EVALUATING THE SURFACE PROPERTIES OF HA COATING ON Co-Cr BASED ALLOY SUBSTRATE HIRBOD MALEKI KHEIMEH SARI This project report submitted in partial fulfilment of the requirements for the award of the degree of Master of Engineering (Mechanical Advanced Manufacturing Technology) Faculty of Mechanical Engineering Universiti Teknologi Malaysia JANUARY 2014

Upload: others

Post on 20-Jun-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

EVALUATING THE SURFACE PROPERTIES OF HA COATING ON Co-Cr

BASED ALLOY SUBSTRATE

HIRBOD MALEKI KHEIMEH SARI

This project report submitted in partial fulfilment of

the requirements for the award of the degree of

Master of Engineering (Mechanical – Advanced Manufacturing Technology)

Faculty of Mechanical Engineering

Universiti Teknologi Malaysia

JANUARY 2014

Page 2: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

iii

Dedicated to…

My beloved parents, Mahdokht and Ardeshir

My cherished brother

And lastly my dear family and friends

Thanks for your endless support

Page 3: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

iv

ACKNOWLEDGEMENT

Apart from the efforts of mine, the success of any project largely depends on

the inspiration and guidance comes from people around. I take this opportunity to

declare my appreciation of the people who have been instrumental in accomplishing

this project. I would like to extend my sincere thanks to all technical staffs at

Production and Material Science Laboratories. I am deeply indebted to Assoc. Prof.

Dr. Izman bin Sudin for his support and constant supervision as well as for providing

necessary information and equipment regarding to my project.

I would like to express my gratitude towards my parents for their kind co-

operation and encouragement which aided me in completion of my thesis. My thanks

and appreciations also go to my colleagues in developing the project and people who

have willingly helped me out with their abilities.

Page 4: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

v

ABSTRACT

Hydroxyapatite (HA) is the main structural component of natural bone and

due to its excellent biocompatibility and bioactivity it can be used in biomedical

application as a coating layer for metallic implants to help formation of chemical

bonding at HA/bone interface and work as a protective layer against ion release from

a metallic prosthesis. In this study, HA bioactive coating was created using sol-gel

method on the high carbon CoCrMo substrate. Although sol-gel is simple and cost

effective method with capability to control chemical composition and able to coat on

the complex-shape implants, massive cracks of HA sol-gel coated layer on implants

are still the major issue. Cracks can be minimized by changing the viscosity,

composition or variation in heat treatment procedure. In this study, Na3PO4 and

CaCl2 were used as the main precursors in sol-gel preparation. The sol-gel was

centrifuged at three different speeds (1500, 1750 and 2000 rpm). Coated specimens

were sintered at 500°C, 600°C and 700°C for 20 minutes and 1 hour respectively.

HA coated samples were analyzed under FESEM, XRD, AFM and electrochemical

corrosion tests. The initial FESEM test revealed that the best centrifuging speed that

results in a crack free HA coated layer at room temperature is 1750 rpm with

viscosity of 1798 CP. The FESEM and XRD results also revealed that the best

surface morphology with semi-crystalline microstructure belong to the sample

sintered at 600°C for 20 min. Also it is concluded that sintering temperature above

600°C for HA coating on Co-Cr based alloys results in cracks propagation.

Moreover, in terms of surface roughness all coated and sintered samples except the

one sintered at 500°C for 20 min, showed a good result as well. Finally, in terms of

corrosion resistance the sample sintered at 600°C for 20 min showed the corrosion

rate almost 3.5 times lesser than uncoated sample.

Page 5: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

vi

ABSTRAK

Hydroxyapatite (HA) adalah komponen utama struktur semula jadi tulang dan

disebabkan sifat biokompatibiliti dan bioaktivitinya yang sangat baik, ia boleh

digunakan dalam aplikasi bioperubatan sebagai lapisan salutan untuk implan logam

bagi membantu pembentukan ikatan kimia pada antara muka HA/tulang dan

bertindak sebagai lapisan perlindung terhadap pelepasan ion daripada prostesis

logam. Dalam kajian ini, salutan bioaktif HA telah dihasilkan dengan kaedah sol-gel

ke atas substrat CoCrMo berkarbon tinggi. Walaupun kaedah sol-gel adalah kaedah

yang mudah dan kos efektif serta berupaya untuk mengawal komposisi kimia, dan

mampu menyalut salutan ke atas implan yang kompleks, namun keretakan lapisan

sol-gel HA pada implan masih menjadi isu utama. Keretakan boleh diminimumkan

dengan menukar kelikatan, komposisi atau variasi dalam prosedur rawatan haba.

Na3PO4 dan CaCl2 telah digunakan sebagai prekursor utama dalam penyediaan sol-

gel. Sol-gel telah diputar pada 1500, 1750 dan 2000 rpm. Spesimen yang tersalut,

disinter pada 500°C, 600°C dan 700°C, selama 20 minit dan 1 jam. Sampel yang

disaluti HA telah dianalisa dengan FESEM, XRD, AFM dan ujian kakisan

elektrokimia. Ujian awal FESEM mendapati kelajuan putaran yang terbaik adalah

1750 rpm dengan kelikatan 1798 CP pada suhu bilik. Ia dapat menghasilkan lapisan

salutan HA yang bebas dari keretakan. Keputusan FESEM dan XRD juga mendapati

bahawa morfologi permukaan dan mikrostruktur semi-kristal terbaik ialah sampel

yang dibakar pada 600°C selama 20 min. Kesimpulanya, suhu pembakaran melebihi

600°C untuk menyalut HA ke atas Co-Cr berasaskan aloi menyebabakan keretakan.

Selain itu, dari segi kekasaran permukaan semua sampel tersalut dan tersinter kecuali

yang disinter pada 500°C selama 20 min, menunjukkan hasil yang baik, akhir sekali

dari segi ketahanan kakisan sampel disinter pada 600°C selama 20 min menunjukkan

kadar kakisan hampir 3.5 kali lebih kecil daripada sampel yang tidak bersalut.

Page 6: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

vii

TABLE OF CONTENTS

CHAPTER TITLE PAGE

DECLARATION ii

DEDICATION iii

ACKNOWLEDGEMENTS iv

ABSTRACT v

ABSTRAK vi

TABLE OF CONTENTS vii

LIST OF TABLES x

LIST OF FIGURES xi

LIST OF ABBREVIATIONS xiv

LIST OF SYMBOLS xvi

LIST OF APPENDICES xviii

1 INTRODUCTION 1

1.1 Introduction 1

1.2 Problem Statement 4

1.3 Objectives 5

1.4 Scopes of Study 5

2 LITERATURE REVIEW 7

2.1 Introduction 7

Page 7: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

viii

2.2 Classifications of Biomaterials 8

2.2.1 Metallic Biomaterials 8

2.2.2 Titanium Alloys 11

2.2.3 Stainless Steel 13

2.2.4 Cobalt Chromium Alloys 14

2.2.5 Metallic Biomaterials in Human Body 18

2.2.6 Observations on Metal Ion Release 19

2.3 Effect of Corrosion on Metallic Implants 22

2.3.1 Corrosion Behavior of Titanium Based Alloys 24

2.3.2 Corrosion Behavior of Stainless Steel 24

2.3.3 Corrosion Behavior of Cobalt Chromium Based

Alloys 25

2.4 Surface Modification of Metallic Biomaterials 26

2.4.1 Surface Roughness 26

2.5 Hydroxyapatite 28

2.5.1 Mechanical Properties of Hydroxyapatite 29

2.5.2 Structure of Hydroxyapatite 30

2.6 Sol-gel Coating 31

2.6.1 Sol-gel Preparation 33

2.6.2 Dip Coating 36

2.6.3 Spin Coating 38

2.7 Effect of Heat Treatment on Hydroxyapatite 38

2.8 Summary of Literatures 40

3 RESEARCH METHODOLOGY 43

3.1 Introduction 43

3.2 Sample Preparation 44

3.3 Sol-gel Preparation of Hydroxyapatite (HA) 45

3.4 Coating and Heat Treatment 47

3.5 Testing and Analysis Characterization 48

3.6 Electrochemical Corrosion Test 49

3.7 Total Number of Samples 51

3.8 Experimental Equipment 52

Page 8: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

ix

3.8.1 Field Emission Scanning Electron Microscopy

(FESEM) 53

3.8.2 X-Ray Diffraction (XRD) Analysis Machine 53

3.8.3 Atomic Force Microscope (AFM) 54

3.8.4 Muffle Furnace 55

3.8.5 Magnetic Stirrer 56

3.8.6 Centrifuge 56

3.8.7 Viscometer 57

3.8.8 Precision Cutter 58

3.8.9 Ultrasonic Bath 58

3.8.10 Gold Coating Machine 59

4 RESULTS AND DISSCUSION 60

4.1 Introduction 60

4.2 Surface Roughness Test on Pre-coating Samples 61

4.3 Initial FESEM Test 63

4.4 FESEM Test after Heat Treatment 70

4.5 XRD Test 75

4.6 Electrochemical Corrosion Test 79

4.7 AFM Surface Roughness Test 81

5 CONCLUSIONS AND RECOMMENDATIONS 87

5.1 Conclusions 87

5.2 Recommendations for Future Works 88

REFERENCES 90

Page 9: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

x

LIST OF TABLES

TABLE NO. TITLE PAGE

2.1 Mechanical properties of metallic biomaterials 10

2.2 Metals used for orthopedic implant applications 11

2.3 ASTM standards of pure Ti and Ti alloys 12

2.4 Compositions (in wt%) of austenitic stainless steels

used for orthopedic implant fabrication 14

2.5 Chemical composition for HC and LC CoCrMo 16

2.6 ASTM standards of cobalt-chromium alloys 16

2.7 A typical chemical composition of normal human

blood plasma 18

2.8 Properties of HA Sol-gel powders heat treated

at different temperatures 39

2.9 Highlights of literatures 40

3.1 Optimum value of parameters in HA sol-gel preparation 46

3.2 Electrochemical corrosion test parameters 51

3.3 Total number of utilized samples 52

4.1 Result of AFM surface roughness (Ra) test on

pre-coating samples 61

4.2 Phases present in coated layer of samples after heat treatment 77

4.3 Electrochemical corrosion test results 81

4.4 AFM surface roughness (Ra) result after heat treatment 86

Page 10: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xi

LIST OF FIGURES

FIGURE NO. TITLE PAGE

2.1 Schematic illustration of a passive surface 23

2.2 A simple dip coating procedure 37

2.3 XRD pattern of the HA sol-gel films on Ti heat treated at

different temperatures for 1 hr in air. (●) HA; (º) Ti. 39

3.1 General flowchart of the research methodology 44

3.2 Schematic illustration of specimen 45

3.3 General flowchart of sample preparation 45

3.4 Sol-gel preparation 47

3.5 General flowchart of sol-gel preparation 47

3.6 General flowchart of coating and heat treatment procedure 48

3.7 Electrochemical corrosion test machine 50

3.8 Electrochemical corrosion test process 50

3.9 FESEM equipment 53

3.10 XRD machine 54

3.11 AFM machine 55

3.12 Muffle vacuum furnace 55

3.13 Magnetic stirrer 56

3.14 Centrifuge 57

3.15 Viscometer 57

3.16 Precision cutter 58

3.17 Ultrasonic bath 59

3.18 Gold coating machine 59

Page 11: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xii

4.1 Topographic view of the surface for a) bare sample 1,

b) bare sample 2 61

4.2 Complete results of AFM surface roughness test on

a) bare sample 1, b) bare sample 2 62

4.3 Sample 1, centrifuging speed 1500 rpm, magnification

a) 500X, b) 1000X 64

4.4 Sample 2, centrifuging speed 1500 rpm, magnification

a) 500X, b) 1000X 65

4.5 Sample 3, centrifuging speed 1750 rpm, magnification

a) 500X, b) 1000X 66

4.6 Sample 4, centrifuging speed 1750 rpm, magnification

a) 500X, b) 1000X 67

4.7 Sample 5, centrifuging speed 2000 rpm, magnification

a) 500X, b) 1000X 68

4.8 Sample 6, centrifuging speed 2000 rpm, magnification

a) 500X, b) 1000X 69

4.9 Sintered sample at a) 500°C, 20 min b) 500°C, 1 hr 71

4.10 Sintered sample at a) 600°C, 20 min b) 600°C, 1 hr 72

4.11 Sintered sample at a) 700°C, 20 min b) 700°C, 1 hr 73

4.12 EDX analysis of sintered sample at 500°C for 1 hr 74

4.13 EDX analysis of sintered sample at 600°C for 20 min 75

4.14 XRD pattern of commercial HA powder in 100°C 76

4.15 XRD pattern of HA coated layer on samples after

Heat treatment at: a) 500°C, 20 min b) 500°C, 1 hr

c) 600°C, 20 min d) 600°C, 1 hr e) 700°C, 20 min

f) 700°C, 1 hr 78

4.16 Potential Vs current diagram of bare sample 79

4.17 Potential Vs current diagram of sample sintered

at 500°C for 1 hr 80

4.18 Potential Vs current diagram of sample sintered

at 600°C for 20 min 80

4.19 Topographic view of the sample’s surface sintered

at a) 700°C, 20 min b) 700°C, 1 hr c) 600°C, 20 min

d) 600°C, 1 hr e) 500°C, 20 min f) 500°C, 1 hr 82

Page 12: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xiii

4.20 Complete results of AFM surface roughness test on

sintered sample at a) 500°C, 20 min b) 500°C, 1 hr

c) 600°C, 20 min d) 600°C, 1 hr e) 700°C, 20 min

f) 700°C, 1 hr 83

Page 13: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xiv

LIST OF ABBREVIATIONS

AFM Atomic Force Microscopy

AISI American Iron and Steel Institute

ASTM American Society for Testing and Materials

CF Carbon Fiber

CP Centipoise

EDX/ EDS Energy Dispersive X-Ray Spectrometer

FCC Face-Centered Cubic

FESEM Field Emission Scanning Electron Microscopy

g/ml grams per milliliter

GPa Giga Pascal

HA Hydroxyapatite

HC High Carbon

HCP Hexagonal Close-Packed

HF Hydrofluoric acid

HV Vickers Hardness

J/m2 Joule per square meters

LC Low Carbon

ml milliliter

mmol l-1

millimole per litter

mmpy millimeter per year

MPa m1/2

Mega Pascal per square meters

mv/s millivolt per second

rpm revolutions per minute

SP.rr Setpoint ramp rate

TCP Tricalcium Phosphate

Page 14: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xv

TTCP Tetracalcium Phosphate

V Volt

XRD X-Ray Diffraction Analysis

µA micro ampere

Page 15: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xvi

LIST OF SYMBOLS

As Arsenic

Br Bromine

C Carbon

Ca Calcium

Cl Chlorine

Co Cobalt

Cr Chromium

F Fluorine

Fe Iron

Ge Germanium

K Potassium

Klc Plane strain fracture toughness

Mg Magnesium

Mo Molybdenum

Na Sodium

Nb Niobium

Ni Nickel

O Oxygen

P Phosphorus

S Sulfur

Si Silicon

Ta Tantalum

Ti Titanium

V Vanadium

W Tungsten

Page 16: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xvii

Zr Zirconium

μ Dynamic viscosity

Page 17: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

xviii

LIST OF APPENDICES

APPENDIX TITLE PAGE

A FESEM Results after Heat Treatment in 5000X and

10000X Magnification 98

B XRD Patterns of HA Coated Layer after Heat Treatment 105

Page 18: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

1

CHAPTER 1

INTRODUCTION

1.1 Introduction

Biomaterial refers to any substance (other than a drug) or combination of

substances, synthetic or natural in origin, which can be used for any period of time,

as a whole or as a part of a system which treats, augments, or replaces any tissue,

organ, or function of the body [1]. Performance of biomaterials is controlled by two

characteristics of biofunctionality and biocompatibility. Biofunctionality defines the

ability of the device to perform the required function and refers to mechanical

properties of the biomaterial, whereas biocompatibility determines the compatibility

of the material with the body [2].

A wide range of materials encompassing all the classical materials such as

Metals (gold, tantalum, Ti6-Al4-V, 316L stainless steel, Co-Cr Alloys, titanium

alloys), Ceramics (alumina, zirconia, carbon, titania, bioglass, hydroxyapatite(HA)),

Composite (Silica/SR, CF/UHMWPE, CF/PTFE, HA/PE, CF/epoxy, CF/PEEK,

CF/C, Al2O3/PTFE), Polymers, Ultra high molecular weight polyethylene

(UHMWPE), Polyethylene (PE), Polyurethane (PU), Polytetrafuoroethylene (PTFE),

Polyacetal (PA), Polymethylmethacrylate (PMMA), Polyethylene Terepthalate

Page 19: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

2

(PET), Silicone Rubber (SR), Polyetheretherketone (PEEK), Poly lactic acid (PLA),

and Polysulfone (PS) have been investigated as biomaterials.

Researchers also classified materials into several types such as bioinert,

bioactive, biostable, biodegradable and etc. In broad terms, inert (more strictly,

nearly inert) materials prohibit or minimize tissue response. Active materials

encourage bonding to surrounding tissue. Degradable or resorbable materials are

incorporated into the surrounding tissue, or may even dissolve completely over a

period of time. Metals are typically inert, ceramics may be inert, active or resorbable

and polymers may be inert or resorbable [3]. Biomaterials must be nontoxic,

noncarcinogenic, chemically inert, stable, and mechanically strong enough to

withstand the repeated forces of a lifetime.

The physical properties of the materials, their potential to corrode in the

tissue environment, their surface configuration, tissue induction and their potential

for eliciting inflammation or rejection response are all important factors on this area.

The biomaterial discipline has evolved significantly over the past decades. The goal

of biomaterial researches has been continued to develop implant materials that

induce predictable, control-guided and rapid healing of the interfacial tissues both

hard and soft [4]. Very important requirement for any material used in the human

body is biocompatibility which is defined as the ‘‘ability of a material to perform

with an appropriate host response in a specific application”, because it should not

cause any adverse reaction in the body [5].

Mostly metallic biomaterials used as orthopedic prostheses in biomedical

applications. Metallic biomaterials used in bone plate are neither bioactive nor

biodegradable. However, they are the most common biomaterials for manufacturing

medical devices such as hip joints, bone plates and dental implants because they have

good mechanical properties such as Modulus of elasticity, Tensile strength,

Compressive strength, Elongation Metallurgical properties, low cost and also they

are easy to fabrication. Indeed, among metallic biomaterials, stainless steel, cobalt

alloys and titanium alloys have the most applications in orthopedic issues [6].

Page 20: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

3

Among above-mentioned metallic biomaterials CoCrMo alloys are

biocompatible materials and are widely used as orthopedic implant materials in

clinical practice such as hip joint and knee replacement due to their superior

mechanical properties, good wear- and corrosion-resistances. The biocompatibility of

CoCrMo alloys are closely related to their good corrosion resistance due to the

presence of an extremely thin passive oxide film that spontaneously forms on the

alloy surface. XPS analysis reveals that its composition is predominantly Cr2O3 with

some minor contribution from Co and Mo oxides. These films also form on the

surfaces of other metallic biomaterials (stainless steels, titanium and its alloys) and

serve as a barrier to corrosion processes in alloy systems [7].

In spite of the good corrosion resistance of CoCrMo alloys, there is still a

concern about metal ion release from orthopedic implants into the body fluids

(serum, urine, etc.). Metals from orthopedic implant materials are released into

surrounding tissue by various processes, including corrosion, wear and mechanically

accelerated processes such as stress corrosion, corrosion fatigue and fretting

corrosion. Such metal ions and wear debris, concentrated at the implant-tissue

interface, may migrate through the tissue. Research shows that the metal release is

associated with clinical implant failure, osteolysis, cutaneous allergic reactions and

remote site accumulations [8]. One effective approach for preventing and/or reducing

the potentially harmful metal ion release from orthopedic implant materials is coating

the surfaces of these materials.

A bioactive surface coating is capable to support bonding to surrounding

bone. One of the best bioactive compounds which is suitable for coating metallic

biomaterial implants is “Hydroxyapatite”. Hydroxyapatite (HA, Ca10(PO4 )6(OH)2) is

the main structural component of natural bone, and used as an important material for

bone and tooth implants in the biomaterial field. In order to achieve bioactivity for

metal implants (e.g. Co alloys, Ti alloys or stainless steel) as bone substitutes, HA

coating is usually introduced onto their surfaces. Porous HA coating on these metal

substrates can be adopted as bone cements in reconstruction. HA has many biological

profits such as direct bonding to bone and enhancement of new bone formation

around it due to its chemical similarity with hard tissues. HA as a coating also can

reduce the amount of ion release from the metallic substrate. Because HA has poor

Page 21: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

4

mechanical properties and it is weak and brittle without any support, so it is applied

as a coating on an inert metal with good bio-mechanical properties such as CoCrMo

[9].

To date many essential techniques have been used in the preparation of HA

coatings such as plasma spraying, magnetron sputtering, laser ablation, sol-gel,

biomimetic, and electrochemical deposition [10]. Compared to other coating

techniques, the sol-gel technique is one of the thin film methods provides some

benefits over the others such as chemical homogeneity, fine grain structure, and low

processing temperature. Moreover, compared to the other thin film methods, it is

simple and cost efficient, as well as effective for the coating of complex-shaped

implants.

1.2 Problem Statement

Massive cracks of HA sol-gel coated layer on implants are still a major issue

due to releasing harmful ions from the body of implant which can result in adverse

biological reactions in human body. Reduction of cracks can be done by controlling

several parameters such as finding the most appropriate viscosity of sol-gel regarding

to examine different range of centrifuging speed in the procedure of sol-gel

preparation, obtaining the most suitable proportion of sol-gel precursors and finally

applying different range of sintering time and temperature to get the best heat

treatment procedure.

For Co-Cr based alloy as a metallic biomaterial there are some disadvantages

that result in some restrictions in their usage in biomedical applications such as its

corrosion behavior in vivo which made concerns about metal ion release in human

body and its biocompatibility and cell growth on its surface which has not been

reported as good as titanium alloys. Nonetheless, these problems would be solved by

Page 22: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

5

coating implants with biocompatible and corrosion resistant material like

Hydroxyapatite (HA).

Furthermore, there are very limited extensive studies investigating the effect

of coating method and heat treatments on Co-Cr based substrates as compared to

Titanium alloy substrates

1.3 Objectives

Based on problem statement, the main aims of this study are:

1. To determine the feasible parameters for a crack free HA coating on Co-

Cr based substrate.

2. To analyze the surface morphology of the HA coated layer on Co-Cr

based substrate under different coating conditions.

3. To compare the corrosion behavior of Co-Cr based substrate before and

after HA sol-gel coating.

1.4 Scopes of Study

The scopes of this project are narrowed as follow:

1. HC CoCrMo based alloy is used as the substrate material.

2. Sol-gel method is employed for coating HA on the substrate.

3. Na3PO4 and CaCl2 are used as the main precursors of sol-gel preparation.

Page 23: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

6

4. Centrifuging speed of sol-gel solution is varied in three levels (1500, 1750

and 2000 rpm).

5. Sintering temperature and soaking time of HA coated samples are 500ºC,

600ºC and 700ºC at 20 minutes and 1 hour, respectively.

Page 24: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

90

REFERENCES

1. Dee, K.C., D.A. Puleo, and R. Bizios, An introduction to tissue-biomaterial

interactions. 2003: Wiley. com.

2. Malekani, J., et al. Biomaterials in orthopedic bone plates: a review. in

Proceedings of the 2nd Annual International Conference on Materials

Science, Metal & Manufacturing (M3 2011). 2011. Global Science and

Technology Forum.

3. www.azom.com Biomaterials: An Overview.

4. Muddugangadhar, B., et al., Biomaterials for dental implants: An overview.

International Journal of Oral Implantology and Clinical Research, 2011. 2(1):

p. 13-24.

5. Gurappa, I., Characterization of different materials for corrosion resistance

under simulated body fluid conditions. Materials Characterization, 2002.

49(1): p. 73-79.

6. Temenoff, J.S. and A.G. Mikos, Biomaterials: the intersection of biology and

materials science2008: Pearson/Prentice Hall.

7. Contu, F., B. Elsener, and H. Böhni, Corrosion behaviour of CoCrMo

implant alloy during fretting in bovine serum. Corrosion science, 2005. 47(8):

p. 1863-1875.

8. Okazaki, Y. and E. Gotoh, Comparison of metal release from various

metallic biomaterials in vitro. Biomaterials, 2005. 26(1): p. 11-21.

9. Ignjatovic, N. and D. Uskokovic, Synthesis and application of

hydroxyapatite/polylactide composite biomaterial. Applied surface science,

2004. 238(1): p. 314-319.

Page 25: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

91

10. Thian, E., et al., Magnetron co-sputtered silicon-containing hydroxyapatite

thin films—an in vitro study. Biomaterials, 2005. 26(16): p. 2947-2956.

11. Helmus, M.N., D.F. Gibbons, and D. Cebon, Biocompatibility: meeting a key

functional requirement of next-generation medical devices. Toxicologic

pathology, 2008. 36(1): p. 70-80.

12. Narayan, R., Biomedical materials. 2009: Springer.

13. Park, J.B. and J.D. Bronzino, Biomaterials: principles and applications.

2002: crc press.

14. NakaNo, T., Mechanical properties of metallic biomaterials. Metals for

Biomedical Devices, 2010: p. 71.

15. Kokubo, T., H.-M. Kim, and M. Kawashita, Novel bioactive materials with

different mechanical properties. Biomaterials, 2003. 24(13): p. 2161-2175.

16. Rack, H. and J. Qazi, Titanium alloys for biomedical applications. Materials

Science and Engineering: C, 2006. 26(8): p. 1269-1277.

17. Rodriguez-Gonzalez, F.A., Biomaterials in orthopaedic surgery. Corrosion

Engineering, Science and Technology, 2010. 45(2).

18. Ratner, B., et al., Biomaterials science: an introduction to materials in

medicine. San Diego, California, 2004: p. 162-164.

19. Rahaman, M.N., et al., Ceramics for prosthetic hip and knee joint

replacement. Journal of the American Ceramic Society, 2007. 90(7): p. 1965-

1988.

20. Marti, A., Cobalt-base alloys used in bone surgery. Injury, 2000. 31: p. D18-

D21.

21. Fehring, T.K., J.H. Chaffin III, and R.L. Kennedy, Enhanced biocompatible

implants and alloys. 2001, Google Patents.

22. Herrera, M., et al., Effect of C content on the mechanical properties of

solution treated as-cast ASTM F-75 alloys. Journal of Materials Science:

Materials in Medicine, 2005. 16(7): p. 607-611.

23. Niinomi, M., Recent metallic materials for biomedical applications.

Metallurgical and materials transactions A, 2002. 33(3): p. 477-486.

Page 26: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

92

24. Patel, B., et al., Cobalt-based orthopaedic alloys: Relationship between

forming route, microstructure and tribological performance. Materials

Science and Engineering: C, 2012. 32(5): p. 1222-1229.

25. Hosoya, K., et al., A novel covalently crosslinked gel of alginate and silane

with the ability to form bone‐like apatite. Journal of Biomedical Materials

Research Part A, 2004. 71(4): p. 596-601.

26. Yan, Y., et al., Biotribocorrosion of CoCrMo orthopaedic implant materials -

Assessing the formation and effect of the biofilm. Tribology international,

2007. 40(10-12): p. 1728-1728.

27. Souza, J., et al., Do oral biofilms influence the wear and corrosion behavior

of titanium? Biofouling, 2010. 26(4): p. 471-478.

28. Lewis, A., et al., Effect of synovial fluid, phosphate‐buffered saline solution,

and water on the dissolution and corrosion properties of CoCrMo alloys as

used in orthopedic implants. Journal of Biomedical Materials Research Part

A, 2005. 73(4): p. 456-467.

29. Sargeant, A. and T. Goswami, Hip implants - Paper VI - Ion concentrations.

Materials & Design, 2007. 28(1): p. 155-171.

30. Okazaki, Y., et al., Comparison of metal concentrations in rat tibia tissues

with various metallic implants. Biomaterials, 2004. 25(28): p. 5913-5920.

31. Davies, A.P., et al., Metal-specific differences in levels of DNA damage

caused by synovial fluid recovered at revision arthroplasty. Journal of Bone

and Joint Surgery-British Volume, 2005. 87B(10): p. 1439-1444.

32. Jacobs, J.J., et al., Can metal levels be used to monitor metal-on-metal hip

arthroplasties? Journal of Arthroplasty, 2004. 19(8): p. 59-65.

33. MacDonald, S.J., Can a safe level for metal ions in patients with metal-on-

metal total hip arthroplasties be determined? The Journal of Arthroplasty,

2004. 19(8): p. 71-77.

34. Burstein, G.T., C. Liu, and R.M. Souto, The effect of temperature on the

nucleation of corrosion pits on titanium in Ringer's physiological solution.

Biomaterials, 2005. 26(3): p. 245-256.

Page 27: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

93

35. Keegan, G., I. Learmonth, and C. Case, Orthopaedic metals and their

potential toxicity in the arthroplasty patient A REVIEW OF CURRENT

KNOWLEDGE AND FUTURE STRATEGIES. Journal of Bone & Joint

Surgery, British Volume, 2007. 89(5): p. 567-573.

36. Revie, R.W., Corrosion and corrosion control. 2008: John Wiley & Sons.

37. Milosev, I., The effect of biomolecules on the behaviour of CoCrMo alloy in

various simulated physiological solutions. Electrochimica Acta, 2012. 78: p.

259-273.

38. Fathi, M.H., et al., In vitro corrosion behavior of bioceramic, metallic, and

bioceramic–metallic coated stainless steel dental implants. Dental Materials,

2003. 19(3): p. 188-198.

39. Moreno, J.M.C., et al., Surface analysis and electrochemical behavior of Ti-

20Zr alloy in simulated physiological fluids. Materials Science and

Engineering B-Advanced Functional Solid-State Materials, 2013. 178(18): p.

1195-1204.

40. Lewis, G., Properties of open-cell porous metals and alloys for orthopaedic

applications. Journal of Materials Science-Materials in Medicine, 2013.

24(10): p. 2293-2325.

41. Swiontkowski, M.F., et al., Cutaneous metal sensitivity in patients with

orthopaedic injuries. Journal of Orthopaedic Trauma, 2001. 15(2): p. 86-89.

42. Li, M., et al., Corrosion behavior of TiN coated type 316 stainless steel in

simulated PEMFC environments. Corrosion science, 2004. 46(6): p. 1369-

1380.

43. Della Valle, A.G., et al., Late fatigue fracture of a modern cemented forged

cobalt chrome stem for total hip arthroplasty: a report of 10 cases. The

Journal of Arthroplasty, 2005. 20(8): p. 1084-1088.

44. Reclaru, L., et al., Corrosion behaviour of cobalt–chromium dental alloys

doped with precious metals. Biomaterials, 2005. 26(21): p. 4358-4365.

45. Liu, X., P.K. Chu, and C. Ding, Surface modification of titanium, titanium

alloys, and related materials for biomedical applications. Materials Science

and Engineering: R: Reports, 2004. 47(3): p. 49-121.

Page 28: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

94

46. Boyan, B.D., et al., Role of material surfaces in regulating bone and

cartilage cell response. Biomaterials, 1996. 17(2): p. 137-146.

47. Łaczka‐Osyczka, A., et al., Behavior of bone marrow cells cultured on three

different coatings of gel‐derived bioactive glass–ceramics at early stages of

cell differentiation. Journal of biomedical materials research, 1998. 42(3): p.

433-442.

48. Hayashi, K., et al., Effect of surface roughness of hydroxyapatite—coated

titanium on the bone-implant interface shear strength. Biomaterials, 1994.

15(14): p. 1187-1191.

49. Deligianni, D.D., et al., Effect of surface roughness of hydroxyapatite on

human bone marrow cell adhesion, proliferation, differentiation and

detachment strength. Biomaterials, 2000. 22(1): p. 87-96.

50. Paital, S.R. and N.B. Dahotre, Calcium phosphate coatings for bio-implant

applications: Materials, performance factors, and methodologies. Materials

Science and Engineering: R: Reports, 2009. 66(1): p. 1-70.

51. Méndez-Vilas, A., J. Bruque, and M. González-Martín, Sensitivity of surface

roughness parameters to changes in the density of scanning points in multi-

scale AFM studies. Application to a biomaterial surface. Ultramicroscopy,

2007. 107(8): p. 617-625.

52. Jandt, K.D., Atomic force microscopy of biomaterials surfaces and interfaces.

Surface Science, 2001. 491(3): p. 303-332.

53. Corbridge, D.E.C., Phosphorus-an outline of its chemistry, biochemistry and

technology. 1985: Elsevier Science Publishers BV.

54. Hench, L.L. and J. Wilson, An introduction to bioceramics. Vol. 1. 1993:

World Scientific.

55. Wolke, J., et al., In vivo dissolution behavior of various RF magnetron-

sputtered Ca-P coatings on roughened titanium implants. Biomaterials, 2003.

24(15): p. 2623-2629.

56. Stephenson, P., et al., The effect of hydroxyapatite coating on ingrowth of

bone into cavities in an implant. The journal of arthroplasty, 1991. 6(1): p.

51-58.

Page 29: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

95

57. Thangamani, N., K. Chinnakali, and F. Gnanam, The effect of powder

processing on densification, microstructure and mechanical properties of

hydroxyapatite. Ceramics international, 2002. 28(4): p. 355-362.

58. Rezwan, K., et al., Biodegradable and bioactive porous polymer/inorganic

composite scaffolds for bone tissue engineering. Biomaterials, 2006. 27(18):

p. 3413-3431.

59. Wang, C., et al., Phenotypic expression of bone-related genes in osteoblasts

grown on calcium phosphate ceramics with different phase compositions.

Biomaterials, 2004. 25(13): p. 2507-2514.

60. Zhang, S., et al., Adhesion strength of sol–gel derived fluoridated

hydroxyapatite coatings. Surface and Coatings Technology, 2006. 200(22): p.

6350-6354.

61. Wen, C., et al., Hydroxyapatite/titania sol–gel coatings on titanium–

zirconium alloy for biomedical applications. Acta Biomaterialia, 2007. 3(3):

p. 403-410.

62. Heimann, R.B., Plasma-spray coating. 2008: Wiley. com.

63. Kim, H.-W., et al., Hydroxyapatite coating on titanium substrate with titania

buffer layer processed by sol–gel method. Biomaterials, 2004. 25(13): p.

2533-2538.

64. Liu, D.-M., T. Troczynski, and W.J. Tseng, Water-based sol–gel synthesis of

hydroxyapatite: process development. Biomaterials, 2001. 22(13): p. 1721-

1730.

65. Overgaard, S., et al., The influence of crystallinity of the hydroxyapatite

coating on the fixation of implants MECHANICAL AND

HISTOMORPHOMETRIC RESULTS. Journal of Bone & Joint Surgery,

British Volume, 1999. 81(4): p. 725-731.

66. Milella, E., et al., Preparation and characterisation of titania/hydroxyapatite

composite coatings obtained by sol–gel process. Biomaterials, 2001. 22(11):

p. 1425-1431.

Page 30: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

96

67. Fathi, M. and A. Hanifi, Evaluation and characterization of nanostructure

hydroxyapatite powder prepared by simple sol–gel method. Materials letters,

2007. 61(18): p. 3978-3983.

68. Raynaud, S., et al., Calcium phosphate apatites with variable Ca/P atomic

ratio I. Synthesis, characterisation and thermal stability of powders.

Biomaterials, 2002. 23(4): p. 1065-1072.

69. Gross, K., et al., Thin hydroxyapatite coatings via sol–gel synthesis. Journal

of Materials Science: Materials in Medicine, 1998. 9(12): p. 839-843.

70. LeGeros, R., et al., Biphasic calcium phosphate bioceramics: preparation,

properties and applications. Journal of materials science: Materials in

Medicine, 2003. 14(3): p. 201-209.

71. odrıguez-Lorenzo, L., M. Vallet- egı, and J. Ferreira, Fabrication of

hydroxyapatite bodies by uniaxial pressing from a precipitated powder.

Biomaterials, 2001. 22(6): p. 583-588.

72. Liu, J., et al., The influence of pH and temperature on the morphology of

hydroxyapatite synthesized by hydrothermal method. Ceramics international,

2003. 29(6): p. 629-633.

73. Yang, L., et al., Description of surface roughness of sol–gel films/coatings by

X-ray reflectivity technique. Applied Surface Science, 2009. 255(19): p. 8226-

8229.

74. Sonawane, R., S. Hegde, and M. Dongare, Preparation of titanium (IV) oxide

thin film photocatalyst by sol–gel dip coating. Materials Chemistry and

Physics, 2003. 77(3): p. 744-750.

75. Gutoff, E.B. and E.D. Cohen, Coating and drying defects: troubleshooting

operating problems. 2006: Wiley-Interscience.

76. Kim, H.W., et al., Sol‐gel‐modified titanium with hydroxyapatite thin films

and effect on osteoblast‐like cell responses. Journal of Biomedical Materials

Research Part A, 2005. 74(3): p. 294-305.

77. Grandfield, K., et al., Free form fabricated features on CoCr implants with

and without hydroxyapatite coating in vivo: a comparative study of bone

Page 31: EVALUATING THE SURFACE PROPERTIES OF HA ...eprints.utm.my/id/eprint/41833/1/...2.3.1 Corrosion Behavior of Titanium Based Alloys 24 2.3.2 Corrosion Behavior of Stainless Steel 24 2.3.3

97

contact and bone growth induction. Journal of Materials Science: Materials

in Medicine, 2011. 22(4): p. 899-906.

78. Harle, J., et al., Initial responses of human osteoblasts to sol–gel modified

titanium with hydroxyapatite and titania composition. Acta Biomaterialia,

2006. 2(5): p. 547-556.

79. Yan, Y., A. Neville, and D. Dowson, Biotribocorrosion of CoCrMo

orthopaedic implant materials—assessing the formation and effect of the

biofilm. Tribology international, 2007. 40(10): p. 1492-1499.

80. Vidal, C.V. and A.I. Muñoz, Effect of thermal treatment and applied

potential on the electrochemical behaviour of CoCrMo biomedical alloy.

Electrochimica Acta, 2009. 54(6): p. 1798-1809.