ABB MEASUREMENT & ANALYTICS | INFORMATION | INF04/83 REV. G C1300 Advanced circular chart recorders Fitting additional pens Pen capsule (green) (C1900/0122) Pen capsule (black) (C1900/0119) (Kit 1A) r e b m u N t i K e s o p r u P ) J X X 3 1 ( t r a h C e p y T r o l y a T . o N t r a P ) K X X 3 1 ( t r a h C e p y T t n e K . o N t r a P 1 t i K e d a r g p u 2 n e P 0 2 7 1 / 0 0 9 1 C 4 2 7 1 / 0 0 9 1 C A 1 t i K e d a r g p u 4 n e P 1 2 7 1 / 0 0 9 1 C 5 2 7 1 / 0 0 9 1 C 2 t i K e d a r g p u 3 n e P 2 2 7 1 / 0 0 9 1 C 6 2 7 1 / 0 0 9 1 C 3 t i K r e k r a m t n e v e e m i t e u r T 3 2 7 1 / 0 0 9 1 C 7 2 7 1 / 0 0 9 1 C Table 2.1 Pen Upgrade Kits Fig. 2.1 Pen Upgrade Kit 1 or 1A Continued… 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular Chart Recorder. The work must be carried out by a trained technician. Refer to the User Guide (Part No. IM/C1300) for full operating and programming information following upgrades. 2 Pen Upgrade Kit Identification Refer to Table 2.1 and Figs. 2.1 to 2.3 for the contents of pen upgrade kits. Information. Due to variations in pen geometry and position of chart fixings it is not possible to convert instruments from one chart/pen type to another, i.e. Taylor to Kent type or vice versa. The 5 th letter in the instrument code identifies the chart type (see the Ordering Information contained in the Data Sheet, SS/C1300). For example, a Taylor type chart (13XXJ) can be fitted only to instruments with J in the instrument code number. Similarly, when fitting additional or replacement pens, ensure that the correct pen arm is used. Pen arms used with Taylor type charts (13XXJ) are identified by a hole located near the motor end of the arm. This hole is not visible on pen arms used with Kent type charts (13XXK).

Upload: others

Post on 22-Aug-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

— A B B M E A SU R EM ENT & A N A LY TI C S | I N FO R M ATI O N | I N F0 4/8 3 R E V. G

C1300Advanced circular chart recordersFitting additional pens

1

Pen capsule (green) (C1900/0122)Pen capsule (black) (C1900/0119) (Kit 1A)

rebmuNtiK esopruP )JXX31(trahCepyTrolyaT.oNtraP )KXX31(trahCepyTtneK.oNtraP

1tiK edargpu2neP 0271/0091C 4271/0091C

A1tiK edargpu4neP 1271/0091C 5271/0091C

2tiK edargpu3neP 2271/0091C 6271/0091C

3tiK rekramtneveemiteurT 3271/0091C 7271/0091C

Table 2.1 Pen Upgrade Kits

Fig. 2.1 Pen Upgrade Kit 1 or 1A

Continued…

INFORMATIONNo. INF04/83 Rev. F Date January 2014 Product C1300 Advanced Circular Chart Recorders

Manual IM/C1300

Fitting Additional Pens

1 IntroductionThis publication details the fitting of additional pens to the C1300 Advanced Circular Chart Recorder. The work must be carried out bya trained technician. Refer to the User Guide (Part No. IM/C1300) for full operating and programming information following upgrades.

2 Pen Upgrade Kit IdentificationRefer to Table 2.1 and Figs. 2.1 to 2.3 for the contents of pen upgrade kits.

Information. Due to variations in pen geometry and position of chart fixings it is not possible to convert instruments from onechart/pen type to another, i.e. Taylor to Kent type or vice versa. The 5th letter in the instrument code identifies the chart type(see the Ordering Information contained in the Data Sheet, SS/C1300). For example, a Taylor type chart (13XXJ) can be fittedonly to instruments with J in the instrument code number.

Similarly, when fitting additional or replacement pens, ensure that the correct pen arm is used. Pen arms used with Taylor typecharts (13XXJ) are identified by a hole located near the motor end of the arm. This hole is not visible on pen arms used withKent type charts (13XXK).

Page 2: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

Pen capsule (blue) (C1900/0120)

…2 Pen Upgrade Kit Identification

Fig. 2.2 Pen Upgrade Kit 2

Fig. 2.3 Pen Upgrade Kit 3

������

������

����

�����

������

�� ��

��

����

�����

����

������������

���

���

�������

�����

��

�����

� ��

�����

���������

���

������

�����

���

������

����������� !""#""$ %

����&

���������������'��(�)*���+!$&%�,-.���)*���/&* %�,01���2��)*���+/3&%����',4-51���62�1��%�� !""#" *3

�,0�1�� !""#"*"&%

������ ������������������������7��� !""#""$8%

���� ������������������������7��� !""#""$$%

Page 3: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

3.1 Fitting the I/O Module – Figs. 3.1 and 3.2

Warning. Before making any connections, ensure that the power supply, any high voltage-operated control circuits and highcommon mode voltages are switched off.

a) Disconnect the recorder's power supply.

b) Open the door and chart plate.

Caution. The recorder is vulnerable to electrostatic damage. Wear an anti-static strap or place the unit on an anti-staticworkbench.

c) Identify the module position – see Fig. 3.1.

Notes.• Module positions 2 and 3 can also be used for additional I/O modules

(module types 1 and 2) for use with math functions.• The module type is marked on the component side of the PCB.

Fig. 3.1 Module Positions and Functions

� � � � �

���������� �����

���������� �����

��������!����������%

�����������15�����%

"�� ��������! � # $ %

��������#��1,'9����:�;2�1����6�����������<<2�2��,1��5�'�2��-%

����&�����1�(

���������� �<����%

Note. If pen 3 is to be installed (Pen Upgrade Kit 2), refer to Section 3.3 to fit the Display Board and Membrane Overlay.

Continued…

Page 4: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

4

�15=�2���.��#�7�<51�

<���2>?��.�����72�,10�,�<�,�<�#�7�<51��5-2�=4�-2�2���1,0�1-

��-5������72�,1�0�,�<�2-1�',��<�'����'�1?�2��'124

��',������72�,1�0�,�<�2����.��7,2��0�,�<�-544���-�-���&���0�1�(%

*

3

&

…3.1 Fitting the I/O Module – Figs. 3.1 and 3.2d) If necessary, remove and discard appropriate knockout from the recorder's case.

Note. Remove knockouts by striking from the inside of the case, taking great care not to damage any internal components.Alternatively, drill out to 0.875 in. (22.2mm) using a suitable drill bit.

e) Carefully clean out hole and ensure all debris is removed from inside the recorder.

f) Route the leads and cables into the case.

Note. If NEMA 4X hosedown standard is to be maintained, suitable cable glands must be used to prevent water ingress.

g) Fit the I/O module as detailed in Fig. 3.2.

Note. Connections to the terminal block can be madebefore or after it is fitted to the main board

Fig. 3.2 Fitting the I/O Module

Page 5: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

5

3.2 Fitting the Motor and Pen Assembly – Fig. 3.3

Warning. Before making any connections, ensure that the power supply, any high voltage-operated control circuits and highcommon mode voltages are switched off.

a) Disconnect the recorder's power supply.

b) Open the door and chart plate.

Caution. The recorder is vulnerable to electrostatic damage. Wear an anti-static strap or place the unit on an anti-staticworkbench.

Fit the motor and pen assembly as detailed in Fig. 3.3.

Fig. 3.3 Fitting the Motor and Pen Assembly

� � � � �

<���2>?�7�����4�-2�2�������.��7�����41,�� ��',����.��7�����2���.�7�����41,���,�<�-�'5��5-2�=��(��-'��(-�,�<(,-.��-

�����'���.��7���������.��7,2��0�,�<:���-5�2�=�.��=�����(2���2���.��'����'����2-�1�',��<����2�� ��>��.��7,2��0�,�<

�2��4��:���-5�2�=��.�4���,�7�2-�4�-2�2���<,0�6��2�-�12>����0,�

�,��>511?���7�6���.��',01��2�-�>��7��.��7�����41,��,�<�7�����(2��-

"����!���������

"��������������

"����#���������

�<@5-���.��4���4��--5���0?���7�62�=��.��4��,�7�>��7�2�-�?�9��,�<�0��<2�=��.��4���,�7� �>2��4���,�7�,�<���A'.�'9�4��--5���

*

3

*

3

����4���4��--5������ �=�,77�

��� �����

������������������������ ��������������������A����

��� ��������� ������� ��������������� ��������

�2��54=�,<��1,0�1�2�-2<��.��<����=,-9���,��,

*

3

&

���<�>��7�.���

*

3

&

+

/

$

8

!

��'5����.��7�����(2��-�5-2�=��(��',01��2�-�� �>���������4�������>���',01���2�4�-2�2��-

*

Continued…

Page 6: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

6

3.3 Fitting the Display Board and Membrane Overlay – Figs. 3.4 and 3.5

Note. This section is applicable only if installing pen 3 (Pen Upgrade Kit 2).

Warning. Before making any connections, ensure that the power supply, any high voltage-operated control circuits and highcommon mode voltages are switched off.

a) Disconnect the recorder's power supply.

b) Open the door and chart plate.

Caution. The recorder is vulnerable to electrostatic damage. Wear an anti-static strap or place the unit on an anti-staticworkbench.

c) Fit the display board and membrane overlay as detailed in Fig. 3.4.

��'5����.��<2-41,?�0�,�<����.��<����5-2�=�>�5�-�,4��26��-

���1��.��4����'�26��4,4���>��7��.��0,'9��>�.��-�1>A,<.�-26��<2-41,?��6��1,?�,�<:��-5�2�=�'����'����2���,�2��:�2�-�����.��6��1,?�2���.����'�--�2���.���5�-2<���>��.�<��������--�>2�71?�2����4�-2�2���

���7��.��2�-2<���>��.��<���:�45-.��.��'1�,��41,-�2'�01,�92�=�41,���5�(,�<-������7�6���.��-�1>�,<.�-26��01,�92�=�41,���>��7��.��5�-2<���>��.��<����� �7�6��,�?���-2<5,1�,<.�-26��>��7��.��<�����'�--�5-2�=�4����1�57���.���

�41,'�� �.�� <2-41,?� '�6��� 0?��6��-2�=�-��4-��������,�< " * 3

�7�6��-'��(B

���-���-'��(

*B,�<�-12<���.��<2-41,?�'�6���,(,?�>��7��.��<���3

&

+

��-2�2����.��<2-41,?�0�,�<�6����.��1�',�2�=�42�-,<@,'��������,'.��>��.�>�5���26���.�1�-�����.�2�-2<���>��.��<����

/

$

�2�9��.��<2-41,?�0�,�<-�5-2�=�.���200���',01��-54412�<B

8 B��-5�2�=��200���',01�4�-2�2���<�,-�-.�(� !

Fig. 3.4 Fitting the Display Board and Membrane Overlay

Page 7: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

7

…3.3 Fitting the Display Board and Membrane Overlay – Figs. 3.4 and 3.5d) Replace the EPROM as detailed in Fig. 3.5.

e) Close the chart plate and door.

f) Restore the recorder's power supply.

�� ��

�12=��'������'5�A�>>��>���(�� ���(2�.�'5�A�>>�2��-�'9��,�<�4��--���(��� ���>2�71?2����-�'9���

*�-2�=�'.24��)��,'�������1

4��62<�<:�',��>511?��7�6���� ���>��7�2�-

-�'9��

Fig. 3.5 Replacing the EPROM

Page 8: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

8

3.4 Commissioninga) Access to the Commissioning Level is prevented by a security link on the recorder's main board. Set the security link to the OFF

position – see Fig. 3.6.

� � � � �

'����(�)*+ ,�'+&���-�+ +-&�

#

!

'��

-&

-��

��-2�2����.��-�'5�2�?�12�92���.��-���4�-2�2��

#

!

'��

-&

-��

Fig. 3.6 Disabling the Security System

���������������������������������� ���������������������������������������������������������������������������������������

� �����!�������

�������������"�

���������� ���

����������#$##%

�������������������������������������������������������������������������������&����'��()������������������'��()���� ����������&�����*� ��

Instrument SetupPress the key to open the Main Menu.

Highlight Calibration.

Press and hold the key and press the key to access the Commissioning Level.

Highlight Instrument Setup.

Instrument TypeSelect the instrument type required:

� 3 ) )'.,����?4�

� � �.�--�1����,�<��'.,��-C � ,?1����?4��'.,��-�� 3DDC%� � ������?4��'.,��-�� 3DD�%

�570����>�4��- � ����4������<%* � (��4��-����<:�=����%3 � .����4��-����<:�=����:�015�%& � ��5��4��-����<:�=����:�015�:�01,'9%

Continued on next page.

b) Close the chart plate and door.

c) Restore the recorder's power supply.

3.4.1 Instrument Setup

Page 9: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

9

…3.4.1 Instrument Setup

���������� ��������+�(���

������������

��������&���(��(

�������������"�

������ �,������

����������-.�'/

�&�������� ���

��������� ����

���������� ���

����������#$##%

� �����!�������

������������"�

�������������

�������������"�

True Event Pen

Note. For use only when installing Pen Upgrade Kit 3 (True Time Event Marker).

Select YES to enable the true time event option.

Instrument Type UpdateDisplayed to indicate that the instrument type has been updated.

Company StandardSelect YES to reset all configuration parameters to the Company Standard configuration.

Note. YES is displayed momentarily to indicate that all configuration parameters have beenreset to Company Standard. The display then reverts to NO.

Line RejectionSet to frequency of local mains power supplies to ensure maximum noise rejection on analoginputs.

Security TypeSelect type of Configuration Level security required:

A – Security code protectedB – No access

Totalizer Log ResetSelect YES to reset all totalizer logs.

Note. YES is displayed momentarily to indicate that all totalizers have been reset to Zero. Thedisplay then reverts to NO.

Return to top of page.

Page 10: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

10

3.4.2 Hardware Calibration

�������������������������������������������������������������������������������&����'��()������������������'��()���� ����������&�����*� ��

&���������������

��������������0

&���������������

��������������0

��&�������0�'���

��������������.

��&�������0���)

��������������.

Hardware CalibrationPress the key to open the Main Menu.

Highlight Hardware Calibration.

Select Calibration FunctionSelect the pen to calibrate.

Set Pen x HighThe selected pen moves to its 100% position on the chart. Use the and keys to positionthe pen on the 100% line on the chart.

Set Pen x LowThe selected pen moves to its 0% position on the chart. Use the and keys to positionthe pen on the 0% line on the chart.

Return to top of page.

Page 11: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

11

3.4.3 Display Setup and Test

Note. This Section is applicable only if Pen Upgrade Kit 2 (Pen 3) has been installed.

�������������������������������������������������������������������������������&����'��()������������������'��()���� ����������&�����*� ��

�����������#

�����������#..�1

�����������#

�����������#..�1

�����������2

�����������#..�1

��3��������#

�����������#..�1

��3��������2

�����������#..�1

���������4�(�

���������"�����

Display Setup & TestPress the key to open the Main Menu.

Highlight Display Setup & Test.

Contrast 1Set the required screen contrast for display 1.

Contrast 2Set the required screen contrast for display 2.

Brightness 1Set the required brightness for display 1.

Brightness 2Set the required brightness for display 2.

Display ModeSet the required display mode:

Normal – White text on blue backgroundInverted – Blue text on white background

Return to top of page.

Page 12: Fitting Additional Pens - ABB Group · 2020. 4. 24. · Fitting Additional Pens 1 Introduction This publication details the fitting of additional pens to the C1300 Advanced Circular

INF

04

/83

Rev

. G

09

.20

18—We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document.

We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents – in whole or in parts – is forbidden without prior written consent of ABB. ©ABB 2018

—ABB Limited Measurement & AnalyticsHoward Road, St. NeotsCambridgeshire, PE19 8EU, UKTel: +44 (0)1480 475 321Fax: +44 (0)1480 217 948Email: [email protected]

abb.com/measurement

ABB Limited Measurement & Analytics125 E. County Line Road, Warminster, PA 18974USATel: +1 215 674 6000Fax: +1 215 674 7183

12

INF0

4/83

Rev

. F

The Company’s policy is one of continuous productimprovement and the right is reserved to modify the

information contained herein without notice.

Printed in UK (01.14)

© ABB 2014

ABB LimitedHoward Road, St. NeotsCambridgeshirePE19 8EUUKTel: +44 (0)1480 475321Fax: +44 (0)1480 217948

ABB Inc.125 E. County Line RoadWarminsterPA 18974USATel: +1 215-674-6000Fax: +1 215-674-7183

ABB has Sales & Customer Supportexpertise in over 100 countriesworldwide

www.abb.com

4 Completiona) Disconnect the recorder's power supply.

b) Open the door and chart plate.

c) Set the security link to the ON position – see Fig. 3.6.

d) Write the upgrade kit number on the label supplied with the upgrade kit and attach the label to the recorder – see Fig. 4.1.

������������������������������������� �� ���� ����������� ��������

�����������������������������������������������

����� ­��� �������������������­��­������������������������

����������������������

� ������������������������

Fig. 4.1 Attaching Upgrade Kit Label

e) Close the chart plate and door.

f) Restore the recorder's power supply.

f) Refer to Section 3 of the User Guide (IM/C1300) and configure the recorder.