PRE-CLEANING (in procedure room) MANUAL HIGH-LEVEL DISINFECTION 3 I RINSING AFTER CLEANING WITH ENZYMATIC FLUID V RINSING AFTER MANUAL DISINFECTION ALCOHOL FLUSH V STORAGE 1. Immediately wipe down the insertion tube and distal tip with a soft cloth soaked in detergent solution as per manufacturer’s recommendation. 2. Flush the air and water channels (and auxiliary channel if applicable) for 10 seconds. 3. Suction detergent solution through the suction channel. Alternate suctioning detergent solution and air several times until the solution is visibly clean. Finish by suctioning air. 4. Attach the waterproof cap securely to the EVE connector. Detach all removable components. Transport the endo- scope and removable components in a covered container to the cleaning room. FUJINON PRESSURE GUAGE LT-7 5 10 15 20 25 30 35 40 FUJINON PRESSURE GUAGE LT-7 5 10 15 20 25 30 35 40 FUJINON PRESSURE GUAGE LT-7 5 10 15 20 25 30 35 40 I CLEANING WITH ENZYMATIC FLUID LEAK TESTING 1. Immerse the endoscope and removable components (air/water and suction buttons, and auxiliary J-Port connector 2 ) in freshly prepared warm enzymatic detergent. Follow the detergent manufacturer’s instructions. Thoroughly clean all external surfaces and removable components. 2. Brush all cylinders thoroughly. 3. Brush all channels of endoscope until there is no debris on the cleaning brush. 4. Brush straight back into the suction channel through to the distal end of the insertion tube. 5. Brush the suction channel by lining the brush with the dot of the endoscope housing, then place the brush at a 1 o’clock angle. 6. Brush down the biopsy channel through to the distal end. 7. Attach the Fujinon cleaning adapter supplied with the endoscope and flush all channels 2 three times. 500 Series 500 Series 450 Series 250 Series FUJINON ENDOSCOPE GUIDE CLEANING HIGH-LEVEL DISINFECTION STORAGE 1. Ensure the leak tester has no fluid/moisture inside the tubing, connector or bulb. 2. Attach the leak tester to the vent port and pressurize the endoscope to 18.5 – 21.5kPa (140 –160 mmHg). Flex the distal portion of the endoscope in all directions, observing for a drop in pressure. If no leaks are found, release pressure and disconnect the leak tester. Leak test should last a minimum of 30 seconds. 3. If a leak is detected, follow the manufacturer’s instructions 1 to send the endoscope in for repair. Warning: This guide is a summary of the steps necessary to properly clean your Fujinon endoscope. Follow the detailed instructions described in Fujinon’s Cleaning, Disinfection and Storage Manual. Important: • Wear appropriate personal protective equipment. • Thoroughly clean the endoscope to ensure effective disinfection/sterilization. • Visually inspect the equipment after cleaning. If debris remains, ensure the debris is removed and repeat the procedure. 1 Refer to Delayed Reprocessing Post Procedure Guidelines Care & Handling document for returning the scope for repair. 2 Auxiliary J-Port connectors and channels are found only on colonoscopes and therapeutic gastroscopes. All channels must be flushed and properly reprocessed, including air outlet channel on DBE endoscopes. 3 If using an automated endoscope reprocessor (AER) for chemical disinfection, refer to AER manufacturer’s instructions. When using an AER a “cool down period” must be allowed for the endoscope to return to room temperature inside and out. Failure to allow for “cool down” may result in damage to the endoscope. 1. Rinse the endoscope and all removable components with potable water. Flush water through all channels 2 three times using the Fujinon cleaning adapter. Purge water from all channels using forced air. 1. Immerse the endoscope and removable components in high-level disinfectant solution. 2. Verify the minimum effective concentration of the disinfectant solution PRIOR to use. 3. Follow the disinfectant manufacturer’s instructions for soaking parameters and other detailed instructions. 4. Inject disinfectant solution into all channels using a clean Fujinon cleaning adapter. Ensure that all channels are filled with the disinfectant and that no air pockets remain within the channels. 1. Thoroughly rinse all surfaces and removable components. Flush sterilized water through all channels 2 using a clean Fujinon cleaning adapter. 2. After rinsing the endoscope and removable components, dry the outside of the endoscope with a clean lint free cloth and flush air through all channels. DO NOT USE a pressurized hose to flush the auxiliary J-Port channel. High pressure may cause damage to the auxiliary J-Port channel. 1. Flush all channels 2 with 70% alcohol followed by air to facilitate drying while in storage. 1. Store all removable components separately to promote aeration of channels and ports. Do not store endoscope in its carrying case. Ensure that the endoscope is hung securely with the insertion tube and light guide cable hanging freely in a well-vented storage cabinet. 2. Remove the protective cap from the end of the EVE connector while storing. 1 2 3 4 5 6 7 8 © 2011 Fujinon Corporation FCP-1/11 Fujinon Inc., 10 High Point Drive, Wayne, NJ 07470 1.800.490.0661 www.fujinonendoscopy.com Fuji-1060 Endoscope Cleaning Poster 12-10-10_Layout 1 12/10/10 3:21 PM Page 1

Upload: others

Post on 27-Jul-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: FUJINON 1 PRE-CLEANING (in procedure room) I RINSING AFTER CLEANING … · 2015-05-05 · detergent manufacturer’s instructions. Thoroughly clean all external surfaces and removable

PRE-CLEANING (in procedure room)

MANUAL HIGH-LEVEL DISINFECTION3

I RINSING AFTER CLEANING WITH ENZYMATIC FLUID

V RINSING AFTER MANUAL DISINFECTION

ALCOHOL FLUSH

V STORAGE

1. Immediately wipe down the insertion tube and distal tipwith a soft cloth soaked in detergent solution as per manufacturer’s recommendation.

2. Flush the air and water channels (and auxiliary channel ifapplicable) for 10 seconds.

3. Suction detergent solution through the suction channel. Alternate suctioning detergent solution and air severaltimes until the solution is visibly clean. Finish by suctioning air.

4. Attach the waterproof cap securely to the EVE connector.Detach all removable components. Transport the endo-scope and removable components in a covered containerto the cleaning room.

FUJINONPRESSURE GUAGE

LT-7

5

10

15

2025

30

35

40

FUJINONPRESSURE GUAGELT-7

5

10

15

2025

30

35

40

FUJINONPRESSURE GUAGE

LT-7

5

10

15

2025

30

35

40

I CLEANING WITH ENZYMATIC FLUID

LEAK TESTING

1. Immerse the endoscope and removable components (air/water and suction buttons, and auxiliary J-Port connector2)in freshly prepared warm enzymatic detergent. Follow the detergent manufacturer’s instructions. Thoroughly clean all external surfaces and removable components.

2. Brush all cylinders thoroughly.

3. Brush all channels of endoscope until there is no debris on thecleaning brush.

4. Brush straight back into the suction channel through to the distal end of the insertion tube.

5. Brush the suction channel by lining the brush with the dot of theendoscope housing, then place the brush at a 1 o’clock angle.

6. Brush down the biopsy channel through to the distal end.

7. Attach the Fujinon cleaning adapter supplied with the endoscopeand flush all channels2 three times.

500 Series

500 Series

450 Series 250 Series

FUJINONENDOSCOPEGUIDECLEANINGHIGH-LEVEL DISINFECTIONSTORAGE 1. Ensure the leak tester has no fluid/moisture inside the tubing, connector or bulb.

2. Attach the leak tester to the vent port and pressurize the endoscope to 18.5 –21.5kPa(140–160 mmHg). Flex the distal portion of the endoscope in all directions, observingfor a drop in pressure. If no leaks are found, release pressure and disconnect the leaktester. Leak test should last a minimum of 30 seconds.

3. If a leak is detected, follow the manufacturer’s instructions1 to send the endoscope in for repair.

Warning:This guide is a summary of the steps necessary to properlyclean your Fujinon endoscope. Follow the detailed instructionsdescribed in Fujinon’s Cleaning, Disinfection and Storage Manual.

Important:• Wear appropriate personal protective equipment.

• Thoroughly clean the endoscope to ensure effective disinfection/sterilization.

• Visually inspect the equipment after cleaning. If debris remains, ensure the debris is removed and repeat the procedure.

1 Refer to Delayed Reprocessing Post Procedure Guidelines Care& Handling document for returning the scope for repair.

2 Auxiliary J-Port connectors and channels are found only on colonoscopes and therapeutic gastroscopes. All channels must be flushed and properly reprocessed, including air outlet channel on DBE endoscopes.

3 If using an automated endoscope reprocessor (AER) for chemical disinfection,refer to AER manufacturer’s instructions. When using an AER a “cool downperiod” must be allowed for the endoscope to return to room temperature inside and out. Failure to allow for “cool down” may result in damage to theendoscope.

1. Rinse the endoscope and all removable components with potable water. Flush waterthrough all channels2 three times using the Fujinon cleaning adapter. Purge waterfrom all channels using forced air.

1. Immerse the endoscope and removable components inhigh-level disinfectant solution.

2. Verify the minimum effective concentration of the disinfectant solution PRIOR to use.

3. Follow the disinfectant manufacturer’s instructions forsoaking parameters and other detailed instructions.

4. Inject disinfectant solution into all channels using a cleanFujinon cleaning adapter. Ensure that all channels arefilled with the disinfectant and that no air pockets remainwithin the channels.

1. Thoroughly rinse all surfaces and removable components. Flush sterilized waterthrough all channels2 using a clean Fujinon cleaning adapter.

2. After rinsing the endoscope and removable components, dry the outside of the endoscope with a clean lint free cloth and flush air through all channels. DO NOT USEa pressurized hose to flush the auxiliary J-Port channel. High pressure may causedamage to the auxiliary J-Port channel.

1. Flush all channels2 with 70% alcohol followed by air tofacilitate drying while in storage.

1. Store all removable components separately to promote aeration of channels andports. Do not store endoscope in its carrying case. Ensure that the endoscope is hungsecurely with the insertion tube and light guide cable hanging freely in a well-vented storage cabinet.

2. Remove the protective cap from the end of the EVE connector while storing.