in-silico bifurcation stenting: plaque composition ... · arterial wall thickness (tzucu 2001,...

18
10 th European Bifurcation Club meeting - Bordeaux, France - 17 th and 18 th October 2014 In-silico bifurcation stenting: plaque composition / bifurcation angle vs plaque shift J.J. Wentzel C. Chiastra, F. Iannaccone Erasmus MC, Rotterdam, The Netherlands Politecnico di Milano, Milano, Italy Gent University, Gent, Belgium

Upload: others

Post on 01-Aug-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

10th European Bifurcation Club meeting - Bordeaux, France - 17th and 18th October 2014

In-silico bifurcation stenting: plaque composition / bifurcation

angle vs plaque shift

J.J. Wentzel C. Chiastra, F. Iannaccone

Erasmus MC, Rotterdam, The Netherlands

Politecnico di Milano, Milano, Italy Gent University, Gent, Belgium

Page 2: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

2

Aim

3 bifurcation geometries with different distal angles* are investigated

* Distal angle = 57.3° ± 10.0° calculated on LAD, RCA, LCX (mainly LAD, 92.2%) (n = 153 patients) by Elsaban et al. 2013 Elsaban et al. J Invasive Cardiol 2013; 25:118-122

45 deg 57.5 deg 70 deg

To study the influence of distal angle / plaque composition on plaque shift / carina shift because of provisional stenting

2 different type of plaques:

-  fibrotic

-  lipid

Create model with plaque included

Novel bifurcation phantoms for validation of quantitative coronary angiography algorithms. Girasis C, … Serruys ,Wentzel JJ. CCI, 2011, 2011

Page 3: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Bifurcation phantom studies

Designed bifurcation phantom

3

Novel bifurcation phantoms for validation of quantitative coronary angiography algorithms. Girasis C, … Serruys ,Wentzel JJ. CCI, 2011, 2011 May 1;77(6):790-7.

1.  Diameter ratio : finet law 2.  Length lesion 3.  Location MLD 4.  Bifurcation angle

Page 4: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

4

Intima

Media

Adventitia Plaque

1 2 3 4 5 6 7 8 9 10 11 12 13 14

Model : LAD / D1 Bifurcation

Arterial wall thickness (Tzucu 2001, Holzapfel 2005):

intima + media thickness = 0.1 lumen Diameter (healthy)

intima thickness/wall thickness = 0.26

media thickness/total wall thickness = 0.35

adventitia thickness/wall thickness = 0.39

Plaque burden > 40% at location of lumen narrowing (Glagov 1987)

Page 5: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

1 2 3 4

5

6 7

8 9 10 11 12

13 14

* * * * * * * * * * * * * *

5

PB = Plaque burden (up to the external elastic membrane)

Intima

Media

Adventitia Plaque

1 PB = 35.4% 2 PB = 41.1% 3 PB = 85.9% 4 PB = 87.6% 5 PB = 90.8%

6 PB = 71.2% 7 PB = 82.4% 8 PB = 86.6% 9 PB = 88.8% 10 PB = 90.7%

11 PB = 91.2% 12 PB = 82.5% 13 PB = 66.4% 14 PB = 43.5%

* * * * *

* * * * *

* * * *

Model : LAD / D1 Bifurcation

Page 6: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

6

Computer simulations Mechanical properties of plaque components

Loree et al. J Biomech, 1994; 2: 1995-204 Holzapfel et al. Am J Physiol Heart Circ Physiol, 2005; 289: H2048-H2058

Isotropic hyperelastic behavior with ideal plasticity to mimic vessel

damage

(Lor

ee 1

994)

Arterial wall

Isotropic hyperelastic behavior with ideal plasticity to mimic plaque

rupture

Fibrotic / Lipid plaque

(Hol

zapf

el 2

005)

Page 7: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

7

Methods :Computer simulations Stent and balloon •  Multi-Link 8 (Abbott Laboratories, Abbott Park, IL, USA)

•  Balloon:

-  Bare-metal stent, Co-Cr alloy

-  Size: 3x18 mm

-  Modeled as a straight tube using Kiousis

approach*

-  Calibrated using the manufacturer

compliance chart from 10 atm to 14 atm

(nominal pressure = 10 atm, burst pressure = 18 atm)

Kiousis et al. Int J Num Methods Eng, 2008; 75:826-855

Free expansion simulation

Manufacturer test

Pressure [atm]

Dia

met

er [m

m]

Page 8: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

8

Computer simulated provisional stenting

1.  Insertion of 3x18mm Multi-Link 8 stent

2.  Balloon expansion at 14 atm (Final diameter = 3.22 mm)

3.  Final configuration after

balloon deflation

Note: Post-dilatation with a 3.5x12 balloon is needed (work in progress)

Page 9: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Results – fibrotic plaque

9

•  Geometrical configuration pre/post

PRE

POST

45 deg 57.5 deg 70 deg

α = 45.0° α = 57.5° α = 70.0°

α = 43.1° α = 51.6° α = 65.9°

α

Distal angle

-  Distal angle decreases after stent deployment

-  Lumen shape of side branch is effected

Page 10: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Results: plaque shift

10

•  Volumetric analysis

-  SB lumen : Distal angle ↑↑ SB ostial compromise ↓↓

-  SB plaque : Distal angle ↑↑ Plaque shift ↓↓ (absent with distal angle = 70°)

FIBROTIC PLAQUE

SB ostial compromise = lumen volume decrease in the SB ostial segment after MB stenting Plaque shift = plaque volume increase in the SB ostial segment after MB stenting Carina shift = vessel volume decrease in the SB ostial segment after MB stenting

Distal angle ↑↑ Area Polygon of confluence ↑↑ Plaque shift ↓↓

5 mm

SB

DMB

PMB

Page 11: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Results – lipid plaque

11

•  Geometrical configuration pre/post

PRE

POST

45 deg 57.5 deg 70 deg

α = 45.0° α = 57.5° α = 70.0°

α = 38.4° α = 49.4° α = 65.4°

α

Distal angle

-  Distal angle decreases after stent deployment

-  Lumen shape of side branch is effected

-  The decrease of distal angle is greater in the lipid cases than the fibrotic cases

Page 12: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Results: plaque shift

12

•  Volumetric analysis

FIBROTIC PLAQUE LIPID PLAQUE

SB ostial compromise = lumen volume decrease in the SB ostial segment after MB stenting Plaque shift = plaque volume increase in the SB ostial segment after MB stenting Carina shift = vessel volume decrease in the SB ostial segment after MB stenting

-  SB lumen : SB ostial compromise and plaque shift more severe in lipid cases

-  SB plaque : Carina shift present only in the lipid case with distal angle = 70°

-  Lumen size of side branches marginally changes after stenting but lumen shape is effected

Page 13: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Results: plaque shift

13

•  Volumetric analysis

FIBROTIC PLAQUE LIPID PLAQUE

SB ostial compromise = lumen volume decrease in the SB ostial segment after MB stenting Plaque shift = plaque volume increase in the SB ostial segment after MB stenting Carina shift = vessel volume decrease in the SB ostial segment after MB stenting

-  SB lumen : SB ostial compromise and plaque shift more severe in lipid cases

-  SB plaque : Carina shift present only in the lipid case with distal angle = 70°

-  Lumen size of side branches marginally changes after stenting but lumen shape is effected

Page 14: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Results: plaque shift

14

•  Volumetric analysis

FIBROTIC PLAQUE LIPID PLAQUE

SB ostial compromise = lumen volume decrease in the SB ostial segment after MB stenting Plaque shift = plaque volume increase in the SB ostial segment after MB stenting Carina shift = vessel volume decrease in the SB ostial segment after MB stenting

-  SB ostial compromise and plaque shift more severe in lipid cases

-  Carina shift present only in the lipid case with distal angle = 70°

-  Lumen size of side branches marginally changes after

stenting but lumen shape is effected

Page 15: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Conclusions/Discussion (I)

15

•  Development coronary bifurcation model including plaque

dimensions

•  Quantification of plaque shift by volumetric analysis ! same

order of magnitude as Xu et al. Circ cardiovasc int 2012

•  Plaque shift depends on plaque composition and bifurcation

angle

•  The largest plaque shift in lipid rich

plaque

•  The largest plaque shift in

small bifurcation angle

Page 16: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

16

• Lumen size of SIDE BRANCH does marginally change -->

more influence on lumen shape

Simulation Xu et al. 2012

Xu et al. Circ Cardiovasc Interv, 2012; 25 657-662

PRE

POST

45 LIP

Conclusions/Discussion (I)

Page 17: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall

Acknowledgements

J.J. Wentzel, E. Regar, R. van Geuns, A. van der Steen, P. Serruys Erasmus MC, Rotterdam, The Netherlands C. Chiastra, F. Migliavacca, G. Dubini Politecnico di Milano, Milano, Italy F. Iannaccone, M. de Beule, P. Mortier Gent University, Gent, Begium

Page 18: In-silico bifurcation stenting: plaque composition ... · Arterial wall thickness (Tzucu 2001, Holzapfel 2005): intima + media thickness = 0.1 lumen Diameter (healthy) intima thickness/wall