1Produktvorstellung – SECUTEST SIII+ Dieter Feulner
Electrical safety Test for appliances
Farhaan Ahmed
2Produktvorstellung – SECUTEST SIII+ Dieter Feulner
Agenda
Secutest SIII General setup
Measurement
Software
3Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+general
SECUTEST SIII
4Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+general
Universal test instrument for testing the electrical safety of portable electrical equipment in commercial and medical applications.
In order to evaluate electrical safety, tests are performed for protective conductor connections, insulation resistance and leakage currents.
5Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+general
Automatic adaptation to either 110 V / 60 Hz or 230 V / 50 Hz mains power
Automatic recognition of safety class (I, II) and mains connection errors
Menu-driven test sequences: fully automated or manual
Safety for the user thanks to integrated personal protection - suitable for persons with basic electro-technical instruction
6Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+general
Convenient memory and report generating functions (optional)
Data interface for PC, printer and barcodes
Report generation with printer module (PSI module), DA-II printer adapter or PC software
User interface in various languages: English, French, German, Italian, Spanish, Czech, Dutch
Parallel test sockets for devices under test without mains plug
7Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+general
Patient leakage current: AC and DC components measured separately
Equivalent leakage current measurement with 230 V test voltage
Function test with power analysis
Temperature measurement
Current measurement (with optional clip-on ammeter)
Voltage and resistance measurement
8Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+general
Dimensions: 292 x 130 x 243 mm, weight: 4.0 e.q. 4.5 kg
Including automatic detection and indication of whether or not the probe is connected to PE
Including direct read-out of measurement data via the RS 232 interface (direct printing)
Including calibration certificate in accordance with DKD
9Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+Tester schematic
Push-buttons (left and right) for releasingthe handle from its snap-in position
Test socket forconnecting the device under test
Push-buttons(left and right)for releasing the lid
Lid
Compartment for probe and accessories
Carrying handle and tilt stand
OptionalSECUTEST PSIalphanumeric keypad,printer & memory
OptionApplied parts for measuringMedical devices
Optional service socket for connecting a PC or printer
Optional additional unit for high voltage test
10Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ front plate
Jacks for Live / Neutral /PE conductor
Jacks for connectingthe probe
Function selectorswitch
Scroll key formenu and parameterselection
LCD-displaygraphical withbacklight
Signal lamp for mainsconnection error
Key for startingtest sequence
Help key
Key switchingtest voltage to the test soecket
Signal lamp forthe function test
Jacks forapplied parts
11Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ rotary switch
12Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Principle work with cursor
13Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Setup
General Setup
14Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Setup
Save the settings
Change the limits
Make settings for all test sequences (standard’s)
Look at the stored test result
Information about the software of the tester
15Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Limits
Set backgound ligth time
Duration for each test in a sequence
Nominal Voltage for leakage current calculation
Test of short circuit in mains
Wait after power on to start measurement
Do not ask any user input after starting the test
16Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Test Sequence
Test stops if measurement is out of limit range
For PSI only
Intrinsic error of tester is included in test result
Warning if IT system is present
Internal beeper
Detectiong of PE connection
Measured value is printed via RS232 immediately
17Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Protocols
Save the test result in a protocol
Print the test result via RS232
Show all stored protocols(and select one)
Print all stored protocols via RS232
Delete all stored protocols
18Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Service
Information about internal Software of the tester
Release of Software, Customer
Set time and date
19Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUX
Measurements
20Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ R-PE
Test current can be set to 200 mA and (10) 25 A
21Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ ISO / HV passive leakage
Measuring the insulation resistance
Measuring the protective leakage current with an alternative method
For Medical equipment only
High voltage – Insulation test (with DC voltage)
22Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ active Leakage current
Protective earth current
Touch current (current from accessible parts to ground – measured with the probe)
For medical equipment only
Alternative measurement for protective earth current
For medical devices only (overall leakage current)
23Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Function Test
Power supply of device is measured
24Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Multimeter
Using the Tester like a Multimeter
Voltage measurement with the probe (against PE)
Voltage measurement withtest leads
(Low) Resistance measurements
25Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUX
Measurements with auxiliary
Current measurement with a clamp
Protective earth with a clamp
Temperature measurement with PT100 or PT 1000
26Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO
Auto mode
Predefined test sequences
27Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO
Test Sequence IEC 60335
For routine test
28Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 60335 START
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
29Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 60335 SETUP
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
30Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 60335 CONFIGURE
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
31Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 60335 PE
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
32Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 60335 HV
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
33Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 60335 Leakage Current
Measurement (type of leakage current)
Measurement Value
Test Voltage
Measurment conditions
34Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK Start screen
The test starts automatically if device is switched on
Push return button to change the Type
35Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK 60335
The test starts automatically if device is switched on
Push return button to change the Type
36Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK 60335 SETUP
The test starts automatically if device is switched on
Push return button to change the Type
37Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK 60335 Configure
The test starts automatically if device is switched on
Push return button to change the Type
38Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK ground continuity
The test starts automatically if device is switched on
Push return button to change the Type
39Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK HV
The test starts automatically if device is switched on
Push return button to change the Type
40Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO WIK LEAKAGE Current
The test starts automatically if device is switched on
Push return button to change the Type
41Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Service and Repair
Test Sequence DIN VDE 0701
For Maintenance and Repair
42Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
43Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701 visual
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
44Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701 PE
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
45Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701 RISO
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
46Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701 Leakage
Leakage current active or passive, depending on type of device
Devices with elctronic switches shall be measured active in both directions of mains supply (Differential + Touch current)
47Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701 result
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
48Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ AUTO 0701 Function Test
Select :
standard
Connection of the device
Type of device
ID-Numer of device
Change settings for this test sequence
49Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Testing of 3-phase devices
For three-phase devices use the AT3-III:
• fully automatic test sequence
• high voltage test (optional)
• residual current measurement
50Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+
Software
51Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ SECUTEST (P)SI
Enter customer information.
Enter repair informationenter further informaton of the appliance.
PRINT STOREor
Activates the „PSI-menue“
52Produktvorstellung – SECUTEST SIII+ Dieter Feulner
internal Test Report using free software Z853
53Produktvorstellung – SECUTEST SIII+ Dieter Feulner
internal Test Report using free software Z853
54Produktvorstellung – SECUTEST SIII+ Dieter Feulner
Print direct
55Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Software and documentation
Test ProtocolTesting of the Electrical Safety according to IEC601.1
Order number: Customer:aaaaAenggleisfff95081 erlangen
Contractor:GMC-Instruments
Device: defibrilatorccc Manufacturer:draegerType: oem Protection class: Inventory no.: aaa Isolated applied parts: Type Test according to IEC 601.1 Date of test: 01.02.01Visual inspection:[ ] Protective conductor (only for class I equipment)[ ] Casing and mechanical parts[ ] Insulating parts[ ] Terminal connectors including plugs[ ] Markings[ ] OtherMeasurement Meas. value Limit value OK
Protective conductor continuity +0.063Ohm <0.200OhmInsulation resistance >+310.0MOh >2.000MOhmDifferential currentEarth leakage current +0.016mA <0.500mAEarth leakage current single fault condition +0.037mA <1.000mAHousing leakage current +000.5µA <100.0µAHousing leakage current single fault condition +022.8µA <0.500mAPatient leakage current AC +000.4µA <0.100mAPatient leakage current AC single fault condition +001.2µA <0.500mAPatient current DC +000.2µA <10.00µAPatient current DC single fault condition +000.1µA <050.0µAPatient current with mains on applied parts +000.4µA <05.00mAAuxiliary patient current ACAuxiliary patient current AC single fault conditionAuxiliary patient current DCAuxiliary patient current DC single fault conditionDielectric strength test [ ]Functional test[ ] Functional and safety testDate of next test : Test equipment used Type:
SECUTEST 0751/601Manufacturer:GOSSEN-METRAWATT
Signature:Testing person:
_____________________________________________________Place: Date 10.03.02
Data-management andReport generation
with MS Office
PC.doc-word
56Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ Software and documentation
Maintenance and servicemanagement program
PS3
57Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+ summary
•Easy to carry (lightwight, compact)
•fully automatically test sequence, even for 3-phase devices
•excellent price-performance ratio by customer-defined configuration
•Easy to handle, online-help•Leakage current AC and DC•High voltage test up to 6kV (DC)•0,2; 10 or 25 Ampere test current•Function test with power measurement•Internal memory up to 125 protocols
58Produktvorstellung – SECUTEST SIII+ Dieter Feulner
SECUTEST SIII|+