Beaverstate Dental, Inc.

Evolution Series Dental Chair

Service Manual

Includes: Installation Instructions Operation Instructions Troubleshooting Illustrated Parts Breakdown (IPB's) Schematics

900-024

ENJOY RELIABILITY

Table of ContentsInstallation ServiceEvolution Series Dental Chair (110V/220V) ............................................................................... 4

WARRANTYThe Evolution Chair is guaranteed against defects in material and workmanship for a period of two (2) years after date of purchase. Faulty parts will be repaired or replaced; no claim for labor or consequential damages will be allowed. All replacement parts claimed defective will be billed in the usual manner, such invoices being subject to adjustment after the alleged defective material has been inspected by the manufacturer. This guarantee does not cover damage resulting from the use of cleaning, disinfecting, or sterilization chemicals and processes. The warranty does not cover damage resulting from improper installation or maintenance, accident, misuse, or accidents incurred in shipping and handling. All claims against the freight carrier must be initiated at the time the damaged items are received. The claim is the responsibility of the customer.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

3

Evolution Series Dental ChairCleaning and DisinfectingInfection control in the dental ofce continues to be a high priority for our customers and end users. We will not attempt to specify the required intervals for disinfection nor can we state what is the overall best surface disinfectant. Please refer to the Infection Control Recommendations published by the American Dental Association for further reference.

Barrier Technique Chemical Disinfection

Beaverstate Dental, Inc. strongly advocates the barrier technique be used whenever possible to preserve the nish and appearance of the equipment. Wherever possible disposable barriers should be used and changed between patients. The barrier technique will ensure maximum long term durability of the surfaces and nishes of equipment. Regardless of the chemical disinfectant used, it is imperative that the equipment be thoroughly washed with mild soap and warm water at least once per day. This wash down will minimize the harmful effects of chemical disinfectant residues being allowed to accumulate on the equipment. When using chemical disinfectants always pay strict attention to the disinfectant manufacturers directions. When using concentrated disinfectants measure the concentrate carefully and mix according to package directions. Disinfectant solutions that are relatively harmless to surfaces at their recommended strengths can be corrosive at higher than recommended dilution ratios.

Unacceptable Disinfectants

These disinfectants will harm the surface nishes of Beaverstate Dental equipment and are not recommended. USE OF THESE PRODUCTS WILL VOID YOUR WARRANTY. Chemical Composition Strong Phenols/Phenol-Alcohol Combinations Sodium Hypochlorite/Household Bleach Alcohol Household Cleaners Trade Names Lysol, Lysol 2, Lysol Professional Coe Foam, Coe Spray Pump, Vitaphone, Omni II Clorox, Ajax, Purex

*Conditionally Acceptable Disinfectants

These disinfectants have been found to be the least harmful to the equipment surfaces. None of these products are recommended for chair upholstery. Use the barrier technique. Chemical Composition Iodophors** Mild Phenols Glutaraldehyde/Phenol sprays Synergized Super Quat Phenol/Water Sprays Trade Names Biocide, Aseptic-IDC Wescodyne, SD5, Promedyne, Ido-Five Procide ES, Asepti-Steryl Aerosol Sterall Spray, Coldspor Dis-cide, Cavicide, Kleenaseptic Top-Cide, Sporicidin Pump Spray, Birex se

*Beaverstate Dental, Inc. makes no representation as to the disinfectant efcacy of these products. We make no warranty, expressed or implied, that these disinfectants will not damage the surface nishes. Damage and discoloration of the surface nishes is not covered under the warranty. **Iodophor based disinfectants will cause yellow staining on many surfaces. Regular washing with soap and water will minimize this staining. Iodophor neutralizers such as Promedyne are also available.

4

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

EVOLUTION SERIES - DIMENSIONS

6"

7"

44"

Seat

35" H 19.5"L

14.5"

16.5"

71" - 78"

120

35"

43"29.5"

23-30"Range Measurement When Headrest Extended

29.25"

20"

7"

26.5" 20"

30.75"

16.50"23"

IL39.eps

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

5

DESCRIPTION OF FUNCTIONS FOR THE KEYPAD CONTROL

300-165 Control Pad Green Switches Pre-Position Programs See 1 2 3

Gray Switch Auto Return Exit

Manual Back Down Switch

Manual Base Up Switch

Manual Back Up Switch

IL44.eps

Manual Base Down Switch

Green Switches:

These switches are used to place the patient in any of the ve (5) programmed operating positions.

Green Switches (P1-P5): These switches are used to set the programs in the memory. To start programming, the chair must be in the exit position (base down and back up). Manually run the chair to the rst chosen position. Push (P1) until an audible tone is heard. This sets the (P1) position in the memory. Continue to your next chosen position and program each position in the same manner. Note: After the programming of a position. The chair must be returned to the exit position before the next position can be programmed. To do this press the Auto Return Exit button.

Green Switches (P1-P5): To reprogram a position, manually move the chair to the new position and press the appropriate program button until an audible tone is heard. The new position is now in the program memory. Foot Control: When using the foot control, the green preposition button on the foot control will be automatically set to the program button (P1-5) last used. To change the pre-position on the foot control, simply push the desired program button on the chair back and this will automatically change the foot control preposition button to that program. Note: ALL SWITCHES OPERATE ON A 5-VOLT DC SYSTEM.

6

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

EVOLUTION SERIES - BACK UPHOLSTERY & MEMBRANE PAD INSTALLATION1. Remove the four (4) screws which hold the back shroud in place.

2.

Place upholstery on the chair back as shown. A. Snap the two (2) upper upholstery aps into place. Be sure that the cut-outs in these aps are positioned over the spacers (located towards the bottom of the chair back.) You may have to stretch the upholstery to accomplish this.

(A)

(A)

3.

Attach the membrane pad to the circuit board. A. You will see an arrow () marked on both the male and female connectors. Please make sure to connect the ribbon and circuit board so the arrows () on each connector correspond with each other.

4.

Replace the shroud and secure it with four (4) screws which were removed in Step 1. Arrow

Arrow

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

7

EVOLUTION SERIES - P.C. BOARD ELECTRONICS DIAGRAM (115V)F1 1/4 A P1 P3 T1 S1 S2 P2 P4 S3 BR2 IN G N D 0UT 78L05 +5V

100uf

C3

C4

.1uf

VR1

120VAC:12VAC 6A F2 GND

+5V R1 220 1

U2 6 R2 330 4 R3 MOC3063 1K R4 39 1/2 W BASE DOWN Q2

+5V R9 1

U4 6 R10 330 2 4 R11 MOC3063 1K BACK UP

.01 1KV

1 AC WHITE 2 AC BLACK

2 BASE DOWN CON1. SH1.2

BACK UP CON. SH1.3

R12 39 1/2 W

J1

3 BACK UP A.C. 4 BASE DOWN A.C. 5 BACK DOWN A.C. 6 BASE UP A.C.

+5V R5 220 2 BASE UP CON. SH1.1 1

U3 6 R6 330 4 R7 MOC3063 1K R8 39 1/2 W BASE UP Q4

+5V R13 1

U5 6 R14 330 2 4 . MOC3063 R15 1K BACK DWN

.01 1KV

BACK DOWN CON SH1.4

R16 39 1/2 W

OPTIONAL

RP2 +5V 470 X 6

+5V

1

2

3

4

5

6

D1 GROUP A SENSE S5 AUTO PREPOSITION S4 PROGRAM SET U1 S3 AUTO EXIT 1 2 3 BASE DOWN BASE UP S2 BACK DOWN BACK UP S1 4 5 6 7 8 1K X 8 2 3 JP1 TEST GND 1 4 +5V 1K

D2

D3

D4

10K

P2OR

1

6 7 8 9 10 11 12 13

J2

4 5 6 7 8

P3

V 4 RB0 D MCLR 3 RB1 D RA4 2 RB2 RA3 1 RB3 RA2 18 RB4 RA1 17 RB5 RA0 16 RB6 OSC1 V 15 RB7 S OSC2 S 5 X1

BASE UP CON. BASE DOWN CON. BACK UP CON. BACK DOWN CON.

SH2.1 SH2.2 SH2.3 SH2.4

16C84-04/P 1 2 3

10K

10K

455 KHz C1 150pf C2 150pf

J3

4 5 6 7 8 GND

8

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

.01 1KV

220

Q3

C7

C9

.01 1KV

220

.1uf

C5

VAL

Q1

C6

C8

EVOLUTION SERIES - CHAIR WIRING DIAGRAM0

BackP3

P1

P2 P4 P5

300-165 Membrane Switch

J3

300-060 Diode Board

300-001 Foot Control300-046ChassisGND

Chair J2

Foot J1

300-028

300-089 Switch Cable

Ground Power Common

Power Supply Cord

GN BLK WH

Base UP1 2 3

OR

BL YL OR BR

1 3 5 7

2 4 6 8

BLK RD WH GN

OR

See Note 3Base Motor Assy.

GN YL BL RD

4 5 6 7

See Note 1

BR OR YL BL

7 5 3 1

8 6 4 2

GN 7 8 WH 5 6 RD 3 4 BLK 1 2 Circuit Board

GN GND

8

Back Motor

BLK WH RD OR GN

Supply Cable

BLK WH RD OR GN

9 10

RD RDBase Down

Plug-InOR RD RD BLK BLK WH

See Note 2

300-045

WH BLK BLK RD RD OR

1 1 2 2 3 3 4 4 5 5 6 6 Connector 1 1 2 2 3 3 4 4 Connector

WH WH BLK BLK RD RD OR

Base

300-028

Base/Top Supply Cable

RD WH BLK GN

RD WH BLK GN

Note 1: 110V - Back Tilt Motor Assy. - P/N 300-067 220V - Back Tilt Motor Assy. - P/N 300-270 Note 2: 110V - P.C. Board - P/N 300-057 220V - P.C. Board - P/N 300-248 Note 3: 110V - Base Motor Assy. - P/N 300-047 220V - Base Motor Assy. - P/N 300-256

Top

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

9

EVOLUTION SERIES - P.C. BOARD ELECTRONICS DIAGRAM (220V)F1 1/4 A P1

T1

78L05 S1 S2 S3 BR2100uf C3 C4 .1uf

+5V

P2 P3P4

IN VAL

G N D

0UT.1uf C5

VR1

230VAC:12VAC 3A F2 GND

+5V

R1 220

1

6

R2 3302.2M R3 .01 1KV

+5V

R17 220

1

6

R18 3302.2M .01 1KVR33 .01 1KV

21 AC WHITE 2 AC BLACK

4 MOC30636

2 U6 1 MOC3063

4

R19

C6

360K

360K

U2 1

Q2 Q22.2M

62.2M R20

Q1 Q1

360K

360K

5 BACK DOWN A.C. 6 BASE UP A.C.

2 BASE DOWN CON. SH1.2

4

2 BACK UP CON. SH1.3

4

R5 U3MOC3063

R21 U7MOC3063

1K

1K

+5V

R9 220

1

6

R10 3302.2M R11 .01 1KV

+5V

R26 220

1

6

R27 3302.2M

2U4 1

4 MOC30636

2 U8 1 MOC3063

4

R28

C7

360K

360K

R14

Q4 Q42.2M

62.2M

Q3 Q3

R31

39 1/2W

2 BASE UP CON. SH1.1

4

R13 U5MOC3063

2 BACK DOWN CON. SH1.4

4

R30MOC3063

1K

U9

1K

OPTIONAL +5V RP2 +5V 470 X 6

10K

GROUP A SENSE S5 AUTOPREPOSITION S4

1

2

3

4

5

6

JP1 TEST GND +5V

1K D1 D2 D3 D4

PROGRAM SET

1U1

4

S3 AUTO EXIT

1 2 3

P2OR

1 2 3

1K X 8

6 7 8 9 10 11 12 13

BASE DOWN BASE UP S2 BACK DOWN BACK UP S1

4 5 6 7 8

J2

4 5 6 7 8

P3

V 4 RB0 D MCLR 3 RB1 D RA4 2 RB2 RA3 1 RA2 RB3 18 RA1 RB4 17 RA0 RB5 16 OSC1 RB6 V 15 RB7 S OSC2 S

BASE UP CON. BASE DOWN CON. BACK UP CON. BACK DOWN CON.

SH2.1 SH2.2 SH2.3 SH2.4

16C84-04/P110K 10K

5

X1

2 3

455 KHz C1 150pf C2 150pf

J3

4 5 6 7 8

GND

10

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

39 1/2W

R16

360K

360K

R12

R29

R15

R32

C9

R23

R7

R25

R4

R8

4 BASE DOWN A.C.

39 1/2W

39 1/2W

J1

3 BACK UP A.C.

R22

R6

C8

EVOLUTION SERIES - TOP MOTOR/BASE MOTOR TEST

See Note 1

BR OR YL BL

7 5 3 1

8 6 4 2

GN 7 8 WH 5 6 RD 3 4 BLK 1 2 Circuit Board

GN

GND

Back Motor

Plug-In

See Note 2 WH WH BLK BLK RD RD OR Base/Top Supply Cable (300-045)

OR RD RD BLK BLK WHWH BLK BLK RD RD OR

1 1 2 2 3 3 4 4 5 5 6 6 Connector A

Test Side

Wire Color Legend BLK GN WH RD OR BR Black Green White Red Orange Brown

RD WH BLK GN

1 1 2 2 3 3 4 4 Connector B

RD WH BLK GN

Note :

Separate connector A Perform all tests on (test side) of connector A Test 2 Jump 2-5 Back will lower if motor is good Test 4 Jump 2-6 Base will raise if motor and limit switch are good

Note 1: 110V - Back Tilt Motor Assy. - P/N 300-067 220V - Back Tilt Motor Assy. - P/N 300-270 Note 2: 110V - P.C. Board - P/N 300-057 220V - P.C. Board - P/N 300-248

Test 1 Jump 2-3 Back will raise if motor is good. Test 3 Jump 2-4 Base will lower if motor and limit switch are good

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

11

EVOLUTION SERIES - BASE MOTOR/LIMIT SWITCH TEST

Chassis300-046

Ground Power Common

GNDGN BLK WH Base UP1 2 3 4 5 6 7 8 9 10

300-028

Power Supply Cord

OROR

See Note 1

Base Motor Assy.

GN YL BL RD

BLK WH RD OR GN300-045

Supply Cable

BLK WH RD OR GN

RD RD

Base Down300-028Wire Color Legend

Note 1: 110V - Back Tilt Motor Assy. - P/N 300-067 220V - Back Tilt Motor Assy. - P/N 300-270

Test 1 Jump 2-7 Base will raise if limit switch is good.

Test 2 Jump 2-6 Base will lower if limit switch is good.

BLK GN WH RD OR BR -

Black Green White Red Orange Brown

12

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

EVOLUTION SERIES - BASE ASSEMBLY

Base/Top Supply Cable (300-045)

See Note 1

See Note 2

Power Supply Cord (300-046)

Note 1: 110V - Base Capacitor - P/N 300-273 220V - Base Capacitor - P/N 300-284 Note 2: Base shaft should be cleaned and lubricated every 24 months.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

13

EVOLUTION SERIES -BASE LIMIT ADJUSTMENTS

See Note 1

12.5" Max. Down28.5" Max. UpBase Up Limit Switch (300-028) and Base Down (300-028)

Note 1: 110V - Base Motor Assy. - P/N 300-047 220V - Base Motor Assy. - P/N 300-256 Note 2: CAUTION do not exceed Maximum up or down limit setting.

14

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

EVOLUTION SERIES -SHROUD ASSEMBLY

6 1

5

2 1

4 2

43

IL66.eps

3

ITEM PART DESCRIPTION 1 1 2 3 4

PART NUMBER QTY 1 1 1 1 1

Wideback Shroud ............................................. 300-286 Slingback Shroud 300-285 Bellows Cover.................................................. 300-417 Edge Molding for Bellows 300-428 Fabric Bellows ................................................. 300-418

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

15

EVOLUTION SERIES - FOOT CONTROL - PARTS LIST

4

IL207.eps

5 Pre-Position Sensor - Yellow Power - Black Back Up - Green Base Down - Red Orange - Exit White - Base Up Brown - Back Down Blue - Pre-Position

ITEM PART DESCRIPTION 1 2 3 4 5

PART NUMBER QTY 1 2 2 1 1

Screw................................................................ 001-208 Clamp ............................................................... 300-054 Screw................................................................ 001-205 Clamp ............................................................... 300-025 Foot Control ..................................................... 300-001

To retrot foot control to chair in eld

16

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

EVOLUTION SERIES -TOP ASSEMBLYFoot Control (300-001)

See Note 1

See Note 2

See Note 4

See Note 3

Serial Number

Note 1: 110V - Back Tilt Motor Assy. - P/N 300-067 220V - Back Tilt Motor Assy. - P/N 300-270 Note 2: 110V - Top Capacitor - P/N 300-282 220V - Top Capacitor - P/N 300-283 Note 3: 110V - P.C. Board - P/N 300-057 220V - P.C. Board - P/N 300-248 Note 4: Toe motor shaft should be cleaned and lubricated every 24 months.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

17

BASE MOTOR BRAKE REPLACEMENT INSTRUCTIONS WARNINGThis procedure is dangerous! There is a potential for injury including the loss of ngers or limbs. It is extremely important that you follow these directions carefully. READ THROUGH ENTIRE DIRECTIONS FIRST If you do not feel qualied to perform this procedure, contact a qualied service person.

Steps 1-10 Shows Transmission removal 1. Before starting this job, you will need several items: 1 screw - 1/4 diameter with 20 threads per inch - any type head 1 2x 4 - cut to 24 in length 1 pair of needle nose pliers 1 claw hammer (possibly) 1 open end wrench - 1/4 wide 1 open end wrench - 11/32 wide 1 roll of masking tape (no duct tape) 1 - 1/8 Allen wrench 1 - 3/32" Allen wrench

2.

Elevate the chair fully up, as shown.

3.

Release the bellows assembly. Remove the edge molding, as shown. Remove three (3) 11/32" nylon lock nuts and washers from the bottom, one (1) in front and two (2) in rear.

4.

Collapse the bellows assembly fully up, as shown. Secure the collapsed bellows using masking tape, as shown. Do not use duct tape as it could leave marks on the equipment.

18

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

BASE MOTOR BRAKE REPLACEMENT INSTRUCTIONS5. THIS STEP IS VERY IMPORTANT! Insert the 2"x 4" into the scissor lift as shown. It should make solid contact with the top plate and the lower bellows support.

DANGERThis brace must not be bumped in any way during this procedure. If this brace becomes unstable, the entire chair may fall. Keep ngers away from scissor lift joints as much as possible as there is a potential for injury.

Lower the chair down until it is snug with the 2" x 4". If the scissor lift shifts backward because you have lowered the chair down too far, raise it back up and lower it down until the chair is in the correct position. Unplug the chair from the wall socket. 6. Next, remove the pin that links the motor power screw to the chair base. This connection may be tight. If necessary, use the hammer to tap the pin out.

7.

Using the Allen wrench, loosen the two (2) set screws above the motor pivot pins.

8.

Take the 1/4 screw and thread it partway into the hole in the motor pivot pin. If necessary, use the claw hammer to pull the pin out of the motor. Repeat this procedure for the motor pivot pin on the other side of the chair.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

19

BASE MOTOR BRAKE REPLACEMENT INSTRUCTIONS

9.

Remove the motor assembly and position it, as shown.

10 Using the open end wrench, remove the three (3) screws which attach the electric motor to the transmission. End transmission removal

11. Take note of how the brake is positioned in the transmission. The die cast part of the brake assembly should be facing out. Remove the two (2) 3/32" allen head screws that hold the brake assembly in the transmission.

12. Remove the brake assembly using a pair of needle nose pliers. Insert the new brake assembly (die cast side out) into the transmission. Note: Oil should not be present in the transmission cavity. If oil is present, the transmission must be sent back to Beaverstate Dental for resealing before installing the new brake.

13. Take the coupling (black rubber piece) and insert it onto the same shaft as the brake assembly. Next, engage the motor with the transmission.

20

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

BASE MOTOR BRAKE REPLACEMENT INSTRUCTIONS

14. Replace the three (3) screws (with washers) and reattach the motor to the transmission. Note: The transmission power screw should be directly opposite the motor hump.

15. Replace the motor to its original position. Be careful to align all of the holes, so the shafts and pins will install easily. Note: The wire bundle must not extend past the lower beam or the scissors lift will damage the wire bundle.

16. Replace the two (2) motor pivot pins and the pin at the end of the transmission power screw. These pins may t a bit tight, if so, lightly tap the pins with the hammer if needed. Do not use excessive force on the pins. If you are having trouble reinstalling a pin, adjust the hole position and try again. Excessive force will damage the chair. Tighten the set screws above the motor pivot pins. Plug the chair back in and run it to the full up position. This should allow you to remove the wood support. Once this is done, cycle the chair up and down a few times to make sure the motor has been installed correctly. If there is a problem, run the chair back up, reinstall the wood support as in Step 5, and x any bad connections made between the motor and chair. 17. Finally, remove the tape and reinstall the bellows. Refer to Step 3.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

21

BASE MOTOR REPLACEMENT INSTRUCTIONS WARNINGThis procedure is dangerous! There is a potential for injury including the loss of ngers or limbs. It is extremely important that you follow these directions carefully. READ THROUGH ENTIRE DIRECTIONS FIRST If you do not feel qualied to perform this procedure, contact a qualied service person.

1.

Before starting this job, you will need several items: 1 screw - 1/4 diameter with 20 threads per inch - any type head 1 2x 4 - cut to 24 in length 1 pair of needle nose pliers 1 claw hammer (possibly) 1 open end wrench - 1/4 wide 1 open end wrench - 11/32 wide 1 roll of masking tape (no duct tape) 1 - 1/8 Allen wrench 1 - 3/32" Allen wrench 1 - Philips screwdriver

2.

Elevate the chair fully up, as shown.

3.

Release the bellows assembly. Remove the edge molding, as shown. Remove three (3) 11/32" nylon lock nuts and washers from the bottom, one (1) in front and two (2) in rear.

4.

Collapse the bellows assembly fully up, as shown. Secure the collapsed bellows using masking tape, as shown. Do not use duct tape as it could leave marks on the equipment.

22

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

BASE MOTOR REPLACEMENT INSTRUCTIONS5. THIS STEP IS VERY IMPORTANT! Insert the 2"x 4" into the scissor lift as shown. It should make solid contact with the top plate and the lower bellows support.

DANGERThis brace must not be bumped in any way during this procedure. If this brace becomes unstable, the entire chair may fall. Keep ngers away from scissor lift joints as much as possible as there is a potential for injury.

Lower the chair down until it is snug with the 2" x 4". If the scissor lift shifts backward because you have lowered the chair down too far, raise it back up and lower it down until the chair is in the correct position. Unplug the chair from the wall outlet. 6. Next, remove the pin that links the motor power screw to the chair base. This connection may be tight. If necessary, use the hammer to tap the pin out.

7.

From left to right, disconnect the following: screw #1 (two green wires), #4 (yellow wire), #6 (red & blue wires), #7 (orange & red wires).

8.

Using the Allen wrench, loosen the two (2) set screws above the motor pivot pins.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

23

BASE MOTOR REPLACEMENT INSTRUCTIONS

9.

Take the 1/4 screw and thread it partway into the hole in the motor pivot pin. If necessary, use the claw hammer to pull the pin out of the motor. Repeat this procedure for the motor pivot pin on the other side of the chair.

10. Remove motor assembly. To install new motor, reverse steps 1-10.

24

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

UPPER/BACK TILT MOTOR REMOVAL INSTRUCTIONS

1.

Recline backrest to full down position. Raise chair to full up position.

2.

UNPLUG CHAIR

3.

Remove seat assembly by unscrewing one (1) screw from each side of seat assembly.

4.

The motor cable is secured with three (3) screws, remove them.

5.

Lift wire harness out of PC board wiring slot and disconnect 4 pin connector.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

25

UPPER MOTOR REMOVAL INSTRUCTIONS

6.

Unplug green (ground) wire from circut board.

7.

Unclip capacitor from metal retainer.

8.

Remove e-clip. Push pin in, 3/4 of the way through motor tube bushing.

Hold backrest with one hand and move it slightly to remove tension from motor shaft, slightly rotate motor tube away from pin. With motor removed, backrest will rest against lower shroud.

26

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

UPPER/BACK TILT MOTOR REMOVAL INSTRUCTIONS

9.

Remove two (2) 5/32" hex head motor mount screws at front of chair. Support motor to avoid falling and causing damage.

10. Check for proper adjustment of motor (see adjustment instructions) before installing replacement motor.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

27

UPPER/BACK TILT MOTOR ADJUSTMENT INSTRUCTIONS

1.

Backrest down - With the motor tube run all the way in to the limit switch stop, there should be 3/8" to 5/8" distance between the tube and the motor gear housing. To adjust the distance, manually unscrew the tube off the threaded shaft. Turn a half turn and screw it back on. Check the distance indicated in step 1. Make sure motor is at limit stop.

2.

Backrest up - With the motor tube run all the way out to limit switch stop, there should be 7" to 7 1/8" between the tube and the motor gear housing. Make sure the distance in step 1 is correct then adjust the distance in step 2. If distance in step 2 is incorect, see steps 3 through 8.

3.

Remove motor from chair if it isn't already. See removal instructions.

4.

Hold motor tube so it won't spin. Run tube all the way out to limit stop. Note: You will need to connect the 4 pin connector to another power source

28

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

UPPER/BACK TILT MOTOR ADJUSTMENT INSTRUCTIONS

5.

Remove two (2) screws that hold the front round cover on the motor.

6.

Loosen the two (2) screws that hold the nylon cam in place.

7.

Rotate cam clockwise for more travel or counter clockwise for less travel. NOTE: The outer cam is the only one that adjusts. It is the limit stop for the backrest up tube out.

Mo

Less Travel

Travel re

8.

Hold on to the tube and run it in to stop. Install in chair.

800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com

29

Beaverstate Dental, Inc.115 S. Elliott Rd. Newberg OR 97132 800-237-2303 503-538-8756 503-538-2845 Fax www.beaverstatedental.com