http:// quality control in radiation therapy, a new concept: dosimetry check wendel dean renner math...

45
http://www.MathResolution s.com 1 Quality Control in Radiation Therapy, A New Concept: Dosimetry Check Wendel Dean Renner Math Resolutions, LLC FDA Cleared April 27, 2001 US Patent 6,853,702

Upload: dwight-mathews

Post on 26-Dec-2015

214 views

Category:

Documents


0 download

TRANSCRIPT

http://www.MathResolutions.com

1

Quality Control in

Radiation Therapy,A New Concept: Dosimetry Check

Wendel Dean RennerMath Resolutions,

LLCFDA Cleared April 27, 2001 US Patent 6,853,702

http://www.MathResolutions.com

2

Radiation Therapy Needs Feedback

• The quality control procedure should tell you if there is a mistake.

• Diligence in hunting for a mistake is a weak quality control system.

• Without feedback, how do you know there has been no mistake?

http://www.MathResolutions.com

3

The New Method:Dosimetry

Check

• Take a calibrated picture of each treatment field.

• Download the treatment plan (Dicom RT or RTOG format).

• Recompute the dose using individual measured field fluence of each beam.

• Compare the dose to the plan dose.

http://www.MathResolutions.com

4

Use an Available Electronic Device to Take a Calibrated Picture of Each

Field as Used for Treatment

• Use the Amorphous Silicon Flat-panel Electronic Portal Imaging Device (EPID) on the Varian, Elekta, or Siemens Accelerator in Integration Mode.

• Use the MapCheck diode array.• Use an ion chamber array such as the

PTW729 or MatriXX.

http://www.MathResolutions.com

5

Use X-ray Film to Take a Calibrated Picture of Each Field as Used for

Treatment•Film at 100cm on couch or in blocking tray.

•Use 1 to 3 mm of copper on top of x-ray film.

•Can use step wedge or dosimeter to calibrate.

•Can rescale the curve.

•The beam should be set up as for treatment.

•1 percent accuracy is achievable for confidence.

http://www.MathResolutions.com

6

Calibrate to “Relative” Monitor Units

A Unit for Fluence

• 100 mu for a 10x10 field is 100 relative monitor units.

• 100 mu for a 40x40 field with scatter collimator factor of 1.05 is 105 relative monitor units.

• 100 mu for a 20x20 field with a wedge factor of 0.5 and scatter collimator factor of 1.02 is 51.0 relative monitor units.

RMU is an intensity fluence of the x-ray radiation.

Examples of relative monitor units:

http://www.MathResolutions.com

7

Beam 4 of Seven Field Head and Neck Case to be Shown

http://www.MathResolutions.com

8

Compute the Dose

• Download the plan in Dicom RT or RTOG format. Need CT scans, beam positions, 3D dose matrix, outlined regions of interest.

• Otherwise need the CT scan files and must position the beams to the same isocenter and angles.

• Must associate each field with the measured field picture.

http://www.MathResolutions.com

9

Dose Comparison Tools

Show plan dose and recomputed dose together.

http://www.MathResolutions.com

10

Transverse Plane

http://www.MathResolutions.com

11

Dose Comparison Tools: Gamma Method

3% - 3 mm criteria. Red tinted area is >= criteria.

http://www.MathResolutions.com

12

Gamma Method: Sagittal Plane

3% - 3 mm criteria. But note all outside of target.

http://www.MathResolutions.com

13

Dose Comparison Tools: Compare profile through patient.

http://www.MathResolutions.com

14

Gamma Volume Histogram

92.7% of the body volume at 20% of the dose or more has a gamma value <= 1.0

http://www.MathResolutions.com

15

3D Gamma Surface: 3%-3mm

http://www.MathResolutions.com

16

Dose Comparison Tools

Show dose difference between plan and recomputed plan, here 196 cG out of 6543 (3%) is tinted red. But we are not

considering distance.

http://www.MathResolutions.com

17

Dose Comparison Tools

•Show dose difference in 3D perspective room view.

•Red region is an overdose difference.

•Cyan shows under dose difference.

http://www.MathResolutions.com

18

Dose Difference Volume Histogram

For the body volume selected here.

http://www.MathResolutions.com

19

What Does It Take?

• Fifteen minutes (or treatment time) to do a dry run without the patient.

• Convert EPID images to fluence, a few minutes.

• Download plan, a few minutes.• Run our program “Dosimetry Check,” print

dose distribution, compare dose. Probably 30 to 45 minutes.

• Someone needs to look at the results and approve the treatment.

http://www.MathResolutions.com

20

What Do You Get?

• Dose shown on patient anatomy in 2D and 3D.

• Dose computed dependent upon the measured field fluence.

• A check on everything in the beam.• Confidence that the dosimetry and

delivery is correct.

http://www.MathResolutions.com

21

What You Don’t Get:

• Does not say anything about whether the fields are on the patient correctly.

• Probably insensitive to using the wrong energy.

http://www.MathResolutions.com

22

Why?

• Above all else, do no harm.

http://www.MathResolutions.com

23

Two Kinds of Errors:

• Systematic errors (in your procedures, data, etc.).

• Random errors (in executing the procedures).

http://www.MathResolutions.com

24

Systematic Errors

• Can be found by testing your procedures.

• Irradiating a phantom and measuring the dose constitutes a feedback system.

• Without this feedback, you could miss a problem.

http://www.MathResolutions.com

25

Random Errors

• Without feedback you will never know for sure that the treatment was correct for each patient.

http://www.MathResolutions.com

26

Overdose

• The patient reaction will tell you too late!

http://www.MathResolutions.com

27

Under Dose

• Might never be noticed.• Results in lost opportunity to

effectively treat the cancer.

http://www.MathResolutions.com

28

Do we really knowthe extent of mistakes that occur in radiation

therapy?

http://www.MathResolutions.com

29

Effective Quality Assurance

• We want to maximize feedback.• We want to maximize the amount of

information that is measured in the treatment room.

• We want to minimize the amount of information we use that is NOT measured in the room.

http://www.MathResolutions.com

30

Monitor Unit Check

• Redundancy is good.• But there is no feedback.

http://www.MathResolutions.com

31

Present Methods with Feedback

• Diode surface measurement at one point on each field to be compared to computed dose.

• Irradiate a cylindrical or square phantom using the plan, recompute the plan to same phantom to compare.

http://www.MathResolutions.com

32

Diode Surface Measurement

• This is a simple QA procedure, yet a surprising number of centers do not do it. But this only checks the dose at one point.

• Wedge could still be in wrong direction.• Compensator or IMRT could be wrong.• Blocks and/or field shape could be wrong

(should also pick up from film review, but some redundancy would be good).

• Dose distribution could still be wrong.

http://www.MathResolutions.com

33

Phantom

• Quality controls how the beams add up.• Maximizes the information measured in

room.• Can only show dose where measured in

the phantom.• Comparison of dose is somewhat abstract

in relation to the patient.• Assumption that dose in patient will be

right might not hold.• Does not tell you if beams are on the

patient correctly.

http://www.MathResolutions.com

34

Our new method:Dosimetry Check

• Shows the dose on the patient anatomy. • Evaluation of dose is not abstract.• Can handle arc therapy (but we presently

don’t have a device to measure a field that changes shape during gantry rotation).

• Can show dose in any plane or 3D dose cloud.• Same limitation of only checking dosimetry.• Easy and fast to do.

http://www.MathResolutions.com

35

Dosimetry Check Adds a Powerful Tool for Quality Control

• May replace some less effective procedures.

• May compliment others.

http://www.MathResolutions.com

36

Math Resolutions, LLC, is offering software to accomplish the new

method.

• FDA 510(k) K010225.• Software, manuals,and information

available at our website: http://MathResolutions.com.

• U.S. Patent 6,853,702 on this process.

http://www.MathResolutions.com

37

Measuring the field with X-Ray

film

• One could use a device instead of film.• Optimally the field should be measured

during treatment. That would require a device that the patient could be treated through.

• Film must be digitized.

http://www.MathResolutions.com

38

Use 1 to 3 mm Copper for X-ray Film

• We have found that using 1 to 3 mm of copper to shield for contamination electrons gives accurate results. 1 mm for 6 MV, 3 for 18 MV

• Testing is done by using the measured fluence to compute dose profiles at depth that can be compared to measured profiles.

http://www.MathResolutions.com

39

Film to Use

• Use V film up to 125 mu.• Use EDR2 film if over 125 mu.• Consider the effective mu for IMRT.• Limits depend upon film digitizer.• Range depends upon the distance you

use.

http://www.MathResolutions.com

40

Kodak EDR2 Film in Ready Pack

Curve crosses the density of 3.0 at 650 monitor units at 100cm.

http://www.MathResolutions.com

41

Kodak Therapy V Film

Film crosses density of 3.00 at 200 mu at 100 cm

http://www.MathResolutions.com

42

Calibration Procedures for X-Ray Film

• Include ion chamber or diode in bolus stack.• Or use a step wedge.• Run curve to calibrate the step wedge once.• Thereafter use the step wedge with each

case.• Or consider rescaling the calibration curve

from a single test exposure.• The goal is accuracy to 1 percent for

measuring relative mu

http://www.MathResolutions.com

43

Example IMRT Casecomputed from films

http://www.MathResolutions.com

44

Sagittal, IMRT

http://www.MathResolutions.com

45

Coronal, IMRT