expiratory cassette cleaning wall manual.pdf
TRANSCRIPT
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Disinfection Washer-DisinfectorSERVO-i/s
Caution!All personnel should be aware of the riskof parts being infected when disassembling and cleaning the ventilator.
0123
Wiping and DiscardingWhenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
Rinsing the cassette in water (<35°C/95°F) immediately after use may be a useful alternative to disinfection. Immediate rinsing can remove particles and reduces the risk of cross contamination between patients.
Important!
Important!
Rinse Washer-Disinfector
Position and flow
85-95°C (185-203°F)
If rinsing before disinfection is not included in the washer disinfector program:
Drying alternatives
12-24 hours
Run the cassettein a SERVO-i/s witha testlung
Shake/Tilt
5-7 times 10 min
Drying cabinet
1 hour maximum 70°C (158°F).
If available, connect a 22 mm hose to expiratory inlet.
Room air
Caution!Never dry the cassette by applying high pressure air as you may damage the internal tubing.
The water pressure in the washer-disinfector should not exceed 1.5 bar.The maximum water flow should not exceed 10 l/min.The water must have free passage through the cassette.
Note.
Rinse the parts thoroughly in water (<35°C/95°F).Let the water flow through the parts.
© Maquet Critical Care AB 2012. All rights reserved. • MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1201 • Rev. 09 English • Routine wall cleaning diagram 3.2.0 English
GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences. MAQUET specializes in solutions, therapies and products for surgical interventions and intensive care
Disinfection Manually
Caution!All personnel should be aware of the riskof parts being infected when disassembling and cleaning the ventilator.
SERVO-i/sRefer to the Cleaning and Maintenance User’s Manual for detailed information.
0123
Wiping and DiscardingWhenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
Important!
Rinse Drying alternatives
12-24 hours
Run the cassettein a SERVO-i/s witha testlung
Shake/Tilt
5-7 times 10 min
Drying cabinet
1 hour maximum 70°C (158°F).
If available, connect a 22 mm hose to expiratory inlet.
Room air
Caution!Never dry the cassette by applying high pressure air as you may damage the internal tubing.
Important!
Disinfectant
Rinse
Rinse the parts thoroughly in water to remove all traces of disinfectant. Let the water flow through the parts. Carefully shake and tilt the cassette, turn upside down and repeat, at least 5-7 times.
Let the parts soak in a disinfectant:Ethyl- or isopropyl alcohol 70%Cidex OPAHexanios G+R, Aniosyme DD1, Gigazyme Plus or Anioxyde 1000.
Important!Follow the disinfectant manufacturer’s recommendations and instructions.
Rinsing the cassette in water (<35°C/95°F) immediately after use may be a useful alternative to disinfection. Immediate rinsing can remove particles and reduces the risk of cross contamination between patients.
Important!
Rinse the parts thoroughly in water (<35°C/95°F). Let the water flow through the parts.
© Maquet Critical Care AB 2012. All rights reserved. • MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1201 • Rev. 09 English • Routine wall cleaning diagram 3.2.0 English
GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences. MAQUET specializes in solutions, therapies and products for surgical interventions and intensive care
15 min 121°C (250°F)
Wiping and DiscardingWhenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
Important!
Rinse the cassette in water (<35°C/95°F). Let the water flow through the parts
Before placing the expiratory cassette in an autoclave, make sure that no water remains inside the cassette.
Rinse
Dry
5-7 times
Shake/Tilt
4 min 134°C (273°F)
Autoclave
Sterilization is normally not necessary for the expiratory cassette but, when applied, use validated processes only.
Instrument parts are typically autoclaved at a temperature of 134°C (273°F) for 4-7 minutes.
Rubber parts are typically autoclaved at a temperature of 121°C (250°F) for 15 minutes.
Note.
Drying alternatives
12-24 hours
Run the cassettein a SERVO-i/s witha test lung
Shake/Tilt
5-7 times 10 min
Drying cabinet
1 hour maximum 70°C (158°F).If available, connect a 22 mm hose to expiratory inlet.
Room air
Caution!Never dry the cassette by applying high pressure air as you may damage the internal tubing.
Important!
12-24 hours
Sterilization procedure
Caution!All personnel should be aware of the riskof parts being infected when disassemblingand cleaning the ventilator.
(Note. Autoclaving reduces the lifetime of the expiratory cassette)
0123
SERVO-i/sRefer to the Cleaning and Maintenance User’s Manual for detailed information.
Drying cabinet
1 hour maximum 70°C (158°F).If available, connect a 22 mm hose to expiratory inlet.
Room air
© Maquet Critical Care AB 2012. All rights reserved. • MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1201 • Rev. 09 English • Routine wall cleaning diagram 3.2.0 English
GETINGE GROUP is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, GETINGE and MAQUET. ArjoHuntleigh focuses on patient mobility and wound management solutions. GETINGE provides solutions for infection control within healthcare and contamination prevention within life sciences. MAQUET specializes in solutions, therapies and products for surgical interventions and intensive care
Refer to
the Cleaning
and M
aintenance User’s M
anual for d
etailed
inform
ation.
When S
ervo Duo G
uard filter (or eq
uivalent) is used, the
only cleaning needed
is to wip
e the ventilator and
removab
le parts w
ith a soft lint-free cloth moistened
in soap
& w
ater or detergent-b
ased d
isinfectant. N
ote. In case of m
ore contaminated
surfaces use ethyl alcohol (70%
) or isoprop
yl alcohol (70%). A
void contact
with electrical contacts
Wip
e off/d
iscardHazard
ousw
aste(infectious)
Hazard
ousw
aste(infectious)
Whenever p
ossible, clean and
disinfect the eq
uipm
ent im
med
iately after use. Do not leave foreign p
articles, such as b
lood or saliva, to d
ry on the equip
ment.
Cautio
n!A
ll personnel should
be aw
are of the risk of p
arts being infected
when d
isassemb
ling and
cleaning the ventilator.
Imp
ortant!
Cleaning procedure w
hen Servo D
uo Guard filter is used
0123
SE
RV
O-i/s
© M
aquet C
ritical Care A
B 2012. A
ll rights reserved. • M
AQ
UE
T reserves the right to mod
ify the design and
specifications contained
herein without p
rior notice. Ord
er No. 65 15 014
• Printed
in Sw
eden • 1201 • R
ev. 09 English • R
outine wall cleaning d
iagram 3.2.0 E
nglishG
ETIN
GE
GR
OU
P is a lead
ing global p
rovider of p
roducts and
system
s that contribute to q
uality enhancement and
cost efficiency w
ithin healthcare and life sciences. W
e operate und
er the three b
rands of A
rjoHuntleigh, G
ETIN
GE
and M
AQ
UE
T. ArjoH
untleigh focuses on p
atient mob
ility and w
ound m
anagement solutio
ns. G
ETIN
GE
provid
es solutions for infection control w
ithin healthcare and
contamination p
revention within life sciences. M
AQ
UE
T sp
ecializes in solutions, therapies and
prod
ucts for surgical interventions and
intensive care