gm32, online data sheet - sick.com€¦ · ... ex version, passive 242 (9.53) 223 ......

17
Online data sheet GM32 IN-SITU GAS ANALYZERS

Upload: nguyenduong

Post on 02-Jul-2018

223 views

Category:

Documents


0 download

TRANSCRIPT

On

lin

e d

ata

sh

ee

t

GM32

IN-SITU GAS ANALYZERS

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

Ordering information

Type Part no.

GM32 On request

Other models and accessories www.sick.com/GM32

15267

certified

14181certified

Detailed technical data

Classifications

ECl@ss 5.0 27150202

ECl@ss 5.1.4 27150202

ECl@ss 6.0 27150202

ECl@ss 6.2 27150202

ECl@ss 7.0 27150202

ECl@ss 8.0 27150202

ECl@ss 8.1 27150202

ECl@ss 9.0 27150202

ETIM 5.0 EC001190

ETIM 6.0 EC001190

UNSPSC 16.0901 41115403

Dimensional drawing (Dimensions in mm (inch))

Purge air attachment, Ex version, passive

242 (9.53)

223 (8.78)

14 (0.55)18 (0.71)

14

2 (

5.5

9)

Ø 40 (1.57)

45° 45°

Ø 22.4 (0.88)

Ø 200 (7.87)

Ø 215.8 (8.50)

Ø 254 (10.00)

Ø 14 (0.55)

28.4 (1.12)

Ø 200 (7.87)

Ø 239 (9.41)

Ø 18 (0.71)

45° 45°

2 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Purge air fixture

Ø 254 (10.00)

Ø 215.8 (8.50)

Ø 200 (7.87)

45° 45°

Ø 22.4 (0.88)

Ø 18 (0.71)

R 239.9 (9.44)

18 (0.71) 14 (0.55)

242 (9.53)

223 (8.78)

Ø 239 (9.41)

Ø 200 (7.87)

Ø 14 (0.55)

23

6.7

(9

.32

)

181.8 (7.16)

45° 45°

Ø 40

(1.57)

SLV4-2 purge air unit, 2BH1300

Ø 10

(0.39)

55

0 (

21

.65

)

47

0 (

18

.50

)4

0

(1.5

7)

550 (21.65)

470 (18.50)

128 (5.04)

257 (10.12)

Ø 4

0

(1.5

7)

40

(1.57)

375.7 (14.79)

45

°

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 3

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

SLV6-1 purge air unit, 2SH1300, ATEX 3G

550 (21.65)

470 (18.50)40 (1.57)

65

0 (

25

.59

)

57

0 (

22

.44

)

30

.5 (

1.2

0)

40

(1

.57

)

287.2 (11.31)

Ø 1

0 (0.3

9)

Ø 40 (1.57)

42.6 (1.68)200.3 (7.89)

Connection unit; Ex-version

15

8 (

6.2

2)

575 (22.64)

134 (5.28)53.5

(2.11)

181 (7.13)

545 (21.46)

515 (20.28)

40

0 (

15

.75

)

30

(1

.18

)

23

0 (

9.0

6)

30

(1

.18

)

34

0 (

13

.39

)

R = 4

.5 (.

18)

ø 9

(.3

5)

34

0

(13

.39

)

4 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Connection unit; standard version

15

8 (

6.2

2)

450 (17.72)

420 (16.54)

390 (15.35)

34

0 (

13

.55

)

40

0 (

15

.75

)

Ø 9

(.3

5)

30

(1

.18

)

30

(1

.18

)

24

(0

.94

)R =

4.5

(.18)

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 5

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

GM32 reflector unit

161 (6.34)

28

0 (

11

.02

)

146 (5.75)

135 (5.31)

291 (11.46)

Ø 200

(7.87)

45°

14

4 (

5.6

7)

35

.5

(1.4

0)

45°

R = 295(11.61)

Max. 120°

6 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Gas-testable measuring probe (GPP)

Ø 254 (10.00)

Ø 215.8 (8.50)

Ø 22.4

(0.8

8)

45°

Ø 18 (0.71)45°

Ø 200 (7.87)

B

A238 (9.37)

36

0 (

14

.17

)

23

3 (

9.1

7)

Ø 2

54

(10

.00

)

Ø 1

14

.3

(4.5

0)

L2

L1

L3

B

10 (0.39)

227 (8.94)

A

GPP measuring probes L3

227 477 727 977

Probe length,nominal

L1 L2 Weight

900 914 353 103 --- --- 27

1,500 1,624 1,063 813 563 313 35

2,000 2,108 1,547 1,297 1,047 797 42

2,500 2,608 2,047 1,797 1,547 1,297 50

All dimensions in mm, all weights in kgApplication-specific lengths available on request

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 7

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

Open measuring probe (GMP)

Ø 200 (7.87)

Ø 40

(1.57)

Ø 14 (0.55)

36

0 (

14

.17

)

23

3.2

(9

.18

)

176.9 (6.98)144 (5.67)

45° 45°

Ø 254 (10.00)

Ø 200 (7.87)

Ø 18 (0.71)

Ø 215.8 (8.50)

45° 45°

L1

L2 L3

Ø 1

14

.3

(4.5

0)

10 (0.39)

227 (8.94)

55

(2.1

7)

76.4 (3.01)

BA

BA

GMP measuring probes L3

250 500 750 1,000 1,250 1,500 1,750

Probelength, nominal

L1 L2 Weight

900 935 296 --- --- --- --- --- --- 21

1,500 1,644 1,005 755 505 255 --- --- --- 24

2,000 2,128 1,489 1,239 989 739 489 239 --- 30

2,500 2,628 1,988 1,738 1,488 1,238 988 738 488 35

All dimensions in mm, all weights in kgApplication-specific lengths available on request

8 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Sender/receiver unit; Ex-version

359 (14.13)315 (12.40)

200(7.87)

163.6

(6.44)

12

0 (

4.7

2)

17

9.7

(7

.07

)

45

9.5

(1

8.0

9)

33

9 (

13

.35

)

144 (5.67) 131 (5.16)

Max. 180°

R =

204

(8.0

3)

Max

.105°

R =

38

4 (

15

.12

)

R =

422 (1

6.61)

Max. 180°

409.5 (16.12)

87.6

(3.45)

≈ 4

50

(1

7.7

2)

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 9

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

Sender/receiver unit; standard version

359 (14.13)315 (12.40)

200(7.87)

163.6 (6.44)1

20

(4

.72

)

17

9.7

(7

.07

)

45

9.5

(1

8.0

9)

33

9 (

13

.35

)

144 (5.67) 131 (5.16)

Max. 180°

R =

204

(8.0

3)

Max

.105°

R =

38

4 (

15

.12

)

R =

422 (1

6.6

1)

Max. 180°

10 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Connection typesCross-duct version; Ex-version

7

8

9

ßà

á

6

21

4

5

3

â

ãä

①  Connection unit with Ex-p control unit for Zone 1 (2G) and Zone 2 (3G)②  Sender/receiver unit③  Reflector unit④  Purge air unit⑤  SCU-P100 control unit (option)

No. Cable Length Remarks Part no.

6 Power supply (versions 2G and 3G) From customer: 115 / 240 V AC, 50 / 60 Hz —

7 Ex-p controller — connection unit(only version 2G)

Included in connection unit —

8 Power supply (only version 3G) From customer: 115 / 240 V AC, 50 / 60 Hz —

9 Connection unit — sender/receiver unit 5 m10 m

Order separately 20576662057667

10 Differential pressure switch — connection unit From customer —

11 Differential pressure switch — connection unit From customer —

12 Input pressure/temperature From customer —

13 Ethernet PC/network (glass fiber cable) From customer —

14 Inputs/outputs From customer: terminal connections —

15 Connection unit — SCU-P100 (CAN cable) From customer —

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 11

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

Measuring probe version; Ex-version

6

7

8

9

ß

5

21

3

4 â

à

á

①  Connection unit with Ex-p control unit for Zone 1 (2G) and Zone 2 (3G)②  Sender/receiver unit with measuring probe (shown: GMP measuring probe)③  Purge air unit④  SCU-P100 control unit (option)

No. Cable Length Remarks Part no.

5 Power supply (versions 2G and 3G) From customer: 115 / 240 V AC, 50 / 60 Hz —

6 Ex-p controller — connection unit(only version 2G)

Included in connection unitt —

7 Power supply (only version 3G) From customer: 115 / 240 V AC, 50 / 60 Hz —

8 Connection unit — sender/receiver unit 5 m10 m

Order separately 20576662057667

9 Diffrential pressure switch — connection unit From customer —

10 Input pressure/temperature From customer —

11 Ethernet PC/network (glass fiber cable) From customer —

12 Inputs/outputs From customer: terminal connections —

13 Connection unit — SCU-P100 (CAN cable) From customer —

12 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Cross-duct version; standard version

ä

æ

7

ß

9

8

ã

â

á

à

21

45

6 3

å

①  Reflector unit②  Sender/receiver unit③  Purge air unit (for sender/receiver unit)④  Connection unit⑤  SCU-P100 control unit (option)⑥  Purge air unit (for reflector unit)

No. Cable Length Comments Part no.

7 Filter monitoring SLV4(reflector unit)

5 m Included in reflector unit purge air attachment 2032143

8 PT1000 temperature sensor Available as option or provided by customer —

9 Sender/receiver unit — reflector unit(CAN cable)

12 m24 m

Order separately 20208612027031

10 Filter monitoring SLV4(sender/receiver unit)

5 m Included in sender/receiv-er unit purge air attachment

2032143

11 Sender/receiver unit — purge air attachment(CAN cable)

0.8 m Included in sender/receiv-er unit purge air attachment

2023704

12 Sender/receiver unit — connection unit(CAN cable)

10 m20 m

Order separately 20287862045422

13 Power supply sender/receiver unit 10 m20 m

Order separately 20465482046549

14 Ethernet cable — PC/network From customer —

15 Inputs/outputs From customer: terminal connections —

16 Power supply connection unit From customer: 100 - 240 V AC; 50/60 Hz —

17 Connection unit — SCU-P100 From customer —

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 13

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

Measuring probe version; standard version

à

â

5

6

ß

9

8

7

1

23

4

á

①  Sender/receiver unit with measuring probe (shown: GMP measuring probe)②  Connection unit③  SCU-P100 control unit (option)④  Purge air unit (only for GMP measuring probe)

No. Cable Length Remarks Part no.

5 Filter monitoring SLV4 5 m Included in measuringprobe purge air attachment

2032143

6 Power supply for GPP measuring probe 10 m Included in measuringprobe purge air attachment;

From customer: 115 / 230 V AC, 50 / 60 Hz

2017519

7 Sender/receiver unit — purge air attachment(CAN cable)

0,8 m Included in measuringprobe purge air attachment

2023704

8 Sender/receiver unit — connection unit(CAN cable)

10 m20 m

Order separately 20287862045422

9 Power supply sender/receiver unit 10 m20 m

Order separately 20465482046549

10 Ethernet cable — PC/network From customer —

11 Inputs/outputs From customer: terminal connections —

12 Power supply connection unit From customer: 100 ... 240 V AC, 50 / 60 Hz —

13 Connection unit — SCU-P100 (CAN cable) From customer —

14 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

GM32IN-SITU GAS ANALYZERS

Instruction for installationCross-duct version

47

0

Min

. 7

00

mm

470

L

5

6

8

4

32

1

7

ø1

33

min. 30

Mounting example on a vertical duct or stack①  Recommended free space: approx. 1.5 m (min. 1 m)②  Weather hood (option)③  Sender/receiver unit④  Distance flange - flange⑤  Active measuring path⑥  Mounting flange, L=240 mm/500 mm/1,000 mm⑦  Reflector unit⑧  Attachment for purge air unit (at the customer): e.g. 4 steel pipes, 50 x 5 mmProject planning is extremely important for trouble-free mounting and commissioning of the measuring device. The most importantrequirements and preparations at the site of installation for determining the measuring site for the GM32-Cross Duct Version aredescribed as follows:Unimpeded inlet and outlet zone taken into considerationAlways ensure “distance flange – flange” and “active measuring path” values are maintained!Plan suitable free space for mounting and maintenance activities, for example, for tilting the sender/receiver unitArrange for sufficient weatherproofing for the device components, if mounting outdoorsThe site of installation must be sufficient for the weight of the sender/receiver unit and reflectorArrange clearance of minimum 0.5 m in direction of flow upstream from the device for the calibration supports, in order that themeasuring device and calibration probe cannot interfere with each other

2018-06-22 09:01:13 | Online data sheet

Subject to change without notice

GAS ANALYZERS | SICK 15

ABCDEF

HIJKLMNOPQRST

GM32IN-SITU GAS ANALYZERS

Measuring probe version

30

L

50

Min

. 7

00

mm

47

0

Ø1

33

5

432

1

Mounting example on a vertical duct or stack①  Recommended free space: approx. 1.5 m (min. 1 m)②  Weather hood (option)③  Sender/receiver unit④  Mounting flange, L=240 mm/500 mm/1,000 mm, tilt tube: approx. 1° towards the duct, so that condensate can drain off.⑤  Attachment for purge air unit (at the customer): e.g. 4 steel pipes, 50 x 5 mmProject planning is extremely important for trouble-free mounting and commissioning of the measuring device. The most importantrequirements and preparations at the site of installation for determining the measuring site for the GM32 measuring probe versionare described as follows:Unimpeded inlet and outlet zone taken into considerationPlan suitable free space for mounting and maintenance activities, for example, for tilting the sender/receiver unit and extendingthe probeArrange for sufficient weatherproofing for the device components, if mounting outdoorsThe site of installation must be sufficient for the weight of the sender/receiver unit including the measuring probeArrange clearance of minimum 0.5 m in direction of flow upstream from the device for the calibration supports, in order that themeasuring device and calibration probe cannot interfere with each other

16 GAS ANALYZERS | SICK Online data sheet | 2018-06-22 09:01:13

Subject to change without notice

Onlin

e da

ta s

heet

SICK AG | Waldkirch | Germany | www.sick.com

SICK At A GlAnCeSICK is one of the leading manufacturers of intelligent sensors and sensor solutions for industrial applica-tions. A unique range of products and services creates the perfect basis for controlling processes securely and efficiently, protecting individuals from accidents and preventing damage to the environment.

We have extensive experience in a wide range of industries and understand their processes and require-ments. With intelligent sensors, we can deliver exactly what our customers need. In application centers in Europe, Asia and North America, system solutions are tested and optimized in accordance with customer specifications. All this makes us a reliable supplier and development partner.

Comprehensive services complete our offering: SICK lifetime Services provide support throughout the ma-chine life cycle and ensure safety and productivity.

For us, that is “Sensor Intelligence.”

WOrldWIde preSenCe:Contacts and other locations – www.sick.com