feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed...

6
Journal of Chromatography A, 835 (1999) 237–242 Short communication Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography * Ya-Ling Tsai, Tiing Yu Department of Applied Chemistry, National Chiao Tung University, 1001 Ta-Hsueh Road, Hsinchu 30050, Taiwan Received 22 September 1998; received in revised form 7 December 1998; accepted 7 December 1998 Abstract Newly synthesized environmentally benign refrigerant 1,1,1,2-tetrafluoroethane (R134a) was used as the mobile phase in high-speed counter-current chromatography (HSCCC) in this study. R134a remains as a liquid under mild pressure and becomes a gas at atmospheric pressure. Accordingly, operation of HSCCC equipped with regular polytetrafluoroethylene tubing was satisfactory. Since R134a turned into gas at ambient temperature and pressure, further concentration for the collected fractions after elution was not required. A mixture of four standard components of Angelicae radix was separated to demonstrate the feasibility of using this new mobile phase in HSCCC. 1999 Elsevier Science B.V. All rights reserved. Keywords: Mobile phase composition; Counter-current chromatography; Tetrafluoroethane 1. Introduction CCC has been recognized as a useful preparative separation technique. In the last few years, two new Counter-current chromatography (CCC) is a form techniques named pH-peak-focusing and pH-zone- of liquid chromatography. The retention of the liquid refining CCC have been developed and proved to be stationary phase in the column requires no solid highly efficient preparative methods [1–4]. Refer to support; therefore permanent adsorption of analytes the literature [5–8] for thorough description of does not occur. Separation is entirely attributable to theory and application for CCC and high-speed CCC partitioning of analytes between the stationary and (HSCCC). Recently we modified a solar coaxial mobile phases. Conventionally, the solvent systems CCC apparatus suitable for using supercritical fluid are made by mixing two immiscible liquids. After carbon dioxide as the mobile phase [9]. Separations thorough saturation of each other, one of the two can be accomplished in a more versatile way through portions is used as the mobile phase and the other as fine-tuning the solvating power of supercritical fluid the stationary phase. Without support from the solid carbon dioxide by changing the temperature and packing material, liquid stationary phases are re- pressure [10]. In addition, the supercritical fluid tained in the separation tubing by either gravitational carbon dioxide becomes a gas when leaving the or centrifugal force. Due to the large capacity of the column, which spares further concentration of the stationary phase allowed in the separation tubing, collected fractions. The advantages are, however, overshadowed by the fact that pressure-resistant * tubing and valves are required for the operation. Corresponding author. Tel.: 1886-3-5731673; fax: 1886-3- 5723764; e-mail: [email protected] Furthermore, polyether ether ketone (PEEK) tubing 0021-9673 / 99 / $ – see front matter 1999 Elsevier Science B.V. All rights reserved. PII: S0021-9673(98)01059-0

Upload: ya-ling-tsai

Post on 02-Jul-2016

213 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography

Journal of Chromatography A, 835 (1999) 237–242

Short communication

Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as themobile phase in high-speed counter-current chromatography

*Ya-Ling Tsai, Tiing YuDepartment of Applied Chemistry, National Chiao Tung University, 1001 Ta-Hsueh Road, Hsinchu 30050, Taiwan

Received 22 September 1998; received in revised form 7 December 1998; accepted 7 December 1998

Abstract

Newly synthesized environmentally benign refrigerant 1,1,1,2-tetrafluoroethane (R134a) was used as the mobile phase inhigh-speed counter-current chromatography (HSCCC) in this study. R134a remains as a liquid under mild pressure andbecomes a gas at atmospheric pressure. Accordingly, operation of HSCCC equipped with regular polytetrafluoroethylenetubing was satisfactory. Since R134a turned into gas at ambient temperature and pressure, further concentration for thecollected fractions after elution was not required. A mixture of four standard components of Angelicae radix was separatedto demonstrate the feasibility of using this new mobile phase in HSCCC. 1999 Elsevier Science B.V. All rights reserved.

Keywords: Mobile phase composition; Counter-current chromatography; Tetrafluoroethane

1. Introduction CCC has been recognized as a useful preparativeseparation technique. In the last few years, two new

Counter-current chromatography (CCC) is a form techniques named pH-peak-focusing and pH-zone-of liquid chromatography. The retention of the liquid refining CCC have been developed and proved to bestationary phase in the column requires no solid highly efficient preparative methods [1–4]. Refer tosupport; therefore permanent adsorption of analytes the literature [5–8] for thorough description ofdoes not occur. Separation is entirely attributable to theory and application for CCC and high-speed CCCpartitioning of analytes between the stationary and (HSCCC). Recently we modified a solar coaxialmobile phases. Conventionally, the solvent systems CCC apparatus suitable for using supercritical fluidare made by mixing two immiscible liquids. After carbon dioxide as the mobile phase [9]. Separationsthorough saturation of each other, one of the two can be accomplished in a more versatile way throughportions is used as the mobile phase and the other as fine-tuning the solvating power of supercritical fluidthe stationary phase. Without support from the solid carbon dioxide by changing the temperature andpacking material, liquid stationary phases are re- pressure [10]. In addition, the supercritical fluidtained in the separation tubing by either gravitational carbon dioxide becomes a gas when leaving theor centrifugal force. Due to the large capacity of the column, which spares further concentration of thestationary phase allowed in the separation tubing, collected fractions. The advantages are, however,

overshadowed by the fact that pressure-resistant* tubing and valves are required for the operation.Corresponding author. Tel.: 1886-3-5731673; fax: 1886-3-5723764; e-mail: [email protected] Furthermore, polyether ether ketone (PEEK) tubing

0021-9673/99/$ – see front matter 1999 Elsevier Science B.V. All rights reserved.PI I : S0021-9673( 98 )01059-0

Page 2: Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography

238 Y.-L. Tsai, T. Yu / J. Chromatogr. A 835 (1999) 237 –242

which worked smoothly for the anti-twist mechanism polar analytes were more readily eluted using R134ain the slow-speed solar coaxial apparatus failed to as the mobile phase than methanol-modified carbonremain intact in HSCCC [11]. dioxide [12,15].

Alternatively 1,1,1,2-tetrafluoroethane (R134a) isemployed as the mobile phase to replace commonliquids in this work. R134a is one of the hydro- 2. Experimentalfluorocarbons that were synthesized to substitute theozone-depleting chlorofluorocarbon refrigerants. 2.1. CCC instrumentation and proceduresCantrell and Blackwell demonstrated that R134apossessed different selectivities and solvation power The instrumental layout is shown in Fig. 1. Aon retention than carbon dioxide and methanol-modi- Pharma-Tech (Baltimore, MA, USA) Model CCC-fied carbon dioxide when used as the mobile phase in 1000 HSCCC was housed in an oven with a 618Cchromatography [12]. In addition, Blackwell and precision. Three column modules were wound fromStringham [13] examined the mixture of carbon a 29.5-m long PTFE tube (2.54 mm O.D.31.59 mmdioxide and R134a as the mobile phase of near- I.D.), that gave a total volume of 177 ml. The ovencritical chromatography. This mobile phase showed was adjusted to the experimental temperature. Thestrong temperature and compositional dependence on stationary phase was then delivered to fill the columnboth retention and selectivity. The same research by an Isco 260D syringe pump (Lincoln, NE, USA).group did several other works to demonstrate the After the excessive stationary phase was discarded,promise of using R134a as mobile phase or modifiers R134a was drawn from the liquid tank (|6.4 bar) toin super- and sub-critical fluid chromatography in the pump and pressurized to the desired pressure.analytical applications [14–17]. Jackson and Fulton While the CCC rotor was revolving at 800 rpm, the[18] examined the possibility of using R134a and mobile phase was pumped into the column. Thesome hydrochlorofluorocarbons as extraction sol- stationary phase was gradually replaced by thevents for polar components in the presence of R134a at a flow-rate of about 2 ml /min and collectedsurfactants. Roth investigated the possibility to em- in a graduated cylinder via Path 1. When the mobileploy these alternative refrigerants as super-critical phase flow reached the outlet, a burst of R134a vaporextraction solvents from their thermodynamic pros- was observed and the flow-rate indicated on thepect [19]. pump head also increased drastically. The flow-rate

R134a remains as a liquid under relatively low was then gradually stabilized. The stationary phasepressure; thus the operation using a CCC apparatus was collected a further 10 min and the collectedwith polytetrafluoroethylene (PTFE) tubing is volume was used to roughly estimate the phaseadequate. The liquid evaporates into a gas at room retention ratio. The experiment was finished if onlytemperature and pressure, and separates from ana- the stationary phase retention was to be measured.lytes directly after coming out of the column. Thus, If a chromatographic run was followed, the mobileit also possesses the advantage of using supercritical phase was allowed to flow one more hour afterfluid carbon dioxide as the mobile phase in CCC. stationary phase displacement to reach hydrodynamicThe operational pressure for the fluid to have an equilibrium, because the two phases were not pre-adequate solvent strength, however, would be much saturated like in the conventional operation. Then thelower than supercritical fluid CO . We examined the R134a flow was switched using the six-port switch to2

feasibility of using R134a as the mobile phase in the pass through a Bio-rad (Hercules, CA, USA) 1801separation of a mixture of four standard components UV–Vis HPLC detector then to a tube, via Path 2.of Angelicae radix using HSCCC in this work. The flow-rate was fine-adjusted using a meteringAlthough components separated in this study are of valve (Valve V2). It usually took several minutes torelatively low polarity, R134a may potentially also attain a steady baseline signal after path switch tobe used to elute more polar compounds due to its allow a sample injection. The attached outlet stain-higher dipole moment (2.06 D) than that of CO (0 less steel tubing (1.59 mm O.D.30.254 mm I.D.)2

D) [18]. In fact, it has been reported that even more immersed in |2 ml methanol was used for collecting

Page 3: Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography

Y.-L. Tsai, T. Yu / J. Chromatogr. A 835 (1999) 237 –242 239

Fig. 1. Schematic diagram of the instrument layout.

components of interest. A 100-ml sample solution 3. Results and discussionwas injected and the effluent was monitored at 212nm. The sample collecting was operated under a Because a PTFE tubing was used as the separationhood. column, we tried to keep the pressure as low as

possible. After careful testing we found steady flowcould be obtained when the applied pressure was

2.2. Chemicals above 10 bar. Since the upper pressure limit for thePTFE tubing was rated 20.7 bar (300 p.s.i.), the

Five standard components of Angelicae radix operational pressure range was therefore about 10–were examined in this study. Their structures are 20 bar for this system. Because R134a is a gas atshown in Fig. 2. p-Cymene (.95%) was obtained ambient temperatures and pressures, it is not possiblefrom TCI (Tokyo, Japan); trans-ferulic acid (99%), to measure the partition coefficient (K) in the solventbergapten (99%) and safrole (97%) from Aldrich system in the conventional manner. We therefore(Milwaukee, WI, USA); HPLC-grade isopimpinellin assumed the polarity of R134a was similar to that offrom Extrasynthese (Genay Cedex, France). The chloroform. Since the target compounds were ofsample solutions were prepared in methanol. HPLC- medium polarity, a R134a–methanol–water two-grade methanol and ethyl acetate were purchased phase system was considered a reasonable selection.from Mallinckrodt (Paris, KY, USA). Refrigerant Both the head-to-tail and tail-to-head elution modes1,1,1,2-tetrafluoroethane (99.9%) was purchased [5–7] were tested. No significant phase retention wasfrom Ausimont (Milan, Italy). observed for the tail-to-head flow direction. Phase

Page 4: Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography

240 Y.-L. Tsai, T. Yu / J. Chromatogr. A 835 (1999) 237 –242

acid. However, this component was not eluted out ofthe column after 3 h using the aqueous stationaryphases of three different ratios of methanol (10, 40and 60%, v/v). In order to reduce the dissociation ofthis acid, the pH of the aqueous solution was thusadjusted to acidic. Under pH of 2.3, 3.5, 4.6 and 5.9the mobile phase was still unable to elute thiscomponent. Since the property of R134a was as-sumed to be close to that of chloroform, the partitioncoefficient of trans-ferulic acid was measured usinga methanol–water–chloroform (6:4:4) system. The Kvalue found was 0.74, which suggested a reasonableelution time for trans-ferulic acid. The result con-tradicted the assumption of the similarity betweenR134a and chloroform.

3.2. Separation for a mixture of bergapten, p-cymene, isopimpinellin and safrole

With a low ratio of methanol in water, it was notpossible to resolve the p-cymene and safrole pair,and the bergapten and isopimpinellin pair well. The

Fig. 2. Structures of bergapten, trans-ferulic acid, isopimpinellin, capacity factors of all the components were found top-cymene and safrole. increase with increasing methanol content in the

stationary phase. Two chromatograms using station-retention ratios (S ) were examined under different ary phases of slightly different methanol portions areF

compositions using the head-to-tail mode, and the shown in Fig. 3. As can be seen, the resolution of theestimated results are shown in Table 1. two pairs using the stationary phase of methanol–

water (70:30) edged that of using methanol–water3.1. Elution of trans-ferulic acid (65:35) at the expense of longer elution time. In

order to obtain better resolution, an even higherExtraction or chromatography of polar analytes methanol content was apparently required. However,

suffers from the low solvating power using super- higher methanol proportion in the aqueous solutioncritical fluid carbon dioxide. A polar modifier usually lowered the retention volume drastically (see Tableis added to enhance the analyte solubility in the fluid. 1), and thus was prohibited. In fact, the baselineThe relatively high dipole moment of R134a might signal of using methanol–water (70:30) exhibitedact as a successful eluent for the polar trans-ferulic higher noise level than that of using methanol–water

(65:35). The hydrodynamic stability became difficultto reach with higher methanol content. The smallTable 1peak shown in Fig. 3a, becoming a shoulder ofPhase retention ratios under different stationary compositionsp-cymene in Fig. 3b, could be impurities in theMethanol–water S (%)Fp-cymene sample. It usually appeared as an in-

10:90 81 dependent peak in most chromatographic runs. The20:80 81

dip appearing just before the component p-cymene50:50 81was due to the sample solvent and was used to60:40 80

70:30 76 calculate the phase retention ratio. Since the mobile80:20 Negligible and stationary phases were not pre-saturated, the90:10 Negligible stationary phase depletion was expected throughout

Page 5: Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography

Y.-L. Tsai, T. Yu / J. Chromatogr. A 835 (1999) 237 –242 241

temperatures and pressures in this study were farbelow the critical temperature (101.58C) and pres-sure (40.4 bar) [20,21], very small density changeswere expected according to the equation of statederived by Sorner and Strom [22], and thus thesolvation power of R134a. Accordingly, it is notsurprising to observe the negligible effects on vary-ing temperature and pressure.

In conclusion, we successfully demonstrated theprospect of using R134a as the mobile phase forseparating samples of relatively low polarity withoutmodifying the commercialized instrument. However,one gives up many advantages or introduces compli-cations for spontaneous evaporation related to: parti-tion coefficient measurement, mobile phase pre-equilibration, pressure and temperature conditions,choice of mobile phase, and potential air pollution.

Acknowledgements

Financial support (contract no: NSC-88-2113-M-009-003) from the National Science Council ofTaiwan is appreciated. We also thank Professor I.-M.Chu for the generous loan of the CCC apparatus.

Fig. 3. Chromatograms of a mixture of bergapten (B) (500mg/ml), isopimpinellin (I) (500 mg/ml), p-cymene(P) (250 mg/ml) and safrole (S) (500 mg/ml). (a) Stationary phase, methanol– Referenceswater (65:35), S 581%; (b) stationary phase, methanol–waterF

(70:30), S 577%. The effluent was monitored on-line by a UVF

[1] A. Weisz, A.L. Scher, K. Shinomiya, H.M. Fales, Y. Ito, J.detector at 212 nm. Experimental conditions: apparatus, HSCCCAm. Chem. Soc. 116 (1994) 704.centrifuge; column, semipreparative multilayer coil, 1.59 mm I.D.,

[2] A. Weisz, A.L. Scher, Y. Ito, J. Chromatogr. A 732 (1996)177-ml coil volume; flow-rate, 2.0 ml /min; injection volume, 100283.ml; under 20 bar and 258C

[3] Y. Ma, Y. Ito, Anal. Chem. 68 (1996) 1207.[4] Y. Ma, Y. Ito, J. Chromatogr. A 771 (1997) 81.

the elution. Accordingly, the phase-retention ratio at [5] N.B. Mandava, Y. Ito (Eds.), Countercurrent Chromatog-the end of the elution should be different from the raphy: Theory and Practice, Marcel Dekker, New York,

1988.calculated value using the solvent front. It is also[6] W.D. Conway, Countercurrent Chromatography: Apparatus,noted that the composition of the stationary phase at

Theory and Applications, VCH, New York, 1990.the completion of the elution would be also different[7] Y. Ito, W.D. Conway (Eds.), High-Speed Countercurrent

from that at the beginning due to the same reason. Chromatography, Wiley-Interscience, New York, 1996.Although both separations, shown in Fig. 3, were [8] Y. Ito, CRC Crit. Rev. Anal. Chem. 17 (1986) 65.

performed under 20 bar and 258C, experiments at 10 [9] T. Yu, S.-E. Li, Y.-H. Chen, H.P. Wang, J. Chromatogr. A 724(1996) 91.bar and 258C, 15 bar and 258C, and 20 bar and 208C,

[10] T. Yu, Y.-H. Chen, J. Chromatogr. A 790 (1997) 31.were actually also performed using the stationary[11] Y.-L. Tsai, Master Thesis, National Chiao Tung University,

phase of methanol–water (65:35). The results (not Taiwan, 1998.shown) displayed no significant difference from the [12] G.O. Cantrell, J.A. Blackwell, J. Chromatogr. A 782 (1997)chromatogram in Fig. 3a. Since the operational 237.

Page 6: Feasibility study of using refrigerant 1,1,1,2-tetrafluoroethane as the mobile phase in high-speed counter-current chromatography

242 Y.-L. Tsai, T. Yu / J. Chromatogr. A 835 (1999) 237 –242

[13] J.A. Blackwell, R.W. Stringham, J. Chromatogr. A 796 [18] K. Jackson, J.L. Fulton, Langmuir 12 (1996) 5289.(1998) 355. [19] M. Roth, Anal. Chem. 68 (1996) 4474.

[14] J.A. Blackwell, R.W. Stringham, Chirality 9 (1997) 693. [20] K. Fujiwara, S. Nakamura, M. Noguchi, J. Chem. Eng. Data[15] J.A. Blackwell, L.E. Schallinger, J. Microcol Sep. 6 (1994) 43 (1998) 55.

551. [21] H. Aoyama, G. Kishizawa, H. Sato, K. Watanabe, J. Chem.[16] J.A. Blackwell, R.W. Stringham, J. High Resolut. Chroma- Eng. Data 41 (1996) 1046.

togr. 20 (1997) 631. [22] M. Sorner, K.H.U. Strom, Can. J. Chem. Eng. 73 (1995)[17] J.A. Blackwell, R.W. Stringham, Anal. Chem. 69 (1997) 854.

4608.