focus on hemodynamic monitoring and circulatory assist devices (relates to chapter 66, “nursing...
TRANSCRIPT
Focus on Hemodynamic Monitoring and Circulatory Assist Devices
(Relates to Chapter 66, “Nursing Management: Critical Care,” in the textbook)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
2
Hemodynamic Monitoring
Measurement of pressure, flow, and oxygenation within the cardiovascular system
Includes invasive and noninvasive measurements
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
3
Hemodynamic Monitoring
Invasive and noninvasive measurements Systemic and pulmonary
arterial pressures Central venous pressure (CVP) Pulmonary artery wedge
pressure (PAWP) Cardiac output (CO)/cardiac
index (CI)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
4
Hemodynamic Monitoring
Invasive and noninvasive measurements (cont’d) Stroke volume (SV)/stroke
volume index (SVI) O2 saturation of arterial
blood (SaO2) O2 saturation of mixed
venous blood (SvO2)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
5
Hemodynamic MonitoringGeneral Principles
CO: volume of blood pumped by heart in 1 minute
CI: CO adjusted for body size SV: volume ejected with each
heartbeat SVI: SV adjusted for body
size
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
6
Hemodynamic MonitoringGeneral Principles
Systemic vascular resistance (SVR) and pulmonary vascular resistance (PVR) Opposition to blood flow by
systemic and pulmonary vasculature
Preload, afterload, and contractility determine SV.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
7
Hemodynamic MonitoringGeneral Principles
Preload: volume of blood within ventricle at end of diastole
Afterload: forces opposing ventricular ejection Systemic arterial pressure Resistance offered by aortic
valve Mass and density of blood to be
movedCopyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
8
Hemodynamic MonitoringGeneral Principles
Contractility: strength of ventricular contraction
PAWP: measurement of pulmonary capillary pressure; reflects left ventricular end-diastolic pressure under normal conditions
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
9
Hemodynamic MonitoringGeneral Principles
CVP: right ventricular preload or right ventricular end-diastolic pressure under normal conditions, measured in right atrium or in vena cava close to heart
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
10
Principles of Invasive Pressure Monitoring
Equipment must be referenced and zero-balanced to environment, and dynamic response characteristics optimized.
Referencing: positioning transducer so zero reference point is at level of atria of heart or phlebostatic axis
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Components of Pressure Monitoring System
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc. 11
Fig. 66-3. Components of a pressure monitoring system. The cannula, shown entering the radial artery, is connected via pressure (nondistensible) tubing to the transducer. The transducer converts the pressure wave into an electronic signal. The transducer is wired to the electronic monitoring system, which amplifies, conditions, displays, and records the signal. Stopcocks are inserted into the line for specimen withdrawal and for referencing and zero-balancing procedures. A flush system, consisting of a pressurized bag of intravenous fluid, tubing, and a flush device, is inserted into the line. The flush system provides continuous slow (approximately 3 mL/hr) flushing and provides a mechanism for fast flushing of lines.
12
Identification of the Phlebostatic Axis
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-4. Identification of the phlebostatic axis. A, Phlebostatic axis is an external landmark used to identify the level of theatria in the supine patient. The phlebostatic axis is defined as the intersection of two imaginary lines: one drawn vertically through the fourth intercostal space at the sternum, and another drawn horizontally through the midchest, halfway between the outermost anterior and outermost posterior points of the chest. B, As the backrest of the supine patient is elevated, the phlebostatic axis remains at the same anatomic location, becoming progressively elevated from the floor. The zero reference point must be repositioned with changes in backrest elevation, in order to keep it at the phlebostatic level.
13
Principles of Invasive Pressure Monitoring
Zeroing: confirms that when pressure within system is zero, monitor reads zero During initial setup of
arterial line Immediately after insertion
of arterial line
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
14
Principles of Invasive Pressure Monitoring
Zeroing (cont’d) When transducer has
been disconnected from pressure cable or pressure cable has been disconnected from monitor
When accuracy of values is questioned
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
15
Principles of Invasive Pressure Monitoring
Optimizing dynamic response characteristics involves checking that equipment reproduces, without distortion, a signal that changes rapidly.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
16
Principles of Invasive Pressure Monitoring
Optimizing dynamic response (cont’d) Perform dynamic response
test (square wave test) every 8 to 12 hours and
When system is opened to air
When accuracy of values is questioned
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
17
Dynamic Response Test (Square Wave Test)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-5. Optimally damped system. Dynamic response test (square wave test) using the fast flush system: normal response.
18
Types of Invasive Pressure Monitoring Continuous arterial
pressure monitoring Acute
hypertension/hypotension Respiratory failure Shock Neurologic shock
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
19
Types of Invasive Pressure Monitoring Continuous arterial
pressure monitoring (cont’d) Coronary interventional
procedures Continuous infusion of
vasoactive drugs Frequent ABG sampling
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
20
Arterial Pressure Monitoring High- and low-pressure alarms
based on patient’s status Measure at end of expiration. Risks/complications
Hemorrhage Infection Thrombus formation Neurovascular impairment Loss of limb
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
21
Arterial Pressure Monitoring
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-6. A, Simultaneously recorded electrocardiogram (ECG) tracing and B, systemic arterial pressuretracing. Systolic pressure is the peak pressure. The dicrotic notch indicates aortic valve closure. Diastolicpressure is the lowest value before contraction. Mean pressure is the average pressure over timecalculated by the monitoring equipment.
22
Arterial Pressure Monitoring Continuous flush irrigation
system Delivers 3 to 6 mL of
heparinized saline per hour Maintains line patency Limits thrombus formation
Assess neurovascular status distal to arterial insertion site hourly.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
23
Pulmonary Artery Pressure Monitoring Guides management of patients
with complicated cardiac, pulmonary, and intravascular volume problems. PA diastolic (PAD) pressure and
PAWP: indicators of cardiac function and fluid volume status
Monitoring PA pressures allows therapeutic manipulation of preload.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
24
Pulmonary Artery Pressure Monitoring PA flow-directed catheter
Distal lumen port in PA Samples mixed venous blood
Thermistor lumen port near distal tip Monitors core temperature Thermodilution method
measuring CO
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
25
PA Catheter
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-7. Pulmonary artery (PA) catheter. A, Illustrated catheter has five lumens. When properly positioned,the distal lumen exit port is in the PA and the proximal lumen ports are in the right atrium and right ventricle.The distal and one of the proximal ports are used to measure PA and central venous pressures, respectively.A balloon surrounds the catheter near the distal end. The balloon inflation valve is used to inflate the balloonwith air to allow reading of the pulmonary artery wedge pressure. A thermistor located near the distal tipsenses PA temperature and is used to measure thermodilution cardiac output when solution cooler thanbody temperature is injected into a proximal port. B, Photo of an actual catheter.
26
Pulmonary Artery Pressure Monitoring Additional lumina
Right atrium or right atrium and right ventricle
Right atrium port Measurement of CVP Injection of fluid for CO measurement
Blood sampling
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
27
Pulmonary Artery Pressure Monitoring Additional lumina (cont’d)
Second proximal port Infusion of fluids and drugs Blood sampling
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
28
Pulmonary Artery Pressure Monitoring When measurements are
obtained PA: at end expiration PAWP: by inflating balloon with
only enough air until PA waveform changes to a PAWP waveform
Balloon should be inflated slowly and for no more than four respiratory cycles or 8 to 15 seconds.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
29
PA Waveforms During Insertion
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-9. Change in pulmonary artery pressure (PAP) waveform to pulmonary artery wedge pressure(PAWP) waveform with balloon inflation. The balloon is inflated while observing the bedside monitor forchange in the waveform. Balloon inflation (arrow) in patient with a normal PAWP.
30
Effect of Overinflated Balloon
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-10. Balloon inflation (arrow) in patient with elevated wedge pressure. Overwedging of balloon (balloon has been overinflated). The danger of overinflating the balloon is that the pulmonary artery (PA)vessel may rupture from the pressure of the balloon. PAP, Pulmonary artery pressure.
31
Central Venous Pressure Monitoring Measurement of right
ventricular preload Obtained from
PA catheter using one of the proximal lumina
Central venous catheter placed in internal jugular or subclavian vein
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
32
Central Venous Pressure Waveforms
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-11. Cardiac events that produce the central venous pressure (CVP) waveform with a, c, and v waves.The a wave represents atrial contraction. The x descent represents atrial relaxation. The c wave represents the bulging of the closed tricuspid valve into the right atrium during ventricular systole. The v wave representsatrial filling. The y descent represents opening of the tricuspid valve and filling of the ventricle.
33
Measuring Cardiac Output
Intermittent bolus thermodilution method
Continuous cardiac output method
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
34
Measuring Cardiac Output
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-12. Normal cardiac output curve. Cardiac output is calculated from the temperature change in the pulmonary artery when a fixed volume and known temperature of a solution is injected into the proximal port in the right atrium. The nurse should observe the curve during injection to make sure that it is smooth.
35
Measuring Cardiac Output
SVR, SVRI, SV, and SVI can be calculated when CO is measured. ↑ SVR
Vasoconstriction from shock Hypertension ↑ release or administration of epinephrine or other vasoactive inotropes
Left ventricular failure
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
36
Measuring Cardiac Output
SVR, SVRI, SV, and SVI calculated (cont’d) ↓ SVR
VasodilationShock states and drugs that ↓ afterload
Changes in SV are becoming more important indicators of pumping status of heart than other parameters.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
37
Noninvasive Hemodynamic Monitoring Impedance cardiography (ICG)
Continuous or intermittent, noninvasive method of obtaining CO and assessing thoracic fluid status
Impedance-based hemodynamic parameters (e.g., CO, SV, SVR) are calculated from Zo, dZ/dt, MAP, CVP, and ECG.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
38
Noninvasive Hemodynamic Monitoring Major indications
Early signs and symptoms of pulmonary or cardiac dysfunction
Differentiation of cardiac or pulmonary cause of shortness of breath
Evaluation of causes and management of hypotension
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
39
Noninvasive Hemodynamic Monitoring Major indications (cont’d)
Monitoring after discontinuing a PA catheter or justification for inserting a PA catheter
Evaluation of pharmacotherapy Diagnosis of rejection
following cardiac transplantation
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
40
Venous Oxygen Saturation PA and CVP catheters can
measure oxygen saturation of hemoglobin in venous blood–mixed venous oxygen saturation. SvO2, ScvO2
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
41
Venous Oxygen Saturation SvO2/ScvO2 reflects
balance between oxygenation of arterial blood, tissue perfusion, and tissue oxygen consumption (VO2).
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
42
Venous Oxygen Saturation Normal SvO2/ScvO2 at rest
is 60% to 80%.
↓ in SvO2/ScvO2 ↓ arterial oxygenation Low CO Low hemoglobin level ↑ oxygen consumption or
extractionCopyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
43
Venous Oxygen Saturation ↑ in SvO2/ScvO2
May indicate clinical improvement (e.g., improved arterial oxygen saturation)
Worsening clinical condition (e.g., sepsis)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
44
Complications With PA Catheters Infection and sepsis
Asepsis for insertion and maintenance of catheter and tubing mandatory
Change flush bag, pressure tubing, transducer, and stopcock every 96 hours.
Air embolus (e.g., disconnection)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
45
Complications With PA Catheters Ventricular dysrhythmias
During PA catheter insertion or removal
If tip migrates back from PA to right ventricle
PA catheter cannot be wedged. May need repositioning
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
46
Complications With PA Catheters Pulmonary infarction or PA
rupture Balloon rupture (e.g.,
overinflation) Prolonged inflation Spontaneous wedging Thrombus/embolus formation
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
47
Preventing PA Rupture and Pulmonary Infarction
Check PA pressure waveforms often for signs of catheter occlusion, dislocation, or spontaneous wedging.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
48
Preventing PA Rupture and Pulmonary Infarction
Continually flush system with a slow infusion of heparinized saline solution.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
49
Noninvasive Arterial Oxygenation Monitoring
Pulse oximetry Continuous method of
determining arterial oxygenation: SpO2
Monitoring SpO2 may ↓ frequency of ABG sampling
Normally 95% to 100%
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
50
Noninvasive Arterial Oxygenation Monitoring
Pulse oximetry Monitoring SpO2 may ↓ frequency
of ABG sampling (cont’d) Measurements may be difficult if patients are hypothermic, receiving IV vasopressors, or experiencing hypoperfusion.
Alternate locations for placement of probe: forehead, earlobe
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
51
Nursing Management
Obtain baseline data. General appearance Level of consciousness Skin color/temperature Vital signs Peripheral pulses Urine output
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
52
Nursing Management
Correlate baseline data with data obtained from biotechnology (e.g., ECG; arterial, CVP, PA, and PAWP pressures; SvO2/ScvO2).
Single hemodynamic values are rarely significant.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
53
Nursing Management
Monitor trends and evaluate whole clinical picture.
Goals Recognize early clues. Intervene before problems
develop or escalate.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
54
Circulatory Assist Devices (CADs) Decrease cardiac work and
improve organ perfusion when drug therapy fails.
Provide interim support when Left, right, or both ventricles
require support while recovering from injury (e.g., myocardial infarction)
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
55
Circulatory Assist Devices (CADs) Provide interim support
when (cont’d) Heart requires surgical repair
and patient must be stabilized (e.g., ruptured septum)
Heart has failed and patient needs cardiac transplantation
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
56
Intraaortic Balloon Pump (IABP) Provides temporary
circulatory assistance ↓ afterload Augments aortic diastolic
pressure Outcomes
Improved coronary blood flow Improved perfusion of vital
organs
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
57
IABP Machine
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-13. Intraaortic balloon pump machine.
58
Intraaortic Balloon Pump (IABP) Consists of
Sausage-shaped balloon Pump that inflates and
deflates balloon Control panel for
synchronizing balloon inflation with cardiac cycle
Fail-safe features
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
59
IABP
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-14. Intraaortic balloon pump. A, During systole the balloon is deflated, which facilitates ejection of theblood into the periphery. B, In early diastole, the balloon begins to inflate. C, In late diastole, the balloon istotally inflated, which augments aortic pressure and increases the coronary perfusion pressure with the endresult of increased coronary and cerebral blood flow.
60
Intraaortic Balloon Pump (IABP) Referred to as
counterpulsation Timing of balloon inflation
is opposite to ventricular contraction.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
61
Intraaortic Balloon Pump (IABP) Complications of IABP
therapy Vascular injury
Dislodging of plaque Aortic dissection Compromised distal circulation
Thrombus and embolus formation
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
62
Intraaortic Balloon Pump (IABP) Complications of IABP
(cont’d) Mechanical complications
Improper timing of balloon inflation
↑ afterload ↓ CO Myocardial ischemia ↑ myocardial oxygen demand
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
63
Intraaortic Balloon Pump (IABP) To decrease risks of IABP
therapy, Obtain cardiovascular,
neurovascular, and hemodynamic assessments every 15 to 60 minutes, based on patient’s status
Keep patient immobile and limited to side-lying or supine position with HOB <45 degrees
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
64
Intraaortic Balloon Pump (IABP) Decreasing risks of IABP
(cont’d) Leg with IABP catheter must
not be flexed at hip to avoid kinking or dislodgement of catheter.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
65
Ventricular Assist Devices (VADs) Provide short and longer-
term support for failing heart Allow greater mobility than
IABP Inserted into path of flowing
blood to augment or replace action of ventricle
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
66
Ventricular Assist Devices (VADs) VADs can
Be implanted (e.g., peritoneum) or positioned externally
Provide biventricular support
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
67
Schematic Diagram of Biventricular Assist Device
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Fig. 66-15. Schematic diagram of a biventricular assist device.
68
Ventricular Assist Devices (VADs) Indications for VAD
therapy Extension of
cardiopulmonary bypass Failure to wean Postcardiotomy cardiogenic shock
Bridge to recovery or cardiac transplantation
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
69
Ventricular Assist Devices (VADs) Indications for VAD
therapy (cont’d) Patients with New York
Heart Association Classification IV who have failed medical therapy
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
70
Nursing Management Circulatory Assist Devices
Similar to care for patient with an IABP Observe patient for bleeding,
cardiac tamponade, ventricular failure, infection, dysrhythmias, renal failure, hemolysis, and thromboembolism.
Patient may be mobile and will require an activity plan.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
71
Nursing Management Circulatory Assist Devices
Goal Recovery through ventricular
improvement Heart transplantation Artificial heart implantation
Many patients will die or choose to terminate device, causing death.
Psychologic support for patient and caregiver is essential.
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
72
The pulmonary artery waveform of a patient with a pulmonary artery catheter is blunted. The nurse notifies the health care provider, recognizing that:
1. The balloon is overinflated.2. The catheter may be occluded by a thrombus.3. The catheter is wedged in a pulmonary capillary.4. The catheter has migrated from the pulmonary artery to the right ventricle.
Audience Response Question
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
Case Study
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc. 73
74
Case Study
78-year-old man is admitted to ICU in acute decompensated heart failure.
Pulmonary artery catheter and arterial lines inserted
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
75
Discussion Questions
1.Why is it important to keep the transducer at the level of the phlebostatic axis?
2.How can you determine if the blood pressure is accurate?
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
76
Discussion Questions
3. Why would you monitor the mean arterial pressure?
4.What measurements would be abnormal in someone with acute decompensated heart failure?
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.
77
Discussion Questions
5. If the PA/PAWP is elevated, what could this indicate?
6.What should the nurse do if the PA waveform looks wedged?
7.What purpose does thermodilution serve?
Copyright © 2011, 2007 by Mosby, Inc., an affiliate of Elsevier Inc.