field desorption, field ionisation

Upload: hey80milion

Post on 14-Apr-2018

235 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 field desorption, field ionisation

    1/13

    IONISATION TECHNIQUESIN MASS

    SPECTROSCOPYField Ionisation (FI) and Field Desorption (FD)

  • 7/27/2019 field desorption, field ionisation

    2/13

    FIELD IONISATION (FI)AND FIELD DESORPTION (FD)

    H. D. Becky in 1969 introduced the concept of FD-

    MS for analysis of large organic molecules.

    FI and FD are "soft" ionisation methods; i.e. theincrease in the internal energy transferred to the

    analyte molecules during ionisation is minimal and

    the sunsequent fragmentation is largely reduced.

  • 7/27/2019 field desorption, field ionisation

    3/13

    FIELDIONIZATION (FI)

    Field ionization (FI) is the generation of M+ ionsby removal of electrons, primarily from gas samplemolecules, using a high electric field.

    This generally occurs at a sharp edge or tip that isbiased to a high electrical potential

  • 7/27/2019 field desorption, field ionisation

    4/13

    During Field Ionization, sample molecules become ionized

    by the quantum tunneling of a valence electron as they

    pass close to the tips of emitter electrodes.

    The electrodes are essentially a large collection of carbon

    micro-needles, deposited on tungsten wire, surrounded by a

    very high electric field within the ion source.

    These micro-needles allow very high electric field strengths

    to be applied to a molecule.

    The mechanism of ionisation is based on the fact that when

    a molecule is subjected to a very high electric field, a

    valence electron tunnels through the potential barrier and is

    removed from the molecule. The resulting ion is therefore

    M+.

  • 7/27/2019 field desorption, field ionisation

    5/13

    FIELD IONIZATION

    Field Ionization offers the following benefits:

    Suitable for substances that can be introduced into

    the source via a gas chromatograph or heated

    direct insertion probe

    Energetically gentle ionization process providing a

    large proportion of molecular ions from many

    different substances

    Spectra exhibit some fragmentation due to the heat

    used for sample volatilization

  • 7/27/2019 field desorption, field ionisation

    6/13

    FIELDDESORPTION (FD)

    Field desorption (FD) is a method for emitting ionsinto the gas phase.

    Sample spread on an emitter is heated while a highelectric field is applied.

    Ions are then emitted by the tunneling, ion-moleculereactions, thermal fusion effects,

    and other phenomenon

    occurring on the emitter

    surface and around the

    whisker ends. The ionization phase depends strongly on the

    sample material and the spread condition.

  • 7/27/2019 field desorption, field ionisation

    7/13

    FIELD DESORPTION

    Field Desorption analysis feature the following:

    Suitable for high molecular mass and/or thermally labilesubstances such as polymers, peptides, carbohydratesand organic or inorganic salts

    A solution of the sample is applied to the emitter before

    it is introduced into the ion source The emitter is mounted on the tip of the axial sample

    introduction probe

    Samples undergo little or no fragmentation during FDionization

    Ionization maximizes the production of high-qualitymolecular mass information for improved compoundidentification/confirmation

    FD-MS has proven capabilities in analyzing nonioniclow- to medium polarity compounds

  • 7/27/2019 field desorption, field ionisation

    8/13

    FIELD IONIZATION (FI) VERSUS FIELD

    DESORPTION (FD)

  • 7/27/2019 field desorption, field ionisation

    9/13

    FIELD IONIZATION (FI) VERSUS FIELD

    DESORPTION (FD)

    FI : sample is heated

    in a vacuum so as to

    volatilize it onto an

    ionization surface. FI issuited for use with

    volatile, thermally

    stable compounds.

    FI sources are

    arranged to function

    also as FD sources

    FD : the sample is

    placed directly onto

    the surface (dipping

    emitter in an analyte

    solution) before

    ionization but FD is

    needed for nonvolatile

    and/or

    thermally labilesubstances.

  • 7/27/2019 field desorption, field ionisation

    10/13

    FIELD IONIZATION (FI) VERSUS FIELD

    DESORPTION (FD)

    FI Spectrum for D-Glucose FD Spectrum for D-Glucose

  • 7/27/2019 field desorption, field ionisation

    11/13

    LIQUID INTRODUCTION FIELD DESORPTION

    IONISATION

    (LIFDI)

    A major advance was made by H B Linden in 2000 byincorporating a fused silica capillary to allow liquidintroduction of the analyte onto the FD emitter, hence theacronym LIFDI.

    LIFDI is regarded as a convenient tool to ionize any solid,liquid, or gaseous species transferred to the emitter insideof the ion source through a capillary without breakingvacuum.

    This development made the FD technique much more

    user friendly and, because the emitter remains in place,the fragile emitters last much longer.

    The fused silica capillary also enables air and moisturesensitive molecules to be easily introduced without

    exposure to ambient air.

    http://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.htmlhttp://www.udel.edu/chem/ms/lifdi.html
  • 7/27/2019 field desorption, field ionisation

    12/13

    The entire LIFDI sample preparation is done by

    dipping the capillary into the sample solution for 1-2

    s. A volume of 40 nL is aspirated automatically and

    forced through the capillary.

    Upon arrival at the emitter, osmotic and capillary

    forces between the dendrites distribute the solution

    over the entire emitter.

    The small volume of solvent is evaporated in the

    vacuum within seconds, followed by acquiringspectra at a total sample prep of less than 30 s.

  • 7/27/2019 field desorption, field ionisation

    13/13

    APPLICATIONS

    Identification of Organometallic compounds using FieldDesorption Ionization Technique.

    Qualitative and quantitative analysis for drugs and

    endogenous compounds, e.g., tranquillizers,

    immunosuppressive and antitumour agents and free

    amino acids, in human body fluids, biocides, e.g.,

    phenylureas, carbamates, organophosphorus and

    organometallic compounds, in river water, and natural and

    synthetic products, e.g., saponins, chlorophyll and

    deoxyribonucleotides, can be performed Field desorption mass spectrometric profile analysis can

    be used as a technique for the detection of

    biotransformation products of xenobiotics in crude urine

    extracts.