hamilton arc system - insatech - forsiden · arc view handheld arc wi sensor adapter – 6 –...

8
Sensing Solutions that Empower and Propel HAMILTON ARC SYSTEM

Upload: nguyenthu

Post on 25-Sep-2018

220 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

Sensing Solutions that Empower and Propel

HAMILTON ARC SYSTEM

Page 2: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

The true power of the HAMILTON ARC familyHAMILTON has delivered quality products to laboratories for more than 50 years, including products such as syringes, laboratory devices, automation platforms and sensors.

In the world of process analytics, HAMILTON is known for sensors (pH, oxygen, redox, and conductivity), and for its first-class choice of process armatures, completed with their selection of process connectors.

HAMILTON ARC revolutionizes the integration of sensors by rethinking communication between sensors, end users and process control systems. The functionality of a traditional transmitter has been replaced by a microprocessor within the sensor. ARC sensors bypass the transmitter and communicate directly to the control system. Sensors are available in all standard pH, dissolved oxygen and conductivity formats.

ARC Vital: Eliminate the cost of the transmitter and get more reliable readings

For users in established facilities, ARC sensors can communicate without the use of a transmitter with 4-20 mA devices or digitally with a computer. In this area the advantage of the ARC system is in lab calibration with ARC View Handheld, retained records, and the proven quality and accuracy of HAMILTON sensors. The 4-20 mA signal is more stable and reliable than the low currents or voltages in classical oxygen, conductivity or pH measurement.

ARC Versa: In lab pre-calibrations, sensor management with ARC View Handheld and wireless communication option

Combine cost savings and reliability of ARC Vital with the convenience of ARC View Handheld and ARC Wi. The ARC View Handheld device facilitates in-lab pre-calibration, configuration of the HAMILTON ARC sensors, and sensor management. Wired or wireless communication will make handling and maintenance of the sensors and process control more reliable, simple and efficient.

ARC Vision: Digital signal transmission and wireless monitoring of multiple sensorsFor new facilities and installations the ARC sensors are truly revolutionary. In digital mode several ARC sensors can run and be maintained in parallel. Wired or wireless communication will make installation, handling and maintenance of the sensors and process control more reliable, simple and efficient. ARC View Handheld enables wireless, handheld, local monitoring of all sensors in process as well as in-lab calibration of sensors prior to installation into the process.

The bridge between process parameter and process controlThree ARC concepts meet customer requirements from established research and pilot labs to new production facilities and installations:

HAMILTON’s ARC Vital, ARC Versa, and ARC Vision systems offer an optimal solution for every use case.

– 2 –

HAMILTON ARC System

Page 3: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

– 3 –

Sensing Solutions that Empower and Propel

Advanced solution for new installations that are running a digital network. The ARC sensors are connected through Modbus to the PLC, and the ARC View Handheld provides onsite monitoring and calibration

· Provides the premium digital solution for new levels of performance.

· Eliminates downtime, increases process safety and accuracy.

Products: ARC sensors (digital mode), ARC View Handheld, and ARC Wi Sensor Adapter.

CUTTING-EDGE TECHNOLOGY

SIMPLICITY AND QUALITY MULTIPLE CONNECTIVITY

ARC VitalPremium quality sensors solution for applications where the PLC is in close proximity to the sensor and no local display is required.

· Saves cost of ownership.

· Optimizes your analog system through delivering directly from the sensor a robust signal interface.

Products: ARC sensors (4-20mA mode) and ARC View Handheld for in lab calibration.

Which ARCare you?

ARC VersaOnline sensor management solution for applications where the sensor is positioned far from the PLC and local reading would be enabled through a wireless connection to the ARC View Handheld.

· Empowers you to get the most out of the combination of analog and digital technologies at your disposal.

· Gives onsite personnel the ability to stay mobile as they monitor data directly from the sensors.

Products: ARC sensors (4-20mA mode), ARC View Handheld, and ARC Wi Sensor Adapter.

4 -

20 m

A

PLC

Mod

bus

PLC

4 -

20 m

A

PLC

ARC Vision

Page 4: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

• PHERMLYTEreferenceelectrolytefactoryprepressurizedforaclog-freediaphragm and minimized diaphragm potentials

• HighperformanceHPCOATRAMICdiaphragm

• EVEREF-FReferencecartridgeforsilver-freeelectrolytes

• Poisonresistant“PHI“pHglass

• Monitoringofallsensorfunctions,statusofthesensorquality(glassresistance,reference resistance)

• Excellentlinearity.Independantofcableconditionsandlength

• Suitableforsteamsterilization,autoclavingandCIP

• AllpartsincontactwithmediaareFDA-compliant

• Easytocleanduetotheforward-facing,flush arrangementofelectrodes,EHEDGcertifiedwithHygienicSocket

• Monitoringofallsensorfunctions,statusofthesensorquality(cellconstant)

• Transmitterless-directdigitaland/oranalogsignalbyprovenVP8.0connectorhead.

• PG13.5processthreadconnection.

• On-lineandinlabpre-calibration,autorecognitionofthecalibrationstandard(internallystoreddataforcalibrationstandards).

• ConfigurableusingtheRS485interfacewithPC,ARCViewHandheldorbyusingtheModbusRTUprotocolfromtheprocesscontrolsystem.

• Savesallprocessrelevantinformation:sensoridentityandcalibrationdata,operatinghours,cleaningandsterilizationcycles,sensorquality,errorsandwarnings.

– 4 –

HAMILTON ARC Sensors

Optical Oxygen Sensor: VISIFERM DO ARC

• Precisemeasurementoptics,stableupto130°C

• Doesnotrequireelectrolyteandpolarization

• Temperature-resistantelectronicsbuiltintoa12mmshaft

• Easily replaceable sensor cap containing the sensing element

• Monitoring of all sensor functions and status of the replaceable sensor cap

• Suitableforsteamsterilization,autoclavingandCIP

Technical Data

Range of measurement: 0.05 - 300% air saturationDetectionlimit: 0.01%-volResponsetimet98%: <30sat25°C,fromairtonitrogenOxygen consumption: NoneRequired flow: NoneOperationtemperature: -10-130°C;thesensorprovides

noDOreadingabove80°CPressurerange: 0-12barMeasurement principle: Oxygen dependant luminescenceMediumcontactedmaterials: Stainlesssteel316L/DIN1.4435 FDA-approvedSiliconeandEPDMSurface quality of steel: Ra = 0.4 μm (N5)Interface: Configurable4-20mAinterface(DOor

temperature),digitalRS485ModbusRTUinterface

Conductivity Sensor: Conducell 4USF ARCRangeofmeasurement: 1-500000μS/cmExpectedcellconstant: c≈0.36cm-1

Operationtemperature: 0-130°CPressure range: 0 - 20 barTemperature sensor: YesMeasurement principle: Four pole contactingMediumcontactedmaterials: Stainlesssteel316L/DIN1.4435 FDA-approvedPEEK FDA-approvedEPDMSurface quality of steel: Ra = 0.4 μm (N5)Interface: 2x 4-20 mA interface (conductivity and

temperature),digitalRS485ModbusRTUInterface

pH Sensor: Easyferm Plus ARC

Rangeofmeasurement: pH0-14Zero point: 0 ± 20 mVOperationtemperature: 0-130°CPressurerange: 0-6barTemperature sensor: YesMeasurement principle: Potential measured against referenceShaftmaterial: GlassMembraneglass: HAMILTONtype“PHI”glassO-ringmaterial: FDA-approvedEPDMElectrolyte: Pressurized PHERMLYTEDiaphragmmaterial: HPCOATRAMICInterface: 2x 4-20 mA interface (pH and

temperature) , digital RS485 ModbusRTUInterface

Unique Features of all ARC Sensors

Page 5: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

• Suitedforapplicationsinbio-technology,pharmaceutical industry, food industry and beverage industry

• CleanabilitytestedandcertifiedaccordingtoEHEDGcriteria

• BiocompatibiltytestedandcertifiedbyMDT

• Odorlessandtastelessgelelectrolyte(testedbyTÜVRheinland)

• Small sensitivity todisturbancesandeasily calibratedbecauseofHAMILTON“HB”pHglass

• Monitoringofallsensorfunctions,statusofthesensorquality(glassresistance,reference resistance)

pH Sensor: Easyferm Food ARC

pH Sensor: Polilyte Plus ARC

• POLISOLVE PLUS reference electrolyte allows highly reproducible and stablemeasurements over long periods of time

• Minimaldiffusionpotential

• 2SINGLEPOREs fordirect contactbetweenPOLISOLVEPLUSelectrolyteandmedium

• HAMILTON“H“ pH glass, stable for long periods plus a patented referencesystem

• Upside-downmountingpossible

• Monitoringofallsensorfunctions,statusofthesensorquality(glassresistance,reference resistance)

pH Sensor: pHeasy ARC

• ARCempowersfullyintegratedaccuracymonitoring

• Extends service intervals and increases reliability of pH measurementapplications

• Upside-downmountingpossible

• 2 SINGLE POREs for clog-free contact between measurement medium andPOLISOLVE electrolyte

• Monitoringofallsensorfunctions,statusofthesensorquality(glassresistance,reference resistance, Checkref potential)

Rangeofmeasurement: pH0-14Zero point: 0 ± 20 mVOperationtemperature: 0-130°CPressurerange: 0-6barTemperature sensor: YesMeasurement principle: Potential measured against referenceShaftmaterial: GlassMembraneglass: HAMILTONtype“PHI”glassO-ring material: FPMElectrolyte: POLISOLVEDiaphragmmaterial: SINGLEPOREInterface: 2x 4-20 mA interface (pH and

temperature),digitalRS485ModbusRTUInterface

– 5 –

HAMILTON ARC Sensors

Rangeofmeasurement: pH0-14Zero point: 0 ± 20 mVOperationtemperature: 0-130°CPressurerange: 0-6barTemperature sensor: YesMeasurement principle: Potential measured against referenceShaftmaterial: GlassMembraneglass: HAMILTONtype“HB”glassO-ringmaterial: FDA-approvedEPDMElectrolyte: PressurizedFOODLYTEDiaphragmmaterial: HPCOATRAMICInterface: 2x 4-20 mA interface (pH and

temperature),digitalRS485ModbusRTUInterface

Rangeofmeasurement: pH0-14Zero point: 0 ± 20 mVOperationtemperature: 0-130°CPressurerange: 0-6barTemperature sensor: YesMeasurement principle: Potential measured against referenceShaftmaterial: GlassMembraneglass: HAMILTONtype“H”glassO-ring material: FPMElectrolyte: POLISOLVEPLUSDiaphragmmaterial: SINGLEPOREInterface: 2x 4-20 mA interface (pH and

temperature),digitalRS485ModbusRTUInterface

Technical Data

Page 6: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

ARC View Handheld

ARC Wi Sensor Adapter

–6–

HAMILTONARCDevices

Permissibleoperatingtemp.: 5-45°CMax.transmittingpower: +0dBmMax.datatransmissionrate: 256kbpsCommunicationprotocol: ModbusRTUFrequencyrange: 2.45GHz(IEEE802.15.4)Max.rangeinopenarea: ~100mMax. range in process environment: ~20 mDimensions(WxDxH): 140x35x85mmWeight: 340 gPower supply: •OnArcViewDockingwithplug-inpowersupplyunit: 100-240V~/50-60Hz•Batterycontinuousoperation: 9hProtectionRating: IP 67 (exceptARCView Dock, power

adapter and sensor cable)USBinterface: OnARCViewDockPowersupplyunitwithplugadaptersforEU,GB,USA,AUS

Technical Data

This device empowers the operator to monitor measurement values, calibrate ARCsensorsandconfigurevariousparameters.ThestandardversionofARCViewHandheld can communicate with up to 30 ARC sensors over wireless channel paralleltotheprocesscontrolsystem’smaininterface.WithARCViewDocktheARC View Handheld is connected over cable to a single ARC sensor and can be usedasalaboratorydesktopcalibration,configurationandmonitoringtool.

• On-line wireless sensor monitoring (up to 30 HAMILTON ARC sensors)

• Displaysensordata

• Sensor calibration

• Sensorconfiguration

• Visualization of the sensor status

• Datamanagement

This wireless adapter is mounted between the VP head of an ARC sensor and the VP sensor cable. The ARC Wi provides the wireless communication between the ARC sensor and ARC View Handheld unit.

• ARCWiisplug&playforeachandeveryARCsensortype

• Configuration-freeinstallation

• AnalogsensorsignalsarebypassedthroughtheARCWiSensorAdapter

• Routingfunctionsforthedigitalsignalsbetweenwirelessandwiredconnections(PLC)

• Visualindicationofanactivewirelessconnectionandsensorfailure

Permissibleoperatingtemp. 5-80°CMax.transmittingpower: +0dBmMax.datatransmissionrate: 256kbpsCommunicationprotocol: ModbusRTUFrequencyrange: 2.45GHz(IEEE802.15.4)Max.rangeinopenarea: ~100mMax. range in process environment: ~20 mDimensions(WxDxH): 75x30x70mmWeight: 110gPower supply: With VP sensor cable

ARCViewHandheldDock

ARC View Handheld

Technical Data

Page 7: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

Product name Order numberEasyfermPlusARC120 242091

Easyferm Plus ARC 225 242 092

Easyferm Plus ARC 325 242 093

Easyferm Plus ARC 425 242 094

PolilytePlusARC120 242111

Polilyte Plus ARC 225 242112

Polilyte Plus ARC 325 242113

Polilyte Plus ARC 425 242114

EasyfermFoodARC120 242120

Easyferm Food ARC 225 242121

Easyferm Food ARC 325 242122

Easyferm Food ARC 425 242123

pHeasyARC120 242154

pHeasy ARC 225 242155

pHeasy ARC 325 242156

pHeasy ARC 425 242157

Conducell4USFARC120 242159

Conducell4USFARC225 242160

Conducell4USFARC325 242161

Conducell4USFARC425 242162

VisifermDOARC120 242163

VisifermDOARC225 242164

VisifermDOARC325 242165

VisifermDOARC425 242166

ARC View Handheld Package 242180

ARC Wi Sensor Adapter 242170

–7–

Ordering Information

includes:

ARC View Handheld•

ARCViewHandheldDock•

ARCCableVP8•

ARC Memory Stick•

Power Adapter•

ARC View Handheld Package

Page 8: HAMILTON ARC SYSTEM - Insatech - forsiden · ARC View Handheld ARC Wi Sensor Adapter – 6 – HAMILTON ARC Devices Permissible operating temp.: 5 - 45°C Max. transmitting power:

P/N:691069/0103-2010

Your HAMILTON representative:

http://www.hamiltoncompany.com

Europe, Africa, Asia: [email protected] America, South America, Pacific Rim:

[email protected]

HAMILTON Bonaduz AGViaCrusch8CH-7402Bonaduz,SwitzerlandToll-Free:00800-660-660-60Telephone:+41-81-660-60-60Fax:[email protected]

HAMILTON Company4970EnergyWayReno,Nevada89502USAToll-Free:800-648-5950Telephone:+1-775-858-3000Fax:[email protected]

HAMILTON Quality Products:Laboratory ElectrodesDURACAL™pHBuffersConductivity SensorsConductivity StandardsDissolvedOxygenSensorsArmatures for SensorsMICROLITER™-SyringesGASTIGHT®-SyringesInstrument SyringesSoftGrip™-PipettesMiniature ValvesBioLevitatorDeCapper

Laboratory Automation for:DrugDiscoveryGenomicsProteomicsForensicsInVitroDiagnostics