flow measurement uncertainty - analysis based on field grade components
TRANSCRIPT
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
1/12
INTRODUCTION
This paper provides a guide to estimating the un-
certainty of a flow measurement made with field
grade instruments. It is applicable to calibrated or
uncalibrated turbine or ultrasonic based measure-
ments. Pressure and temperature measurements con-
tribute uncertainty, the uncertainty components of
each instrument are described. A set of typical
manufacturers specifications are used as an example.
The effects of ambient temperature on both mea-
surements are described.
A flow computer acquires signals from the pres-
sure and temperature transducers and converts the
readings into engineering units. A gas chromatograph
and the equation of state contribute uncertainty tothe determination of compressibility under flowing
base conditions. The components that contribute un-
certainty are described and typical values are pro-
posed.
The process of combining uncertainty components
is illustrated based on hypothetical values. The con-
cepts of sensitivity coefficient, expanded uncertainty
and coverage factor are described. The data are or-
ganized such that the relative contributions of the
components can be readily compared. The value of
this organizational structure is illustrated by compar-
ing calibrated and uncalibrated turbine meters.
The paper concludes with a brief discussion of
advanced topics. These include topics from flow
measurement, instrument calibration and measure-
ment uncertainty.
Flow Measurement Uncertainty - Analysis Based on Field Grade Components
Thomas Kegel
Colorado Engineering Experiment Station, Inc., (CEESI)Nunn, Colorado USA
MEASUREMENT UNCERTAINTY
When a measurement is made there are two im-
portant values associated with the result of the mea-
surement process. The first is the numerical value of
the variable being measured, the second is the un-
certainty associated with that numerical value. This
section provides an introduction to measurement un-
certainty by briefly describing the paper standards
and terminology.
Paper Standards
The uncertainty analysis procedure commonly used
with flow measurement processes is described in the
standards ANSI/ASME MFC-2M1, ANSI/ASME
PTC 19.12, and ISO 51683. All three of these stan-dards are based on an uncertainty analysis method
developed for comparing results of rocket engine
tests4.
A document more recently published by ISO de-
scribes an uncertainty analysis procedure originally
developed to compare the test data of different cali-
bration laboratories5. From a practical point of view,
the newer ISO method does not differ significantly
from the older method of Ref. 1-3. It has been
adopted by the National Institute of Standards and
Technology (NIST) as well as a number of large cor-
porations as a formal policy6. In addition, the latest
revision of the standard ANSI/PTC 19.17includes
the ISO method. It is likely that future revisions of
flow measurement uncertainty standards will adopt
the ISO method. This paper therefore teaches the
application of that method.
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
2/12
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
3/12
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
4/12
222222 41.006.020.014.026.0 ,,,,'Vbu
results in uVb
= 0.54 % for the ultrasonic meter.
The expanded uncertainty5 is the product of stan-
dard uncertainty and a coverage factor. That prod-
uct defines the confidence level. Most measurements
are stated at 95% confidence with a coverage factor
of 2.0. In the present example the expanded uncer-
tainties are 2.00.68 % = 1.36 % for the turbine meter
and 2.00.54 % = 1.08 % for the ultrasonic meter.
Using a coverage factor of 2.0 in the present ex-
ample is a simplifying assumption. A different value
may be required in a real uncertainty analysis. The
details represent an advanced topic beyond the scope
of this paper.
PRESSURE MEASUREMENT
The hypothetical pressure transmitter is config-
ured as follows:
Upper Range Limit (URL) = 2500 psig
Span = 1400 psig (20 mA)
Zero = 500 psig (4 mA)
Measured Pressure = 1054 psig (13.85 mA)
The uncertainty in pressure measurement is made
up of six components:
1. Combined Performance Specification
2. Stability
3. Ambient Temperature Effect
4. Calibration Process
5. Barometric Pressure
6. Data Acquisition
Determining numerical values for the six compo-
nents is discussed below. The behavior of four of the
components is shown in Figure 1. The x-axis is pres-
sure, the y-axis is the uncertainty expressed as a per-
cent of reading. The uncertainty decreases as the
pressure increases, this is due to full scale effects.
Such effects are present in all measurement pro-
cesses. In some processes the relative magnitude may
be very small and the effect may not be apparent.
The components of uncertainty for pressure mea-
surement are combined using Eq. 4, the units of u
are psi. Combining the numerical values:
222222 99.012.047.018.222.1 ,,,,'Pfu
results in uPf
= 2.73 psig which is 0.26% of 1068 psia.
The uncertainty must be expressed as a percent of
absolute (psia) not gage (psig) units because the pres-
sure value is used to calculate compressibility.
Combined Performance Specification
In the present example the vendor claims the ac-
curacy is 0.15% of span, the term accuracy is
said to include the four components listed below.
Regardless of vendor terminology, the user should
make sure that all four components are included.
1. Repeatability
2. Reproducibility
3. Linearity
4. Hysteresis
In the present example the span is fixed and so
this uncertainty component can be calculated in units
of psig. Doing so results in:
psig1400
%100
%15.0span%15.0 -!
"
#$
%
&.'.'U
= 2.10 psig [5a]
u= 0.58(2.10)=1.22 psig [5b]
0.0
0.2
0.4
0.6
0.8
500 700 900 1100 1300 1500
Pressure [psig]
StandardUncertainty[%
] Total
Stability
Accuracy
Data Acq
Amb Temp
Figure 1: Pressure Measurement
Uncertainty Components
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
5/12
Repeatability and Reproducibility
Repeatabilityis the ability of the transmitter to
measure a constant pressure over a short period
of time. Reproducibility is the ability of the trans-
mitter to measure the same pressure subject to varia-
tions that occur over a long period of time. A short
period of time might be measured in minutes while a
long period of time might be months.
Linearity
The sensitivity is the ratio of pressure transmit-
ter input (psi) to output (mA). In the present example
the sensitivity is 56.25 [psi/mA]. The linearity de-
fines the consistency of the sensitivity value over the
pressure range. Stated differently, the linearity mea-
sures the degree to which the input and output arerelated by a straight line.
Hysteresis
This effect accounts for a dependence of the out-
put on the direction change of the input. An example
would be the backlash present in a gearing system. A
good machinist will eliminate backlash by first back-
ing out the tool and then moving it into the part. In a
pressure transmitter the hysteresis arises from the
mechanical properties of the sensor. Hysteresis con-tributes measurement uncertainty because the direc-
tion of change of a measureand prior to a measure-
ment is generally unknown.
Stability
The vendor claims the stability is 0.25% of
URL. Calculating uncertainty in pressure units re-
sults in:
psig2500%100
%25.0
URL%25.0 -!"
#
$%
&
.'.'U
= 3.75 psig [6a]
u= 0.58(3.75) =2.18 psig [6b]
Ambient Temperature Effect
It has been proposed that all measuring devices
are first thermometers. Every instrument exhibits
sensitivity to changes in ambient temperature. In a
pressure transmitter the ambient temperature effects
the mechanical properties of the elastic element. The
uncertainty component accounts for the magnitude
of the effect. In the present example the vendor states
an uncertainty of:
) *span%35.0URL%45.0 ,.'U [7]
for every 100F change in the ambient temperature.
In the present example the pressure transducer is
installed in an environment where the temperature
variation is controlled to within 2.5F which repre-
sents a total potential temperature change of 5F. The
uncertainty due to the ambient temperature effect
therefore becomes:
) * !!
"
#
$$
%
&-,.'
F100
F5span%35.0URL%45.0
!
!
U
= 0.81 psig [8a]
u= 0.58(0.81) = 0.47 psig [8b]
Calibration Process
In the present example the role of the calibration
process is to maintain the transmitter within the origi-
nal performance specifications. No additional uncer-tainty is assumed to be contributed by this compo-
nent.
Barometric Pressure
In the present example a fixed value of baromet-
ric pressure is assumed. This value is calculated based
on the elevation of the meter station. It is assumed
that the barometric pressure can vary by 0.2 psia
due to meteorological conditions. This variation rep-
resents an uncertainty in the fixed value
Data Acquisition
The 4-20 mA transmitter output is converted into
engineering units by a flow computer. This process
adds uncertainty in the conversion of the analog sig-
nal to digital. The vendor claims an accuracy of
0.1% of full scale. The term %full scale from
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
6/12
the computer vendor is assumed to mean the same
as %span from the transducer vendor. The con-
version process to pressure units is similar as that
used with Eq. 5:
psig1400%100
%15.0
span%10.0 -!"
#
$%
&
.'.'U
= 1.71 psig [9a]
u= 0.58(1.71) =0.99 psig [9b]
TEMPERATURE MEASUREMENT
The hypothetical temperature transmitter is con-
figured as follows:
Span = 100 F
Zero = 0 F
Measured Temperature = 50 F (12 mA)
The uncertainty in temperature measurement is
made up of the following components:
1. Combined Performance Specification
2. Stability
3. Ambient Temperature Effect
4. Calibration Process
5. RTD Probe
6. Self Heating
7. Heat Transfer Effects
8. Data Acquisition
The determination of numerical values is described
below. Combining the values:
) * 222222 59.001.037.003.015.02 ,,,,'Tfu
results in uTf
= 0.73 F which is 0.14% of 510 R. The
uncertainty must be expressed as a percent of abso-
lute (R) units because the temperature value is used
to calculate compressibility.
Combined Performance Specification
In the present example the vendor claims the ac-
curacy is 0.25% of span:
F100%100
%25.0span%25.0 -!
"
#$%
&.'.'U
= 0.25F [10a]
u= 0.58(0.25)=0.15F [10b]
The term accuracy is assumed to include the
same four components as discussed with the pres-
sure measurement.
Stability
In the present example the vendor claims the sta-
bility is 0.25% of span. The standard uncertainty is u
= 0.15 F.
Ambient Temperature Effect
The vendor states an uncertainty of:
0.25%span)F4.0( ,.'U [11]
for every 50F change in the ambient temperature.
In the present example the temperature transducer
is installed in an environment where the temperature
variation is controlled to within 2.5F which repre-
sents a total potential temperature change of 5F. The
uncertainty due to the ambient temperature effecttherefore becomes:
) * !!
"
#
$$
%
&-,.'
F50
F5span%25.0F4.0
!
!
U
= 0.05 F [14a]
u= 0.58(0.05) = 0.03 F [14b]
Calibration Process
In the present example the role of the calibration
process is to maintain the transmitter within the origi-nal performance specifications. No additional uncer-
tainty is assumed to be contributed by this compo-
nent.
RTD Probe
The vendor claims an uncertainty that varies with
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
7/12
temperature. In the current example that value is U
= 0.63 F (u= 0.37 F)
Self Heating
When current flows through a resistor heat is gen-
erated. The heat results in a potential measurement
offset that is estimated as uncertainty. In the present
example this effect is assumed to be included in the
RTD probe uncertainty.
Heat Transfer Effects
The transmitter output indicates the temperature
of the tip of the temperature sensor RTD. The de-
sired value is the temperature of the flowing gas. If
there is a temperature difference between the flow-
ing gas and the pipe wall heat transfer will result.Heat transfer is present between the RTD and pipe
wall due to conduction through the thermowell as well
as direct radiation. The result is to drive the RTD
temperature closer to the pipe wall temperature. Heat
transfer due to convection is present between the
thermowell and the flowing gas. The convection will
drive the RTD temperature toward that of the flow-
ing gas. The difference between the RTD reading
and the flowing gas temperature (Tht) is a measure-
ment error, the magnitude must be estimated as an
uncertainty component.
The numerical value of Tht
is a function of gas
density, gas velocity, and thermowell geometry as well
as the difference between flowing gas and pipe wall
(Tgw
)10. The relationship between Tht
and gas ve-
locity is shown in Figure 2 based on the following
conditions:
thermowell: 0.5 OD, 0.25 ID, 3 length withinpipe
pipe wall temperature = 60 F and 80 F
gas temperature = 50 F
gas pressure = 1068 psia
In the present example it is assumed that the gas
velocity is 20 ft/s and the pipe wall temperature is 80
F. Under such conditions Tht = 0.02 F (u= 0.01
F).
Data Acquisition
The 4-20 mA transmitter output is converted intoengineering units by the same flow computer as used
for pressure measurement. The standard uncertainty
is u = 0.59 F.
COMPRESSIBILITY
The hypothetical gas composition used in the
present example is contained in column 2 of Table 1.
The process of determining uncertainty in calculated
compressibility is made up of four components:
1. State Equation
2. Chromatograph Repeatability and Reproducibil-
ity
3. Chromatograph Calibration Standard
4. Sensitivity Coefficients
The determination of numerical values is described
below. Combining the values:
222222 002.006.0,19.006.0 ,',' ZbZf uu
result in uZf= 0.20 % and uZb= 0.06 %.
State Equation
The AGA 8 equation of state11provides a value
of compressibility given inputs of pressure, tempera-
ture and gas composition. The uncertainty in the state
equation is U= 0.1% (u= 0.06%).
Figure 2: Heat Transfer Effects in
Temperature Measurement
0.1
1
10
1 10Velocity [ft/s ]
TemperatureDifference,/
Tht
[F]
/Tgw= 30F
/Tgw= 10F
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
8/12
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
9/12
VOLUME AT FLOWING CONDITIONS
The volume flow is measured using either a tur-
bine or ultrasonic meter. The uncertainties for using
these meters are specified by the AGA 714and AGA
915standards. For operation at higher flowrates the
uncertainties are 1.0% (u= 0.58%) for the turbine
meter and 0.7% (u = 0.41%) for the ultrasonic
meter. Performance of either meter at lower flowrates
will have higher uncertainties, the low flow condi-
tions are not considered in this example.
REDUCING THE UNCERTAINTY
The uncertainty values for all sixteen components
for turbine meter based measurement are contained
in Table 2. Column 2 shows uncertainty components
expressed as percentages of mean values. Column 3show the percent contribution to combined uncertainty
made by each component, they are calculated based
on %100
2
-!!
"
#
$$
%
&
y
xi
u
u. The summary in Table 2 is a pow-
erful tool in allocating measurement resources based
on measurement uncertainty. The turbine meter, for
example, is the largest contributor to uncertainty. It
represents the best return on an investment intended
to reduce uncertainty. The ambient temperature ef-
fects, on the other hand, contribute very little uncer-tainty. The overall uncertainty cannot be reduced by
investing in reduced uncertainty in ambient tempera-
ture effects.
The current example concludes with a hypotheti-
cal calibration of the turbine meter. The calibration
data are contained in Figure 3, the y-axis is percent
shift in K Factor from a nominal values. The data are
well within AGA 7 specifications, the uncertainty can
be reduced as a result. The fitted curve is slightly
nonlinear, the K Factor changes with flowrate. Thiswill require a more complex flow computer algorithm.
The solid lines in Fig. 3 represent the 95% confidence
interval, 95% of the data are within the interval. The
interval width is 0.2% (u = 0.1%) which accounts
for random effects present during the calibration pro-
cess. The uncertainty of the hypothetical calibration
facility flowrate measurement is 0.25% (u=0.13%)
which accounts for systematic effects associated with
the facility itself. In the current example these two
components make up the uncertainty associated with
the turbine meter measurement. The recalculated un-
certainty is uVb
= 0.39%, the details are contained in
Table 3. Having reduced the turbine meter uncer-
tainty, it can be seen that other components now domi-
nate the uncertainty. The gas chromatograph and pres-
sure transducer stability are two examples.
GENERAL DISCUSSION
It is important to emphasize that the analysis pre-
sented above is intended as an example only. Appli-
cable numerical values need to be determined for a
real uncertainty analysis. In addition, several poten-
tial uncertainty components have been neglected in
the interest of simplifying the example. These include,but are not limited to:
flowmeter installation effects
gas sampling
wet gas
pulsating flow
Considerable analysis was required to estimate
uncertainties of components that ended up contribut-
ing very little uncertainty. Two examples are the heat
transfer and the chromatographic analysis determi-nation of base compressibility. Unfortunately, the
magnitude of an uncertainty component is unknown
until the analysis is complete.
Figure 3: Turbine Meter Calibration
Data
-0.6
-0.3
0.0
0.3
0.6
20 40 60 80 100
Flowrate [%max]
KFactorShift[%
]
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
10/12
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
11/12
tnenopmoCdradnatS
]%[ytniatrecnU ]%[noitubirtnoC
erusserP
noitacificepSecnamrofrePdenibmoC 411.0 08.2
ytilibatS 402.0 59.8
tceffEerutarepmeTtneibmA 440.0 24.0
erusserPcirtemoraB 110.0 30.0
noitisiuqcAataD 390.0 58.1
erutarepmeT
noitacificepSecnamrofrePdenibmoC 920.0 91.0
ytilibatS 920.0 91.0
tceffEerutarepmeTtneibmA 600.0 10.0
eborPDTR 370.0 31.1
stceffErefsnarTtaeH 200.0 < 10.0
noitisiuqcAataD 611.0 88.2
ytilibisserpmoCesaB
noitauqEetatS 060.0 77.0
hpargotamorhC 200.0 < 10.0
ytilibisserpmoCgniwolF
noitauqEetatS 060.0 77.0
hpargotamorhC 091.0 67.7
emuloVgniwolF
eulaV7AGA 085.0 72.27
Table 2: First Summary of Uncertainty Components
-
8/11/2019 Flow Measurement Uncertainty - Analysis Based on Field Grade Components
12/12
tnenopmoC
dradnatS
]%[ytniatrecnU ]%[noitubirtnoC
erusserP
noitacificepSecnamrofrePdenibmoC 411.0 34.8
ytilibatS 402.0 39.62
tceffEerutarepmeTtneibmA 440.0 52.1
erusserPcirtemoraB 110.0 80.0
noitisiuqcAataD 390.0 55.5
erutarepmeT
noitacificepSecnamrofrePdenibmoC 920.0 65.0ytilibatS 920.0 65.0
tceffEerutarepmeTtneibmA 600.0 20.0
eborPDTR 370.0 04.3
stceffErefsnarTtaeH 200.0 < 10.0
noitisiuqcAataD 611.0 56.8
ytilibisserpmoCesaB
noitauqEetatS 060.0 33.2
hpargotamorhC 200.0 < 10.0
ytilibisserpmoCgniwolF
noitauqEetatS 060.0 33.2
hpargotamorhC 091.0 33.32
emuloVgniwolF
yrotarobaL 521.0 01.01
ytilibicudorpeRdnaytilibataepeR 001.0 64.6
Table 3: Second Summary of Uncertainty Components