functioning, but the outputs have not been fully tested. The cardio2
parameter is not functional giving ‘ERR’ message displayed. Units
NiBP pump is functional, but has not been fully tested. Unble to check
units SPO2 as unit does not include an SPO2 probe. The printer is
functional. Unit does not include an event marker. Unit is in fair
physical condition.
Unit includes: (1x) Toco transducer, (1x) US transducer and (1x) NiBP
HG.115/116.v2
cuff
1008 Hoskins Cot and
Mattress x 5
The cots are all in good condition.
1019 Kodak Dryview 8150
Laser Imager
Unit is able to power up, with units touchscreen display showing to be
in working order. Unit has not been tested. Unit is in good physical
condition.
Version: 1.1.b04
1026 Kodak Dryview 8900
Laser Imager
Powers up but unable to complete the system boot-up (message on
display panel reads: invalid system disk). No further checks could be
carried out. The front corner of the outer casing of the unit has been
damaged otherwise it is in reasonable physical condition.
1030 Cots with Pentaflex
Paediatric
Mattresses x2
Fair physical condition
1031 Cots with Pentaflex
Paediatric
Mattresses x2
Fair physical condition
1032 Cots with Pentaflex
Paediatric
Mattresses x2
Fair physical condition
1033 Cots with Pentaflex
Paediatric
Mattresses x2
Fair physical condition
1035 Arjo Grey Electric
Tilt Table
Unit is able to power up, however units footswitch is damaged causing
the up and down movement to only be able to go down with its button
getting stuck in place. Units motorised forward/backward tilt showed
to be functional. The mechanical head tilt is functional. Unit is in fair
physical condition.
1040 Thermo Klenz
Washers - 1015547
The unit is able to power up, with its display in working order and its
function buttons looking to respond. Unit has not been tested. Unit is in
satisfactory physical condition.
1041 Thermo Klenz
Washers - 1015546
The unit is able to power up, with its display in working order and its
function buttons looking to respond. Unit is displaying message for
‘maintenance’. Unit has not been tested. Unit is in satisfactory physical
HG.115/116.v2
condition.
1042 Thermo Klenz
Washers - 1015548
The unit is able to power up, with its display in working order and its
function buttons looking to respond. Unit is displaying message for
‘maintenance’. Unit has not been tested. Unit is in satisfactory physical
condition
1043 Thermo Klenz
Washers - 1015558
The unit is able to power up, with its display in working order and its
function buttons looking to respond. Unit has not been tested. Unit is in
satisfactory physical condition.
1046 Kodak Carestream
Dryview 8150 Laser
Imager
Powers up and completes the initialisation process successfully.
Display panel and touch screen interface are in good working order and
the system menus are accessible. No further checks carried out. The
unit is in reasonable physical condition. Software version: 1.1.B04
1052 Kodak Dryview 5800
Laser Imager
Powers up and completes the initialisation process successfully.
Display panel and function keys are in good working order and the
system menus are accessible. Includes 2x film trays (8x10-20x25 and
14x17-35x43). No further checks carried out. The unit is in good
physical condition. Software version: 2.4.B01P1.
1053 Kodak Dry View
5800 Laser Imager
Powers up and completes the initialisation process successfully.
Display panel and function keys are in good working order and the
system menus are accessible. Includes 3x film trays (8x10-20x25,
10x12-25x30 and 14x17-35x43). No further checks carried out. The
unit is in good physical condition. Software version: 2.3.B02.
1054 Lisa 500 Autoclave
and Printer - 090592
W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch
screen are fully functional. A vacuum tested cycle was carried out and
passed with no problem. Printer is working but need ribbon and paper.
The interior of the chamber and door gasket are in good condition. The
memory card MMC- SD and trays are not included. Unit is in good
physical condition with some scuff and marks on the covers. Counter at
1556. Type B autoclave, Year 2009
1055 Lisa 500 Autoclave
and Printer
W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch
screen are fully functional. A vacuum tested cycle was carried out and
passed with no problem. Printer is working fine. The interior of the
chamber and door gasket are in good condition. The memory card
MMC- SD and trays are not included. Unit is in good physical
condition with some scuff and marks on the covers. Counter at 1578.
Type B autoclave, Year 2009
HG.115/116.v2
1056 Lisa 500 Autoclave
and Printer
W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch
screen are fully functional. A vacuum tested cycle was carried out and
unit failed the test – error A390 was displayed. Printer is working fine.
The interior of the chamber and door gasket are in fair condition. The
memory card MMC- SD and trays are not included. Unit is in fair
physical condition with some scuff and marks on the covers. Counter at
1674. Type B autoclave, Year 2009
1057 Lisa 500 Autoclave
and Printer
W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch
screen are fully functional. A vacuum tested cycle was carried out and
passed with no problem. However, a ‘Clock Battery Low’ message is
displayed on the screen. Printer is working fine. The interior of the
chamber and door gasket are in good condition. The memory card
MMC- SD and trays are not included. Unit is in good physical
condition with some scuff and marks on the covers. Counter at 1641.
Type B autoclave, Year 2009
1061 Agfa Drystar 3000
X-Ray Printer
Unit is able to power up, with it able to pass through its initial
initialisation successfully. Units display is functional, with its function
buttons looking to respond. Unit has not been test. Unit is in
satisfactory physical condition.
Software version: R1.95
1062 Marquette Mac 5000
ECG Recorder with
Leads - 16673
Units power indication light illuminates however units display does not
show to be coming on. unable to carry out any further checks. Unit
includes 12wire ECG. Unit is in fair physical condition.
1064 Stryker Transfer
Trolley
Units mechanical height adjustment is in working order, with the its
mechanical tilt also in working order. Units wheels are freely movable
with the brake lock functional. The cot sides are in place. The mattress
is in place and in fair physical condition. Unit is in satisfactory
physical condition.
1075 Maquet Electric
Operating Table
with Qty of Extra
Accessories and
Extensions
The operating table is in reasonable physical condition with all
cushions in place and the transport cart is also included. Remote
control is included but only the up/down motorised movement of the
table is functional, other functions appear to be not working (the motor
starts when the keys are activated from the remote control but no
movement is detected). The battery is in working condition.
Accessories included with the table consists of extensions, head rests
and leg rests.
HG.115/116.v2
1078 Huntleigh Akron
Electric Patient
Couch - Green
Unit looks to be powering up, with its electrical height adjustment in
working order. The mechanical movements also show to be in working
order. The wheels are freely movable and the brake locks are
functional. Unit includes (1x) footswitch which is functional. Unit is
in fair physical condition
1080 Luxo, Daray x 2 and
Heine HL5000
Examination Lights-
100587, 6321, 126511
The two Daray examination lights are not powering up, the Luxo
examination light and the Heine wall mounted examination light are
both able to power up and produce a light output. All of the units
mechanical movements are work however there locks are not. All four
of the examination lights have bad damage such as cracks and
scratches to casing etc, bad dents to casing etc, loose casing etc. Lot is
in below satisfactory physical condition.
1081 Green Electric
Dialysis Chair
Unit powers up, with its electrical height adjustment in working order.
Units mechanical movements are also in working order. Units wheels
are freely movable and its brakes are functional. Unit includes (1x)
remote which is functional. Unit is in satisfactory physical condition.
1082 HNE Akron Electric
Couch - Red
Unit powers up, with its electrical movements in working order. Units
mechanical movements are also in working order. Units wheels are
movable and the brakes function. Unit includes (1x) remote which is
functional. Unit is in reasonable physical condition.
1085 Medtronic Lifepak
20 Defib and Leads -
13251 Serial
Number 30493063
The defibrillator powers up and is able to pass through its initial self-
test successfully. Units function buttons, rotary knob and display are in
working order. Unit does not include any paddles but does include (1x)
AED cable. Units output was tested using a defibrillator analyser and
all readings taken showed to be within set limits. Unit showed to be
functional in both manual and AED modes. Unit was able to charge to
360J. The defibrillator is biphasic. The audio voice prompts are
functional. Unit includes pacer function which showed to be functional
but however out of limits. Unit includes (1x) 3 wire ECG and units
ECG was verified as functional. Units SPO2 parameter has not been
tested but unit includes (1x) SPO2 trunk Units printer is functional but
the battery is not. Unit is in good physical condition.
Manufacture date: 2002
1088 Pentax EPM-3500
Camera
Control/Lightsource
- 055808 Serial
Number E0011304
The processor powers up, but unable to determine if units function
buttons are functional, none of the indication lights are on. The unit has
not been tested. Unit is in satisfactory physical condition.
Manufacture date: 2004
HG.115/116.v2
Lamp hours: 0.0
Total hours: 1693.5
1095 SLE 2000 Infant
Ventilator
The ventilator is able to power up, with its numerical displays in
working order. Unit is missing its reset function button but the rest of
units function buttons are in working order, also units CPAP/PEEP
regulator is damaged and not functional. The ventilator showed to be
functional when it was tested on a medical compressor. However the
PEEP cannot be adjusted and CPAP mode is not able to function as the
CPAP/PEEP regulator is not working. The rest of the vent modes
(SIMV, PTV and CMV) showed to be in working order. Units output
has not been verified. Unit is able to be calibrated to 21%. Unit does
not come with any O2 or air hoses, also unit is missing its stand. The
ventilator is in reasonable physical condition.
Vent modes: SIMV, PTV, CMV and CPAP
Operation hours: 4563.25h
1111 Welch Allyn CP10
ECG Recorder with
ECG Lead
Unit is able to power up, units function buttons and display are in
working order. Unit has not been fully tested. Unit includes 12 wire
ECG. No paper to check the printer. The battery is functional. There is
a rattling sound with something loose inside of the unit. Unit is not on
a trolley. Unit is in fair physical condition.
Software version:1.06
1115 Ascena Syringe
Pumps x 18 - 7525,
103094, 054391,
101273, 118485,
7044, 101577, 101621
The syringe drivers are in satisfactory physical condition. 4 of the units
have damaged/broken cases and 2 of the units have parts removed from
inside. Not tested.
1124 Endoscopy Stack
Trolley with Sony
Trinitron Monitor
and Olympus
Surgical CO2
Insufflator
Unable to test lot as the processor does not include a camera head. The
processor does however power up, with its indication lights in working
order and its function buttons looking to respond but not verified. The
Sony monitor also powers up but again has not been tested. The lot is
in satisfactory physical condition.
1125 Fisher and Paykel
Neopuff Infant
Resuscitator
Unit powers up, with its function buttons and numerical displays in
working order. Unit showed to heat up well in ‘prewarm’, ‘Manual’
and ‘Baby’ modes. With the heat adjustment in ‘Manual’ and ‘Baby’
modes functional. ‘Baby’ mode showed to be in working order. Unit
includes (1x) skin temperature probe which is functional, however its
reading has not been verified. Units examination light and timer are
both functional. Units electrical height adjustment is functional but
does jolt when raising up the unit. The Neopuff part of the unit has not
HG.115/116.v2
been checked. Unit includes an O2 hose and a mattress. Unit is in
satisfactory physical condition.
1126 Fisher and Paykel
Neopuff Infant
Resuscitator
Unit powers up, with its function buttons and numerical displays in
working order. Unit showed to be able to heat up well in ‘prewarm’,
‘manual’ and ‘baby’ modes. With the heat adjustment in ‘manual’ and
‘baby’ modes functional. However after a brief period of time the unit
displays alarm message ‘sensor disconnected’ with the numerical
display stopping, this happens when the skin temperature probe is
actually in place. ‘Baby’ mode is not fully functional. Unit includes
(1x) skin temperature probe. Units examination light and timer are both
functional. Units electrical height adjustment is functional. The
Neopuff part of the unit has not been checked. Unit includes an O2
hose and a mattress. Unit is in satisfactory physical condition.
Manufacture date: 2004-11
1129 Siemens Servo
Ventilator 300 and
Parts
The ventilator powers up, with units function buttons, indication lights
and numerical displays in working order. The ventilator showed to be
functional when it was tested using a medical compressor. With the
BPM, tidal volume and minute volume all matching up accordingly.
Units output has not been verified. Unit showed to be functional in
both volume and pressure modes. Units O2 was able to calibrate to
21%. Units battery functions but alarms for being low. Unit includes
O2 and air hoses. The ventilator is in satisfactory physical condition
with t however not including a stand and there are also chips to
paintwork around edges of unit.
Operation hours: 30950h
1130 Siemens Servo
Ventilator 300 and
Parts
The ventilator powers up, with units function buttons, indication lights
and numerical displays in working order. The ventilator showed to be
functional when it was tested using a medical compressor. With the
BPM, tidal volume and minute volume all matching up accordingly.
Units output has not been verified. Unit showed to be functional in
both volume and pressure modes. Units O2 was able to calibrate to
21%. Units battery functions but alarms for being low. Unit includes
O2 and air hoses. The ventilator is in satisfactory physical condition
with it however not including a stand and there are also chips to
paintwork around edges of unit.
Operation hours: 31147h
1131 Drager Photo
Therapy 4000 and
8000 Therapy
Drager photo therapy lamp 4000 & 8000. The 4000 has both lamp type
coming on and the counter is at 125831hrs. The 8000 lamp is working
fine with the fan and alarm tested ok. Counter at 296639hrs. Both stand
HG.115/116.v2
Lamps and castors are in fair condition.
1143 SLE 2000 HFO
Ventilator
The ventilator powers up, with units main display and numerical
displays in working order. Units function buttons look to be
responding. Unable to test the ventilator as the unit is missing its
expiratory valve. Unit includes O2 and air hoses. Units blended gas
output connection is badly damaged. The ventilator is in satisfactory
physical condition.
Vent modes: SIMV, PTV, CMV and CPAP
Operation hours: 25618.91h
1144 SLE 2000 Infant
Ventilator
The ventilator is able to power up, with its function buttons, numerical
displays and gauges in working order. The ventilator showed to be
functional when tested with a medical compressor. All ventilation
modes showed to be in working order. Units output has not been
verified. Units oxygen was able to calibrate to 21%. Unit includes O2
and air hoses. The ventilator is in satisfactory physical condition.
Vent modes: SIMV, PTV, CMV and CPAP
Operation hours: 6839.96h
1145 Dantec Dynamic
Keypoint EMG
Stimulator
The unit is able to power up, with units function buttons and monitor
functional. The unit has not been tested and unit does not include any
accessories or leads. Unit is able to enter its main menu. The unit is in
reasonable physical condition.
V2.12
1151 Engell
Fridge/Freezer
The unit is able to power up, the unit shows signs it is able to start to be
able to cool down, however the unit has not been fully tested. Unit is
fair physical condition.
1153 Kodak Dryview 8900
Laser Imager
Unit is able to power up, with its touchscreen display in working order.
The unit has not been tested. Unit is in satisfactory physical condition.
Version: DV8900_2.14
1154 Dometic Pharmacy
Refrigerator
The unit is able to power up, with the unit not showing any signs it is
able to cool down, however the unit has not been fully tested. Unit
includes its key for its door lock. Unit is in satisfactory physical
condition.
1156 Fixed Height Patient
Couch - Black
The couch is in satisfactory physical condition, with some rips at
corners of the cushions. The back movement is freely movable.
HG.115/116.v2
1158 Grey Electric
Dialysis Chair
The unit is able to power up, with units electrical movements showing
to be in working order. Unit includes (1x) remote. Units wheels show
to be freely movable. Units cushions have some rips to the corners.
Unit is in reasonable physical condition.
1169 HP Pagewriter 200I
ECG Recorder with
ECG Lead
Unit powers up, with units function buttons looking to respond and its
display functional however there are lots of pixels missing from it. Unit
has not been tested. Unit include 12 wire ECG. Unit does not include
any paper to test the printer. The battery is functional. Unit is not on a
trolley. Unit is in fair physical condition.
1175 Flat Screen Monitors
x 3
The two flat screen monitors both power up, however the they have
large cracks/damage to the screens and the two monitors are not
functional. The Dell monitor is able to power up, not fully tested. Lot is
in poor physical condition as the screens are badly damaged.
1180 Braun SCT 2
Tourniquets x 2 -
14822, 18867
Both units appear to have their gauges and regulators intact. Lot has
not been tested. Both units are complete with their correct hoses. Lot is
in reasonable physical condition with chips to paint work.
1190 Trolley with Siemens
Unity Audiology
Test Equipment
The unity PC probe Mic is able to power up, with it not being tested
and neither has the PC hearing instruments analyser. There are no
complete leads or accessories with lot so unable to test unit. The Hi-pro
powers up. Lot is in fair physical condition.
1192 FMS Duo with
Arthroscopy Fluid
Management System
with Integrated
Shower and Foot
Switch
Unit powers up, with its displays functional and its function buttons
looking to respond. Unit has not been tested. Unit is in complete and
missing its shaver. Unit includes (1x) foot control but unable to verify
if it is responding. Unit is in fair physical condition, with the suction
pumps cover missing.
1207 Quayle Deroter and
Audio/Hearing Aid
Fitting Tool with
Dust Extractor, Foot
Pedal and Drill Head
- MS3031,
CP0027975
The polisher is able to power up, with the polisher and fans functional.
Some of the vacuum pumps wires have been cut so unable to power up
unit. Unit includes a micro drill, with its hand piece included, the micro
drill has not been tested. Unit includes (1x) footswitch which has
exposed cables wires where the mains cable is coming away from the
footswitch. Unit includes a stand for the polisher. Unit includes a box
of hearing accessories (not checked). Lot is in reasonable physical
condition.
1218 Medilink System
Treatment Unit -
Cased
Unit is able to power up, with its function buttons and displays in
working order. Unit has not been fully tested, but module modes are
selectable (ultrasound and shortwave). Unit includes (2x) ultrasound
HG.115/116.v2
transducers, both 1.1 and 3.3 MHz with crystal areas (7.0cm2 and
0.79cm2). Unit also includes shortwave (1x) disk electrode. Unit is in
fair physical condition.
Version: v1.20
1227 Braun Tourniquet x
2
Both tourniquets are in fair condition with the regulators. Manometer
and pneumatic switch appears to be functional. Units were not tested
with medical air.
1229 Datex Normocap
200 x 2
Only one of the gas monitor are working. The working ones has its
display and function keys working fine. Co2 function is working but
battery failure and air leak detected is displayed. There is a piece of the
cover chipped off near the water trap assembly. The second unit is not
powering up at all.
1231 Aesculap Drill Unit
Type: GA 244
The main drill power up but there is no foot switch or hand piece to
test functionality. The irrigation pump is not powering up. Lot in
average physical condition.
1248 Mobile Eye Chart Unit powers up, with snellen back light is illuminating. Unit is missing
its plastic caster for the stand. Unit is in poor physical with damage to
the plastic graphs.
1249 Lisa 500 Autoclave
and Printer
W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch
screen are fully functional. A vacuum tested cycle was carried out and
passed with no problem. ‘Dust filter needs replacement. Printer is
working fine. The interior of the chamber and door gasket are in good
condition. The memory card MMC- SD and trays are not included.
Unit is in good physical condition with some scuff and marks on the
covers. Counter at 1560. Type B autoclave, Year 2009
1250 Lisa 500 Autoclave
and Printer
W&H Sterilisation, Model: LISA 517 – Unit power up ok, LCD/touch
screen are fully functional. A vacuum tested cycle was carried out and
passed with no problem. Printer is working fine. The interior of the
chamber and door gasket are in good condition. The memory card
MMC- SD and trays are not included. Unit is in good physical
condition with some scuff and marks on the covers. Counter at 1578.
Type B autoclave, Year 2009