    How the New ISO Particle Count

    Standard Will Affect You

    Leonard Bensch, Pall Corporation

    Tags: particle counting, contamination control

    ISO Revises Standards for Particle Counting and Reporting

    Fluid contamination is a primary root cause of mechanical component wear and reduction in system reliability

    and performance. Because of this, most companies have established programs to monitor and control

    contamination. These programs include particle counting. Three new or revised ISO fluid power standards

    will have a major impact on such programs. The changes are primarily the result of a new method for

    calibrating liquid automatic particle counters (APCs) that is traceable to the National Institute of Standards

    and Technology (NIST), but unfortunately produces a substantially different calibration. This article presents

    an introduction to the new calibration method including the impact on resultant particle sizes, particle counts

    and contamination classes.

    AUTOMATIC PARTICLE COUNTER CALIBRATION

    APCs have generally replaced optical microscopy as the primary method for quantifying particulate

    contamination in fluid wetted systems. In the late 1960s a calibration procedure was developed to ensure that

    particle counts obtained with an APC agreed as closely as possible with counts obtained by optical

    microscopy. The method, using AC Fine Test Dust (ACFTD), ultimately became International Standard ISO

    4402 [1], and is still in use today by most oil analysis laboratories around the world.

    For some time, industry had been aware that when more sophisticated instruments such as scanning

    electron microscopes are used to analyze the ACFTD, a substantial increase in the numbers o f particles

    (especially below 10 m) are reported compared to the distribution based on optical microscopes, given in

    ISO 4402. Also, in 1992, AC Rochester discontinued manufacturing ACFTD. Therefore National Fluid Power

    Association (NFPA) and ISO standards committees began to develop a revised APC calibration method

    based on a new contaminant whose particle size distribution could be NIST traceable. ISO 12103-A3 [2]

    Medium Test Dust (ISO MTD) was selected as the best candidate. In 1993, NIST undertook an effort to certify

    the particle size distribution of suspensions of ISO MTD as a reference material for APC calibration. This

    effort resulted in NIST Standard Reference Material SRM 2806 consisting of a suspension of ISO MTD in

    MIL-H-5606 hydraulic fluid. Their analyses show a significant difference in the particle size distribution of ISO

    MTD as measured with an electron microscope compared to previous results with an APC calibrated per ISO

    4402 (see Figure 1).

    Figure 1shows significantly more particles at sizes below about 10 m were observed by NIST than with theACFTD calibration. The higher particle counts are a result of the enhanced sensitivity of scanning electron

    microscopy when compared to optical microscopy performed in the 1960s. At sizes larger than 10 m, fewer

    particles were observed by NIST. This is primarily because NIST reported the projected area equivalent

    spherical diameter of particles, which is smaller than the longest chord dimension used to generate the

    published ACFTD size distribution. The two methods intersect around the 10 m level.

    ISO has updated the current APC calibration procedure to use the SRM 2806. The revised method was

    approved on December 9, 1999, as ISO/FDIS 11171 [3]. This calibration procedure includes many other

    enhancements to ensure better resolution, accuracy, repeatability and reproducibility; however, the effect of

    the new ISO MTD dust and NIST counts will have the largest impact. Also, a new designation, m(c) is

    recommended for reporting particle sizes using the new ISO 11171 procedure, with the (c) referring to

    certified sizes traceable to NIST, so that the results are easily distinguished from those obtained with ISO

    4402.

    Related Articles

    Particle Counting - Peril or Prize?

    Advice for Measuring Oil Filter Life

    Identifying Turbine Oil Cleanliness Levels

    Value of Particle Counting in Oil Analysis

    Home | Buyers Guide | Glossary | Events | Bookstore | Newsletters | Browse Topics

    GREASES HYDRAULICS INDUSTRIAL LUBRICANTS SYNTHETICS FILTRATION STORAGE/HANDLING OIL ANALYSIS

    Tweet Like 2 Share

    the New ISO Particle Count Standard Will Affect You http://www.machinerylubrication.com/Read/12/iso-parti

    7/2/2016

  • 7/25/2019 How the New ISO Particle Count Standard Will Affect You

    2/4

    ISO has also developed a new procedure, ISO 11943 [4], for calibration and verification of online particle

    counting and dilution systems. ISO 11943 is applied to online particle counters used for contamination

    analysis of operating systems in the field as well as filter test systems.

    REDEFINITION OF PARTICLE SIZES

    Based on the size distribution of ISO MTD in Figure 1, for particle sizes below about 10 m, the particle size

    determined by NIST for a given particle count is greater than the corresponding particle size per ISO 4402.

    The difference between the methods increases as the particle size decreases. For example, the particle

    count for 1 m (ISO 4402 calibration) of about 2000 particles/g corresponds to a particle size of 4.2 m(c)

    (NIST size distribution). Thus, below about 10 m, particle sizes with NIST (ISO 11171) calibration will appear

    higher than the old definition (ISO 4402), e.g., 2 m is 4.6 m(c) and 5 m is 6.4 m(c). Above about 10 m,

    the opposite relationship exists; new particle sizes will appear lower than the old definition, e.g., 15 m is

    about 13.6 m(c). The relationship between particle sizes defined by the ISO 4402 and NIST size

    distributions for ISO MTD is shown in Table 1. Note that these changes will not impact particle counts

    obtained with optical microscopy, as these counting methods have not changed.

    Because of possible differences between automatic particle counters and the accuracy of their original APC

    calibration, these relationships may vary slightly and must be determined for each APC to be used. It should

    also be noted that even though a particle size or count may change because of the new calibration, the actual

    contamination level in a system will not be influenced and will remain the same.

    IMPACT OF NEW CALIBRATION ON PARTICLE COUNTS

    Particle counts obtained with an APC calibrated with the new ISO 11171 procedure will differ from the

    corresponding particle counts obtained with the APC calibrated per ISO 4402 at any particular size. Users of

    particle count data must be made aware of the APC calibration method and how to interpret results when

    using the new calibration method. As a first approximation, historical automatic particle count data may be

    converted from ISO 4402 sizes to the new NIST sizes using Table 1.

    The values in Table 2illustrate the effect of the two calibration procedures on particle counts. As shown in the

    table, if one changes from ISO 4402 to the new ISO 11171 calibration without making adjustments to the

    sizes being monitored, significant differences arise in particle concentrations. For particle sizes smaller than

    about 10 m, apparent increases in particle concentration will be reported which could prompt unnecessary

    action. The magnitude of the difference increases with decreasing particle size. For sizes larger than 10 m,

    the reverse occurs and apparent decreases in concentration result from the new calibration. Failure to

    recognize this variation as the result of a change in calibration, rather than a change in contamination level,

    may lead to misinterpretation of particle count data and inappropriate action.

    CHANGES TO THE ISO CODE

    On December 2, 1999, ISO approved modifications to the ISO 4406 coding method based on the new

    calibration procedure. For APC counts, the revised procedure, ISO FDIS 4406.2 [5], uses three code

    numbers, corresponding to the concentrations of particles larger than 4 m(c), 6 m(c) and 14 m(c) with the

    new calibration method. The new 6 m(c) and 14 m(c) sizes correspond closely to ISO 4402 sizes of 5 mand 15 m (see Table 1). These sizes were chosen so that no significant shift in code number occurs due to

    changes in the APC calibration method. For optical microscopy the calibration is unchanged and the two

    digits will remain the same as before at 5 m and 15 m. Thus, the second two digits of the actual code will

    be similar regardless of the calibration or measurement technique used.

    In some instances, companies had previously adopted for their own use a 3-digit form of the ISO 4406 code

    with the third digit representing 2 m. For these companies the change to three digits used in the new ISO

    4406:1999 will be relatively easy because the second two digits will typically remain the same as before (only

    the particle sizes are adjusted). Because the first digit for the new code will represent 4 m(c) rather than 4.6

    m(c) which more closely corresponds to the old 2 m, an increase in the particle count and ISO code for this

    digit may be noticed. As a rule of thumb, specifications should be increased in the first digit by about 1 code

    level with the second two digits remaining unchanged. For example, a specification of 17/15/12 based on 2,

    5, and 15 m (ACFTD calibration) should be changed to ISO 18/15/12 based on 4, 6, and 14 m(c) (NIST

    calibration).

    Figure 2reflects the impact on the ISO 4406 code for automatic particle counts conducted on two typical fluid

    samples. Note that for the new NIST calibration per ISO 11171, the cleanliness classes are determined by

    particle counts > 4, 6, and 14 m(c). In each case, the second two digits of the code are nearly identical to

    the previous ISO 4406 2-digit code. Indeed, differences only result when observed particle concentrations

    correspond to transitions between two code numbers. As a result of the change in size classes with the

    revised code, the ISO code designation for a particular sample will not change appreciably, regardless of

    whether the new or old calibration method is used. This permits equipment users and others who specify

    cleanliness levels in terms of an ISO code to maintain their historical databases and their existing cleanliness

    specifications without change.

    Changes In NAS 1638 - In order to implement the new NIST calibration with minimal impact on NAS

    cleanliness classes, it has been proposed to revise the reference particle size ranges in NAS 1638 [6] based

    on the particle size relationship in Table 1; therefore, the actual particle sizes measured would be similar

    White Papers

    Controlling Gearbox Lubricant Contamination

    Buyers Guide

    Oil Analysis Laboratories

    Oil Analysis Instruments

    Oil Sampling Hardware

    Poll

    Have you used the ISO

    cleanliness code to set target

    alarms for system

    cleanliness?

    Yes

    No

    the New ISO Particle Count Standard Will Affect You http://www.machinerylubrication.com/Read/12/iso-parti

    7/2/2016

  • 7/25/2019 How the New ISO Particle Count Standard Will Affect You

    3/4

    regardless of the method of calibration.

    As an example, 5 m per ISO 4402 calibration is replaced with 6 m(c) per NIST calibration. Because optical

    microscopic particle counting procedures are not being changed, no changes are required in the cleanliness

    standards provided the size ranges counted correspond to the original size ranges set by the standards. In

    addition to the change in particle size, it has also been proposed that NAS 1638 be changed to utilize

    cumulative particle counts.

    Table 3shows the changes proposed in the NAS 1638 cleanliness classes for APC analysis. It should be

    noted that, because the first size range has been extended to 1 m or 4 m(c), new contamination limits are

    included.

    CONCLUSIONS

    Three new or revised ISO standards will have an impact on particle counting and reporting standards - ISO

    11171, ISO 11943 and ISO 4406. The new APC calibration method, ISO 11171, utilizes NIST certified

    calibration samples, instead of obsolete ACFTD. ISO 11943 for online APC calibration is based on the

    revised calibration method and validates the entire on-line particle counting system, including a dilution

    system when used. ISO 4406, the revised solid contamination level code, uses the new calibration method

    and adds a third size, 4 m(c), to the code for APC results. Sizes for the revised code were chosen to allow

    users of the code to retain their historical data base and leave cleanliness level specifications based on the

    code unchanged.

    The adoption and implementation of these new and revised international standards will cause the following to

    occur:

    Apparent particle sizes when using an APC calibrated per ISO 11171 will change;

    Particle counts when using an APC calibrated with ISO 11171 will change;

    Cleanliness codes will change unless APC reported particle sizes are adjusted in accordance with the

    revised ISO 4406:1999 or the proposed revised NAS 1638.

    The projected changes in each of these parameters have been discussed in this

    paper. Users should always keep in mind that any reported changes are an artifact

    of the methods utilized and actual contamination levels will remain the same as

    before.

    REFERENCES

    1) ISO 4402, Hydraulic fluid power - Calibration of automatic-count instruments for particles suspended in

    liquids - Method using classified AC Fine Test Dust contaminant, 1991.

    2) ISO 12103, Road vehicles - Test dust for filter evaluation - Part 1: Arizona test dust, 1997.

    3) ISO FDIS 11171, Hydraulic fluid power Calibration of liquid automatic particle counters, 1999.

    4) ISO 11943, Hydraulic fluid power - on-line liquid automatic particle-counting systems - Methods of

    calibration and validation, 1999.

    5) ISO FDIS 4406.2, Hydraulic fluid power - Fluids - Code for defining the level of contamination of solid

    particles, 1999.

    6) NAS 1638, Cleanliness Requirements of Parts Used in Hydraulics Systems, 1992.

    AUTHOR INFORMATION

    Dr. Leonard Bensch

    Vice President, Pall Corporation

    25 Harbor Park Drive

    Port Washington, NY 11050

    Phone: 516-484-3600

    the New ISO Particle Count Standard Will Affect You http://www.machinerylubrication.com/Read/12/iso-parti

    7/2/2016

  • 7/25/2019 How the New ISO Particle Count Standard Will Affect You

    4/4

    Practicing Oil Analysis(5/2000)

    Related Articles

    Particle Counting - Peril or Prize?

    Advice for Measuring Oil Filter Life

    Identifying Turbine Oil Cleanliness Levels

    Value of Particle Counting in Oil Analysis

    Get 60 Questions to Ask Your Lubricant

    Supplierabsolutely FREE when you sign up

    for our free Lube-Tips newsletter. Email:

    Oil Analysis BasicsPresents the fundamentals of oil analysis for

    machinery condition monitoring in an easy to

    understand format. Learn everything from how to

    take a proper oil sample to how to select a test slate

    for your applications.

    Table of Contents| Read Excerpts

    Machinery Oil AnalysisPresents the practice of oil analysis as a condition

    monitoring tool for machines. Covers each step in

    the sampling, testing and diagnostic process.

    Includes latest advances in technology and

    instrumentation.

    Table of Contents| Read Excerpts

    Services Subscribe | Contact Us | Privacy Policy | RSS | Advertise

    Quick Links Home | Buyers Guide | Glossary | Events | Bookstore | Newsletters | Browse Topics

    NORIA CORPORATION MACHINERY LUBRICATION | RELIABLE PLANT

    the New ISO Particle Count Standard Will Affect You http://www.machinerylubrication.com/Read/12/iso-parti

    7/2/2016