hplc application recent development stationary phase material dr suman presentation
TRANSCRIPT
2
Chromato-graphy / -Chromato-graphy / -graph / -gram / -graphergraph / -gram / -grapher
• Chromatography:Chromatography: Separation Separation techniquetechnique
• Chromatograph:Chromatograph: InstrumentInstrument• Chromatogram:Chromatogram: Obtained Obtained
“Picture/Spectra”“Picture/Spectra”
• Chromatographer:Chromatographer: PersonPerson
Invention of Invention of Chromatography by M. Chromatography by M.
TswettTswettEther
CaCO3
Chlorophyll
ChromatoChromatography
ColorsColors
Comparing Comparing Chromatography to the Chromatography to the
Flow of a River... Flow of a River...
Base
Water flow
Light leaf
Heavy stone
5
Mobile Phase / Mobile Phase / Stationary PhaseStationary Phase
• A site in which a moving A site in which a moving phase (phase (mobile phasemobile phase) and ) and a non-moving phase a non-moving phase ((stationary phasestationary phase) make ) make contact via an interface contact via an interface that is set up.that is set up.
• The affinity with the The affinity with the mobile phase and mobile phase and stationary phase varies stationary phase varies with the solute. with the solute. SeparationSeparation occurs due to occurs due to differences in the speed differences in the speed of motion.of motion.
Strong
Weak
Mobile Mobile phasephase
StationarStationary phasey phase
7
Separation Process and Separation Process and
ChromatogramChromatogram for Column for Column ChromatographyChromatography
Ou
tpu
t co
nce
ntr
ati
on
Time
ChromatogramChromatogram
8
Pump
Sample injection unit(injector)
Column
Column oven(thermostatic
column chamber)
Detector
Eluent (mobile phase)
Drain
Data processorDegasser
Flow Channel Diagram for High Flow Channel Diagram for High Performance Liquid Performance Liquid
ChromatographChromatograph
11
ChromatogramChromatogram
tR
t0
Inte
nsi
ty o
f d
ete
ctor
sig
nal
Time
Peak tR : Retention time
hA
t0 : Non-retention time
A : Peak areah : Peak height
Retention Factor, Retention Factor, kk
tR
t0
Str
en
gth
of
dete
ctor
sig
nal
Time
tR: Retention timet0: Non-retention time
0
0R
t
ttk
13
Theoretical Plate Theoretical Plate Number, Number, NN
W
W1/2
H1/2
H
2
.21
R
R
/
R
W
t
W
2
2
2
545
16
Area
Ht
tN
14
Resolution, Resolution, RRSS
2,2/11,2/1
RR
21
RRS
12
12
18.1
)(21
hh WW
tt
WW
ttR
tR1 tR2
W1 W2
W1/2h,1 W1/2h,2 h1/2
16
Objectives of Objectives of PretreatmentPretreatment
• To improve sensitivity and selectivityTo improve sensitivity and selectivity• To improve the accuracy of To improve the accuracy of
quantitative valuesquantitative values• To protect and prevent the To protect and prevent the
deterioration of columns and analytical deterioration of columns and analytical instrumentsinstruments
• To simplify measurement operations To simplify measurement operations and proceduresand procedures
• To stabilize target substancesTo stabilize target substances
17
Filtration and Centrifugal Filtration and Centrifugal SeparationSeparation
• In general, filter every In general, filter every sample before injection!sample before injection!
• It is convenient to use a It is convenient to use a disposable filter with a disposable filter with a pore diameter of pore diameter of approx. 0.45 approx. 0.45 µµm.m.
• Centrifugal separation Centrifugal separation is applicable for is applicable for samples that are samples that are difficult to filter.difficult to filter.
Filter Syringe
18
DeproteinizationDeproteinization
• PrecipitationPrecipitation– Addition of organic solvent (e.g., Addition of organic solvent (e.g.,
Acetonitrile)Acetonitrile)– Addition of acid (e.g., Trichloroacetic Addition of acid (e.g., Trichloroacetic
acid, Perchloric acid)acid, Perchloric acid)– Addition of heavy metal or neutral Addition of heavy metal or neutral
salt(Ammonium Sulfate)salt(Ammonium Sulfate)
• UltrafiltrationUltrafiltration
19
Solid Phase ExtractionSolid Phase Extraction
(1)Conditioning
(2)Sample addition
(3)Rinsing
(4)Elution
Solvent with low elution strength
Solvent with high elution strength
Target componentUnwanted components
20
Pre-Column Pre-Column DerivatizationDerivatization
• OPA Reagent (Reacts with Primary OPA Reagent (Reacts with Primary Amines)Amines)
o-phthalaldhyde(OPA)
+ R-NH2 N-R
S-R’
R’-SH
NO2
2,4-dinitrophenylhydrazine(2,4-DNPH)
+
CHO
CHO
2,4-DNPH (Reacts with Aldehydes and Ketones)
O2N
NHNH2
C=OR
R’NO2
O2N
NHN=C
H+
R
R’
HPLC ApplicationsHPLC Applications
Chemical
Environmental
Pharmaceuticals
Consumer Products
Clinical
PolystyrenesDyesPhthalates
TetracyclinesCorticosteroidsAntidepressantsBarbiturates
Amino Acidsvitaminshomocysteine
BioscienceProteinsPeptidesNucleotides
LipidsAntioxidantsSugars
Polyaromatic HydrocarbonsInorganic Ionsherbicides
23
Qualitative AnalysisQualitative AnalysisIdentification based on retention timeIdentification based on retention time
Identification based on retention factor and response
Identification based on sample spiking
Spectral Peak Identification by PDA The use of selective detectors and spectrometers can greatly increase the confidence in the peak assignment. Detector systems such as Diode Array UV Spectrometers are able to record unique spectra for each peak within the sample chromatogram. The spectra may be recorded in ‘real time’ as the eluent can be directly introduced into the detector system.
Peak Purity
Peak purity can be established by taking the ratio of two signals (wavelengths) across the peak(s) of interest. If the peak is pure, then the ratio of the two signals should be constant across the peak. If the peak is impure, then the ratio between the two signals (wavelengths) will change across the peak as the spectral differences caused by the interfering peak change the signal ratio.
Spectral Characterization Mass Spectrometric detectors can be configured to produce fragmentation patterns that can be assigned to analyte moieties, so building up a ‘picture’ of the analyte molecule. The spectral peaks and patterns combined with the molecular (pseudomolecular) weight acquired can be used to characterise the analyte molecule.
29
Quantitative AnalysisQuantitative Analysis
• Quantitation performed with peak Quantitation performed with peak area or height.area or height.
• Calibration curve created Calibration curve created beforehand using a standard.beforehand using a standard.
• - External standard method- External standard method– Absolute calibration curve methodAbsolute calibration curve method– Internal standard methodInternal standard method– Standard addition methodStandard addition method
Peak Height or Peak Area
Peak area is mostly used. In HPLC peaks may be tailed. In this case, peak heights may vary (although area will remain constant), area vales are more repeatable. A disadvantage of peak-area methods is that they are more affected by neighboring peaks. For trace analysis, when the peak of interest is very small, use peak height for calculations, this reduces the error sustained in small changes in peak start and end time variation.
Area %/ Height % (Normalization)
External Standard Quantitation The ESTD procedure uses absolute response factors. The response
factor is normally calculated as amount / area of the analyte in the calibration sample. Response factors are normally viewed as a single point calibration curve, using the origin to determine the second point for the regression line – this approach is often called ‘Single Point Calibration’.
37
Calibration Curve for Calibration Curve for Absolute Calibration Absolute Calibration
Curve MethodCurve Method
C1
C4
C3
C2
ConcentrationArea
A1
A2
A3
A4C1 C2 C3 C4
A1
A2
A3
A4
Concentration
Peak a
rea
Calibration curveCalibration curve
A line of best fit (regression line) is used to join the points of the curve obtained. The line of best fit is usually given a ‘Correlation Coefficient’ which is the square root of the regression coefficient and gives a measure of how well the data points fit a straight line. The intercept of the regression equation indicates systematic error – a large positive or negative value may indicate an inherent error within the sample preparation, concentration or analysis .The slope of the line indicates the analytical ‘sensitivity’.
The Matrix EffectThe matrix effect problem occurs when the unknown sample contains many impurities.If impurities present in the unknown interact with the analyte to change the instrumental response or themselves produce an instrumental response, then a calibration curve based on pure analyte samples will give an incorrect determinationSuggested method:1. Internal Standard Method2. Standard Addition Methods
Internal Standard Method
Internal standards are used to adjust for variations in analytical response due to instrumental and/or matrix effects and variations in the amount of sample due to variable injection volumes.•It must have similar chemical properties to the target substance.•Its peak must appear relatively near that of the target substance.•It must not already be contained in the actual samples.•Its peak must be completely separated from those of other sample components. •It must be added at a concentration that will produce a peak-area or peak height ratio of about unity with compound(s) of interest• It must be stable; unreactive with sample components, column packing, or mobile phase• It is desirable for it to be commercially available in high purity•Are added in the concentration range 0.3 – 0.5 portion of the expected MAXIMUM analyte concentration
Internal standards are used to adjust for variations in analytical response due to instrumental and/or matrix effects and variations in the amount of sample due to variable injection volumes.•It must have similar chemical properties to the target substance.•Its peak must appear relatively near that of the target substance.•It must not already be contained in the actual samples.•Its peak must be completely separated from those of other sample components. •It must be added at a concentration that will produce a peak-area or peak height ratio of about unity with compound(s) of interest• It must be stable; unreactive with sample components, column packing, or mobile phase• It is desirable for it to be commercially available in high purity•Are added in the concentration range 0.3 – 0.5 portion of the expected MAXIMUM analyte concentration
Drug Substance Internal StandardCholesterol compounds 19-hydroxycholesterolPhenolic acid 2,3,4 trihydroxybenzoic acidThuringiensin Adenosine MonophosphateOxolinic Acid Nalidixic AcidChlormadinone Acetate Medroxyprogesterone-17-AcetateAcetylsalicylic Acid Salicylic AcidSorbitol Methyl NonadecanateSimvastatin LovastatinDiazepam And Oxazepam Prazepam
43
Calibration Curve for Calibration Curve for Internal Standard Internal Standard
MethodMethod
C1
C4
C3
C2
Concentration Area
A1
A2
A3
A4C1/CIS C2 /CIS C3 /CIS C4 /CIS
A1/AIS
A2 /AIS
A3 /AIS
A4 /AIS
Concentration of target substance / Concentration of
internal standard
Are
a f
or
targ
et
sub
stan
ce /
Are
a f
or
inte
rnal st
an
dard
Calibration curveCalibration curveTarget
substanceInternal standard
CIS
CIS
CIS
CIS
AIS
AIS
AIS
AIS
44
Advantages of Advantages of Internal Standard Internal Standard
Method (1)Method (1)• Not affected by inconsistencies in injection Not affected by inconsistencies in injection
volume.volume.
10 µL injected
9 µL injected
CX / CIS
AX / AIS
XIS
XIS
Same Same area ratioarea ratio
45
Advantages of Advantages of Internal Standard Internal Standard
Method (2)Method (2)• Not affected by the pretreatment Not affected by the pretreatment
recovery rate.recovery rate.
100% recovery
rate
90% recovery rate
CX / CIS
AX /
AIS
XIS
XIS
Same Same area ratioarea ratio
Standard Addition Standard Addition MethodsMethods
Better method to use when matrix effects can be Better method to use when matrix effects can be substantialsubstantial
Standards are added directly to aliquots of the Standards are added directly to aliquots of the sample, therefore matrix components are the same.sample, therefore matrix components are the same.
Procedure:Procedure:
•Obtain several aliquots of sample (all with the Obtain several aliquots of sample (all with the same volume).same volume).
• known and unknown are the same analyteknown and unknown are the same analyte
•Spike the sample aliquots ==> add different Spike the sample aliquots ==> add different volume of standards with the same concentration volume of standards with the same concentration to the aliquots of sampleto the aliquots of sample
•Dilute each solution (sample + standard) to a fixed Dilute each solution (sample + standard) to a fixed volumevolume
•Measure the analyte concentrationMeasure the analyte concentration
48
Standard addition usually involves the addition of an
increasing increments of a standard solution to sample
aliquots of the same size (spiking).
49
Glucose in blood serum is Glucose in blood serum is determined spectrophotometrically by determined spectrophotometrically by the formation of a the formation of a colored complex colored complex with with o-toludineo-toludine. . Six identical Six identical 50.0 50.0 L samples L samples of of blood serum were treated with blood serum were treated with increasing amount of known increasing amount of known standardsstandards of glucose. of glucose. The following results were obtained:The following results were obtained:What is the glucose concentration in What is the glucose concentration in the blood serum samples.the blood serum samples.
Example of Standard Addition Data
The magnitude of the x-intercept is the concentration of the glucose in the original solution. To understand why this is so, consider the absorbance at the following two values:•at x = 0, the value of y is the absorbance of the solution with no added standard (i.e., it corresponds to the concentration of silver that we ultimately want).•at the x-intercept, there is no absorbance.Thus, the magnitude of the difference between x=0 and the x-intercept is the concentration of silver that is needed to produce the signal for the original solution of interest! Our job now is to determine the x-intercept.
The magnitude of the x-intercept is the concentration of the glucose in the original solution. To understand why this is so, consider the absorbance at the following two values:•at x = 0, the value of y is the absorbance of the solution with no added standard (i.e., it corresponds to the concentration of silver that we ultimately want).•at the x-intercept, there is no absorbance.Thus, the magnitude of the difference between x=0 and the x-intercept is the concentration of silver that is needed to produce the signal for the original solution of interest! Our job now is to determine the x-intercept.
51
The value of glucose in the original sample is found by The value of glucose in the original sample is found by determiningdetermining the x-intercept by extrapolation, i.e., the x-intercept by extrapolation, i.e., finding the value of x finding the value of x when y = 0. when y = 0.
Or by using the values shown by the Linear Fit and the form of Or by using the values shown by the Linear Fit and the form of the the linear equationlinear equation
y = mx + by = mx + b0 = 0.006580 = 0.00658xx + 0.229 + 0.229
Solving for x, Solving for x, xx = -0.229/0.00658 = = -0.229/0.00658 = - 34.8- 34.8 g.g.
The amount of glucose in unknown glucose sample, that had The amount of glucose in unknown glucose sample, that had no standard glucose was added, is no standard glucose was added, is 34.80 34.80 g.g.
Since the volume was 50.0 Since the volume was 50.0 L, the [Glucose] = 34.8 L, the [Glucose] = 34.8 g / 50 g / 50
L orL or
1 1 g/g/L. This could also be expressed as [Glucose] = 0.696 L. This could also be expressed as [Glucose] = 0.696
mg/mLmg/mL
Microporous Silica Particles
Bonded Stationary Phase Bonded Stationary Phase
New generation of organo-silane material incorporates ethylene bridges into porous silica. Tetraethoxysilane + Bis(tetraethoxysilyl)ethane = Polyethoxysilane. • Provide pH stability from 1-12 • Five times more durability than earlier hybrids. •The homogenous surface offers some steric selectivity.
Ethylene Bridged Hybrid [BEH] HPLC Column.
Ethylene Bridged Hybrid [BEH] HPLC Column.
BEH C18 BEH ShieldRP18 BEH C8 BEH Phenyl BEH HILIC
BEH Amide
Ligand Type
Trifunctonal C18
Monofunctional Embedded Polar
Trifunctional C8
Trifunctional Phenyl-Hexyl
Unbonded BEH Particle
Trifunctional Carbamoyl
Ligand Density
3.1μmol/m2 3.3μmol/m2 3.2μmol/m2 3.0μmol/m2 n/a 7.5 μmol/m2
Carbon Load
18% 17% 13% 15% Unbonded 12%
Endcap Style
Proprietary TMS Proprietary Proprietary N/A None
pH Range 1-12 2-11 1-12 1-12 1-9 1-11
Temp.Limit 60 oC 45oC 60 oC 60 oC 45 oC 90 oCPore Diameter
130AO 130AO 130AO 130AO 130AO 130AO
Surface Area
185m2/g 185m2/g 185m2/g 185m2/g 185m2/g 185m2/g
Particle Sizes
1.7,2.5,3.5,5μm
1.7,2.5,3.5,5μm 1.7,2.5,3.5,5μm
1.7,2.5,3.5,5μm
1.7,2.5,3.5,5μm
1.7,2.5,3.5,5 μm
Charged Surface Hybrid columnWhen the silica surface, including both inner-pore area and outer-pore surface, is covalently modified by silyl ligands bearing charges, an opposite charges—can be permanently attached to the outer-surface by electrostatic attraction while being excluded from entering the pores. As a result, a material with a charged inner-pore surface and an oppositely-charged exterior surface can be created. CSH 130 C18= BEH 130 Base particle + low level of basic moieties + tri functional C18 / End cap. •Positive surface charge and Acidic pH. •Used for impurity profiling in stability indicating methods and/or forced degradation studies .The advantages of CSH Technology include:· Superior peak shape for basic compounds· Increased loading capacity· Rapid column equilibration after changing mobile phase pH· Improved batch to batch reproducibility· Exceptional stability at low and high pH.
Charged Surface Hybrid columnWhen the silica surface, including both inner-pore area and outer-pore surface, is covalently modified by silyl ligands bearing charges, an opposite charges—can be permanently attached to the outer-surface by electrostatic attraction while being excluded from entering the pores. As a result, a material with a charged inner-pore surface and an oppositely-charged exterior surface can be created. CSH 130 C18= BEH 130 Base particle + low level of basic moieties + tri functional C18 / End cap. •Positive surface charge and Acidic pH. •Used for impurity profiling in stability indicating methods and/or forced degradation studies .The advantages of CSH Technology include:· Superior peak shape for basic compounds· Increased loading capacity· Rapid column equilibration after changing mobile phase pH· Improved batch to batch reproducibility· Exceptional stability at low and high pH.
High strength silica columnWhen compared to ethylene-bridge-containing BEH and CSH particles, the higher silanophilicity of the 100% silica HSS particle offers significant advantages including increase polar compound retention and significantly different selectivity. Additionally, the HSS particle possesses the mechanical strength necessary to tolerate UPLC pressures up to1034 bar.It is designed to provide different selectivity for basic compounds when compared to traditional high coverage, fully endcapped C18 chemistries. This is due to the increased silanol activity on the silica particle surface when bonded at intermediate ligand densities with no endcapping.
High strength silica columnWhen compared to ethylene-bridge-containing BEH and CSH particles, the higher silanophilicity of the 100% silica HSS particle offers significant advantages including increase polar compound retention and significantly different selectivity. Additionally, the HSS particle possesses the mechanical strength necessary to tolerate UPLC pressures up to1034 bar.It is designed to provide different selectivity for basic compounds when compared to traditional high coverage, fully endcapped C18 chemistries. This is due to the increased silanol activity on the silica particle surface when bonded at intermediate ligand densities with no endcapping.
-synonyms: solid core particles, Fused-core particles-Most recent available particle size is 1.7 m (core 1.25 m, layer 0.23 m)
Solid Core HPLC Column
Solid CorePorous silica
Core shell columns and Van Deemter
Van Deemter equation: H = A +B/u + CuA-value (Eddy dispersion) - narrow particle size distribution in addition to an enhanced roughness of their surface compared to porous particles, leading to a smaller A-coefficient by about 40%.B-value (Longitudinal diffusion) - Compared to totally porous particles, the Fused-Core particles have a much shorter diffusion path because of the solid core.; 20% decrease in comparison with porous particles.C-value (Mass transfer resistances) – solid core, impenetrable by analytes cause shorter diffusion path – C-value is reduced.
Vendors and Available phasesVendor Column/product name
Average particle diameter (μm)
Shell thickness (μm)
Stationary phase chemistry
Advanced Material Technology
Halo 2.7 0.50C18, C8, HILIC, RP-amide, phenylhexyl, pentafluorophenyl
Advanced Material Technology
Halo Peptide-ES 160 Å
2.7 0.50 C18
Agilent Poroshell 300 5 0.25 C18, C8, C3Agilent Poroshell 120 2.7 0.50 EC-C18, SB-C18
Sigma–Aldrich Ascentis Express 2.7 0.50C18, C8, HILIC, RP-amide, phenylhexyl, pentafluorophenyl
Sigma–AldrichAscentis Express Peptide-ES 160 Å
2.7 0.50 C18
Phenomenex Kinetex2.61.7
0.350.23
C18, XB-C18, C8, HILIC, pentafluorophenyl
Macherey-Nagel Nucleoshell 2.7 0.5 RP-18, HILIC
Thermo Scientific Accucore 2.6 0.50C18, aQ, RP-MS, HILIC, phenylhexyl, pentafluorophenyl
Sunniest SunShell 2.6 0.5 C18
Commercially not available
Eiroshell1.71.71.7
0.350.250.15
C18
2D HPLCChromatographic technique in which the injected sample is separated by passing through two different separation stages. This is done by injecting the eluent from the first column onto a second column. Typically the second column has a different separation mechanism, so that bands that are poorly resolved from the first column may be completely separated in the second column. (For instance, a C18 column may be followed by a phenyl column.) Alternately, the two columns might run at different temperatures. The second stage of the separation must be run much faster than the first, since there is still only a single detector. Can also adopt in GC, CE, SEC,SFC. Multidimensional Protein Identification Technology (MudPIT) is a separation technique which utilized two chromatography techniques back to back and can be coupled to silica capillary. Typically the column consist of strong cation exchanger (SCX) followed by reversed phase (RP) material.
2D HPLCChromatographic technique in which the injected sample is separated by passing through two different separation stages. This is done by injecting the eluent from the first column onto a second column. Typically the second column has a different separation mechanism, so that bands that are poorly resolved from the first column may be completely separated in the second column. (For instance, a C18 column may be followed by a phenyl column.) Alternately, the two columns might run at different temperatures. The second stage of the separation must be run much faster than the first, since there is still only a single detector. Can also adopt in GC, CE, SEC,SFC. Multidimensional Protein Identification Technology (MudPIT) is a separation technique which utilized two chromatography techniques back to back and can be coupled to silica capillary. Typically the column consist of strong cation exchanger (SCX) followed by reversed phase (RP) material.