Page 1: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Page 2: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Protons - Superior Dose Distribution

The Bragg Peak is representative of a delivery of a full, localized, and uniform dose of energy in the target R

elat

ive

Dos

e 3

2

1

0 10 20 30Depth in Tissue (cm)

4

60

Page 3: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

System Components

The Synchrotron Accelerator

The Beam Transport System

The Stationary Beam Gantries (Isocentric

Gantries) Patient Alignment and

Digital Imaging Systems Facility Control System

Page 4: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Facility LayoutThe Loma Linda University Medical Center Proton Treatment CenterStationary Beam:

Has two branches: •Eye Tumors •Head and Neck Tumors

The Injector:Protons are stripped out of the nucleus of hydrogen atoms and sent to the accelerator.

Synchrotron: (Accelerator)A ring of magnets, 20 ft. in diameter, through which protons circulate in a vacuum tube.

Beam Transport System:Carries the beam from the accelerator to one of four treatment rooms. This system consists of several bending and focusing magnets which guide the beam around corners and focus it to the desired location.

The Gantries:Resembling giant ferris wheels can rotate around the patient and direct the proton beam to a precise point. Each gantry weights about 90 tons and stands 3 stories tall. It supports the bending and focusing magnets to direct the beam.

Steel-reinforced concrete walls are up to 15 feet thick.

Page 5: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Proton Accelerator Control System

Conforma 3000 Integrated Solution Treatment Control

System (TCS) Accelerator Control

System (ACS) Treatment Workstation Integrated and

Automated

“The Brain”Software & Electronics for Treatment Precision

AccelerateDeliverControl

Page 6: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Accelerator Control - Advantages

Automatic Beam Centering in Treatment Nozzle Key to Active Beam Scanning

Rapid Beam Line Switching <10 sec.

Accurate, Reproducible Beam Energy Within 0.1 MeV

Patented Safety Systems Ensure Beam is Delivered at the Right Place & Time

Page 7: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Gantry Treatment Room

Integrated and automated Rotates >360o

Treatment from multiple “portals” or angles

Isocenter is contained within a sphere to 1.0 mm in radius

Automated set up Designed for functionality

Page 8: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Conforma 3000 Passive Scattering Beam Delivery System

Compensator (Bolus) and

Block (Aperture)

Range Modulation

WheelSecond

ScattererFirst

Scatterer

Dose andFlatness

Monitor(s)

X-rayTube Profile

Monitor

MovableSnout

(cones)

Beam

Page 9: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Conforma 3000 Active Beam SystemIntensity Modulated Proton Therapy

OptionalCompensator (Bolus) and

Block (Aperture)

MovableSnout

(cones)

Redundant Dose

Detectors

Radiographic Imaging

Panels (2)

Scanning Magnets

Beam

IMPT or Active Beam Scanning (ABS)

Supports Compensator-less treatments

Variable energy controls penetration depth

Capable of continuous (raster scanning) &

discrete point based (“Spot” scanning)

Page 10: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Intensity Modulated Proton Therapy

Ability to Treat Complex Volumes & Multiple Target Areas

“Spot Scanning” Technique Beam scans from the

top down Radiation deposited

from side to side

Page 11: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Modular Design (4 gantry, 2 Stationary)

Page 12: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Treatment Planning Software

Page 13: Experience, Expertise and a Commitment to Excellence™

Experience, Expertise and a Commitment to Excellence™

Contact Information

Jon Slater,

President & CEO

909.799.8333

[email protected]

Daryl Anderson

Chief Financial Officer

909.799.8342

[email protected]

David Lesyna

VP Engineering

909.799.8312

[email protected]

Mike Martin

VP of Sales

513.518.8037

[email protected]

Ken Rice

Director of Computer Science

909.799.8318

[email protected]

www.optivus.com