Volumetric Gas Adsorption Apparatus for the Measurement of Physical Adsorption and Desorption Isotherms The adsorptive gas supply valve (3) is closed and the vacuum (2) and sample (7) valves are open allowing the manifold and sample tube to be evacuated. The sample tube is not in the cold bath, so the sample is at ambient temperature. When the necessary vacuum is achieved, valves 2 and 7 close and the cold bath is raised, cooling the sample to the analysis temperature. The illustration to the right represents a generic volumetric physical adsorption analyzer in its most elementary form. The critical components are: 1) Analysis manifold of accurately known volume and temperature 2) Vacuum system with valve to manifold 3) Source of adsorptive gas (typically, N 2 ) with valve to manifold 4) Pressure transducer and temperature sensor 5) Means for recording the signal from the transducer and temperature sensor 6) A sample tube of precisely known free or void-space 7) Sample tube connected to analysis manifold 8) Means to reduce the temperature of the sample when required, (typically to liquid nitrogen (LN 2 ) temperature). Valve 3 is opened momentarily to charge the manifold to a pressure (P m ) slightly above vacuum, preparing the instrument to dispense a dose of adsorptive onto the sample. The quantity of gas (n m ) in the manifold can be determined from the univer- sal gas law Preparation Charging the Manifold www.micromeritics.com Micromeritics Instrument Corporation One Micromeritics Dr., Norcross, GA 30093-1877, USA US Sales: 770.662.3633, International Sales: 770.662.3660 Customer Orders: 770.662.3636, Fax: 770.662.3696 Part # xxx/xxxxx/00 $42.00 Presented by Micromeritics Instrument Corporation Dosing Valve 7 is opened allowing some of the gas to enter the sample tube. Equilibration Some quantity of gas (n ads ) will be adsorbed by the sample and removed from the gas phase. Pressure is monitored un- til it stabilizes, indicating adsorption has equilibrated. The equilibration pressure (P e )is recorded. Quantity Adsorbed The quantity of gas (n e ) remaining in the combined manifold and sample tube volume (V m + V s ) can be calculated from the universal gas law. This is complicated by the vertical temperature profile in the sample tube, one portion essential- ly being at ambient temperature and another portion being at the temperature of the cold bath typically LN 2. . The calculation of n e is made traceable by a free-space mea- surement typically performed prior to the analysis and which characterizes the sample tube volume in regard to ‘warm’ and ‘cold’ volumes. Once n e is determined, the quantity of gas adsorbed by the sample at P e is n ads = n m - n e This establishes the point on the isotherm (P e , n ads ). Valve 7 closes and valve 3 opens, and the manifold is charged to a pressure slightly higher than P e after which the dosing and quilibration processes are repeated. This cycle continues until the analysis pressure is near satu- ration pressure at which time the complete adsorption iso- therm has been developed. The desorption isotherm is mea- sured by a step-wise reduction in pressure until the a low pressure over the sample is achieved. At that point, most of the physically adsorbed molecules will have been desorbed from the surface. n m = PmVm RT (3) Adsorptive Gas (nitrogen) and Valve (2) Vacuum Pump and Valve (7) Sample Valve (6) Sample Tube (evacuated) (8) Cold Bath Reservoir (down position) Sample ‘Warm’ Volume ‘Cold’ Volume N 2 3 2 Vacuum Pressure = P m 6 Manifold Volume = V m Sample Tube Volume = V s Pressure = P i LN 2 1) Preparation 3) Sample Dosing 2) Charging the Manifold (5) Signal Recorder N 2 (3) (2) Pressure = P e (6) and (1) Analysis Manifold (evacuated) Amorphous Silica Alumina, 11nm pores Adsorption Desorption Quantity adsorbed, std. cm 3 /g 400 200 0 100 300 0 1.00 0.50 0.25 0.75 p/p o Type IV isotherm: Nitrogen adsorption and desorption on amorphous silica alumina. Pressure Equilibration Vacuum Temperature Sensor and Pressure Transducer (4)

Upload: others

Post on 06-Jun-2020

25 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: for the Measurement of Physical Adsorption and Desorption ... · Adsorption Desorption Quantity adsorbed, std. cm 3 /g 400 200 0 100 300 0 0.25 0.50 0.75 1.00 p/po Type IV isotherm:

Volumetric Gas Adsorption Apparatusfor the Measurement of Physical Adsorption and Desorption Isotherms

The adsorptive gas supply valve (3) is closed and the vacuum (2) and sample (7) valves are open allowing the manifold and sample tube to be evacuated. The sample tube is not in the cold bath, so the sample is at ambient temperature.When the necessary vacuum is achieved, valves 2 and 7 close and the cold bath is raised, cooling the sample to the analysis temperature.

The illustration to the right represents a generic volumetric physical adsorption analyzer in its most elementary form. The critical components are:1) Analysis manifold of accurately known volume and temperature2) Vacuum system with valve to manifold3) Source of adsorptive gas (typically, N2) with valve to manifold4) Pressure transducer and temperature sensor5) Means for recording the signal from the transducer and temperature sensor6) A sample tube of precisely known free or void-space7) Sample tube connected to analysis manifold8) Means to reduce the temperature of the sample when required, (typically to liquid nitrogen (LN2) temperature).

Valve 3 is opened momentarily to charge the manifold to a pressure (Pm) slightly above vacuum, preparing the instrument to dispense a dose of adsorptive onto the sample. The quantity of gas (nm) in the manifold can be determined from the univer-sal gas law

Preparation

Charging the Manifold

www.micromeritics.comMicromeritics Instrument Corporation

One Micromeritics Dr., Norcross, GA 30093-1877, USAUS Sales: 770.662.3633, International Sales: 770.662.3660

Customer Orders: 770.662.3636, Fax: 770.662.3696 Part # xxx/xxxxx/00 $42.00

Presented by Micromeritics Instrument Corporation

Dosing

Valve 7 is opened allowing some of the gas to enter thesample tube.

Equilibration

Some quantity of gas (nads) will be adsorbed by the sample and removed from the gas phase. Pressure is monitored un-til it stabilizes, indicating adsorption has equilibrated. The equilibration pressure (Pe)is recorded.

Quantity Adsorbed

The quantity of gas (ne) remaining in the combined manifold and sample tube volume (Vm + Vs) can be calculated from the universal gas law. This is complicated by the vertical temperature profile in the sample tube, one portion essential-ly being at ambient temperature and another portion being at the temperature of the cold bath typically LN2..The calculation of ne is made traceable by a free-space mea-surement typically performed prior to the analysis and which characterizes the sample tube volume in regard to ‘warm’ and ‘cold’ volumes. Once ne is determined, the quantity of gas adsorbed by the sample at Pe is

n ads = nm- ne

This establishes the point on the isotherm (Pe, nads). Valve 7 closes and valve 3 opens, and the manifold is charged to a pressure slightly higher than Pe after which the dosing and quilibration processes are repeated.This cycle continues until the analysis pressure is near satu-ration pressure at which time the complete adsorption iso-therm has been developed. The desorption isotherm is mea-sured by a step-wise reduction in pressure until the a low pressure over the sample is achieved. At that point, most of the physically adsorbed molecules will have been desorbed from the surface.

nm = PmVm RT

(3) Adsorptive Gas (nitrogen) and Valve

(2) Vacuum Pump and Valve

(7) Sample Valve

(6) Sample Tube (evacuated)

(8) Cold Bath Reservoir (down position)

Sample

‘Warm’ Volume

‘Cold’ Volume

N2 3

2

Vacuum

Pressure = Pm

6Manifold Volume = Vm

Sample Tube Volume = VsPressure = Pi

LN2

1) Preparation

3) Sample Dosing

2) Charging the Manifold

(5) Signal Recorder

N2 (3)

(2)

Pressure = Pe(6)

and

(1) Analysis Manifold (evacuated)

Amorphous Silica Alumina, 11nm pores

Adsorption Desorption

Qua

ntity

ads

orbe

d, s

td. c

m3/g

400

200

0

100

300

0 1.000.500.25 0.75

p/po

Type IV isotherm:

Nitrogen adsorption and desorption on amorphous silica alumina.

Pressure Equilibration

Vacuum

Temperature Sensor and Pressure Transducer

(4)