Ultra-fastPLATINblue2 UHPLC from KNAUER
Fast
PLATINblue UHPLC�ultimate�performance�HPLC�with�an�extended�pressure�range�up�to�1000�bar�(15000�psi)�at�fl�ow�rates�<�1.5�ml/min�
�minimal�dead�volume�of�25�µl�with�SmartMix�gradient�mixer
2
2
Short cycle timefast�gradients
�high-performance�autosampler�for�fast�injection�(15�s�or�<��0�s�with�needle�wash)
�optimized�communication�and�software�operations�for�short�initialization�times
2
2
2
High throughput�process�up�to�7���samples�in�2�x�3�4-well�plates�or�up�to����samples�in�standard�1.��ml�vials
2
Speed and productivity
Fast detection�data�acquisition�rates�up�to�200�Hz�provide�higher�resolution�and�peak�capacity,�as�well�as�a�more�accurate�integration
2
Increase your laboratory’s productivity by taking advantage of the benefi ts of PLATINblue ultra high-performance LC (UHPLC).
With�PLATINblue�UHPLC,�you�can�now�use��conventional�HPLC�columns�as�well�as�columns�packed�with�sub-2�µm�particles�on�one�system�to�achieve:
�faster�separations,�thanks�to�the�use�of�shorter�columns�with�equal�or�higher�separation�performance
�higher�resolution�and�unmatched�separation�effi�ciency�at�equivalent�column�lengths
2
2
2
mA
Um
AU
00
minmin
00
5050
100100
150150
200200
250250
300300
350350
Separation�of�benzoic�acid�esters
Comparison of UHPLC and HPLC
mA
Um
AU
0.960.96
minmin
0.980.98 113838
4040
4242
4444
4646
4848
5050
80 Hz80 Hz50 Hz50 Hz
20 Hz20 Hz
5 Hz5 Hz
Effect of data acquisition rate on the detection of a fast eluting peak
Enlarged�detail,�chromatogram�overlay
3
! TipThe PLATINblue system includes eluent fi lters for protection against particulate matter. When working with columns packed with sub-2 µm particle sizes, samples should also always be fi ltered through a 0.2 µm membrane.
PreciseHigh-quality PLATINblue components were designed to work together perfectly to enable accurate measurement with optimal reproducibility – a prerequisite for analysis results you can rely on.
The sharp and narrow peaks produced by fast chromatography in conjunction with low-noise and low-drift detectors ensure sensitive measurements.
Sensitivity and reproducibility
Ultimate detection sensitivity�electronically�controlled�pump�drive�provides�for�low�pulsation�eluent�fl�ow�and�a�stable�baseline
�high�light�intensity�over�the�full�UV-Vis�range�through�the�combination�of�a�long-life�“high�brightness”�D2�with�a�tungsten-halogen�lamp
�extremely�low�noise�(5�µAU)�and�drift�(<�50�µAU/h)�specifi�cations
�excellent�linearity�of�>�3�AU�for�precise�simultaneous�determination�of�main�substances�and�trace�components
2
2
2
2
4
!TipA very high injection peak can be disadvantageous in fast chromato-graphy since the fi rst sample peaks elute very quickly after the injection. Dissolve the sample in eluent whenever possible to minimize this disturbance.
pea
k ar
ep
eak
areaa
0.010.01
concentration [ng/µlconcentration [ng/µl]]
0.10.1 1110010000
1000010000
100000100000
1010 100100
10000010000000
Linearity of detection over 3 orders of magnitude
anthracene�in�ACN/H2O
ReproducibilityPrecision�can�only�be�as�good�as�the�weakest�link�in�the�chain.�All�critical�components�of�the�PLATINblue�system�were�optimized�for�the�demanding�requirements�of�UHPLC,�resulting�in�a�system�with�excellent�reproducibility.
Optimized:Pump� ✓Eluent�mixing� ✓Injection� ✓Column� ✓Detection� ✓Data�processing� ✓
5
mA
Um
AU
0.00.0
minmin
0.50.5 1.01.000
1010
2020
3030
4040
5050
6060
1.51.5 2.02.0
7070
8080
Six�successive�separations�using�online�mixing�of�a�ternary�eluentEluent:�3�%�CH3COOH/MeOH/ACN�=�7�/20/2Flow�rate:�1�ml/min,�Temp.:�30�°C,�Wavelength:�225�nm
Retention time<�0.3�%�RSD�Peak areas<�1.1�%�RSD�with�partial-loop�injection�(<�0.3�%�RSD�with�full-loop�injection)
Excellent reproducibility of overall system
Easy to useSimple to operate and to service
So that you can stay focused on what is really important, the PLATINblue ultra high-performance LC system is equipped with many useful features that make operation more convenient, relieve you from time-consuming routine checks and assist you in your daily work.
Simple lamp replacement thanks to detachable side panel
�
Comfortable operation�easy�to�read�color�LCD�display
intuitive�touchscreen�control*
�automatic�confi�guration�of�communication�with�PC�and�software�eliminates�time-consuming�manual�setup
�automatic�selection�of�optimal�detector�integration�time
2
2
2
2
TipThe lifespan of the detector lamps can be conserved by having one or both turned off during interruptions in analysis and switched back on automatically when resuming measurements.!
Easy access for quick maintenance
�solid�glass�doors�provide�convenient�access�to�system�components
�removable�side�panel�for�easy�lamp�exchange�and�maintenance�
Reliable and GLP compliant fast�Ethernet�communication
automatic�pump�head�detection�(RFID)
automatic�lamp�detection
automatic�wavelength�verifi�cation
comprehensive�GLP�reporting�functions
Safety and protection�system�leak�sensor,�minimum�pressure�shut-off�and�eluent�level�control�feature�in�ChromGate®�software�provide�protection�against�leaks�and�running�dry
�soft-start�fl�ow�for�protection�of�columns�and�system�components
*�touchscreen�version�optional
2
2
2
2
2
2
2
2
2
7
Intelligent
Smart product features that save time
�automatic�system�configuration�for�the��PLATINblue�system�via�the�instrument�network
navigation�with�tree�structure�for�quick�orientation
Only�ONE�software�for�your�chromatography�laboratory�is�required�to�control�all�KNAUER�instruments�and�a��wide�range�of�chromatography�instrumentation�from�other�manufacturers.�Your�lab�benefits�from�reduced�training�expenses,�uniform�operability�and�harmonized�data�reporting.�
Flexible reporting functions �report�layouts�are�easily�customized�via�multiple�options�in�Custom�Report�feature
�predefined�standard�reports�for�quick�output�of�results
2
2
2
2
User-friendly, scalable and compliant
Automatic configuration of PLATINblue systems makes installation simple
Scalability and compliance
�scalable�from�a�single�instrument��to�a�multi-user�enterprise-wide��solution�thanks�to�Client/Server��functionality
FDA�21�CFR�Part�11�compliance
Easy to use�simple�method�creation�via�software��“wizards”
�“sequence�wizard”�provides�assistance��and�saves�time�when�programming��a�sequence�of�analyses
Convenient diode-array features
clear�presentation�of�3-D�data
�easy�creation�of�spectrum�libraries��and�fast�spectrum�search
2
2
2
2
2
2
ChromGate® PLATINblue Edition is our easy-to-use software solution for laboratories with high demands for functionality, flexibility and scalability, as well as GLP compliance. Based on EZChrom Elite™ software, ChromGate® PLATINblue Edition software adds extended KNAUER-specific functions.
�
�
PLATINblue
2 PLATINblue P-1 Ultra High Pressure Pump
Ultra�high�pressure�pump�optimized�for�fast�liquid�chromatography,�capable�of�pressures�up�to�1000�bar�(15000�psi)�at�fl�ow�rates�up�to�1.5�ml/min.�The�electronically�controlled�pump�drive�provides�for�low�pulsation,�a�highly�precise�fl�ow�rate�and�a�stable�baseline.�
2 PLATINblue M-1 Modular Eluent Manager
Combination�module�to�add�quaternary�gradient�functionality�and�an�online�degasser�for�four�eluents�to�the�ultra�high�pressure�pump.�Additionally,�the�M-1�can�be�equipped�with�an�A/D-D/A�interface�for�adding�devices�with�an�analog�input/output.
2 PLATINblue UV-1 Multiple Wavelength Detector
High�sensitivity�detector�for�data�acquisition�rates�up�to�200�Hz�with�dual-lamp�functionality�for�a�wavelength�range�of�1�0–�00�nm.�Allows�simultaneous�measurement�of�up�to���wavelengths.�Best�in�class�for�noise�and�drift�values.�With�an�unsurpassed�linearity�of�3�AU,�trace�components�can�be�determined�accurately�even�when�main�components�are�present.
10
Component overview
2 PLATINblue T-1 Column Temperature Manager*
This�contact�oven�provides�for�accurate�thermostatting�over�the�entire�length�of�up�to�4�columns�from�10�to�150�°C.�The�eluent�is�equilibrated�before�reaching�the�column�and�is�cooled�before�reaching�the�detector�to�minimize�noise�and�drift.��A�stable�column�temperature�not�only�improves�reproducibility�but�considerably�improves�the�resolution�of�many�HPLC�methods�as�well.
*�ask�for�availability
2 PLATINblue AS-1 Autosampler
Fast�and�extremely�accurate�UHPLC�autosampler�can�handle�pressures�up�to�1000�bar�(15000�psi)�for�injection�of�samples�from�well�plates�and�standard�vials.�Supports�full�and�partial-loop�fi�lling�as�well�as�“microliter-pickup”�for�fl�exible�selection�of�injection�volumes.�The�tray�cooling�option�allows�for�analysis�of�temperature-sensitive�samples.�
2 PLATINblue PDA-1 Diode Array Detector*
This�extremely�sensitive�diode�array�detector�with�a�data�acquisition�rate�up�to�100�Hz�with�dual-lamp�confi�guration�has�a�wavelength�range�of�1�0–�00�nm.�Use�spectrum�data�to�determine�peak�purity�or�to�facilitate�the�identifi�ca-tion�of�unknown�substances.
11
TipPLATINblue components are available with either a touchscreen display or with a small status display for operation with software. Both types can be freely combined in one system.
!
VersatileWide range of application areas
PharmaceuticalDetermination�of�paracetamol�and�typical�impurities�produced�during�synthesis�
�1�� 4-aminophenol� 2�� paracetamol�3�� N-(4-hydroxyphenyl)propanamide�4�� N-(2-hydroxyphenyl)acetamide�5�� N-(3-chloro-4-hydroxyphenyl)acetamide���� 1-(4-hydroxyphenyl)ethanoneoxime�7�� N-phenylacetamide���� 4-(acetylamino)phenyl�acetate���� 4-nitrophenol�10�� chloroacetanilide�11�� 1-(2-hydroxyphenyl)ethanone
PLATINblue systems can be flexibly used to increase the throughput and sensitivity of many applications. Benefit from the advantages of high-speed UHPLC applications without having to completely convert your established HPLC methods. The following are only a few examples of the applications possible with PLATINblue.
EnvironmentalSeparation�of�benzene�derivatives
FoodQuick�determination�of�tocopherols
400
200
0
mA
U
0 1 2 3 4 5min
1
2
345 678 91011
Separation columnBlueOrchid�C1��1.��µm100�x�2��mm
Separation conditions Eluent:�� A:�phosphate�� � buffer�� � B:�ACNFlow�rate:���50�µl/minMode:�� gradientPressure:�� ��50�bar�
(13�00�psi)Detection:��254�nm�� � (�0�Hz,�0.005�s)Temp.:� 50�°C
1�� thiourea2�� toluene3�� propylbenzene4�� butylbenzene
0
mA
U
min
12
3
4
50
100
150
0 1 2 3
Separation columnBlueOrchid�C1��1.��µm100�x�2�mm
Separation conditionsEluent:�� ACN/H2O�� � �5:15�(v/v)Flow�rate:��1000�µl/minPressure:�� 720�bar��� � (10400�psi)Detection:��254�nm��� � (�0�Hz,�0.005�s)Temp.:� 35�°C
1� trans-retinol2�� alpha-tocopherol3�� beta-tocopherol4�� gamma-tocopherol5�� delta-tocopherol��� vitamin�D2
mA
U
0
min
1
2
34
0
5
610
20
30
40
1 2 3 4 5
Separation columnBlueOrchid�Si�1.��µm�0�x�2�mm
Separation conditions Eluent:�� � �heptane/
butanol�1000:4�(v/v)
Flow�rate:��1700�µl/minPressure:�� �350�bar��
(50�0�psi)Detection:��2�0�nm��� � (50�Hz,�0.01s)Temp.:� 25�°C
12
Clinical diagnosticsDetermination�of�sedatives,�e.g.�barbiturates
Selectivity testEngelhardt�test�in�less�than���minutes��
1�� Barbital2�� Luminal3�� Prominal4�� Revonal
mA
U
min
12
34
0 1 2 3 4
0
20
40
60
Separation columnBlueOrchid�C1��1.��µm�0�x�2�mm
Separation conditions Eluent:�� �MeOH/H2O�
�0:40�(v/v)Flow�rate:��500�µl/minPressure:�� ��2��bar�
(11��0�psi)Detection:���254�nm�
(�0�Hz,�0.005�s)Temp.:� 40�°C
1�� uracil2�� aniline3�� phenol4�� p-ethylaniline5�� N,N-dimethylaniline��� benzoic�acid�ethyl�ester7�� toluene��� ethylbenzene
Separation columnBlueOrchid�C1��1.��µm100�x�2�mm
Separation conditions Eluent:� �MeOH/H2O�
49:51�(v/v)Flow�rate:��430�µl/minPressure:�� ��0�bar�� � (12�00�psi)Detection:���254�nm�
(�0�Hz,�0.005s)Temp.:� 40�°C
mA
U
0 2 4 6 8 10min
123
4
5 67
8
25
0
50
75
100
13
Specifi cally designed for ultra high-performance liquid chroma-tography applications, KNAUER UHPLC columns packed with 2 µm or sub-2 µm particles make high-resolution separations possible. Their exceptional stability in both acidic and alkalic eluents make them universally applicable.
Bluespher columns�packed�with�2�µm�stationary�phase�share�many�of�the�physical�properties�of�columns�packed�with�our�long-established�Eurospher�columns�and�facilitate�therefore�the�scaling�of�existing�applications.
BlueOrchid columns�are�packed�with�an�ultra-pure�silica�gel�with�1.��µm�particles�for�ultimate�resolution.�These�columns�exhibit�a�particularly�low�backpressure.
Effi cientKNAUER UHPLC columns
FactsP-1 Ultra High Pressure Pump
Flow rate range� �0.01–3.00�ml/min,�� � adjustable�in�0.001�ml�� � incrementsMax. pressure� �1000�bar�(15000�psi)�at��
<�1.5�ml/min,��00�bar��(�700�psi)�at�<�3.0�ml/min
Flow rate accuracy� ±�1%Flow rate precision � <�0.1�%�RSDHigh-pressure gradient � up�to�4�pumpsLow-pressure gradient up�to�4�eluentsUltra-efficient mixing� yes,�SmartMix�included
M-1 Modular Eluent Manager
Low-pressure mixing/control�of�4�eluentsgradient moduleDegasser module� 4�channels,�Teflon®�AF,��� � high�efficiencyA/D-D/A interface 24-bitLAN� �T10/100�Ethernet�router�
included
UV-1 Multiple Wavelength Detector
Wavelength range 1�0–�00�nm�(UV-Vis)Wavelength accuracy*� <�1�nmLamps� “high-brightness”�D2�and� � tungsten-halogenMax. data acquisition rate 200�HzChannels� max.��Noise*� ±5�µAUDrift*� <�50�µAU/hLinearity*� >�3�AU
*�according�to�ASTM
T-1 Column Temperature Manager
Temperature range� 10–150�°CTemperature stability� +/–�0.1�°CEluent preheating� equilibration�to�temperature� � of�column�compartmentPostcolumn cooling� yesMaximum heating rate ��°C/min
Specifications – PLATINblue
Pump P-1 High-quality materials – stainless steel and glass
Easy access of detector flow cell
14
PDA-1 Photo Diode Array Detector
Wavelength range� 1�0–�00�nm�(UV-Vis)Lamps� “high-brightness”�D2�and��� � tungsten-halogenMax. data acquisition rate 100�HzChannels� �Diodes� 512Noise*� ±5�µAUDrift*� 500�µAU/h
*�according�to�ASTM
15
AS-1 Autosampler
Ultra high pressure up�to�1000�bar�(15000�psi)injectionSample capacity� �max.�7���positions�with��
2�x�3�4-well�plates�or�����standard�1.��ml�vials
Injection cycle time� �15�s,�� � <��0�s�with�needle�wash��Injection modes� �full-loop,�partial-loop�and�
“microliter-pickup”Precision� <�0.3�%�RSD�full-loop�� � injectionCarryover� ��<�0.05�%�with�needle�washSample tray cooling � 4–22�°C
Tech
nica
l�dat
a�ar
e�su
bjec
t�to
�cha
nge�
with
out�
notic
e.
Pump P-1 with SmartMix mixer and pump head
Rear viewTubing guide
Support
www.knauer.net
HPLC · SMB · Osmometry
Wissenschaftliche�GerätebauDr.�Ing.�Herbert�Knauer�GmbHHegauer�Weg�3�141�3�Berlin,�Germany
Phone:�� +4�-(0)30-�0�727-0Telefax:� +4�-(0)30-�015010E-Mail:� [email protected]:�www.knauer.net
Fast and reliable
Excellent support starts at the product development stage. That’s why we developed PLATINblue instruments with ultimate reliability in mind. Each component was designed for maximum dependabi-lity and durability, using only the best materials available.
To keep all of the components in your PLATINblue system running with the highest performance, we offer a comprehensive support and maintenance program to ensure your system’s reliable operation.
All support requests will be answered within�two�hours.�Guaranteed!�If�a�problem�arises,�your�personal�PLATINblue�contact�person�will�contact�you�within�24�hours�with�a�detailed�plan�for�resolution�of�the�problem.�KNAUER�management�is�also�at�your�disposal�and�can�be�contacted�by�telephone�when�necessary.�If�so�required,�we�will�solve�the�problem�at�your�site�–�quickly�and�without�complication.�
Our warranty for�PLATINblue�products�has�been�extended�to�three�years.�In�conjunction�with�a�service�agreement,�you�are�entitled�to�an�all-inclusive�full�service�plan.�The�reliable�and�continuous�operation�of�your�PLATINblue�system�is�our�highest�priority.
So you can get off running right from the start,�we�combine�the�PLATINblue�system�installation�with�training�for�you�and�your�team.�A�follow-up�training�session�within�the�following�six�months�is�included�in�the�service�package�at�no�extra�cost.
We will accompany you when�working�out�your�methods,�step�by�step.�Our�application�experts�are�available�Monday�through�Friday�to�assist�you�with�any�questions�you�might�have.
3 Year
24 hService
Con
sult
ing
Installation & Training
Warranty3 Year
24 hService
Con
sult
ing
Installation & Training
Warranty
3 Year
24 hService
Con
sult
ing
Installation & Training
Warranty
3 Year
24 hService
Con
sult
ing
Installation & Training
Warranty
©�KNAUER�April�200�