Identification and Quantification of Pesticides in Fruit and Vegetables with Accurate Mass Database using Gas Chromatography Coupled to High Resolution Orbitrap Mass Spectrometry Samanta Uclés, Víctor Cutillas , Amadeo R. Fernández-Alba Agrifood Campus of International Excellence (ceiA3), European Union Reference Laboratory for Pesticide Residues in Fruits and Vegetables. University of Almería (Spain); e-mail: [email protected] ABSTRACT To protect and prevent damages in fruit and vegetables crops, the addition of some pesticides may be needed, thus pesticide residues can be present in the final product. For this reason, to assess the food safety for consumers in fruit and vegetables is an important issue in laboratories. Typically food samples are analyzed by using high performance or ultra high performance liquid chromatography (HPLC or UHPLC) and/or gas chromatography (GC) coupled to mass spectrometry (MS) to identify and quantify target compounds. The most common MS systems in routine analysis are based on triple quadrupole (QqQ) analyzers. But, even reaching low detection limits, these techniques have some limitations consequence of their unit mass resolution. That limitation can drive to analytical difficulties consequence of interferences from the matrix. That can represent an incorrect identification and quantification. This limitation can be overcome by applying high resolution accurate mass spectrometers (HRAMS) instruments such as Orbitrap-MS. The present work is focused in the evaluation of GC-Orbitrap-MS, which has a high resolving power (working routinely at 60000 resolution FWHM at m/z 200, but able to achieve up to 120000) with mass accuracies lower than 5 ppm, according to European Union analytical quality control (EU AQC) procedures, but they were usually < 1 ppm . This evaluation was done by analyzing different representative commodities of fruits and vegetables spiked with 50 different pesticides. The GC-Orbitrap-MS obtained very low limits of identification, below 5 µg/kg in all cases, good linearity (residuals < 20 % for every calibration level and R 2 > 0.99) and extremely good mass accuracy (< 1 ppm for around 90 %) for almost all combinations compound/matrix. Repeatability also met criteria described in EU AQC procedures (RSD < 20 %). THE AUTHORS ACKNOWLEDGE FUNDING SUPPORT FROM THE EUROPEAN COMMISSION, DG SANTE, SPECIFIC AGREEMENT No 7 OF FRAMEWORK PARTHERSHIP AGREEMENT No SANCO/2005/FOOD SAFETY/0025 PESTICIDES IN FRUIT AND VEGETABLES AND DOMINIC ROBERTS, FROM THERMO FISHER SCIENTIFIC, FOR HIS TECHNICAL SUPPORT Conclusions This first evaluation shows the identification power for pesticide residues in fruit and vegetables of this HRAMS platform, being able to evaluate all target selected compounds at 10 µg/kg, in agreement with the EU AQC procedures. Due to the wide linear range is possible to inject real samples with high concentration levels for pesticide residues with no saturation phenomenon, and so on to adequately quantify them. Mass accuracies are < 5 ppm (requirement in EU AQC procedures) and keep constant along chromatographic peak. This work is an starting point to increase the number of pesticides included in the database for being applied in a higher number of commodities. Experimental Centrifuge 5 minutes at 3700 rpm Extraction Method: Citrate QuEChERS 10 g of Sample + 10 ml of AcN Analysis 4 g MgSO 4 + 1 g NaCl + 1 g NaCit·2H 2 O +0.5 g NaCit·1.5H 2 O Centrifuge 5 minutes at 3700 rpm Shake 4 min automatically Take 5 mL + 750 mg MgSO4 + 125 mg PSA Shake 4 min automatically Add 40 µL Formic Acid (5 % in AcN) GC-MS Conditions GC Parameters: Injection volume: 1 µL Splitless mode Column: TG-5SILMS with 5 m Safeward (30 m x 0.25 mm x 0.25 µm) Carrier gas: Helium Constant flow: 1.2 mL/min Total run time: 35.6 min Oven temperature program: MS Parameters: Full Scan Positive mode Electron Impact Resolution: 60000 FWHM AGC: 1·10 6 Scan Range: 50-550 m/z System: TRACE 1300 Series GC coupled to a QExactive GC Orbitrap MS (Thermo) 0 50 100 150 200 250 300 350 0 10 20 30 40 T (ºC) T (min) The pesticide database employed for this study has been generated by two approaches: 1) obtaining the ions from high resolution NIST library, when it is present, or 2) by injecting standard solutions in pure solvent at a high concentration level (500 µg/kg), observing their generated experimental ions and calculating their theoretical mass ions with the help of the molecular formula. TIC for Standard Mix in Solvent at 500 µg/kg Extracted Ion Chromatogram for Acrinathrin (known ions with nominal mass) at 500 µg/kg m/z 209 m/z 153 m/z 181 Acrinathrin C26H21F6NO5 O O O O O CF 3 CF 3 N With Elemental Composition tool, present in Xcalibur Qual Browser software, introduce the exact mass observed in the spectrum and the maximum number of atoms for the molecule for the elemental composition calculation. Then we can estimate the theoretical ion to be included in the database Pesticide Database Substracted espectrum (24.26 min) Pesticide Database Pesticide List Results 0 10 20 30 40 50 60 70 80 90 100 1 2 5 10 20 50 100 200 500 Percentage of Compounds (%) C (µg/kg) Compounds Identified Tomato Orange Leek 0 10 20 30 40 50 60 70 80 90 < 20 % 20-50 % > 50 % Percentage of Compounds (%) Matrix Effect Orange Leek A total of 50 different pesticides were selected and spiked in 3 matrices (orange, tomato and leek) at different concentration levels. Identification criteria were: the presence of at least 2 ions, with the correct retention time and mass accuracy < 5 ppm. Applying the database built for these compounds, all pesticides were identified to SANTE requirements at 10 µg/kg, except for bromopropylate in leek (20 µg/kg) and chlorothalonil in leek. The limit of detection for the main quantifying ion was much lower. Results provided good linearity with residuals < 20 % for each calibration level and R 2 values > 0.99. This system does not saturate for concentrations/conditions tested. Comparing obtained slopes to tomato matrix, orange presented around 75 % of pesticides with no matrix effect (ME < |20| %). However, leek showed the highest percentage of compounds in the strong matrix effect range (ME > |50| %). Repeatability was also tested by injecting 5 replicates at 10 µg/kg concentration level and expressed as RSD (%). As average, the result was around 90 % of compounds with RSDs < 5 %. Vinclozolin Standard in Tomato at 10 µg/kg (Ion Overlap) m/z 212.0028 m/z 178.0418 m/z 197.9872 m/z 284.9954 y = 17515x - 1E+06 R² = 0,997 y = 12328x - 58829 R² = 0,999 y = 19437x + 48815 R² = 0,998 0,0E+00 2,0E+07 4,0E+07 6,0E+07 8,0E+07 1,0E+08 1,2E+08 0 100 200 300 400 500 600 Area C (µg/kg) Linearity (Vinclozolin) 0 10 20 30 40 50 60 70 80 90 100 Tomato Orange Leek Percentage of Compounds (%) RSDs, 10 µg/kg (n = 5) < 5 % 5-10 % > 10 % (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.47 ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.47 ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.00ppm) (0.00 ppm) (0.00 ppm) (0.00 ppm) (0.47 ppm) (0.00 ppm) (0.00 ppm) (0.47 ppm) Base Peak for Vinclozolin (212.0028) Standard in Tomato at 10 µg/kg 0 10 20 30 40 50 60 70 80 > 2 Percetange of Compounds (%) Mass Accuracy (ppm) Mass Accuracy at 10 µg/kg Tomato Orange Leek 0.5 0.5 < x 1 1 < x 1.5 1.5 < x 2 Full Scan Spectrum for Vinclozolin Standard in Tomato at 10 µg/kg (14.005 min) Mass accuracy for all pesticides/matrices was also evaluated. As can be seen for 90-95 % of the pesticides mass accuracies were 1 ppm. The software employed for data treatment was TraceFinder, which provides mass accuracies for each selected ion in the database in the apex of the peak. From base peak chromatogram for HCB (point to point) we can observe mass accuracy is rather constant along all the chromatographic peak. Full Scan Spectrum for Vinclozolin Standard in Library (NIST) Mass spectrum obtained for each pesticide can be also compared with a mass spectrum registered in a library, as NIST, for example. # Rt (min) Compound 27 14.90 Fenpropidin 28 15.36 Fenpropimorph (NIST) 29 16.34 Fipronil (NIST) 30 12.03 HCB 31 11.92 HCH, alpha- (NIST) 32 12.46 HCH, beta- (NIST) 33 12.62 HCH, gamma- (lindane) (NIST) 34 22.12 Iprodione 35 18.42 Kresoxim-methyl (NIST) 36 14.27 Metalaxyl (NIST) 37 18.33 Myclobutanil (NIST) 38 19.47 Oxadixyl (NIST) 39 14.09 Parathion-methyl (NIST) 40 16.09 Pendimethalin (NIST) 41 10.40 Phenylphenol, 2- (NIST) 42 13.33 Pirimicarb (NIST) 43 16.69 Procymidone (NIST) 44 12.41 Propazine (NIST) 45 12.93 Pyrimethanil (NIST) 46 12.64 Terbutylazine (NIST) 47 22.35 Tetramethrin (NIST) 48 14.11 Tolclofos-methyl (NIST) 49 21.35 TPP (NIST) 50 11.37 Trifluralin (NIST) 51 14.01 Vinclozolin (NIST) # Rt (min) Compound 1 24.70 Acrinathrin 2 30.69 Azoxystrobin (NIST) 3 22.43 Bifenthrin (NIST) 4 9.26 Biphenyl 5 22.55 Bromopropylate (NIST) 6 18.39 Bupirimate (NIST) 7 19.31 Chlorobenzilate (NIST) 8 12.99 Chlorothalonil (NIST) 9 11.40 Chlorpropham (NIST) 10 15.16 Chlorpyrifos (NIST) 11 13.93 Chlorpyrifos-methyl (NIST) 12 24.30 Cyhalothrin, lambda- (NIST) 13 27.64 Cypermethrin (NIST) 14 19.51 DDD pp'- (NIST) 15 18.16 DDE pp'- 16 19.58 DDT op'- (NIST) 17 20.75 DDT pp'- (NIST) 18 12.73 Diazinon (NIST) 19 8.23 Dichlorvos (NIST) 20 18.33 Dieldrin (NIST) 21 17.50 Endosulfan, alpha- (NIST) 22 19.32 Endosulfan, beta- (NIST) 23 20.59 Endosulfan, sulfate- (NIST) 24 29.50 Esfenvalerate (SS, RR) (NIST) 25 28.03 Etofenprox (NIST) 26 14.73 Fenitrothion (NIST) m/z Experimental Elemental Composition m/z Theoretical 181.0648 C13H9O 181.0648 209.0835 C14H11ON 209.0835 153.0699 C12H9 153.0699 114.0339 C8H4N 114.0338

Upload: ngothu

Post on 08-Sep-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Identification and Quantification of Pesticides in Fruit ... · This evaluation was done by analyzing different representative commodities of fruits and vegetables spiked with 50

Identification and Quantification of Pesticides in Fruit and Vegetables with Accurate Mass Database using Gas Chromatography Coupled to

High Resolution Orbitrap Mass Spectrometry Samanta Uclés, Víctor Cutillas, Amadeo R. Fernández-Alba

Agrifood Campus of International Excellence (ceiA3), European Union Reference Laboratory for Pesticide Residues in Fruits and Vegetables. University of Almería (Spain); e-mail: [email protected]

ABSTRACTTo protect and prevent damages in fruit and vegetables crops, the addition of some pesticides may be needed, thus pesticide residues can be present in the final product. For this reason, to assess the food safety for consumers in fruit and vegetables is an important issue in laboratories.

Typically food samples are analyzed by using high performance or ultra high performance liquid chromatography (HPLC or UHPLC) and/or gas chromatography (GC) coupled to mass spectrometry (MS) to identify and quantify target compounds. The most common MS systems in routine analysis are based on triple quadrupole (QqQ) analyzers. But, even reaching low detection limits, these techniques have some limitations consequence of their unit mass resolution. That limitation can drive to analytical difficulties consequence of interferences from the matrix. That can represent an incorrect identification and quantification. This limitation can be overcome by applying high resolution accurate mass spectrometers (HRAMS) instruments such as Orbitrap-MS.

The present work is focused in the evaluation of GC-Orbitrap-MS, which has a high resolving power (working routinely at 60000 resolution FWHM at m/z 200, but able to achieve up to 120000) with mass accuracies lower than 5 ppm, according to European Union analytical quality control (EU AQC) procedures, but they were usually < 1 ppm . This evaluation was done by analyzing different representative commodities of fruits and vegetables spiked with 50 different pesticides.

The GC-Orbitrap-MS obtained very low limits of identification, below 5 µg/kg in all cases, good linearity (residuals < 20 % for every calibration level and R2 > 0.99) and extremely good mass accuracy (< 1 ppm for around 90 %) for almost all combinations compound/matrix. Repeatability also met criteria described in EU AQC procedures (RSD < 20 %).

THE AUTHORS ACKNOWLEDGE FUNDING SUPPORT FROM THE EUROPEAN COMMISSION, DG SANTE, SPECIFIC AGREEMENT No 7 OF FRAMEWORK PARTHERSHIP AGREEMENT No SANCO/2005/FOOD SAFETY/0025 PESTICIDES IN FRUIT AND VEGETABLES

AND DOMINIC ROBERTS, FROM THERMO FISHER SCIENTIFIC, FOR HIS TECHNICAL SUPPORT

Conclusions This first evaluation shows the identification power for pesticide residues in fruit and vegetables of this HRAMS platform, being able to evaluate all target selected compounds at 10 µg/kg, in agreement with the EU AQC procedures.

Due to the wide linear range is possible to inject real samples with high concentration levels for pesticide residues with no saturation phenomenon, and so on to adequately quantify them.

Mass accuracies are < 5 ppm (requirement in EU AQC procedures) and keep constant along chromatographic peak.

This work is an starting point to increase the number of pesticides included in the database for being applied in a higher number of commodities.

Experimental

Centrifuge 5 minutes at 3700 rpm

Extraction Method: Citrate QuEChERS

10 g of Sample + 10 ml of AcN

Analysis

4 g MgSO4 + 1 g NaCl +1 g NaCit·2H2O +0.5 g NaCit·1.5H2O

Centrifuge 5 minutes at 3700 rpm

Shake 4 min automatically

Take 5 mL + 750 mg MgSO4 + 125 mg PSA

Shake 4 min automatically

Add 40 µL FormicAcid (5 % in AcN)

GC-MS Conditions

GC Parameters: • Injection volume: 1 µL• Splitless mode• Column: TG-5SILMS with 5 m Safeward (30 m x 0.25 mm x 0.25 µm)• Carrier gas: Helium• Constant flow: 1.2 mL/min• Total run time: 35.6 min• Oven temperature program:

MS Parameters: • Full Scan• Positive mode• Electron Impact• Resolution: 60000 FWHM• AGC: 1·106

• Scan Range: 50-550 m/z

System:TRACE 1300 Series GC coupled to a QExactive GC Orbitrap MS (Thermo)

0

50

100

150

200

250

300

350

0 10 20 30 40

T (º

C)

T (min)

The pesticide database employed for this study has been generated by two approaches: 1) obtaining the ions from high resolution NIST library, when it is present, or 2) by injecting standard solutions in pure solvent at a high concentration level (500 µg/kg), observing their generated experimental ions and calculating their theoretical mass ions with the help of the molecular formula.

TIC for Standard Mix in Solvent at 500 µg/kg

Extracted Ion Chromatogram for Acrinathrin(known ions with nominal mass) at 500 µg/kg

m/z 209

m/z 153

m/z 181

AcrinathrinC26H21F6NO5

OO

O

O

O

CF3

CF3

N

With Elemental Composition tool, present in XcaliburQual Browser software, introduce the exact mass observed in the spectrum and the maximum number of atoms for the molecule for the elemental composition calculation. Then we can estimate the theoretical ion to be included in the database

Pesticide Database

Substracted espectrum (24.26 min)

Pesticide Database

Pesticide List

Results

0102030405060708090

100

1 2 5 10 20 50 100 200 500

Perc

enta

ge o

f Com

poun

ds (%

)

C (µg/kg)

Compounds Identified

Tomato

Orange

Leek

0

10

20

30

40

50

60

70

80

90

< 20 % 20-50 % > 50 %

Perc

enta

ge o

f Com

poun

ds (%

)

Matrix Effect

Orange

Leek

A total of 50 different pesticides were selected and spiked in 3 matrices (orange, tomato and leek) at different concentration levels. Identification criteria were: the presence of at least 2 ions, with the correct retention time and mass accuracy < 5 ppm.

Applying the database built for these compounds, all pesticides were identified to SANTE requirements at 10 µg/kg, except for bromopropylatein leek (20 µg/kg) and chlorothalonil in leek. The limit of detection for the main quantifying ion was much lower.

Results provided good linearity with residuals < 20 % for each calibration level and R2 values > 0.99. This system does not saturate for concentrations/conditions tested.

Comparing obtained slopes to tomato matrix, orange presented around 75 % of pesticides with no matrix effect (ME < |20| %). However, leek showed the highest percentage of compounds in the strong matrix effect range (ME > |50| %).

Repeatability was also tested by injecting 5 replicates at 10 µg/kg concentration level and expressed as RSD (%). As average, the result was around 90 % of compounds with RSDs < 5 %.

Vinclozolin Standard in Tomato at 10 µg/kg(Ion Overlap)

m/z 212.0028m/z 178.0418m/z 197.9872m/z 284.9954

y = 17515x - 1E+06R² = 0,997

y = 12328x - 58829R² = 0,999

y = 19437x + 48815R² = 0,998

0,0E+00

2,0E+07

4,0E+07

6,0E+07

8,0E+07

1,0E+08

1,2E+08

0 100 200 300 400 500 600

Are

a

C (µg/kg)

Linearity (Vinclozolin)

0

10

20

30

40

50

60

70

80

90

100

Tomato Orange Leek

Perc

enta

ge o

f Com

poun

ds (%

)

RSDs, 10 µg/kg (n = 5)

< 5 %

5-10 %

> 10 %

(0.00 ppm)(0.00 ppm)

(0.00 ppm)

(0.00 ppm)

(0.47 ppm)

(0.00 ppm)

(0.00 ppm)

(0.00 ppm)

(0.00 ppm)

(0.47 ppm)

(0.00 ppm)

(0.00 ppm)

(0.00 ppm)

(0.00 ppm)

(0.00 ppm)(0.00ppm)

(0.00 ppm)

(0.00 ppm)

(0.00 ppm)

(0.47 ppm)(0.00 ppm)

(0.00 ppm)

(0.47 ppm)

Base Peak for Vinclozolin(212.0028) Standard in Tomato at 10 µg/kg

0

10

20

30

40

50

60

70

80

£ 0.5 0.5 < x £ 1 1 < x £ 1.5 1.5 < x £ 2 > 2Perc

etan

ge o

f Com

poun

ds (%

)

Mass Accuracy (ppm)

Mass Accuracy at 10 µg/kg

Tomato

Orange

Leek

≤ 0.5 0.5 < x ≤ 1 1 < x ≤ 1.5 1.5 < x ≤ 2

Full Scan Spectrum for Vinclozolin Standard in Tomato at 10 µg/kg (14.005 min)Mass accuracy for all pesticides/matrices was also

evaluated. As can be seen for 90-95 % of the pesticides mass accuracies were ≤ 1 ppm. The software employed for data treatment was TraceFinder, which provides mass accuracies for each selected ion in the database in the apex of the peak.

From base peak chromatogram for HCB (point to point) we can observe mass accuracy is rather constant along all the chromatographic peak.

Full Scan Spectrum for Vinclozolin Standard in Library (NIST)

Mass spectrum obtained for each pesticide can be also compared with a mass spectrum registered in a library, as NIST, for example.

# Rt (min) Compound27 14.90 Fenpropidin28 15.36 Fenpropimorph (NIST)29 16.34 Fipronil (NIST)30 12.03 HCB31 11.92 HCH, alpha- (NIST)32 12.46 HCH, beta- (NIST)33 12.62 HCH, gamma- (lindane) (NIST)34 22.12 Iprodione35 18.42 Kresoxim-methyl (NIST)36 14.27 Metalaxyl (NIST)37 18.33 Myclobutanil (NIST)38 19.47 Oxadixyl (NIST)39 14.09 Parathion-methyl (NIST)40 16.09 Pendimethalin (NIST)41 10.40 Phenylphenol, 2- (NIST)42 13.33 Pirimicarb (NIST)43 16.69 Procymidone (NIST)44 12.41 Propazine (NIST)45 12.93 Pyrimethanil (NIST)46 12.64 Terbutylazine (NIST)47 22.35 Tetramethrin (NIST)48 14.11 Tolclofos-methyl (NIST)49 21.35 TPP (NIST)50 11.37 Trifluralin (NIST)51 14.01 Vinclozolin (NIST)

# Rt (min) Compound1 24.70 Acrinathrin2 30.69 Azoxystrobin (NIST)3 22.43 Bifenthrin (NIST)4 9.26 Biphenyl5 22.55 Bromopropylate (NIST)6 18.39 Bupirimate (NIST)7 19.31 Chlorobenzilate (NIST)8 12.99 Chlorothalonil (NIST)9 11.40 Chlorpropham (NIST)

10 15.16 Chlorpyrifos (NIST)11 13.93 Chlorpyrifos-methyl (NIST)12 24.30 Cyhalothrin, lambda- (NIST)13 27.64 Cypermethrin (NIST)14 19.51 DDD pp'- (NIST)15 18.16 DDE pp'-16 19.58 DDT op'- (NIST)17 20.75 DDT pp'- (NIST)18 12.73 Diazinon (NIST)19 8.23 Dichlorvos (NIST)20 18.33 Dieldrin (NIST)21 17.50 Endosulfan, alpha- (NIST)22 19.32 Endosulfan, beta- (NIST)23 20.59 Endosulfan, sulfate- (NIST)24 29.50 Esfenvalerate (SS, RR) (NIST)25 28.03 Etofenprox (NIST)26 14.73 Fenitrothion (NIST)

m/zExperimental

Elemental Composition

m/zTheoretical

181.0648 C13H9O 181.0648209.0835 C14H11ON 209.0835153.0699 C12H9 153.0699114.0339 C8H4N 114.0338