EvaluationoftheBodyFluidCycleoftheYumizen®H2500Analyzer(HoribaMedicalSAS)AnaïsInquel1,MartinBroly1,MyriamChevalier1,KhadijaViau1,MarieCBéné1,EliseTomini2,CamilleDebord1,EliseDarrambide2,MarionEveillard11.Serviced’HématologieBiologique,CHUdeNantes,France2.SociétéHoribaMedicalSAS,Montpellier,France
INTRODUCTION
CONCLUSION
RESULTS
METHODSNinety-eightsamplesincludingcerebro-spinal(6),pleural(18),ascitic(5),synovial(54),pericardic(2)andbronchoalveolarliquid(BAL,13)werecollectedover10weeksinaprospectivestudyatNantesUniversityHospital (figure 1). The samples were analyzed in duplicate by the Yumizen®H2500 to determine thenumber of leukocytes and erythrocytes, the partial formula (% of mononuclear cells (%MN) and % ofpolymorphonuclearcells(%PN))andtodetectabnormalcells.ResultswerecomparedtoaconventionalmicroscopiccountandtotheXN10(Sysmex®)referenceanalyzerinbodyfluidmode.Both instrumentswerecalibratedpriortothesamplesanalysisusingexternalcalibrators.Passing-Bablokregressionwasusedtoanalyzeresults.
Seventy-fouroutof98sampleswereusedtocompareleukocyteanderythrocytecount.ElevensampleswereexcludedbecausetheywereoutsidetherangeoflinearityoftheanalyzersandtheBALwereexcludedduetoviscosityrelatedflags.Thecorrelationcoefficientbetweenthe2analyzerswas0.991for leukocyte (figure2)and0.955betweenYumizen®H2500andtheconventionalchambercountingmethod(figure3).Discrepantresultswereobservedinlipemicsamplesandsamplescontainingcrystals.Forerythrocytecounts,thecorrelationcoefficientwas0.996betweenbothanalyzers.
Seventy-onesampleswereretained.Comparisonoftheanalyzersforthe%ofPMNyieldedaslopeof0.931,interceptof+0.18andacorrelationcoefficientof0.933(figure4A).Thecoefficientwas0.914betweenthemanualmodeandYumizen®H2500.Thepresenceofabnormalplasmacellsinthesamplecauseddiscrepantresultsbetweenbothanalyzers(figure4B,figure6)andbetweentheYumizen®H2500andmanualcount.The%MNcorrelationcoefficientbetweenthetwoanalyzers,andbetweentheYumizen®H2500andthemanualmodewasrespectively0.933(figure5)and0.917.
LeukocyteandErythrocytecount Differentialcount
ThisstudyconfirmedthegoodanalyticalperformanceoftheYumizen®H2500incomparisonwithconventionalmicroscopiccountaswellaswiththeXN10analyzer.Inaddition,datafromthisstudyallowedthescattergramthresholdsusedfortheleukocytecountandthedifferentialtobeoptimized
Automatedanalysisofbiologicalfluidsisnowrequiredtoensuretherobustnessandreliabilityofresultsreporting.Inthisstudy,weevaluatedtheperformanceofthebodyfluidcycleoftheYumizen®H2500.
Figure2:LeukocyteCorrelationbetweenYumizen®H2500andSysmexXN10®analyzer Figure3:LeukocyteCorrelationbetweenYumizen®H2500andConventionalManualCount
Figure1:Distributionofsamplestype
Figure4:A:CorrelationofPMN%betweenYumizen®H2500andSysmexXN10®analyzer,B:BiasonPMNfromYumizen®H2500andSysmexXN10®analyzer Figure5:CorrelationofMN%betweenYumizen®H2500andSysmexXN10®analyzer
A B
Figure6:ExamplesofscattergrammofYumizen®H2500forBFcontainingatypicalcells;A:Pleuraleffusionwith24%ofplasmacells,B:Asciticwith13%ofmesothelialcells.
A
B
0.0
2000.0
4000.0
6000.0
0.0 2,000.0 4,000.0 6,000.0
S105-Dx
XN-9000
WBCBF/TC-BF
y=0,92x+15,39
r² = 0,991
0
2000
4000
6000
8000
10000
12000
0 2,000 4,000 6,000 8,000 10,000 12,000
BFW
BC
WBc Kov a
Correlation BFWBC / Kova
y=0,812x+16,34
r² = 0,955
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
100.0
0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 100.0
BF
PN
%
PN% (XN-10)
Correlation BFPN% vs PN% (XN-10)
y=0,931x+0,18
r² = 0,933
Plasmacel ls
Plasmacel ls
-45
-35
-25
-15
-5
5
15
25
0 200 400 600 800 1000
BiaisPN%=f(BFWBC)vs.XN
0.0
10.0
20.0
30.0
40.0
50.0
60.0
70.0
80.0
90.0
100.0
0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 80.0 90.0 100.0
S105
-Dx
MN%- XN10
y=0.939x+6.64
r² = 0.933
CorrelationBFMN%vsMN%(XN-10)