friction and wear measurments of bovine articular
TRANSCRIPT
FRICTION AND WEAR MEASURMENTS OF BOVINE ARTICULAR CARTILAGE AGAINST NON-NATIVE MATERIALS
Sevan R. Oungoulian
Submitted in partial fulfillment of the
requirements for the degree
of Doctor of Philosophy
in the Graduate School of Arts and Sciences
COLUMBIA UNIVERSITY
2015
©2015
Sevan R. Oungoulian
All rights reserved
ABSTRACT
FRICTION AND WEAR MEASURMENTS OF BOVINE ARTICULAR CARTILAGE AGAINST NON-
NATIVE MATERIALS
Sevan R. Oungoulian
The three studies reported in this thesis investigate the friction and wear properties of articular cartilage.
The aim of the first study presented was to examine the functional properties and biocompatibility
of glutaraldehyde-fixed bovine articular cartilage for potential use as a resurfacing material in joint
arthroplasty. Treated cartilage disks were fixed over a range of glutaraldehyde concentrations and
incubated, along with an untreated control group, at 37 °C for up to 28 days. The equilibrium
compressive modulus increased nearly twofold in the treated samples when compared to day 0 control,
and maintained that property value from day 1 to day 28; the minimum friction coefficient did not change
significantly with fixation and incubation time, whereas the time constant for the frictional response
decreased twofold at most. Live explants co-cultured with fixed explants showed no qualitative difference
in cell viability over 28 days of incubation. Cartilage-on-cartilage frictional measurements were performed
under the configuration of migrating contact for a subset of treated explants over a period of 28 days
exhibited either no significant difference or slightly lower friction coefficient values than the untreated
control group. These results suggest that a properly titrated glutaraldehyde treatment can reproduce the
desired functional properties of native articular cartilage and maintain these properties for at least 28 days
in-vitro.
The aim of the second study was to determine whether the latest-generation particle analyzers
are capable of detecting cartilage wear debris generated during in vitro loading experiments that last 24 h
or less, by producing measurable content significantly above background noise levels otherwise
undetectable through standard biochemical assays. Immature bovine cartilage disks were tested against
glass using reciprocal sliding under unconfined compression creep for 24 h. Control groups were used to
assess various sources of contamination. Results demonstrated that cartilage samples subjected to
frictional loading produced particulate volume significantly higher than background noise and
contamination levels at all tested time points. The particle counter used was able to detect very small
levels of wear), whereas no significant differences were observed in biochemical assays for collagen or
glycosaminoglycans among any of the groups or time points.
The aim of the final study was to measure the wear response of immature bovine articular
cartilage tested against glass or alloys used in hemiarthroplasties. Two cobalt chromium alloys and a
stainless steel alloy were selected for these investigations. The surface roughness of one of the cobalt
chromium alloys was also varied within the range considered acceptable by regulatory agencies.
Cartilage disks were tested in a configuration that promoted loss of interstitial fluid pressurization to
replicate conditions believed to occur in hemiarthroplasties. Results showed that considerably more
damage occurred in cartilage samples tested against smooth stainless steel and rough cobalt chromium
alloys compared to smooth glass, and smooth cobalt chromium alloys. The two materials producing the
greatest damage also exhibited higher equilibrium friction coefficients. Cartilage damage occurred
primarily in the form of delamination at the interface between the superficial tangential zone and the
transitional middle zone, with much less evidence of abrasive wear at the articular surface. These results
suggest that cartilage damage from frictional loading occurs as a result of subsurface fatigue failure
leading to the delamination. Surface chemistry and surface roughness of implant materials can have a
significant influence on tissue damage, even when using materials and roughness values that satisfy
regulatory requirements.
i
TABLE OF CONTENTS
LIST OF FIGURES ....................................................................................................................................... iv
LIST OF TABLES .......................................................................................................................................... v
Chapter 1: INTRODUCTION ................................................................................................................... 1
1.1 Cartilage Lubrication ..................................................................................................................... 1
1.1.1 Boundary Lubrication ................................................................................................................ 1
1.1.2 Interstitial Fluid Pressurization .................................................................................................. 1
1.2 Cartilage Wear .............................................................................................................................. 2
1.3 Hemiarthroplasty ........................................................................................................................... 2
Chapter 2: GLUTARALDEHYDE FIXED CARTILAGE AS A RESURFACING MATERIAL ................... 4
2.1 Introduction .................................................................................................................................... 4
2.2 Materials and Methods .................................................................................................................. 6
2.2.1 Study Design ............................................................................................................................. 6
2.2.2 Sample Harvest ......................................................................................................................... 6
2.2.1 Glutaraldehyde Treatment ........................................................................................................ 7
2.2.2 Mechanical Testing ................................................................................................................... 8
2.2.3 Biocompatibility Testing ............................................................................................................ 9
2.3 Results ........................................................................................................................................ 10
2.3.1 Study 2-1 ................................................................................................................................. 10
2.3.2 Study 2-2 ................................................................................................................................. 11
2.4 Discussion ................................................................................................................................... 11
2.5 Acknowledgements ..................................................................................................................... 13
2.6 Figures ........................................................................................................................................ 14
ii
Chapter 3: CARTILAGE WEAR DEBRIS MEASUREMENTS VIA PARTICLE ANALYZER ................ 23
3.1 Introduction .................................................................................................................................. 23
3.2 Materials and Methods ................................................................................................................ 23
3.2.1 Sample Harvest ....................................................................................................................... 23
3.2.2 Solution Preparation ................................................................................................................ 23
3.2.3 Mechanical Testing ................................................................................................................. 24
3.2.4 Particulate Analyzer ................................................................................................................ 24
3.2.5 Particulate Assay..................................................................................................................... 25
3.2.6 Particulate Imaging.................................................................................................................. 25
3.2.7 Statistical Analysis................................................................................................................... 26
3.1 Results ........................................................................................................................................ 26
3.1.1 Particulate Number.................................................................................................................. 26
3.1.2 Particulate Volume .................................................................................................................. 26
3.1.3 Particulate Assay and Imaging ................................................................................................ 26
3.2 Discussion ................................................................................................................................... 26
3.3 Acknowledgements ..................................................................................................................... 28
3.5 Figures ........................................................................................................................................ 29
Chapter 4: INFLUENCE OF BEARING MATERIAL AND ROUGHNESS ON CARTILAGE
DEGENERATION ....................................................................................................................................... 34
4.1 Introduction .................................................................................................................................. 34
4.2 Materials and Methods ................................................................................................................ 35
4.2.1 Material Preparation ................................................................................................................ 35
4.2.1 Sample Harvest ....................................................................................................................... 36
iii
4.2.2 Mechanical Testing ................................................................................................................. 36
4.2.3 Particulate Testing................................................................................................................... 37
4.2.4 Biochemical Testing ................................................................................................................ 38
4.2.5 Histological Staining ................................................................................................................ 38
4.3 Results ........................................................................................................................................ 38
4.3.1 Material Roughness Testing ................................................................................................... 38
4.3.2 Mechanical Testing ................................................................................................................. 38
4.3.3 Particulate Testing................................................................................................................... 39
4.3.4 Biochemical Testing ................................................................................................................ 39
4.3.5 Histological Staining ................................................................................................................ 39
4.4 Discussion ................................................................................................................................... 39
4.5 Acknowledgements ..................................................................................................................... 40
4.6 Figures ........................................................................................................................................ 41
Chapter 5: CONCLUSION .................................................................................................................... 48
5.1 Motivation .................................................................................................................................... 48
5.2 Glutaraldehyde Fixed Cartilage As A Resurfacing Material ........................................................ 48
5.3 Cartilage Wear Debris Measurements Via Particle Analyzer ..................................................... 49
5.4 Influence of Artificial Bearing Material on Cartilage Degeneration ............................................. 49
Chapter 6: REFERENCES .................................................................................................................... 51
iv
LIST OF FIGURES
Figure 2-1: Schematic representation of bovine calf articular cartilage harvest location ........................... 14
Figure 2-2: Representative biocompatibility culture dish with viable and glutaraldehyde treated cartilage 15
Figure 2-3: Representative FEBio curve-fit of experimental stress-relaxation test ..................................... 16
Figure 2-4: Summary of mechanical properties for Study 2-1 .................................................................... 17
Figure 2-5: Qualitative and quantitative biocompatibility from chondrocyte viability ................................... 18
Figure 2-6: Frictional properties for Study 2-2 ............................................................................................ 19
Figure 2-7: India ink staining of representative Study 2-2 tibial plateau strips ........................................... 20
Figure 2-8: Intact immature bovine knee condyles ..................................................................................... 21
Figure 2-9: Potential clinical approach for bioprosthetic arthroplasties ...................................................... 22
Figure 3-1: Schematic for TEST, CTRL, ENVR (cross-sectional views), and BASE .................................. 30
Figure 3-2: Wear particulate measurements for bovine articular cartilage explants................................... 31
Figure 3-3: Wear particulate confocal z stack and angled z stack.............................................................. 32
Figure 3-4: Particulate size (A) and volume (B) distribution for representative 24 h TEST solution .......... 33
Figure 4-1: Schematic representation of bovine calf articular cartilage harvest location ........................... 41
Figure 4-2: Frictional properties of cartilage against different counterface materials ................................. 42
Figure 4-3: Representative qualitative macroscopic comparison of cartilage tissue .................................. 43
Figure 4-4: Summary of cartilage mechanical properties and wear measurements .................................. 44
Figure 4-5: Summary of cartilage biochemical properties .......................................................................... 45
Figure 4-6: Representative histological images of top region ..................................................................... 46
v
LIST OF TABLES
Table 3-1: Biochemical assay measurements of cartilage wear debris ...................................................... 29
Table 4-1: Roughness measurements for orthopaedic material platens .................................................... 47
vi
ACKNOWLEDGEMENTS
My graduate experience here at Columbia University has been extraordinarily rewarding and fulfilling.
There are many people who have contributed to and helped shape this experience.
I would like to express gratitude to Dr. Gerard Ateshian. To have worked under his guidance has
been a great privilege. His unwavering sense of academic rigor and discipline set a high bar that I have
strived to reach in my graduate career. His kindness, patience, and encouragement along the way
provided a model of managerial action. The persistence of his intellectual curiosity coupled with his desire
for deep understanding is quite extraordinary. Of much personal importance, I would like to thank him for
his trust in the exploration of new scientific ideas.
I would like to thank members of my doctoral committee for their valuable insights and comments
to this work: Dr. Clark Hung, Dr. Kristin Myers, Dr. Christopher Ahmad, and Dr. Jeffrey Kysar. In
particular, I would like to thank Dr. Clark Hung for his discussions and insight to tissue and biochemical
interactions at the cellular level. I have thoroughly enjoyed our discussions and know that this body of
research would not be complete without your valuable advice.
Dr. Michael Albro was an outstanding role model first as a graduate student and eventually as a
post-doctoral researcher at Columbia. Mike has been both a source of tremendous help, friendship, and
inspiration. I feel very fortunate for having been able to work with him.
I would like to thank my close friends and current colleagues, Brian Jones, Robert Nims, Alex
Cigan, Krista Durney, Chieh Hou, and Brandon Zimmerman. I will deeply miss working beside you and
wish you all the very best. I would also like to thank Vikram Rajan, Dr. Matteo Caligaris, Dr. Clare Canal,
Dr. Hui Chen, Charlie Yongpravat, Dr. Elizabeth Oswald, Andrea Tan, Sonal Sampat, Adam Nover,
Brendan Roach, Amy Silverstein, and Kyoko Yoshida, it has been a pleasure knowing and working with
all of you. Thanks to Keith Yeager for providing a great collaborative partner in the area of engineering
design, for granting me wide use of facilities, and for his support and access to valuable professional
opportunities. I have also been fortunate in having worked with tremendous undergraduate and graduate
student research volunteers who have taught me as much about being a mentor as I hope to have taught
them about rubbing together bits of cartilage. Thank you for making this experience at Columbia
memorable.
vii
My friends and family deserve so very much appreciation for all of their continued love and
support: to my parents Semik and Laura Oungoulian for their unyielding encouragement for success in all
of my pursuits, to Ani and Shaunt Oungoulian for their much appreciated sense of humor, and to Sarah
Dymkar Brandt, for her incredible warmth, love, and care.
1
CHAPTER 1: INTRODUCTION
1.1 Cartilage Lubrication
1.1.1 Boundary Lubrication
The primary function of articular cartilage is to serve as the bearing material in diarthrodial joints,
transmitting loads while minimizing friction and wear. The friction coefficient of cartilage has been
characterized extensively in the literature, using standard measurements of normal and tangential forces
acting across a sliding interface [1-8]. In the mid 20th century, the idea that cartilage tribology was
mediated by the mechanism of boundary lubrication was proposed by Sir John Charnley [9]. Boundary
lubricants are typically thought of as molecules that coat the surface of a bearing and effect a low friction
coefficient due to molecular-level repulsive forces [10]. Many investigators since have explored the
tribological function of different synovial fluid constituents, such as hyaluronan [11-17], lubricin [7, 8, 18-
22], and surface-active phospholipids [13, 14, 23-29]. Homology was discovered between lubricin [30, 31]
and superficial zone protein, a more recently identified molecule that is expressed by chondrocytes in the
superficial region of the tissue [23, 32]. This discovery suggests that cartilage constituents may also
contribute to boundary lubrication.
1.1.2 Interstitial Fluid Pressurization
The concept that pressurization of the interstitial fluid of articular cartilage could play a dominant role in
cartilage lubrication has been the subject of more recent theoretical and experimental work. Analysis has
helped formulate a theoretical framework that would embody this mechanism for friction coefficient
reduction. It has been proposed that the joint articular surfaces come in direct contact, but that the
pressurized interstitial fluid of this highly porous tissue supports most of the contact load. Cartilage
porosity is especially high at the articular surface, so the contact area over which collagen are in direct
contact with apposing collagen fibers is very low [33]; with load transfer between collagen and fluid, or
fluid and fluid, occurring over the bulk of the remaining contact area. This results in a very low friction
coefficient while the interstitial fluid pressure is elevated. The friction coefficient will rise as the portion of
2
the load supported by the pressurized interstitial fluid diminishes, and as the portion of the load supported
by direct contact between solid constituents increases. Agreement between theoretical predictions and
and experiment, much of which was performed previously in this lab, is very strong [33-38].
Important insight into the reason for low in-vivo friction can be gained from the investigation of this
interstitial fluid pressurization mechanism. It has since been shown that under conditions of steady motion
between curved articular cartilage surfaces, that interstitial fluid pressurization remains elevated [39].
Under conditions that promote a continuously migrating contact area, the friction coefficient remains low
[39]. This finding is significant when considered in the context that during normal function of diarthrodial
joints, conditions exist that promote migrating contact areas on at least one of the apposing articular
surfaces. This has been demonstrated in experimental studies of articular contact during normal activities
of daily living in many different diarthrodial joints [40-44].
1.2 Cartilage Wear
As mentioned previously, the friction coefficient of cartilage has been characterized extensively in the
literature [1-8]. However, characterization of cartilage wear and damage has proved to be more difficult,
and a standard method for measuring microscopic wear has yet to be agreed upon. Qualitative
observations of cartilage wear debris have been made [45-52]; however, fewer studies report quantitative
measurements of cartilage wear. The primary quantitative approaches proposed to date include
biochemical assaying of cartilage and test solutions [53-57], characterization of changing articular layer
thickness [54, 58-60], and changes in surface roughness [7, 57, 61-65]. Each of these methods has
limitations in determining the minute amounts of wear expected in the normal in-vivo environment.
Without a standard method for quantitative measurement, the complex relationship between tissue
damage and wear and friction coefficient is still poorly understood, and determining the underlying
physical and biological events that lead to these conditions remains elusive [3, 66-69]. It is generally
assumed that low friction is associated with low wear and damage, but as seen with the first generation of
artificial hips, this association is not straightforward [70-72].
1.3 Hemiarthroplasty
3
The surgical procedure in which half of a damaged synovial joint is replaced with a non-native bearing is
called hemiarthroplasty. According to the Centers for Disease Control, over 360,000 hemiarthroplasty
procedures are performed each year in the US, [73]. Hemiarthroplasties are less invasive, require less
surgical time and result in lower blood loss than total arthroplasties [74, 75], however the value over total
joint arthroplasty still subject to debate [74-82]. Some indications for hemiarthroplasty include fracture of
the femoral or humeral head [83-85] and fixation failure due to osteonecrosis [76, 86, 87], while
contraindications include disease or damage to the apposing joint surface [80]. Hemiarthroplasty patient
outcomes are sometimes less successful than total arthroplasties, and often require revision, due in part
to accelerated erosion of the articular surface [59, 88-92] although reasons for this acceleration in
damage is unclear.
Extending the insight gained by our understanding of the importance of migrating contact areas on
maintaining an elevated interstitial fluid pressure in articular cartilage, and resulting low friction coefficient,
to joint hemiarthroplasties might perhaps help explain this accelerated wear. It is possible that traditional
hemiarthroplasties compromise joint lubrication by reducing the migrating contact pattern.
4
CHAPTER 2: GLUTARALDEHYDE FIXED CARTILAGE AS A
RESURFACING MATERIAL
2.1 Introduction
Total joint arthroplasty is a commonly applied and generally effective surgical treatment for debilitating
arthritic conditions, accounting for over 1 million surgical procedures in the US alone during 2010
according to the CDC [73]. The development of an alternative resurfacing material to impermeable metals
or ultra-high molecular weight polyethylene remains an important engineering achievement that can
potentially improve the long-term clinical outcome of arthroplasty procedures by providing a more
cartilage-like bioprosthetic replacement [93, 94]. Recent theoretical developments and corroborating
experimental evidence support the idea that interstitial fluid pressurization of the porous native cartilage is
critical to the maintenance of low friction [38, 95, 96]. These findings suggest that joint resurfacing with a
porous biologically stable material has the potential to outperform traditional nonporous implant materials,
because of improved tribological properties. Though porous hydrogels and electrospun polymers have
been considered [63, 97-101], chemically fixed cartilage allografts or xenografts may exhibit better
functional properties [102-106].
Bioprosthetic heart valves have been successfully used clinically since the 1960s [107-109].
These valves typically consist of human or porcine heart valve, or bovine pericardial tissue that has been
chemically treated. Glutaraldehyde is most commonly used for fixation and sterilization, though it is
sometimes supplemented with alcohol or gamma radiation (when using low glutaraldehyde
concentrations) [110, 111]. Glutaraldehyde causes cross-linking of proteins in the valve tissue
extracellular matrix, which has been shown to alter mechanical properties [110, 112-115], though fixed
valves are able to maintain a satisfactory level of function when implanted, as evidenced from their
clinical success. Fixation under a pressure differential has also been advocated, in order to alter both the
configuration under which cross-linking occurs and the resulting stress-strain response of the fixed tissue
[113, 116-118].
5
Similar treatment has been advocated for musculoskeletal connective tissues [119-122], most
notably for the knee meniscus [103-105, 123-128]. These studies have similarly demonstrated that
glutaraldehyde treatment may alter mechanical properties [103, 128], though proper titration of the
concentration may control this variation [103, 104]. Animal studies have been performed which use whole
menisci as bioprosthetic replacements to the native meniscus [123, 125, 126] or as osteochondral
allografts [106, 124, 127], showing encouraging biomechanical function and integration. An additional
human case study was reported to have successfully used a glutaraldehyde fixed bovine meniscus
xenograft for the treatment of avascular necrosis of the scaphoid proximal pole, with satisfactory
performance reported after five years without the occurrence of inflammation [127]. Fewer studies have
been reported on the glutaraldehyde fixation of articular cartilage: One study demonstrated that fixation
increased the resistance of the tissue to degradation by collagenase [129], and another study
demonstrated decreased wear and improved tissue stabilization in vitro, following fixation [130].
Previous basic science findings have clearly established the structure-function relationships of
articular cartilage that promote elevated interstitial fluid pressurization: The tissue must be highly porous
but exhibit low permeability, to prevent rapid loss of interstitial fluid pressurization [35, 36, 131]; it must
exhibit a higher modulus in tension than compression in order to enhance this pressurization, by
restricting the ability of the tissue to expand laterally when compressed axially [34, 131-133]. In particular,
it is known from theory [134] and experiments [39] that elevated interstitial fluid pressurization and
decreased friction can be achieved when the dimensionless Peclet number for cartilage sliding contact is
much greater than unity (Pe>>1). The Peclet number is given by Pe=Vh/HAk, where V is the velocity of
the migrating contact area, h is a characteristic dimension of the contacting articular layers (cartilage
thickness or contact radius), HA is its aggregate modulus and k its hydraulic permeability. It is also known
that interstitial fluid pressurization is enhanced when the tensile modulus of the porous material is much
greater than its compressive modulus [34, 35, 131, 135]. Finally, it is known that the friction coefficient
decreases with increasing porosity [33, 38].
Therefore, though glutaraldehyde treatment is known to alter the mechanical and transport
properties of soft tissues, these alterations may be titrated by adjusting the glutaraldehyde concentration
[103, 104] to control the alteration in Pe such that it remains significantly above unity. In this case,
6
interstitial fluid pressurization can remain elevated. Many alternative materials, such as hydrogels, do not
share all the critical structure-function relationships found in articular cartilage, which are needed to
promote elevated interstitial fluid pressurization as outlined above. Furthermore, methods to improve their
functionality using alternative hydrogel formulations and fiber reinforcement may be constrained by
biocompatibility issues.
The aim of this study was to examine the functional properties and biocompatibility of glutaraldehyde-
fixed bovine articular cartilage over several weeks of incubation at body temperature to investigate its
potential use as a resurfacing material in joint arthroplasty. The hypothesis was that successful fixation
treatments may preserve functional properties of native tissue over long-term culture at body
temperature.
2.2 Materials and Methods
2.2.1 Study Design
Two studies were performed, each with glutaraldehyde treatment and incubation time as factors. In Study
2.1, the effects of glutaraldehyde stabilization on mechanical properties and tissue biocompatibility were
examined over a period of 28 days. Compressive and tensile moduli, hydraulic permeability, friction
parameters, and biocompatibility (via cell viability) were compared over time (1, 7, 14, or 28 days) to
optimize glutaraldehyde treatment concentration (0% control, 0.02, 0.20, or 0.60% glutaraldehyde) for
tissue maintenance. In Study 2-2, findings from Study 2-21 were applied and additional mechanical and
qualitative tests were performed to evaluate functional performance over time (1, 7, 14, or 28 days) at the
optimal glutaraldehyde treatment concentration (0% control and 0.20% glutaraldehyde) with more
physiologically relevant friction test conditions.
2.2.2 Sample Harvest
Experiments were conducted using articular cartilage explants harvested from matched left-right pairs of
healthy tibial plateaus and opposing condyles of 2-3 months old calf knee joints obtained from a local
abattoir. In Study 2-1, full thickness cartilage discs (4 mm) were excised from the tibial plateaus of eight
joints (Figure 2-1A). Samples were randomized to friction testing, mechanical testing, and biocompatibility
7
groups. Those samples in the friction and mechanical testing groups were placed with the articular
surface down on a freezing stage microtome (Leica Instruments #SM2400, Nusslock, Germany),
embedded and frozen in a water soluble sectioning gel (Thermo Scientific #1310APD, Rockford, IL), then
trimmed of bone and deep zone tissue to obtain samples inclusive of an intact superficial zone with a final
thickness of 1.1 ± 0.03 mm. These testing samples were then randomized into various treatment groups
and test type. Samples were rinsed with and transferred to filtered phosphate buffered saline (PBS)
supplemented with an isothiazolone based biocide (Sigma-Aldrich #46878-U, St. Louis, MO). Viable
explants for biocompatibility testing were manually trimmed of bone and deep zone tissue then
transferred to dishes containing Dulbecco’s Modified Eagle Medium (DMEM) supplemented with 10%
fetal bovine, amino acids (1 X minimum essential amino acids, 1 X nonessential amino acids), buffering
agents (10 mM HEPES, 10 mM sodium bicarbonate, 10 mM TES, 10 mM BES), and antibiotics (100
U/mL penicillin, 100 µg/mL streptomycin) and cultured at 37° C and 15% CO2 (Sigma Aldrich, St. Louis,
MO).
In Study 2-2, full thickness cartilage strip-shaped explants (28 mm 10 mm ~1 mm), one from
each medial and lateral sides of the tibial plateau, were harvested along with the opposing femoral
condyles from nine joints (Figure 2-1B). Tibial plateau strips were trimmed of bone and deep zone tissue
as described in Study 2-1 to obtain samples inclusive of an intact superficial zone and with a final
thickness of 1.1 ± 0.06 mm. Full thickness cartilage discs (10 mm) were punched from the condyles for
use as opposing articulating faces. Samples were rinsed with and transferred to filtered PBS with biocide
in the same manner as listed for Study 2-1. Condyle discs were randomized within each joint, wrapped in
PBS soaked gauze and stored at -20 °C for a period no greater than 30 days until use.
2.2.1 Glutaraldehyde Treatment
All glutaraldehyde solutions were prepared from 8% stock glutaraldehyde (Sigma Aldrich #G7526, St.
Louis, MO) and PBS. In Study 2-1, specimens were randomized to four glutaraldehyde treatment groups
(0% control, 0.02, 0.20, or 0.60% glutaraldehyde, n=20 per group per time point, of which 8 were used for
friction testing, 8 for mechanical testing, and 4 for biocompatibility) and placed in 1000 mL (80 samples
per solution concentration) of the appropriate glutaraldehyde supplemented PBS solution for 24 h at 8 °C
8
[103] . After treatment, samples were rinsed with and transferred to 1000 mL sterile PBS with biocide for
a 48 h desorption, rinsed again, and transferred to fresh PBS with biocide where they remained at 37°C
for the remainder of the incubation time (1, 7, 14, or 28 days). Sample groups were rinsed and stored
separately to prevent potential cross contamination. For this study, a total of 340 specimens were tested
(20 at day 0 and 80 at each of days 1, 7, 14 and 28). In Study 2-2, tibial plateau strips were randomized
to two treatment groups (0% control, or 0.20% glutaraldehyde, n = 9 per group, for a total of 18
specimens) and treated with glutaraldehyde according to the protocol listed for Study 2-1 and transferred
to fresh PBS with biocide where they remained at 37°C for the remainder of the incubation time (with
friction tests repeated on the same samples at 1, 7, 14, and 28 days). Samples were randomized by tibial
plateau side, with either treatment or control assigned to either the medial or lateral side for each joint.
2.2.2 Mechanical Testing
A custom two-axis loading device equipped with a six-degrees-of-freedom load cell (JR3 Inc. #20E12A4,
Woodland, CA) was used for all mechanical testing. The device was controlled with custom LabVIEW
software (National Instruments Corporation #LabVIEW 2010, Austin, TX) and an associated motion
controller (National Instruments Corporation #7354, Austin, TX) to provide continuous measures of
specimen thickness, position, applied loads, and reaction loads.
In Study 2-1, unconfined compression stress-relaxation tests were performed on a subset of
samples (n=8 per group from each incubation time point) to evaluate the equilibrium compressive (E-Y)
and tensile moduli (E+Y) and hydraulic permeability (k). For each sample, a tare strain was applied (10%
strain with 4800 s relaxation time). After equilibrium was reached, a stress relaxation test was performed
(to 15% strain from pre-tare reference configuration, with 2400 s relaxation time). Stress-relaxation data
were curve-fitted as previously described [136]. Briefly, finite element models were constructed in FEBio
[137] that incorporated the geometry of the specimens. The cartilage was modeled as a biphasic material
containing intrinsically incompressible fluid and solid phases [138]; the composite porous solid phase
consisted of a compressible neo-Hookean solid ground matrix reinforced by a continuous, random
distribution of fibril bundles sustaining tension only, with a linear stress–strain response; the hydraulic
permeability was assumed constant [136, 139]. A least-square parameter optimization routine in FEBio
9
was used to curve-fit the stress-relaxation data and extract the specimen E-Y, k, and (fibril tensile
modulus) values. These values were substituted into a finite element analysis of equilibrium uniaxial
tension to produce the tissue tensile modulus E+Y [136].
Friction measurements against smooth glass (Edmund Optics Inc. #27-834, Barrington, NJ) were
performed on another subset of samples (n=8), using a creep loading configuration (4.45 N) with
reciprocal sliding motion (±5.0 mm at 1 mm/s, Figure 2-1A) [38]. The temporal response of the friction
coefficient was evaluated from the normal and frictional load responses over 3600 s. The minimum
friction coefficient, µmin, equilibrium friction coefficient, µeq, and time constant to reach equilibrium, tµ, were
calculated as described previously [38]. Samples were submerged in PBS throughout all tests and
allowed to recover for a duration equal to or greater than the test time. A two-way ANOVA on the effects
of treatment and incubation time, followed by post-hoc testing using Tukey correction, was performed on
the moduli, hydraulic permeability, and friction parameters.
In Study 2-2, quantitative friction measurements and qualitative articular surface staining was
performed on tibial plateau strip samples. Tibial plateau strips were affixed to 60 mm petri dishes (Becton
Dickinson #351007, Franklin Lakes, NJ) with 6 µl of cyanoacrylate (Henkel Corporation #Loctite 420,
Rocky Hill, CT). Custom indexing features were attached to each dish allowing for repeatable positioning
at subsequent test time points. Friction measurements were performed against the untreated matching
condyle explant in a creep configuration with a prescribed 4.45 N load and sliding with translation range
of ±5 mm and velocity of (1 mm/s) at room temperature (Figure 2-1B). The response of the friction
coefficient was evaluated from the normal and frictional load responses over 3600 s. The initial friction
coefficient, µinitial, and final friction coefficient, µfinal, were calculated as described previously [39]. Samples
were submerged in PBS throughout all tests and allowed to recover for a duration equal to or greater than
the test time. A two-way ANOVA on the effects of treatment and incubation time, followed by post-hoc
testing using Tukey correction, was performed on friction parameters.
2.2.3 Biocompatibility Testing
Viable explants (n=4 per treatment group for each incubation time point) in Study 2-1 were co-cultured
with an equal number of fixed explants for each glutaraldehyde concentration (Figure 2-2). Samples were
10
cultured in supplemented DMEM, changed three times a week. At each time point, viable samples from
each treatment group were sliced in half and stained for chondrocyte viability (Invitrogen #L3224,
Eugene, OR). Sample cross sections were scanned on a confocal microscope (Leica Microsystems #TCS
SP5, Buffalo Grove, IL) and combined for assessment. Viable and non viable cells were counted with the
use of software after image processing [140] (NIH #ImageJ V1.44p, Bethesda, MD). Cross sectional
specimen areas were measured using sample outlines and viable to non viable cell ratios were
normalized to sample cross sectional image area. A two-way ANOVA on the effects of treatment and
incubation time, followed by post-hoc testing using Tukey correction, was performed on the cell viability
ratio.
2.3 Results
2.3.1 Study 2-1
Glutaraldehyde treated samples showed an immediate change in appearance similar to that reported for
other tissues [103, 125, 128]. Treated samples appeared to yellow with increasing glutaraldehyde
treatment concentration and felt firmer to the touch. Curve-fitting of the stress-relaxation response to the
theoretical model produced fits with coefficients of determination that did not differ among treatment
groups or time points, producing an average and standard deviation of R2=0.987 ± .011 (representative fit
in Figure 2-3). The resulting E-Y displayed a dose-dependent increase with concentration immediately
upon fixation (p ≤ 0.034, Figure 2-4), followed by maintenance of compressive stiffness over the study
duration for the higher 0.20% and 0.60% glutaraldehyde concentrations (p ≤ 0.040) and a maintenance of
native (day 0 control) levels for the 0.02% concentration (p ≥ 0.446). With the exception of day 7 at the
0.20% glutaraldehyde concentration, the difference between same day control and higher 0.20% and
0.60% glutaraldehyde concentrations (p ≤ 0.036) was also maintained over the test duration. E+Y
displayed a similar increase with glutaraldehyde concentration with significant differences detected
between 0.60% glutaraldehyde and both day 0 and same day control values at day 7, 14, and 28 (p ≤
0.015, Figure 2-4). No differences were detected between any other factor level combinations (p ≥ 0.229).
No significant differences were detected between any factor level combinations for k (p ≥ 0.577).
11
Significant increases in µeq from native values were observed immediately following glutaraldehyde
treatment, at the higher 0.20% and 0.60% treatment concentrations (p ≤ 0.003, Figure 2-4), although
values returned close to native levels by day 14 (p ≥ 0.360). A downward trend in the time constant tµ was
observed with incubation time, with a significant difference from native values detected only for the 0.02%
treatment concentration at day 14 and day 28 (p ≤ 0.016, Figure 2-4). No significant differences were
detected across glutaraldehyde concentrations or incubation times for µmin (p ≥ 0.473, Figure 2-4). No
qualitative or quantitative differences were detected in cell viability at any time point for any tested
glutaraldehyde concentration (p ≥ 0.848, Figure 2-5).
2.3.2 Study 2-2
Friction coefficient measurements in the migrating contact configuration provided an additional measure
of long-term performance in a more physiologic loading condition. Initial friction coefficient µinitial remained
near native values for the 0.20% glutaraldehyde treated tibial plateau strips (p ≥ 0.783, Figure 2-6). At day
14, control group values were different from both native and same day 0.20% glutaraldehyde treated
group values, although the difference between same day values was not detected at day 28. Final friction
coefficient µfinal values mirrored µinitial values with 0.20% glutaraldehyde treated group maintaining near
native values over the test duration (p ≥ 0.288, Figure 2-6), and a significant difference detected
between day 14 control group and native levels (p ≤ 0.0267, Figure 2-6). After testing on day 28, surface
fibrillation was assessed qualitatively with india ink staining (Figure 2-7). All nine of the control group
samples displayed evidence of fibrillation over the contact area, while none of the glutaraldehyde-treated
test group samples showed signs of superficial zone fibrillation.
2.4 Discussion
Glutaraldehyde fixation of bovine articular cartilage xenografts offers the potential for creating a highly
functional resurfacing material for hemiarthroplasties in humans, much like the use of glutaraldehyde-
treated bovine heart valves has provided a practical treatment for heart disease. The results of this study
demonstrate that glutaraldehyde fixation of cartilage over a range of concentrations produces only
moderate changes in its mechanical properties (Figure 2-4 & Figure 2-6) and is biocompatible with live
12
tissue in co-culture (Figure 2-5). The equilibrium tensile and compressive moduli increases twofold, as
expected from the crosslinking effect of the glutaraldehyde treatment; however, the hydraulic permeability
remains unchanged; the equilibrium friction coefficient returns to control values after 7 days; the minimum
friction coefficient is nearly unchanged, and the time constant for the frictional response decreases
twofold at most, consistent with the twofold increase in the moduli and the resulting effect on the
characteristic time constant for loss of interstitial fluid pressurization [141]. When loaded in a migrating
contact configuration, glutaraldehyde fixed tissue performs almost identically to native tissue (Figure 2-6).
This outcome is consistent with the fact that the frictional response under migrating contact is dependent
on the tissue’s ability to sustain interstitial fluid pressurization continuously, which is achievable when the
Peclet number remains significantly greater than unity [39]; based on the measured material properties of
control and glutaraldehyde tissue, a sliding velocity of V=1 mm/s and an estimated contact radius of a~2
mm, the Peclet number Pe=V a/E+Y k decreased from ~130 to ~65 between control and treated samples,
remaining significantly greater than unity. Furthermore, it is found that glutaraldehyde-treated tissue is
better able to withstand repeated friction loading than untreated tissue (Figure 2-7).
Previous studies have reported results from the use of autografts for cartilage resurfacing
applications, but donor tissue supply is limited [142-147] and problems still exist with donor site morbidity
and incongruence [148]. Commercial hydrogel product testing has shown inadequate connection to
subchondral bone [149, 150]. When compared to alternative porous resurfacing materials such as
hydrogels or the current generation of engineered cartilage constructs, glutaraldehyde-treated cartilage
exhibits mechanical properties much more similar to native tissue. Most importantly, results of the 28-day
long incubation of fixed tissues at body temperature reported here suggest that they may be used
successfully under in vivo conditions.
Taken together, these highly encouraging results suggest that a properly titrated glutaraldehyde
treatment (0.20% in this case) can reproduce the desired functional properties of native articular cartilage,
and maintain these properties for at least 28 days while being incubated at body temperature. The
results of this study motivate further investigations of the potential use of glutaraldehyde-fixed xenografts
for joint resurfacing. Immature cartilage from bovine joints can be easily harvested nearly whole, by
‘popping’ it off the underlying bone as illustrated in Figure 2-8. The resulting articular layer has a clean
13
and smooth underlying surface. This fortuitous ability to harvest whole articular layers from immature
joints may facilitate their use as resurfacing materials, following glutaraldehyde treatment, with a suitable
biocompatible adhesive to attach it to a substrate (such as medical octyl cyanoacrylates) (Figure 2-9). In
a carefully controlled supply of bovine joints from animals raised for this specific purpose, animals can be
sacrificed at an age when the size of the joint (such as the bovine humeral head diameter) matches the
desired dimensions.
2.5 Acknowledgements
Research reported in this section was supported by the National Institute of Arthritis and Musculoskeletal
and Skin Diseases of the National Institutes of Health under Award Number R01 AR043628.
14
2.6 Figures
Figure 2-1: Schematic representation of bovine calf articular cartilage harvest location and friction testing schematic.
(A) Study 2-1 includes articular cartilage discs harvested from the tibial plateau, fixed with glutaraldehyde, and tested
against a smooth glass counterface. (B) Study 2-2 includes articular cartilage strips harvested from the tibial plateau,
fixed with glutaraldehyde, and tested against unfixed convex discs from the opposing condyles.
15
Figure 2-2: Representative biocompatibility culture dish with viable and glutaraldehyde treated cartilage explants.
Explants were cultured for 28 days in cell culture media at 37° C and 15% CO2. Samples from each treatment group
were sectioned and stained for chondrocyte viability at 1, 7, 14, and 28 days.
16
Figure 2-3: Representative FEBio curve-fit of experimental stress-relaxation test (5% strain) used to obtain model
parameters for E-Y, k, and ξ (0% control sample at day 14, R2=0.988).
17
Figure 2-4: Summary of mechanical properties for Study 2-1 as a function of glutaraldehyde concentration. µeq =
equilibrium friction coefficient; tµ = time constant for the rise of the friction coefficient to its equilibrium value; µmin =
minimum friction coefficient, typically achieved at the earliest time point of the friction response; E-Y = equilibrium
compressive modulus; E+Y = equilibrium tensile modulus; k = hydraulic permeability. Significant differences (p ≤ 0.05)
from native day 0 values indicated with †, differences between groups indicated with *.
18
Figure 2-5: Qualitative and quantitative biocompatibility from chondrocyte viability in sectioned co-cultured explants at
day 1 and day 28 for 0.0% control, 0.02, 0.2, and 0.6% glutaraldehyde treated explants. Samples from each
treatment group were sectioned and stained for chondrocyte viability at 1, 7, 14, and 28 days. No significant
differences were detected between groups for the ratio of viable to non viable cells normalized to tissue area.
19
Figure 2-6: Frictional properties for Study 2-2, migrating contact configuration. µinitial = friction coefficient at start of
test; µfinal = friction coefficient at end of test (1 h). Significant differences (p ≤ 0.05) from day 0 values indicated with †,
differences between groups indicated with *.
20
Figure 2-7: India ink staining of representative Study 2-2 tibial plateau strips shows surface fibrillation of the control
(CTRL) group without any apparent superficial zone damage to the 0.2% glutaraldehyde (GLUT) treated sample
group after friction testing at day 28.
21
Figure 2-8: Intact immature bovine knee condyles (A). ‘Popped’ condylar articular layer (B). Cross-section of popped
articular layer, showing the underlying surface (C).
22
Figure 2-9: Potential clinical approach for bioprosthetic arthroplasties, using glutaraldehyde-fixed xenografts. (A)
Natural shoulder joint. (B) A glutaraldehyde-treated bovine articular layer may be affixed to a hemispherical stainless
steel substrate supported on a metal stem, and inserted into the humerus in analogy to standard hemiarthroplasties.
23
CHAPTER 3: CARTILAGE WEAR DEBRIS MEASUREMENTS
VIA PARTICLE ANALYZER
3.1 Introduction
Although qualitative observations of cartilage wear debris have been made [45-52]; quantitative
measurements of cartilage wear have proven to be more challenging, with only a few studies having
reported such measurements. The primary quantitative approaches proposed to date include biochemical
assaying of cartilage and test solutions [54-57, 151], characterization of changing articular layer thickness
[54, 58-60], and changes in surface roughness [7, 57, 61, 62, 64, 65, 101]. One study examining
polyethylene wear debris in hip arthroplasty reported the use of an automated particle analyzer [152]. The
aim of this study was to test whether latest-generation particle analyzers are capable of detecting
cartilage wear debris generated during in vitro loading experiments that last 24 h or less, by producing
measurable content significantly above background noise levels.
3.2 Materials and Methods
3.2.1 Sample Harvest
Articular cartilage cylindrical explants were harvested from the tibial plateau of 2-3 month old calf knee
joints (n=9) obtained from a local abattoir. Full thickness cartilage discs (4 mm) were excised, placed
with articular surface down on a freezing stage microtome (Leica Instruments #SM2400, Nusslock,
Germany), embedded and frozen in a water soluble sectioning gel (Thermo Scientific #1310APD,
Rockford, IL), and trimmed of bone and deep zone tissue to obtain samples inclusive of an intact
superficial zone with a final thickness between 1.2 to 1.4 mm (n=9 per treatment group). After
preparation, samples were stored at -20° C for a period no greater than 30 days until use. Cartilage
sample pairs harvested from adjoining regions were randomized into test and control groups to produce a
study design with repeated measure.
3.2.2 Solution Preparation
24
Phosphate buffered saline solution (PBS) supplemented with a protease inhibitor cocktail (Thermo
Scientific #78438, Rockford, IL) and bactericide (Sigma-Aldrich #46878-U, St. Louis, MO) was used as a
base solution for all testing. Base solution was passed through a 0.1 µm pore filter (Millipore
#SLVV033RS, Billerica, MA) to eliminate background particulate, and care was taken to prevent
contamination. Four solution groups were prepared, one each for the loaded cartilage wear (TEST),
unloaded cartilage control (CTRL), environmental contamination control (ENVR), and base solution
contamination control (BASE) groups (Figure 3-1).
3.2.3 Mechanical Testing
Test and control cartilage samples were thoroughly rinsed with base solution prior to testing and
submerged in 10 mL of base solution, which was aspirated and replaced with fresh solution at 1, 2, 6, and
24 h time points. Test cartilage samples were subjected to friction using a smooth glass slide. The glass
slide surface roughness was measured, Ra = 0.003±0.002 µm, using an optical profilometer with a 10X
objective and 1.81 X 1.36 mm scan area (Zygo #NV5000, Middlefield, CT). Measurements were
performed in an unconfined compression creep configuration with a prescribed 4.48 N load (equivalent to
0.36 MPa) and intermittent sliding with translation range of ±10 mm and sliding velocity of 1 mm/s at room
temperature, as used previously for the TEST group [7]. Control cartilage samples were placed in dishes
without loading platens to determine the combined effects of natural particulate shedding from cartilage
and environmental contamination for the CTRL group. Dishes without cartilage samples were used to
determine environmental solution contamination for the ENVR group and were sampled in the same way
as TEST and CTRL groups. The BASE solution particulate content was also assessed to verify
cleanliness over time Figure 3-1.
3.2.4 Particulate Analyzer
Solutions were tested with a particle sizing and counting analyzer configured to measure particulates from
0.6 to 60μm in diameter using two separate aperture tubes (30 μm and 100 μm) with a 300 μL total
aspiration volume (Beckman Coulter #Multisizer 4, Brea, CA). After testing with the 100 µm aperture tube,
solutions were then screened through a 10 µm filter (Spectrum Labs #148102, Rancho Dominguez, CA)
prior to testing with the 30 µm aperture, to prevent clogging (the 10 µm filters are first cleaned by copious
25
flushing with BASE solution). Counting analyzer calibration verification for both particulate size and
number was performed per manufacturer specifications. Solution samples were placed in clean
containers for testing, and agitated throughout test duration to keep particulates in suspension during
characterization. For each group, cumulative particulate content was calculated at each time point by
adding results from preceding time points.
3.2.5 Particulate Assay
To concentrate wear particulate in preparation for biochemical assay, remaining sample solutions were
centrifuged at 14,000 RCF for 45 minutes. The supernatant was removed and replaced with filtered
deionized water to decrease the overall salt content prior to lyophilization. This centrifugation rinse cycle
was repeated twice for each solution sample prior to lyophilization. Wear particulate lyophilisate was
digested with Proteinase K for 12 hours at 60° C in 0.1 M sodium acetate containing 5 mM EDTA, 1 mM
iodacetamide, 10 µg/mL pepstatin A, and 50mM Tris. Minimal digestion solution volumes were used (250
µL) to maximize particulate concentration for increased assay sensitivity. Solution digests were then
assayed for sulfated glycosaminoglycan (GAG) content using the 1,9-dimethylmethylene blue dye-binding
assay [153]. Hydroxyproline content was determined from aliquots of the Proteinase K digest product.
Samples were hydrolyzed by heating with equal volume of 12M HCl at 107° C for 18 hours and dried. The
hydrolysates were assayed for hydroxyproline content using a colorimetric assay [154]. Collagen content
was calculated from the aliquot fraction and hydroxyproline to total collagen ratio (7.64) [155].
3.2.6 Particulate Imaging
Representative samples were obtained for TEST, CTRL, and ENVR solutions at the 24 h time point and
concentrated according to the method above. Solutions were then incubated in 20 μM dichlorotriazinyl
aminofluorescein (Invitrogen #D16, Eugene, OR) for 30 min then mixed one to one with a water soluble
mounting gel to prevent particulate motion during imaging (Sigma-Aldrich #C 0612, St. Louis, MO). Gel
solutions were mounted on coverslips for confocal imaging and sealed to prevent evaporation [156]. Z
image stacks were obtained on a confocal microscope (Leica Microsystems #TCS SP5, Buffalo Grove, IL)
26
and combined (NIH #ImageJ V1.44p, Bethesda, MD) for qualitative visualization. Before and after testing,
images were taken of the TEST cartilage tissue samples to assess potential macroscopic damage.
3.2.7 Statistical Analysis
A two-way analysis of variance was performed for the factors of treatment (TEST, CTRL, ENVR, BASE)
and time (1, 2, 6, 24 h) using repeated measures, with type I error set at =0.05 and significance set at
p≤0.05. Post-hoc testing of the means used Tukey correction. Analyses were performed for particulate
number, particulate volume, GAG content, and collagen content (SAS Institute #SAS V9.2, Cary, NC).
3.1 Results
3.1.1 Particulate Number
TEST solution cumulative wear particulate number was found to be significantly greater than CTRL,
ENVR, and BASE solutions at all time points tested (p≤0.005, Figure 3-2A). No differences were detected
between CTRL, ENVR, and BASE solutions at any time points (p≥.97).
3.1.2 Particulate Volume
TEST solution cumulative wear particulate volume, similar to particulate number, was found to be
significantly greater than CTRL, ENVR, and BASE solutions at the 2, 6, and 24 hour time points (p<.042)
(Figure 3-2B). No differences were detected between CTRL, ENVR, and BASE solutions at any time
points (p≥.90).
3.1.3 Particulate Assay and Imaging
No differences were detected between TEST, CTRL, ENVR, and BASE solutions for either GAG (p≥.87)
or collagen content (Table 3-1, p≥.93). Qualitative visualizations from confocal imaging regarding particle
geometry (Figure 3-3) matched size distributions obtained via particle analyzer (Figure 3-4A, Figure
3-4B).
3.2 Discussion
27
The reported measurements of this study clearly demonstrate that the cartilage samples subjected to
frictional loading produce particulate content that is significantly higher than background noise and
contamination levels (Figure 3-2A, Figure 3-2B). The experimental design of this study accounted for
shedding of cartilage debris in the absence of loading, as may occur from natural enzymatic degradation
or other similar mechanisms. The design also accounted for contamination from the testing environment,
such as dust particles from the air or debris from the dishes and fluid handling equipment. By enforcing a
clean testing environment, and minimizing enzymatic degradation using protease inhibitors, it was found
that environmental contamination was negligible at all time points, in comparison to the wear produced
from frictional loading. Confocal images provide direct visual evidence of the debris characterization from
the particle counter (Figure 3-3).
Though the amount of cartilage wear observed in the TEST group was significantly higher than
background noise and contamination, it should be noted that it represented only a minute amount of loss
from the cartilage sample. As noted from Figure 3-2B, the particulate volume at 24 h averaged less than
2106 µm
3, which is less than 0.02 percent of the total volume of cartilage (1610
9 µm
3). This outcome is
not surprising, considering the relatively short 24 h duration of loading used in this test, when compared
to the normal lifetime of cartilage. It is thus remarkable that the particle analyzer used in this study is able
to detect such small but significant wear levels, otherwise not detected using standard biochemical
assays (Table 3-1). This high sensitivity will considerably facilitate future studies of the wear response of
cartilage under a variety of testing conditions, allowing investigators to test various wear-related
hypotheses.
The lubrication regime achieved in the testing conditions of this study is a combination of
interstitial fluid pressurization and boundary lubrication [157, 158]. As shown in prior studies, upon
application of the load, the interstitial fluid in the tissue pressurizes and supports most of the applied load,
producing a low friction coefficient [38]. As the pressure subsides, the load becomes supported by the
solid matrix, thus resulting in a boundary lubrication regime and a high friction coefficient [38]. For a 4 mm
diameter plug, the characteristic time for this loss of fluid pressurization is ~1400 s [141]. Therefore, for
the experiments in this study, boundary lubrication prevailed over most of the testing duration.
28
Furthermore, the magnitude of wear reported in this study is only representative of the specific testing
conditions used here. Other experimental conditions will likely produce differing amounts of wear.
In summary, the results of this study establish unequivocally that latest-generation particle
analyzers are capable of detecting very low wear levels in cartilage experiments conducted over a period
no greater than 24 h.
3.3 Acknowledgements
Research reported in this section was supported by the National Institute Of Arthritis And Musculoskeletal
And Skin Diseases of the U.S. National Institutes of Health under Award Number R01AR043628. The
authors would like to thank Mr. Aleksander V. Navratil and Dr. Elon J. Terrell for training and use of the
optical profilometer.
29
3.5 Figures
Table 3-1: Biochemical assay measurements of cartilage wear debris showing no detected difference in either
glycosaminoglycan (A, p≥.87) or collagen (B, p≥.93) content among the TEST, CTRL, ENVR, or BASE groups at any
time point (n=8 per group, 1, 2, 6, 24 h).
30
Figure 3-1: Schematic for TEST, CTRL, ENVR (cross-sectional views), and BASE (15mL centrifuge tube) solution
configurations.
31
Figure 3-2: Wear particulate measurements for bovine articular cartilage explants showing significant (* p ≤ 0.05)
increases in particulate number (A) and volume (B)when compared to unloaded CTRL, ENVR, and BASE groups at
respective time points (n=9 per group, 1, 2, 6, 24 h). No difference between CTRL, ENVR, or BASE groups was
detected (A: p≥.97, B: p≥.90).
32
Figure 3-3: Wear particulate confocal z stack and angled z stack showing qualitative agreement between observed
size distribution and particle analyzer measurements for representative 24 h TEST solution.
33
Figure 3-4: Particulate size (A) and volume (B) distribution for representative 24 h TEST solution showing a high
number of micron sized particles.
34
CHAPTER 4: INFLUENCE OF BEARING MATERIAL AND
ROUGHNESS ON CARTILAGE DEGENERATION
4.1 Introduction
Hemiarthroplasty is a surgical procedure that replaces half of a diarthrodial joint with an artificial bearing,
leaving the apposing articular cartilage layer intact. According to the Centers for Disease Control, there
are over 360,000 hemiarthroplasty operations performed each year in the US, [73]. Common indications
for hemiarthroplasty include acute bone fracture of the proximal femoral head of the hip [85], or proximal
humeral head of the shoulder [83, 84]; failed internal fixation from osteonecrosis of the diarthrodial joint
head [76, 86, 87]; and low activity expectations [79]. In all cases, hemiarthroplasty is indicated only if the
non-resurfaced native articular layer shows no significant signs of disease or damage [80].
Hemiarthroplasties are less invasive, require less surgical time and result in lower blood loss than total
arthroplasties [74, 75]. The value of hemiarthroplasty over total joint arthroplasty however is still subject to
debate [76-78], most notably for displaced femoral neck fractures [74, 79, 80], and the treatment of
primary shoulder osteoarthritis [75, 81, 82]. Patient outcomes are sometimes shown to be less successful
than with total arthroplasties, and often require revision, due in part to erosion of the articular surface [59,
88-92]. There is a general uncertainty about the cause of seemingly accelerated cartilage wear in
hemiarthroplasties.
The low friction coefficient of healthy cartilage, resulting primarily from boundary lubrication and
fluid load support, has been characterized extensively in the literature [1, 2, 4-6, 157, 159, 160]. In
particular, it is known from theory [39, 134] and experiments [39] that elevated interstitial fluid
pressurization and decreased friction can be achieved. Surface lubricating molecules acting as boundary
lubricants, such as lubricin, are known to be important as well [161]. Qualtiative and quantitative methods
for wear measurement via particulate analysis [48-50, 52, 162], biochemical assay [56, 57, 163], histology
[46, 53, 163], electron microscopy [47], magnetic resonance and radiological methods have been
developed. However, the complex relationship between friction coefficient and tissue damage and wear is
still poorly understood, and determining the underlying physical and biological events that lead to these
35
conditions remains elusive [3, 66-69]. It is generally assumed that low friction is associated with low wear
and damage, but as seen with the first generation of artificial hips, this association is not straightforward
[70-72]. Further understanding of this relationship between friction, damage, and wear is crucially
important to the common clinical procedure of hemiarthoplasty.
Material surface chemistry and geometry generally play an important role in tribology, and
advancements in material science and surface preparation have produced materials, surface coatings,
and polishing techniques for artificial joint bearing surfaces thought to minimize the degradative effects of
wear and damage to native articular cartilage in the in-vivo environment. Some investigators have
suggested that using different artificial materials for resurfacing the joint might produce less wear [164-
168]. However, a comprehensive set of quantitative data is lacking, many of the previous in-vitro studies
apply loading conditions that encourage accelerated wear, and use artificial material counterface
geometry that doesn’t meet FDA guidelines for surface roughness of artificial joint components in contact
with native tissue [167, 168]. This deviation from accepted guidelines for surface roughness of implanted
artificial joint bearing components, severely limits the potential clinical impact of those resulting material
performance comparisons.
The aim of this study was to explore use of materials used in hemiarthroplasty to glass, a material
commonly used in in-vitro studies. Two different cobalt chromium alloys along with a stainless steel alloy
accepted by the FDA for use as bearing counterfaces were selected for comparison. Material surface
preparation was conducted and tested in accordance to FDA guidelines to ensure that surface roughness
would be comparable to implanted joint components. A test configuration that promotes loss of fluid
pressurization over time was specifically selected to determine whether the equilibrium friction coefficient
is responsible for wear and damage. The hypothesis was that low equilibrium friction coefficient would be
associated with low levels of cartilage tissue wear.
4.2 Materials and Methods
4.2.1 Material Preparation
36
Cobalt-chromium low carbon (CoCrLC), cobalt-chromium high carbon (CoCrHC), and 316 low vacuum
melt stainless steel (316SS) orthopaedic material samples were obtained as cylindrical rod sections to be
used as friction platens (M. Vincent & Associates, Bloomington, MN). Platen samples were ground flat
and parallel using a surface grinder (Acer #AGS-1020AHD, Anaheim, CA). Platen samples were held in a
custom jig to maintain surface flatness, then ground and polished using successively finer silicon carbide
abrasives and final alumina oxide slurry polish steps. A group of CoCrLC platen samples was withheld
from the final polishing step to create a group with a rougher surface (CoCrLC-Ra25nm). Uncoated
optical borosilicate glass discs were used for a glass platen group (Edmund Optics #43-892, Barrington
NJ). Surface roughness measurements were performed before and after cartilage friction testing with an
optical profilometer to ensure surface finish compliance to both ISO 21534:2007, and ASTM F2033
standards with average roughness Ra less than 0.05µm for all platen samples. Six measurements were
taken on each surface.
4.2.1 Sample Harvest
Experiments were conducted using articular cartilage explants harvested from matched left-right pairs of
healthy tibial plateaus and opposing condyles of 2-3 months old calf knee joints obtained from a local
abattoir. Full thickness cartilage discs (4 mm) were excised from the tibial plateaus of eight joints,
placed with the articular surface down on a freezing stage microtome (Leica Instruments #SM2400,
Nusslock, Germany), embedded and frozen in a water soluble sectioning gel (Thermo Scientific
#1310APD, Rockford, IL), then trimmed of bone and deep zone tissue to obtain samples inclusive of an
intact superficial zone with a final thickness of 1.6 ± 0.05 mm (Figure 4-1). Samples were then
randomized into various treatment groups and test type. Samples were rinsed with and transferred to
filtered phosphate buffered saline (PBS) supplemented with an isothiazolone based biocide (Sigma-
Aldrich #46878-U, St. Louis, MO) and protease inhibitors.
4.2.2 Mechanical Testing
A custom two-axis loading device equipped with a six-degrees-of-freedom load cell (JR3 Inc. #20E12A4,
Woodland, CA) was used for all mechanical testing. The device was controlled with custom LabVIEW
37
software (National Instruments Corporation #LabVIEW 2010, Austin, TX) and an associated motion
controller (National Instruments Corporation #7354, Austin, TX) to provide continuous measures of
specimen thickness, position, applied loads, and reaction loads.
Friction measurements against all platens were performed on samples (n=8), using an
unconfined compression creep loading configuration (2.22 N) with reciprocal sliding motion (±5.0 mm at 1
mm/s, Figure 4-1) [38]. Sample solutions were collected for wear particulate samples as described
previously [162]. The temporal response of the friction coefficient was evaluated from the normal and
frictional load responses over 4 h. The friction coefficient, µ was calculated as described previously [38].
Samples were submerged in PBS throughout all tests and allowed to recover for a duration equal to or
greater than the test time. Before and after test sample wet weights and macroscopic images were taken.
A two-way ANOVA on the effects of treatment and test time, followed by post-hoc testing using Tukey
correction, was performed on minimum friction coefficient µmin, friction coefficient at 98% creep
displacement (µ98%), and friction coefficient at 4 h (µ4h).
Unconfined compression stress-relaxation tests were performed on samples (n=8) after friction
testing to evaluate the equilibrium compressive modulus (EY). For each sample, a tare load was applied
(0.025N with 360 s relaxation time). After equilibrium was reached, a stress relaxation test was performed
(to 5% strain from pre-tare reference configuration, with 1200 s relaxation time). A one-way ANOVA on
the effects of treatment, followed by post-hoc testing using Tukey correction, was performed on swelling
ratio (after test to before test wet weight ratio) and compressive modulus.
4.2.3 Particulate Testing
Friction test bath solutions were tested with a particle sizing and counting analyzer configured to measure
particulates from 0.6 to 60μm in diameter using two separate aperture tubes (30 μm and 100 μm) with a
300 μL total aspiration volume (Multisizer 4, Beckman Coulter, Brea, CA). Counting analyzer calibration
verification for both particulate size and number was performed per manufacturer specifications. Solution
samples were placed in clean containers for testing, and agitated throughout the test duration to keep
particulates in suspension during characterization. Solution preparation and sample rinsing procedures
were performed as described previously [162]. A one-way ANOVA on the effects of treatment, followed by
38
post-hoc testing using Tukey correction, was performed on total particulate number and total particulate
volume.
4.2.4 Biochemical Testing
After mechanical testing, samples were bisected and weighed, and half of each construct was lyophilized
and digested in 0.5 mg/mL proteinase K solution (Fisher Scientific) overnight at 56 °C [155]. Sulfated
GAG content was determined by dimethylmethylene blue spectrophotometric assay [153, 169], collagen
content was assessed by orthohydroxyproline assay of acid-hydrolysates of proteinase K digests [169]
assuming a 1:7.64 OHP-to-collagen mass ratio [155]. Biochemical contents were normalized to tissue
sample pre-test weights. A one-way ANOVA on the effects of treatment, followed by post-hoc testing
using Tukey correction, was performed on GAG and collagen content.
4.2.5 Histological Staining
Remaining bisected samples were fixed, sectioned, and stained for collagen with safranin-o [169].
4.3 Results
4.3.1 Material Roughness Testing
No differences in individual platen group surface roughness were detected after friction testing (p ≤ 0.007,
Table 4-1). As intended, differences between the rough CoCrLC-Ra25nm group and all other platen
groups were detected both before (p ≤ 0.003, Table 4-1) and after (p ≤ 0.002, Table 4-1) friction testing.
4.3.2 Mechanical Testing
No difference in minimum friction coefficient (µmin) was detected between any of the tested counterfaces
(Figure 4-2A, p ≥ 0.72). Differences were detected in friction coefficient at 98% creep displacement (µ98%)
and at 4 hours (µ4h) between the 316SS group and Glass, CoCrHC, and CoCrLC (p ≤ 0.01); and between
the CoCrLC – Ra 25nm and Glass, CoCrHC, and CoCrLC (Figure 4-2B, p ≤ 0.005). No differences were
detected between Glass, CoCrHC, and CoCrLC (Figure 4-2B, p ≥ 0.27) or between 316SS and CoCrLC –
39
Ra 25nm (Figure 4-2B, p ≥ 0.57). After friction testing, gross swelling in the top region was evident for
many of the samples in the 316SS and CoCrLC – Ra 25nm groups (Figure 4-3).
Swelling ratio differences were detected between the 316SS group and Glass, CoCrHC, and
CoCrLC (Figure 4-4A p ≤ 0.01), and between the CoCrLC–Ra 25nm and Glass, CoCrHC, and CoCrLC
(Figure 4-4A p ≤ 0.003).
4.3.3 Particulate Testing
Differences between Glass and CoCrLC-Ra25nm, and Glass and 316SS, were detected for both wear
particulate number (Figure 4-4C p ≤ 0.023) and particulate volume (Figure 4-4D p ≤ 0.043), however no
differences were detected between any other groups (p ≥ 0.27).
4.3.4 Biochemical Testing
Post friction test GAG content differences were detected between CoCrLC-Ra25nm and Glass, CoCrLC,
and 316SS groups (Figure 4-5A p ≤ 0.041), although no other differences were detected between groups
(Figure 4-5B p ≥ 0.98).
4.3.5 Histological Staining
After sectioning and staining, nearly all samples exhibited swelling of the top layer (Figure 4-6A). Some
samples in groups exhibiting high swelling ratio (316SS and CoCrLC-Ra25nm) showed delamination of
the top layer (Figure 4-6B).
4.4 Discussion
The results of this experimental study demonstrate that artificial bearing materials commonly used in
hemiarthroplasty may induce damage of the apposing cartilage surface, even when prepared to FDA
approved surface roughness levels and when tested using physiologic loads. These results suggest that
316SS, when compared to CoCr, should not be used as a bearing surface against articular cartilage,
contradictory to results previously published [168]. Swelling ratio tended to increase with equilibrium
friction coefficient, although the difference in wear particulate generation was not as apparent. Swelling
40
ratio more than doubled when surface roughness doubled from approximately 0.013µm to 0.027µm for
the CoCrLC to CoCrLC-Ra25nm groups, even though both were well under the 0.050µm limit required by
the FDA. Decreases in equilibrium compressive modulus mirrored swelling of the top layer (Figure 4-4).
Previous studies have shown that the layer just beneath the superficial zone of both neonatal bovine and
adult human cartilage has a low shear modulus [170, 171]. Macroscopic images show general swelling of
the top region (Figure 4-3), and while microscopic histological sections show little surface fibrillation,
failure of the region just beneath the surface near the depth of lowest shear properties [170, 171] is seen
(Figure 4-6). This could be taken to suggest that cartilage degradation associated with osteoarthritis might
initiate from the damage of this layer as opposed to wear at the surface.
Taken together, these results suggest that selection of artificial joint bearing material and surface
preparation may have an effect on the long term success of hemiarthroplasties, and that equilibrium
friction coefficient is indeed related to tissue damage. The results motivate further investigations on the
difference between articular cartilage wear and damage, and of current guidelines for the surface
preparation of orthopaedic bearing materials.
4.5 Acknowledgements
Research reported in this section was supported by the National Institute of Arthritis and Musculoskeletal
and Skin Diseases of the National Institutes of Health under Award Number R01 AR043628.
41
4.6 Figures
Figure 4-1: Schematic representation of bovine calf articular cartilage harvest location and friction testing schematic.
Articular cartilage discs harvested from the tibial plateau, trimmed of subchondral bone, and tested against flat
counterfaces of different material composition and roughness.
42
Figure 4-2: Frictional properties of cartilage against different counterface materials and roughness. Minimum friction
coefficient (µmin) (A) and a comparison of friction coefficient at time of 98% creep displacement and final time at 4h
between tissue friction test counterface materials of different composition and roughness (B). Significant differences
(p ≤ 0.05) between groups indicated with *.
43
Figure 4-3: Representative qualitative macroscopic comparison of cartilage tissue samples before and after friction
testing for glass and 316SS counterface materials. Top, front, and side views are shown. After friction testing,
swelling of the upper region is evident and highlighted by arrows in the 316SS after test images.
44
Figure 4-4: Summary of cartilage mechanical properties and wear measurements. Swelling ratio (A) (after test to
before test wet weight ratio), equilibrium compressive modulus EY (B), and wear debris particulate number (C) and
volume measurements (D) for cartilage tissue friction testing. Significant differences (p ≤ 0.05) between groups
indicated with *.
45
Figure 4-5: Summary of cartilage biochemical properties. GAG content (A) and collagen content after friction testing
normalized to sample initial wet weight. Significant differences (p ≤ 0.05) between groups indicated with *.
46
Figure 4-6: Representative histological images of top region. Cartilage tissue samples after friction testing for glass
(A) and 316SS (B) counterface materials stained with safranin-o showing damage and delamination of top layer for
316SS tested sample.
47
Table 4-1: Roughness measurements for orthopaedic material platens before and after tissue friction testing. No
differences were detected on a per material basis before and after testing for all materials, and between smooth
materials (Glass, CoCrHC, CoCrLC, and 316SS, p ≥ 0.98) or on a per material basis before and after testing (p ≥
0.99). Differences between the rough CoCrLC Ra 25nm group and all others were detected (p ≤ 0.01).
Glass CoCrHC CoCrLC CoCrLC 316SSRoughness - Ra Ra 25nmBefore Test [µm] 0.008 ± 0.005 0.011 ± 0.004 0.013 ± 0.003 0.027 ± 0.010 0.010 ± 0.004
After Test [µm] 0.010 ± 0.003 0.012 ± 0.003 0.012 ± 0.002 0.025 ± 0.008 0.012 ± 0.005
48
CHAPTER 5: CONCLUSION
5.1 Motivation
This work has been motivated by the desire to further understand degradation of articular cartilage. Most
of the past literature has focused on measurement of friction coefficient as a proxy for damage, however
understanding of osteoarthritis cannot be improved without better understanding of the mechanisms for
cartilage degradation and their relationship to friction. More specifically, the studies described in this
thesis have been motivated by the following questions:
- Can a porous material capable of sustaining interstitial fluid pressurization, such as
glutaraldehyde fixed articular cartilage, be used for functional resurfacing of damaged
articular cartilage?
- Are low levels of articular cartilage wear, associated with in-vivo conditions, possible to
measure using readily available particulate analysis equipment?
- Is material selection and the associated surface roughness characteristics important for the
performance of artificial joint components used in hemiarthroplasties?
5.2 Glutaraldehyde Fixed Cartilage As A Resurfacing Material
It was seen that glutaraledhyde fixation of bovine articular cartilage over a range of concentrations
produced only moderate changes in its mechanical properties and is biocompatible with live tissue in co-
culture. When loaded in a migrating contact configuration, glutaraldehyde fixed tissue performed almost
identically to native tissue. Furthermore, it is found that glutaraldehyde-treated tissue is better able to
withstand repeated friction loading than untreated tissue. These encouraging results suggest that a
properly titrated glutaraldehyde treatment can reproduce the desired functional properties of native
articular cartilage, and maintain these properties for at least 28 days while being incubated at body
temperature.
This study represents a necessary step before proceeding with animal experiments. Chondrocyte
cell viability was the only measurement of biocompatibility reported here. In vivo, it is possible that
glutaraldehyde-fixed cartilage wear debris within the synovial capsule might provoke an immunologic
49
response that could lead to joint pain, cell death, or further tissue degradation. Furthermore, the explant
testing configuration employed in this in vitro study does not exactly reproduce the response of a whole
joint to physiological load magnitudes. Finally, the failure characteristics of glutaraldehyde-treated
cartilage remain to be investigated. In vivo investigations might address some of the implications of this
study regarding the immunologic response from residual glutaraldehyde diffusion and fixed cartilage
debris generation within the closed synovial capsule.
5.3 Cartilage Wear Debris Measurements Via Particle Analyzer
The measurements reported here clearly demonstrate that the cartilage samples subjected to frictional
loading produce wear particulate content that is significantly higher than background noise. Confocal
images provide direct visual evidence of the debris characterization from the particle counter. Though the
amount of cartilage wear observed in the friction test group was significantly higher than background
noise and contamination, it should be noted that it represented only a minute amount of loss from the
cartilage sample. The particulate volume after 24 h of testing averaged less than 0.02 percent of the total
volume of cartilage. This outcome is not surprising, considering the relatively short 24 h duration of
loading used in this test, when compared to the normal lifetime of cartilage. It is thus remarkable that the
particle analyzer used in this study is able to detect such small but significant wear levels, otherwise not
detected using standard biochemical assays. This high sensitivity will considerably facilitate future studies
of the wear response of cartilage under a variety of testing conditions, allowing investigators to test
various wear-related hypotheses. The results of this study establish unequivocally that latest-generation
particle analyzers are capable of detecting very low wear levels in cartilage experiments conducted over a
period no greater than 24 h.
5.4 Influence of Artificial Bearing Material on Cartilage
Degeneration
The results of this experimental study demonstrate that artificial bearing materials commonly used in
hemiarthroplasty may induce damage of the apposing cartilage surface, even when prepared to FDA
approved surface roughness levels and when tested using physiologic loads. Swelling tended to increase
50
with equilibrium friction coefficient, although the difference in wear particulate generation was not as
apparent. Swelling more than doubled when surface roughness doubled even though all tested materials
were well under the limit required by the FDA. Decreases in equilibrium compressive modulus mirrored
swelling of the top layer. Macroscopic images show general swelling of the top region, and while
microscopic histological sections show little surface fibrillation, failure of the region just beneath the
surface near the depth of lowest shear properties [170, 171] is seen. This could be taken to suggest that
cartilage degradation associated with osteoarthritis might initiate from the damage of this layer as
opposed to wear at the surface.
These results suggest that selection of artificial joint bearing material and surface preparation may
have an effect on the long term success of hemiarthroplasties, and that equilibrium friction coefficient is
related to tissue damage. The results motivate further investigations on the difference between articular
cartilage wear and damage, and reevaluation of current guidelines for the surface preparation of
orthopaedic bearing materials.
51
CHAPTER 6: REFERENCES
1. Jones, E.S., Joint lubrication. Lancet, 1934. 1(5783): p. 1426-1427.
2. McCutchen, C.W., The frictional properties of animal joints. Wear : an international journal on the
science and technology of friction lubrication and wear, 1962. 5(1): p. 1-17.
3. Stachowiak, G.W., A.W. Batchelor, and L.J. Griffiths, Friction and Wear Changes in Synovial Joints.
Wear : an international journal on the science and technology of friction lubrication and wear, 1994.
171(1-2): p. 135-142.
4. Forster, H. and J. Fisher, The influence of loading time and lubricant on the friction of articular
cartilage. Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in
medicine, 1996. 210(2): p. 109-19.
5. Ateshian, G.A. and V.C. Mow, Lubrication and wear of diarthroidal joints. Basic orthopaedic
biomechanics & mechano-biology. 2005, Philadelphia, PA: Lippincott Williams & WIlkins.
6. Naka, M.H., et al., Evaluation of the effect of collagen network degradation on the frictional
characteristics of articular cartilage using a simultaneous analysis of the contact condition. Clin
Biomech (Bristol, Avon), 2005. 20(10): p. 1111-8.
7. Jay, G.D., et al., Association between friction and wear in diarthrodial joints lacking lubricin. Arthritis
and rheumatism, 2007. 56(11): p. 3662-9.
8. Elsaid, K.A., G.D. Jay, and C.O. Chichester, Reduced expression and proteolytic susceptibility of
lubricin/superficial zone protein may explain early elevation in the coefficient of friction in the joints
of rats with antigen-induced arthritis. Arthritis Rheum, 2007. 56(1): p. 108-16.
9. Charnley, J., The Lubrication of Animal Joints in Relation to Surgical Reconstruction by
Arthroplasty. Annals of the Rheumatic Diseases, 1960. 19(1): p. 10-19.
10. Bowden, F.P. and D. Tabor, The friction and lubrication of solids. Oxford classic texts in the
physical sciences. 2001, Oxford, New York: Clarendon Press ; Oxford University Press. xviii, 374 p.
11. Bell, C.J., E. Ingham, and J. Fisher, Influence of hyaluronic acid on the time-dependent friction
response of articular cartilage under different conditions. Proceedings of the Institution of
Mechanical Engineers. Part H, Journal of engineering in medicine, 2006. 220(1): p. 23-31.
52
12. Forsey, R.W., et al., The effect of hyaluronic acid and phospholipid based lubricants on friction
within a human cartilage damage model. Biomaterials, 2006. 27(26): p. 4581-90.
13. Hills, B.A. and M.K. Monds, Enzymatic identification of the load-bearing boundary lubricant in the
joint. Br J Rheumatol, 1998. 37(2): p. 137-42.
14. Schmidt, T.A., et al., Boundary lubrication of articular cartilage: role of synovial fluid constituents.
Arthritis Rheum, 2007. 56(3): p. 882-91.
15. Swann, D.A., et al., Role of hyaluronic acid in joint lubrication. Ann Rheum Dis, 1974. 33(4): p. 318-
26.
16. Walker, P.S., et al., Behaviour of synovial fluid on surfaces of articular cartilage. A scanning
electron microscope study. Ann Rheum Dis, 1969. 28(1): p. 1-14.
17. Walker, P.S., et al., Mode of aggregation of hyaluronic acid protein complex on the surface of
articular cartilage. Ann Rheum Dis, 1970. 29(6): p. 591-602.
18. Jay, G.D., Characterization of a bovine synovial fluid lubricating factor. I. Chemical, surface activity
and lubricating properties. Connect Tissue Res, 1992. 28(1-2): p. 71-88.
19. Jay, G.D., K. Haberstroh, and C.J. Cha, Comparison of the boundary-lubricating ability of bovine
synovial fluid, lubricin, and Healon. Journal of biomedical materials research, 1998. 40(3): p. 414-8.
20. Swann, D.A., H.S. Slayter, and F.H. Silver, The molecular structure of lubricating glycoprotein-I, the
boundary lubricant for articular cartilage. J Biol Chem, 1981. 256(11): p. 5921-5.
21. Zappone, B., et al., Molecular aspects of boundary lubrication by human lubricin: effect of disulfide
bonds and enzymatic digestion. Langmuir, 2008. 24(4): p. 1495-508.
22. Gleghorn, J.P., et al., Boundary mode lubrication of articular cartilage by recombinant human
lubricin. J Orthop Res, 2009. 27(6): p. 771-7.
23. Gale, L.R., et al., Boundary lubrication of joints - Characterization of surface-active phospholipids
found on retrieved implants. Acta Orthopaedica, 2007. 78(3): p. 309-314.
24. Hills, B.A., Oligolamellar Lubrication of Joints by Surface-Active Phospholipid. Journal of
Rheumatology, 1989. 16(1): p. 82-91.
25. Hills, B.A., Boundary lubrication in vivo. Proceedings of the Institution of Mechanical Engineers.
Part H, Journal of engineering in medicine, 2000. 214(1): p. 83-94.
53
26. Hills, B.A. and B.D. Butler, Surfactants Identified in Synovial-Fluid and Their Ability to Act as
Boundary Lubricants. Annals of the Rheumatic Diseases, 1984. 43(4): p. 641-648.
27. Jay, G.D. and C.J. Cha, The effect of phospholipase digestion upon the boundary lubricating ability
of synovial fluid. Journal of Rheumatology, 1999. 26(11): p. 2454-2457.
28. Pickard, J.E., et al., Investigation into the effects of proteins and lipids on the frictional properties of
articular cartilage. Biomaterials, 1998. 19(19): p. 1807-1812.
29. Schwarz, I.M. and B.A. Hills, Surface-active phospholipid as the lubricating component of lubricin.
Br J Rheumatol, 1998. 37(1): p. 21-26.
30. Flannery, C.R., et al., Articular cartilage superficial zone protein (SZP) is homologous to
megakaryocyte stimulating factor precursor and is a multifunctional proteoglycan with potential
growth-promoting, cytoprotective, and lubricating properties in cartilage metabolism. Biochemical
and Biophysical Research Communications, 1999. 254(3): p. 535-541.
31. Jay, G.D., et al., Homology of lubricin and superficial zone protein (SZP): products of
megakaryocyte stimulating factor (MSF) gene expression by human synovial fibroblasts and
articular chondrocytes localized to chromosome 1q25. J Orthop Res, 2001. 19(4): p. 677-87.
32. Schumacher, B.L., et al., A novel proteoglycan synthesized and secreted by chondrocytes of the
superficial zone of articular cartilage. Arch Biochem Biophys, 1994. 311(1): p. 144-52.
33. Ateshian, G.A., H.Q. Wang, and W.M. Lai, The role of interstitial fluid pressurization and surface
porosities on the boundary friction of articular cartilage. Journal of Tribology-Transactions of the
Asme, 1998. 120(2): p. 241-248.
34. Park, S., et al., Cartilage interstitial fluid load support in unconfined compression. Journal of
Biomechanics, 2003. 36(12): p. 1785-96.
35. Soltz, M.A. and G.A. Ateshian, Experimental verification and theoretical prediction of cartilage
interstitial fluid pressurization at an impermeable contact interface in confined compression. Journal
of Biomechanics, 1998. 31(10): p. 927-934.
36. Soltz, M.A. and G.A. Ateshian, Interstitial fluid pressurization during confined compression cyclical
loading of articular cartilage. Ann Biomed Eng, 2000. 28(2): p. 150-9.
54
37. Krishnan, R., et al., Removal of the superficial zone of bovine articular cartilage does not increase
its frictional coefficient. Osteoarthritis Cartilage, 2004. 12(12): p. 947-55.
38. Krishnan, R., M. Kopacz, and G.A. Ateshian, Experimental verification of the role of interstitial fluid
pressurization in cartilage lubrication. Journal of Orthopaedic Research, 2004. 22(3): p. 565-70.
39. Caligaris, M. and G.A. Ateshian, Effects of sustained interstitial fluid pressurization under migrating
contact area, and boundary lubrication by synovial fluid, on cartilage friction. Osteoarthritis
Cartilage, 2008. 16(10): p. 1220-7.
40. Ahmed, A.M., D.L. Burke, and A. Yu, In-vitro measurement of static pressure distribution in synovial
joints--Part II: Retropatellar surface. J Biomech Eng, 1983. 105(3): p. 226-36.
41. Caligaris, M., Migrating contact area promotes low friction and wear in articular joints. 2010,
Columbia University. p. xiv, 144 leaves, bound.
42. Walker, P.S. and J.V. Hajek, The load-bearing area in the knee joint. Journal of Biomechanics,
1972. 5(6): p. 581-9.
43. Huberti, H.H. and W.C. Hayes, Patellofemoral contact pressures. The influence of q-angle and
tendofemoral contact. The Journal of bone and joint surgery. American volume, 1984. 66(5): p. 715-
24.
44. von Eisenhart, R., et al., Quantitative determination of joint incongruity and pressure distribution
during simulated gait and cartilage thickness in the human hip joint. Journal of Orthopaedic
Research, 1999. 17(4): p. 532-539.
45. Simon, W.H., Wear properties of articular cartilage in vitro. Journal of Biomechanics, 1971. 4(5): p.
379-89.
46. Meachim, G., Light microscopy of Indian ink preparations of fibrillated cartilage. Annals of the
Rheumatic Diseases, 1972. 31(6): p. 457-64.
47. Clarke, I.C., R. Contini, and R.M. Kenedi, Friction and Wear Studies of Articular-Cartilage -
Scanning Electron-Microscope Study. Journal of Lubrication Technology-Transactions of the Asme,
1975. 97(3): p. 358-368.
48. Podsiadlo, P. and G.W. Stachowiak, Numerical-Analysis of Shape of Wear Particles from Arthritic
and Asymptomatic Synovial Joints. Journal of Orthopaedic Rheumatology, 1995. 8(3): p. 155-160.
55
49. Podsiadlo, P., M. Kuster, and G.W. Stachowiak, Numerical analysis of wear particles from non-
arthritic and osteoarthritic human knee joints. Wear : an international journal on the science and
technology of friction lubrication and wear, 1997. 210(1-2): p. 318-325.
50. Stachowiak, G.W. and P. Podsiadlo, Analysis of wear particle boundaries found in sheep knee
joints during in vitro wear tests without muscle compensation. Journal of Biomechanics, 1997.
30(4): p. 415-419.
51. Ozturk, H.E., et al., The effect of surface-active phospholipids on the lubrication of osteoarthritic
sheep knee joints: Friction. Tribology letters, 2004. 16(4): p. 283-289.
52. Peng, Z., Osteoarthritis diagnosis using wear particle analysis technique: Investigation of
correlation between particle and cartilage surface in walking process. Wear : an international
journal on the science and technology of friction lubrication and wear, 2007. 262(5-6): p. 630-640.
53. Lipshitz, H., R. Etheredge, 3rd, and M.J. Glimcher, In vitro wear of articular cartilage. The Journal of
bone and joint surgery. American volume, 1975. 57(4): p. 527-34.
54. Lipshitz, H. and M.J. Glimcher, In vitro studies of the wear of articular cartilage--II. Characteristics of
the wear of articular cartilage when worn against stainless steel plates having characterized
surfaces. Wear : an international journal on the science and technology of friction lubrication and
wear, 1979. 52(2): p. 297-339.
55. Marnell, P. and R.K. White, Quantitative-Analysis of Joint Lubrication. Wear : an international
journal on the science and technology of friction lubrication and wear, 1980. 61(2): p. 203-218.
56. Bank, R.A., et al., A simplified measurement of degraded collagen in tissues: application in healthy,
fibrillated and osteoarthritic cartilage. Matrix biology : journal of the International Society for Matrix
Biology, 1997. 16(5): p. 233-43.
57. Verberne, G., et al., Techniques for assessment of wear between human cartilage surfaces. Wear :
an international journal on the science and technology of friction lubrication and wear, 2009.
266(11-12): p. 1216-1223.
58. Bae, W.C., et al., Indentation testing of human cartilage - Sensitivity to articular surface
degeneration. Arthritis and rheumatism, 2003. 48(12): p. 3382-3394.
56
59. Parsons, I.M.t., P.J. Millett, and J.J. Warner, Glenoid wear after shoulder hemiarthroplasty:
quantitative radiographic analysis. Clinical orthopaedics and related research, 2004(421): p. 120-5.
60. Temple-Wong, M.M., et al., Biomechanical, structural, and biochemical indices of degenerative and
osteoarthritic deterioration of adult human articular cartilage of the femoral condyle. Osteoarthritis
Cartilage, 2009. 17(11): p. 1469-76.
61. Forster, H. and J. Fisher, The influence of continuous sliding and subsequent surface wear on the
friction of articular cartilage. Proceedings of the Institution of Mechanical Engineers. Part H, Journal
of engineering in medicine, 1999. 213(4): p. 329-45.
62. Graindorge, S.L. and G.W. Stachowiak, Changes occurring in the surface morphology of articular
cartilage during wear. Wear : an international journal on the science and technology of friction
lubrication and wear, 2000. 241(2): p. 143-150.
63. Northwood, E., J. Fisher, and R. Kowalski, Investigation of the friction and surface degradation of
innovative chondroplasty materials against articular cartilage. Proceedings of the Institution of
Mechanical Engineers. Part H, Journal of engineering in medicine, 2007. 221(H3): p. 263-279.
64. Katta, J., et al., Effect of nominal stress on the long term friction, deformation and wear of native
and glycosaminoglycan deficient articular cartilage. Osteoarthritis Cartilage, 2009. 17(5): p. 662-8.
65. Li, L., et al., Computational wear simulation of patellofemoral articular cartilage during in vitro
testing. Journal of Biomechanics, 2011. 44(8): p. 1507-13.
66. Brandt, K.D., et al., Yet more evidence that osteoarthritis is not a cartilage disease. Ann Rheum
Dis, 2006. 65(10): p. 1261-4.
67. Kaufman, K.R., et al., Gait characteristics of patients with knee osteoarthritis. Journal of
Biomechanics, 2001. 34(7): p. 907-15.
68. Wluka, A.E., et al., The determinants of change in tibial cartilage volume in osteoarthritic knees.
Arthritis Rheum, 2002. 46(8): p. 2065-72.
69. Schouten, J.S., F.A. van den Ouweland, and H.A. Valkenburg, A 12 year follow up study in the
general population on prognostic factors of cartilage loss in osteoarthritis of the knee. Ann Rheum
Dis, 1992. 51(8): p. 932-7.
70. Charnley, J., Tissue Reactions to Polytetrafluorethylene. Lancet, 1963. 2(732): p. 1379-&.
57
71. Charnley, J., The long-term results of low-friction arthroplasty of the hip performed as a primary
intervention. Journal of Bone and Joint Surgery-British Volume, 1972. 54(1): p. 61-76.
72. Griffith, M.J., et al., The classic: socket wear in Charnley low friction arthroplasty of the hip. Clinical
orthopaedics and related research, 2006. 448: p. 3-9.
73. CDC/NCHS National Hospital Discharge Survey: Number, rate, and standard error of all-listed
surgical and nonsurgical procedures for discharges from short-stay hospitals,
by selected procedure categories: United States, 2010. National Health Statistics Report 2013; 1-3].
Available from: http://www.cdc.gov/nchs/data/nhds/4procedures/2010pro4_numberrate.pdf.
74. Blomfeldt, R., et al., A randomised controlled trial comparing bipolar hemiarthroplasty with total hip
replacement for displaced intracapsular fractures of the femoral neck in elderly patients. Journal of
Bone and Joint Surgery-British Volume, 2007. 89B(2): p. 160-165.
75. Gartsman, G.M., T.S. Roddey, and S.M. Hammerman, Shoulder arthroplasty with or without
resurfacing of the glenoid in patients who have osteoarthritis. Journal of Bone and Joint Surgery-
American Volume, 2000. 82A(1): p. 26-34.
76. Baumgarten, K.M., C.J. Lashgari, and K. Yamaguchi, Glenoid resurfacing in shoulder arthroplasty:
indications and contraindications. Instr Course Lect, 2004. 53: p. 3-11.
77. Lindstrand, A., et al., The patella in total knee arthroplasty: resurfacing or nonresurfacing of patella.
Knee Surg Sports Traumatol Arthrosc, 2001. 9 Suppl 1: p. S21-3.
78. Orfaly, R.M., et al., A prospective functional outcome study of shoulder arthroplasty for
osteoarthritis with an intact rotator cuff. J Shoulder Elbow Surg, 2003. 12(3): p. 214-21.
79. Cuckler, J.M. and J.R. Tamarapalli, An algorithm for the management of femoral neck fractures.
Orthopedics, 1994. 17(9): p. 789-92.
80. Rodriguez-Merchan, E.C., Displaced intracapsular hip fractures: hemiarthroplasty or total
arthroplasty? Clin Orthop Relat Res, 2002(399): p. 72-7.
81. Boileau, P., et al., Arthroplasty of the shoulder. Journal of Bone and Joint Surgery-British Volume,
2006. 88(5): p. 562-75.
82. Smith, K.L. and F.A. Matsen, 3rd, Total shoulder arthroplasty versus hemiarthroplasty. Current
trends. Orthop Clin North Am, 1998. 29(3): p. 491-506.
58
83. Connor, P.M. and D.F. D'Alessandro, Role of hemiarthroplasty for proximal humeral fractures. J
South Orthop Assoc, 1995. 4(1): p. 9-23.
84. Loew, M., et al., Influence of the design of the prosthesis on the outcome after hemiarthroplasty of
the shoulder in displaced fractures of the head of the humerus. Journal of Bone and Joint Surgery-
British Volume, 2006. 88(3): p. 345-50.
85. Giliberty, R.P., Hemiarthroplasty of the Hip Using a Low-Friction Bipolar Endoprosthesis. Clinical
orthopaedics and related research, 1983(175): p. 86-92.
86. Barnes, C.L., D.N. Collins, and C.L. Nelson, Cup arthroplasty, surface replacement arthroplasty,
and femoral head resurfacing for osteonecrosis. Semin Arthroplasty, 1991. 2(3): p. 222-7.
87. Grecula, M.J., et al., Endoprostheses for osteonecrosis of the femoral head. A comparison of four
models in young patients. Int Orthop, 1995. 19(3): p. 137-43.
88. Champion, L.M. and S.A. McNally, Dislocation after revision of hemiarthroplasty to total hip
replacement. Injury, 2004. 35(2): p. 161-4.
89. Haidukewych, G.J., T.A. Israel, and D.J. Berry, Long-term survivorship of cemented bipolar
hemiarthroplasty for fracture of the femoral neck. Clinical orthopaedics and related research,
2002(403): p. 118-26.
90. Hasan, S.S., et al., Characteristics of unsatisfactory shoulder arthroplasties. J Shoulder Elbow
Surg, 2002. 11(5): p. 431-41.
91. Hedtmann, A. and A. Werner, [Shoulder arthroplasty in rheumatoid arthritis]. Orthopade, 2007.
36(11): p. 1050-61.
92. Sperling, J.W., R.H. Cofield, and S.P. Steinmann, Shoulder arthroplasty for osteoarthritis secondary
to glenoid dysplasia. The Journal of bone and joint surgery. American volume, 2002. 84-A(4): p.
541-6.
93. Huey, D.J., J.C. Hu, and K.A. Athanasiou, Unlike bone, cartilage regeneration remains elusive.
Science, 2012. 338(6109): p. 917-21.
94. Nooeaid, P., et al., Osteochondral tissue engineering: scaffolds, stem cells and applications. J Cell
Mol Med, 2012. 16(10): p. 2247-70.
59
95. Ateshian, G.A., et al., An asymptotic solution for the contact of two biphasic cartilage layers. Journal
of Biomechanics, 1994. 27(11): p. 1347-60.
96. Ateshian, G.A., et al., Finite deformation biphasic material properties of bovine articular cartilage
from confined compression experiments. Journal of Biomechanics, 1997. 30(11-12): p. 1157-64.
97. Castro, N.J., S.A. Hacking, and L.G. Zhang, Recent progress in interfacial tissue engineering
approaches for osteochondral defects. Ann Biomed Eng, 2012. 40(8): p. 1628-40.
98. Katta, J.K., et al., Friction and wear behavior of poly(vinyl alcohol)/poly(vinyl pyrrolidone) hydrogels
for articular cartilage replacement. Journal of biomedical materials research, 2007. 83A(2): p. 471-
479.
99. McClure, G., et al., Determination of lubricating film thickness for permeable hydrogel and non-
permeable polyurethane layers bonded to a rigid substrate with particular reference to cushion form
hip joint replacements. Proceedings of the Institution of Mechanical Engineers. Part H, Journal of
engineering in medicine, 1996. 210(2): p. 89-93.
100. Northwood, E. and J. Fisher, Erratum to "A multi-directional in vitro investigation into friction,
damage and wear of innovative chondroplasty materials against articular cartilage" (vol 22, pg 834,
2007). Clinical Biomechanics, 2007. 22(10): p. 1132-1132.
101. Northwood, E. and J. Fisher, A multi-directional in vitro investigation into friction, damage and wear
of innovative chondroplasty materials against articular cartilage. Clinical Biomechanics, 2007. 22(7):
p. 834-842.
102. Akens, M.K., et al., In vitro studies of a photo-oxidized bovine articular cartilage. J Vet Med A
Physiol Pathol Clin Med, 2002. 49(1): p. 39-45.
103. Hunter, S.A., et al., Effects of matrix stabilization when using glutaraldehyde on the material
properties of porcine meniscus. Journal of Biomedical Materials Research Part A, 2003. 67A(4): p.
1245-1254.
104. Hunter, S.A., et al., Meniscal material properties are minimally affected by matrix stabilization using
glutaraldehyde and glycation with ribose. Journal of Orthopaedic Research, 2005. 23(3): p. 555-
561.
60
105. Hunter, S.A., et al., Biomechanical and biologic effects of meniscus stabilization using triglycidyl
amine. Journal of Biomedical Materials Research Part A, 2010. 93(1): p. 235-42.
106. Shahgaldi, B.F., et al., Repair of Cartilage Lesions Using Biological Implants - a Comparative
Histological and Biomechanical Study in Goats. Journal of Bone and Joint Surgery-British Volume,
1991. 73(1): p. 57-64.
107. Daly, R.C., et al., Long-term results of aortic valve replacement with nonviable homografts.
Circulation, 1991. 84(5 Suppl): p. III81-8.
108. Magilligan, D.J., The future of bioprosthetic valves. ASAIO Trans, 1988. 34(4): p. 1031-2.
109. Siddiqui, R.F., J.R. Abraham, and J. Butany, Bioprosthetic heart valves: modes of failure.
Histopathology, 2009. 55(2): p. 135-44.
110. Hilbert, S.L. and V.J. Ferrans, Porcine aortic valve bioprostheses: morphologic and functional
considerations. J Long Term Eff Med Implants, 1992. 2(2-3): p. 99-112.
111. Schoen, F.J., Cardiac valve prostheses: pathological and bioengineering considerations. J Card
Surg, 1987. 2(1): p. 65-108.
112. Arcidiacono, G., A. Corvi, and T. Severi, Functional analysis of bioprosthetic heart valves. Journal
of Biomechanics, 2005. 38(7): p. 1483-90.
113. Billiar, K.L. and M.S. Sacks, Biaxial mechanical properties of the native and glutaraldehyde-treated
aortic valve cusp: Part II--A structural constitutive model. J Biomech Eng, 2000. 122(4): p. 327-35.
114. Gratzer, P.F., C.A. Pereira, and J.M. Lee, Solvent environment modulates effects of glutaraldehyde
crosslinking on tissue-derived biomaterials. Journal of biomedical materials research, 1996. 31(4):
p. 533-43.
115. Sacks, M.S. and C.J. Chuong, Orthotropic mechanical properties of chemically treated bovine
pericardium. Ann Biomed Eng, 1998. 26(5): p. 892-902.
116. Flomenbaum, M.A. and F.J. Schoen, Effects of fixation back pressure and antimineralization
treatment on the morphology of porcine aortic bioprosthetic valves. J Thorac Cardiovasc Surg,
1993. 105(1): p. 154-64.
117. Vesely, I., Analysis of the Medtronic Intact bioprosthetic valve. Effects of "zero-pressure" fixation. J
Thorac Cardiovasc Surg, 1991. 101(1): p. 90-9.
61
118. Wells, S.M. and M.S. Sacks, Effects of fixation pressure on the biaxial mechanical behavior of
porcine bioprosthetic heart valves with long-term cyclic loading. Biomaterials, 2002. 23(11): p.
2389-99.
119. Ersek, R.A. and A.G. Delerm, Processed irradiated bovine cartilage for nasal reconstruction. Ann
Plast Surg, 1988. 20(6): p. 540-6.
120. Gibeault, J.D., et al., Use of cross-linked bovine pericardium as a disc replacement in the rabbit
temporomandibular joint. J Oral Maxillofac Surg, 1989. 47(8): p. 828-33.
121. Heatley, F.W. and P.R. Allen, Glutaraldehyde treated bovine fibrocartilage arthroplasty for the
scaphoid: a case history and discussion. Biomaterials, 1986. 7(4): p. 305-7.
122. Kangesu, L., T.E. Goodacre, and P.R. Stanley, Survival of irradiated glutaraldehyde preserved
bovine cartilage in nasal reconstruction: a retrospective study. Br J Plast Surg, 1991. 44(7): p. 483-
5.
123. Canham, W. and W. Stanish, A study of the biological behavior of the meniscus as a transplant in
the medial compartment of a dog's knee. Am J Sports Med, 1986. 14(5): p. 376-9.
124. Heatley, F.W. and W.J. Revell, The Use of Meniscal Fibrocartilage as a Surface Arthroplasty to
Effect the Repair of Osteochondral Defects - an Experimental-Study. Biomaterials, 1985. 6(3): p.
161-168.
125. Heatley, F.W. and W.J. Revell, The Use of Glutaraldehyde-Treated Bovine Meniscal Fibrocartilage
to Replace Osteochondral Defects of the Dog Patella. Journal of Bone and Joint Surgery-British
Volume, 1986. 68(5): p. 842-842.
126. Powers, D.L., M.E. Davenport, and P.J. Wisnewski, Glutaraldehyde-cross-linked meniscal
allografts: clinical, gross, and histological results. J Invest Surg, 1988. 1(4): p. 249-57.
127. Revell, W.J. and F.W. Heatley, Functional Restoration of an Articular Surface Using a Heterotopic
Xenograft - Biology of Host-Implant Interactions in the Canine Patella. Biomaterials, 1988. 9(2): p.
173-180.
128. Wisnewski, P.J., D.L. Powers, and J.M. Kennedy, Glutaraldehyde-cross-linked meniscal allografts:
mechanical properties. J Invest Surg, 1988. 1(4): p. 259-66.
62
129. Stanescu, R. and V. Stanescu, In vitro protection of the articular surface by cross-linking agents. J
Rheumatol, 1988. 15(11): p. 1677-82.
130. Radin, E.L., et al., Factors influencing articular cartilage wear in vitro. Arthritis Rheum, 1982. 25(8):
p. 974-80.
131. Soltz, M.A. and G.A. Ateshian, A Conewise Linear Elasticity mixture model for the analysis of
tension-compression nonlinearity in articular cartilage. J Biomech Eng, 2000. 122(6): p. 576-86.
132. Huang, C.Y., V.C. Mow, and G.A. Ateshian, The role of flow-independent viscoelasticity in the
biphasic tensile and compressive responses of articular cartilage. J Biomech Eng, 2001. 123(5): p.
410-7.
133. Huang, C.Y., et al., Experimental verification of the roles of intrinsic matrix viscoelasticity and
tension-compression nonlinearity in the biphasic response of cartilage. J Biomech Eng, 2003.
125(1): p. 84-93.
134. Ateshian, G.A. and H. Wang, A theoretical solution for the frictionless rolling contact of cylindrical
biphasic articular cartilage layers. Journal of Biomechanics, 1995. 28(11): p. 1341-55.
135. Caligaris, M., et al., Investigation of the frictional response of osteoarthritic human tibiofemoral
joints and the potential beneficial tribological effect of healthy synovial fluid. Osteoarthritis Cartilage,
2009. 17(10): p. 1327-32.
136. Huang, A.H., et al., Sliding contact loading enhances the tensile properties of mesenchymal stem
cell-seeded hydrogels. Eur Cell Mater, 2012. 24: p. 29-45.
137. Maas, S.A., et al., FEBio: finite elements for biomechanics. J Biomech Eng, 2012. 134(1): p.
011005.
138. Mow, V.C., et al., Biphasic creep and stress relaxation of articular cartilage in compression? Theory
and experiments. J Biomech Eng, 1980. 102(1): p. 73-84.
139. Ateshian, G.A., The role of interstitial fluid pressurization in articular cartilage lubrication. Journal of
Biomechanics, 2009. 42(9): p. 1163-76.
140. Williams, S.K., et al., Prolonged storage effects on the articular cartilage of fresh human
osteochondral allografts. The Journal of bone and joint surgery. American volume, 2003. 85-A(11):
p. 2111-20.
63
141. Carter, M.J., I.M. Basalo, and G.A. Ateshian, The temporal response of the friction coefficient of
articular cartilage depends on the contact area. Journal of Biomechanics, 2007. 40(14): p. 3257-60.
142. Erdil, M., et al., Osteochondral autologous graft transfer system in the knee; mid-term results. Knee,
2013. 20(1): p. 2-8.
143. Hangody, L., et al., Arthroscopic autogenous osteochondral mosaicplasty for the treatment of
femoral condylar articular defects. A preliminary report. Knee Surg Sports Traumatol Arthrosc,
1997. 5(4): p. 262-7.
144. Hangody, L., et al., Mosaicplasty for the treatment of articular cartilage defects: application in
clinical practice. Orthopedics, 1998. 21(7): p. 751-6.
145. Hangody, L., et al., Autologous osteochondral mosaicplasty. Surgical technique. J Bone Joint Surg
Am, 2004. 86-A Suppl 1: p. 65-72.
146. Hangody, L., et al., Autologous osteochondral grafting--technique and long-term results. Injury,
2008. 39 Suppl 1: p. S32-9.
147. Imade, S., et al., Effectiveness and limitations of autologous osteochondral grafting for the
treatment of articular cartilage defects in the knee. Knee Surg Sports Traumatol Arthrosc, 2012.
20(1): p. 160-5.
148. Ahmad, C.S., et al., Biomechanical and topographic considerations for autologous osteochondral
grafting in the knee. Am J Sports Med, 2001. 29(2): p. 201-6.
149. Lange, J., et al., [Results of SaluCartilage implantation for stage IV chondral defects in the knee
joint area]. Unfallchirurg, 2006. 109(3): p. 193-9.
150. Meyer, C., et al., [Dislocation of artificial cartilage (SaluCartialge)]. Unfallchirurg, 2005. 108(2): p.
163-6.
151. Lipshitz, H., R. Etheredge, and M.J. Glimcher, Invitro Wear of Articular-Cartilage .1. Hydroxyproline,
Hexosamine, and Amino-Acid Composition of Bovine Articular-Cartilage as a Function of Depth
from Surface - Hydroxyproline Content of Lubricant and Wear Debris as a Measure of Wear. The
Journal of bone and joint surgery. American volume, 1975. A 57(4): p. 527-534.
64
152. Maloney, W.J., et al., Isolation and Characterization of Wear Particles Generated in Patients Who
Have Had Failure of a Hip-Arthroplasty without Cement. The Journal of bone and joint surgery.
American volume, 1995. 77(9): p. 1301-1310.
153. Farndale, R.W., D.J. Buttle, and A.J. Barrett, Improved quantitation and discrimination of sulphated
glycosaminoglycans by use of dimethylmethylene blue. Biochimica et biophysica acta, 1986.
883(2): p. 173-7.
154. Stegeman.H and K. Stalder, Determination of Hydroxyproline. Clinica Chimica Acta, 1967. 18(2): p.
267-&.
155. Hollander, A.P., et al., Increased damage to type II collagen in osteoarthritic articular cartilage
detected by a new immunoassay. Journal of Clinical Investigation, 1994. 93(4): p. 1722-32.
156. Krahn, K.N., et al., Fluorescently labeled collagen binding proteins allow specific visualization of
collagen in tissues and live cell culture. Analytical Biochemistry, 2006. 350(2): p. 177-85.
157. Gleghorn, J.P. and L.J. Bonassar, Lubrication mode analysis of articular cartilage using Stribeck
surfaces. Journal of Biomechanics, 2008. 41(9): p. 1910-8.
158. Gleghorn, J.P., et al., Analysis of frictional behavior and changes in morphology resulting from
cartilage articulation with porous polyurethane foams. Journal of Orthopaedic Research, 2010.
28(10): p. 1292-9.
159. Jay, G.D., D.A. Harris, and C.J. Cha, Boundary lubrication by lubricin is mediated by O-linked
beta(1-3)Gal-GalNAc oligosaccharides. Glycoconjugate Journal, 2001. 18(10): p. 807-815.
160. Lee, D.W., X. Banquy, and J.N. Israelachvili, Stick-slip friction and wear of articular joints. Proc Natl
Acad Sci U S A, 2013. 110(7): p. E567-74.
161. Gleghorn, J.P., et al., Boundary mode frictional properties of engineered cartilaginous tissues -
Discussion with reviewers. European Cells and Materials, 2007. 14: p. 28-29.
162. Oungoulian, S.R., et al., Articular cartilage wear characterization with a particle sizing and counting
analyzer. J Biomech Eng, 2013. 135(2): p. 024501.
163. Byers, P.D., et al., Histological and biochemical studies on cartilage from osteoarthrotic femoral
heads with special reference to surface characteristics. Connect Tissue Res, 1977. 5(1): p. 41-9.
65
164. Clarke, I.C., Wear-screening and joint simulation studies vs. materials selection and prosthesis
design. Crit Rev Biomed Eng, 1982. 8(1): p. 29-91.
165. Foy, J.R., et al., Effect of phospholipidic boundary lubrication in rigid and compliant
hemiarthroplasty models. Proceedings of the Institution of Mechanical Engineers. Part H, Journal of
engineering in medicine, 1999. 213(1): p. 5-18.
166. Maistrelli, G., V. Sessa, and V.L. Fornasier, Response of the articular cartilage to weight-bearing:
comparison of hemiarthroplasty with ceramic and cobalt-chromium head in dogs. Ital J Orthop
Traumatol, 1991. 17(3): p. 387-93.
167. Patel, A.M. and M. Spector, Tribological evaluation of oxidized zirconium using an articular cartilage
counterface: A novel material for potential use in hemiarthroplasty. Biomaterials, 1997. 18(5): p.
441-447.
168. Chan, S.M.T., et al., Friction and Wear of Hemiarthroplasty Biomaterials in Reciprocating Sliding
Contact With Articular Cartilage. Journal of Tribology-Transactions of the Asme, 2011. 133(4).
169. Kelly, T.A., et al., Spatial and temporal development of chondrocyte-seeded agarose constructs in
free-swelling and dynamically loaded cultures. Journal of Biomechanics, 2006. 39(8): p. 1489-97.
170. Buckley, M.R., et al., High-resolution spatial mapping of shear properties in cartilage. Journal of
Biomechanics, 2010. 43(4): p. 796-800.
171. Buckley, M.R., et al., Mapping the depth dependence of shear properties in articular cartilage.
Journal of Biomechanics, 2008. 41(11): p. 2430-7.