Hemodynamic Monitoring and
Volume Management
Lakhmir S. Chawla, MD
Division of Renal Diseases and Hypertension
![Page 2: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/2.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Overview
• Blood Pressure Monitoring
• Assessment of Intravascular Volume
• Volume Responsiveness
– Where are you on the Starling Curve?
• Cardiac Output Monitoring
![Page 3: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/3.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Measurement of Blood Pressure
• Stethescope plus BP Cuff
• Dinamap
• Arterial line
![Page 4: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/4.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Old School
•Advantages
– Accurate systolic and
diastolic measurement
– Individual validation not
needed
•Disadvantages
– Cumbersome
– Operator variability
![Page 5: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/5.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Oscillometric Blood
Pressure Measurement
•Advantages
– Easy to use
– Faster than manual
– Measures MAP, not SBP or DBP
•Disadvantages
– Does not measure systolic or
diastolic directly; uses ratio
formula to estimate SBP and
DBP
– Assumes NSR
– Motion Artifact
![Page 6: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/6.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Overestimate Underestimate
(Too much pressure to occlude) (Too little pressure to occlude)
![Page 7: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/7.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 8: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/8.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Skipped Values in the
DINAMAP algorithm
• Skipped SBP
– 89, 119, 120, 124, 125, 130, 140, 141, 150,
160, 170, 180, 190, and 200 mm Hg
• Skipped HR
– 95, 99, 103, 106, and 109 bpm.
Skip Patterns in DINAMAP-Measured Blood Pressure in 3 Epidemiological Studies
Kathryn M. Rose. (Hypertension. 2000;35:1032.)
![Page 9: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/9.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Invasive Arterial BP Monitoring
![Page 10: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/10.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 11: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/11.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Central Peripheral
Effects of Systolic and Diastolic
BP
• SBP rises (approx 20-30 mm Hg)
• DBP falls
• Pulse Pressure Widens
![Page 12: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/12.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 13: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/13.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 14: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/14.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Quality Assurance with Arterial
Lines
• Square Root Test
• Waveform assessment
![Page 15: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/15.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Overdamped
decreased SPB, increased DPB
Check for clot, large air bubbles, soft tubing
Underdamped
Increased SPB, ringing artifacts in waveform
Remove bubbles, often small, replace tubing with larger bore,
add damping device
Hemodynamic monitoring : invasive and noninvasive clinical application 3rd ed.
Gloria Oblouk Darovic. Philadelphia : W.B. Saunders Co., 2002.
![Page 16: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/16.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Summary of Arterial Pressure
Assessment
• If you are not measuring the BP yourself, think prescribe in MAP terms, not SBP or DBP
• When using an arterial line, know its location, evaluate its wave form
• As we become more and more technology dependent
Trust, BUT Verify
![Page 17: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/17.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Assessment of Intravascular
Volume
• Bedside- Physical Exam
– JVP
– Orthostatics
• Central Venous Pressure
• Pulmonary Artery Wedge Pressure
• Stroke Volume
• Stroke Volume Variation
![Page 18: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/18.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Physical Exam
• Jugular Venous Pressure
– Observer Variability
– Difficult in obese and edematous patients
– Trending problematic
• Orthostatics
– Can be helpful
– Difficult to do in critically ill patients
– Specific but not sensitive
– Result Binary; Following trend difficult
![Page 19: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/19.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Physical Exam
• Does not reliably predict intravascular
volume status
• Classic Triad - Jugular venous
distension, crackles on auscultation,
and peripheral edema
•Crit Care Med 2001; 29: 1502-1512
•N Engl J Med 1983; 308:263–267
•Crit Care Med 1984; 12:107–112
•Crit Care Med 2001; 29:1635–1636
•Intensive Care Med 2001; 27:1238
![Page 20: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/20.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Portable CXR
• The classic radiographic signs of volume overload (pleural effusions, septal lines, peribronchial cuffing, or distribution of lung densities) are poor predictors of LVEDP
•Crit Care Med 2001; 29: 1502-1512
•Crit Care Med 2001; 29:1635–1636
![Page 21: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/21.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Central Venous Pressure (CVP)
• Advantages
– Can be trended
– Objective Measure
• Disadvantages
– Not helpful in the absence of cardiac output measure
– Significantly affected by respiration and PEEP
Michard F et al Chest 2002; 121:2000–2008
Kumar A et al Crit Care Med 2004; 32:691–699
Shippy CR et al Crit Care Med 1984; 12:107–112
![Page 22: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/22.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Central Venous Pressure
• Does NOT predict cardiac
preload or volume status
•Michard F et al Chest 2002; 121:2000–2008
•Kumar A et al Crit Care Med 2004; 32:691–699
•Shippy CR et al Crit Care Med 1984; 12:107–112
![Page 23: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/23.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Pulmonary Capillary Wedge
(Occlusion ) Pressure
• Value generated by right sided heart catheterization (typically balloon guided)
• Under idealized conditions, PCWP (PAOP) represents left ventricular end diastolic pressure (LVEDP)
• LVEDP is a surrogate for LV end diastolic volume which should represent pre-load
![Page 24: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/24.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 25: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/25.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Pulmonary Capillary Wedge Pressure
• Advantages
– Measures left sided pressures as
compared CVP
• Disadvantages
– Somewhat invasive & more side effect than
CVP
– Accuracy dependent on optimal placement
– Wedge pressure must be read by
experienced operator
![Page 26: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/26.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Pulmonary Capillary Wedge Pressure
Contrary to expectations
• PCWP fails to correlate with LVEDV or
stroke volume
• The wedge is NOT a good measure of intravascular volume status.
![Page 27: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/27.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
The Evidence
•Kumar A et al Crit Care Med 2004 Vol. 32, No. 3
• Sakka SG, Intensive Care Med 1999; 25:
•843–846
• Jardin F, Intensive Care Med 1994; 20:550–554
•Calvin JE, Crit Care Med 1981; 9:437–443
•Calvin JE,. Surgery 1981; 90:61–76
•Diebel L, J Trauma 1994; 37:
•950–955
•Martyn JA,. J Trauma 1981; 21:
•619–626
• Lichtwarck-Aschoff M,. Intensive
•Care Med 1992; 18:142–147
•Tousignant CP, Anesth Analg 2000; 90:351–355
•Thys DM,. Anesthesiology 1987; 67:630–634
•Ellis RJ, J Thorac Cardiovasc Surg 1979; 78:605–613
• Hoeft A, Anesthesiology 1994; 81:76–86
•Douglas PS, Ann Thorac Surg 1987; 44:31–34
•Hansen RM, Anesthesiology 1986; 64:764–770
•Tuman KJ, J Cardiothorac Vasc Anesth 1995; 9:2–8
•Buhre W, Eur J Anaesthesiol 1999; 16:11–17
•Dennis JW, J Vasc Surg 1992; 16:372–377
![Page 28: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/28.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
A, Initial CVP
and SVI
B, Δ in CVP and
SVI in response to
saline
C, Initial PWP and SVI
D, Δ in PWP and SVI
in response to saline.
Pulmonary artery occlusion
pressure and central venous
pressure fail to predict ventricular
filling volume, cardiac
performance, or the response to
volume infusion in normal
subjects
Kumar, Anand MD; Critical Care
Medicine: Volume 32(3) March
2004 pp 691-699
![Page 29: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/29.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
•2-D echo LVEDVI compared to SVI
![Page 30: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/30.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Volume Responsiveness
Increase in cardiac output with
volume bolus
CO = HR x SV
•Stroke Volume corrected for BSA = SVI
![Page 31: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/31.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 32: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/32.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Renal Considerations
• Pre-renal (low FeNa)
– Not necessarily volume depleted
– Decreased Renal Perfusion Pressure
RPP = CO x RVR
CO = cardiac output
RVR = Renal Vascular Resistance
Optimize RPP
Flow is preferable to resistance
Dobutamine versus phenylephrine
![Page 33: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/33.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
If Physical Exam, CVP, and
PCWP do not determine volume
status, how do we optimize our
patients?
Build a Starling curve one patient at a time
Assess the effect of volume infusion on stroke volume
![Page 34: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/34.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Volume seeking versus room for diuresis
![Page 35: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/35.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Stroke Volume Variation
• Based on clinical observation that respi-
phasic variation of arterial pressure in
intra-vascular volume depletion
• Quantitative Measure
• Available on Pulse Contour Systems
•Anesth Analg. 2003 May;96(5):1254-7.
•Eur J Anaesthesiol. 2004 Feb;21(2):132-8.
•Chest. 2005 Aug;128(2):848-54
•Anesth Analg. 2001 Apr;92(4):984-9
![Page 36: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/36.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
ROC curve for CVP, PAOP, spontaneous venous pulse, and pulse pressure variation
Δ PP, change in pulse pressure; Δ SP, change is systolic pressure; PAOP, pulmonary artery
occlusion pressure; RAP, right arterial pressure. Michard F, Boussat S, Chemla D, et al.: Relation between respiratory changes in arterial pulse pressure and fluid
responsiveness in septic patients with acute circulatory failure.
Am J Respir Crit Care Med 2000, 162:134–138.
![Page 37: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/37.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
ROC curves comparing the ability of the pulmonary artery occlusion pressure (PAOP), the left ventricular end-diastolic area index (EDAI), and the delta down component (dDown) of the positive pressure ventilation-induced arterial systolic pressure variation
Tavernier B et al. Anesthesiology. 1998 Dec;89(6):1313-21.
![Page 38: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/38.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Summary
Measures of Cardiac Output and Volume
Responsiveness
•Cardiac Output
– PAC thermodilution
– Pulse Contour
•Volume Responsiveness
– Transpulmonary
thermodilution (GEDVi)
– Stroke Volume Variation
•Pulse Contour – Offers CO and SVV monitoring
•PAC – CO monitoring, CVP, and PCWP
•GEDVi – Offers GEDVi, CO, and SVV
![Page 39: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/39.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Summary: How to monitor volume status
• Physical Exam
• Filling pressures
– CVP and PCWP
• Chest Radiograph
• Echocardiography
• Quantitative Measures of Cardiac Output
![Page 40: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/40.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
In order to follow stroke volume,
an accurate measure of cardiac
output is required
![Page 41: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/41.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
How to monitor cardiac output
• Thermodilution
– Pulmonary Artery Catheterization
– Transpulmonary – GEDV thermodilution
• Lithium Dilution
• Impedance Plethysmography
• Pulse Contour
• Esophageal Doppler
• TEE
![Page 42: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/42.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 43: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/43.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Impedance Cardiography (ICG)
• Alternating Current is Transmitted
Through the Chest
• Current Seeks Path of Least Resistance: The Blood Filled Aorta
• The ICG Measures the Baseline Impedance to this Current
• With Each Heartbeat, Blood Volume and Velocity in the Aorta Change
• ICG Measures the Corresponding Change in Impedance
• ICG Uses the Baseline and Changes in Impedance to Measure and Calculate Hemodynamic Parameters
![Page 44: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/44.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Aortic Blood Flow
Changes Impedance
Aorta ICG Waveform
![Page 45: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/45.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Global End Diastolic Volume
Index
• Requires CVP and thermodilution
arterial line
• Cold solution infused via CVP and
thermodilution curve generated
• This parameter is closely linked to intra-
thoracic blood volume
• Predicts volume responsiveness
Michard et al, CHEST 2003; 124:1900 –1908
![Page 46: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/46.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 47: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/47.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Li Dilution Cardiac Output • Technique
– Involves injecting small amount of Li into CVP and detecting dissappearance with special monitor on arterial line
• Reliabiltiy – As less invasive
– As good as PAC
• Drawbacks – Cumbersome
– Requires Li injectate
Crit Care Med 1997 25(11):1796-800
British Journal of Anaesthesia, 1993, Vol. 71, No. 2 262-266
![Page 48: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/48.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
![Page 49: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/49.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Pulse Contour Cardiac Output
• Utilizes Standard Arterial Line
• Calculates Stroke Volume
![Page 50: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/50.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Effect of vascular tone • The algorithm looks for characteristic
changes in the arterial pressure waveform that reflect changes in tone (i.e., MAP, Skewness, Kurtosis)
![Page 51: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/51.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Trending Stroke Volume
• Blood pressure is sampled at 100 Hz
• Changes in stroke volume will result in corresponding changes in the pulse pressure
• Whole waveform measure of the pulse pressure is achieved by taking the standard deviation of the sampled points of each beat
• Std(BP) Pulse Pressure Stroke Volume
• SV estimates are averaged over 20 seconds
PP SV
SBP
DBP
![Page 52: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/52.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Summary
• Most Reliable CO Monitor
– PA Catheter
– Pulse Contour
• Minimally Invasive
– LiCO
– Pulse Contour
• Non-Invasive
– Bioimpedance
![Page 53: Hemodynamic Monitoring and Volume Management€¦ · Hemodynamic Monitoring and Volume Management ... 121:2000–2008 ... •Tousignant CP, Anesth Analg 2000;](https://reader030.vdocuments.net/reader030/viewer/2022021806/5bb6db0c09d3f2a4338c3de4/html5/thumbnails/53.jpg)
DIVISION OF CRITICAL CARE MEDICINE GWUMC
Conclusion
• Optimal assessment of intra-vascular volume status requires quantitative measures of cardiac output
• PA catheters reliably provide this function
• Echocardiography is also reliable
• Less invasive technologies are evolving and appear promising