i.1 _ hvm2003e

Upload: cunconfuny

Post on 16-Oct-2015

12 views

Category:

Documents


0 download

DESCRIPTION

I.1 _ HVM2003e Documents

TRANSCRIPT

  • Technical specifications subject to change, HVM2003e.doc, 10/12 Page 1/2

    HILO-TEST GmbH, Am Hasenbiel 42, D-76297 Stutensee-Karlsruhe, Tel. 07244/20500-0, www.hilo-test.de

    High-voltage measurement equipment

    HVM 2001, HVM 2003

    The high-voltage measurement instruments HVM 2001 / HVM 2003 consist of a measuring amplifier with digital display for universal use, a built-in high-voltage divider for direct measurement up to 10 kV / optional 20 KV and an external input for an external high-voltage divider, HVT***RCR, for measurement and display of voltages up to 200 kV.. The high-voltage measurement system HVM 2001 is designed for measuring ac- and dc- voltages. The high-voltage measurement system HVM 2003 allows in addition measurement of the peak value of lightning and switching pulse voltages. Optionally the unit can be delivered with a DKD calibration certificate. Technical specification: HVM 2001 HVM 2003

    Direct high-voltage-input 10 kV=/ 7kV 10 kV=/ 7kV

    Direct high-voltage-input, optional 20 kV=/14kV 20 kV=/14kV

    External input, Rs = 1000 k 100V100% 100V100% Calibration possibility of the preamplifier 2 % 2 %

    Analogue output, Rs = 5 k 5V100% 5V100%

    1mA100% 1mA100% Weighting of input voltage: DC Average Average

    AC + Peak/2 + Peak/2

    AC - Peak/2 - Peak/2 Pulse, LI & SI + Peak Pulse, LI & SI - Peak Accuracy, AC/DC 1 % 5 Digit 1 % 5 Digit Accuracy, Pulse 2.5 % 5 Digit Bandwidth 0 - 10 kHz -1000 kHz Measuring input internal/external, switch able, Range: 19.99 / 199.9 BCD-output in multiplex mode, optically isolated 15-way-D-connector Mains power supply 230 V/50 Hz/10 VA Dimensions 485*90*300 mm Weight 3 kg

  • Technical specifications subject to change, HVM2003e.doc, 10/12 Page 2/2

    HILO-TEST GmbH, Am Hasenbiel 42, D-76297 Stutensee-Karlsruhe, Tel. 07244/20500-0, www.hilo-test.de

    optional with:

    Broad-band High-voltage Dividers for HVM 2001, HVM 2003 HVT 120 RCR, HVT 240 RCR The Broadband high-voltage divider HVT 120 RCR / HVT 240 RCR are state-of-the-art measuring equipment with excellent high-frequency transmission characteristics. The high-voltage impedance consists of a precise metal film resistor and a high-voltage capacitor with series damping resistor in parallel. The measuring cable with the termination network is an essential part of the divider. The

    termination network can be adjusted for an input impedance of 1 M // 10 - 30 pF of the unit connected. Technical specification HVT 120 RCR HVT 240 RCR max. Input voltage : DC voltage 120 kV= 240 kV=

    AC voltage, 50 Hz 90 kV 180 kV Pulse voltage, 1.2/50s 200 kV 360 kV

    Dielectric Strength: AC voltage 50 Hz, 0.3 sec 120 kV 250 kV Pulse voltage, 1.2/50s 300 kV 600 kVs Test voltage: 10 pulses 1.2/50 s, pos./neg, each 250 kV 500 kV

    High voltage impedance 270 M 360 M

    //100 pF+2k //180pF+3.6k

    Nominal load of the low voltage arm of the divider 1.0 M, 10-30 pF 1.0 M, 10-30 pF Measuring cable with termination network 10 m 20 m long Divider ratio 2000 : 1 1 % (or customer specific) Short time stability 0.1% Proximity effect, for wall distance > 2 * height of divider 0.2%

    Temperature drift in the range of 5 - 40C 100 ppm/K Ambient temperature for normal operation 17 C - 23 C Rise time 35 ns 20% 50 ns 20% Over shoot < 5% < 5% Dimensions: Height 1.2 m 2.15 m Socket 360 mm 1200X1200 mm 4 wheels, 2 looka. weight ca. 25 kg ca. 40 kg The dividers can also be used for high-voltage measurement of ac-, dc- and pulse voltages with an oscilloscope.