GEO 600 STATUS. Andreas Freise on behalf of Hartmut Grote for the LSC. Gravitational Waves: New Frontier Seoul, 16.01.2013 LIGO-G1300030. Ground based detector network. GWIC ‚The future of gravitation wave astronomy‘. GEO 600 optical design. GEO 600: detector and technology development. - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/1.jpg)
GEO 600 STATUS
Andreas Freise on Andreas Freise on behalf of Hartmut behalf of Hartmut Grote for the LSCGrote for the LSCGravitational Waves: New FrontierSeoul, 16.01.2013 LIGO-G1300030
![Page 2: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/2.jpg)
![Page 3: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/3.jpg)
![Page 4: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/4.jpg)
Ground based detector network
GWIC ‚The future of gravitation wave astronomy‘
![Page 5: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/5.jpg)
GEO 600 optical design
![Page 6: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/6.jpg)
GEO 600: detector and technology development
• Unique position as detector and technology test bed• GEO 600 more freely examines new techniques• GEO pioneers techniques– Signal-recycling– Monolithic fused silica suspensions– Electrostatic actuators– Squeezing & its control
• GEO techniques have propagatedto LIGO / VIRGO detectors
![Page 7: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/7.jpg)
GEO High Frequency (HF)Interesting high frequencyregion limited by shot noise
![Page 8: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/8.jpg)
GEO-HF upgrade program• Implementation started in 2009• Seismic, thermal, and unexplained noise below 500 Hz• Goal: Improve sensitivity at frequencies > 500 Hz
→ Improve shot noise limited sensitivity
Frequency [Hz]
Str
ain
[1
/sq
rt(H
z)]
![Page 9: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/9.jpg)
GEO-HF program components
• Output Mode Cleaner (OMC)• Switch to Homodyne / DC readout• Signal-recycling configuration change• Squeezed vacuum injection• Light power increase• `Astrowatch’ whenever possible!
![Page 10: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/10.jpg)
Output Mode Cleaner (OMC)
Dark fringe contrast defect, due to mode mismatch and misalignment of the interferometer arms
![Page 11: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/11.jpg)
• OMC housed in vacuum tank• Multi-stage seismic isolation
starting at 1 Hz
Output Mode Cleaner (OMC)
![Page 12: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/12.jpg)
• Design: (optimised for DC readout)– Quasi-monolithic– Four mirror cavity– Length actuation via PZT
After OMC (≈6 mW):
Before OMC (≈60 mW):
Output Mode Cleaner (OMC)
![Page 13: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/13.jpg)
Heterodyne vs DC readout
• Heterodyne readout– GW information in beat
of GW-SBs ↔ RF-SBs– (Almost) no carrier
at IFO output
• DC readout– GW information in beat
of GW-SBs ↔ carrier– Some carrier light at the
output required– GW signal directly in detected
power
![Page 14: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/14.jpg)
Signal Recycling
Signal Recycling: •makes GEO resonant to GWs•Allows shaping of the shot noise•Influences higher-order modes in the dark fringe
QuickTime™ and aPNG decompressor
are needed to see this picture.
![Page 15: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/15.jpg)
Signal Recycling Change (2010)
Increase transmission of Signal Recycling mirror to reduce shot noise at high frequencies
![Page 16: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/16.jpg)
Changed signal-recycling
• More HOMs withreduced signal-recycling mirror reflectivity
R=98 % R=90 %
![Page 17: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/17.jpg)
Changed signal-recycling
• Analyze interferometer output beam using the OMC
Mode order
Power [mW]
0 6.3
1 7.7
2 10.6
3 1.2
4 0.6
5 1.7
6 3.6
7 3.7
8 3.6
9 1.1
10 0.8
11 0.5
![Page 18: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/18.jpg)
Thank you
Signal Recycling: Sensitivity Comparison
![Page 19: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/19.jpg)
Squeezed vacuum injection
• Directly reduces shot noise• (Frequency independent) squeezing is similar to an
increase in laser power• 10 dB squeezing source developed at the AEI
![Page 20: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/20.jpg)
GEO 600 and Squeezed Light
Mode Matching
T=10%
![Page 21: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/21.jpg)
![Page 22: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/22.jpg)
20
Squeezing Phase Lock
Start with SQZ master to IFO main laser (fiber/PLL)
Error signal in refl. of IFO(Coherent Control signal)
different ports / possibilities
Feedback to SQZ master frequency (DC-2kHz)
Low-freq. Error signal (<0.01Hz) from noise dither
![Page 23: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/23.jpg)
23
Optical Loss
(Faraday Rotators: 0.65% loss)
Detection
Efficiency:
0.626
![Page 24: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/24.jpg)
24
Stable operation of squeezing!
![Page 25: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/25.jpg)
25
Spectrum (August 2012)
![Page 26: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/26.jpg)
26
Next steps on Squeezing (GEO)
Need Automatic Alignment (started) Need OPA temp set-point control Lower OMC loss (->2%), high-QE PD (->1%) New Phase noise budget after OMC change Lower det. Noise (main IFO) ->6dB look realistic, perhaps little more...
![Page 27: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/27.jpg)
Percent of time operated in Astrowatch (Science) mode: 70%Percent of Science mode, operated with squeezed light: 90%
![Page 28: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/28.jpg)
29
Burst DetectionHubble image of Betelgeuse (@200 pc)
![Page 29: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/29.jpg)
28
Detector Status from Nov 17, 2011 to Oct 15, 2012
![Page 30: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/30.jpg)
... endStar trails over GEO 600, by Jerome Degallaix
Thank you!
![Page 31: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/31.jpg)
... end
![Page 32: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/32.jpg)
• Squeezing–Try new control schemes–Try new optical / low loss components–Automatic alignment
• OMC–Try alternative alignment control techniques
• Thermal compensation–Test mass side heaters (recently installed)
–Compensation of beamsplitter lens with thermal radiation
• Astrowatch!
Next Steps
![Page 33: GEO 600 STATUS](https://reader033.vdocuments.net/reader033/viewer/2022051620/568139c0550346895da1670d/html5/thumbnails/33.jpg)
Timeline
• 07-2009: Tuned signal-recycling• 09-2009: DC-readout• 12-2009: OMC installation• 04-2010: Squeezing installation• 05-2011: Output optic suspension• 06-2011: Use 5W input power, Routine Squeezing• 06-2011 to 09-2011: Joint GEO - VIRGO run• 09-2011: Laser upgrade• 11-2012: IMC upgrade, baffle installation, side
heater • 2013: Input mode cleaner change, TCS, power
increase, mid freq. Noise hunting, baffle installation