High position resolution gamma-ray imagers consisting of a monolithic MPPC array with submillimeter pixelized scintillator crystals Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa Yamamura, Sigeyuki Nakamura, Nobuyuki Kawabata, Hirokazu Ikeda, Kei Kamada Abstract—We report on the development of two versatile, high spatial resolution gamma-ray imagers for medical imaging. One is a compact gamma-ray camera, the other is a tweezers type coin- cidence imaging system. These applications consisting of a large- area monolithic Multi-Pixel Photon Counter (MPPC) and submil- limeter pixelized scintillator matrices. The MPPC array has 4 × 4 channels with a three-side buttable, very compact package. Each channel has a photosensitive area of 3 × 3 mm 2 and 3600 Geiger mode avalanche photodiodes (APD). For a typical operational gain of 7.5× 10 5 at + 20 degrees, gain fluctuation over the entire MPPC device is only ± 5.6%, and dark count rates (as measured at the 1 p.e. level) amount to 400 kcps per channel. We particularly selected Ce-doped (Lu,Y)2(SiO4)O (Ce:LYSO) and a brand-new scintillator, Ce-doped Gd3Al2Ga3O12 (Ce:GAGG) due to their high light yield and density. To improve the spatial resolution, these scintillators were fabricated to 22 × 22 or 15 × 15 matrices of 0.5 × 0.5 mm 2 pixels. These scintillator matrices were coupled to the MPPC array with an acrylic light guide with 1 mm thick, and signals were read out using the charge division resis- tor network, which compiles signals into four position-encoded analog outputs. The spatial resolution of 1.2 mm was achieved with the compact gamma-ray camera using collimated 57 Co source, and a radiography image of a bearing was successfully obtained. On the other hand, the spatial resolution of 1.1 mm was achieved with the coincidence imaging system using a 22 Na source. Furthermore the experimental measurements for a PET scanner was performed, and the spatial resolution of 0.91 mm was achieved. These results suggest that the gamma-ray imagers has excellent potential for their uses as a high spatial medical imaging, and also be promising for positron emission tomography (PET). Manuscript received ????? T. Kato, J. Kataoka, T. Nakamori, A. Kishimoto are with the Research Institute for Science and Engineering, Waseda University, Shinjuku, Tokyo 169-8555, Japan (e-mail: [email protected]; [email protected]; [email protected]; daphne3h- [email protected]). S. Yamamoto is with Nagoya University Graduate School of Medicine, 1- 1-20, Daikominami, Higashi-ku, Nagoya-shi, Aichi 461-8673, Japan (e-mail: [email protected]). K. Sato, Y. Ishikawa, K. Yamamura, S. Nakamura, N. Kawabata are with Hamamatsu Photonics, K.K., 1126-1, Ichino-cho, Higashi-ku, Hamamatsu- shi, Shizuoka 435-8558, Japan (e-mail: [email protected]; yoshi- [email protected]; [email protected]; [email protected]; kawa- [email protected]). H. Ikeda is with ISAS/JAXA, 3-1-1, Yoshinodai, Chuo-ku, Sagamihara-shi, Kanagawa, 252-5210, Japan (e-mail; [email protected]). K. Kamada is with Materials Research Laboratory, Furukawa Co., Ltd., 1-25-13, Kannondai, Tsukuba, Ibaraki 305-0856, Japan (e-mail; k- [email protected]). I. I NTRODUCTION Positron emission tomography (PET) imaging is a well- established method of detecting cancers and diagnosing Alzheimer’s in its early stages [1]. Currently, many advan- tageous aspects of PET combined with Magnetic Resonance Imaging (MRI) have being proposed (MRI-PET), and with prototypes now being tested as MRI produces an excellent soft-tissue contrast and anatomical detail without additional radiation [2]–[4]. However, a Photo-Multiplier Tube (PMT) incorporated in conventional PET scanners is difficult to use within the high magnetic field of MRI. Moreover, the large- size PMT-based PET not only complicates use in narrow MRI tunnels but also limits the spatial resolution far from the theoretical limits of PET resolution. A Multi-Pixel Photon Counter (MPPC), also known as a Silicon Photo-Multiplier (SiPM), is a promising semiconductor photodetector for PET. It is insensitive to magnetic fields and compact. In addition, it has a gain comparable to that of PMTs at up to the 10 5 10 6 level, resulting in good signal-to-noise (S/N) ratio and excellent timing property [5]. These great advantages make the MPPC an ideal photosensor for MRI- PET as well as for Time Of Flight (TOF) applications [6], [7]. A high-resolution MRI-PET/TOF-PET technique utilizing the MPPC array is now being developed. We previously devel- oped and tested a monolithic, three-side buttable 4 × 4 MPPC array with submillimeter pixelized Ce-doped (Lu,Y) 2 (SiO 4 )O (Ce:LYSO) and Ce-doped Gd 3 Al 2 Ga 3 O 12 (Ce:GAGG) scintil- lator matrices [8]. In the position histograms, each scintillator pixel is clearly resolved, suggesting the possibility of its use for submillimeter high resolution gamma-ray imaging application. In this paper, we developed two MPPC-based gamma-ray imagers; one is a compact gamma-ray camera [9], the other is a tweezers type coincidence imaging system [10]. Furthermore, we evaluated a spatial resolution of prototype gantry for a PET scanner. II. MATERIALS AND METHODS A. 4 × 4 MPPC array Fig. 1 shows a picture of the monolithic 4 × 4 MPPC array [11] developed in this paper. The MPPC array was designed and developed for future applications in nuclear medicine (such as PET scanners) by Hamamatsu Photonics K.K.. Each channel has a photosensitive area of 3 × 3 mm 2 and 60 × 60

Upload: others

Post on 18-Aug-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

High position resolution gamma-ray imagersconsisting of a monolithic MPPC array withsubmillimeter pixelized scintillator crystals

Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, YoshitakeIshikawa, Kazuhisa Yamamura, Sigeyuki Nakamura, Nobuyuki Kawabata, Hirokazu Ikeda, Kei Kamada

Abstract—We report on the development of two versatile, highspatial resolution gamma-ray imagers for medical imaging. One isa compact gamma-ray camera, the other is a tweezers type coin-cidence imaging system. These applications consisting of a large-area monolithic Multi-Pixel Photon Counter (MPPC) and submil-limeter pixelized scintillator matrices. The MPPC array has 4× 4channels with a three-side buttable, very compact package. Eachchannel has a photosensitive area of 3× 3 mm2 and 3600 Geigermode avalanche photodiodes (APD). For a typical operationalgain of 7.5× 105 at + 20 degrees, gain fluctuation over the entireMPPC device is only ± 5.6%, and dark count rates (as measuredat the 1 p.e. level) amount to ≤ 400 kcps per channel. Weparticularly selected Ce-doped (Lu,Y)2(SiO4)O (Ce:LYSO) anda brand-new scintillator, Ce-doped Gd3Al2Ga3O12 (Ce:GAGG)due to their high light yield and density. To improve the spatialresolution, these scintillators were fabricated to 22× 22 or 15× 15matrices of 0.5× 0.5 mm2 pixels. These scintillator matrices werecoupled to the MPPC array with an acrylic light guide with 1 mmthick, and signals were read out using the charge division resis-tor network, which compiles signals into four position-encodedanalog outputs. The spatial resolution of 1.2 mm was achievedwith the compact gamma-ray camera using collimated 57Cosource, and a radiography image of a bearing was successfullyobtained. On the other hand, the spatial resolution of 1.1 mmwas achieved with the coincidence imaging system using a 22Nasource. Furthermore the experimental measurements for a PETscanner was performed, and the spatial resolution of 0.91 mmwas achieved. These results suggest that the gamma-ray imagershas excellent potential for their uses as a high spatial medicalimaging, and also be promising for positron emission tomography(PET).

Manuscript received ?????T. Kato, J. Kataoka, T. Nakamori, A. Kishimoto are with the Research

Institute for Science and Engineering, Waseda University, Shinjuku,Tokyo 169-8555, Japan (e-mail: [email protected];[email protected]; [email protected]; [email protected]).

S. Yamamoto is with Nagoya University Graduate School of Medicine, 1-1-20, Daikominami, Higashi-ku, Nagoya-shi, Aichi 461-8673, Japan (e-mail:[email protected]).

K. Sato, Y. Ishikawa, K. Yamamura, S. Nakamura, N. Kawabata are withHamamatsu Photonics, K.K., 1126-1, Ichino-cho, Higashi-ku, Hamamatsu-shi, Shizuoka 435-8558, Japan (e-mail: [email protected]; [email protected]; [email protected]; [email protected]; [email protected]).

H. Ikeda is with ISAS/JAXA, 3-1-1, Yoshinodai, Chuo-ku, Sagamihara-shi,Kanagawa, 252-5210, Japan (e-mail; [email protected]).

K. Kamada is with Materials Research Laboratory, Furukawa Co.,Ltd., 1-25-13, Kannondai, Tsukuba, Ibaraki 305-0856, Japan (e-mail; [email protected]).

I. INTRODUCTION

Positron emission tomography (PET) imaging is a well-established method of detecting cancers and diagnosingAlzheimer’s in its early stages [1]. Currently, many advan-tageous aspects of PET combined with Magnetic ResonanceImaging (MRI) have being proposed (MRI-PET), and withprototypes now being tested as MRI produces an excellentsoft-tissue contrast and anatomical detail without additionalradiation [2]–[4]. However, a Photo-Multiplier Tube (PMT)incorporated in conventional PET scanners is difficult to usewithin the high magnetic field of MRI. Moreover, the large-size PMT-based PET not only complicates use in narrow MRItunnels but also limits the spatial resolution far from thetheoretical limits of PET resolution.

A Multi-Pixel Photon Counter (MPPC), also known as aSilicon Photo-Multiplier (SiPM), is a promising semiconductorphotodetector for PET. It is insensitive to magnetic fields andcompact. In addition, it has a gain comparable to that of PMTsat up to the 105∼ 106 level, resulting in good signal-to-noise(S/N) ratio and excellent timing property [5]. These greatadvantages make the MPPC an ideal photosensor for MRI-PET as well as for Time Of Flight (TOF) applications [6],[7].

A high-resolution MRI-PET/TOF-PET technique utilizingthe MPPC array is now being developed. We previously devel-oped and tested a monolithic, three-side buttable 4× 4 MPPCarray with submillimeter pixelized Ce-doped (Lu,Y)2(SiO4)O(Ce:LYSO) and Ce-doped Gd3Al2Ga3O12 (Ce:GAGG) scintil-lator matrices [8]. In the position histograms, each scintillatorpixel is clearly resolved, suggesting the possibility of itsuse for submillimeter high resolution gamma-ray imagingapplication. In this paper, we developed two MPPC-basedgamma-ray imagers; one is a compact gamma-ray camera [9],the other is a tweezers type coincidence imaging system [10].Furthermore, we evaluated a spatial resolution of prototypegantry for a PET scanner.

II. MATERIALS AND METHODS

A. 4× 4 MPPC array

Fig. 1 shows a picture of the monolithic 4× 4 MPPC array[11] developed in this paper. The MPPC array was designedand developed for future applications in nuclear medicine(such as PET scanners) by Hamamatsu Photonics K.K.. Eachchannel has a photosensitive area of 3× 3 mm2 and 60× 60

Page 2: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

Fig. 1. Photo of the 4× 4 MPPC array developed in this paper.

)5 1

Gai

n (

Bias voltage [V]

71.5 72.0 72.5

4

6

8

10A-1

71.5 72.0 72.5

4

6

8

10B-1

71.5 72.0 72.5

4

6

8

10C-1

71.5 72.0 72.5

4

6

8

10D-1

71.5 72.0 72.5

4

6

8

10A-2

71.5 72.0 72.5

4

6

8

10B-2

71.5 72.0 72.5

4

6

8

10C-2

71.5 72.0 72.5

4

6

8

10D-2

71.5 72.0 72.5

4

6

8

10A-3

71.5 72.0 72.5

4

6

8

10B-3

71.5 72.0 72.5

4

6

8

10C-3

71.5 72.0 72.5

4

6

8

10D-3

71.5 72.0 72.5

4

6

8

10A-4

71.5 72.0 72.5

4

6

8

10B-4

71.5 72.0 72.5

4

6

8

10C-4

71.5 72.0 72.5

4

6

8

10D-4

6.66.8

77.27.47.67.8

88.2

6.6

6.8

7

7.2

7.4

7.6

7.8

8

8.2

)5 10×Gain (

AB

CD

12

34

Fig. 2. left: Gain variation as a function of bias voltage for all pixelsfrom 71.4 to 72.4 V, measured at +20 degrees. right: Gain distribution at theoperation voltage of 72.01 V

Geiger mode avalanche photodiodes (APDs) arranged witha pitch of 50 µm. The gap between each channel is only0.2 mm thanks to the monolithic structure. The MPPC arrayis placed on a surface-mounted package measuring 14.3 by13.6 mm, and fabricated into a three-side buttable structure,that is, the distance from the photosensitive area to the edgeof the package is only 500 µm. An excellent gain uniformity(± 5.6%) (Fig. 2) and very low dark count rates (≤400 kcp,due to the 1 p.e. level) have been achieved at an averaged gainof 7.5× 105, measured at +20 degrees. Table I lists the otherbasic characteristics of the MPPC array.

Also, the energy and time resolutions were obtained as11.5± 0.5% (FWHM at 662 keV photoelectric peak) and493± 22 ps (FWHM), respectively when the MPPC arraywere optically coupled with a Ce:LYSO scintillator [8].

B. Scintillators

To fabricate gamma-ray imaging applications, we selectedCe:LYSO and Ce:GAGG scintillators. Ce:LYSO, which is oneof the most popular scintillator at present in medical imaging,has features such as high light yield (75% of Tl:NaI), shortscintillation decay time (40 nsec) and high density (7.4 g/cm3)greater than Bi12Ge3O20 (BGO) (7.1 g/cm3) [12]. However,Ce:LYSO contains a considerable amount of self radiationemitted from 176Lu. Alternatively, a brand-new scintillator,Ce:GAGG also have very high light yield and short scintillatordecay time, and it is noteworthy that Ce:GAGG is absence ofself radiation [13]. Table II lists the other basic characteristicsof the Ce:LYSO and the Ce:GAGG scintillators.

Fig 3 (right) shows 137Cs spectra obtained by us-ing a 3× 3× 10 mm3 Ce:LYSO and Ce:GAGG (Fig 3(left)) crystals with a 50 µm-type 3× 3 mm2 MPPC

TABLE ISPECIFICATION OF THE 4× 4 MPPC ARRAY AT + 25 DEG

Parameters SpecificationNumber of elements [ch] 4× 4

Effective active area / channel [mm] 3× 3Pixel size of a Geiger-mode APD [µm] 50

Number of pixels / channel 3600Typical photon detection efficiency1(λ=440 nm) [%] 50

Typical dark count rates / channel [kcps] ≤ 400Terminal capacitance / channel [pF] 320

Gain (at operation voltage) 7.5× 105

1 including cross-talk and after-pulse contributions

TABLE IIBASIC CHARACTERISTICS OF THE CE:LYSO AND CE:GGAG

SCINTILLATORS

Ce:LYSO Ce:GGAGDensity [g/cm3] 7.10 6.63

Light yield [photons/MeV] 25,000 46,000Decay time [nsec] 40 88(91%) and 258(9%)

Peak wavelength [nm] 420 520

(Hamamatsu:S10362-33-050C), measured at +20 degrees. TheMPPC was operated at the gain of 7.5× 105. Typically,MPPCs are most sensitive within the range of 350-500 nm[14]. In this sense, a emission of the Ce:GAGG, peaking at520 nm is not favorable, but the output charges from theMPPC with the Ce:GAGG was about 21% larger than thatof the Ce:LYSO due to the high light yield of Ce:GAGG. TheEnergy resolution for the 662 keV photoelectric peak were9.9% and 7.9% for the Ce:LYSO and the Ce:GAGG afterlinearity correction [15], respectively. Their high light yieldshould realize the better energy resolution, and moreover, goodspatial resolution when they are fabricated to small pixels andread out by a charge division resistor network [9].

C. Configuration of compact gamma camera

The compact gamma camera consists of the MPPC arraywith a Ce:GAGG scintillator matrix, namely, 22× 22 matrix of0.5× 0.5 mm2 pixels (Fig. 4 (left)). Each pixel is divided witha reflective BaSO4 layer 0.1 mm thick, and the total size of thescintillator matrix is 13.1× 13.1× 10 mm3. The scintillatormatrix were optically coupled to the MPPC array with anacrylic light guide 1 mm thick, which distributes scintillatorphotons across multiple MPPC array channels.

The signals from the MPPC array was compiled into fourposition-encoded analog outputs by using a charge divisionresistor network [8]. The four signals were fed into QuadLinear FAN IN/OUT (Phillips MODEL 6954) and divided intotwo lines. One was directly fed into the charge sensitive ADC(HOSHIN V005; hereafter CSADC), with the other beingsummed over four signals to generate a trigger with the non-update discriminator (Technoland N-TM 405).

The tungsten sheet, which is the complex material of thepowder of W and the resins and has a stopping power equalto a Pb, with a 1 mm diameter and 3 mm long hole wasused as a gamma-ray collimator. The collimator was placedat the distance of 12 mm from the surface of the Ce:GAGGmatrix (Fig 4 (right)). The Ce:GAGG matrix was not enclosedany shields. The spatial resolution of this gamma camerawas evaluated by means of taking images of the collimated

Page 3: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

Charge [pC]0 50 100 150 200 250 300 350 400

Nor

mal

ized

cou

nts

1

10

210

310Ce:LYSO

Ce:GGAG

Fig. 3. left: Photo of the 3× 3× 10 mm3 Ce:LYSO and the Ce:GAGGscintillators. right: Energy spectra of 137Cs source. Red line and dashedblack line represent the Ce:GAGG and Ce:LYSO, respectively.

Fig. 4. left: Photo of the 22× 22 of 0.5× 0.5 mm2 Ce:GAGG matrix.right: Configuration of the compact gamma camera.

122 keV gamma-ray irradiated by a 57Co source. The 57Cowas placed at the distance of 66 mm from the surface oftungsten collimator, and its position was flexibly controlledparallel to the tungsten collimator using the X-stage (SGSP 20-85, Sigma Koki) to evaluate the resolution performance, whilethe accuracy of the stage controllers was 1 µm/pulse. Theimages were taken by changing the position of 57Co sourceat 3.46 mm intervals (corresponding to a viewing angle of3 degrees). The image quality without the collimator was alsoevaluated by taking a radiography image of a small (10 mmdiameter) ball bearing made of stainless steel, irradiated bythe 57Co source.

D. Configuration of tweezers type imaging system

The tweezers type imaging system has two phosphor sand-wich (phoswich) gamma-ray detector blocks consisting of theMPPC arrays, Ce:LYSO and Ce:GAGG scintillator matrices.These scintillator matrices are composed of 15× 15 matricesof 0.5× 0.5 mm2 pixels optically separated by BaSO4 layer0.1 mm thick (Fig 5 (left)). The total size of the scintillatormatrices are 9.7× 9.7× 5 mm3, and the configurations ofeach matrix is completely matched. The Ce:LYSO matrix wascoupled to the MPPC array with the acrylic light guide 1 mmthick, and the Ce:GAGG matrix was coupled to the other sideof the Ce:LYSO matrix (Fig 5 (right)). The detector block iscapable of two-layer Depth of Interaction (DoI) measurementby identifying in which scintillator matrices the event occured.Two detector blocks were then attached on acrylic tongs, andform a tweezers type coincidence gamma-ray imager (Fig 6).

Output signals from the MPPC arrays were fed into acoincidence DAQ system developed by ESPEC TECHNOCORP. after compiled four position-encoded analog outputs

Fig. 5. left: Photos of the 15× 15 of 0.5× 0.5 mm2 Ce:LYSO (left side)and Ce:GAGG (right side) matrices. right: Configuration of the two-layerphoswich gamma-ray detector block. The MPPC array, the acrylic light guide,the Ce:LYSO and the Ce:GAGG scintillator matrices are optically coupledeach other.

Fig. 6. Photo of the developed tweezers type coincidence imager using a pairof MPPC arrays coupled with the 0.5× 0.5 mm2 Ce:LYSO and Ce:GAGGscintillator matrices.

(x direction; X+ and X−, y direction; Y+ and Y−) bythe summing operational amplifiers [9], [16]. The compiledsignals were digitized by a 100 M samples/s ADC (AD9218BST-105), and then, processed by field-programmable gatearrays (FPGAs). When the digital signals were over thethreshold of digital comparator, the signals are integratedwith two different integration time (130 ns and 320 ns). Thepositional distributions were calculated by the anger-logic;x = (X+/(X++X−)) and y = (Y+/(Y++Y−)), and the energywas delivered from the sum of four signals. If the timingsignals from the MPPC arrays coincidence within the time-window of 20 ns and their energies are within the energy-window of 511± 102.2 keV, the HIT address, timing and validflag are stored in a memory for use in creating list-mode data.

In a performance test, a 22Na source (0.593 MBq at themeasurement date) was set between the two detector blocks.The 22Na source was doped within the central 0.25 mm φregion and could hence be regarded as a point source.

E. Configuration of prototype gantry for a PET scanner

We plan to fabricate a PET scanner using the MPPC arraysand the fine scintillator matrices. In a preliminary performancetest, two MPPC-based PET detectors consisting of the MPPCarrays, 0.5× 0.5 mm2 Ce:LYSO and Ce:GAGG scintillatormatrices described in section II-D were used. The photo ofthe experimental setup is provided in Fig 7. The 0.25 mmφ 22Na point source was located between the MPPC-basedPET detectors, and its position was then flexibly controlled

Page 4: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

Fig. 7. Experimental setup of the prototype system for a PET scanner. Thedistance between two detector blocks is 70 mm.

-40 -30 -20 -10 0 10 20 30 40

-40

-30

-20

-10

0

10

20

30

40

0

200

400

600

800

1000

1200

1400

1600

1800

X position (a.u.)

Y p

ositi

on (

a.u.

)

-40 -30 -20 -10 0 10 20 30 40

-40

-30

-20

-10

0

10

20

30

40

0

100

200

300

400

500

600

700

X position (a.u.)

Y p

ositi

on (

a.u.

)

Fig. 8. Flood images of 22× 22 of 0.5× 0.5 mm2 Ce:GAGG scintillatormatrix with a 137Cs (left) and a 57Co (right) source.

by the X-stage and the θ-stage (SGSP 80Y-AW, Sigma Koki),whose accuracy was 2.5× 10−3 deg/pulse. Coincidence eventswere taken at ten positions by changing the roll angle (θ) at18 degree intervals from 0 to 162 degrees. At each step, datawere taken for 10 minutes.

III. RESULTS

A. Compact gamma-ray camera

1) flood image: Fig 8 shows flood images obtained for the22× 22 Ce:GAGG scintillator matrix by irradiation a 137Csand a 57Co source. The flood images show overlapping peaksof the side pixels, as the total size of scintillator matrix isa bit larger than the sensitive area of the MPPC array. Andscintillation photons from the side pixels were only deliveredinto the side channels of the MPPC array. However, the centralpixels were successfully resolved in the flood images. In theflood image of the 137Cs source, the peaks are narrowerthan that of the 57Co source. This is because the 662 keVgamma-ray irradiated by 137Cs source generate greater amountof charges from the MPPC array. The greater the chargesinvolved, naturally, the better positional statistical error.

We extracted energy spectra from only resolved pixels byselecting events around the corresponding peaks in the floodimages. Fig 9 shows example spectra of the 137Cs and the 57Cosource after the linearity correction. The energy resolution for662 keV and 122 keV were 14.2% and 21.8%, respectively.

Energy [keV]0 100 200 300 400 500 600 700 800

Cou

nts

210

310

Energy [keV]0 20 40 60 80 100 120 140 160 180 200

Cou

nts

10

210

310

Fig. 9. Example energy spectra of a 137Cs and a 57Co source for0.5× 0.5 mm2 Ce:GAGG matrix extracted from the red square in Fig 8.

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

-4 -3 -2 -1 0 1 2 3 4-4

-3

-2

-1

0

1

2

3

4

X position (mm)

Y p

ositi

on (

mm

)

Fig. 10. Images of collimated 57Co source. From upper left to lower right,the 3.46 mm movement of the 57Co source are presented.

A energy window of 122 keV± FWHM/2 (corresponding to∼13 keV) was applied when the collimated 57Co and theradiography images were taken.

2) pinhole image: The images of the collimated 57Cosource was shown in Fig 10. The 3.46 mm movement of the57Co source was clearly distinguished. The spatial resolutionat the center of field of view was 1.2 mm (FWHM) (withoutcorrection of the collimator hole diameter of 1 mm).

3) radiography image: Fig 11 (right) shows the radiogra-phy image of the ball bearing, which was smoothed by thegaussian filter of σ = 0.15 mm. The structure of the ballbearing including 1 mm ball can be clearly resolved.

B. Tweezers type coincidence imager

1) flood image: Fig 12 show flood image results obtainedfor 15× 15 Ce:LYSO and Ce:GAGG scintillator matricesindividually coupled to the MPPC array. Fig 13 show exampleenergy spectra extracted from the flood images. The energyresolutions for 662 keV, which were not corrected for non-linearity, were 14.0% and 9.4% for Ce:LYSO and Ce:GAGG,respectively.

2) signal selection: Fig 14 shows the wave form corre-sponding to 511 keV photoelectric absorption by the Ce:LYSOand Ce:GAGG. The decay time of the signals, which in-cludes the scintillator decay time and the MPPC response,are approximately 70 ns and 150 ns for the Ce:LYSO and theCe:GAGG, respectively. Fig 15 (left) shows the 2D energyplot; x-axis and y-axis correspond to the 130 ns and 320 nsADC channel, respectively. For the 130 ns integration time,the signals of the Ce:LYSO is same or a little bit larger thanthat of Ce:GAGG as shown in Fig 14, while, for the 320 ns

Page 5: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

Fig. 11. left: Photo of the 10 mm diameter ball bearing. right: Radiographyimage of the ball bearing smoothed by the gaussian filter of σ = 0.15 mm.

50 100 150 200 250 300 350 400 45050

100

150

200

250

300

350

400

450

0

20

40

60

80

100

120

140

X position (a.u.)

Y p

ositi

on (

a.u.

)

50 100 150 200 250 300 350 400 45050

100

150

200

250

300

350

400

450

0

20

40

60

80

100

120

X position (a.u.)

Y p

ositi

on (

a.u.

)

Fig. 12. Flood images of 15× 15 of 0.5× 0.5 mm2 Ce:LYSO (left) andCe:GAGG (right) scintillator matrices with a 137Cs source.

integration time, the signals of the Ce:GAGG is larger. As aresult, the events of each scintillator trace two different linesin 2D energy plot. Fig 15 (right) show the 130 ns to 320 nsADC channel ratio. If the Fast(130 ns)/Slow(320 ns) ratio isover 0.84, the event is regarded as being originated from theCe:LYSO, and conversely if the Fast/Slow ratio is under 0.84,the event is regarded as the Ce:GAGG event. Fig 16 show theenergy spectra after event selection, and each scintillator eventcan be effectively distinguished.

3) Image reconstruction: Fig 17 show the sinogram trans-formed from the list mode data. The two interaction positionsin each coincidence event, which define the Lines of Response(LoR), were randomly determined according to the uniformdistribution within the hit pixels. The simple planar imagereconstruction [17] were performed with the intersection ofthe all LoRs and the plane centrally located between the twodetectors. Fig 18 shows the planar image of the 22Na pointsource. The non-DoI image were reconstructed by reckoningthe Ce:LYSO and the Ce:GAGG pixels as a single pixel. Whenthe DoI information was applied, the spatial resolution were1.1 mm (x direction) and 1.4 mm (y direction), which wasslightly better than the case of non-DoI (1.2mm (x direction)and 1.5 mm (y direction)).

C. Prototype gantry for a PET scanner

Maximum Likelihood-Expectation Maximization (MLEM)[18] method was used to reconstruct images. Fig 19 showssinograms and the resultant reconstructed images obtained forthe center and off-center (3.0 mm), respectively. The radialspatial resolution was estimated at 0.91 and 0.94 mm for thecenter and 3.0 mm off-center, respectively.

ADC ch

0 20 40 60 80 100 120

Cou

nts 210

ADC ch

0 20 40 60 80 100 120

Cou

nts

10

210

Fig. 13. Example energy spectra for Ce:LYSO (left) and Ce:GAGG (right)matrices, extracted from the red square in Fig 12.

nsec0 50 100 150 200 250 300 350

mV

-140

-120

-100

-80

-60

-40

-20

0

Ce:LYSO

Ce:GAGG

130ns 320ns

Fig. 14. Wave forms of the Ce:LYSO and the Ce:GAGG scintillator,corresponding to 511 keV photoelectric absorption event. Red line and dashedblack line represent the Ce:GAGG and Ce:LYSO, respectively.

IV. CONCLUSION

In this work, we developed two versatile high spatialresolution gamma-ray cameras, which both achieved ∼1 mmspatial resolutions. These gamma-ray cameras consist of the4× 4 MPPC arrays and submillimeter pixelized scintillatormatrices. The MPPC arrays have fine gain uniformity of± 5.6% and very low dark count rates of ≤ 400 kcps asmeasured at +20 degrees. Moreover, its compactness due tothe monolithic and three-side buttable structure is suit for thecompact medical imaging devices. For the compact gammacamera, 0.5× 0.5 mm2 Ce:GAGG scintillator matrix wasused. The spatial resolution of 1.2 mm was achieved for thecollimated 57Co image at the center of the field of view. For thetweezers type coincidence imager, 0.5× 0.5 mm2 Ce:LYSOand Ce:GAGG scintillator matrices were used for phoswichdetector. In the 2D energy plot, optical signals from theCe:LYSO and the Ce:GAGG can be effectively distinguished.The spatial resolution of 1.1 mm was achieved for the simpleplanar image reconstruction when the DoI information wasapplied. In the preliminary measurement for a PET scanner,we used two detectors consisting of the MPPC arrays, theCe:LYSO and the Ce:GAGG scintillator matrices with chang-ing the roll angle. The radial spatial resolutions of 0.91 mm(center) and 0.94 mm (3 mm off-center) were achieved froma MLEM reconstructed images. These results suggest that a

Page 6: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

130ns ADC channel

0 5 10 15 20 25 30 35 40

320n

s A

DC

cha

nnel

0

5

10

15

20

25

30

35

40

50

100

150

200

250

300

350

400

450

500

Ce:LYSOC

e:G

AG

G

100×130ns ADC channel / 320ns ADC channel

0 20 40 60 80 100 120

Cou

nts

0

500

1000

1500

2000

2500

3000

Ce:GAGG Ce:LYSO

Fig. 15. left: 2D energy plot. x and y axis represent the 130 ns and 320 nsADC channel, respectively. right: 130 ns to 320 ns ADC channel ratio. Rightpeak and left peak represent the Ce:GAGG and Ce:LYSO events, respectively.

ADC channel

0 10 20 30 40 50 60 70 80 90

Cou

nts

0

100

200

300

400

500

600

700Ce:LYSO

Ce:GAGG

Sum

Fig. 16. Energy spectra distinguished from 2D energy plot. Black line, boldred line and dashed bold blue line represent the sum, Ce:GAGG and Ce:LYSOspectra, respectively.

100

200

300

400

500

600

00

14

14

detector1

dete

ctor

2

Fig. 17. left: Coincidence diagram constructed between two detector blocks.right Schematic chart for the experimental setup, seen from the top view.

-4 -3 -2 -1 0 1 2 3 4

-4

-3

-2

-1

0

1

2

3

4

0

50

100

150

200

250

300

350

400

450

X position [mm]

Y p

ositi

on [m

m]

-4 -3 -2 -1 0 1 2 3 4

-4

-3

-2

-1

0

1

2

3

4

0

100

200

300

400

500

X position [mm]

Y p

ositi

on [m

m]

Fig. 18. The reconstructed planar images applied non-DoI (left) and DoI(right) information.

-6 -4 -2 0 2 4 60

20

40

60

80

100

120

140

160

180

0

20

40

60

80

100

120

140

160

180

200

r [mm]

)θan

gle

(

-6 -4 -2 0 2 4 60

20

40

60

80

100

120

140

160

180

0

20

40

60

80

100

120

140

160

r [mm]

)θan

gle

(

-6 -4 -2 0 2 4 6-6

-4

-2

0

2

4

6

0

100

200

300

400

500

X position [mm]

Y p

ositi

on [

mm

]

-6 -4 -2 0 2 4 6-6

-4

-2

0

2

4

6

0

50

100

150

200

250

300

X position [mm]

Y p

ositi

on [

mm

]

Fig. 19. Sinograms and reconstructed images by MLEM algorithm, measuredwith a 22Na point source placed at center (left) and off-axis positionx = 3 mm (right).

monolithic MPPC array coupled with submillimeter pixelizedCe:LYSO and Ce:GAGG matrices could be promising as highspatial resolution medical imaging, and have encouraged usto develop MPPC-based modules for use in submillimeterresolution PET scanners.

REFERENCES

[1] William W. Moses, “Trends in PET imaging”, Nucl. Instr. and Meth. A471 (2001) 209-214.

[2] Raymond R. Raylman, et al., “Initial tests of a prototype MRI-compatiblePET imager”, Nucl. Instr. and Meth. A 569 (2006) 306-309.

[3] Hao Peng, et al., “Proof-of-principle study of a small animal PET/field-cycled MRI combined system using conventional PMT technology”, Nucl.Instr. and Meth. A 612 (2010) 412-420.

[4] Siichi Yamamoto, et al., “Design and performance from an integratedPET/MRI system for small animals”, Ann Nucl Med (2010) 24:89-98

[5] Takeshi Nakamori, et al., “Development of a gamma-ray imager using alarge area monolithic 4× 4 MPPC array for a future PET scanner”, JINST7 (2012) C01083.

[6] A. Nassalski, et al., “Multi Pixel Photon Counters (MPPC) as an Alterna-tive to APD in PET Applications”, IEEE Trans. Nucl. Sci. 57 no.3 (2010)1008-1014.

[7] Chang Lyong Kim, et al., “Multi-Pixel Photon Counters for TOF PETDetector and Its Challenges”, IEEE Trans. Nucl. Sci. 56 no.5 (2009) 2580-2585.

[8] T. Kato, et al., “A novel gamma-ray detector with submillimeter res-olutions using a monolithic MPPC array with pixelized Ce:LYSO andCe:GGAG crystals”, Nucl. Instr. and Meth. A, in press (2012).

[9] Seiichi Yamamoto, et al., “Development of a high-resolution Si-PM-basedgamma camera system”, Phys. Med. Biol., 56 (2011) 7555-7567.

[10] Seiichi Yamamoto, et al., “Development of a tweezers-type coincidenceimaging detector”, Ann Nucl Med (2008) 22:387-393

[11] K. Sato, et al., “Application Oriented Development of Multi-PixelPhoton Counter (MPPC)”, IEEE Trans. Nucl. Sci. Conference Record,(NSS/MIC) (2010) 243-245.

[12] Ioannis G. Valais, et al., “Evaluation of the light emission efficiency ofLYSO:Ce scintillator under X-ray excitation for possible applications inmedical imaging”, Nucl. Instr. and Meth. A 569 (2006) 201-204.

[13] Kei Kamada, et al., “2 inch diameter single crystal growth and scin-tillator properties of Ce:Gd3Al2Ga3O12”, Journal of Crystal Growth 352(2012) 88-90.

Page 7: High position resolution gamma-ray imagers consisting of a ... · Takuya Kato, Jun Kataoka, Takeshi Nakamori, Aya Kishimoto, Seiichi Yamamoto, Kenichi Sato, Yoshitake Ishikawa, Kazuhisa

[14] K. Yamamoto, et al., “Development of Multi-Pixel Photon Counter(MPPC)”, IEEE Trans. Nucl. Sci Conference Record, N24-292 (2007)1511-1515.

[15] T. Kato, et al., “Development of a large-area monolithic 4× 4 MPPCarray for a future PET scanner employing pixelized Ce:LYSO and Pr:LuAGcrystals”, Nucl. Instr. and Meth. A 638 (2011) 83-91.

[16] Seiichi Yamamoto, et al., “Development of a Si-PM-based high-resolution PET system for small animals”, Phys. Med. Biol. 55 (2010)5817-5831.

[17] P. Beltrame, et al., “The AX-PET demonstrator-Design, construction andcharacterization”, Nucl. Instr, and Meth. A 654 (2011) 546-599.

[18] Lucas Parra, et al., “List-mode likelihood: EM Algorithm and ImageQuality Estimation Demonstrated on 2-D PET”, IEEE Trans. Med. Imag.17, no.2 (1998) 228-235.