Poster Design & Printing by Genigraphics ® - 800.790.4001 Further Measurement to Test Electron Conversion Theory: 116 In Measurement for Detector Calibration Sondra Miller 1,2 ; Ninel Nica, PhD 1 ; John Hardy, PhD 1 ; John Goodwin 1 Internal Conversion Coefficient (ICC) Preparation and Irradiation of the Source Impurity Identification Preliminary Results Calibrating Detector for 119m Sn Decay Schemes ABSTRACT Precise internal conversion coefficients (ICCs) are vital to the study of nuclear decay schemes, determining transition rates, spin and parity designations, and branching ratios. However, there are very few experimental tests of the calculated ICC's and in fact there are only ~10 measurements available with errors of less than 1%. Such a paucity of data complicates scientists' efforts to determine what theoretical calculations should be used to model the ICC. The goal of our present experiment is to determine the α k for the 65.7-keV M4 transition in 119Sn. However, the energy of the 119Sn x-rays is below the energy range that our HPGe detector is accurately calibrated for. The β-decay of 116In populates states in 116Sn which produce a few strong transitions with well established conversion coefficients. This allows us to calibrate our detector at the energy of the Sn x-rays, which is an essential requirement for the measurement of the 119Sn ICC. x-ray e- ~10 -18 s e- ~10 -17 s -ray L shell K shell Hole 119m Sn: Number of Photons Detected with Given Energies 24.5-25.1 [keV] Sn x-rays Uncertain efficiency calibration 151. mm and x-rays 116 In: Number of Photons Detected with Given Energies 138 [keV] and 418 [keV] peaks: α k values well-known to calibrate Sn x-rays Nuclear de-excitation energy leads to γ-ray emission or to electron emission Processes occur competitively Electron emission leaves hole; filled by higher level electron with emission of an x-ray ICC (α) measures ratio of electrons versus γ- rays emitted Can be expressed as the sum of ratios for each energy shell 1 Texas A&M University, Cyclotron Institute 2 Florida A&M University, FAMU/FSU College of Engineering 1294: 116In 1508: 116In 1369: 24Na 1072: 116In Bgd Bgd Bgd Bgd 116In: Number of Photons Detected at Given Energies; Source Nuclei of Peaks Labeled Values of α k are well-known for these two peaks Theoretical calculations agree on these values 116In ß decays to 116Sn leading to 138 [keV] and 418 [keV] γ rays and to Sn x-rays Trim excess tape to minimize impurities Adhesive Mylar tape In + 2NO 3 (aq) Adhesive Mylar tape Source ready to be irradiated 7.0 mm 0.5 in In(NO 3 ) 2 (aq) placed on Mylar tape to produce micron-thick film of indium nitrate In is isotopically purified to minimize impurities Adhesive Mylar tape placed on film following solvent evaporation Created two sources: A and B Irradiated sources by neutron activation at theTexas A&M Nuclear Science Center Measured resulting x and -rays Germanium detector Texas A&M University Cyclotron Institute Relative photopeak efficiencies calibrated to 0.15% above 50 [keV] 4 spectra recorded between ~2 and ~20 hours after activation Theoretical Values Experimental Value Good agreement between experimental and theoretical values. Data has had the room background photon counts subtracted to minimize impurities in spectra. Typical Impurities: Other elements with similar Z values, other naturally occurring isotopes Elements from the mylar substrate Other phenomena: Escape peaks; Compton background 119m Sn: α k Calculations Decay Scheme of 116 In (and 116 Sb) Transitions to 116 Sn. 24-25 [keV] Sn x-rays

Upload: others

Post on 25-Nov-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Poster Design & Printing by Genigraphics® - 800.790.4001

Further Measurement to Test Electron Conversion Theory: 116In Measurement for Detector Calibration

Sondra Miller1,2; Ninel Nica, PhD1; John Hardy, PhD1; John Goodwin1

Internal Conversion Coefficient (ICC)

Preparation and Irradiation of the Source

Impurity Identification

Preliminary ResultsCalibrating Detector for 119mSn

Decay Schemes

ABSTRACTPrecise internal conversion coefficients (ICCs) are vital to the study of nuclear decay schemes, determining transition rates, spin and parity designations, and branching ratios.

However, there are very few experimental tests of the calculated ICC's and in fact there are only ~10 measurements available with errors of less than 1%. Such a paucity of data complicates scientists' efforts to determine what theoretical calculations should be used to model the ICC.

The goal of our present experiment is to determine the αk for the 65.7-keV M4 transition in 119Sn. However, the energy of the 119Sn x-rays is below the energy range that our HPGe detector is accurately calibrated for.

The β-decay of 116In populates states in 116Sn which produce a few strong transitions with well established conversion coefficients. This allows us to calibrate our detector at the energy of the Sn x-rays, which is an essential requirement for the measurement of the 119Sn ICC.

x-ray

e-

~10-18 s

e-

~10-17 s

-ray

L shell

K shell

Hole

119mSn: Number of Photons Detected with Given Energies

24.5-25.1 [keV] Sn x-raysUncertain efficiency calibration

151. mm

and x-rays

116In: Number of Photons Detected with Given Energies

138 [keV] and 418 [keV] peaks: αkvalues well-known to calibrate Sn

x-rays

Nuclear de-excitation energy leads to γ-ray emission or to electron emission Processes occur competitively Electron emission leaves hole; filled by higher level electron with emission of an x-ray

ICC (α) measures ratio of electrons versus γ-rays emitted Can be expressed as the sum of ratios for each energy shell

1Texas A&M University, Cyclotron Institute

2Florida A&M University, FAMU/FSU College of Engineering

1294: 116In

1508: 116In

1369: 24Na

1072: 116In

BgdBgd

Bgd Bgd

116In: Number of Photons Detected at Given Energies; Source Nuclei of Peaks Labeled

Values of αk are well-known for these two peaks Theoretical calculations agree on these values

116In ß decays to 116Sn leading to 138 [keV] and 418 [keV] γ rays and to Sn x-rays

Trim excess tape to minimize impurities

Adhesive Mylar tape

In + 2NO3 (aq)

Adhesive Mylar tape

Source ready to be irradiated

7.0 mm

0.5 in

In(NO3)2(aq) placed on Mylar tape to produce micron-thick film of indium nitrate

In is isotopically purified to minimize impurities

Adhesive Mylar tape placed on film following solvent evaporationCreated two sources: A and B

Irradiated sources by neutron activation at theTexas A&M Nuclear Science Center

Measured resulting x and -raysGermanium detector

Texas A&M University Cyclotron Institute

Relative photopeak efficiencies calibrated to 0.15% above 50 [keV]

4 spectra recorded between ~2 and ~20 hours after activation

Theoretical ValuesExperimental Value

Good agreement between experimental and theoretical values.

Data has had the room background photon counts subtracted to minimize impurities in spectra.

Typical Impurities:Other elements with similar Z values, other naturally occurring isotopesElements from the mylar substrateOther phenomena: Escape peaks; Compton background

119mSn: αk Calculations

Decay Scheme of 116In (and 116Sb) Transitions to 116Sn.

24-25 [keV] Sn x-rays