icd introduction 2004. 去顫器之適應症 – 中央健保局...
Post on 20-Dec-2015
327 views
TRANSCRIPT
ICD vs. Drugs: Life Expectancy
Larsen GC, Pauker SG, et al. J Am Coll Cardiol. 1992;19(6):1323-1334.
0
1
2
3
4
5
6
7
ConventionalAntiarrhythmic
Therapy
ICD Therapy
LifeExpectancy
Estimate(No. Years))
Amiodarone
2.5
3.9
6.1
Lead Measurements
Parameter Acute Lead System Chronic Lead System
Ventricular capture threshold
< 1V < 3V
Pacing impedance 200-1000 Ohms 200-1000 Ohms
Defibrillation impedance
< 80 Ohms < 80 Ohms
Filter R-wave amplitude
> 5mV (Sinus rhythm) > 3mV (sinus rhythm)
Slew rate > 0.75V/sec 0.45V/sec
ICD Generator
Audiotranducer
Microcircuitry
Magnet switch
Telemetry
High voltageoutput Sense
amplifier
Pace output
ms
Battery
HV capacitor
HVO circuit
3-6 V 10-mAsec
750V 30-A
Biphasic Polarity
ReversedPolarity
Initial Polarity
Tilt =65%
Tilt =65%
Vi=
750V
V2=92V
V1=263V
D1=7 ms
Longivity of ICD
• 5 years background monitoring
• 5 years pacing (single chamber, 100% pacing)
• 200 shocks
Life of ICD Battery
BOL (3.2-3.3 V)
ERI (2.5V) EOL (2.3-2.4 V)
Cel
l Vol
tage
(V
)
Time (months)
BOL: beginning of lifeERI: elective replacement indicatorEOL: end of life
ICD Lead
silicone
polyurethane
P/S electrodes
RV defibrillation electrode
RA/SVC defibrillation electrode
Silver
MP35NPl-Ir
Operation of ICD System
Monitor RhythmPost-shock Bradycardia
Cardioversion
AntiTachycardia Pacing DefibrillationTachycardia
BradycardiaAntibradycardia
Pacing
Burst / Ramp
Anti-Tahcycardia Pacing
Redetection
Attempt 1 with 3 pulses Attempt 2 with 4 pulses
300ms
300ms
300ms
300ms
300ms
Burst
300ms
292ms
284ms
Ramp Burst
Ramp
Detection Algorithm
Rate
Duration
Redetection Duration
Post-Shock Duration
90-200 bpm
1-60 sec
1-15 sec
1-60 sec
110-220 bpm
1-30 sec
1-15 sec
1-30 sec
PARAMETER VT-1 ZONE VT ZONE VF ZONE
130-250 bpm
1-15 sec
1 sec
1 sec
Tachycardia Detection
• Rate detection zones– Automatic gain / threshold control
• Detection enhamcements / Tachycardia discrimination– Sudden onset– Interval stability– Dual chamber detection– Automatic morphology analysis
Automatic Gain Control
0.43 mV
3.44 mV
“Fast AGC” exponential decay within a dynamic range, using sinus template
Maximum sensitivity for dynamic range
“Slow AGC” using one of 12 dynamic ranges
8:1 (6:1 in atrium)
Separately programmable maximum sensitivity for device
0.27 mV
Detection Enhancement - OnsetVF
VT
VF
VT
Analysis Analysis
1 2
1. Gradual transition, Onset not met2. Duration ends, therapy inhibited
1 2
1. Abrupt transition, Onset met2. Duration ends, therapy initiated
Pivot interval
600600
Onset = 175 ms
Stage 2 is met4 of 4 comparisons (intervals to baseline average)are > Onset threshold.
Largest CL decrease is > Onset threshold
Stage 1 is met
572 456 785 410400 405 412
VT VT VT VTVS VS VTVS VS VS VT VTVT
Detection met
Baseline Avg.
Detection Enhancement - Stability
VT Detect//
//
Stability = 10 ms
7 8 9 10 9 9 10 88 8 9 9
Duration metStable (9 < 10)Initiate ATP therapy
DurationStability Analysis
Variance (ms)
Weighted average
Detection Enhancement – A & V Rate
Duration met, gradual
Ventricle
Atrium
Detection metStart Duration
V Rate > A Rate170 bpm > 90 bpm
A Rate
V Rate
Initiate therapy
Concurrent Processing
//
//
P.S. Stability
P.S. Sustained Rate DurationDuration
V Rate < A Rate
Stability
A > AFib Rate Threshold
Tachy episode
declared (8/10)
P.S. AFib Rate Threshold
P.S. Duration
TREAT
Sudden
Unstable
AFib Rate Threshold False
V Rate > A Rate True
Onset
P.S. V Rate > A Rate
PR Logic in GEM® DR inappropriately detected for Sinus Tachycardia due to PR interval > 50% of RR interval.
PR Logic withheld using the Sinus Tach criteria for 17 beats before eventually delivering inappropriate VT therapy.
PR Logic™ (before enhancement)In Patient with Long PR interval
GEM® III’s new programmable 1:1 VT-ST boundary at nominal setting will perform identically to GEM DR in Sinus Tach episodes with PR interval > 50% of RR interval.
PR Logic at nominal settings inappropriately detected for same Sinus Tachycardia episode with Long PR interval.
VT-ST boundary 50%
Nominal PR Logic™ (VT-ST boundary at 50%)In Patient with Long PR interval
GEM® III DR appropriately withholds therapy for Sinus Tachycardia with a Long PR interval when 1:1 VT-ST boundary is programmed to 66%.
Enhanced PR Logic appropriately withholds detection for patients with a long PR interval.
VT-ST boundary 66%
Enhanced PR Logic™ (VT-ST boundary at 66%) In Patient with Long PR interval
Enhanced PR Logic™ PerformanceTheoretical Performance with Programmable VT-ST
boundary**
*18 beat self-terminating run: double tach in VT zone & regularity unmet
Positive PredictiveValue:95.2%
True Positive:711 episodes
False Positive:36 episodes
False Negative:1 episode*
(non-sustained)
True Negative:218 episodes
VT/VF SVTActual Rhythm
SV
TV
T/V
F
100% sensitivity (sustained
VT/VF)
85.8%incremental specificity
Det
ecte
d R
hyth
m
1919
**By programming the VT-ST boundary in GEM® III, 19 of the 22 inappropriate therapies would now be correctly withheld (1 patient’s P-wave fell in Junctional zone and would not be reclassified)