indico 100l service manual - 740914

136
CPI Canada Inc. PRE-INSTALLATION 1 SERVICE AND INSTALLATION 2 INSTALLATION MANUAL INTERFACING, PROGRAMMING, AND CALIBRATION 3 ACCEPTANCE TESTING 4 TROUBLESHOOTING 5 REGULAR MAINTENANCE 6 THEORY OF OPERATION 7 SPARES 8 SCHEMATICS 9 The original version of this manual (Jun 03,1999) has been drafted in the English language by: Communications & Power Industries communications & medical products division. X-Ray Generator Service Manual Ch # 740914-00 Rev. B Page 0-1

Upload: arturo-jimenez-terrero

Post on 21-Oct-2015

605 views

Category:

Documents


27 download

TRANSCRIPT

Page 1: Indico 100L Service Manual - 740914

CPI Canada Inc.

PRE-INSTALLATION

1

SERVICE AND

INSTALLATION

2

INSTALLATION MANUAL

INTERFACING, PROGRAMMING, AND CALIBRATION

3

ACCEPTANCE TESTING

4

TROUBLESHOOTING

5

REGULAR MAINTENANCE

6

THEORY OF OPERATION

7

SPARES

8

SCHEMATICS

9

The original version of this manual (Jun 03,1999) has been drafted in the English language by: Communications & Power Industries communications & medical products division.

X-Ray Generator Service Manual Ch # 740914-00 Rev. B Page 0-1

Page 2: Indico 100L Service Manual - 740914

CPI Canada Inc.

(This page intentionally left blank.)

Page 0-2 Rev. B X-Ray Generator Service Manual Ch # 740914-00

Page 3: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1

CHAPTER 1

PRE-INSTALLATION

CONTENTS:

Section 1.1.0 INTRODUCTION ..........................................................................................................................................1-2

1.1.1 Generator Description ...............................................................................................................................1-2 1.1.2 Features ....................................................................................................................................................1-2 1.1.3 Radiographic Performance........................................................................................................................1-2 1.1.4 Environmental Specifications ....................................................................................................................1-3 1.1.5 Applicable Standards ................................................................................................................................1-3 1.1.6 Electromagnetic Compatibility (EMC) .......................................................................................................1-7

1.2.0 SAFETY ........................................................................................................................................................1-7 1.2.1 Safety and Warning Symbols ....................................................................................................................1-7 1.2.2 Safety Notices and Warnings....................................................................................................................1-8 1.2.3 Safety Warning Labels ..............................................................................................................................1-9

1.3.0 PREPARING FOR INSTALLATION............................................................................................................1-14 1.3.1 Generator Heat Output............................................................................................................................1-14 1.3.2 Generator Power Requirements .............................................................................................................1-14 1.3.3 Generator Ground Requirements............................................................................................................1-17 1.3.4 Locating and Mounting the Generator.....................................................................................................1-17 1.3.5 Seismic Centers for the Generator..........................................................................................................1-20 1.3.6 Pre-Installation Checklist.........................................................................................................................1-20 1.3.7 Tools and test Equipment Required........................................................................................................1-21

1.4.0 COMPATIBILITY LISTING..........................................................................................................................1-21 1.5.0 GENERATOR LAYOUT AND MAJOR COMPONENTS ............................................................................1-22 1.6.0 CUSTOMER SUPPORT .............................................................................................................................1-23

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-1

Page 4: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.1.0 INTRODUCTION

This chapter summarizes the main features of the 30 kW, 37.5 kW, and 50 kW, 100 kHz X-ray generator (performance, regulatory and compatibility). Safety information is provided, along with environmental, room and installation requirements. This chapter concludes with a pre-installation checklist and a diagram showing the major component layout. The information in this chapter is provided in order for the installer to be able to plan the site layout prior to installation of the generator.

1.1.1 Generator Description

Depending on model, the X-ray generator is intended for use in radiographic only stationary X-ray systems used for medical, chiropractic, or veterinary diagnostic purposes. The X-ray generator consists of a main power cabinet, and a control console. The main power cabinet contains the HT tank and control circuits, the filament drivers, the low speed starter, and interface connections to the room equipment. The control console allows the operator to select the technique factors, image receptors, etc., and to initiate an X-ray exposure. The console also contains the AEC interface connections.

1.1.2 Features

The following are the main features of and the options available for the generator:

• Integral low speed starter, compatible with X-ray tubes with type “R” stator. • 24 VAC 150 watts power source for collimator lamp. • 24 VDC 75 watts power source for system locks. • Optional AEC. Interfaces with 2 A.I.D. ion chambers.

1.1.3 Radiographic Performance kVp range: 40 to 150 kVp output capability (125

kVp maximum for some models). kVp steps: variable in 1 kVp steps. kVp accuracy: ± 5 %. Ripple (kV): 5 % p-p over the full operating range. Risetime (10-90%): <1.5 ms. Time range: 0.004 - 6 seconds. mAs range 1.0 to 420 mAs (large focus), 0.1 to

100 mAs (small focus) for 30 and 37.5 kW generators 1.0 to 600 mAs (large focus), 0.1 to 100 mAs (small focus) for 50 kW generators.

mAs accuracy: ± (10 % + 0.2) mAs. mA range 25 to 300 mA (30 kW generators), 25

to 375 mA (37.5 kW generators), 25 to 500 mA (50 kW generators).

Coefficient of linearity: 0.05 (station to station) mAs. Coefficient of reproducibility: kV, mAs ≤ 0.04.

Page 1-2 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 5: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.1.4 Environmental Specifications

OPERATINGAmbient temperature range 10 to 40 °C (50 to 104 °F). Relative humidity 20 to 80%, non condensing. Atmospheric pressure range 500 to 1060 hPa (375 to 795 mm Hg). TRANSPORT AND STORAGEAmbient temperature range -25 to 70 °C (-13 to 158 °F). Relative humidity 5 to 95%, non condensing. Atmospheric pressure range 500 to 1060 hPa (375 to 795 mm Hg).

1.1.5 Applicable Standards

This series of X-ray generators complies with some or all of the following regulatory and design standards as indicated in the table below. VZW2551 / 2552

SERIES VZW2554 SERIES

REGULATORY STANDARD 30 kW 37.5 kW 30 kW 37.5 kW 50 kW1) UL187. √ √ 2) FDA Center for Devices & Radiological Health

(CDRH) - 21 CFR subchapter J (USA). √ √ √ √ √

3) Radiation Emitting Devices Act - C34 (Canada). √ √ √ √ √ 4) Medical Devices Regulations (Canada). √ √ √ 5) EC Directive 93/42/EEC concerning Medical

Devices (European Community). √ √ √

6) EN 60601-1/IEC 60601-1, EN 60601-2-7/IEC 60601-2-7, CSA 601.1, UL2601.1.

√ √ √

7) EN 60601-1-2:2001/IEC 60601-1-2:2001. √ √ √ 8) EN 60601-1-4/IEC 60601-1-4, EN ISO 14971. √ √ √ √ √

A) SAFETY

• UL187. • FDA Center for Devices & Radiological Health (CDRH) - 21 CFR title 21 subchapter J (USA). • Radiation Emitting Devices Act - C34 (Canada). • Medical Device Regulations (Canada). • EC Directive 93/42/EEC concerning Medical Devices (European Community). • EN 60601-1/IEC 60601-1, EN60601-2-7/IEC 60601-2-7, CSA 601.1, UL2601.1

-Type of protection against electric shock: Class I equipment. -Degree of protection against electric shock: Not classified. -Degree of protection against harmful ingress of water: Ordinary equipment. -Mode of operation: Continuous operation with intermittent loading (standby - exposure). -Equipment not suitable for use in presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.

• EN 60601-1-4/IEC 60601-1-4, EN ISO 14971.

NOTE: All referenced standards are considered to be at the latest revision.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-3

Page 6: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.1.5 Applicable Standards (Cont) B) EMC (EN 60601-1-2:2001/IEC 60601-1-2:2001)

Guidance and manufacturer’s declaration – electromagnetic emissions The VZW2554 series of X-ray generators is intended for use in the electromagnetic environment specified below. The customer or the user of the VZW2554 series should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance

RF emissions CISPR 11

Group 1 The VZW2554 series of X-ray generators use RF energy only for their internal functions. Therefore, the RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class A (The VZW2554 series of X-ray generators in

combination with shielded location)

Harmonic emissions

IEC 61000-3-2

Not Applicable

Voltage fluctuations/

flicker emissions IEC 61000-3-3

Not Applicable

The VZW2554 series of X-ray generators must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that exits the shielded location, a minimum RF filter attenuation of 40dB from 30 MHz to 230 MHz and 47dB from 230 MHz to 1 GHz. (The minimum at 30 MHz is 40dB and the minimum at 230 MHz is 47dB). The VZW2554 series is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

NOTE It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specifications.

Page 1-4 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 7: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.1.5 Applicable Standards (Cont)

Guidance and manufacturer’s declaration – electromagnetic immunity The VZW2554 series of X-ray generators is intended for use in the electromagnetic environment specified below. The customer or the user of the VZW2554 series should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD) IEC 61000-4-2

± 6 kV contact ± 8 kV air

± 6 kV contact ± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst IEC 61000-4-4

± 2 kV for power supply lines ± 1 kV for

input/output lines

± 2 kV for power supply lines ± 1 kV for

input/output lines

Mains power quality should be that of a typical commercial or hospital environment.

Surge IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

± 1 kV differential mode

± 2 kV common mode

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruption,

and voltage variations on power supply

input lines IEC 60601-4-11

< 5 % UT

(> 95 % dip in UT)for 0.5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT)

< 5 % UT (> 95 % dip in UT)

for 5 s

< 5 % UT

(> 95 % dip in UT)for 0.5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT)

< 5 % UT (> 95 % dip in UT)

for 5 s

Mains power quality should be that of a typical commercial or hospital environment. If the user of the VZW2554 series X-ray generator requires continued operation during power mains interruptions, it is recommended that the X-ray generator be powered from an uninterruptible power supply or battery.

Power frequency (50/60 Hz)

IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment

NOTE: UT is the A.C. mains voltage prior to application of the test level.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-5

Page 8: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.1.5 Applicable Standards (Cont)

Guidance and manufacturer’s declaration – electromagnetic immunity The VZW2554 series of X-ray generators is intended for use in the electromagnetic environment specified below. The customer or the user of the VZW2554 series should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Conducted RF IEC 61000-4-6

3 Vrms

150 kHz to 80MHz

3 Vrms

150 kHz to 80MHz

The VZW2554 series of X-ray generators must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that enters the shielded location, a minimum RF filter attenuation of 40dB from 30 MHz to 230 MHz and 47dB from 230 MHz to 1 GHz. (The minimum at 30 MHz is 40dB and the minimum at 230 MHz is 47dB.)

Radiated RF IEC 61000-4-3

3 V/m 80MHz to 2.5

GHz

3 V/m 80MHz to 2.5

GHz

Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3 V/m.a

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 These guidelines may not apply all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. NOTE 2 It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specification. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VZW2554 series of X-ray generators is used exceeds the applicable RF compliance level above, the X-ray generator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the X-ray generator.

Page 1-6 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 9: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.1.6 Electromagnetic Compatibility (EMC)

In accordance with the intended use, some models of this series of X-ray generators comply with the European Council Directive concerning Medical Devices as per the table in this section. The CE marking affixed to this product signifies this. One of the harmonized standards of this Directive defines the permitted levels of electromagnetic emission from this equipment and its required immunity from the electromagnetic emissions of other devices. It is not possible, however, to exclude with absolute certainty the possibility that other high frequency electronic equipment, which is fully compliant to the EMC regulations, will not adversely affect the operation of this generator. If the other equipment has a comparatively high level of transmission power and is in close proximity to the generator, these EMC concerns (the risk of interference) may be more pronounced. It is therefore recommended that the operation of equipment of this type such as mobile telephones, cordless microphones and other similar mobile radio equipment be restricted from the vicinity of this X-ray generator.

1.2.0 SAFETY

1.2.1 Safety and Warning Symbols

The following advisory symbols are used on the safety warning labels, and/or on circuit boards, and/or on the operator console.

High voltage symbol used to indicate the presence of high voltage.

Warning symbol used to indicate a potential hazard to operators, service personnel or to the equipment. It indicates a requirement to refer to the accompanying documentation for details.

Radiation exposure symbol used on operator console. Lights to indicate that an exposure is in progress. This is accompanied by an audible tone from the console.

WARNING THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.

Radiation warning label on operator console. Never allow unqualified personnel to operate the X-ray generator.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-7

Page 10: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.2.2 Safety Notices and Warnings

WARNING: THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS AND OPERATING INSTRUCTIONS ARE OBSERVED.

WARNING: PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY GENERATORS ARE THE RESPONSIBILITY OF USERS OF SUCH GENERATORS. CPI CANADA INC. PROVIDES INFORMATION ON ITS PRODUCTS AND ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTER-SALE OPERATING AND SAFETY PRACTICES.

THE MANUFACTURER ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT MAINTAINED OR SERVICED ACCORDING TO THIS SERVICE AND INSTALLATION MANUAL, OR FOR ANY GENERATOR THAT HAS BEEN MODIFIED IN ANY WAY.

THE MANUFACTURER ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY RADIATION OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM POOR OPERATING TECHNIQUES OR PROCEDURES.

X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending over periods of many months or even years. Operators and service personnel should avoid any exposure to the primary beam and take protective measures to safeguard against scatter radiation. Scatter radiation is caused by any object in the path of the primary beam and may be of equal or less intensity than the primary beam that exposes the film. No practical design can incorporate complete protection for operators or service personnel who do not take adequate safety precautions. Only authorized and properly trained service and operating personnel should be allowed to work with this X-ray generator equipment. The appropriate personnel must be made aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of excessive exposure to X-ray radiation during system operation.

Do not connect unapproved equipment to the rear of the console. J1 and J12 on the CPU board are used for interconnect cables to the generator main cabinet. J1 and J2 on the optional AEC board are the AEC inputs from the AEC chamber(s). J5 and J6 are CAEC inputs ONLY. INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE.

CAUTION: DO NOT EXCEED THE TUBE MAXIMUM OPERATING LIMITS. INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS GENERATORS ARE OPERATED WITHIN PUBLISHED SPECIFICATIONS.

Page 1-8 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 11: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.2.2 Safety Notices and Warnings (Cont)

WARNING: HAZARDOUS VOLTAGES EXIST INSIDE THE GENERATOR `WHENEVER THE MAIN POWER DISCONNECT IS SWITCHED ON. THESE AREAS INCLUDE THE MAIN FUSEHOLDER AND ASSOCIATED CIRCUITS ON THE POWER INPUT BOARD.

THE CONSOLE ON/OFF SWITCH DOES NOT DISCONNECT THE MAIN POWER FROM THE ABOVE AREAS INSIDE THE GENERATOR.

THE DC BUSS CAPACITORS, LOCATED IN THE MAIN CABINET PRESENT A SAFETY HAZARD FOR AT LEAST 5 MINUTES AFTER THE POWER HAS BEEN REMOVED FROM THE UNIT. CHECK THAT THESE CAPACITORS ARE DISCHARGED BEFORE SERVICING THE GENERATOR.

1.2.3 Safety Warning Labels

This subsection defines the safety labels used inside and outside the generator cabinet.

NOTE: THESE LABELS AND WARNINGS ARE PROVIDED TO ALERT SERVICE PERSONNEL THAT SERIOUS INJURY WILL RESULT IF THE HAZARD IDENTIFIED IS IGNORED.

WARNING: SWITCH OFF THE MAIN POWER DISCONNECT AND ALLOW SUFFICIENT TIME FOR ALL CAPACITORS TO DISCHARGE BEFORE REMOVING ANY COVERS.

WARNING: IF ANY BARRIERS OR COVERS MUST BE REMOVED FOR SERVICE, TAKE ALL REQUIRED PRECAUTIONS WITH RESPECT TO THE HAZARD(S) AND IMMEDIATELY REPLACE THE BARRIERS / COVERS WHEN THE NEED FOR REMOVAL IS COMPLETED.

This information is provided to help you establish safe operating conditions for both you and your X-ray generator. Do not operate this X-ray generator except in accordance with these instructions, and any additional information provided by the X-ray generator manufacturer and / or competent safety authorities.

WHEN CONNECTING ADDITIONAL EQUIPMENT TO THE GENERATOR, IT IS THE RESPONSIBILITY OF THE INSTALLER TO VERIFY COMPLIANCE TO ALL LOCAL REGULATORY STANDARDS FOR SAFETY, EMC, AND HAZARD ANALYSIS / RISK ASSESSMENT OF THE FINAL SYSTEM CONFIGURATION.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-9

Page 12: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.2.3 Safety Warning Labels (Cont)

Caution HV Behind Cover Label

This label is attached to a cover over the inverter board(s). The inverter assembly is connected to the main DC bus and will have high voltage applied at all times that the generator is switched on. This assembly will remain energized for 5 minutes after either the generator is switched off via the console, or the main disconnect is switched off. This label is also attached to a cover over the main input fuses on the power input board. This area will have mains voltage applied as long as the main disconnect is switched on.

Weight Label

This label is attached to the main cabinet and to the HT oil tank, and states the approximate weight of the main cabinet and the HT oil tank. This label cautions against attempting to lift those assemblies without proper assistance.

Page 1-10 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 13: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.2.3 Safety Warning Labels (Cont)

DC Buss Label

This label is attached to the main access panel on the generator cabinet. The internal capacitors may hold a lethal charge for up to 5 minutes after the console or the main power disconnect is switched off. Do not remove the cover for a minimum of 5 minutes after the power has been switched off.

WARNING: WAIT A MINIMUM OF 5 MINUTES AFTER THE INPUT MAINS POWER HAS BEEN REMOVED BEFORE REMOVING ANY COVERS OR ACCESS PANELS. ONCE THE COVER(S) / PANEL(S) ARE REMOVED CHECK THAT THE VOLTAGE ACROSS THE DC BUS CAPACITORS IS LESS THAN 48 VDC BEFORE SERVICING.

Caution HV Exposed Label

This label is attached to the main access panel on the generator cabinet. High voltage is present within this cabinet at all times that the mains power is switched on.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-11

Page 14: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.2.3 Safety Warning Labels (Cont)

Caution HV Label

This label is attached to the resonant assembly. The resonant board will have exposed high voltage at all times that the DC bus is charged.

Danger HV Label

This label is attached to the primary terminals on the HT oil tank. These terminals may be energized at all times that the generator is switched on, and for 5 minutes after the console or the main disconnect is switched off.

Customer Interface Board

110 VAC is present on this board at all times that the generator is switched ON.

AEC Board

Some versions of AEC board will have high voltage present. This only applies to AEC boards fitted with a high voltage supply for ion chambers, and can be recognized by the presence of the high voltage supply components within the dashed outline on the right side of the board. No high voltage hazard exists on this board if these components are not fitted.

Page 1-12 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 15: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.2.3 Safety Warning Labels (Cont)

Power Input Board

HIGH VOLTAGE HAZARD: High voltage is present on this board whenever the generator is connected to live AC mains. The DC bus capacitors will remain charged for up to 5 minutes after the generator has been switched off. BURN HAZARD: Power input boards for single phase generators are fitted with several high power resistors that operate at temperatures sufficient to cause skin burn. Ensure that these resistors have cooled sufficiently after the power has been switched off before servicing.

Primary of Power Supply Auxiliary Transformer

FUSE RATING: Fuse F1 is located on the generator chassis, to the left of the main input fuses on the power input board. Single phase generators: FNM-3 (3A 250V). Three phase generators: FNQ-2 (2A 500V).

Primary of 24 V Interface / Console Transformer

FUSE RATING (three phase generators): On three phase generators, F2 is located on the generator chassis, to the left of the main input fuses on the power input board. (For single-phase generators, the fuse for the primary of this transformer is on the power input board).

Fuse type and rating (three phase generators): FNQ-1 (1A 500V).

Low Speed Starter Board

HIGH VOLTAGE HAZARD: High voltage is present on this board whenever the generator is switched on.

BURN HAZARD: This board is fitted with high power resistors that operate at temperatures sufficient to cause skin burn. Ensure that these resistors have cooled sufficiently after the power has been switched off before servicing.

Inverter Board

HIGH VOLTAGE HAZARD: The inverter board(s) are connected to the main DC bus and will have high voltage applied at all times that the generator is switched on. This assembly will remain energized for up to 5 minutes after the generator is switched off, or the main disconnect is switched off.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-13

Page 16: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.2.3 Safety Warning Labels (Cont)

HT Tank

HIGH VOLTAGE HAZARD: High voltage may be present at the primary terminals on the tank lid board, at the output high voltage connectors, and at the mA/mAs measuring jacks if the shorting link is opened for mA/mAs measurements. The generator must not be energized if the leads that attach to the primary of the high voltage transformer are disconnected.

TORQUE NOTICE: Do not over-tighten the nuts on the feed through terminals for the primary of the HT transformers. Over tightening may damage the HT tank.

1.3.0 PREPARING FOR INSTALLATION

1.3.1 Generator Heat Output

The maximum heat output of the main cabinet is 500 watts, with 20 watts maximum heat output for the console. Both the main cabinet and the console are convection cooled. Cooling slots underneath and on the sides of the main cabinet must not be obstructed, and sufficient clearance must be provided to allow adequate air flow for proper cooling. The console should never be covered while it is switched on.

1.3.2 Generator Power Requirements

30 kW Single Phase Line Voltage 230 VAC ± 10%, 1 phase.

Line Frequency 50/60 Hz.

Momentary Current 185 Amps.

Standby Current 5 Amps.

37.5 kW Single Phase Line Voltage 230 VAC +10% -0%, 1 phase.

Line Frequency 50/60 Hz.

Momentary Current 235 Amps.

Standby Current 5 Amps.

30 kW Three Phase Line Voltage 400 VAC ± 10%, 3 phase.

Line Frequency 50/60 Hz.

Momentary Current 60 Amps.

Standby Current 5 Amps.

Page 1-14 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 17: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.3.2 Generator Power Requirements (Cont)

37.5 kW Three Phase Line Voltage 400 VAC ± 10%, 3 phase.

Line Frequency 50/60 Hz.

Momentary Current 75 Amps.

Standby Current 5 Amps.

50 kW Three Phase Line Voltage 400 VAC ± 10%, 3 phase.

Line Frequency 50/60 Hz.

Momentary Current 100 Amps.

Standby Current 5 Amps.

The following table defines the power line requirements for the generators.

NOTE: THE FOLLOWING TABLE CONTAINS RECOMMENDED VALUES FOR THE WIRE SIZES BETWEEN THE MAINS DISCONNECT AND THE GENERATOR. THE ACTUAL VALUES USED AT AN INSTALLATION ARE DEPENDENT ON THE QUALITY OF THE INPUT LINE (VOLTAGE LEVEL) THE CURRENT REQUIREMENTS AND THE LENGTH OF THE CABLE RUN AND MUST BE CONFIRMED BY THE INSTALLER. FINAL SELECTION OF GENERATOR INPUT WIRE AND DISCONNECTS AS WELL AS THE CABLING FROM THE DISTRIBUTION TRANSFORMER TO THE MAINS DISCONNECT, MUST MEET THE REQUIREMENTS OF THE LOCAL ELECTRICAL CODES AND IS USUALLY DETERMINED BY HOSPITAL/CONTRACTOR ENGINEERING. ALL THE RATINGS LISTED CONSIDER THE GENERATOR REQUIREMENTS ONLY. THE INSTALLER MUST MAKE THE NECESSARY COMPENSATION FOR ADDITIONAL LOAD REQUIREMENTS A POOR QUALITY INPUT LINE MAY RESULT IN THE INSTALLER HAVING TO DERATE THE GENERATOR'S MAXIMUM POWER

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-15

Page 18: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.3.2 Generator Power Requirements (Cont)

Mains Voltage

Minimum Recommended

Mains Disconnect to Generator

15 ft/5 m max)

Generator Momentary

Line Current

Minimum Recommended

Generator Service Rating

Minimum Recommended

Distribution Transformer

Rating

Minimum Recommended

Ground Wire Size

Apparent Mains

Resistance

30 kW 1∅ Generator

230 VAC

#4

(21 mm2)

185 A 120 A 45 kVA #4

(21 mm2) 0.06 Ω

37.5 kW 1∅ Generator

230 VAC

#2

(33 mm2)

235 A 120 A 60 kVA #2

(33 mm2) 0.048 Ω

30 kW 3∅ Generator

400 VAC #6

(13.3 mm2)

60 A 100 A 40 kVA #6

(13.3 mm2) 0.30 Ω

37.5 kW 3∅ Generator

400 VAC #6

(13.3 mm2)

75 A 100 A 50 kVA #6

(13.3 mm2) 0.24 Ω

50 kW 3∅ Generator

400 VAC #6

(13.3 mm2)

100 A 100 A 65 kVA #6

(13.3 mm2) 0.17 Ω

• Recommended Service Disconnect: Refer to the preceding table. • All wiring and grounding must comply with local electrical codes. • All wiring must be copper. • The disconnect switch shall be located within reach of the operator.

Page 1-16 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 19: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.3.3 Generator Ground Requirements POWER LINE:

• A suitable ground must be connected from the disconnect switch to the main generator ground, located to the left of the main fuse block on the power input board. The ground wire is typically part of the line cord, and the current capacity of the ground conductor must normally be equal to or greater than that of the line conductors.

• If a neutral line is provided with the system, under no circumstances is it to be used for ground purposes. The ground conductor must carry fault currents only.

X-RAY TUBE HOUSING:

• A copper ground cable, #10 AWG (6 mm2) or greater, is to be connected from the X-ray tube’s housing to the H.T. tank ground stud (located at the top of the H.T. tank).

STATOR CABLE:

• Use of a shielded stator cable is recommended. The shield for the stator cable must be properly grounded at the chassis ground studs as per chapter 2.

1.3.4 Locating and Mounting the Generator The main generator cabinet is self standing and does not need to be supported. However, the installation should meet the following requirements:

• The floor must be flat and level. • The generator installation area must be clean and free of dirt or debris. • If required, the generator may be anchored to the floor via the hold-down brackets (figure 1-1) or via

the holes located in the base of the cabinet (figure 1-2). • Sufficient room must be provided to allow access to the rear and side panels for installation and

service. See figure 1-3 for recommended clearances. • A cable conduit should be provided from the control console to the main cabinet to allow routing of

the control cable if required. Allow for a 2 inch conduit. • The control console is normally freestanding on a desk or shelf. It may be anchored if necessary.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-17

Page 20: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.3.4 Locating and Mounting the Generator (Cont)

LEVELLINGFEET

(4 PLACES)

SPACERSSUPPLIED BYCUSTOMER(4 PLACES)

TOPVIEW

20.1 (511)

FILE: IN_MTG.CDR

16.1 (409)

3.3

(85)13

.9 (3

53)

DIMENSIONS ARE IN INCHES (MM)

0.39 (10 mm ) DIA

FRONT

Figure 1-1: Anchoring plates for securing the generator

TOPVIEW

15.0 (380)

16.1 (409)

FILE: IN_MTG2.CDR

12.3

(312

)

DIMENSIONS ARE IN INCHES (MM)

0.63 (16 mm ) DIA

FRONT

1.5 (38)

Figure 1-2: Holes in base of cabinet for securing the generator

Page 1-18 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 21: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.3.4 Locating and Mounting the Generator (Cont)

Figure 1-3 shows recommended clearances around the generator.

Figure 1-3: Generator clearances

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-19

Page 22: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.3.5 Seismic Centers for the Generator

Figure 1-4 shows the seismic center location for the generator main cabinet.

FRONTLEFT SIDE

7(178)

6(152)

13.7

(348

)

DIMENSIONS ARE IN INCHES(MM)

INDICO L SEISMIC.CDR

Figure 1-4: Generator seismic centers

1.3.6 Pre-Installation Checklist

Before starting the generator installation, review the following checklist.

CHECK √ DESCRIPTION Is there an unloading area to transport the generator from the delivery truck to

the inside of the building? If the installation is not on the same floor as the delivery entrance, is there an

elevator available? Is there a transport dolly or similar device to move the generator? Do any regulatory bodies need to be notified prior to installation? If movers are required, have arrangements for time and equipment been

completed? Are lifting straps or some other suitable device available to lift the generator off

the shipping pallet?

Page 1-20 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 23: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.3.7 Tools and test Equipment Required

The following is a checklist of recommended tools and test equipment for installation and calibration of the generator.

CHECK √ DESCRIPTION General handtools for installation: Wrenches, nut drivers, assortment of

screwdrivers, pliers, etc. If the generator is to be anchored to the floor, suitable hardware and drills, drill

bits etc must be available. A supply of connectors for wiring: terminal lugs, caps, splices etc. A calibrated DVM which indicates true RMS voltages. Dual trace memory oscilloscope with a minimum 20 MHz bandwidth;

appropriate leads, probes, etc. Device for measuring true kVp. This may be a Dynalyzer equivalent or a non-

invasive meter such as the Keithley TRIAD system. A calibrated radiation meter with detectors that will allow for R/min and uR

type measurements (or uGy and Gy/min). A suitable mA / mAs meter. A strobe or reed type tachometer to verify that the anode is rotating up to

speed. A sufficient selection of patient absorbers to allow AEC calibration if this option

is fitted. A suggested selection is 3/4 inch Al (quantity 2); 1 mm of Cu (quantity 8), Water in containers of 5.0. 10.0, 15.0 cm thickness.

Vapor proof compound for the HT terminations.

1.4.0 COMPATIBILITY LISTING

This X-ray generator is compatible with the following equipment: X-RAY TUBES: Refer to the compatibility statement/product description at the end of this chapter, for compatibility. NOTE THE STATOR WARNING BELOW.

NOTE: THE LOW SPEED STARTER IS COMPATIBLE WITH “R” TYPE STATORS ONLY.

AEC DEVICES: An AEC option is available to allow interfacing with A.I.D. and Comet ion chambers. Refer to compatibility statement/product description at the end of this chapter, for compatibility. BUCKY: Liebel-Flarsheim 110V Bucky or equivalent, Summit J500 Bucky or equivalent.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-21

Page 24: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc. 1.5.0 GENERATOR LAYOUT AND MAJOR COMPONENTS

The dimensions and weight of the generator are shown in the table below.

ITEM LENGTH WIDTH HEIGHT WEIGHT Main cabinet in shipping pack 32 (815) 31 (787) 51 (1295) 210 (95) Main cabinet unpacked 18.1 (459) 16.1 (408) 41 (1041) 200 (90) Control console 17 (432) 8.75 (225) 3 (77) 5 (2.25)

The above dimensions are inches (mm), weights are in pounds (Kg). Figures 1-4 and 1-5 show the major components located inside the main cabinet.

REAR

TOP VIEWOF

GENERATORCABINET

FRONT

NOTE: FIGURES 1-5AND 1-6 SHOW THE

VIEWS REFERENCEDIN THE FIGURE ABOVE

RIGHT SIDE

LEFT SIDE

HT OIL TANK

FILE: SM_SUBID.CDR

AUXILIARY BOARD

FILAMENT BOARD

INVERTER BOARD(DEPENDING ON MODEL,MAY USE ONE OR TWOINVERTER BOARDS)

DC BUS ASSEMBLY(1 PHASE UNITS ONLY)

CUSTOMER INTERFACEBOARD

POWER SUPPLYCONTROL BOARD

Figure 1-5: Major generator subassemblies view 1

Page 1-22 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 25: Indico 100L Service Manual - 740914

CPI Canada Inc. Pre-installation 1 1.5.0 GENERATOR LAYOUT AND MAJOR COMPONENTS (Cont)

RESONANTBOARD

POWER INPUTBOARD

MAIN INPUTFUSES

POWER SUPPLYAUXILIARY

TRANSFORMER

24V INTERFACE/CONSOLE POWER

TRANSFORMER

AUXILIARYPOWER FUSE

LOW SPEEDSTARTERBOARD

ACCESS COVERSHOWN REMOVED

SM_SUB1.CDR Figure 1-6: Major generator subassemblies view 2

1.6.0 CUSTOMER SUPPORT Any questions regarding X-ray generator operation should be directed to the Customer Support Department as shown on the inside cover page of this manual.

X-Ray Generator Service Manual Ch # 740914-01 Rev. N Page 1-23

Page 26: Indico 100L Service Manual - 740914

1 Pre-installation CPI Canada Inc.

(This page intentionally left blank)

Page 1-24 Rev. N X-Ray Generator Service Manual Ch # 740914-01

Page 27: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2

CHAPTER 2

INSTALLATION

CONTENTS:

2.1.0 INTRODUCTION ...........................................................................................................................................2-2 2.2.0 UNPACKING..................................................................................................................................................2-2 2.3.0 LIFTING THE GENERATOR .........................................................................................................................2-3 2.4.0 EQUIPMENT PLACEMENT...........................................................................................................................2-4

2.4.1 Main Cabinet ..............................................................................................................................................2-4 2.4.2 Control Console..........................................................................................................................................2-4

2.5.0 WIRING TO THE GENERATOR....................................................................................................................2-5 2.5.1 Control Console..........................................................................................................................................2-5 2.5.2 Hand Switch (Optional) ..............................................................................................................................2-6 2.5.3 Stator Connections.....................................................................................................................................2-7 2.5.4 Thermal Switch...........................................................................................................................................2-8 2.5.5 Power Line Mains.......................................................................................................................................2-8 2.5.6 High Tension Cables ..................................................................................................................................2-9 2.5.7 X-Ray Tube Housing Ground.....................................................................................................................2-9 2.5.8 Room Equipment......................................................................................................................................2-10 2.5.9 Safety Interlocks.......................................................................................................................................2-10

2.6.0 PROGRAMMING THE LOW SPEED STARTER ........................................................................................2-10 2.6.1 Low Speed Starter Tube Select Table .....................................................................................................2-10 2.6.2 Low Speed Starter Boost Voltage Selection ............................................................................................2-11 2.6.3 Low Speed Starter Run Voltage Selection...............................................................................................2-12 2.6.4 Low Speed Starter Boost Time Selection ................................................................................................2-13

2.7.0 INITIAL RUN-UP..........................................................................................................................................2-13 2.7.1 Initial Voltage Measurements ...................................................................................................................2-13 2.7.2 Single Phase Primary Tap Selection........................................................................................................2-15 2.7.3 Three Phase Primary Tap Selection ........................................................................................................2-16

2.8.0 TUBE MA AUTO CALIBRATION.................................................................................................................2-17 2.9.0 FINAL CHECKS...........................................................................................................................................2-17 2.10.0 UTILITY SOFTWARE ................................................................................................................................2-17

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-1

Page 28: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.1.0 INTRODUCTION

This Chapter contains instructions for unpacking, positioning, and cabling the X-ray generator, allowing for initial power-up and tube auto calibration.

2.2.0 UNPACKING

WARNING: THE GENERATOR MAIN CABINET WEIGHS APPROXIMATELY 210 POUNDS (95 KG) IN ITS SHIPPING CONTAINER.

THE HT OIL TANK IS LOCATED IN THE MAIN CABINET. ONE PERSON SHOULD

NOT ATTEMPT TO LIFT OR MOVE THIS ASSEMBLY WITHOUT PROPER EQUIPMENT OR ASSISTANCE.

1. Inspect the pack for evidence of shipping damage. If there is evidence of shipping damage, note

this in the event that a damage claim is justified. 2. Remove the cardboard outer pack. See the cautionary note below before removing the pack.

CAUTION: OPEN THE CARDBOARD PACK CAREFULLY. SHARP TOOLS MAY DAMAGE THE CONTENTS.

3. Set aside the cardboard pack(s). 4. Unscrew the bolts that secure the generator to the shipping pallet. Carefully lift the generator

from the pallet. Refer to 2.3.0 for the procedure for lifting the generator. 5. Inspect for internal and external shipping damage. 6. Unscrew the leveling feet at the bottom of the generator by a minimum of 1 ½ in. (35 mm). This

will provide the required airflow underneath the generator cabinet and allow room to make leveling adjustments when the generator is placed in its final location.

7. Unpack the control console. 8. Unpack the manuals and any other paperwork that may be packed with the generator. 9. Keep the shipping containers. In case of shipping damage, place the unit(s) back in its shipping

pack and notify the carrier and the Customer Support Department as shown on the inside cover page of this manual.

Page 2-2 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 29: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.3.0 LIFTING THE GENERATOR

One of the following methods is recommended for lifting or moving the generator. Proper equipment or sufficient help is necessary to lift the unit. Refer to figure 2-1 for identification of the panels referenced below. • Lifting straps may be used with a hoist to raise the generator. The straps should be placed under the

generator cabinet and along the four sides to properly support the cabinet. • Remove the main access panel and the rear wiring access panel from the generator cabinet

Temporarily disconnect the ground wire(s) from the cover(s). The metal cross brace found at the front of the cabinet (below the circuit board mounting panel) may be used, in conjunction with the lip at the upper edge of the opening for the rear wiring access panel, as lifting points. Two people (one on each side) will be required to lift the generator cabinet. Be sure to properly reconnect the ground leads when the covers are reinstalled.

Figure 2-1: Generator lifting points

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-3

Page 30: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.4.0 EQUIPMENT PLACEMENT

2.4.1 Main Cabinet

Place the generator cabinet in a location that will allow the following: • Easy front and side access for service and sufficient clearance at the rear for room interface cables.

Refer to Chapter 1. • Air circulation - a minimum height of 1 ½ in. (35 mm) is recommended to allow airflow underneath the

generator. • Stable footing - the leveling feet at the bottom of the cabinet will be used to prevent movement during

normal operation. • Close proximity to service disconnect boxes - cables should not be on the floor where they could be

stepped on. After the generator has been placed in it’s final location, adjust the leveling feet such that the generator is level and stable. As noted previously, the leveling feet must be unscrewed by a minimum of 1 ½ in. (35 mm) to allow for proper airflow underneath the generator.

2.4.2 Control Console Position the control console in its intended location and ensure that it is stable. Refer to Chapter 1. • If the console is located on a shelf, supply index pins or equivalent hardware to the base of the

console to prevent slipping. • Ensure that the console is mounted at a height and angle to allow easy viewing of the displays. • Leave sufficient slack in the cabling to the console to allow for future service and maintenance.

NOTE: DO NOT LOCATE THE CONTROL CONSOLE WHERE X-RADIATION MAY BE PRESENT DURING INSTALLATION OR OPERATION.

NOTE: IT IS RECOMMENDED THAT THE CONSOLE CABLES NOT BE DISASSEMBLED FOR SYSTEM INSTALLATION. HOWEVER, IF THIS IS ABSOLUTELY NECESSARY TO ROUTE THE CABLES, PLEASE ENSURE CAREFUL AND CORRECT REASSEMBLY OF THE CONNECTOR SHELL TO AVOID ANY POSSIBILITY OF PINCHING OF THE INTERNAL WIRES BY THE SHELL.

YOU MAY CHOOSE TO TEMPORARILY LOCATE THE CONSOLE NEAR THE GENERATOR FOR INITIAL PROGRAMMING AND CALIBRATION. IF THIS IS SO, PLEASE COMPLETE THE FINAL CONSOLE INSTALLATION PER THIS SECTION WHEN THE GENERATOR INSTALLATION IS COMPLETED.

Page 2-4 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 31: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.5.0 WIRING TO THE GENERATOR

Unless specified otherwise, it is suggested that all cables be routed into the generator main cabinet through the wiring access panel on the rear of the cabinet. The cables should be secured to the lip on the access panel using tie wraps or equivalent fasteners. All cables should be routed away from high voltage areas in the main cabinet, and dressed and secured neatly in place. Cables should be cut to the correct length if possible as excess cabling may contribute to EMI/RFI problems. For those cables that cannot be cut to the correct length (HT cables and console cables for example), try to minimize the area inside any loops of excess cable, as these loops are in effect an antenna. EXCESS LENGTHS OF CABLING MUST NEVER BE BUNDLED UP AND STORED INSIDE THE GENERATOR.

2.5.1 Control Console 1. Locate and unbundle the 37 conductor console cable. The free end of this cable is routed out of

the generator main cabinet through the access panel on the rear of the cabinet. Connect the free end of this cable to J1 on the rear of the console. Ensure that the screw locks are fully tightened to secure the connector to the console. For reference, the other end of this cable is connected to J1 on the power supply control board.

2. Locate and unbundle the 25 conductor console cable. The free end of this cable is routed out of

the generator main cabinet through the access panel on the rear of the cabinet. Connect the free end of this cable to J12 on the rear of the console. Ensure that the screw locks are fully tightened to secure the connector to the console. For reference, the other end of this cable is connected to J6 on the customer interface board.

3. Figure 2-2 shows the designations of the connectors on the rear panel of the control console.

Do not connect unapproved equipment to the rear of the console. J1 and J12 on the CPU board are used for the interconnect cables to the generator main cabinet. J1 and J2 on the AEC board are for conventional ion chamber AEC inputs, J5 and J6 are CAEC inputs used for special applications only. J1, J2, J5, and J6 are only available if the optional AEC board is fitted. AEC chamber compatibility must be confirmed per chapter 3. INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE.

Figure 2-2: Rear of control console

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-5

Page 32: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.5.2 Hand Switch (Optional)

Route the hand switch cable into the console via the cutout at the rear of the console. If using a CPI supplied hand switch assembly, plug the 3-pin connector on the end of the hand switch cable into J9 on the CPU board (observe orientation). If using a hand switch assembly without a mating connector for J9, the hand switch connections may be made to the J9 solder pads on the CPU board in the console. Pin assignments are as follows:

CPU BOARD DEFINITION J9-1 X-ray J9-2 Prep J9-3 Common

The hand switch cable should be dressed and secured as required to prevent strain on the terminations during normal use of the generator.

Page 2-6 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 33: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.5.3 Stator Connections

Refer to figure 2-3 for the X-ray tube stator and thermal switch connections. 1. Route the X-ray tube stator cable to the stator terminal block. This is shown in figure 2-3.

USE OF A SHIELDED STATOR CABLE IS RECOMMENDED. THE SHIELD FOR THE STATOR CABLE MUST BE PROPERLY GROUNDED AT THE CHASSIS GROUND STUD SHOWN IN FIGURE 2-3.

2. Connect the wires to the stator terminal block as shown. The tube thermal switch will normally be

connected to the THERMAL SWITCH connections on the stator terminal block, but may optionally be connected to the room interface board (refer to INPUTS in Chapter 3).

Figure 2-3: Stator connections to generator

STATOR LABEL DESIGNATIONS TERMINAL DESIGNATIONS FOR SUMMIT TUBES ONLY COMMON 09 MAIN 07 SHIFT 08 THERMAL SWITCH (Term 1) T5 THERMAL SWITCH (Term 2) T6

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-7

Page 34: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.5.4 Thermal Switch

Refer to 2.5.3 for the X-ray tube thermal switch connections.

2.5.5 Power Line Mains

WARNING: TO AVOID ELECTRICAL SHOCK, ENSURE THAT THE AC MAINS DISCONNECT IS LOCKED IN THE OFF POSITION, AND THAT ALL MAINS CABLES ARE DE-ENERGIZED BEFORE CONNECTING TO THE GENERATOR.

Refer to Chapter 1 for generator power and generator power line requirements. 1. Pass the AC mains cable through the access hole located at the lower rear of the main cabinet. 2. Use an appropriate cable clamp to secure the mains cable at the cabinet entrance. 3. Temporarily remove the safety cover from the main fuses. Strip sufficient cable jacket to allow the

ground wire to reach the main ground connector located at the left side of the main fuse block. Refer to figure 2-4.

4. Connect the ground wire to the chassis ground connector, and connect the mains power wires to

the terminals on the bottom of the main fuse holder (3 wires for 3 phase systems, 2 wires for single phase systems as shown). Be sure to replace the main fuse safety cover after all connections are made and properly tightened.

5. DO NOT SWITCH ON MAINS POWER AT THIS TIME.

Figure 2-4: Generator mains connection

Page 2-8 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 35: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.5.6 High Tension Cables

The X-ray tube should be mounted in its normal fixture i.e. tube stand or other device. 1 Verify that the HT cable terminations are clean, in good condition, and coated with vapor proof

compound. 2. Remove the dust caps which cover the high voltage terminals on the HT oil tank. 3. Connect the anode and cathode cables to the HT tank. Use the right angle connectors for the HT

transformer end. Ensure that the cables are plugged into the proper connectors on the HT tank. Refer to figure 2-5.

4. Be sure that the HT cable connectors are tight and there is no play between the connector

insulator and the screw down ring.

Figure 2-5: HV connector identification

2.5.7 X-Ray Tube Housing Ground

A separate ground wire (10 AWG, 6mm2) must be connected from the X-ray tube housing to one of the ground studs on the HT tank. Refer to figure 2-5 for the location of these ground studs. These ground locations may have other ground wires already connected, ensure that these existing ground wires are not disconnected when making the X-ray tube ground connection. Failure to make this ground connection may result in intermittent operation and/or exposure errors.

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-9

Page 36: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.5.8 Room Equipment

Refer to Chapter 3 for connection of the Buckys, interlocks, collimator lamp and system locks power, and optional AEC chambers. It is suggested that these items not be connected until after the initial run-up of the generator is complete, and the tube auto calibration routine has been performed as described near the end of this Chapter.

2.5.9 Safety Interlocks The room door interlock switch must be wired to the generator as described in INPUTS in Chapter 3 prior to power up. This switch will provide a closed contact when the door is closed.

2.6.0 PROGRAMMING THE LOW SPEED STARTER

PLEASE BE SURE TO READ AND UNDERSTAND THIS SECTION FULLY BEFORE PROCEEDING. Before continuing, note the part number of the low speed starter in the generator. Part number 732752-00 has a 33 µF phase-shift capacitor; part number 732752-01 has a 12.5 µF capacitor, and part number 732752-02 has a 45 µF capacitor. Confirm that the phase shift capacitor is compatible with the desired tube(s), as listed in table 1 in Supplement 746026 which follows chapter 2. The starter boost voltage is set to approximately 240 VAC, (except where specifically noted in table 1 in Supplement 746026), and the starter run voltage is selectable to be 52, 73, or 94 VAC. Boost times are selectable in software to be either 1.2, 1.5, 1.9, or 2.5 seconds. Therefore, the generator may be configured to be compatible with stator types as per table 1 only. If the desired tube type is not listed, please contact CPI product support for assistance.

WARNING: 240 VAC IS PRESENT ON THE LOW SPEED STARTER BOARD AT ALL TIMES THAT THE GENERATOR IS SWITCHED ON. TAKE APPROPRIATE PRECAUTIONS WHEN SERVICING THIS BOARD

2.6.1 Low Speed Starter Tube Select Table

See Supplement 746026 which immediately follows this chapter.

Page 2-10 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 37: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.6.2 Low Speed Starter Boost Voltage Selection

Follow the steps below to verify and configure the correct low speed starter boost voltage. 1. Confirm the required BOOST VOLTAGE for the selected tube type per table 1 in supplement

746026, which follows chapter 2. The requirement for the vast majority of tubes in table 1 is 240 V. For tubes that require 120 V boost, this is noted along with the run voltage, boost time, etc for that tube. • Refer to figure 2-6 (power supply auxiliary transformer). Locate this transformer under the

power input board; refer to chapter 1 for details. • If there is only one connection on the 120V tap on the power supply auxiliary transformer, the

boost voltage is set to 240 VAC. • If there are two connections on the 120V tap, the boost voltage is set to 120 VAC. • Confirm the correct boost voltage at F1 on the low speed starter board with a suitable

voltmeter. • Go to 2.6.3 if the boost voltage is properly set for the selected tube type and therefore does

NOT need to be changed.

2. If the boost voltage tap on the power supply auxiliary transformer needs to be changed from 240V to 120V, proceed as follows: • Loosen the clamping screws for the 240V tap and for the 120V tap. • Move the boost voltage lead from the 240V tap to the 120V tap. The boost voltage lead is the

lead that connects directly to J1 on the low speed starter board. Do not disturb the other leads on the 120V and / or 240V taps.

• Ensure that the wires are positioned between the two metal clamps on the transformer terminal block.

• Retighten both of the clamping screws. • Confirm the correct boost voltage at F1 on the low speed starter board with a suitable

voltmeter.

3. If the boost voltage tap on the power supply auxiliary transformer needs to be changed from 120V to 240V, proceed as follows: • Loosen the clamping screws for the 120V tap and for the 240V tap. • Move the boost voltage lead from the 120V tap to the 240V tap. The boost voltage lead is the

lead that connects directly to J1 on the low speed starter board. Do not disturb the other leads on the 120V and / or 240V taps

• Ensure that the wires are positioned between the two metal clamps on the transformer terminal block.

• Retighten both of the clamping screws. • Confirm the correct boost voltage at F1 on the low speed starter board with a suitable

voltmeter.

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-11

Page 38: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.6.3 Low Speed Starter Run Voltage Selection

Follow the steps below to verify and configure the correct low speed starter run-voltage. 1. Note the required RUN VOLTAGE for the selected tube type per table 1 in Supplement 746026,

which follows chapter 2. Then confirm the low speed starter run-voltage in the generator. This is determined by the run-voltage setting on the power supply auxiliary transformer. Refer to figure 2-6. • Locate this transformer under the power input board; refer to chapter 1 for details. • Locate the run-voltage output on the transformer. This will be set to the 52V, 73V, or 94V tap,

and must match the required run-voltage for the selected tube in table 1. • Go to 2.6.4 if the run-voltage setting is correct for the selected tube type, and therefore does

NOT need to be changed.

2. If the run-voltage setting on the power supply auxiliary transformer needs to be changed, proceed as follows: • Loosen the clamping screws for the current run-voltage tap, and for the required run-voltage

tap. • Move the run-voltage output lead from the current tap position to the required tap position

(52V, 73V, or 94V). Ensure that the wire is positioned between the two metal clamps on the transformer terminal block.

• Retighten both of the clamping screws.

Figure 2-6: Low speed run-voltage tap selection

Page 2-12 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 39: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.6.4 Low Speed Starter Boost Time Selection

Follow the steps below to verify and configure the low speed starter boost time.

1. Ensure that jumper JW1 on the auxiliary board is in the “.15S” position. This sets the hardware boost time to the minimum value. The boost time is then set in software as described below.

2. Refer to chapter 3, section 3.5.6 for the procedure for setting the actual boost time.

NOTE: CONFIRM PROPER LOW SPEED STARTER CONFIGURATION USING A SUITABLE TACHOMETER BEFORE MAKING ANY EXPOSURES.

2.7.0 INITIAL RUN-UP

This section describes the procedure for initial power-on of the generator after it has first been installed.

PLEASE OBSERVE THE FOLLOWING POINTS REGARDING THE MAIN DISTRIBUTION TRANSFORMER: • IF USING A DISTRIBUTION TRANSFORMER WITH AN ISOLATED SECONDARY, THE

SECONDARY WINDING MUST BE A WYE (STAR) CONFIGURATION WITH THE CENTER POINT GROUND REFERENCED. DO NOT USE A DELTA CONFIGURED SECONDARY AS THERE IS NO GROUND REFERENCE IN THIS CONFIGURATION.

• IF USING AN AUTOTRANSFORMER TYPE DISTRIBUTION TRANSFORMER, THE A.C. INPUT TO THE TRANSFORMER MUST BE GROUND REFERENCED.

2.7.1 Initial Voltage Measurements 1. Verify that the mains voltage and current capacity is correct for the generator installation per

GENERATOR POWER REQUIREMENTS in Chapter 1 of this manual. 2. Temporarily remove the safety cover over the main input fuses in the generator. 3. If the mains supply is compatible with the generator, switch on the main breaker and/or the

disconnect switch and check for the following voltages:

NOTE: DO NOT SWITCH THE GENERATOR ON AT THIS TIME (ONLY THE AC MAINS TO THE GENERATOR IS TO BE SWITCHED ON).

WARNING: USE EXTREME CARE IN MEASURING THESE VOLTAGES. ACCIDENTAL CONTACT WITH MAINS VOLTAGES MAY CAUSE SERIOUS INJURY OR DEATH.

MAINS VOLTAGE WILL BE PRESENT INSIDE THE GENERATOR CABINET, EVEN WITH THE CONSOLE SWITCHED OFF.

THE DC BUS CAPACITORS PRESENT A SAFETY HAZARD FOR A MINIMUM OF 5 MINUTES AFTER THE GENERATOR HAS BEEN SWITCHED OFF. CHECK THAT THESE CAPACITORS ARE DISCHARGED BEFORE TOUCHING ANY PARTS IN THE GENERATOR.

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-13

Page 40: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.7.1 Initial Voltage Measurements (Cont)

4. Measure and record the voltage across the main line fuses in the generator. Single phase units

will only use one set of voltage measurements.

L1 phase to L2 phase: ______ VAC. L1 phase to ground: ______ VAC. L1 phase to L3 phase: ______ VAC. L2 phase to ground: ______ VAC. L2 phase to L3 phase: ______ VAC. L3 phase to ground: ______ VAC.

5. Are these voltages within specification for the unit as per Chapter 1? For 3 phase units, the

phase to ground voltage should be 230 V ± 10 % for 400 V units, and 277 V ± 10 % for 480 V units. For single phase 230 V units, the line to ground voltage should be 115 V ± 10 %.

___ Check.

BEFORE CONTINUING, REFER TO SECTION 2.7.2 OR 2.7.3 FOR THE PROCEDURE FOR ADJUSTING LINE VOLTAGE TAPS ON THE 24 V INTERFACE / CONSOLE POWER TRANSFORMER AND ON THE POWER SUPPLY AUXILIARY TRANSFORMER.

6. Switch OFF the mains power to the generator. Verify that there is no line voltage present, then

replace the safety cover on the main input fuse block. 7. Switch the mains power ON again, then switch the generator ON at the console. 8. Verify that the red LED (DC BUSS OK) located on the power input board in the main cabinet is lit. ___ Check. 9. Switch off the generator, then switch back ON after approximately 10 seconds. Observe the

startup sequence on the console APR display. • The generator type / model may be briefly displayed. • The software revision will be shown while the console runs a series of diagnostic tests.

10. Verify that the following LED’s on the auxiliary board in the main cabinet are lit:

D2 CNTCTR CLSD D3 S.S. OK D29 +12V D30 -12V D31 +/- 35V

Page 2-14 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 41: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.7.2 Single Phase Primary Tap Selection

THIS SECTION APPLIES TO SINGLE PHASE GENERATORS ONLY. ENSURE THAT THE AC MAINS IS SWITCHED OFF AND LOCKED OUT, AND THAT ALL CAPACITORS ARE DISCHARGED BEFORE PROCEEDING. 1. Note the position of the 208 / 240V tap on the power supply auxiliary transformer (this is factory

set to 240 VAC). Refer to figure 2-6, this shows the locations of the transformer taps. 2. Based on the line voltage measured in section 2.7.1, step 4, set the tap referenced in the

previous step as follows. Refer to figure 2-7, this shows the line voltage taps schematically. • Use the 208V tap if the line voltage is 215 VAC or less. • Use the 240V tap if the line voltage is 216 VAC or higher.

Figure 2-7: Schematic, primary of power supply aux transformer

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-15

Page 42: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc. 2.7.2 Single Phase Primary Tap Selection (Cont)

3. Note the position of the 208 / 230V tap on the 24V interface / console power transformer (this is

factory set to 230 VAC). Refer to figure 2-8, this shows the transformer as viewed from the left side of the generator.

4. Based on the line voltage measured in section 2.7.1, step 4, set the 24V interface / console

power transformer primary voltage tap as follows: • Black to pin 4 (208V tap) if the line voltage is 215 VAC or less. • Black to pin 6 (230V tap) if the line voltage is 216 VAC or higher.

Figure 2-8: 24V interface / console power transformer, terminal view 2.7.3 Three Phase Primary Tap Selection

THIS SECTION APPLIES TO 400 VAC THREE PHASE GENERATORS ONLY. NO TAP CHANGE IS NEEDED ON 480 VAC UNITS. ENSURE THAT THE AC MAINS IS SWITCHED OFF AND LOCKED OUT, AND THAT ALL CAPACITORS ARE DISCHARGED BEFORE PROCEEDING. 1. Note the position of the 380 / 400V tap on the power supply auxiliary transformer (this is factory

set to 400 VAC). Refer to figure 2-6, this shows the locations of the transformer taps. 2. Based on the line voltage measured in section 2.7.1, step 4, set the tap referenced in the

previous step as follows. Refer to figure 2-7, this shows the line voltage taps schematically. • Use the 380V tap if the line voltage is 389 VAC or less. • Use the 400V tap if the line voltage is 390 VAC or higher.

Page 2-16 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 43: Indico 100L Service Manual - 740914

CPI Canada Inc. Installation 2 2.8.0 TUBE MA AUTO CALIBRATION

It is recommended that the generator be tested at this point with only the rotor and high tension cables connected. The generator should be able to complete an X-ray tube calibration and seasoning cycle without other equipment connected to the generator (other than the basic interlocks as noted below). This will allow for easier fault isolation as each section of the system is connected and tested.

Before being able to make X-ray exposures, the room door interlock must be closed and the thermal switch must be closed.

Prior to beginning tube auto calibration (per the procedure in chapter 3), the tube(s) used in this installation must be properly selected, and the generator kV and kW limits should be programmed. Refer to chapter 3. It is recommended that the tube(s) be conditioned (seasoned) before beginning tube auto calibration; refer to chapter 6.

2.9.0 FINAL CHECKS

The room interface connections may now be completed. These items are described in 2.5.8. • When finished all wiring, check that all connections are tight and secure. • Check that all cables are dressed neatly inside the main cabinet, and secured as necessary. • Reconnect any grounds that have been removed from covers. Then reinstall all covers before placing

the generator into service.

NOTE: THE INSTALLER SHOULD ENSURE THAT ALL CABLE CONNECTIONS TO THE GENERATOR ARE SECURE, AND ALL CABLES EXTERNAL TO THE GENERATOR ARE ADEQUATELY PROTECTED AGAINST ACCIDENTAL DISCONNECTION.

2.10.0 UTILITY SOFTWARE

The optional CPI utility software allows for data communication between the console and an external computer. This provides a “virtual storage oscilloscope” that displays key waveforms and displays data not otherwise available, and allows uploading and downloading of EEPROM, tube, and APR data. It also allows editing of film / screen names and APR text. A computer with an RS-232 port (i.e. laptop) and a serial cable are required to run the utility software and to interface to the console. The serial cable to connect to the console is a special cable that must be ordered from the factory. The part number of this cable to order is 735184-00.

X-Ray Generator Service Manual Ch # 740914-02 Rev. P Page 2-17

Page 44: Indico 100L Service Manual - 740914

2 Installation CPI Canada Inc.

(This page intentionally left blank)

Page 2-18 Rev. P X-Ray Generator Service Manual Ch # 740914-02

Page 45: Indico 100L Service Manual - 740914

CPI Canada Inc X-Ray Tube Stator Compatibility Tables

SUPPLEMENT

X-RAY TUBE STATOR COMPATIBILITY TABLES

CONTENTS:

1.0 INTRODUCTION............................................................................................................................................. 2 2.0 LOW SPEED STARTER TUBE SELECT TABLE........................................................................................... 3 3.0 DUAL SPEED STARTER TUBE SELECT TABLE ......................................................................................... 7

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 1

Page 46: Indico 100L Service Manual - 740914

X-Ray Tube Stator Compatibility Tables CPI Canada Inc 1.0 INTRODUCTION

This supplement contains the X-Ray Tube Stator Compatibility Tables for the Low Speed Starter and the Dual Speed Starter. The Dual Speed Starter table does not apply to the Indico 100L model.

Page 2 Rev. H X-Ray Generator Service Manual Supplement 746026-00

Page 47: Indico 100L Service Manual - 740914

CPI Canada Inc X-Ray Tube Stator Compatibility Tables 2.0 LOW SPEED STARTER TUBE SELECT TABLE

TABLE 1: TUBE TYPES (LOW SPEED STARTER) TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

RUN VOLTAGE

BOOST TIME

SHIFT CAPAC

LOW SPD STARTER PART NO.

CGR Statorix 200A/240/260 550 (50/110 Ω 24 slot)

MN640 M641 MN641 M643 MS740 MSN742 RN620 RN640 RSN722 RSN742

94 VAC 2.5 Sec 12.5 µF 732752-01

Chirana Rotax KA 125 20/50 Ω stator

RIK-T 12/50 0.8/2.0

73 VAC 1.5 Sec 33 µF 732752-00

Chirana Rotax KA 125 20/20/20 Ω stator

RIK-T 30/50 1.2/2.0

52 VAC 1.5 Sec 100 µF 52 mH Inductor

732752-04

Comet DO7 25/50 Ω stator

DX7 52 VAC 1.5 Sec 33 µF 732752-00

Comet DO9 25/50 Ω stator

DX9 0.6/2.0 DX9 1.2/2.0

52 VAC 1.5 Sec 33 µF 732752-00

Comet DO9 25/50 Ω stator

DX91HS DX92HS DX93HS

52 VAC 1.5 Sec 33 µF 732752-00

Comet DO10 25/50 Ω stator

DX10HS 0.6/1.0 DX10HS 1.0/2.0

52 VAC 1.5 Sec 33 µF 732752-00

Comet DX10 25/50 Ω stator

DX101HS DX104HS DX105HS

52 VAC 1.5 Sec 33 µF 732752-00

Comet XSTAR8 XSTAR74 25/50 Ω stator

XST-8 XST-74

52 VAC 1.5 Sec 33 µF 732752-00

Dunlee PX1300 3” anode “S” stator (15/30 Ω)

PX1302 PX1312

52 VAC 2.5 Sec 33 µF 732752-00

Dunlee PX1400 4” anode “S” stator (15/30 Ω)

PX1429 PX1431 PX1436 PX1482 PX1483

52 VAC 2.5 Sec 33 µF 732752-00

GE Maxiray 75 3” anode 23/23 Ω equal impedance “E” stator

0.6/1.0 11° 0.6/1.5 15° 1.0/2.0 15°

73 VAC 1.5 Sec 45 µF 732752-02

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 3

Page 48: Indico 100L Service Manual - 740914

X-Ray Tube Stator Compatibility Tables CPI Canada Inc 2.0 LOW SPEED STARTER TUBE SELECT TABLE (Cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

RUN VOLTAGE

BOOST TIME

SHIFT CAPAC

LOW SPD STARTER PART NO.

GE Maxiray 100 4” anode 23/23 Ω equal impedance “E” stator

0.3/1.0 11° 0.6/1.0 11° 0.6/1.2 11° 0.6/1.5 11° 0.6/1.5 12° 0.6/1.25 12.5° 1.0/2.0 15°

73 VAC 1.5 Sec 45 µF 732752-02

Gilardoni Rotagil S/AS

AR1130 0.6/1.2 AR2050 0.6/1.2 AR3000-1 1.0/2.0 AR3060 1.0/2.0 AR30100 0.6/2.0 AR40100 0.6/1.3 AR9000-1 0.6/1.3 AR9000-2 0.6/1.3

52 VAC 1.5 Sec 33 µF 732752-00

IAE C52 Super C100 20/40 Ω stator

RTC 600HS RTC 700HS RTC1000HS RTM 101H/HS RTM 102H/HS

52 VAC 1.5 Sec 33 µF 732752-00

IAE C52 C352 C52 Super 20/40 Ω stator

RTM 90H/HS RTM 92H/HS

52 VAC 1.5 Sec 33 µF 732752-00

IAE C52 C352 20/40 Ω stator

RTM 78H/HS AP DX104 X40 X50/AH

52 VAC 1.5 Sec 33 µF 732752-00

Philips ROT350 ROT351

RO 12/30 (0.6/1.2) RO 17/50 (0.6/1.3) RO 20/50 (0.6/1.2) RO 30/50 (1.2/1.8)

52 VAC 1.5 Sec 33 µF 732752-00

Siemens Opti-150 “S” stator (14/37 Ω)

30/52R 20/40

52 VAC 2.5 Sec 45 µF 732752-02

Toshiba Rotanode XH-106V-2 XS-RB stator

E7252X E7813X

52 VAC 1.5 Sec 33 µF 732752-00

52 VAC 1.5 Sec 33 µF 732752-00 Toshiba Rotanode XH-106V-2 XS-AL stator

E7239X E7242X E7252X E7813X

Note: The starter boost voltage is 120 VAC. This is only available with one-tube generators.

Page 4 Rev. H X-Ray Generator Service Manual Supplement 746026-00

Page 49: Indico 100L Service Manual - 740914

CPI Canada Inc X-Ray Tube Stator Compatibility Tables 2.0 LOW SPEED STARTER TUBE SELECT TABLE (Cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

RUN VOLTAGE

BOOST TIME

SHIFT CAPAC

LOW SPD STARTER PART NO.

Toshiba Rotanode XH-121 XS-RB stator

E7132X E7239X E7240X E7299X

52 VAC 1.5 Sec 33 µF 732752-00

Toshiba Rotanode XH-126

E7242FX 52 VAC 1.5 Sec 33 µF 732752-00

Toshiba XH-157

E7254X E7254FX E7255X

73 VAC 1.5 Sec 33 µF 732752-00

Varian B100 DX52 Std “R”stator (16/50 Ω)

A102 A132 A142

52 VAC 1.5 Sec 33 µF 732752-00

Varian B100 “STD” stator (16/50 Ω)

A102 A132 A142

52 VAC 1.5 Sec 33 µF 732752-00

Varian B130 B150 Std “R”stator (16/50 Ω)

A192 A272 A282 A286 A292 G256 G292

52 VAC 1.5 Sec 33 µF 732752-00

Varian Diamond Std “R”stator (16/50 Ω)

RAD13 RAD14 0.3/1.2 RAD14 0.6/1.2 RAD14 0.6/1.5

52 VAC 1.5 Sec 33 µF 732752-00

Varian Emerald Std “R” stator

RAD 8 RAD 68 RAD 74

52 VAC 1.5 Sec 33 µF 732752-00

Varian Saphire Std “R” stator

RAD21 RAD56 0.6/1.0 RAD56 0.6/1.2 RAD60 RAD92 RAD94

52 VAC 1.5 Sec 33 µF 732752-00

Varian DX62 300-400 kHu, “STD” stator (23/56 Ω)

A192B A197 A256 A272 A282 A286 A292

52 VAC 1.5 Sec 33 µF 732752-00

Varian DX62U Universal 300-400 kHu, configured as “STD” or “R” stator (15/36 Ω)

A192B A197 A256 A272 A282 A286 A292

52 VAC 1.5 Sec 33 µF 732752-00

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 5

Page 50: Indico 100L Service Manual - 740914

X-Ray Tube Stator Compatibility Tables CPI Canada Inc

(This page intentionally left blank)

Page 6 Rev. H X-Ray Generator Service Manual Supplement 746026-00

Page 51: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 7

3.0 DUAL SPEED STARTER TUBE SELECT TABLE

The Dual Speed Starter synthesizes its output frequencies independently of the line frequency and will operate all tubes at 60 and 180 Hz, unless a particular tube only has published ratings to operate at 150 Hz, as indicated in the table. In this case, the starter will output 150 Hz when set for these tubes.

TABLE 2: TUBE TYPES (HIGH SPEED STARTER) TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

CGR Statorix 200A/240/260 550 (50/110 Ω 24 slot)

MN640 M641 MN641 M643 MS740 MSN742 RN620 RN640 RSN722 RSN742

00111 450 150 120 3.0 sec 3.0 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table)

3.4 uF N/A 733317-07 * 735925-07

Comet DO7 25/50 Ω stator

DX7 00000 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Comet DO9 25/50 Ω stator

DX9 0.6/2.0 DX9 1.2/2.0

00000 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Comet DO10 25/50 Ω stator

DI104 0.3/0.8 DI104 0.6/1.0 DI104 0.6/1.3 DI104 0.6/1.8

00011 420 80 150 3.0 sec 1.8 sec 240 80 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Page 52: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 8

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

Comet DO10 25/50 Ω stator

DI106 0.3/0.8 DI106 0.6/1.0 DI106 0.6/1.3

10011 420 80 150 3.0 sec 2.2 sec 240 80 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Comet DO700WX 25/50 Ω stator

DI700 0.6/1.0 DI700 1.0/1.8

10011 420 80 150 3.0 sec 2.2 sec 240 80 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Dunlee PX1300 3” anode “S“ stator (15/30 Ω)

PX1302 PX1312

11100 240 120 100 3.0 sec 2.3 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Dunlee PX1400 4” anode “S” stator (15/30 Ω) (see note 2 at end of this table)

PX1429 PX1431 PX1436 PX1482 PX1483

01100 240 120 100 3.0 sec 4.5 sec 240 70 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Dunlee PX1400 4” anode “S” stator (15/30 Ω) (see note 2 at end of this table)

PX1429 PX1431 PX1436 PX1482

10000 400 120 100 3.0 sec 1.9 sec 240 70 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Dunlee PX1400 4” anode “Q” stator (6/12 Ω)

PX1429 PX1431 PX1436 PX1482 PX1483

10110 340 60 100 3.0 sec 1.0 sec 240 70 20 uF 60 uF 733317-02 * 735925-02

Page 53: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 9

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

GE Maxiray 75 (3” anode) 23/23Ω equal impedance “E” stator

1.0/2.0 15° 0.6/1.0 11°

01110 400 90 80 2.0 sec 0.9 sec 230 70 7.5 uF 47 uF 733317-03 * 735925-03 (See note 5 at end of this table)

GE Maxiray 100 (4” anode) 23/23 Ω equal impedance “E” stator

0.3/1.0 11° 0.6/1.0 11° 0.6/1.2 11° 0.6/1.5 11° 0.6/1.25 12.5° 1.0/2.0 15°

00101 400 90 80 3.0 sec 1.0 sec 230 70 7.5 uF 47 uF 733317-03 * 735925-03 (See note 5 at end of this table)

Gilardoni Rotagil

AR11-30 AR30-60

10101 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 220 50 Hz

60 50 Hz

6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Gilardoni Rotagil A/A5

AR20-50 AR30-100 AR40-100

01101 340150 Hz

60 150 Hz

80 3.0 sec 1.4 sec 220 50 Hz

60 50 Hz

6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

IAE C52 Super C100 20/40 Ω stator

RTC 600HS RTC 700HS RTC 1000HS RTM 101H/HS RTM 102H/HS

11011 440 100 120 1.8 sec 1.0 sec 220 80 5 uF 45 uF 733317-05 * 735925-05 (See note 5 at end of this table)

IAE C52 C352 C52 Super 20/40 Ω stator

RTM 90H/HS RTM 92H/HS

11011 440 100 120 1.8 sec 1.0 sec 220 80 5 uF 45 uF 733317-05 * 735925-05 (See note 5 at end of this table)

Page 54: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 10

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

IAE C52 C352 20/40 Ω stator

RTM 78H/HS AP DX104 X40 X50/AH

11011 440 100 120 1.8 sec 1.0 sec 220 80 5 uF 45 uF 733317-05 * 735925-05 (See note 5 at end of this table)

IAE tubes AND standard “R” stator tubes

Dual speed starters 733317-08 / 735925-08 are special configurations. These may be jumper-configured to support IAE tubes that require dual speed starter 733317-05 / 735925-05 per this table, or to support “R” type stators that use dual speed starter 733317-01 / 735925-01 per this table. Please refer to note 3 at the end of this section before attempting to use this dual speed starter configuration.

5 uF 6 uF

30 uF 31 uF

733317-08 * 735925-08

Philips ROT350 ROT351

RO 12/30 (0.6/1.2) RO 17/50 (0.6/1.3) RO 20/50 (0.6/1.2) RO 30/50 (1.2/1.8)

00000 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Philips ROT350 ROT351

SRO 25/50 (0.6/1.0) SRO 33/100 (0.6/1.2)

00000 400 100 100 3.0 sec 1.4 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table)

6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Philips ROT350 ROT351

RO 20/50 (0.6/1.2)

00000 400 100 100 3.0 sec 1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Page 55: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 11

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

Philips ROT500

SRM 03/10 (0.3/1.0)

01111 420 50 150 3.0 sec 1.8 sec 240 50 12.5 uF 37.5 uF 733317-10 * 735925-10

Philips ROT500

SRM 10/80 (0.3/1.0)

01111 420 50 150 3.0 sec 1.8 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table)

12.5 uF 37.5 uF 733317-10 * 735925-10

Siemens Biangulix 8500 RPM (configured for 150 Hz operation)

BI 125/20/40 11101 400 150 Hz

90 150 Hz

80 3.0 sec 1.9 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table)

5 uF N/A 733317-04

Siemens Opti-150 “S” stator

30/52R 20/40

00000 400 90 80 3.0 sec 1.9 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table) 240 50

5 uF 45 uF 733317-05 * 735925-05 (See note 5 at end of this table)

Siemens OptiTop 150/40/80HC-100L 150/40/80HC-102L 14/37 Ω stator

01011 400 80 120 3.0 sec 1.0 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table) 240 80

5 uF 30 uF 733317-06 * 735925-06 (See note 5 at end of this table)

Siemens Optilix SV150/40/80C-100L SV150/30/50C-100L

01011 400 80 120 3.0 sec 1.0 sec HIGH SPEED OPERATION ONLY (See note 1 at end of this table) 240 80

5 uF 30 uF 733317-06 * 735925-06 (See note 5 at end of this table)

Siemens RAY12S_1, RAY14S_1

11111 400 50 150 3.0 sec 1.4 sec 240 50 5 uF 30 uF 733317-13 * 735925-13 (See note 4 at end of this table)

Page 56: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 12

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

Toshiba XH-106V-2

E7252X E7813X

00100 400 100 100 3.0 sec 1.2 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Toshiba XH-106V-2 XS-AL stator

E7239X E7242X E7252X E7813X

10111 220 60 80 3.0 sec 1.2 sec 130 50 6 uF 46 uF 733317-09 * 735925-09 (See note 5 at end of this table)

Toshiba XH-121

E7132X E7239X E7240X E7299X

00000 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Toshiba XH-126

E7242FX 00000 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Toshiba XH-157

E7254X E7254FX E7255X

10011 420 80 150 3.0 sec 2.2 sec 240 80 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Page 57: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 13

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

Varian B100 DX52 Std “R” stator (16/50 Ω)

A102 A132 A142

01000 400 100 100 3.0 sec 1.0 sec 240 60 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Varian B100 “Q” stator (6/11 Ω)

A102 A132 A142

11010 290 70 60 3.0 sec 0.8 sec 150 50 20 uF 60 uF 733317-02 * 735925-02

Varian B130 B150 Std “R” stator (16/50 Ω)

A192 A272 A282 A286 A292 G256 G292

00000 400 100 100 3.0 sec 1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Varian B130 B150 “Q” stator (6/11 Ω)

A192 A272 A282 A286 A292 G256 G292

00110 290 70 60 3.0 sec 1.3 sec 150 50 20 uF 60 uF 733317-02 * 735925-02

Varian B160 B180 “R” Stator (16/50 Ω)

G1082 G1086 G1092 G1582 G1592

11110 400 100 150 3.0 sec 5.0 sec 240 70 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Page 58: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 14

3.0 Dual Speed Starter Tube Select Table (cont)

TUBE TYPE (HOUSING)

TUBE TYPE (INSERT)

CODE Switches 1........5

H.S. START VOLTS

H.S. RUN

VOLTS

H.S. BRAKE VOLTS

H.S. BRAKE

TIME

BOOST TIME

L.S. START VOLTS

L.S. RUN

VOLTS

H.S. SHIFT

CAPAC

L.S. SHIFT

CAPAC

DUAL SPD STARTER PART NO.

Varian Diamond Std “R” stator

RAD13 RAD14 0.3/1.2 RAD14 0.6/1.2 RAD14 0.6/1.5

00100 400 100 100 3.0 sec 1.2 sec 240 50 6 uF 31 uF 733317-01 * 735925-01(See note 5 at end of this table)

Varian Emerald Std “R” stator

RAD 8 RAD 68 RAD 74

00000 LOW SPEED OPERATION ONLY (See note 1 at end of this table)

1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Varian Saphire Std “R” stator (20/50 Ω)

RAD21 RAD56 0.6/1.0 RAD56 0.6/1.2 RAD60 RAD92 RAD94

10100 400 100 100 3.0 sec 2.3 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Varian DX62 300-400 kHu, “STD” stator (23/56 Ω)

A192B A197 A256 A272 A282 A286 A292

00000 400 100 100 3.0 sec 1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Varian DX62U Universal 300-400 kHu, configured as “STD” or “R” stator (15/36 Ω)

A192B A197 A256 A272 A282 A286 A292

00000 400 100 100 3.0 sec 1.4 sec 240 50 6 uF 31 uF 733317-01 * 735925-01 (See note 5 at end of this table)

Page 59: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 15

3.0 Dual Speed Starter Tube Select Table (cont)

THE DUAL SPEED STARTER USES MODULATION STRATEGIES TO OBTAIN THE DESIRED OUTPUTS. MEASURED VOLTAGES MAY NOT AGREE WITH THOSE LISTED IN THE TABLE. HOWEVER, THE CURRENTS FLOWING IN THE STATOR WINDINGS ARE EQUIVALENT TO THOSE THAT WOULD EXIST IF THE STATOR WAS EXCITED WITH THESE VOLTAGES.

* Dual speed starter part numbers 733317-XX are used in 400 VAC 3φ generators / power supplies and in 480 VAC 3φ generators / power supplies with a line

adjusting transformer. Dual speed starter part numbers 735925-XX are used in 230 VAC 1φ generators / power supplies and in 480 VAC 3φ direct input generators / power supplies (using no line adjusting transformer).

NOTE 1: Tube types designated as low speed only or high speed only must be programmed for low speed only or high speed only operation.

Refer to chapter 3, the section Tube Selection for details (Indico 100 only). For some tubes listed as high speed only, the starter may be capable of low speed operation but the Manufacturers data sheet for the Insert lists high speed only and therefore the low speed parameters that affect the Housing/Stator are shown in Grey text.

NOTE 2: Two settings are shown for the Dunlee PX1400 tube with “S” stator. The settings with the 4.5 sec boost times are per older Dunlee data sheets. The values

with the 1.9 sec boost times are per newer Dunlee data sheets (May 2002). The installer must verify tube compatibility per the appropriate Dunlee tube data sheet.

NOTE 3: Dual speed starters 733317-08 / 735925-08 have a jumper selectable phase shift capacitor that allows selection of 5 uF and 30 uF capacitors for operation

with IAE tubes that use dual speed starter 733317-05 / 735925-05 as per table 2, or this may be set to 6 uF and 31 uF for operation with “R” type stators that use dual speed starter 733317-01 / 735925-01 as per table 2. THIS STARTER IS TYPICALLY FACTORY SET TO THE 5 uF AND 30 uF POSITION. CONFIRM PROPER CONFIGURATION PER THE SECTION “CONFIGURING DUAL SPEED STARTER 733317-08 / 735925-08” IN CHAPTER 2 BEFORE PROCEEDING.

NOTE 4: Dual Speed Starters 733317-13 / 735925-13 are special configurations. These may be programmed to support IAE tubes that require Dual Speed Starter

733317-05 / 735925-05 per this table, or to support ‘R’ type stators that use Dual Speed Starter 733317-01 / 735925-01 per this table. This configuration supports tubes requiring a 5 or 6uF High Speed Starter capacitor. Capacitor selection is made by relays located on the Dual Speed

Starter.

Page 60: Indico 100L Service Manual - 740914

CPI Canada Inc. X-Ray Tube Stator Compatibility Tables

X-Ray Generator Service Manual Supplement 746026-00 Rev. H Page 16

3.0 Dual Speed Starter Tube Select Table (cont) NOTE 5: New generator configurations may substitute the dual speed starter part numbers listed in table 2 with the alternate versions per the table below:

LISTED TYPE ALTERNATE 733317-01, 735925-01 733317-08, 12, 13, 735925-08, 12, 13 733317-03, 735925-03 733317-12, 735925-12 733317-05, 735925-05 733317-13, 735925-13 733317-06, 735925-06 733317-13, 735925-13 733317-08, 735925-08 733317-13, 735925-13 733317-09, 735925-09 733317-12, 13, 735925-12, 13

Page 61: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3

CHAPTER 3

INTERFACING, PROGRAMMING, AND CALIBRATION

CONTENTS:

3.1.0 INTRODUCTION....................................................................................................................................................3-2 3.2.0 WIRING TO INPUTS & OUTPUTS........................................................................................................................3-2

3.2.1 Inputs ..................................................................................................................................................................3-2 3.2.2 Outputs ...............................................................................................................................................................3-4 3.2.3 Inputs & Outputs (simplified schematic) .............................................................................................................3-5 3.2.4 AEC Interconnect ...............................................................................................................................................3-6 3.2.5 AEC Left / Right (Field A / Field C) Reversal .....................................................................................................3-7

3.3.0 GENERATOR PROGRAMMING ...........................................................................................................................3-8 3.3.1 Generator Utility Functions (overview) ...............................................................................................................3-8

3.4.0 LEVEL 1 UTILITIES ...............................................................................................................................................3-8 3.4.1 APR Save ...........................................................................................................................................................3-9 3.4.2. Date and Time .................................................................................................................................................3-10 3.4.3. Exit...................................................................................................................................................................3-10 3.4.4 Utilities ..............................................................................................................................................................3-10

3.5.0 LEVEL 2 UTILITIES .............................................................................................................................................3-10 3.5.1 Calibration ........................................................................................................................................................3-12 3.5.2 AEC Adjustment ...............................................................................................................................................3-12 3.5.3 Focus Disable...................................................................................................................................................3-12 3.5.4 Exp. Counter.....................................................................................................................................................3-13 3.5.5 LCD mode ........................................................................................................................................................3-13 3.5.6 Configuration ....................................................................................................................................................3-13 3.5.7 Tube Selection..................................................................................................................................................3-14 3.5.8 Receptor Conf. .................................................................................................................................................3-15

3.6.0 HARDWARE PROGRAMMING...........................................................................................................................3-16 3.6.1 Speaker Volume ...............................................................................................................................................3-16 3.6.2 LCD Backlight Brightness adjustment..............................................................................................................3-16

3.7.0 AEC PROGRAMMING AND CALIBRATION.......................................................................................................3-17 3.7.1 AEC Limitations: Minimum and Maximum Exposure Times ............................................................................3-17 3.7.2 Precalibration Notes .........................................................................................................................................3-17 3.7.3 Navigating the AEC Adjustment menus ...........................................................................................................3-18 3.7.4 AEC Calibration (Table Bucky).........................................................................................................................3-18 3.7.5 AEC Calibration (Wall Bucky)...........................................................................................................................3-23

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-1

Page 62: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.1.0 INTRODUCTION

This Chapter describes the interfacing of the generator to the Bucky(s), interlocks, collimator lamp and tube stand locks, and to the AEC chambers if the AEC option is fitted. Programming and AEC calibration of the generator is also described in this Chapter.

3.2.0 WIRING TO INPUTS & OUTPUTS

NOTE: THE INSTALLER MUST PROVIDE THE NECESSARY INTERFACING CABLES FOR WIRING TO THE GENERATOR INPUTS AND OUTPUTS DESCRIBED IN THIS SECTION.

WARNING: LINE VOLTAGE IS PRESENT INSIDE THE GENERATOR AT ALL TIMES THAT THE MAIN DISCONNECT IS SWITCHED ON. FOR SAFETY, THE MAIN DISCONNECT SHOULD BE SWITCHED OFF AND LOCKED OUT WHILE CONNECTING ROOM EQUIPMENT.

3.2.1 Inputs The Bucky inputs, S.I.D. interlocks, room door interlock, and thermal switch inputs are opto coupled. This means that a relay contact, transistor, or other low impedance (≤ 100 Ω, 10 mA current rating) switching device must be connected across each of those inputs. Table 3-1 below defines the pinouts, polarity at the terminals and the logic condition required for that input. If using a directional switching device, such as a transistor, the polarity of the voltage seen by the switching device must be observed. This is shown in the table below. Refer also to figure 3-1. This is a pictorial drawing of the Bucky inputs and outputs, the interlock inputs, and the thermal switch inputs. Refer to 3.2.4 for details on AEC connections.

WARNING: DO NOT DEFEAT THE 40″ OR 72″ S.I.D. INTERLOCKS, OR THE ROOM DOOR INTERLOCK UNLESS THE CORRESPONDING DEVICE IS NOT PRESENT. CONSULT THE APPLICABLE REGULATIONS BEFORE BYPASSING ANY EXPOSURE INTERLOCKS. DO NOT VIOLATE ANY REGULATIONS FOR X-RAY SAFETY.

NEVER BYPASS THE X-RAY TUBE THERMAL SWITCH INTERLOCK. OBSERVE IONIZING RADIATION PERSONAL PROTECTION AT ALL TIMES.

Page 3-2 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 63: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.2.1 Inputs (Cont)

NOTE: IN THIS MANUAL, THE WALL BUCKY IS REFERRED TO AS BUCKY 1, AND THE TABLE BUCKY IS BUCKY 2.

CUSTOMER INTERFACE

BOARD DEFINITION

J2-5 (+) J2-6 (-)

B1: B2:

Bucky 1 ready. A contact closure from pin 5 to 6 indicates BUCKY 1 READY.

J3-5 (+) J3-6 (-)

B1: B2:

Bucky 2 ready. A contact closure from pin 5 to 6 indicates BUCKY 2 READY.

J3-1 (+) J3-2 (-)

Door interlock. A contact closure from pin 1 to 2 is required to make an exposure. This indicates that the room door is closed.

J2-7 (+) J2-8 (-)

40 inch (100 cm) S.I.D. interlock. A contact closure from pin 7 to 8 indicates the tube is at a 40 inch (100 cm) S.I.D.

J2-9 (+) J2-10 (-)

72 inch (180 cm) S.I.D. interlock. A contact closure from pin 9 to 10 indicates the tube is at a 72 inch (180 cm) S.I.D.

J3-4 (+) J3-3 (-)

Thermal switch. A contact closure from pin 4 to 3 is required to make an exposure. This indicates normal operation of the X-ray tube.

Table 3-1: Generator inputs

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-3

Page 64: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.2.2 Outputs

Table 3-2 shows the Bucky and auxiliary power outputs from the generator. Refer also to figure 3-1.

CUSTOMER INTERFACE

BOARD DEFINITION

J2-1 B4: 110 VAC return (Bucky 1). J2-2 B6: Ground (Bucky 1). J2-3 B8: 110 VAC out (Bucky 1). J2-4 B3: 110 VAC out (Bucky 1 start). J3-7 B4: 110 VAC return (Bucky 2). J3-8 B6: Ground (Bucky 2). J3-9 B8: 110 VAC out (Bucky 2). J3-10 B3: 110 VAC out (Bucky 2 start). J4-1 J4-2

24 VAC @ 150 watts output for collimator light (unswitched). This output is rated for intermittent (collimator lamp) duty only, <10% duty cycle.

J4-5 (+) J4-6 (-)

24 VDC @ 45 watts output for tube stand locks (unswitched).

Table 3-2: Generator outputs

IN SOME INSTALLATIONS, BUCKY START RELAYS K1 OR K2 ON THE CUSTOMER INTERFACE BOARD MAY DRIVE THE INPUTS OF OPTO COUPLERS OR OTHER LOW CURRENT DEVICES ON THE BUCKY. THE LEAKAGE CURRENT THROUGH THE R-C SNUBBER THAT IS CONNECTED ACROSS THE RELAY CONTACTS (C2 AND R2 FOR K1, C11 AND R9 FOR K2) MAY BE SUFFICIENT TO ENERGIZE THE BUCKY INPUTS WHEN THE RELAYS ARE OPEN.

IF THIS IS EXPERIENCED, THE SNUBBER SHOULD BE DEFEATED. REMOVING THE SERIES RESISTOR, PART OF THE R-C SNUBBER, WILL ELIMINATE THIS PROBLEM.

Removing R2 open-circuits the snubber across the K1 contacts, removing R9 defeats the snubber across the contacts of K2.

Page 3-4 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 65: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.2.3 Inputs & Outputs (simplified schematic)

1

1

7

1

AC RETURN

FROM 24 VACSOURCE

FROM 24 VDCSOURCE

AC RETURN

+24 VDC

+24 VDC

+5 VDC

+5 VDC+24 VDC

+24 VDC

+24 VDC

+24 VDC

+24 VDC

+24 VDC

BUCKY 1 START

BUCKY 2 START

BUCKY 1 STARTCOMMAND

BUCKY 2 STARTCOMMAND

BUCKY 1 READY

BUCKY 2 READY

DOOR INTERLOCK

40" S.I.D. INTERLOCK

72" S.I.D. INTERLOCK

THERMAL SWITCH

110 VAC

110 VAC

BUCKY 1OUTPUTS

BUCKY 2OUTPUTS

40" SIDINTERLOCK *

BUCKY 1READY *

BUCKY 2READY *

72" SIDINTERLOCK *

THERMALSWITCH *

DOORINTERLOCK *

24 VAC OUTFOR COLLIMATOR

LAMP

24 VDC OUTFOR TUBE

STAND LOCKS

J2

J3

J4

2

2

8

2

3

5

9

7

4

6

10

8

5

5

9

4

6

6

10

3

FILE: SM_IO.CDR

+-

REFER TO TABLE3-1 FOR REQUIRED

INPUT LOGIC LEVELS*

Figure 3-1: J2, J3, J4 inputs and outputs (customer interface board)

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-5

Page 66: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.2.4 AEC Interconnect

NOTE: IF YOUR GENERATOR USES “CAEC” AEC CHAMBER(S), REFER TO SEPARATE SUPPLEMENT AT FRONT OF THIS MANUAL IN CONJUNCTION WITH THIS MANUAL.

Table 3-3 and figure 3-2 show the functions and signals at each pin of the AEC connectors J1 and J2 on the AEC board in the console. The AEC connections are directly compatible with ion chambers as per the compatibility statement / product description in chapter 1.

AEC BOARD (J1 TABLE, J2

WALL) DEFINITION

Pin 1 No connection. Pin 2 Field select A. This is configurable to select field C if required,

refer to 3.2.5. Pin 3 Field select B. Pin 4 Reset / start. Pin 5 AEC input signal from chamber. Pin 6 Field select C. This is configurable to select field A if required,

refer to 3.2.5. Pin 7 -15 VDC to chamber. Pin 8 +15 VDC to chamber. Pin 9 Ground.

Table 3-3: AEC board inputs / outputs

FLD A/C

FLD B

FLD C/A

RESET (START)

AEC RAMP

FILE: SM_AEC.CDR

-

+

123456789

+15 VDC

-15 VDC

TO AECCHAMBER

JW4

JW6

JW3

JW5N/CDRIVERCIRCUIT

Figure 3-2: AEC interconnect (simplified schematic)

Page 3-6 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 67: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.2.5 AEC Left / Right (Field A / Field C) Reversal

For normal AEC chamber orientation, the left field is designated field A and the right field is designated field C. This convention is reversed if the AEC chamber is reversed. Refer to configuration A and configuration B in figure 3-3 and table 3-4. Configuration C applies only to a limited number of AEC chamber models. Refer to the AEC chamber manufacturers literature to determine your chamber type.

Figure 3-3: AEC chamber configurations

CHAMBER MODEL CONFIGURATION Refer to compatibility statement / product description in chapter 1.

Refer to manufacturers data.

Table 3-4: Ion chamber compatibility and identification

The AEC board must be configured such that the left and right field select signals correctly select AEC fields A and C. This is done by inserting jumpers on the AEC board per table 3-5. Please confirm that these jumpers are correctly installed to match the chamber configuration in use. BOTH AEC CHAMBERS MUST BE CONFIGURATION A, OR CONFIGURATION B. THE AEC BOARD JUMPERS CONFIGURE THE LEFT / RIGHT (FIELD A / C) SELECTION FOR BOTH AEC CHAMBERS

CONFIGURATION AEC BOARD JUMPERS

Configuration A JW3, JW4 Configuration B JW5, JW6

Table 3-5: AEC chamber identification

NOTE: DO NOT JUMPER ANY COMBINATIONS OTHER THAN JW3 AND JW4, OR JW5 AND JW6. DO NOT REMOVE OR RECONFIGURE ANY OTHER FACTORY INSTALLED JUMPERS ON THE AEC BOARD.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-7

Page 68: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.3.0 GENERATOR PROGRAMMING

Programming of the generator is performed using the operator console. All programming menus are displayed in the LCD display window on the console. The nine “soft key” buttons (1 to 9 in the figure below) are used to navigate through the programming screens and to select and enter values in this section. In this section, SELECT means to press the button adjacent to the desired function shown in the LCD display window. TOGGLE means to press the stated button repeatedly to cycle through the available selections.

Figure 3-4: Programming / calibration mode reference

3.3.1 Generator Utility Functions (overview)

This X-ray generator has two levels of utility functions. The level 1 utilities are available to the operator, and include minor functions such as setting the clock and enabling / disabling the APR editing function. Level two utilities are password protected and include setup and calibration functions.

3.4.0 LEVEL 1 UTILITIES To access the level 1 utilities follow the steps below:

Step Action Result 1. Start with the generator switched OFF. 2. Press and hold the menu button [5] while

pressing the POWER ON button on the console until the console beeps, (approximately 10 seconds).

The generator will go through its start-up sequence, then display the screen shown below. This is the level 1 utility menu. The PREP and X-RAY LEDs are briefly lit during the console start-up sequence. This is a diagnostic test, and allows confirmation that the LEDs are functional.

3. If an error message is presented after the start-up sequence, press OK and / or RESET to access the level 1 utility menu.

Page 3-8 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 69: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.4.0 LEVEL 1 UTILITIES (Cont)

The level 1 utility menu presents the user with 4 options.

APR SAVE OFF EXITDATE & TIME UTILITIES

In most cases, pressing the menu button [5] will return to the previous screen or the previous menu. This will assist in navigating through the screens in this section. The table below shows the functions available within each of the options in the level 1 utility menu. APR SAVE OFF • Enables/disables the ability to make and then

save changes to APR techniques (kV, mA etc, but NOT APR text changes).

DATE & TIME • Allows setting of the date and time. EXIT • Returns to normal operating mode (non setup /

programming mode). UTILITIES • Accesses the level 2 utility functions.

3.4.1 APR Save

Use these steps to enable / disable the ability to make and then save changes to APR techniques: Step Action

1. From the level 1 utility menu select APR SAVE OFF. This is the default when the generator is switched on, the display will show APR SAVE ON if the APR save function has subsequently been changed to ON.

2. In the APR EDIT AND SAVE CONTROL screen, a prompt APR SAVE OFF will display. Press the adjacent button to toggle the selection of APR SAVE OFF or APR SAVE ON. When the desired APR edit mode is selected, press the menu button [5] to return to the level 1 utility menu.

3. With APR SAVE ON, press EXIT to return to the main APR screen. 4. The procedure for making changes to the APR when in the normal operating mode is described

in the operator’s manual. 5. To reset the APR SAVE ON mode, switch the console OFF, then ON again. This will disable

the ability to make further APR changes until the APR SAVE mode is subsequently enabled.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-9

Page 70: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.4.0 LEVEL 1 UTILITIES (Cont)

3.4.2. Date and Time

Use these steps to set the date and time:

Step Action 1. From the level 1 utility menu select DATE & TIME. 2. Press ← or → to select the parameter to change (month, day, year, hour, or minute). The

selection over the caret symbol ^ is the parameter that will be set in the next step. 3. Press UP or DOWN to set the selected parameter. Time is set in 24 hour mode. 4. Repeat steps 2 and 3 to set the remaining parameters.

Note that the time does not increment when in the setup utility. 5. Press the menu button [5] to return to the main level 1 utility screen.

3.4.3. Exit Use this step to access the APR menu. This is the console normal operating mode.

Step Action 1. From the level 1 utility menu select EXIT. The console will enter the normal operating mode.

3.4.4 Utilities Use these steps to access the level 2 utilities:

Step Action 1. From the level 1 utility menu select UTILITIES. 2. Enter the password at the prompt by pressing the button sequence [1], [2], [3], [4]. Refer to

figure 3-4. 3. A security screen will be presented. Select EXIT to return to the main level 1 utility screen.

Select ACCEPT to display the main level 2 utility screen as per 3.5.0.

3.5.0 LEVEL 2 UTILITIES To access the level 2 utilities follow the steps in 3.4.4. The first screen of the level 2 utilities will be displayed.

The level 2 utility menus present the user with 16 options. Screen 1 of the level 2 utilities is shown below:

CALIBRATION EXP. COUNTERAEC ADJUSTMENT [MORE]EXPOSURE LOG FOCUS DISABLE

Page 3-10 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 71: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.5.0 Level 2 Utilities (Cont)

Press [MORE] on the above screen to access screen 2 of the level 2 utilities, shown below:

LCD MODE FILM SPEEDCONFIGURATION [MORE]TUBE SELECTION FILM PROGRAM

Press [MORE] on the above screen to access screen 3 of the level 2 utilities, shown below:

MAS STEP SIZE RECEPTOR CONF. TUBE SEASONING

The table below shows the functions available within the level 2 utility menus. CALIBRATION • Selects the tube auto calibration routine. AEC ADJUSTMENT • Selects AEC calibration mode. EXPOSURE LOG • This function is not available. FOCUS DISABLE • Disables one or both focal spots. EXP. COUNTER • Accesses the exposure counter. [MORE] • Accesses screen 2 of the level 2 utilities. LCD MODE • Selects normal or reverse video for the LCD

display. CONFIGURATION • Allows the AEC function to be enabled or

disabled. Also allows setting of AEC type (if used), setting of maximum kV, setting of filament standby current, selection of stator boost time (prep time), and setting of maximum output power (some models).

TUBE SELECTION • Allows the desired X-ray tube to be selected. Selecting a new X-ray tube erases the tube calibration data.

FILM PROGRAM • This function is not available. FILM SPEED • This function is not available. [MORE] • Accesses screen 3 of the level 2 utilities MAS STEP SIZE • This function is not available. RECEPTOR CONF. • Allows programming of the wall or table

receptors. TUBE SEASONING • This function is not available.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-11

Page 72: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.5.1 Calibration

The calibration function automatically establishes calibration curves relating filament current to tube current (mA) at three or four kV settings, depending on generator type. This allows the generator to predict the filament temperature (filament current) that will be required for any mA value over the full kV operating range. It is recommended that the X-ray tube be seasoned before attempting auto calibration, particularly if the tube has not been used for some time. Refer to Chapter 6 for the procedure to do this.

Use these steps to perform the X-ray tube auto calibration function:

Step Action 1. Ensure that the correct X-ray tube type is selected before beginning X-ray tube auto calibration.

Refer to TUBE SELECTION in this Chapter to select the tube type. Ensure compatibility of the tube stator with the generator before proceeding, and verify all stator connections.

2. From the level 2 utility menus select CALIBRATION. 3. Select the focal spot to calibrate. Start with small (SM FOCUS). 4. Press and hold the X-ray button (or use the optional handswitch) to begin the calibration

procedure. 5. When calibration is finished on the selected focal spot, COMPLETE.. will be displayed. Release

the exposure switch at this point. 6. Select the next focal spot to calibrate and repeat steps 4 and 5. 7. Should the exposure switch be released before auto-calibration is complete, an error message

will be presented. Exposures are not permitted on a tube which has not been calibrated on the selected focal spot.

8. Press the menu button [5] to return to screen 1 of the level 2 utility screens.

3.5.2 AEC Adjustment

Refer to AEC PROGRAMMING AND CALIBRATION at the end of this Chapter for the AEC calibration procedure. Tube auto calibration must be completed before attempting AEC calibration.

Exposure Log This function is not available.

3.5.3 Focus Disable

This function provides the ability to disable focal spots should one of the filaments in the X-ray tube become damaged in the field. If both focal spots are disabled, an error message will be displayed.

Use these steps to disable focal spot(s):

Step Action 1. From the level 2 utility menus select FOCUS DISABLE. 2. Select DISABLE SF to disable the small focus. Select DISABLE LF to disable the large focus.

The ENABLED / DISABLED status of each focal spot is shown on the left side of the LCD display during this programming step, and also during start up of the generator.

3. Press the menu button [5] to return to screen 1 of the level 2 utility screens.

Page 3-12 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 73: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.5.4 Exp. Counter

This function allows the number of exposures taken on each focal spot of the X-ray tube to be viewed.

Use these steps to view the exposure counter:

Step Action 1. From the level 2 utility menus select EXP. COUNTER. 2. The exposure count for the small focus (SF) and large focus (LF) may be viewed on the LCD

display. 3. Press the menu button [5] to return to screen 1 of the level 2 utility screens.

[More]

This accesses screen 2 of the level 2 utilities.

3.5.5 LCD mode

This function allows setting of normal or reverse video on the LCD display.

Use these steps to set the normal or reverse video:

Step Action 1. From the level 2 utility menus select LCD MODE. 2. Select NORMAL for normal video, or INVERSE for reverse video. 3. Press the menu button [5] to return to screen 2 of the level 2 utility screens.

3.5.6 Configuration

This function allows global enabling or disabling of the AEC function, setting of the optional AEC type, setting of the maximum kV, and in some models setting of the maximum generator output power.

Use these steps to set the above parameters:

Step Action 1. From the level 2 utility menus select CONFIGURATION. 2. Toggle button [1] to select AEC INHIBITED if the AEC function is to be disabled. This is

normally selected if the optional AEC board is not fitted. Select 1 FIELD ION AEC for single field ion chamber(s), or ION CHAMBR TYPE 1 or ION CHAMBR TYPE 2 for three field ion chambers per the table below. The selection made in this step applies to both AEC chambers.

AEC Selection Compatible AEC chambers and preamp

Start & Field Select (Active level)

RAMP

ION CHAMBR TYPE 1 AID ICX 153, 159, 162, 192 with preamp 6067I. AID ICX 924 with CAEC preamp 61137E.

0 V + Ramp

ION CHAMBR TYPE 2 Comet 3 field Ion chamber with preamp XP3-VSC. Vacutec 70125, 701125, 701135 with preamp 900.

+12 VDC + Ramp

Refer to the compatibility statement / product description in chapter 1 for the compatible AEC device properties if your AEC chamber and preamplifier is not listed in the above table.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-13

Page 74: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.5.6 Configuration (Cont)

Step Action 3. Press the KV MAX: UP button to increase the maximum kV. Press the DOWN button to

decrease the maximum kV. The maximum kV may be adjusted between the limits of 80 and 125 kV or 80 to 150 kV depending on application. If the maximum kV is changed, the X-ray tube will need to be auto calibrated. Refer to the section Calibration in this chapter for the auto calibration procedure.

4. DENSITY selection is not available. 5. Select 30kW, 32.5kW, or 37.5kW for 1 phase generators, or 50kW for 3 phase 50 kW

generators. Some single phase models will only allow the selection of 30 kW. Where selection of 32.5 and 37.5 kW IS available in the generator, the line voltage must be within the range of 230 to 253 volts AC. This function is not available on all models. IF THE POWER LEVEL IS CHANGED, THE CALIBRATION FUNCTION WILL NEED TO BE PERFORMED ON BOTH FOCAL SPOTS.

6. Select FIL. STDBY. Toggle to set the approximate filament standby current. A: 2.00 amps B: 2.36 amps C: 2.60 amps D: 2.90 amps This must be set below the X-ray tube emission level, consistent with the X-ray tube manufacturers recommended filament standby current.

7. Select PREP TIME. Toggle to set the approximate prep time. A: 2.5 seconds B: 1.9 seconds C: 1.5 seconds D: 1.2 seconds This must be set high enough to ensure that the anode is fully boosted before making an exposure. Refer to the X-ray tube data sheet. Confirm the prep time setting using a suitable tachometer to ensure proper anode RPM.

8. Press the menu button [5] to return to screen 2 of the level 2 utility screens. 3.5.7 Tube Selection

This function allows the desired X-ray tube to be selected from those displayed on the console. Other tube types may be available by downloading new tube data to the EEPROM on the CPU board.

Use these steps to select the desired X-ray tube:

Step Action 1. From the level 2 utility menus select TUBE SELECTION. 2. Select the desired tube by pressing the adjacent button. The selected tube type will be enclosed

within the > and < symbols. NOTE THAT CHANGING THE TUBE TYPE WILL ERASE EXISTING CALIBRATION DATA.

3. Press the menu button [5] to return to screen 2 of the level 2 utility screens.

Page 3-14 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 75: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3

Film Program This function is not available. Film Speed This function is not available. [MORE]

This accesses screen 3 of the level 2 utilities. MAS STEP SIZE This function is not available.

3.5.8 Receptor Conf.

This function allows the wall and table image receptors to be programmed. For each of these receptors, AEC may be ENABLED or DISABLED, and the wall Bucky or table Bucky may be ENABLED or DISABLED. If the wall or table Bucky is ENABLED, the wall or table Bucky selector(s) may be selected AND the generator will look for a Bucky ready signal. If the wall or table Bucky is DISABLED, the wall or table Bucky selector(s) may be selected BUT the generator will not require a Bucky ready signal to make an exposure. This selection is typically made if a grid cabinet is used.

Use these steps to program the image receptors:

Step Action 1. From the level 2 utility menus select RECEPTOR CONF. 2. Toggle button [3] to select AEC: ENABLED or AEC: DISABLED for the wall Bucky. DISABLED

will not allow AEC to be selected for the wall Bucky. 3. Toggle button [4] to select BUCKY: ENABLED or BUCKY: DISABLED for the wall Bucky.

DISABLED means that the generator will not require a Bucky ready signal to make an exposure.

4. Toggle button [8] to select AEC: ENABLED or AEC: DISABLED for the table Bucky. DISABLED will not allow AEC to be selected for the table Bucky.

5. Toggle button [9] to select BUCKY: ENABLED or BUCKY: DISABLED for the table Bucky. DISABLED means that the generator will not require a Bucky ready signal to make an exposure.

6. Press the menu button [5] to return to screen 3 of the level 2 utility screens.

TUBE SEASONING This function is not available.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-15

Page 76: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.6.0 HARDWARE PROGRAMMING

3.6.1 Speaker Volume

The speaker in the console may be set to one of three loudness levels. To change the current speaker volume, perform the following steps.

step Action 1. Switch the generator OFF at the console. 2. Turn the console on its face. A soft, clean cloth should be placed on the work surface to protect

the console from damage while it is face down. 3. Remove and temporarily set aside the screws at the bottom of the console. Carefully remove

the bottom of the console enclosure. 4. Locate JW3 and JW4 on the CPU board in the console. These are near the speaker LS1 at the

bottom of the CPU board. 5. Insert or remove jumpers JW3 and JW4 per the table below for the desired speaker volume. 6. Both jumpers are installed when shipped from the factory. If one or both jumpers is removed to

reduce the speaker volume, it is suggested that they be stored in a safe place in case they need to be reinstalled at a later time.

7. Reposition the console bottom and replace the hardware that was removed in step 3.

JW3, JW4 VOLUME

JW3 and JW4 removed Low

JW3 or JW4 installed Medium

JW3 and JW4 installed High

3.6.2 LCD Backlight Brightness adjustment

The brightness of the LCD display may be adjusted as desired. To change the LCD brightness, perform the following steps.

step Action 1. Follow steps 1 to 3 in section 3.6.1. 2. Locate R2 on the CPU board. Refer to figure 6-2 (Chapter 6). 3. Switch ON the console and observe the LCD display brightness. TAKE ALL APPROPRIATE

PRECAUTIONS WITH THE CONSOLE SWITCHED ON. HIGH VOLTAGE IS PRESENT IN THE AREA OF THE LCD BACKLIGHT AND THE LCD BACKLIGHT POWER SUPPLY.

4. Adjust R2 to set the desired LCD brightness. 5. Switch OFF the console. 6. Reposition the console bottom and replace the hardware that was removed in previous steps.

Page 3-16 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 77: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.7.0 AEC PROGRAMMING AND CALIBRATION

This section applies to generators with optional AEC only. This section contains background information relevant to AEC operation. It is strongly suggested that this be read and understood prior to beginning AEC calibration.

3.7.1 AEC Limitations: Minimum and Maximum Exposure Times

• Due to various delays inherent in the AEC chamber and generator electronics, minimum AEC exposure times should be greater than 10 ms.

• AEC exposures should normally be kept well under one second. When X-ray techniques are used that result in longer exposures, the film density will not be correct due to film reciprocity failure.

• Care must be exercised when using low kV values with table Buckys because most table tops and grids absorb considerable radiation in the range of 60 - 65 kVp. This will adversely affect AEC operation.

3.7.2 Precalibration Notes Please note the following points before beginning AEC calibration:

• All setup functions per the preceeding sections must be completed before beginning AEC calibration.

• The generator must be known to be calibrated before beginning AEC calibration. • All components (X-ray tubes, collimators, AEC chambers, etc) used during AEC calibration must be those

which will be used during procedures, and must be positioned as they will be in actual use of the X-ray room.

• During AEC calibration, always ensure that the central ray is centered relative to the image receptor. • Prior to placing the water absorber, ensure that the collimator is opened sufficiently to cover ALL fields on

the AEC pickup device. • Ensure that the absorber is positioned such as to fully cover the X-ray field. The absorber must extend a

minimum of 3/8″ (10 mm) beyond the X-ray field.

• The AEC chamber(s) must be calibrated, and the AEC fields must be balanced before proceeding. This should be done per the AEC chamber manufacturers procedure.

• AEC calibration is performed using the center field only, (left and right fields are not selectable during AEC

calibration). • A maximum of three separate AEC calibration curves may be established (for three film screen

combinations). Each AEC calibration curve has five breakpoints, normally 55 kVp, 65 kVp, 80 kVp, 100 kVp, and 125 kVp. These are the factory defaults referenced in the following steps, but may be changed if required by pressing kVp ↑ or ↓.

• AEC breakpoint calibration is initially performed for film / screen combination 1, displayed when entering the AEC ADJUSTMENT menu. When this is complete, AEC calibration may be performed for the next film / screen combination. The third and last film / screen combination is calibrated next, if desired.

• The labels for the three film screen combinations may be set to the site requirements using a laptop computer and the Service Software. In this procedure, the three film screen combinations will be referred to as F/S 1, F/S 2, and F/S 3 (film / screen 1, film / screen 2, and film / screen 3).

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-17

Page 78: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.7.3 Navigating the AEC Adjustment menus

The first screen displayed within the AEC ADJUSTMENT menu will show breakpoint #1, normally 55 kVp, for F/S 1. Pressing → will scroll through the F/S 1 kVp breakpoints in the sequence 55, 65, 80, 100, and 125 kVp. Pressing → again will change to F/S 2, starting with the 55 kVp breakpoint. Continuing to press → will scroll through the breakpoints for F/S 2. Pressing → again will change to F/S 3, starting with the 55 kVp breakpoint. Continuing to press → will scroll through the breakpoints for F/S 3. Pressing → again starts the above cycle over, pressing ← reverses the above sequence. Pressing GAIN ADJ when requested opens a submenu which allows setting of the AEC gain for F/S 1, F/S 2, and F/S 3.

3.7.4 AEC Calibration (Table Bucky)

CAUTION: THE PROCEDURES IN THESE SECTIONS REQUIRE X-RAY EXPOSURES. TAKE ALL SAFETY PRECAUTIONS TO PROTECT PERSONNEL FROM X-RADIATION.

The following is an overview of the steps that will be performed in sections 3.7.4 and 3.7.5. • Set up the equipment in preparation for table bucky AEC calibration, including water phantom absorbers. • Set the AEC reference to a mid value at the 80 kVp knee breakpoint for film / screen 1. Adjust the AEC

gain to achieve the correct film density. • Once the AEC gain is correctly set at the 80 kVp knee breakpoint, set the correct film density at 55 kVp, 65

kVp, 100 kVp, and 125 kVp by adjusting the AEC reference at each of those breakpoints. • Repeat the previous two steps for film / screen 2 and film / screen 3, if used, except the density

adjustments for all five kVp breakpoints are now made by only adjusting the AEC reference at each of those breakpoints.

• Repeat the above steps for wall bucky AEC calibration except no generator AEC adjustments are made for wall bucky AEC calibration. The AEC gain setting and AEC reference settings made during table bucky calibration are used for the wall bucky AEC chamber also. The wall bucky AEC chamber gain settings must therefore be set to match the gain of the table bucky AEC chamber.

Step Action SETUP FOR AEC CALIBRATION 1. Confirm that the table bucky AEC chamber is connected to the correct input on the

AEC board per chapter 2. Ensure that the AEC chamber is properly calibrated and the fields are balanced before continuing.

2. Align the tube stand and table Bucky such that the central ray is centered relative to the image receptor.

3. Open up the collimator just enough to expose all fields of the AEC pickup. Ensure that the central ray remains centered relative to the image receptor.

4. Place 15 cm thickness of water between the collimator and the image receptor. The water must be in a flat, X-ray transparent container, sufficiently large to ensure that the X-ray field is fully blocked as noted in 3.7.2.

5. Place the generator into the programming mode. Refer to section 3.3.0. 6. Select AEC ADJUSTMENT from the level 2 utilities. Refer to 3.5.0. Select the table

bucky receptor.

Page 3-18 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 79: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.7.4 AEC Calibration (Table Bucky) Cont Step Action 80 KVP KNEE BREAKPOINT CALIBRATION 7. Select the 80 kVp knee breakpoint #3 for F/S 1. 8. In each step following, where an exposure is made to verify the density, set the B/U

mAs sufficiently high that the AEC exposure is not terminated by the backup timer. 9. Press UP or DOWN to set the AEC REF to 325 (the range is 0 to 1023). 10. Load a test cassette with fresh film and install it in the image receptor. Use a film /

screen combination corresponding to F/S 1. 11. Select GAIN ADJ. Select F/S 1 (note again that F/S 1 is a mnemonic used in this

procedure for the first film / screen being calibrated). Set F/S 1 to an initial gain value (n) as per table 3-6 using the GAIN ↑ or ↓ buttons. Table 3-6 shows recommended starting gain values for 100, 400 and 800 speed film / screens.

12. Press the menu button [5] to return to the AEC breakpoint calibration screen. 13. Make an X-ray exposure. Then develop the film and measure the optical density. 14. If the measured O.D. is not within 5% of the desired value, readjust the AEC gain (as

per step 11) to increase or decrease the density. Increasing the gain value will decrease the density, decreasing the gain value will increase the density.

15. Load the test cassette in step 10 with fresh film and install it in the image receptor. Make an X-ray exposure, then develop the film and measure the optical density.

16. Repeat steps 14 and 15 until the desired O.D. is achieved at the kVp knee breakpoint. 17. Once the desired film density is achieved, record the post mAs, AEC gain adjustment

calibration number, AEC reference number, and O.D. in a copy of table 3-7 for the 80 kVp knee breakpoint.

55 KVP BREAKPOINT CALIBRATION 18. Replace the water in step 4 with 10 cm thickness of water. As before, ensure that the

absorber fully blocks the X-ray field. 19. Select the 55 kVp breakpoint for F/S 1. 20. Load the test cassette in step 10 with fresh film and install it in the image receptor.

Make an X-ray exposure, then develop the film and measure the optical density. 21. If the measured O.D. is not within 5% of the desired value, adjust the AEC REF value

for the 55 kVp breakpoint. DO NOT READJUST THE GAIN ADJ SETTING. Increasing the AEC REF value will increase the density, decreasing the AEC REF value will decrease the density.

22. Repeat steps 20 and 21 until the desired O.D. is achieved at the 55 kVp breakpoint. 23. Record the post mAs, AEC reference number, and O.D. in a copy of table 3-7 for the

55 kVp breakpoint. 65 KVP BREAKPOINT CALIBRATION 24. Select the 65 kVp breakpoint for F/S 1. 25. Load the test cassette in step 10 with fresh film and install it in the image receptor.

Make an X-ray exposure, then develop the film and measure the optical density. 26. If the measured O.D. is not within 5% of the desired value, adjust the AEC REF value

for the 65 kVp breakpoint. DO NOT READJUST THE GAIN ADJ SETTING. 27. Repeat steps 25 and 26 until the desired O.D. is achieved at the 65 kVp breakpoint. 28. Record the post mAs, AEC reference number, and O.D. in a copy of table 3-7 for the

65 kVp breakpoint.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-19

Page 80: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.7.4 AEC Calibration (Table Bucky) Cont Step Action 100 KVP BREAKPOINT CALIBRATION 29. Replace the water in step 18 with 20 cm thickness of water. As before, ensure that the

absorber fully blocks the X-ray field. 30. Select the 100 kVp breakpoint for F/S 1. 31. Load the test cassette in step 10 with fresh film and install it in the image receptor.

Make an X-ray exposure, then develop the film and measure the optical density. 32. If the measured O.D. is not within 5% of the desired value, adjust the AEC REF value

for the 100 kVp breakpoint. DO NOT READJUST THE GAIN AD SETTING. 33. Repeat steps 31 and 32 until the desired OD is achieved at the 100 kVp breakpoint. 34. Record the post mAs, AEC reference number, and OD in a copy of table 3-7 for the

100 kVp breakpoint. 125 kVp BREAKPOINT CALIBRATION 35. Select the 125 kVp breakpoint for F/S 1. 36. Load the test cassette in step 10 with fresh film and install it in the image receptor.

Make an X-ray exposure, then develop the film and measure the optical density. 37. If the measured OD is not within 5% of the desired value, adjust the AEC REF value

for the 125 kVp breakpoint. DO NOT READJUST THE GAIN ADJ SETTING. 38. Repeat steps 36 and 37 until the desired O.D. is achieved at the 125 kVp breakpoint. 39. Record the post mAs, AEC reference number, and O.D. in a copy of table 3-7 for the

125 kVp breakpoint. Step Action FILM / SCREEN 2 BREAKPOINT CALIBRATION 40. Perform F/S 2 calibration only if it is intended to use this film / screen combination. If

this is NOT being used, ensure that this film / screen combination is not used in any anatomical displays.

41. Repeat steps 7 to 39 as described for film / screen 1 except use a test cassette and film to match F/S 2. Be sure that the correct AEC breakpoints are being calibrated, screens 6-10 in the AEC ADJUSTMENT menus calibrate film / screen 2. These screens are accessed by completing the 125 kVp breakpoint for F/S 1, and pressing → as described earlier.

Step Action FILM / SCREEN 3 BREAKPOINT CALIBRATION 42. Perform F/S 3 calibration only if it is intended to use this film / screen combination. If

this is NOT being used, ensure that this film / screen combination is not used in any anatomical displays.

43. Repeat steps 7 to 39 as described for film screen 1 except use a test cassette and film to match F/S 3. Be sure that the correct AEC breakpoints are being calibrated, screens 11-15 in the AEC ADJUSTMENT menus calibrate film / screen 3. These screens are accessed by completing the 125 kVp breakpoint for F/S 2, and pressing → as described earlier.

Page 3-20 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 81: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.7.4 AEC Calibration (Table Bucky) Cont

FILM / SCREEN INITIAL GAIN ADJ VALUE (n) 100 150 400 236 800 244

Table 3-6: Initial gain value vs film / screen speed

The relative AEC voltage gain is expressed by:

Where n ranges from 0 to 254. This means that for low values of n, there is very little slope to the AEC gain curve. Therefore, n needs to change substantially to effect significant gain changes. For high values of n, the AEC gain curve is quite steep, as a result minor changes in n will have a significant effect on AEC gain.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-21

Page 82: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.7.4 Table 3-7: Breakpoint Worksheet

Record the final results in a copy of the table below. The final results are those obtained AFTER films have been developed to verify the correct O.D. at each breakpoint. TABLE BUCKY

FILM SCREEN 1 SPEED = AEC GAIN ADJ CAL # _______ @ kVp _______

#3 BK. POINT = 80 kVp Post mAs = AEC REF NO. = O.D. = #1 BK. POINT = 55 kV Post mAs = AEC REF NO. = O.D. = #2 BK. POINT = 65 kV Post mAs = AEC REF NO. = O.D. = #4 BK. POINT = 100 kV Post mAs = AEC REF NO. = O.D. = #5 BK. POINT = 125 kV Post mAs = AEC REF NO. = O.D. =

FILM SCREEN 2 SPEED = AEC GAIN ADJ CAL # _______ @ kVp _______

#3 BK. POINT = 80 kVp Post mAs = AEC REF NO. = O.D. = #1 BK. POINT = 55 kV Post mAs = AEC REF NO. = O.D. = #2 BK. POINT = 65 kV Post mAs = AEC REF NO. = O.D. = #4 BK. POINT = 100 kV Post mAs = AEC REF NO. = O.D. = #5 BK. POINT = 125 kV Post mAs = AEC REF NO. = O.D. =

FILM SCREEN 3 SPEED = AEC GAIN ADJ CAL # _______ @ kVp _______

#3 BK. POINT = 80 kVp Post mAs = AEC REF NO. = O.D. = #1 BK. POINT = 55 kV Post mAs = AEC REF NO. = O.D. = #2 BK. POINT = 65 kV Post mAs = AEC REF NO. = O.D. = #4 BK. POINT = 100 kV Post mAs = AEC REF NO. = O.D. = #5 BK. POINT = 125 kV Post mAs = AEC REF NO. = O.D. =

WALL BUCKY

FILM SCREEN 1 SPEED =

#3 BK. POINT = 80 kVp Post mAs = O.D. =#1 BK. POINT = 55 kV Post mAs = O.D. =#2 BK. POINT = 65 kV Post mAs = O.D. =#4 BK. POINT = 100 kV Post mAs = O.D. =#5 BK. POINT = 125 kV Post mAs = O.D. =

FILM SCREEN 2 SPEED =

#3 BK. POINT = 80 kVp Post mAs = O.D. =#1 BK. POINT = 55 kV Post mAs = O.D. =#2 BK. POINT = 65 kV Post mAs = O.D. =#4 BK. POINT = 100 kV Post mAs = O.D. =#5 BK. POINT = 125 kV Post mAs = O.D. =

FILM SCREEN 3 SPEED =

#3 BK. POINT = 80 kVp Post mAs = O.D. =#1 BK. POINT = 55 kV Post mAs = O.D. =#2 BK. POINT = 65 kV Post mAs = O.D. =#4 BK. POINT = 100 kV Post mAs = O.D. =#5 BK. POINT = 125 kV Post mAs = O.D. =

Page 3-22 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 83: Indico 100L Service Manual - 740914

CPI Canada Inc. Interfacing, Programming, & Calibration 3 3.7.5 AEC Calibration (Wall Bucky)

The AEC calibration curves that were established in 3.7.4 are used for the wall Bucky as well as the table Bucky. If the calibration on both AEC chambers is similar, the optical density measurements for the wall Bucky should be similar to the table Bucky. If there are significant differences in O.D. between the table Bucky measurements and the wall Bucky measurements under similar conditions, troubleshooting of equipment external to the generator will need to be undertaken.

Step Action SETUP FOR AEC CALIBRATION 1. Confirm that the wall bucky AEC chamber is connected to the correct input on the

AEC board per chapter 2. Ensure that the AEC chamber is properly calibrated and the fields are balanced before continuing.

2. Repeat steps 2 to 6 of 3.7.4, EXCEPT align the tube stand and wall Bucky in step 2, and select the wall Bucky receptor in step 6.

80 KVP KNEE BREAKPOINT VERIFICATION 3. Select the 80 kVp knee breakpoint #3 for F/S 1. DO NOT ADJUST THE AEC REF

VALUE, USE THE SETTING THAT WAS ESTABLISHED FOR TABLE BUCKY CALIBRATION.

4. Load a test cassette with fresh film and install it in the image receptor. Use a film / screen combination corresponding to F/S 1.

5. Make an X-ray exposure. Then develop the film and measure the optical density. 6. If the measured O.D. is not within 5% of the value previously recorded in table 3-7,

adjust the AEC chamber gain per the manufacturers instructions to increase or decrease the density.

7. Repeat steps 5 and 6 until the desired O.D. is achieved at the 80 kVp knee breakpoint.

8. Once the desired film density is achieved, record the post mAs, and O.D. in a copy of table 3-7 for the 80 kVp knee breakpoint.

55 KVP BREAKPOINT CALIBRATION 9. Replace the 15 cm of water with 10 cm thickness of water. As before, ensure that the

absorber fully blocks the X-ray field. 10. Select the 55 kVp breakpoint for F/S 1. DO NOT ADJUST THE AEC REF VALUE,

USE THE SETTING THAT WAS ESTABLISHED FOR TABLE BUCKY CALIBRATION.

11. Load the test cassette in step 4 with fresh film and install it in the image receptor. Make an X-ray exposure, then develop the film and measure the optical density.

12. Verify that the O.D. is within 5 % of the value previously recorded in table 3-7. If this is not so, it indicates that the two AEC chambers are not tracking and troubleshooting must be done at the system level.

13. If the desired O.D. is achieved, record the post mAs, and O.D. in a copy of table 3-7 for the 55 kVp knee breakpoint.

65 KVP BREAKPOINT CALIBRATION 14. Select the 65 kVp breakpoint for F/S 1. DO NOT ADJUST THE AEC REF VALUE,

USE THE SETTING THAT WAS ESTABLISHED FOR TABLE BUCKY CALIBRATION.

15. Load the test cassette in step 4 with fresh film and install it in the image receptor. Make an X-ray exposure, then develop the film and measure the optical density.

16. Verify that the O.D. is within 5 % of the value previously recorded in table 3-7. 17. If the desired O.D. is achieved, record the post mAs, and O.D. in a copy of table 3-7

for the 65 kVp knee breakpoint.

X-Ray Generator Service Manual Ch # 740914-03 Rev. L Page 3-23

Page 84: Indico 100L Service Manual - 740914

3 Interfacing, Programming, & Calibration CPI Canada Inc. 3.7.5 AEC Calibration (Wall Bucky) Cont 100 KVP BREAKPOINT CALIBRATION 18. Replace the 10 cm of water with 20 cm thickness of water. As before, ensure that the

absorber fully blocks the X-ray field. 19. Select the 100 kVp breakpoint for F/S 1. DO NOT ADJUST THE AEC REF VALUE,

USE THE SETTING THAT WAS ESTABLISHED FOR TABLE BUCKY CALIBRATION.

20. Load the test cassette in step 4 with fresh film and install it in the image receptor. Make an X-ray exposure, then develop the film and measure the optical density.

21. Verify that the O.D. is within 5 % of the value previously recorded in table 3-7. 22 If the desired O.D. is achieved, record the post mAs, and O.D. in a copy of table 3-7

for the 100 kVp knee breakpoint. 125 KVP BREAKPOINT CALIBRATION 23. Select the 125 kVp breakpoint for F/S 1. DO NOT ADJUST THE AEC REF VALUE,

USE THE SETTING THAT WAS ESTABLISHED FOR TABLE BUCKY CALIBRATION.

24. Load the test cassette in step 4 with fresh film and install it in the image receptor. Make an X-ray exposure, then develop the film and measure the optical density.

25. Verify that the O.D. is within 5 % of the value previously recorded in table 3-7. 26. If the desired O.D. is achieved, record the post mAs, and O.D. in a copy of table 3-7

for the 125 kVp knee breakpoint. Step Action FILM / SCREEN 2 BREAKPOINT CALIBRATION 27. Verify F/S 2 calibration only if it is intended to use this film / screen combination with

the wall Bucky. 28. Repeat steps 3 to 26 as described for film / screen 1 except use a test cassette and

film to match F/S 2. Be sure that the correct AEC breakpoints are being verified.

Step Action FILM / SCREEN 3 BREAKPOINT CALIBRATION 29. Verify F/S 3 calibration only if it is intended to use this film / screen combination with

the wall Bucky. 30. Repeat steps 3 to 26 as described for film screen 1 except use a test cassette and

film to match F/S 3. Be sure that the correct AEC breakpoints are being verified.

Page 3-24 Rev. L X-Ray Generator Service Manual Ch # 740914-03

Page 85: Indico 100L Service Manual - 740914

CPI Canada Inc Acceptance Testing 4

CHAPTER 4

ACCEPTANCE TESTING

CONTENTS:

4.1.0 INTRODUCTION ...........................................................................................................................................4-2 4.2.0 REQUIRED TEST EQUIPMENT FOR GENERATOR VERIFICATION. .......................................................4-3 4.3.0 CONSOLE OVERVIEW .................................................................................................................................4-3 4.4.0 ACCEPTANCE TESTS (BASIC FUNCTIONS) .............................................................................................4-4

4.4.1 Console Functional Tests...........................................................................................................................4-4 4.4.2 Low Speed Starter Tests............................................................................................................................4-6 4.4.3 Generator kVp, mAs Tests .........................................................................................................................4-6

4.5.0 AEC VERIFICATION......................................................................................................................................4-7 4.6.0 ACCEPTANCE TESTS (REPRODUCIBILITY, LINEARITY AND HVL) ........................................................4-8

X-Ray Generator Service Manual Ch # 740914-04 Rev. E Page 4-1

Page 86: Indico 100L Service Manual - 740914

4 Acceptance Testing CPI Canada Inc 4.1.0 INTRODUCTION

This section details acceptance testing which verifies that the generator is performing within its limits. It is recommended that this be done after the unit is installed, as prescribed in the maintenance schedule in Chapter 6, or when components are replaced which may affect the calibration or X-ray output. Examples of such components are the X-ray tube, HT tank, power supply control board, filament board, CPU board, and the AEC board. Calibration procedures will be supplied with replacement parts in those cases where recalibration is necessary after replacement of that assembly.

WARNING: USE EXTREME CARE IN MEASURING HIGH VOLTAGES. ACCIDENTAL CONTACT MAY CAUSE SERIOUS INJURY OR DEATH.

MAINS VOLTAGE WILL BE PRESENT INSIDE THE GENERATOR CABINET, EVEN WITH THE CONSOLE SWITCHED OFF. THIS VOLTAGE IS EXTREMELY DANGEROUS, USE EXTREME CAUTION.

THE DC BUS CAPACITORS, LOCATED ON THE DC BUS ASSEMBLY, PRESENT A HAZARD FOR A MINIMUM OF 5 MINUTES AFTER THE GENERATOR HAS BEEN SWITCHED OFF. CHECK THAT THESE CAPACITORS ARE DISCHARGED BEFORE TOUCHING ANY PARTS IN THE GENERATOR.

WARNING: THE PROCEDURES IN THIS CHAPTER REQUIRE THE PRODUCTION OF X-RAYS. TAKE ALL APPROPRIATE SAFETY PRECAUTIONS TO PROTECT PERSONNEL FROM X-RAY RADIATION.

WARNING: ALWAYS ENSURE THAT THE EQUIPMENT UNDER TEST AND ALL ASSOCIATED TEST EQUIPMENT IS PROPERLY GROUNDED

ENSURE THAT THE HIGH VOLTAGE CABLES ARE INTACT / UNDAMAGED AND PROPERLY CONNECTED BEFORE ATTEMPTING EXPOSURES

ENSURE THAT THE FOLLOWING ITEMS ARE COMPLETED PRIOR TO PERFORMING THE ACCEPTANCE TESTING:

• The tube auto calibration has been done as per Chapter 3 of this manual. • The programming has been completed as per Chapter 3. • If the installation has AEC; verify that AEC calibration has been completed as per Chapter 3. • Acceptance testing shall only be started after the installation is complete i.e. generator

installed as per Chapter 2.

Page 4-2 Rev. E X-Ray Generator Service Manual Ch # 740914-04

Page 87: Indico 100L Service Manual - 740914

CPI Canada Inc Acceptance Testing 4 4.2.0 REQUIRED TEST EQUIPMENT FOR GENERATOR VERIFICATION.

• kV measuring device such as a Dynalyzer (or equivalent), or a non-invasive kVp meter. • Two channel storage oscilloscope. • mA / mAs meter. • Radiation meter 0-1000 mR and 1-15 R/min. • Lead diaphragm or equivalent to collimate the beam. • General purpose DVM. • Strobe or reed tachometer. • Current probe 0 to 20 amps AC. This is to be used with the oscilloscope to view the stator current

waveform. • A set of HVL filters (see attached supplement). • Calculator.

4.3.0 CONSOLE OVERVIEW

Figure 4-1 shows the designations of the membrane switches as referenced in this procedure.

Figure 4-1: Console switch designations

X-Ray Generator Service Manual Ch # 740914-04 Rev. E Page 4-3

Page 88: Indico 100L Service Manual - 740914

4 Acceptance Testing CPI Canada Inc 4.4.0 ACCEPTANCE TESTS (BASIC FUNCTIONS)

4.4.1 Console Functional Tests

Refer to figure 4-1 for the button designations [1] to [19]

Step Action Result Check 1. Press power ON then power OFF button on

the console. Generator switches on and off.

2. Press power ON again to switch unit on. The PREP and X-RAY LEDS should briefly light. The green LED adjacent to the power ON switch should light.

3. Press OK and / or RESET, if required, to proceed to the main screen shown in figure 4-2. This may be displayed if an error is present when the generator is powered up.

4. From the main APR menu, press APR button [1].

Verify that the available views within that anatomical section are displayed. Refer to figure 4-2 and 4-3.

5. Press the menu button [5] to return to the main menu.

The main menu screen should be displayed.

6. Repeat steps 4 and 5 for APR buttons [2] to [4] and [6] to [9].

The display should be per steps 4 and 5.

7. Select a technique within an anatomical view. See figure 4-4.

The display should be per figure 4-4.

8. Change technique parameters (press the ↑ and ↓ buttons [10], [11], and [12]).

Verify that each of the selected techniques change as appropriate.

9. Press each of the receptor buttons [13] to [15]. Verify that the desired receptor is selected. Receptors not enabled in programming cannot be selected, refer to RECEPTOR CONF. in Chapter 3.

10. Select the TABLE or WALL bucky. Press AEC button [16].

Verify that AEC is selected / deselected by pressing this button. AEC must be enabled in programming in order to select AEC, refer to RECEPTOR CONF. and CONFIGURATION in Chapter 3.

11. Press each of field select buttons [17], [18], and [19] with AEC enabled.

Verify that each field is selected in turn. ION CHAMBER AEC must be selected in order to activate all three fields, refer to CONFIGURATION in Chapter 3.

12. Press the PREP button. The green LED adjacent to the PREP button should light.

13. Press the X-RAY button. The X-RAY warning light adjacent to the X-RAY button should light,

14. Press the power OFF button on the console. Generator switches off.

Page 4-4 Rev. E X-Ray Generator Service Manual Ch # 740914-04

Page 89: Indico 100L Service Manual - 740914

CPI Canada Inc Acceptance Testing 4 4.4.1 Console Functional Tests (Cont)

Depending on generator model, the main menu on your console may show anatomical sections different from those shown below.

CERVICAL GENERALTHORACIC LOWER EXTRMTYLUMBAR UPPER EXTRMTYFULL SPINE SPECIAL PROC

Figure 4-2: Main console menu

Depending on generator model and selected anatomical section, the anatomical views may differ from those shown below.

CERVICAL AP FLEX EXTAPOM NASIUMLATERAL VERTEXOBLIQUE PILLAR

Figure 4-3: Typical anatomical views

The screen shown below is a typical techniques screen in non-AEC mode. CERVICAL F/S: 200 AP AVERAGESID: 40 LG FOCUS TABLE BUCKY TIME: 67mS m A: 150

CM: 24 kVp: 70 mAs: 10.0

Figure 4-4: Typical techniques screen

NOTE: THE SCREENS SHOWN ABOVE ARE EXAMPLES ONLY, ACTUAL SCREENS AND TECHNIQUES MAY VARY IN DETAIL.

X-Ray Generator Service Manual Ch # 740914-04 Rev. E Page 4-5

Page 90: Indico 100L Service Manual - 740914

4 Acceptance Testing CPI Canada Inc 4.4.2 Low Speed Starter Tests Step Action Result Check 1. Connect a current probe to the common lead

of the X-ray tube stator. Switch ON the console. Press and hold the PREP button.

A 60 Hz waveform dropping to less than half amplitude after prep complete.

2. Measure the rotor boost time. Should be approximately 1.5 sec. 3. Use a strobe or reed tachometer and verify

that the tube reaches operating speed at end of boost.

Speed ≥ 3300 RPM.

4. Switch OFF the console. N/A

4.4.3 Generator kVp, mAs Tests

NOTE: A DYNALYZER IS NOT RECOMMENDED FOR MA MEASUREMENTS WITH THIS X-RAY GENERATOR. BANDWIDTH LIMITATIONS OF THE DYNALYZER WILL RESULT IN INACCURATE MA MEASUREMENTS AT MA VALUES LESS THAN APPROXIMATELY 100 MA. MA MEASUREMENTS SHOULD BE MADE WITH AN MA / MAS METER CONNECTED TO THE MA TEST JACKS ON THE HT TANK. EXPOSURE TIMES MUST BE GREATER THAN 100 MS TO ENSURE ACCURATE MEASUREMENTS.

WARNING : TAKE ALL APPROPRIATE PRECAUTIONS WHEN CONNECTING THE MA / MAS METER TO THE HT TANK. ENSURE MAINS POWER IS SWITCHED OFF AND ALL CAPACITORS ARE DISCHARGED BEFORE CONNECTING THE METER. USE THE SAME PRECAUTIONS WHEN DISCONNECTING THE METER. THE MA SHORTING LINK MUST BE IN PLACE ON THE HT TANK AT ALL TIMES EXCEPT WHEN AN APPROVED MA / MAS METER IS CONNECTED.

Step Action Result Check 1. Set up the kVp measuring device to measure

kVp as per the manufacturers instructions. Connect the output of the kVp measuring device to channel 1 of the oscilloscope.

N/A

2. Set up an mA / mAs meter to measure X-ray tube current by temporarily open-circuiting the mA measuring link E17-E18 on the HT tank. Connect the mA / mAS meter to these terminals, following the device manufacturers instructions. Connect the output of the mA / mAS meter to channel 2 of the oscilloscope.

N/A

Page 4-6 Rev. E X-Ray Generator Service Manual Ch # 740914-04

Page 91: Indico 100L Service Manual - 740914

CPI Canada Inc Acceptance Testing 4 4.4.3 Generator kVp, mAs Tests (Cont)

NOTE: TEST EQUIPMENT TOLERANCES MUST BE ALLOWED FOR IN THE MEASUREMENTS IN THE FOLLOWING SECTIONS. LIMITS STATED ARE THE MAXIMUM ALLOWED LIMITS, INCLUDING EQUIPMENT TOLERANCES AND MEASUREMENT ERROR. DISREGARD “CABLE CHARGE” SPIKES AT THE BEGINNING OF THE MA WAVEFORM.

IF HIGH KV EXPOSURES ARE NOT ALLOWED IN THE FOLLOWING STEPS, CHECK THE MAXIMUM KV SETTING THAT HAS BEEN PROGRAMMED. REFER TO “CONFIGURATION” IN CHAPTER 3.

Step Action Result Check If the following mAs steps are not available, the mAs step size may need to be set, refer to MAS

STEP SIZE in Chapter 3. 3. Switch ON the generator and after initialization

select the following technique: Select the NON BUCKY receptor [14] Select kVp = 100, mA = 100, mAs = 10.

N/A

4. Make an exposure and verify the following results. If calculating mAs from mA and time measurements, then measure time at 75% of the peak kVp waveform.

kVp = 100 ± 5.0 kV. mAs = 10 ± 1.5 mAs.

5. Repeat steps 3 and 4 but set the values to kVp = 65, mA = 200, mAs = 20.

kVp = 65 KV ± 3.25 kV. mAs = 20 ± 2.5 mAs.

6. Repeat steps 3 and 4 but set the values to kVp = 120, mA = 200, mAs = 20.

kVp = 120 KV ± 6.0 kV. mAs = 20 ± 2.5 mAs.

7. Select 75 kVp, 150 mA, large focus. Select the mAs values shown below. A: 1.2 mAs. B: 8.0 mAs. C: 26.0 mAs. D: 63.0 mAs.

A: 1.2 mAs ± 0.62 mAs. B: 8 mAs ± 1.3 mAs. C: 26 mAs ± 3.1 mAs. D: 63 mAs ± 6.8 mAs.

8. Select 100 mA, 5 mAs, small focus. Select the kVp values shown below. A: 50 kVp. B: 60 kVp. C: 80 kVp. D: 120 kVp E: 130 kVp (150 kV generators only).

A: 50 kVp ± 2.5 kV. B: 60 kVp ± 3 kV. C: 80 kVp ± 4 kV. D: 120 kVp ± 4.8 kV. E: 130 kVp ± 5.2 kV.

4.5.0 AEC VERIFICATION

This section applies only to generators with AEC. • Review Chapter 3: AEC Adjustment. • Follow the appropriate steps in Chapter 3 to verify the AEC calibration.

X-Ray Generator Service Manual Ch # 740914-04 Rev. E Page 4-7

Page 92: Indico 100L Service Manual - 740914

4 Acceptance Testing CPI Canada Inc 4.6.0 ACCEPTANCE TESTS (REPRODUCIBILITY, LINEARITY AND HVL)

The procedure for performing reproducibility, linearity and HVL testing is contained in a separate document, part number 740917-00 which immediately follows this page.

Page 4-8 Rev. E X-Ray Generator Service Manual Ch # 740914-04

Page 93: Indico 100L Service Manual - 740914

CPI Canada Inc Reproducibility, Linearity, & HVL Testing

SUPPLEMENT

REPRODUCIBILTY, LINEARITY, & HVL TESTING

CONTENTS:

1.0 INTRODUCTION.................................................................................................................................................. 2 2.0 EQUIPMENT SETUP........................................................................................................................................... 3 3.0 REPRODUCIBILITY............................................................................................................................................. 3 4.0 LINEARITY........................................................................................................................................................... 7 5.0 H.V.L. EVALUATION ........................................................................................................................................... 9

X-Ray Generator Service Manual Supplement 740917-00 Rev. A Page 1

Page 94: Indico 100L Service Manual - 740914

Reproducibility, Linearity, & HVL Testing CPI Canada Inc 1.0 INTRODUCTION

This supplement describes reproducibility, linearity, and half - value layer (HVL) tests which may be used to verify performance of medical X-ray generators.

NOTE: THIS SUPPLEMENT DETAILS TYPICAL REPRODUCIBILITY, LINEARITY, AND HVL TESTS. LOCAL REGULATIONS SHOULD ALWAYS BE CONSULTED PRIOR TO PERFORMING THESE TESTS, AS DETAILS MAY VARY IN SOME JURISDICTIONS, OR ADDITIONAL TESTS MAY NEED TO BE PERFORMED.

WARNING: SOME EXPOSURES IN THIS SECTION MUST BE TAKEN AT THE MAXIMUM GENERATOR KVP. THE X-RAY TUBE MUST BE KNOWN TO BE CAPABLE OF OPERATION AT THAT KVP VALUE, AND THE TUBE SHOULD FIRST BE SEASONED TO ENSURE THAT OPERATION AT HIGH KVP VALUES WILL NOT BE PROBLEMATIC.

Page 2 Rev. A X-Ray Generator Service Manual Supplement 740917-00

Page 95: Indico 100L Service Manual - 740914

CPI Canada Inc Reproducibility, Linearity, & HVL Testing 2.0 EQUIPMENT SETUP

1. Place the radiation probe above the table approximately 25 cm (10”). Select an SID of

approximately 100 cm (40”). 2. Place a lead diaphragm over the detector and adjust its height so that the X-ray beam covers the

detector but does not over radiate the sides of the ‘R’ probe. Refer to figure 1.

Figure 1: mR measurement setup 3.0 REPRODUCIBILITY

Calculate reproducibility as follows: 1. Using kV and mA/ms or mAs values per tables 1 to 4, make a series of 5 exposures. 2. Record each of the measured mR values in the appropriate table. Refer to step 3 before

beginning step 2. 3. Record the preselected mAs for each series of exposures in the header of each table. For 3 point

generators, this is the calculated mAs value where mAs = mA X time in seconds (example for 160 mA and 125 ms, mAs = 160 X 0.125 = 20 mAs).

X-Ray Generator Service Manual Supplement 740917-00 Rev. A Page 3

Page 96: Indico 100L Service Manual - 740914

Reproducibility, Linearity, & HVL Testing CPI Canada Inc 3.0 REPRODUCIBILITY (Cont)

4. Calculate and record the average dose mR. 5. Calculate the difference mR - mR for each exposure. 6. Square each difference from the previous step. 7. Calculate the sum of the differences squared. 8. Calculate the standard deviation (S) by using the formula.

9. Calculate reproducibility by dividing S by mR.

10. Table 5 shows example reproducibility calculations. 11. If linearity is to be measured, it is suggested that dose measurements be taken at this time for

entry into tables 6 and 7. Refer to 4.0 LINEARITY for details.

IN TABLES 1 TO 4, 3 POINT MEANS THAT FOR GENERATORS WHERE KV, MA, AND TIME SELECTION IS AVAILABLE, THE KV, MA AND MS VALUES SHOWN SHOULD BE USED. FOR GENERATORS WHERE 2 POINT OPERATION ONLY IS AVAILABLE, THE KV AND MAS VALUES SHOWN SHOULD BE USED.

3 point = Minimum kV, maximum mA, 100 ms. 2 point = Minimum kV, maximum mAs. mAs = __________

EXP No. DOSE (mR) DIFFERENCE DIFFERENCE2

1 2 3 4 5

mR = ________

Calculate each of the differences ie: DIFF1 = mR1 - mR. Repeat for each remaining mR value.

Square each difference. Then calculate the sum of the difference2. Sum of difference2 = _________

Calculate standard deviation (S) using formula at beginning of this section: S = __________

Calculate reproducibility = S = _________ (not to exceed 0.045) mR

Table 1: Reproducibility

Page 4 Rev. A X-Ray Generator Service Manual Supplement 740917-00

Page 97: Indico 100L Service Manual - 740914

CPI Canada Inc Reproducibility, Linearity, & HVL Testing 3.0 REPRODUCIBILITY (Cont)

3 point = Maximum kV, minimum mA, 100 ms. 2 point = Maximum kV, minimum mAs. mAs = __________

EXP No. DOSE (mR) DIFFERENCE DIFFERENCE2

1 2 3 4 5

mR = ________

Calculate each of the differences ie: DIFF1 = mR1 - mR. Repeat for each remaining mR value.

Square each difference. Then calculate the sum of the difference2. Sum of difference2 = _________

Calculate standard deviation (S) using formula at beginning of this section: S = __________

Calculate reproducibility = S = _________ (not to exceed 0.045) mR

Table 2: Reproducibility

3 point = 50% of maximum kV, 250 ms, mA to give 100 - 500 µR (1 - 5 µGy) dose. 2 point = 50% of maximum kV, mAs to give 100 - 500 µR (1 - 5 µGy) dose. mAs = __________

EXP No. DOSE (mR) DIFFERENCE DIFFERENCE2

1 2 3 4 5

mR = ________

Calculate each of the differences ie: DIFF1 = mR1 - mR. Repeat for each remaining mR value.

Square each difference. Then calculate the sum of the difference2. Sum of difference2 = _________

Calculate standard deviation (S) using formula at beginning of this section: S = __________

Calculate reproducibility = S = _________ (not to exceed 0.045) mR

Table 3: Reproducibility

X-Ray Generator Service Manual Supplement 740917-00 Rev. A Page 5

Page 98: Indico 100L Service Manual - 740914

Reproducibility, Linearity, & HVL Testing CPI Canada Inc 3.0 REPRODUCIBILITY (Cont)

3 point = 80% of maximum kV, 250 ms, mA to give 100 - 500 µR (1 - 5 µGy) dose. 2 point = 80% of maximum kV, mAs to give 100 - 500 µR (1 - 5 µGy) dose. MAS = __________

EXP No. DOSE (mR) DIFFERENCE DIFFERENCE2

1 2 3 4 5

mR = ________

Calculate each of the differences ie: DIFF1 = mR1 - mR. Repeat for each remaining mR value.

Square each difference. Then calculate the sum of the difference2. Sum of difference2 = _________

Calculate standard deviation (S) using formula at beginning of this section: S = __________

Calculate reproducibility = S = _________ (not to exceed 0.045) mR

Table 4: Reproducibility

EXAMPLE mAs = _20_

EXP No. DOSE (mR) DIFFERENCE DIFFERENCE2

1 249.0 4.4 19.36 2 245.0 0.4 0.16 3 244.0 0.6 0.36 4 242.0 2.6 6.76 5 243.0 1.6 2.56

mR = _244.6_

Calculate each of the differences ie: DIFF1 = mR1 - mR. Repeat for each remaining mR value.

Square each difference. Then calculate the sum of the difference2. Sum of difference2 = _29.2_

Calculate standard deviation (s) using formula at beginning of this section: S = _2.70_

Calculate reproducibility = S = _0.011_ (not to exceed 0.045) mR

Table 5: Reproducibility

Page 6 Rev. A X-Ray Generator Service Manual Supplement 740917-00

Page 99: Indico 100L Service Manual - 740914

CPI Canada Inc Reproducibility, Linearity, & HVL Testing 4.0 LINEARITY

1. Record two additional series of dose measurements for entry into tables 6 and 7:

• For table 6, use settings per table 3 EXCEPT use an mA (or mAs) value adjacent to the mA (or mAs) setting used in table 3.

• For table 7, use settings per table 4 EXCEPT use an mA (or mAs) value adjacent to the mA (or mAs) setting used in table 4.

• Record the mAs in the header of tables 6 and 7 as per 3.0 step 3. 1. Calculate and record the average dose mR for table 6 and 7. 2. Record the preselected mAs and the average dose values taken from tables 3 and 4, and from

tables 6 and 7, at the top of the next page. 3. Using the appropriate mAs and mR values, calculate X3, X4, X6, and X7 in tables 8 and 9. 4. Calculate the coefficient of linearity, L, as per tables 8 and 9.

mAs = __________

EXP No. DOSE (mR) 1 2 3 4 5

mR = ________

Table 6: Linearity

mAs = __________

EXP No. DOSE (mR) 1 2 3 4 5

mR = ________

Table 7: Linearity

X-Ray Generator Service Manual Supplement 740917-00 Rev. A Page 7

Page 100: Indico 100L Service Manual - 740914

Reproducibility, Linearity, & HVL Testing CPI Canada Inc 4.0 LINEARITY (Cont)

Record the mAs and mR values taken from tables 3, 4, 6, and 7 below.

Table 3 mAs3 = _____ mR3 = _____ Table 4 mAs4 = _____ mR4 = _____ Table 6 mAs6 = _____ mR6 = _____ Table 7 mAs7 = _____ mR7 = _____

mR3 X3 = = ________ mAs3 mR6 X6 = = ________ mAs6 X3 - X6 L = = ________ (not to exceed 0.095) X3 + X6

In the numerator of the above equation, use the absolute value of X3 - X6 (disregard the minus sign).

Table 8: Linearity

mR4 X4 = = ________ mAs4 mR7 X7 = = ________ mAs7 X4 - X7 L = = ________ (not to exceed 0.095) X4 + X7

In the numerator of the above equation, use the absolute value of X4 - X7 (disregard the minus sign).

Table 9: Linearity

Page 8 Rev. A X-Ray Generator Service Manual Supplement 740917-00

Page 101: Indico 100L Service Manual - 740914

CPI Canada Inc Reproducibility, Linearity, & HVL Testing 5.0 H.V.L. EVALUATION

1. Be sure the X-ray source assembly (X-ray tube and beam limiting device) is fully assembled and

functional. 2. Use the test setup as per figure 1. 3. Set the generator as follows: 3 point generators, 80 kV, 200 mA, 50 ms, large focus. For 2 point

generators use 80 kV, 200 mA if this can be set, and 10 mAs. 4. Take a series of three exposures and record the mR values in table 10. Calculate and record the

average of the three exposures. 5. Place 2 mm of Al on top of the lead diaphragm (total of 2 mm added), repeat the exposure and

record the mR value in table 10. 6. Place an additional 1 mm of Al on top of the lead diaphragm (total of 3 mm added), repeat the

exposure; and record the mR value in table 10. 7. Place an additional 3 mm of Al on top of the lead diaphragm (total of 6 mm added), repeat the

exposure; and record the mR value in table 10. 8. The relative transmission for the average of the three mR values where no Al was added is

assigned a value of 1.00. Using that base, assign relative transmission values to the remaining mR values. For example, if the average mR value was 247 and has a relative transmission factor of 1.00, then 162 mR will have a relative transmission of 162 / 247 = 0.66.

9. Plot the relative transmission values in figure 1. This should produce a straight line on the graph

since the X-axis is logarithmic. 10. Interpolate to determine the HVL. The Al thickness at a relative transmission of 0.5 will be the

required HVL value. 11. Repeat steps 4 to 10: 3 point generators, 100 kV, 200 mA, 50 ms, large focus. For 2 point

generators use 100 kV, 200 mA if this can be set, and 10 mAs. Use table 11 to record the values and figure 2 to plot the results

12. Table 12 and figure 3 show example HVL determination.

X-Ray Generator Service Manual Supplement 740917-00 Rev. A Page 9

Page 102: Indico 100L Service Manual - 740914

Reproducibility, Linearity, & HVL Testing CPI Canada Inc 5.0 H.V.L. EVALUATION (Cont)

ADDED ALUMINUM FILTER DOSE (mR) RELATIVE TRANSMISSION 0 0 0 0 (Average of three readings) 1.00 2 1 3

Table 10: HVL dose values 80 kVp

Figure 1: HVL plot 80 kVp

For 80 kVp, the HVL must be ≥ 2.3 mm Al.

Page 10 Rev. A X-Ray Generator Service Manual Supplement 740917-00

Page 103: Indico 100L Service Manual - 740914

CPI Canada Inc Reproducibility, Linearity, & HVL Testing 5.0 H.V.L. EVALUATION (Cont)

ADDED ALUMINUM FILTER DOSE (mR) RELATIVE TRANSMISSION 0 0 0 0 (Average of three readings) 1.00 2 1 3

Table 11: HVL dose values 100 kVp

Figure 2: HVL plot 100 kVp

For 100 kVp, the HVL must be ≥ 2.7 mm Al.

X-Ray Generator Service Manual Supplement 740917-00 Rev. A Page 11

Page 104: Indico 100L Service Manual - 740914

Reproducibility, Linearity, & HVL Testing CPI Canada Inc 5.0 H.V.L. EVALUATION (Cont)

ADDED ALUMINUM FILTER DOSE (mR) RELATIVE TRANSMISSION 0 249 0 244 0 247 0 (Average of above three readings) 247 1.00 2 162 .66 1 131 .53 3 70 .28

Table 12: HVL dose values (example)

Figure 3: HVL plot (example)

By interpolating the thickness of Al at a relative transmission value of 0.5, it can be seen that the HVL is approximately 3.3.

Page 12 Rev. A X-Ray Generator Service Manual Supplement 740917-00

Page 105: Indico 100L Service Manual - 740914

CPI Canada Inc Troubleshooting 5

CHAPTER 5

TROUBLESHOOTING

CONTENTS:

SECTION TITLE 5.1.0 INTRODUCTION....................................................................................................................................................5-2 5.2.0 STATUS AND ERROR CODES.............................................................................................................................5-2

5.2.1 Start-Up Messages .............................................................................................................................................5-2 5.2.2 Status Messages ................................................................................................................................................5-3 5.2.3 Limit Messages...................................................................................................................................................5-3 5.2.4 Error Messages ..................................................................................................................................................5-4

5.3.0 STATUS LED’S ......................................................................................................................................................5-8 5.4.0 MISCELLANEOUS FAULTS................................................................................................................................5-10

5.4.1 Console Failure to Start ....................................................................................................................................5-10

X-Ray Generator Service Manual Ch # 740914-05 Rev. H Page 5-1

Page 106: Indico 100L Service Manual - 740914

5 Troubleshooting CPI Canada Inc

5.1.0 INTRODUCTION

The X-ray generator console will display status messages on the LCD display during normal and abnormal operation of the generator. This Chapter contains tables of those messages and suggests actions to be taken by service personnel to correct any malfunctions that may occur.

5.2.0 STATUS AND ERROR CODES 5.2.1 Start-Up Messages

These messages are displayed at initial power-on only, and indicate the status at the time that the generator is switched on.

MESSAGE DESCRIPTION AEC MEMORY O.K. / ERR

Indicates that the AEC data checksums match. ERR indicates that the AEC calibration data must be recreated by downloading via a laptop computer, or redoing the AEC adjustment procedure as per Chapter 3.

APR MEMORY: O.K. / ERR

O.K. indicates that the APR memory checksums match. ERR indicates an APR error. The APR data set should be recreated by downloading via a laptop computer, or changing any of the APR data and exiting.

CAL MEMORY: O.K. / ERR

O.K. indicates that the tube auto-calibration data checksums match. ERR indicates an auto cal error. The tube will need to be recalibrated.

HEAT SWITCH OPEN, TUBE MAY BE HOT

Indicates that the tube thermal switch is open or not connected.

LF DISABLED Large focus has been disabled in programming. LF UNCAL The large focus is not calibrated, run the CALIBRATION routine. MEMORY TEST: O.K. / ERR

O.K. indicates that the EPROM data checksums match. ERR indicates that the EPROM may be defective and should be replaced.

PROGRAM DATA: O.K. / ERR

Indicates that the mAs steps & error message checksums match. ERR indicates that the mAs step or error message checksums don’t match. This suggests that the software needs to be replaced.

RELEASE EXP. SW.

Indicates that the exposure switch is closed while the generator is being powered-up.

RT CLOCK: O.K. / ERR

O.K. indicates that the time is being incremented by the RT clock. ERR indicates that the battery voltage may be low, or the clock may need setting as per Chapter 3.

SOFTWARE REV Displays the software revision in the console. SF DISABLED Small focus has been disabled in programming. SF UNCAL The small focus is not calibrated, run the CALIBRATION routine. STAMP ERROR: XX Indicates that there was insufficient time to save the time stamp information before

power was lost, or that the lithium battery on the CPU board is low. If this message appears check the battery on the CPU board, or set the clock as per Chapter 3.

TUBE CALIBRATED Both focal spots are calibrated. TUBE DATA: O.K. / ERR

O.K. indicates that the tube insert data checksums match. ERR indicates a tube data error. The tube data must be recreated by downloading via a laptop computer, or possibly the EEPROM is defective and needs to be replaced.

TUBE DISABLED. ENTER UTILITY TO ENABLE.

Both focal spots have been disabled in programming.

Page 5-2 Rev. H X-Ray Generator Service Manual Ch # 740914-05

Page 107: Indico 100L Service Manual - 740914

CPI Canada Inc Troubleshooting 5

5.2.2 Status Messages

These messages indicate the status of the generator. MESSAGE DESCRIPTION

PREPARATION Indicates that PREP has been requested. READY Indicates that the generator is ready to make an exposure. X-RAY EXPOSURE Indicates that an X-ray exposure is in progress.

5.2.3 Limit Messages

These messages indicate that an exposure has been requested that exceeds one or more limits.

MESSAGE PROBLEM ACTION DENSITY LIMIT The generator upper or lower density limit has been

reached. None required.

DUTY CYCLE LIMIT Requested exposure not allowed until the duty cycle limit expires.

Wait for time limit to expire.

FOCUS CHANGED Indicates that the focus has automatically changed while increasing or decreasing the mAs.

None required.

INVALID SETTING Requested setting is invalid, ie attempting to deselect all AEC fields in AEC mode.

Correct the setting.

KVP LIMIT The generator kVp limit has been reached. None required. LF MAS LIMIT The generator large focus mAs limit has been reached. Switch to small focus

for mAs values < 1.0 mAS.

MAS LIMIT The generator mAs limit has been reached. None required. MAX MAS LIMIT The generator power limit has been reached. None required. POWER KV LIMIT Displayed if kV ↑ is pressed when the generator has

reached its power limit. Reduce mAs if higher kVp is required.

POWER MAS LIMIT Displayed if mAs ↑ is pressed when the generator has reached its power limit.

Reduce mAs.

SF MAS LIMIT The generator small focus mAs limit has been reached. Reduce mAs on small focus.

TUBE KW LIMIT Displayed if requested parameter change will violate the tube data.

None required.

CALC. AEC RAMP LIMIT

Requested density correction will require a ramp voltage > 15.0V from AEC device.

Increase KVp and decrease density correction.

LOW RAMP LIMIT Requested density correction will require a ramp voltage below noise threshold of AEC device (≅ 50mV)

Decrease KVp and increase density correction.

X-Ray Generator Service Manual Ch # 740914-05 Rev. H Page 5-3

Page 108: Indico 100L Service Manual - 740914

5 Troubleshooting CPI Canada Inc

5.2.4 Error Messages

These messages indicate that an error has occurred. MESSAGE PROBLEM ACTION

APR MEMORY:ERR. APR data has been corrupted. Reload the APR via the utilty software.

40 INCH ITLK Indicates that the 40 ″ S.I.D. has been selected, but the tube stand is not in the 40 ″ S.I.D. position.

1. Check S.I.D. on tube stand.

2. Check the 25 pin console cable.

72 INCH ITLK Indicates that the 72 ″ S.I.D. has been selected, but the tube stand is not in the 72 ″ S.I.D. position.

1. Check S.I.D. on tube stand.

2. Check the 25 pin console cable.

AEC INTERLOCK Indicates that no AEC device is connected. Connect AEC device or deprogram the AEC option.

BACKUP MAS The AEC backup mAs timer, and not the AEC circuits, terminated the AEC exposure.

1. Check the BUmAs setting, as described on page 5-5.

2. Check the AEC chamber, wiring, and AEC board.

BACKUP TIMER The generator backup timer terminated the exposure at the limit of 6000 ms for non AEC exposures, or the BUmAS limit for AEC exposures.

Increase the mA for non AEC exposures, or increase the BUmAs for AEC exposures.

DEVICE INTERRUPT Indicates that the selected bucky is not ready. Check the Bucky(s). DOOR INTERLOCK Indicates that the room door is open. Check the door

interlock. FAULTY EXP. SW. Indicates that the exposure switch is closed during

power-up. This message is displayed if the switch is not opened immediately after the RELEASE EXP. SW. message in 5.2.1. is displayed.

Check and replace the exposure switch if necessary.

FILAMENT FAULT Indicates that filament current < 2.5 amps detected. 1. Check for open filament in X-ray tube.

2. Check for poor connections in the cathode HT cable.

3. Check the filament supply board.

FOCUS DISABLED Indicates that the requested focal spot has been deprogrammed.

Select the other focal spot, or enable the selected focal spot in programming mode.

HEAT SW. ITR. Indicates that the thermal switch has opened during an exposure.

Allow the X-ray tube to cool.

Page 5-4 Rev. H X-Ray Generator Service Manual Ch # 740914-05

Page 109: Indico 100L Service Manual - 740914

CPI Canada Inc Troubleshooting 5

5.2.4 Error Messages (Cont)

MESSAGE PROBLEM ACTION kV HIGH Indicates measured kV is high by > 15 %. 1. Check power supply

control board and/or HT tank.

2. Re-auto calibrate (per CALIBRATION in Chapter 3). *

3. Consult product support.

kV LOW Indicates measured kV is low by >15 %. 1. Check that the mains voltage is appropriate for the selected output power.

2. Check power supply control board and/or HT tank.

3. Re-auto calibrate (per CALIBRATION in Chapter 3). *

4. Consult product support.

LOW / NO AEC RAMP A low (or no) AEC ramp has been detected from the AEC chamber during the first 50 ms of AEC exposure.

1. Check settings (techniques, receptor selection, etc, and mA / mAs settings. Refer to the note at the bottom of this page).

2. Check AEC chamber, conections, wiring etc.

mA HIGH Indicates measured mA is high by > 20 %. 1. Check for tube arcing.

2. Check power supply control board and/or HT tank.

3. Re-auto calibrate (per CALIBRATION in Chapter 3). *

4. Consult product support.

NOTE: AEC backup mAs = mAs as set in non-AEC mode X 2 (i.e. @ 10 mAs, AEC backup mAs = 20 mAs).

AEC backup time = AEC backup mAs ÷ mA (i.e. 100 mAs ÷ 50 mA = 2 seconds AEC backup time). The “Low / No AEC Ramp” fault limit varies depending on the AEC backup time. Reducing the AEC backup time increases the AEC fault reference voltage. Therefore, for short AEC backup times a larger AEC voltage ramp is require than for long AEC backup times. The formula below illustrates the relationship between AEC backup time and the minimum AEC voltage ramp that must be present at 50

X-Ray Generator Service Manual Ch # 740914-05 Rev. H Page 5-5

Page 110: Indico 100L Service Manual - 740914

5 Troubleshooting CPI Canada Inc

5.2.4 Error Messages (Cont) ms after the start of an AEC exposure.

0.050 X VrefAEC Backup Time( )0.9[ ]=Vramp (Min)

Vref varies from 0 to 5 V, depending on the technique. It can be seen that increasing the AEC backup time decreases the minimum required AEC voltage ramp.

MESSAGE PROBLEM ACTION mA LOW Indicates measured mA is low by > 20 %. 1. Check mains

voltage. 2. Check power supply

control board and/or HT tank.

3. Re-auto calibrate (per CALIBRATION in Chapter 3). *

4. Consult product support.

MANUAL INTERRUPT

The X-ray exposure switch was released before the exposure was finished.

Hold the exposure switch until the exposure is finished.

POWER UP FAILURE, CHECK LINE VOLT.

Indicates that the generator did not power up within 30 seconds of the console being switched on.

1. Check that the mains voltage is not outside the low line specification.

2. Check the 37 conductor console cable.

3. Check the DC bus voltage. The DC bus capacitors may not be charging.

4. Check that the main contactor on the power input board is closed.

P/S FAULT Indicates that the generator is not ready to make an exposure.

1. Check D1 on the auxiliary board. If this is lit, the + or - 12 VDC rails may be low. Check the +/- 12 V supplies on the auxiliary board.

2. Check PREP and kV ENABLE signals on the power supply control board.

3. Consult product support.

Page 5-6 Rev. H X-Ray Generator Service Manual Ch # 740914-05

Page 111: Indico 100L Service Manual - 740914

CPI Canada Inc Troubleshooting 5

5.2.4 Error Messages (Cont)

MESSAGE PROBLEM ACTION P/S NOT READY

Indicates that the main line contactor in the generator is not commanded to close. A DC bus fault may be preventing the DC bus capacitors from charging, in which case the contactor closure will be inhibited.

Check D1 on the auxiliary board. If this is lit, the contactor will not close. Check the DC bus assembly and the soft start circuit on the power input board if this is the case.

ROTOR FAILURE, TURN UNIT OFF

Indicates that the rotor is running at any time other than during prep or during an exposure.

Check rotor board or auxiliary board.

ROTOR FAULT Indicates that the low speed starter board has not sensed adequate stator current.

1. Check the X-ray tube stator wiring.

2. Check the rotor board.

TUBE DISABLED Both focal spots are disabled. Enable one or both focal spots.

TUBE FAULT Indicates a fault has been detected in the kV or mA output, or in the inverter circuits.

Check if D69, D70, D71, D72, or D82 on the power supply control board are lit. If any of the above LED’s are lit, refer to the note at the end of this table for a description of the function of these LED’s.

UNCOMMANDED KV Indicates that kV feedback has been detected when no kV has been requested.

Switch generator OFF. Prevent further use of generator until problem is fixed.

* The generator purges the existing X-ray tube auto calibration data before starting the auto calibration

routine and saving new calibration data. Therefore, auto calibration should be a last resort during general troubleshooting, and should only be done to recalibrate the tube. For example, if a low mA fault is presented, you should ensure that the generator is fully functional, and actually needs recalibration. If calibration is attempted on a partially functional generator, the auto calibration routine may be aborted before any calibration is done, and the generator will inhibit further exposures until the selected mA is calibrated for the selected kV.

X-Ray Generator Service Manual Ch # 740914-05 Rev. H Page 5-7

Page 112: Indico 100L Service Manual - 740914

5 Troubleshooting CPI Canada Inc

5.2.4 Error Messages (Cont) NOTE RE TUBE FAULT ERROR: As briefly described in the preceeding table, LED’s D69, D70, D71, D72, or D82 on the power supply control board indicate the type of fault that caused the stated error.

• D69 indicates high kVp ≥ 130 kVp for 125 kV generators, ≥ 160 kVp for 150 kV generators. If this is lit, consult product support.

• D70 indicates that inverter (primary) current exceeded preset limits. If this is lit, consult product support. • D71 indicates cathode current exceeded preset limits, usually due to an arc. If this is lit, check the X-ray

tube. The X-ray tube may be damaged or simply require seasoning. Refer to Chapter 6 for TUBE CONDITIONING / SEASONING. If this does not solve the problem, check the HT cables and HT connections. Consult product support if the previous steps fail to resolve the problem.

• D72 indicates anode current exceeded preset limits, usually due to an arc. Follow the recommendations for D71.

• D82 indicates an inverter fault. If this is lit, consult product support. 5.3.0 STATUS LED’S

The following list describes the normal states and the functions of the status and diagnostic LED’s on the circuit boards in the generator. Power Input Board (1 phase)

LED NORMAL STATE FUNCTION D2 Lit Indicates that the DC bus voltage is sufficient to energize the

main line contactor. D9 Lit Indicates that +12 VDC is present on the power input board.

This LED does NOT indicate that the +12 VDC rail is within normal limits.

Power Input Board (3 phase) LED NORMAL STATE FUNCTION

DS1 Lit Indicates that the DC bus capacitors are charged. The DC bus must be charged before the main line contactor K5 can close.

DS2 Lit Indicates that +12 VDC is present on the power input board. This LED does NOT indicate that the +12 V rail is within normal limits.

DS4 Lit Indicates that the main contactor (K5) is closed. DS6 Lit Indicates that K2 on the power input board is energized. This

relay supplies AC mains power to the primary of the room interface transformer.

DS7 Lit Indicates that K1 on the power input board is energized. This is the soft-start relay, and is normally energized when the generator is switched on. K1 will open if a DC bus fault is detected.

Page 5-8 Rev. H X-Ray Generator Service Manual Ch # 740914-05

Page 113: Indico 100L Service Manual - 740914

CPI Canada Inc Troubleshooting 5

5.3.0 STATUS LED’s (Cont) Auxiliary Board

LED NORMAL STATE FUNCTION D1 Off Indicates that the + or - 12 VDC rails are low, or that a DC bus

(soft start) fault is present. See D2 under power input board. D2 Lit Indicates that the logic is OK for the main line contactor to close. D3 Lit Indicates that the power input board logic is OK to allow the

main line contactor to close. D29 Lit Indicates that the +12 VDC supply on the auxiliary board is

present. This LED does NOT indicate that the +12 VDC rail is within normal limits.

D30 Lit Indicates that the -12 VDC supply on the auxiliary board is present. This LED does NOT indicate that the -12 VDC rail is within normal limits.

D31 Lit Indicates that the + and - 35 VDC supplies on the auxiliary board are present. This LED does NOT indicate that the these supplies are within normal limits.

Control Board LED NORMAL STATE FUNCTION

D69 Off Refer to “NOTE RE TUBE FAULT ERROR” at end of 5.2.4. D70 Off Refer to “NOTE RE TUBE FAULT ERROR” at end of 5.2.4. D71 Off Refer to “NOTE RE TUBE FAULT ERROR” at end of 5.2.4. D72 Off Refer to “NOTE RE TUBE FAULT ERROR” at end of 5.2.4. D82 Off Refer to “NOTE RE TUBE FAULT ERROR” at end of 5.2.4.

Customer Interface Board

LED NORMAL STATE FUNCTION DS1 Lit Indicates the presence of +24 VDC on this board.

CPU Board LED NORMAL STATE FUNCTION

DS7 Lit Indicates the presence of +24 VDC on this board.

X-Ray Generator Service Manual Ch # 740914-05 Rev. H Page 5-9

Page 114: Indico 100L Service Manual - 740914

5 Troubleshooting CPI Canada Inc

5.4.0 MISCELLANEOUS FAULTS 5.4.1 Console Failure to Start

SYMPTOM: LCD display on console lights up when console switched ON, but start up diagnostics, etc do not display.

SOLUTION: Check battery B1 on CPU board. The console will not start executing the firmware code if the battery is missing, or if the battery voltage is very low ≤ 1 V.

Page 5-10 Rev. H X-Ray Generator Service Manual Ch # 740914-05

Page 115: Indico 100L Service Manual - 740914

CPI Canada Inc Regular Maintenance 6

CHAPTER 6

REGULAR MAINTENANCE

CONTENTS:

Section Title 6.1.0 INTRODUCTION....................................................................................................................................................6-2 6.2.0 SERVICE RECORD...............................................................................................................................................6-3 6.3.0 MAINTENANCE SCHEDULE ................................................................................................................................6-4 6.4.0 OIL FILL / LEVEL CHECK .....................................................................................................................................6-4 6.5.0 CLEANING.............................................................................................................................................................6-6 6.6.0 EPROM / EEPROM REPLACEMENT ...................................................................................................................6-6 6.7.0 BATTERY REPLACEMENT...................................................................................................................................6-6 6.8.0 TUBE CONDITIONING / SEASONING .................................................................................................................6-8 6.9.0 AEC / CAEC BOARD INSTALLATION OR REPLACEMENT................................................................................6-9

X-Ray Generator Service Manual Ch # 740914-06 Rev. C Page 6-1

Page 116: Indico 100L Service Manual - 740914

6 Regular Maintenance CPI Canada Inc 6.1.0 INTRODUCTION

This Chapter provides a recommended schedule for periodic maintenance of the X-ray generator. Instructions are also included for installing an AEC / CAEC board in generators which originally did not have the AEC option installed, and for installing an AEC / CAEC board to upgrade to CAEC (where available). The initial installation date and location, and all service performed on the generator, should be recorded in table 6-1. The record should be as thorough as possible, detailing the scope and type of work that was performed (all service and a record of all replacement parts that were installed). Additionally, the person performing the work should date and sign the record. This information will be invaluable in the future for traceability and to ensure continued compatibility of the generator.

If a major component (such as the HT oil tank or a major circuit board) is replaced, recalibration will be needed. A separate procedure will be included with the spare in those cases, detailing the required calibration procedure. The acceptance test procedure per Chapter 4 should then be performed prior to placing the generator back into service.

WARNING: MAINTENANCE IS TO BE PERFORMED ONLY BY COMPETENT, TRAINED PERSONNEL WHO ARE FAMILIAR WITH THE POTENTIAL HAZARDS ASSOCIATED WITH THIS EQUIPMENT.

NOTE: MAINTENANCE SCHEDULE FREQUENCY MAY BE DICTATED BY CERTAIN REGULATORY REQUIREMENTS OF THE COUNTRY OR STATE IN WHICH THE INSTALLATION IS LOCATED. ALWAYS CHECK THE LOCAL CODES AND REGULATIONS WHEN SETTING THE MAINTENANCE SCHEDULE.

WARNING: ALWAYS SWITCH OFF MAINS POWER TO THE GENERATOR AND WAIT A MINIMUM OF 5 MINUTES FOR CAPACITORS TO DISCHARGE BEFORE BEGINNING ANY PREVENTATIVE MAINTENANCE, INCLUDING CLEANING.

Page 6-2 Rev. C X-Ray Generator Service Manual Ch # 740914-06

Page 117: Indico 100L Service Manual - 740914

CPI Canada Inc Regular Maintenance 6 6.2.0 SERVICE RECORD

INSTALLED BY: __________________ DATE: ______________ LOCATION: _________________________

Service Date Description of Service Performed By

Table 6-1: Installation and service record

X-Ray Generator Service Manual Ch # 740914-06 Rev. C Page 6-3

Page 118: Indico 100L Service Manual - 740914

6 Regular Maintenance CPI Canada Inc 6.3.0 MAINTENANCE SCHEDULE

Maintenance Frequency

Description of Preventative Maintenance

Every 6 Months AND whenever a related certifiable X-ray component is replaced:

1. Clean and re-grease all HV connections using vapor proof compound.

2. Check that all HV connections are tight. 3. Clean the control console, and main cabinet as needed. REFER TO

6.5.0 CLEANING BEFORE PROCEEDING. 4. Perform the X-ray tube auto calibration routine, refer to Chapter 3. 5. Verify the calibration of the generator, refer to Chapter 4. 6. Test the X-ray tube thermal switch circuits in the generator.

Disconnect the tube thermal switch and verify the correct error message, and that X-ray exposures are inhibited.

7. Perform any additional tests required by laws governing this installation.

Every 12 months: 1. Examine the following for any visible damage and replace any damaged components:

• The exterior of the control console, including the membrane switch assembly.

• The cables between the control console and the generator main cabinet.

• The cable(s) between the control console and the AEC chambers (if the AEC option is used).

• The handswitch (if used) and the cables connecting this to the console.

• The HT cables. 2. Open the generator main cabinet and examine the unit for any visible

damage: missing or loose ground connections, oil leaks, damaged cables etc.

3. Ensure that there are no obstructions blocking any of the ventilation holes or louvers on the generator cabinet.

Every 5 years: Replace the lithium battery on the CPU board in the control console. Refer to the spares list in Chapter 8 for the required part number. Refer to 6.7.0 for the battery replacement procedure.

6.4.0 OIL FILL / LEVEL CHECK

The insulating oil level in the HT transformer does NOT require periodic checking under normal conditions. However, if there is evidence of possible oil loss, the procedure for checking the correct oil level follows.

1. Loosen the oil fill plug screw on the tank lid. 2. With the screw sufficiently loosened, remove the rubber (neoprene) plug.

Page 6-4 Rev. C X-Ray Generator Service Manual Ch # 740914-06

Page 119: Indico 100L Service Manual - 740914

CPI Canada Inc Regular Maintenance 6 6.4.0 OIL FILL/LEVEL CHECK (Cont)

3. Use a clean ruler, strip of cardboard, or other equivalent material to determine the oil level --

measured always from the TOP surface of the HT tank’s lid. • Normally the oil level should be between 0.88 - 1.25 inches (22 - 32 mm) from the top of the tank

lid. • If the oil level is between 1.25 - 1.6 inches (32 - 41 mm) from the top of the tank lid, then clean oil

should be added as needed. • If the oil level is greater than 1.6 inches (41 mm) below the top of the tank lid, please consult the

factory.

4. Use only fresh oil, type Shell DIALA AX or equivalent. It is critical that air is not added when topping up the oil. The following procedure is strongly recommended when adding oil. • Use a new clean syringe to remove oil from the container. A 60 cc catheter tip syringe is

recommended. Approximately 60 cc of oil is required to raise the oil level by one millimeter. • Turn the syringe upright and expel any trapped air. • Place the tip of the syringe through the oil-fill plug and into the oil, ensuring that it is below the

surface of the oil. • Gently eject the oil from the syringe into the HT tank, while making sure that the tip of the syringe

remains below the surface of the oil until all of the oil is emptied from the syringe. • Repeat the previous steps until the required amount of oil has been added.

5. Replace the oil fill plug and tighten the screw when finished. Do NOT over tighten the screw. This

should be tightened until the neoprene plug is no longer able to be turned by hand, then tighten the screw another 1 to 1 1/2 turns. Wipe up any oil spills. Dispose of soiled absorber in compliance with government requirements, and ensure conformity to local disposal regulations.

Figure 6-1: HT tank oil fill

X-Ray Generator Service Manual Ch # 740914-06 Rev. C Page 6-5

Page 120: Indico 100L Service Manual - 740914

6 Regular Maintenance CPI Canada Inc 6.5.0 CLEANING

• Never use anything other than mild soap and water to clean plastic surfaces. Other cleaners may damage the plastic.

• Never use any corrosive, solvent or abrasive detergents or polishes. • Ensure that no water or other liquid can enter any equipment. This precaution prevents short circuits and

corrosion forming on components. • Methods of disinfection used must conform to legal regulations and guidelines regarding disinfection and

explosion protection. • If disinfectants are used which form explosive mixtures of gases, these gases must have dissipated before

switching on the equipment again. • Disinfection by spraying is not recommended because the disinfectant may enter the X-ray equipment. • If room disinfection is done with an atomizer, it is recommended that the equipment be switched OFF,

allowed to cool down and covered with a plastic sheet. When the disinfectant mist has subsided, the plastic sheet may be removed and the equipment be disinfected by wiping.

6.6.0 EPROM / EEPROM REPLACEMENT

WARNING: PLEASE TAKE APPROPRIATE ELECTROSTATIC PRECAUTIONS AT ALL TIMES WHEN HANDLING THE EPROM OR EEPROM.

1. Switch the generator OFF at the console and disconnect the mains voltage.

2. Turn the console on its face. A soft, clean cloth should be placed on the work surface to protect the console from damage while it is face down.

3. Remove and temporarily set aside the screws at the bottom of the console. Carefully remove the bottom of the console enclosure.

4. Locate and carefully remove the existing EPROM and / or EEPROM on the CPU board (U7 for the EPROM, U6 for the EEPROM, refer to figure 6-2).

5. Carefully insert the replacement EPROM and / or EEPROM into the socket observing the orientation per figure 6-2.

6. Reposition the console bottom and replace the hardware that was removed in step 3.

6.7.0 BATTERY REPLACEMENT

1. Refer to 6.6.0 for console disassembly instructions to access the CPU board. Refer to figure 6-2 for the battery location.

2. Remove the battery from the holder by gently prying under the battery at the access slot in the battery

holder using a small screwdriver. Slide the battery over the edge of the holder and remove it when it is free.

3. Check the voltage of the new battery prior to inserting it. This should be nominally 3.2V, do not use if it

is under 2.90 V. 4. Wipe the replacement battery with a clean cloth, and ensure that the holder is clean before inserting

the new battery.

Page 6-6 Rev. C X-Ray Generator Service Manual Ch # 740914-06

Page 121: Indico 100L Service Manual - 740914

CPI Canada Inc Regular Maintenance 6 6.7.0 BATTERY REPLACEMENT (cont)

5. Gently lift the spring contact on the holder and insert the replacement battery positive (+) side up 6. A stamp error message may be observed at initial turn-on after replacing the battery. This is normal at

this point and will be cleared by setting the date and time in step 8. 7. Reassemble the console as per 6.6.0. 8. Reset the date and time; refer to Chapter 3 for the procedure to do this.

Figure 6-2: EPROM and battery location

X-Ray Generator Service Manual Ch # 740914-06 Rev. C Page 6-7

Page 122: Indico 100L Service Manual - 740914

6 Regular Maintenance CPI Canada Inc 6.8.0 TUBE CONDITIONING / SEASONING

Tube conditioning or “seasoning” is particularly important for new X-ray tubes or tubes that have not been used for several days. This should be performed before attempting auto calibration, as an unseasoned tube may not operate properly at higher kV values without arcing. Always refer to the X-ray tube manufacturers seasoning procedure, if available. If the X-ray tube manufacturers instructions are not available, the following procedure may be used: IF DEVIATING FROM THE PROCEDURE BELOW, IT MUST BE UNDERSTOOD THAT THE GENERATOR WILL NOT ALLOW EXPOSURES AT KV VALUES HIGHER THAN HAS BEEN AUTO CALIBRATED. FOR EXAMPLE, IF AUTO CALIBRATION HAS ONLY BEEN DONE AT 40 AND 70 KV, THEN EXPOSURES ABOVE 70 KV ARE NOT ALLOWED. TUBE AUTO CALIBRATION IS DONE AT 40 kV, 70 KV, 100 KV, AND 125 KV. THE HIGHER KV VALUES WILL NOT BE CALIBRATED IF LIMITED BY THE “KV MAX” SETTING IN THE UTILITY MENU. 1. In the level 2 utilities menu record the current maximum kV setting, then temporarily set the maximum

kV to 80. Refer to Chapter 3 for the procedure to do this. 2. Start the tube auto calibration sequence, and allow the generator to finish the 70 kV calibration step.

Calibration will not be attempted at higher kV due to the kV limit set in step 1. 3. Season the tube at 70 kV, large focus, 300 mA, by taking approximately 10 exposures of 50 mAs.

These exposures should be taken at the maximum rate that the duty cycle limit will allow. 4. Temporarily set the maximum kV to 110 kV as per step 1. 5. Restart the tube auto calibration sequence, and allow the generator to finish the 100 kV calibration

step. Calibration will not be attempted at higher kV due to the kV limit set in the previous step. THE GENERATOR WILL BEGIN AUTO CALIBRATION AT 40 KV EACH TIME THE AUTO

CALIBRATION SEQUENCE IS RESTARTED IN THIS PROCEDURE. 6. Season the tube at 100 kV, large focus, 300 mA, by taking approximately 10 exposures of 50 mAs.

These exposures should be taken at the maximum rate that the duty cycle limit will allow. 7. Reset the maximum kV to the value that it was in step 1 prior to the temporary change. 8. Season the tube at 120 kV, large focus, 300 mA, if the kV limit allows this, by taking approximately 5

exposures of 50 mAs. These exposures should be taken at the maximum rate that the duty cycle limit will allow.

Page 6-8 Rev. C X-Ray Generator Service Manual Ch # 740914-06

Page 123: Indico 100L Service Manual - 740914

CPI Canada Inc Regular Maintenance 6 6.9.0 AEC / CAEC BOARD INSTALLATION OR REPLACEMENT

1. Switch the generator OFF at the console and disconnect the mains voltage. 2. Turn the console on its face. A soft, clean cloth should be placed on the work surface to protect the

console from damage while it is face down. 3. Remove and temporarily set aside the screws at the bottom of the console. Carefully remove the

bottom of the console enclosure. 4. If an existing AEC board is to be replaced, follow these steps:

a) Locate the four nylon standoffs that secure the AEC board to the CPU board. b) Turn the head of each plunger approximately 90° counterclockwise to release the latches. Refer to

figure 6-3. c) Note the orientation of the existing AEC board, then gently lift the existing AEC board off the nylon

standoffs. d) Orientate the replacement AEC board such that it will fit onto the existing nylon standoffs. Ensure

that the pins on J7 on the CPU board are lined up with the mating connector on the AEC board, then press the AEC board into position such that the J7 connector pins and the AEC board are fully seated. When the AEC board is properly installed, turn the head of each plunger approximately 90° clockwise to secure the board. Ensure that there is a minimum of 1/16” (2 mm) clearance between the heatsink of Q3 on the CPU board and the bottom of the AEC board. If there is insufficient clearance, gently bend Q3 toward J1 on the CPU board until sufficient clearance is available.

e) Reposition the console bottom and replace the hardware that was removed in step 3. f) Connect the AEC cables as per chapter 2, then perform AEC setup and calibration per chapter 3

of the service manual.

INSERT THIS END INTO HOLES IN CPU BOARD

PLUNGER

NYLONSTANDOFF

Figure 6-3: Nylon standoff 5. If an AEC board is to be installed in a generator that previously had no AEC board, follow these steps:

a) Locate the four AEC board mounting holes on the CPU board. These are indicated in figure 6-2.

b) Install a nylon standoff in each mounting hole identified in the previous step. Press the end of the standoff shown in figure 6-3 into the holes in the board, and then fully depress the plunger on the nylon standoffs to secure the standoffs to the board.

X-Ray Generator Service Manual Ch # 740914-06 Rev. C Page 6-9

Page 124: Indico 100L Service Manual - 740914

6 Regular Maintenance CPI Canada Inc 6.9.0 AEC / CAEC BOARD INSTALLATION OR REPLACEMENT (Cont)

c) Orientate the AEC board to be installed with the component side up, and the connectors facing the

rear of the console such that it will fit on the nylon standoffs. Ensure that the pins on J7 on the CPU board are lined up with the mating connector on the AEC board, then press the AEC board into position such that the J7 connector pins and the AEC board are fully seated. When the AEC board is properly installed, turn the head of each plunger approximately 90° clockwise to secure the board. Ensure that there is a minimum of 1/16” (2 mm) clearance between the heatsink of Q3 on the CPU board and the bottom of the AEC board. If there is insufficient clearance, gently bend Q3 toward J1 on the CPU board until sufficient clearance is available.

d) Reposition the console bottom and replace the hardware that was removed in step 3. e) Connect the AEC cables as per chapter 2, then perform AEC setup and calibration per chapter 3

of the service manual.

Page 6-10 Rev. C X-Ray Generator Service Manual Ch # 740914-06

Page 125: Indico 100L Service Manual - 740914

CPI Canada Inc Theory of Operation 7

X-Ray Generator Service Manual Ch # 740914-07 Rev. A Page 7-1

CHAPTER 7

THEORY OF OPERATION

CONTENTS:

7.0 INTRODUCTION...............................................................................................................................................7-2 7.1 FUNCTIONAL OVERVIEW...............................................................................................................................7-2

7.1.1 Power Input Board......................................................................................................................................7-2 7.1.2 Resonant Board .........................................................................................................................................7-2 7.1.3 Low Speed Starter Board ...........................................................................................................................7-2 7.1.4 Auxiliary Board ...........................................................................................................................................7-3 7.1.5 Filament Board ...........................................................................................................................................7-3 7.1.6 Power Supply Control Board......................................................................................................................7-3 7.1.7 Inverter Board.............................................................................................................................................7-4 7.1.8 DC Bus Assembly ......................................................................................................................................7-4 7.1.9 Customer Interface Board ..........................................................................................................................7-4 7.1.10 CPU Board ...............................................................................................................................................7-4 7.1.11 AEC Board (Optional)...............................................................................................................................7-4

7.2.0 HT TANK........................................................................................................................................................7-5

Page 126: Indico 100L Service Manual - 740914

7 Theory of Operation CPI Canada Inc

Page 7-2 Rev. A X-Ray Generator Service Manual Ch # 740914-07

7.0 INTRODUCTION

This Chapter provides an overview of the main functional blocks in the generator.

7.1 FUNCTIONAL OVERVIEW

This X-ray generator contains the power input board, resonant board, low speed starter board, auxiliary board, filament board, power supply control board, inverter board, DC bus assembly, customer interface board, and the HT oil tank in the main cabinet. The console houses the CPU board and optional AEC board. Each of these boards will be described briefly in the following subsections. The HT tank will be briefly discussed in this Chapter also.

7.1.1 Power Input Board

The AC mains power is connected to the main generator fuses on this board. When the main line contactor K5 on the power input board is closed, the mains voltage is switched to the DC bus assembly for rectification and filtering. The mains voltage is also fed to the other generator circuits from this board as required. The soft start circuit, which precharges the DC bus capacitors prior to the main line contactor closing, is located on the power input board. A DC bus voltage detector circuit is also located on this board. This circuit generates a signal that prevents the main contactor from closing should a DC bus fault be detected that prevents the DC bus from charging normally.

7.1.2 Resonant Board The resonant board contains the inverter resonant components (capacitor and inductors), and a current transformer which monitors the inverter current. A sample of the inverter current is then fed to the power supply control board where it is measured to ensure that it is within safe limits. Circuits on the power supply control board also use this sampled inverter current to synchronize the drive pulses to the inverter board.

7.1.3 Low Speed Starter Board The low speed starter board is mounted on the power input board and supplies the boost and run voltages to R type stator tubes. The boost time of approximately 1.5 seconds is controlled by a timer circuit on the auxiliary board. This timer starts when the prep command is received, causing the low speed starter to output approximately 230 VAC to the stator windings in the X-ray tube to rapidly bring the stator up to rated speed. After the 1.5 second boost time has expired, the stator voltage drops to approximately 50 VAC to maintain this speed for the duration of the X-ray exposure. Current sense relays on the low speed starter board monitor the current in both the main and shift windings. If either of these currents is low or absent, a rotor fault signal is sent to the auxiliary board and then to the power supply control board for processing.

Page 127: Indico 100L Service Manual - 740914

CPI Canada Inc Theory of Operation 7

X-Ray Generator Service Manual Ch # 740914-07 Rev. A Page 7-3

7.1.4 Auxiliary Board The auxiliary board contains the + and - 12 VDC supplies, and the + and - 35 VDC supplies, as well as the undervoltage detection circuit for the + and - 12 VDC supplies. The undervoltage fault signal is sent to the power supply control board for further processing by logic circuits on that board. The auxiliary board also processes the signal from the DC bus voltage detector on the power input board. If the DC bus voltage detector senses that the DC bus capacitors are not charging normally, circuits on the auxiliary board will prevent the the main line contactor from closing. The P/S not ready fault signal from the auxiliary board is routed to the power supply control for further processing. The auxiliary board also contains the rotor boost timer and rotor fault detector circuits as discussed in 7.1.3.

7.1.5 Filament Board The filament board contains the filament control, filament driver, filament monitoring, and filament fault circuits. The filament control circuit generates drive signals which cause the filament driver to produce filament primary current. This filament primary current is switched by a relay on the filament board. The primary current is then routed to either the large filament transformer, or to the small filament transformer in the HT tank. The large filament transformer supplies filament current to the large filament in the X-ray tube, while the small filament transformer supplies filament current to the small filament in the X-ray tube. The large / small switching command for the filament select relay originates on the console CPU board, and is routed through the power supply control board to the filament board. The filament current monitoring circuit samples the filament primary current, and is used to adjust the drive signal such that the actual filament current is equal to the demanded filament current. The filament current monitoring circuit will also generate a fault signal if the filament current is less than approximately 2 amps. This filament fault signal is processed by the power supply control board.

7.1.6 Power Supply Control Board The control signals from the CPU board in the console are routed directly to the power supply control board. Logic circuits on the power supply control board ensure that the proper signals and commands from the console are present before an X-ray exposure can be made. These logic circuits also monitor key fault indicators and inhibit an exposure if any of these faults are present. Signals indicating the presence of faults are sent by the power supply control board back to the console, where an error message will be displayed on the console LCD screen. The faults that are reported by the power supply control board include rotor fault, filament fault, tube arc fault (which as noted in Chapter 5 may be generated by any one of five fault detector circuits), and power supply not ready fault. Additionally, a P/S fault signal is generated that will inhibit exposures if any of the previous faults is present, or if the + or - 12 VDC supplies are low. Circuits on the power supply control board generate drive pulses for the inverter board. The power supply control board also contains kV feedback and monitoring circuits which constantly monitor the kVp output and adjust the drive pulses to the inverter such that the actual kVp is equal to the kVp demanded. Therefore, the power supply control board performs the kVp regulation function. Lastly, the power supply control board contains mA feedback and monitoring circuits. These circuits monitor the X-ray tube current and send a voltage which is proportional to the tube current back to the console where it is used by the console to automatically regulate the mA.

Page 128: Indico 100L Service Manual - 740914

7 Theory of Operation CPI Canada Inc

Page 7-4 Rev. A X-Ray Generator Service Manual Ch # 740914-07

7.1.7 Inverter Board The inverter board contains the high current FETs which drive the resonant circuit and the primaries of the HT transformers in the HT tank. The power supply control board generates the inverter drive signal. This drive signal causes the full bridge configured FETs to switch the nominal 600 VDC bus voltage to the HT transformer primaries. The secondary voltage produced by the HT transformers is determined by the frequency at which the DC bus voltage applied to the HT transformers is switched. This is governed by the control board, and is determined by comparing the actual kVp to the demanded kVp and constantly making corrections such that the demanded kVp is equal to the requested kVp.

7.1.8 DC Bus Assembly The DC bus assembly contains the DC bus capacitors and the mains rectifier. The DC bus capacitors are precharged via current limiting resistors on the power input board and the mains rectifier on the DC bus assembly when the generator is switched on. This limits the inrush current drawn by the generator at initial turn-on. Once the capacitors are sufficiently pre-charged, the main contactor will supply mains voltage to the DC bus assembly where it is rectified and filtered by a voltage doubler circuit which produces approximately 600 VDC. This nominal 600 VDC output from the DC bus assembly is fed to the inverter board as described above.

7.1.9 Customer Interface Board The customer interface board contains terminal blocks for the table and wall Bucky, for the S.I.D. interlocks, and for the X-ray tube thermal switch. Also on this board is a terminal block for the 24 VAC output for the collimator lamp and for the 24 VDC output for system locks. The customer interface board connects to the CPU board in the console via a 25 conductor cable.

7.1.10 CPU Board The CPU board in the console contains the microprocessor, EPROM, EEPROM, analog to digital and digital to analog converters, memory decoders and associated circuits. Analog circuits send and receive analog information to and from the power supply control board (kVp demand, kVp feedback, filament demand, mA feedback), and digital circuits send and receive digital information to and from the power supply control board (example X-ray command, prep command, rotor fault, filament fault etc). Signals from the CPU board are also sent to and from the room interface board (room interlocks, thermal switch, Bucky drive and ready signals), to and from the AEC board, to and from the keypad, and to and from the console LCD display.

7.1.11 AEC Board (Optional) The ramp from the AEC chamber is buffered by the AEC board input circuits, then it is fed to a gain stage for further processing. The ramp voltage is then routed to the CPU board, where it is monitored by the CPU. The CPU calculates in software the expected ramp voltage for each technique. The exposure is then terminated by the console when the magnitude of the ramp equals the precalculated value. The AEC board also contains switching circuits which perform AEC field selection, and switch the AEC ramp from either the table Bucky or the wall Bucky to the common buffer and amplifier circuits.

Page 129: Indico 100L Service Manual - 740914

CPI Canada Inc Theory of Operation 7

X-Ray Generator Service Manual Ch # 740914-07 Rev. A Page 7-5

7.2.0 HT TANK The HT tank contains two HT transformers, one for the anode and one for the cathode. The cathode transformer along with its rectifier and filter capacitors generates up to - 62.5 kVp for 125 kV generators, or - 75 kVp for 150 kV generators. The anode transformer along with its rectifier and filter capacitors generates up to + 62.5 kVp for 125 kV generators, or + 75 kVp for 150 kV generators. The kVp output and the X-ray tube current is sampled and fed back to the power supply control board via J9 on the HT tank. The HT tank also has an mA / mAs test jack E17 - E18 which allows the actual tube current to be measured if the shorting link is removed and an approved mA / mAs meter is connected. Filament transformers are also located in the HT tank. The output from the filament board is applied to the primaries of these transformers, the secondaries then couple the filament current to the filaments in the X-ray tube. The filaments float at cathode potential, therefore the filament transformer secondaries also provide the necessary high voltage isolation of the cathode. The HT tank is not field serviceable. The oil must be vacuum filled to purge all trapped air, otherwise the tank will not be able to withstand the rated kVp. This cannot be done in the field.

Page 130: Indico 100L Service Manual - 740914

7 Theory of Operation CPI Canada Inc

Page 7-6 Rev. A X-Ray Generator Service Manual Ch # 740914-07

(This page intentionally left blank)

Page 131: Indico 100L Service Manual - 740914

CPI Canada Inc Spares 8

CHAPTER 8

SPARES

CONTENTS:

Section Title 8.1.0 INTRODUCTION ................................................................................................................................................. 2 8.2.0 SPARE PARTS LIST GENERATOR MAIN CABINET ........................................................................................ 2 8.3.0 SPARE PARTS LIST OPERATOR CONSOLE................................................................................................... 2 8.4.0 SPARE PARTS LIST CONSOLE CABLES......................................................................................................... 2 8.5.0 SPARE PARTS LIST FUSES.............................................................................................................................. 3

X-Ray Generator Service Manual Ch # 740914-08 Rev. L Page 8-1

Page 132: Indico 100L Service Manual - 740914

8 Spares CPI Canada Inc

8.1.0 INTRODUCTION This chapter contains the list of spare parts for the 30, 32.5, 37.5 kW single phase and three phase X-ray generators, and for the 50 kW three phase X-ray generator.

8.2.0 SPARE PARTS LIST GENERATOR MAIN CABINET

DESCRIPTION PART NUMBER NOTE SUGGESTED QTY Power input board Consult factory N/A N/A Resonant Board Consult factory N/A N/A Low speed starter board See note ⇒ 7 1 Auxiliary board 732221-02 2 1 Filament board 731407-00 1 1 Power supply control board 732816-03 3 1 Inverter board Consult factory N/A N/A Customer interface board 735139-00 1 1 Transformer, aux power supply 732417-00 1 1 Transformer, room interface (1 phase units only)

735889-00 N/A 1

Transformer, room interface (3 phase units only)

732415-00 N/A 1

HT tank assembly (complete) Consult factory N/A N/A REFER TO 8.5.0 FOR FUSES

8.3.0 SPARE PARTS LIST OPERATOR CONSOLE

DESCRIPTION PART NUMBER NOTE SUGGESTED QTY CPU board See note ⇒ 5 ⇐ See note AEC board 735206-00 4 ⇐ See note Battery, lithium 3.0V 7412290100 1 1

8.4.0 SPARE PARTS LIST CONSOLE CABLES

DESCRIPTION PART NUMBER NOTE SUGGESTED QTY Cable assembly, 25 conductor, 25 ft 735169-00 6 ⇐ See note Cable assembly, 37 conductor, 25 ft 735170-00 6 ⇐ See note Cable assembly, 25 conductor, 50 ft 735169-01 6 ⇐ See note Cable assembly, 37 conductor, 50 ft 735170-01 6 ⇐ See note

Page 8-2 Rev. L X-Ray Generator Service Manual Ch # 740914-08

Page 133: Indico 100L Service Manual - 740914

CPI Canada Inc Spares 8

8.5.0 SPARE PARTS LIST FUSES FUSE LOCATION FUSE TYPE & PART NO. SUGGESTED QTY Auxiliary board: F1, F2 (12A 250V Slo-Blo) MDA-12, 6713746500 5 Auxiliary board: F3, F4 (7A 250V Slo-Blo) MDA-7, 6713000500 5

Filament board: F1, F2 (4A 250V Slo-Blo)

Customer interface board: F4

GDC-4, 5550033100 5

Low speed starter board: F1, F3 (8A 250V Slo-Blo) S506-8A, 5550037500 5

Low speed starter board: F2 (5A 250V Slo-Blo) GDC-5, 5550035600 5 Power input board. 1 phase: F1, F2 (10A 250V Slo-Blo)

MDA-10, 6713000200 5

Power input board. 1 phase: F4 (3A 250V Slo-Blo) MDL-3, 6713746700 5 Power input board. 1 phase: F6 (1.5A 250V Slo-Blo)

MDL-1.5, 6713541500 5

Power input board. 1 phase: F5 (A, B) 100A 250VAC

NLN-100, 6711906500 5

Power input board. 3 phase: F1, F2 (10A 500V Slo-Blo)

FNQ-10, 6711905500 5

Power input board. 3 phase: F4 (2A 250V Slo-Blo) GDC-2, 5550032600 5

Power input board. 3 phase: F5 (A, B, C) 60A 600VAC

OTS-60, SC3434 5

Customer interface board: F1, F5 (1.6A 250V Slo-Blo)

GDC-1.6, 5550033300 5

Customer interface board: F2 (3.15A 250V Slo-Blo) GDC-3.15, 5550036600 5

Customer interface board: F3 (6.3A 250V Slo-Blo) GDC-6.3, 5550033400 5

Primary of 24V interface / console power transformer: F2 (three phase units only) 1A 500V

FNQ-1, 6711905800 5

Primary of power supply auxiliary transformer: F1 (single phase units) 3A 250V

FNM-3, 6711907400 5

Primary of power supply auxiliary transformer: F1 (three phase units) 2A 500V

FNQ-2, 5550005300 5

NOTE: 1. The part number shown is used on all versions of this X-ray generator. These parts are not model or

option dependent.

2. The part number shown is the suggested replacement for the original auxiliary board. The spares board is “jumper configurable” in the field, and as such will replace the original auxiliary board. This is intended to eliminate the need to stock multiple configurations of this board.

3. The part number shown is the suggested replacement for the original power supply control board. The spares board is “full featured” and will replace the original board in your generator regardless of configuration. This is intended to eliminate the need to stock multiple configurations of this board.

X-Ray Generator Service Manual Ch # 740914-08 Rev. L Page 8-3

Page 134: Indico 100L Service Manual - 740914

8 Spares CPI Canada Inc

4. This item only used on generators fitted with the optional AEC board. Spares should be stocked accordingly.

5. Replacement generator CPU board part number 735225-02 should be ordered for spares use. This board will need jumpers JW7 and JW8 removed for 30 - 37.5 kW generator usage per separate instruction shipped with spare boards. Refer to the table below for spares application of this board.

ORIGINAL CPU BOARD REPLACEMENT CPU BOARD REMARKS 735133-00, 02 735225-02 Note (a) below 735133-03, 04 735225-02 Note (b) below 735225-00, 02 735225-02 Note (c) below

Notes:

(a) 735225-02 will replace 735133-00, 735133-02 BUT new firmware will be required along with the new board. Please consult the factory for details.

JW7 and JW8 on the replacement board must be correctly set, refer to separate instruction shipped with the spare board.

(b) 735225-02 will directly replace 735133-03, 735133-04, new firmware is not required. JW7 and JW8 on the replacement board must be correctly set, refer to separate instruction shipped with the spare board.

(c) 735225-02 is the exact replacement for 735225-00 and 735225-02 when JW7 and JW8 on the replacement board are correctly set, refer to separate instruction shipped with the spare board.

6. Subject cables are available in 25 foot and 50 foot lengths. Spares should be ordered and stocked to suit the desired cable lengths.

7. The low speed starter part number with a 33µF phase shift capacitor for standard “R” type stators is 732752-00. The part number with a 12.5µF phase shift capacitor is 732752-01, and the part number with a 45µF phase shift capacitor is 732752-02.

Page 8-4 Rev. L X-Ray Generator Service Manual Ch # 740914-08

Page 135: Indico 100L Service Manual - 740914

CPI Canada Inc Schematics 9

CHAPTER 9

SCHEMATICS

CONTENTS:

9.1.0 INTRODUCTION ...........................................................................................................................................9-2 9.2.0 FUNCTIONAL SCHEMATIC INDEX..............................................................................................................9-2

X-Ray Generator Service Manual Ch # 740914-09 Rev. E Page 9-1

Page 136: Indico 100L Service Manual - 740914

9 Schematics CPI Canada Inc

9.1.0 INTRODUCTION

This chapter contains the functional schematics for your X-ray generator. Each schematic represents a major function in the generator; the eight functional schematics in this chapter represent all of the major functional blocks in this generator.

9.2.0 FUNCTIONAL SCHEMATIC INDEX

The following functional schematics are included in this manual.

DESCRIPTION DRAWING NUMBER#

DC Bus and Power Distribution MD-0809 System ON MD-0810 Customer Interface MD-0811 X-Ray Exposure MD-0812 KV Control and Feedback MD-0813 Filament Drive and mA Control MD-0814 Low Speed Starter MD-0815 AEC MD-0816

Page 9-2 Rev. E X-Ray Generator Service Manual Ch # 740914-09