Issue date: Valid until
15 March 2017
02 April 2022
G Stones Technical Manager For and on behalf of the Head of
NMO I Stanton Avenue I Teddington I TW11 OJZ I United KingdomTel +44 (0) 20 8943 7272 I Fax +44 (0) 20 8943 7270 I WebNMO is part of the Regulatory Delivery directorate within the Department for Business,
EU-type examination certificate UK/0126/0123 Revision
Issued by:
NMO Notified Body Number 0126
In accordance with the requirements of the (S.I. 2016 No. 1153) which implement, in the Un2014/32/EU, this EU-type examination
Isoil Impianti S.p.A 74, Via Madonna delle RoseITA - 24061 Albano S. Alessandro BG, Italy
In respect of a static liquid fuel meter measuring system road tankers having the following characteristics:
Model designation: Nominal bore: Maximum rate of flow: Minimum rate of flow: Maximum operating pressure:Minimum delivery: Liquids measured:
The necessary data (principal characteristics, alterations, securing, functioningfor identification purposes and conditions (when applicable) are set out in the descriptive annex to this certificate.
This revision replaces previous versions of the certificate.
UK/0126/
15 March 2017
02 April 2022
For and on behalf of the Head of Technical Services Stanton Avenue I Teddington I TW11 OJZ I United Kingdom
Tel +44 (0) 20 8943 7272 I Fax +44 (0) 20 8943 7270 I Web www.gov.uk/government/organisations/regulatoryNMO is part of the Regulatory Delivery directorate within the Department for Business, Energy & Industrial Strategy
type examination certificate Revision 4
Notified Body Number 0126
In accordance with the requirements of the Measuring Instruments Regulations 20which implement, in the United Kingdom,
examination certificate has been issued to:
74, Via Madonna delle Rose
static liquid fuel meter measuring system unloading ships and rail and having the following characteristics:
SMS 75 (50 mm / 2”) / SMS150 (75mm / 3”) 50 mm (2”) / 75 mm (3”) 500 litres / minute (2” system), 1300 litres / minute (3” system) 40 litres / minute (2” system) 100 litres / minute (3” system)
Maximum operating pressure: 10 bar (3” system), 20 bar (2” system) 50 litres (2” system), 100 litres (3” system)
Liquids other than water with viscosity between 0.3 mPa.s. and 750 mPa.s, except Non liquids
The necessary data (principal characteristics, alterations, securing, functioningidentification purposes and conditions (when applicable) are set out in the
descriptive annex to this certificate.
This revision replaces previous versions of the certificate.
UK/0126/0123
MI-005
0135
www.gov.uk/government/organisations/regulatory-delivery Industrial Strategy
Measuring Instruments Regulations 2016 Council Directive
certificate has been issued to:
unloading ships and rail and
SMS 75 (50 mm / 2”) / SMS150 (75mm / 3”)
10 bar (3” system), 20 bar (2” system) litres (2” system), 100 litres (3” system)
Liquids other than water with viscosity between 0.3 mPa.s. and 750 mPa.s, except Non - Newtonian
The necessary data (principal characteristics, alterations, securing, functioning etc) identification purposes and conditions (when applicable) are set out in the
2
Descriptive Annex
1 INTRODUCTION
This pattern is a static meter measuring system for the delivery of liquids with low viscosity stored at atmospheric pressure. It comprises a pump (could be external to the measuring system), a special gas extractor and a meter with a full hose and/or an empty hose. The register may be either mechanical or electronic. The system may be configured as follows:
(a) fixed, mounted at ground level (b) fixed, mounted on a frame (c) fixed, mounted underground (d) mobile, mounted on a frame with wheels
2 CONSTRUCTION
2.1 Hydraulics
The system configuration is shown in Figure 1A, accompanied by a key at Section 9.1. The delivery could be made through:
- Static metallic pipes or - Full or empty flexible hoses (with or without Hose Reel) or - Loading Arms (Top or Bottom execution).
The system can also be used for un-loading applications. In this case the transfer point will be upstream of the measuring system as shown in Figure 1B
2.2 List of special gas extractors
System Manufacturer
Model designation EEC pattern approval number
SMS 75 Isoil Impianti S.p.A FDS 75 I 95.07.01.0001
SMS 150 Isoil Impianti S.p.A FDS 150 I 95.07.01.0002
Note: The special gas extractor is optional when it is demonstrated that: - The pressure at the pump inlet can never fall below either the atmospheric pressure or the saturated vapor pressure of the measured liquid.
- Neither air intake nor gas release can occur in the liquid upstream of the meter.
2.3 List of meters
System Manufacturer
Model designation EEC pattern approval number
SMS 75 Isoil Impianti S.p.A SBM 75 D 81.5.243.19
SMS 150 Isoil Impianti S.p.A SBM 150 D 79.5.243.01
3/13
2.4 List of electronic and mechanical meter registers
System Manufacturer
Model designation EEC pattern approval number / Evaluation certificate
SMS 75 Veeder-Root Counter 7887 D 81.5.243.19
SMS 150 Veeder-Root Counter 7887 D 79.5.243.01
Both Veeder-Root Preset 7889 D 83.5.531.12
Both Veeder-Root Ticket Printer 7888 D83.5.521.09
Both Isoil Impianti S.p.A Vega T LNE - 13954
Both Isoil Impianti S.p.A Vega II LNE – 12081
2.5 Optional Equipment
• Preset valve
• Pulse emitter
• Flow rate indicator
• Flow control Valve
• Temperature Probe
• Other Accessories without metrological effect e.g. pressure gauge etc
• Density meter
2.6 Peripheral devices
The printer is typically an Epson TM-U295 slip printer. It is used for printing delivery tickets showing the date and time of a delivery, the ticket number, the quantity delivered, the price and other operator entered details. The nature and format of the ticket record is set up using the Register menu system. Any simple recipient printer may be used in place of the Epson TM-U295 printer provided:
(i) it bears the CE mark for conformity to the EMC Directive 89/336/EEC;
(ii) it is not capable of transmitting any data or instructions into the Vega T/ Vega II other than for releasing the printout, checking for correct data transmission;
(iii) it prints delivery results and other data as received from the Vega T/ Vega II without any modification or further processing.
3 SUPPORTING DOCUMENTATION
MA 026-04-EN: manual for 2" system with electronic register MA 026-03-EN: manual for 2" system mechanical counter MA 027-10-EN: manual for 3" system with electronic register MA 027-05-EN: manual for 3" system mechanical counter
4/13
4 TECHNICAL DATA
Technical data SMS 75 SMS 150 unit
Nominal bore: 50 75 mm Maximum rate of flow*1: 500 1300 litres/ minute Minimum rate of flow: 40 100 litres/ minute Maximum working pressure: 20 10 bar Minimum working pressure 1 1 bar Minimum delivery : 50 100 litres Liquids measured: Liquids other than water with viscosity between
0.3 mPa.s. and 750 mPa.s, except Non - Newtonian liquids
Liquid Temperature Range*2: (The temperature range is always 60 °C) Climatic environment: - 25 °C to +55 °C
open, condensing Electromagnetic environment:
E2
Mechanical environment: M1 (static installation) Accuracy class: 0.3 or 0.5*3
*1 The maximum rate of flow, however, must not be greater than the maximum rate of flow of any component of the system
*2 The minimum temperature of the range varies in steps of 10 °C, the minimum and maximum temperature must be between - 40°C and +170 °C.
e.g. - 40 °C to +20 °C up to +110 °C to +170 °C.
*3 Accuracy class: the accuracy class is to be selected depending the product and the temperature product as per table 5 of Directive 2014/32/EU, Annex VII (MI-005).
5 APPROVAL CONDITIONS
The certificate is issued subject to the following conditions:
5.1 Legends and inscriptions
5.1.1 The instrument bears the following legends: Manufacturers mark or name Accuracy class Maximum operating pressure Operating temperature range Minimum delivery Flow rate range Serial number Certificate number ‘CE’ marking Supplementary metrology marking Notified body identification number
6 SECURING
6.1 Sealing
The following shall be sealed to prevent unauthorised adjustment or dismantling:
a) the components of the system in accordance with the requirements of the relevant approvals as listed in section 2. Examples are shown in Figures 2 to 5.
b) the plate referred to in 5 shall be sealed or welded to a support of the system.
5/13
7 ALTERNATIVES
There are no alternatives.
8 RECOMMENDED TESTS
8.1 The meter measuring system shall be tested at a minimum of two substantially different rates of flow between the maximum and minimum rates of flow.
8.2 The meter measuring system shall be tested by a minimum of three repeat runs at each rate of flow. Each run shall have a minimum duration of one minute.
8.3 Due regard shall be paid to the viscosity of the liquid used for the verification tests to ensure that the permitted limits of error are not likely to be exceeded when other liquids which the system is intended to measure are metered.
9 ILLUSTRATIONS
Figure 1A Measuring system with flow meter and special gas extractor (model SMS) Figure 1B Measuring System for Unloading Figure 2A Sealing diagram: SMS 75 – MECHANICAL COUNTER Figure 2B Sealing diagram: alternative execution SMS 75 – MECHANICAL
COUNTER Figure 3A Sealing diagram: SMS 75 – ELECTRONIC COUNTER Figure 3B Sealing diagram: alternative execution SMS 75 – ELECTRONIC
COUNTER Figure 4 Sealing diagram: SMS 150 – MECHANICAL COUNTER Figure 5 Sealing diagram: SMS 150 – ELECTRONIC COUNTER
6/13
10 CERTIFICATE HISTORY
ISSUE NO. DATE DESCRIPTION
UK/0126/0123 03 April 2012 Type examination certificate first issued.
UK/0126/0123 Rev 1 25 February 2013 Revision 1 issued. Section 2.1 un-loading application configuration and associated Figure 7 added
UK/0126/0123 Rev 2 27 September 2013 Revision 2 issued. Section 4 minimum working pressure added. Alternative measuring system plate figure 6b added. Reference to this added to section 6. Figure 6 renamed figure 6a.
UK/0126/0123 Rev 3 22 May 2015 Revision 2 issued. Front page product name redefined Section 2.2 Note added Section 2.4 EEC numbers for SMS75 and SMS 150 corrected Section 2.6 references to Vega II added Section 4 *3 note added Section 9 Illustrations renumbered, Figures 6a and 6B replaced by figure 6
UK/0126/0123 Rev 4 15 March 2017 References to 2004/22/EC changed to 2014/32/EU. Section 4: Accuracy Class amended to 0.3 or 0.5 Figure 6 Example of Measuring System plate deleted.
7/13
1) special gas extractor 2) flow meter 3) check valve 4) sight glass of the special gas extractor 5) sight glass 6) sight glass 7) back pressure valve V1) manual valve V2) manual valve
Figure 1a Measuring system with p.d.meter and special gas extractor (model SMS)
8/13
1) special gas extractor 2) flow meter 3) check valve 4) sight glass of the special gas extractor
Figure 1B Measuring System for Unloading
1 P.D. METER SBM 75 2 PRESET REGISTER V.R. 7889 (OPTIONAL) 3 CHECK VALVE 4 PRESET VALVE (OPTIONAL) 5 COUNTER V.R. 7887 6 TICKET PRINTER V.R. 7888 (OPTIONAL) 7 PIPE 8 STRAINER / SPECIAL GAS EXTRACTOR 9 SIGHT GLASS
Figure 2A Sealing diagram: SMS 75 – MECHANICAL COUNTER
9/13
1 P.D. METER SBM 75 2 PRESET REGISTER V.R. 7889 (OPTIONAL) 3 CHECK VALVE 4 PRESET VALVE (OPTIONAL) 5 COUNTER V.R. 7887 6 TICKET PRINTER V.R. 7888 (OPTIONAL) 7 PIPE 8 STRAINER / SPECIAL GAS EXTRACTOR 9 SIGHT GLASS 10 ALTERNATIVE TYPE OF CHECK VALVE WITHOUT PRESET VALVE
Figure 2B Sealing diagram: alternative execution SMS 75 – MECHANICAL COUNTER
10/13
1 P.D. METER SBM 75 2 PULSE EMITTER 3 CHECK VALVE 4 PRESET VALVE (OPTIONAL) 5 VEGA T OR VEGA II ELECTRONIC COUNTER 6 PNEUMATIC ACTUATOR (OPTIONAL) 7 PIPE 8 STRAINER / SPECIAL GAS EXTRACTOR 9 SOLENOID VALVES BOX (OPTIONAL) 10 Pt100 TEMPERATURE PROBE (OPTIONAL), ITS POSITION MAY VARY FROM THE DIAGRAM 11 SIGHT GLASS
Figure 3A Sealing diagram: SMS 75 – ELECTRONIC COUNTER
11/13
1 P.D. METER SBM 75 2 PULSE EMITTER 3 CHECK VALVE 4 PRESET VALVE (OPTIONAL) 5 VEGA T OR VEGA II ELECTRONIC COUNTER 6 PNEUMATIC ACTUATOR (OPTIONAL) 7 PIPE 8 STRAINER / SPECIAL GAS EXTRACTOR 9 SOLENOID VALVES BOX (OPTIONAL) 10 SIGHT GLASS 11 ALTERNATIVE TYPE OF CHECK VALVE WITHOUT PRESET VALVE 12 Pt100 TEMPERATURE PROBE (OPTIONAL), ITS POSITION MAY VARY FROM
THE DIAGRAM
Figure 3B Sealing diagram: alternative execution SMS 75 – ELECTRONIC COUNTER
12/13
1 P.D. METER SBM 150 2 CHECK VALVE 3 PRESET VALVE (OPTIONAL) 4 PRESET REGISTER V.R. 7889 (OPTIONAL) 5 COUNTER V.R. 7887 6 TICKET PRINTER V.R. 7888 (OPTIONAL) 7 PIPE 8 STRAINER / SPECIAL GAS EXTRACTOR 9 SIGHT GLASS 10 ALTERNATIVE TYPE OF CHECK VALVE WITHOUT PRESET VALVE
Figure 4 Sealing diagram: SMS 150 – MECHANICAL COUNTER
13/13
1 P.D. METER SBM 150 2 PULSE EMITTER 3 CHECK VALVE 4 PRESET VALVE (OPTIONAL) 5 VEGA T OR VEGA II ELECTRONIC COUNTER 6 PNEUMATIC ACTUATOR (OPTIONAL) 7 PIPE 8 STRAINER / SPECIAL GAS EXTRACTOR 9 SOLENOID VALVES BOX (OPTIONAL) 10 Pt100 TEMPERATURE PROBE (OPTIONAL), ITS POSITION MAY VARY FROM THE DIAGRAM 11 SIGHT GLASS 12 ALTERNATIVE TYPE OF CHECK VALVE WITHOUT PRESET VALVE
Figure 5 Sealing diagram: SMS 150 – ELECTRONIC COUNTER
Crown copyright 2017 NMO Regulatory Delivery directorate Department for Business, Energy & Industrial Strategy This material may be freely reproduced except for sale