IQ plus® 355Digital Weight Indicator

Version 1.1

Installation Manual

66503

Page 2: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual
Page 3: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Copyright © 2002 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America. Specifications subject to change without notice.

Technical training seminars are available through Rice Lake Weighing Systems. Course descriptions and dates can be viewed at www.rlws.com or obtained by

calling 715-234-9171 and asking for the training department.

Contents

About This Manual ................................................................................................................................... 11.0 Introduction.................................................................................................................................. 1

1.1 Operating Modes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.2 Front Panel Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.3 LED Annunciators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.4 Indicator Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

1.4.1 Toggle Gross/Net Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.2 Toggle Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.3 Zero Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.4 Acquire Tare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.5 Remove Stored Tare Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31.4.6 Print Ticket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2.0 Installation ................................................................................................................................... 42.1 Unpacking and Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.2 Enclosure Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42.3 Cable Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2.3.1 Cable Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52.3.2 Load Cells. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.3.3 Serial Communications and Digital Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.3.4 Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

2.4 Analog Output Module Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.5 Enclosure Reassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.6 Board Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72.7 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

3.0 Configuration.............................................................................................................................. 123.1 Configuration Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

3.1.1 Revolution Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123.1.2 EDP Command Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123.1.3 Front Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13

3.2 Menu Structures and Parameter Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143.2.1 Configuration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153.2.2 Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 173.2.3 Calibration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193.2.4 Serial Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193.2.5 Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203.2.6 Print Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 213.2.7 Digital Input Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223.2.8 Analog Output Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233.2.9 Version Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

4.0 Calibration.................................................................................................................................. 244.1 Front Panel Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244.2 EDP Command Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254.3 Revolution™ Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254.4 More About Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Version 1.1, October 2002

Page 4: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

4.4.1 Adjusting Final Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.4.2 Zero Deadload A/D Counts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264.4.3 Calculating the Span Coefficient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

5.0 EDP Commands.......................................................................................................................... 275.1 The EDP Command Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

5.1.1 Key Press Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 275.1.2 Reporting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285.1.3 The RESETCONFIGURATION Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285.1.4 Parameter Setting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285.1.5 Normal Mode Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30

5.2 Saving and Transferring Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315.2.1 Saving Indicator Data to a Personal Computer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 315.2.2 Downloading Configuration Data from PC to Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

6.0 Print Formatting ......................................................................................................................... 326.1 Print Formatting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326.2 Customizing Print Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

6.2.1 Using the EDP Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 326.2.2 Using the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 336.2.3 Using Revolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

7.0 Appendix .................................................................................................................................... 347.1 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

7.1.1 Displayed Error Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 347.1.2 Using the XE EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

7.2 Status Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357.2.1 Using the P EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 357.2.2 Using the ZZ EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

7.3 Continuous Output (Stream) Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 367.4 ASCII Character Chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 377.5 Front Panel Display Characters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 397.6 Conversion Factors for Secondary Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 407.7 Digital Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

7.7.1 DIGFLx Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417.7.2 DFSENS and DFTHRH Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 417.7.3 Setting the Digital Filter Parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42

7.8 Analog Output Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 427.9 Test Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 437.10 Software Revision History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 447.11 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45

IQ plus 355 Limited Warranty ................................................................................................................ 46

ii IQ plus 355 Installation Manual

Page 5: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

About This ManualThis manual is intended for use by service techniciansresponsible for installing and servicing IQ plus® 355digital weight indicators. This manual applies toindicators using Version 1.1 of the IQ plus 355software. See Section 7.10 on page 44 for a summaryof software changes included in this release.

Configuration and calibration of the indicator can beaccomplished using the indicator front panel keys, theEDP command set, or Version 2.5 or later of theRevolution™ configuration utility. See Section 3.1 onpage 12 for information about configuration methods.

Some procedures described in thismanual require work inside theindicator enclosure. These proceduresare to be performed by qualifiedservice personnel only.

Authorized distributors and their employeescan view or download this manual from theRice Lake Weighing Systems distributorsite at www.rlws.com.

The Operator Card included with this manualprovides basic operating instructions for users of theIQ plus 355. Please leave the Operator Card with theindicator when installation and configuration arecomplete.

1.0 IntroductionThe IQ plus 355 is a single-channel digital weightindicator housed in a NEMA 4X/IP66-rated stainlesssteel enclosure. The indicator front panel consists of alarge (.8 in, 20 mm), six-digit, seven-segment LEDdisplay and five-button keypad. Features include:

• Drives up to eight 350Ω or sixteen 700Ω loadcells

• Supports 4- and 6-wire load cell connections• Two configurable digital inputs• Electronic data processing (EDP) port for full

duplex, RS-232 communications at up to9600 bps

• Printer port for output-only RS-232 and 20mA current loop communications at up to9600 bps

• Optional analog output module provides 0–10VDC or 4–20 mA tracking of gross or netweight values

• Available in 115 VAC and 230 VAC versionsThe IQ plus 355 is NTEP-certified and MeasurementCanada approved for Classes III, III HD, and III L at10,000 divisions. See Section 7.11 on page 45 fordetailed specifications.

1.1 Operating ModesThe IQ plus 355 has three modes of operation:

Normal (weighing) modeNormal mode is the “production” mode of theindicator. The indicator displays gross or netweights as required, using the LED annunciatorsdescribed in Section 1.3 on page 2 to indicatescale status and the type of weight valuedisplayed. Once configuration is complete and alegal seal is affixed to the back of the indicator,this is the only mode in which the IQ plus 355 canoperate.

Setup modeMost of the procedures described in this manualrequire the indicator to be in setup mode,including configuration and calibration.To enter setup mode, remove the large fillisterhead screw from the enclosure backplate. Insert ascrewdriver or a similar tool into the access holeand press the setup switch once. The indicatordisplay changes to show the word CONFIG.

Test modeTest mode provides a number of diagnosticfunctions for the IQ plus 355 indicator. Like setupmode, test mode is entered using the setup switch.See Section 7.9 on page 43 for more informationabout entering and using test mode.

Warning

Introduction 1

Page 6: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

1.2 Front Panel KeypadFigure 1-1 on page 2 shows the IQ plus 355 LED annunciators, keypad, and normal mode key functions.

The symbols shown under the keys (representing up, down, enter, left, right) describe the key functions assignedin setup mode. In setup mode, the keys are used to navigate through menus, select digits within numeric values,and increment/decrement values. See Section 3.1.3 on page 13 for information about using the front panel keysin setup mode.

Figure 1-1. IQ plus 355 Front Panel, Showing LED Annunciators and Normal Mode Key Functions

1.3 LED AnnunciatorsThe IQ plus 355 display uses a set of eight LED annunciators to provide additional information about the valuebeing displayed:

• Gross and Net annunciators are lit to show whether the displayed weight is a gross or net weight.

• Center of zero ( ): Gross weight is within 0.25 graduations of zero. This annunciator lights when thescale is zeroed.

• Standstill ( ): Scale is at standstill or within the specified motion band. Some operations, includingtare functions and printing, can only be done when the standstill symbol is shown.

• lb, kg, oz, and g annunciators indicate the units associated with the displayed value: lb=pounds,kg=kilograms, oz=ounces, g=grams. The displayed units can also be set to short tons (tn), metric tons (t), or NONE (no units informationdisplayed). The lb and kg LEDs function as primary and secondary units annunciators for somecombinations of primary and secondary units. If neither primary nor secondary units are lb, kg, oz, or g, thelb annunciator is lit for primary units, kg for secondary units.

D I G I T A L W E I G H T I N D I C A T O R

Switch betweenprimary and

secondary units

Send data toserial port

Set gross weightto zero

Switch betweengross and netweight display

Acquire tare

NORMAL MODE KEY FUNCTIONS

ZERO GROSSNET

TARE UNITS PRINT

TB/N Units

lb

kg

oz

g

Gross

Net

2 IQ plus 355 Installation Manual

Page 7: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Table 1-1 shows which annunciators are used for all combinations of configured primary and secondary units.For example:

• If the primary unit is pounds (lb) and the secondary unit is kilograms (kg), the lb LED is lit for primaryunits, kg for secondary units.

• If the primary unit is pounds (lb) and the secondary unit is short tons (tn), the lb LED is lit for primaryunits, kg for secondary units. There is no LED for short tons, so the kg LED is used as the secondaryunits annunciator.

• If the primary unit is short tons (tn) and the secondary unit is pounds (lb), the lb LED is lit for primaryunits (tn), and kg is lit for secondary units (lb). Because there is no LED for short tons, the lb and kgLEDs are used as primary and secondary units annunciators.

See Section 3.2.2 for more information about configuring primary and secondary display units.

1.4 Indicator OperationsBasic IQ plus 355 operations are summarized below.

1.4.1 Toggle Gross/Net ModePress the GROSS/NET key to switch the display modefrom gross to net, or from net to gross. If a tare valuehas been entered or acquired, the net value is the grossweight minus the tare.

Gross mode is shown by the Gross annunciator; netmode is shown by the Net annunciator.

1.4.2 Toggle UnitsPress the UNITS key to switch between primary andsecondary units. The units LED to the right of thedisplay is lit.

1.4.3 Zero Scale1. In gross mode, remove all weight from the

scale and wait for the standstill annunciator( ).

2. Press the ZERO key. The center of zero( ) annunciator lights to indicate thescale is zeroed.

1.4.4 Acquire Tare1. Place container on scale and wait for the

standstill annunciator ( ).2. Press the TARE key to acquire the tare weight

of the container. The indicator switches to netmode.

1.4.5 Remove Stored Tare Value1. Remove all weight from the scale and wait for

the standstill annunciator ( ).2. Press the TARE key. The indicator switches to

gross mode, indicating the tare value has beenremoved.

1.4.6 Print Ticket1. Wait for the standstill annunciator ( ).2. Press the PRINT key to send data to the serial

port.

Primary Unit

Secondary Unit

lb kg oz g tn t none

lb lb / lb lb / kg lb / oz lb / g lb / kg

kg kg / lb kg / kg kg / oz kg / g lb / kg

oz oz / lb oz / kg oz / oz oz / g oz / kg

g g / lb g / kg g / oz g / g g / kg

tn lb / kg lb / kg lb / oz lb / g lb / lb lb / kg lb / kg

t lb / kg lb / lb lb / kg

none lb / kg lb / kg lb / lb

Table 1-1. Units Annunciators, Showing Primary / Secondary LEDs Used for All Configurations

Introduction 3

Page 8: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

2.0 InstallationThis section describes procedures for connecting loadcells, digital inputs, and serial communications cablesto the IQ plus 355 indicator. Instructions for fieldinstallation of the analog output option andreplacement of the CPU board are included, alongwith assembly drawings and parts lists for the servicetechnician.

• Use a wrist strap to ground yourself and protectcomponents from electrostatic discharge (ESD)when working inside the indicator enclosure.

• This unit uses double pole/neutral fusing whichcould create an electric shock hazard. Proceduresrequiring work inside the indicator must beperformed by qualified service personnel only.

• The supply cord serves as the power disconnect forthe IQ plus 355. The power outlet supplying theindicator must be installed near the unit and beeasily accessible

2.1 Unpacking and AssemblyImmediately after unpacking, visually inspect the IQplus 355 to ensure all components are included andundamaged. The shipping carton should contain theindicator with attached tilt stand, this manual, and aparts kit. If any parts were damaged in shipment,notify Rice Lake Weighing Systems and the shipperimmediately.

The parts kit contains the items listed below:

• Capacity, identification, and annunciatorlabels. Annunciator labels (PN 53374)provide replacement overlay decals forlabeling primary and secondary units LEDs.

• 6-position screw terminal (PN 70599) forconnector J1 and a 7-position screw terminal(PN 70600) for connector J4 (see Figure 2-3on page 5).

• Two 8-32NC x 7/16 fillister head screws (PN30623). These screws occupy the holes belowand on either side of the setup screw on theindicator backplate (see Figure 2-4 on page7).

• Four 8-32NC x 3/8 machine screws (PN14862) for the indicator backplate (see #24 inFigure 2-5 on page 9).

• Six neoprene washers (PN 45042) forbackplate screws included in the parts kit.

• Four rubber bumpers (“feet”) for the tilt stand,PN 42149.

• Five cable ties, PN 15631.• Two ferrites (PN 66730), used to reduce

susceptibility to radiate electromagnet

interference and EMI installation instructions(PN 67970).

• Three each of grounding clamps (PN 67550),external tooth lock washers (PN 15133), andkep nuts (PN 14676) for cable shieldgrounding against the backplate.

2.2 Enclosure DisassemblyThe indicator enclosure must be opened to connectcables for load cells, communications, digital inputs,and analog output.

The IQ plus 355 has no on/off switch.Before opening the unit, ensure thepower cord is disconnected from thepower outlet.

Ensure power to the indicator is disconnected, thenplace the indicator face-down on an antistatic workmat. Remove the screws that hold the backplate to theenclosure body, then lift the backplate away from theenclosure and set it aside.

2.3 Cable ConnectionsThe IQ plus 355 provides four cord grips for cablinginto the indicator: one for the power cord, three toaccommodate load cell, communications, digitalinputs, and analog output cables. Two of the three freecord grips come with a plug installed to preventmoisture from entering the enclosure. Depending onyour application, remove the plug from any cord gripthat will be used and install cables as required.

NOTE: Because the IQ plus 355 has no on/off switch,the power cord serves as the power disconnect. Thepower outlet must be located close enough to theindicator to allow the operator to easily disconnectpower to the unit.

Figure 2-1 shows the recommended assignments forthe IQ plus 355 cord grips.

Figure 2-1. Recommended Cord Grip Assignments

Caution

Warning

AC Pow

er C

ord

Load

Cell

Cab

le

Serial

Com

mun

icatio

ns

Analog

Out

put,

Digital

Inpu

t

4 IQ plus 355 Installation Manual

Page 9: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

2.3.1 Cable GroundingExcept for the power cord, all cables routed throughthe cord grips should be grounded against theindicator backplate. Do the following to groundshielded cables:

• Use the lockwashers, clamps, and kep nutsprovided in the parts kit to install groundingclamps on the backplate studs adjacent to cordgrips. Install grounding clamps only for cordgrips that will be used; do not tighten nuts.

• Route cables through cord grips and groundingclamps to determine cable lengths required toreach cable connectors. Mark cables to removeinsulation and shield as described below:• For cables with foil shielding, strip insulation

and foil from the cable half an inch (15 mm)past the grounding clamp (see Figure 2-2).Fold the foil shield back on the cable wherethe cable passes through the clamp. Ensuresilver (conductive) side of foil is turnedoutward for contact with the groundingclamp.

• For cables with braided shielding, strip cableinsulation and braided shield from a point justpast the grounding clamp. Strip another halfinch (15 mm) of insulation only to expose thebraid where the cable passes through theclamp (see Figure 2-2).

• For load cell cables, cut the shield wire just pastthe grounding clamp. Shield wire function isprovided by contact between the cable shieldand the grounding clamp.

• Route stripped cables through cord grips andclamps. Ensure shields contact groundingclamps as shown in Figure 2-2. Tightengrounding clamp nuts.

• Finish installation using cable mounts and ties tosecure cables inside of indicator enclosure.

Figure 2-2. Grounding Clamp Attachment for Foil-Shielded and Braided Cabling

Figure 2-3. IQ plus 355 CPU and Power Supply Board

Cord grip

Insulated cable

Foil (silver side out) Grounding clamp

Shield wire (cut)

Length of foil before foldingback on cable insulation

Cut insulation herefor foil-shielded cables

Braid

Cut insulation herefor braided cables

NOTE: Install lockwashersfirst, against backplate,under grounding clamp

Transformer

BR

1

C23 VR1

JMP4

C24

MECCA1

C66

DIGITAL+5V

TEST

J41 2 3 4 5 6 7

SERIAL COMM & DIGITAL IN

ED

PT

GN

D

ED

PR

PR

MA

PR

T

IN2

IN1

F2 F1

Blu

e w

ire

JP4

ANALOGGND

ANALOG+5V

TEST

ANALOG–5V

TEST

+

+

GND

GND

JP1

JP2

+

VR2

C25 +

VR

3

C67

+

–+

Keypad Connector

J7

EEPROM

A/D Conver ter

JP7

Bro

wn

wire

Microprocessor

X2

LED

DR

IVE

R

LED

DR

IVE

R

U4U5To setup switch

ON OFF

SET/ISP

J1

–EXC

+EXC

–SENSE

+SENSE

–SIG

+SIG

6

5

4

3

2

1

LOA

D C

ELL

CO

NN

EC

TOR

Load cellcompensation jumper

Installation 5

Page 10: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

2.3.2 Load CellsTo attach cable from a load cell or junction box,remove connector J1 from the board. The connectorplugs into a header on the board as shown inFigure 2-6 on page 10.

NOTE: Early versions of the IQ plus 355 CPU board useda 7-pin load cell connector. The 7-pin connector is notcompatible with the current 6-pin header.

Wire the load cell cable from the load cell or junctionbox to connector J1 as shown in Table 2-1. If using6-wire load cell cable (with sense wires), removejumpers JP1 and JP2 before reinstalling connector J1(see Figure 2-3). For 4-wire installation, leavejumpers JP1 and JP2 on.

When connections are complete, reinstall connectorJ1 onto the header so that it snaps securely into place.Use two cable ties to secure the load cell cable to theinside of the enclosure.

Setting the Load Cell Compensation JumperThe load cell compensation jumper (above the A/Dconverter location on the CPU board; see Figure 2-3on page 5) must be set ON for load cells withunbalanced bridges. The compensation jumper has theeffect of lowering the positive excitation voltage.Uncompensated unbalanced load cells can causeinstability or calibration errors.

For RL1040 and RL1042 load cells, set thecompensation jumper as follows:

• RL1040 load cells: jumper OFF• RL1042 load cells: jumper ON

For other load cell types, use the following procedureto determine the correct jumper position;

1. Disconnect load cell from indicator and usean ohmmeter to measure the following:

• +EXC to +SIG, +EXC to –SIG• –EXC to +SIG, –EXC to –SIG

Measured values between the excitation lineand each of the signal lines should be within2–3Ω.

2. If the +EXC measurements are ≥ 5% largerthan the –EXC measurements , set thecompensation jumper in the ON position. Ifthe +EXC measurements are < 5% greater (orare less) than the –EXC measurements, set thejumper in the OFF position.

2.3.3 Serial Communications and Digital Inputs To attach serial communications and digital inputcables, remove connector J4 from the board.Connector J4 provides connections for the EDP(Electronic Data Processing) port, printer port, andtwo digital inputs. Connect communications anddigital input cables to connector J4 as shown inTable 2-2.

Once cables are attached, reconnect J4 to the headeron the board (see Figure 2-6 on page 10). Use cableties to secure serial and digital input cables to theinside of the enclosure.

The EDP port supports RS-232 communications only;the printer port provides either active 20 mA output orRS-232 transmission. Both ports are configured usingthe SERIAL menu. See Section 3.0 on page 12 forconfiguration information.

Digital inputs can be set to provide several indicatorfunctions, including all keypad functions. The inputsare active (on) with low voltage (0 VDC) and can bedriven by TTL or 5V logic without additionalhardware. Use the DIG IN menu to configure thedigital inputs.

J1 Pin Function

1 +SIG

2 –SIG

3 +SENSE

4 –SENSE

5 +EXC

6 –EXC

NOTES: • Use grounding procedure described in Section 2.3.1

on page 5.• For 6-wire connections, remove jumpers JP1 and JP2.

Table 2-1. J1 Pin Assignments

Port J4 Pin Label Function

EDP Port 1 EDPT RS-232 TxD

2 GND RS-232 Ground / –20 mA OUT

3 EDPR RS-232 RxD

Printer Port 4 PRMA +20 mA OUT

5 PRT RS-232 TxD

Digital Inputs 6 IN2 Digital Input 2

7 IN1 Digital Input 1

Table 2-2. J4 Pin Assignments

6 IQ plus 355 Installation Manual

Page 11: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

2.3.4 Analog OutputIf the optional analog output module is installed,attach the output cable to connector J1 on the analogoutput board. Table 2-3 lists the analog output pinassignments.

Use the ALGOUT menu to configure and calibrate theanalog output module when cabling is complete. SeeSection 2.4 for information about installing the analogoutput module.

2.4 Analog Output Module InstallationTo install or replace the analog output module, followthe steps listed in Section 2.2 on page 4 for openingthe IQ plus 355 enclosure.

Mount the analog output module on its standoffs inthe location shown in Figure 2-3 on page 5 and plugthe module input into connector J5 on the IQ plus 355board. Connect output cable to the analog outputmodule as shown in Table 2-3, then reassemble theenclosure (Section 2.5).

See Section 7.8 on page 42 for analog outputcalibration procedures.

2.5 Enclosure ReassemblyOnce cabling is complete, position the backplate overthe enclosure and reinstall the backplate screws. Usethe torque pattern shown in Figure 2-4 to preventdistorting the backplate gasket. Torque screws to 15in-lb (1.7 N-m).

Figure 2-4. IQ plus 355 Enclosure Backplate

2.6 Board RemovalIf you must remove the IQ plus 355 CPU board, usethe following procedure:

1. Disconnect power to the indicator. Loosencord grips and remove backplate as describedin Section 2.2 on page 4.

2. Unplug connectors J1 (load cell cable), J4(serial communications and digital inputs), J7(keypad ribbon cable), and JP4 (setup switch).Remove blue and brown power input wires atJP7. If an analog output board is installed,disconnect the analog output cable. SeeFigure 2-3 on page 5 for connector locations.

3. Remove the four nuts from the corners of theCPU board, then lift the board out of theenclosure.

To replace the CPU board, reverse the aboveprocedure. Be sure to reinstall cable ties to secure allcables inside the indicator enclosure.

Pin Signal

1 + Current Out

2 – Current Out

3 + Voltage Out

4 – Voltage Out

Table 2-3. Analog Output Module Pin Assignments

Setup switch access screw

Fillister head screws

10

9

8

7

4

3

1

2

5 6

Torque backplate screws to 15 in- lb (1.7 N-m)

Torque pattern

Installation 7

Page 12: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

2.7 Replacement PartsTable 2-4 lists replacement parts for the IQ plus 355, including all parts referenced in Figures 2-5 through 2-9.

Ref Number PN Description (Quantity) Figure

1 14626 Kep nuts, 8-32NC hex (3) Figure 2-5 on page 9, Figure 2-8 on page 11

2 52211 Display and CPU board assembly, 115 VAC (1) Figure 2-6 on page 10

52210 Display and CPU board assembly, 230 VAC (1)

3 15365 Board mounting spacers (4)

4 39017 Enclosure backplate (1) Figure 2-5 on page 9

5 15626 Cable grips, PG9 (3)

6 30375 Nylon seal rings for cable grips (3)

7 15627 Locknuts, PCN9 (3)

8 19538 Cable grip plugs (2)

9 45042* Sealing washers (4)

10 44676 Sealing washer for setup switch access screw (1)

11 42640 Setup switch access screw, 1/4 x 28NF x 1/4 (1)

12 41965 Power cord assembly, 115VAC (1) Figure 2-5 on page 9, Figure 2-7 on page 11

45254 Power cord assembly, 230VAC (1)

13 41964 Line filter assembly (1) Figure 2-7 on page 11

14 14621 Kep nuts, 6-32NC hex (4) Figure 2-6 on page 10

16 68403 Four-cornered wing knobs for tilt stand (2) Figure 2-5 on page 9

30342 Two-cornered wing knobs for tilt stand (2)

17 29635 Tilt stand (1)

18 15144 Nylon washers for tilt stand, 1/4x1x1/16 (2)

20 15134 Lock washers, internal tooth, No. 8 , Type A (4)

22 50556 Overlay membrane panel (1) Figure 2-9 on page 11

23 50555 Enclosure (1)

24 14862* Screws, 8-32NC x 3/8 (4) Figure 2-5 on page 9

26 45043 Ground wire, 4-in., No. 8 (1)

27 39037 Backplate gasket (1) Figure 2-5 on page 9

28 51974 Setup switch assembly (1) Figure 2-7 on page 11

29 16892 Ground/Earth Label (1)

30 15650* Cable tie mounts (8) Figure 2-6 on page 10

31 45302 Line filter standoffs (2) Figure 2-7 on page 11

33 15131 Lock washers, external tooth, No. 6, Type A (8) Figure 2-6 on page 10

— 70600 7-position connectors for J4, and J12 (2) Figure 2-3 on page 5

— 70599 6-position connectors for J1 (1)

— 53848 200 mA TR5 subminiature fuses (2), 115 VAC F1 and F2 in Figure 2-3 on page 5(See Caution below)53881 100 mA TR5 subminiature fuses (2), 230 VAC

* Additional parts included in parts kit.

For protection against risk of fire, replace fuses only with same type and rating fuse.See Section 7.11 on page 45 for complete fuse specifications.

Table 2-4. Replacement Parts

Caution

8 IQ plus 355 Installation Manual

Page 13: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Figure 2-5. IQ plus 355 Backplate and Tilt Stand Assemblies

PRINT

UNITS

TARE

GROSS

NET

ZERO

410

11

12

6

5

8

16

17

18

24/4X

7

27

20/4X

9/4X

To Line Filter

See Line Filterand

Ground PostAssembly

26

1/3X

Installation 9

Page 14: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Figure 2-6. IQ plus 355 Enclosure and CPU Board

2

3/4X

14/4X

30/8X

33/8X

AC power in from line filter

Cable tie

Brown wire

Blue wire

To setup switch

See JP7 detail belowJ7

J1

JP4

J4

Cable tie mountsfor line filter

Cable tie mountsfor setup switch wires

Line Filter Connection to JP7

10 IQ plus 355 Installation Manual

Page 15: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Figure 2-7. Line Filter AssemblyFigure 2-8. Ground Post Assembly

Figure 2-9. IQ plus 355 Enclosure and Overlay

AC input to CPU board JP713

29

12

Setup switch connectionto JP4 on CPU board

28

26

Ground PostAssembly

31

To power cord

To line filter

To backplateground post 1/3X

PRINT

UNITS

TARE

GROSS

NET

ZERO

22

23

To CPU board J7

Installation 11

Page 16: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.0 ConfigurationTo configure the IQ plus 355 indicator, the indicatormust be placed in setup mode. The setup switch isaccessed by removing the large fillister head screw onthe enclosure backplate. Switch position is changedby inserting a screwdriver into the access hole andpressing the switch.

When the indicator is placed in setup mode, the wordCONFIG is shown on the display. The CONFIG menuis the first of nine main menus used to configure theindicator. Detailed descriptions of these menus aregiven in Section 3.2. When configuration is complete,return to the CONFIG menu and press the (ZERO)key to exit setup mode, then replace the setup switchaccess screw.

3.1 Configuration MethodsThe IQ plus 355 indicator can be configured by usingthe front panel keys to navigate through a series ofconfiguration menus or by sending commands orconfiguration data to the EDP port. Configurationusing the menus is described in Section 3.1.3.

Configura t i on u s ing t he EDP po r t c an beaccomplished using either the EDP command setdescribed in Section 5.0 or Version 2.5 or later of theRevolution™ configuration utility.

3.1.1 Revolution ConfigurationThe Revolution configuration utility provides thepreferred method for configuring the IQ plus 355indicator. Revolution runs on a personal computer toset configuration parameters for the indicator. WhenRevolution configuration is complete, configurationdata is downloaded to the indicator.

Figure 3-1. Sample Revolution Configuration Display

Revolution supports both uploading and downloadingof indicator configuration data. This capability allowsconfiguration data to be retrieved from one indicator,edited, then downloaded to another.

To use Revolution, do the following:

1. Install Revolution on an IBM-compatiblepersonal computer running Windows® 3.11 orWindows 95. Minimum system requirementsare 4MB of extended memory and at least5MB of available hard disk space.

2. With both indicator and PC powered off,connect the PC serial port to the RS-232 pinson the indicator EDP port.

3. Power up the PC and the indicator. Use thesetup switch to place the indicator in setupmode.

4. Start the Revolution program.Figure 3-1 shows an example of one of the Revolutionconfiguration displays.

Revolution provides online help for each of itsconfiguration displays. Parameter descriptionsprovided in this manual for front panel configurationcan also be used when configuring the indicator usingRevolution: the interface is different, but theparameters set are the same.

3.1.2 EDP Command ConfigurationThe EDP command set can be used to configure theIQ plus 355 indicator using a personal computer,terminal, or remote keyboard. Like Revolution, EDPcommand configuration sends commands to theindicator EDP port ; unl ike Revolut ion, EDPcommands can be sent using any external devicecapable of sending ASCII characters over a serialconnection.

EDP commands duplicate the functions availableusing the indicator front panel and provide somefunctions not otherwise available. EDP commands canbe used to simulate pressing front panel keys, toconfigure the indicator, or to dump lists of parametersettings. See Section 5.0 on page 27 for moreinformation about using the EDP command set.

12 IQ plus 355 Installation Manual

Page 17: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.1.3 Front Panel ConfigurationThe IQ plus 355 indicator can be configured using a series of menus accessed through the indicator front panelwhen the indicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.

Figure 3-2. Front Panel Key Functions in Setup Mode

Four front panel keys are used as directional keys to navigate through the menus in setup mode (see Figure 3-2).The UNITS ( ) and PRINT ( ) keys scroll left and right (horizontally) on the same menu level; ZERO ( ) andGROSS/NET ( ) move up and down (vertically) to different menu levels. The TARE key ( ) serves as an Enterkey for selecting parameter values within the menus. A label under each of these keys identifies the directionprovided by the key when navigating through the setup menus.

Figure 3-3. Setup Mode Menu Navigation

Menu Menu Function

CONFIG Configuration Configure grads, zero tracking, zero range, motion band, overload, tare function, sample rate,and digital filtering parameters.

FORMAT Format Set format of primary and secondary units, display rate.

CALIBR Calibration Calibrate indicator. See Section 4.0 on page 24 for calibration procedures.

SERIAL Serial Configure EDP and printer serial ports.

PROGRM Program Set power-up mode, regulatory mode, and consecutive number values.

P FORMT Print Format Set print format used for gross and net tickets. See Section 5.0 for more information.

DIG IN Digital Input Assign digital input functions.

ALGOUT Analog Output Configure analog output module. Used only if analog output option is installed.

VERSION Version Display installed software version number.

Table 3-1. IQ plus 355 Menu Summary

ZERO GROSSNET

TARE UNITS PRINT

B/N Units

Move LEFT /Previous

Move RIGHT /Next

Move UP /Increment Value

Move DOWN /Decrement Value

ENTER Value

SETUP MODE KEY FUNCTIONS

1st LevelParameter

Default value

2nd LevelParameter

Value Value Value

1st LevelParameter

2nd LevelParameter

When moving through values below the first menu level, press to return to the levelabove. Press or to move to the next parameter on the level above.

Configuration 13

Page 18: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

To select a parameter, press or to scroll left orright until the desired menu group appears on thedisplay, then press to move down to the submenu orparameter you want. When moving through the menuparameters, the default or previously selected valueappears first on the display.

To change a parameter value, scroll left or right toview the values for that parameter. When the desiredvalue appears on the display, press to select thevalue and move back up one level. To edit numericalvalues, use the navigation keys to select the digit andto increment or decrement the value (see Figure 3-4).

Figure 3-4. Editing Procedure for Numeric Values

3.2 Menu Structures and Parameter DescriptionsThe following sections provide graphic representations of the IQ plus 355 menu structures. In the actual menustructure, the settings you choose under each parameter are arranged horizontally. To save page space, menuchoices are shown in vertical columns. The factory default setting appears at the top of each column.

Most menu diagrams are accompanied by a table that describes all parameters and parameter values associatedwith that menu. Default parameter values are shown in bold type.

When editing numeric values, press or to change thedigit selected. Press or to increment or decrement thevalue of the selected digit.

Press to save the value entered and return to the level above.

0 0 0 0 0 0

14 IQ plus 355 Installation Manual

Page 19: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.1 Configuration Menu

Figure 3-5. Configuration Menu

CONFIG Menu

Parameter Choices Description

Level 2 submenus

GRADS 10000number

Graduations. Specifies the number of full scale graduations. The value entered must be inthe range 1–100 000 and should be consistent with legal requirements and environmentallimits on system resolution.

To calculate GRADS, use the formula, GRADS = Capacity / Display Divisions.

Display divisions for primary and secondary units are specified on the FORMAT menu.

ZTRKBN OFF0.5D1D3D

Zero track band. Automatically zeroes the scale when within the range specified, as longas the input is within the configured zero range (ZRANGE parameter). Selections are ±display divisions. Maximum legal value varies depending on local regulations.

Table 3-2. Configuration Menu Parameters

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

GRADS

10000 OFF

1D

0.5D

3D

ZTRKBN ZRANGE

1.9%

100%

OFF

2D

1D

3D

MOTBAN

10D

5D

20D

FS+2%

FS+9D

FS+1D

FS

OVRLOA

1

4

2

8

DIGFL1

32

16

64

number

1

4

2

8

DIGFL2

32

16

64

1

4

2

8

DIGFL3

32

16

64

BOTH

PBTARE

NOTARE

KEYED

TAREFN

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

8OUT

32OUT

16OUT

64OUT

DFSENS DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD

15HZ

3.75HZ

30HZ

7.5HZ

SMPRAT

Configuration 15

Page 20: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

ZRANGE 1.9%100%

Zero range. Selects the range within which the scale can be zeroed. The 1.9% selection is± 1.9% around the calibrated zero point, for a total range of 3.8%. Indicator must be atstandstill to zero the scale. Use 1.9% for legal-for-trade applications.

MOTBAN 1D2D3D5D10D20DOFF

Motion band. Sets the level, in display divisions, at which scale motion is detected. Ifmotion is not detected for 1 second or more, the standstill symbol lights. Someoperations, including print, tare, and zero, require the scale to be at standstill. Maximumlegal value varies depending on local regulations.

If OFF is selected, ZTRKBN should also be set to OFF.

OVRLOA FS+2%FS+1DFS+9DFS

Overload. Determines the point at which the display blanks and an out-of-range errormessage is displayed. Maximum legal value varies depending on local regulations.

SMPRAT 15HZ30HZ3.75HZ7.5HZ

Sample rate. Selects measurement rate, in samples per second, of the analog-to-digitalconverter. Lower sample rate values provide greater signal noise immunity.

DIGFL1DIGFL2DIGFL3

2481632641

Digital filtering. Selects the digital filtering rate used to reduce the effects of mechanicalvibration from the immediate area of the scale.

Choices indicate the number of A/D conversions that are averaged to obtain thedisplayed reading. A higher number gives a more accurate display by minimizing theeffect of a few noisy readings, but slows down the settling rate of the indicator. SeeSection 7.7 on page 41 for more information on digital filtering.

DFSENS 8OUT16OUT32OUT64OUT128OUT2OUT4OUT

Digital filter cutout sensitivity. Specifies the number of consecutive readings that must falloutside the filter threshold (DFTHRH parameter) before digital filtering is suspended. IfNONE is selected, the filter is always enabled.

DFTHRH NONE2DD5DD10DD20DD50DD100DD200DD250DD

Digital filter cutout threshold. Specifies the filter threshold, in display divisions. When aspecified number of consecutive scale readings (DFSENS parameter) fall outside of thisthreshold, digital filtering is suspended. If NONE is selected, the filter is always enabled.

TAREFN BOTHNOTAREPBTAREKEYED

Tare function. Enables or disables push-button and keyed tares. Possible values are:

BOTH: Both push-button and keyed tares are enabledNOTARE: No tare allowed (gross mode only)PBTARE: Push-button tares enabledKEYED: Keyed tare enabled

CONFIG Menu

Parameter Choices Description

Table 3-2. Configuration Menu Parameters (Continued)

16 IQ plus 355 Installation Manual

Page 21: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.2 Format Menu

Figure 3-6. Format Menu

FORMAT Menu

Parameter Choices Description

Level 2 submenus

PRIMAR DECPNTDSPDIVUNITS

Specifies the decimal position, display divisions, and units used for the primary units. SeeLevel 3 submenu parameter descriptions.

SECNDR DECPNTDSPDIVUNITSMULTMULEXP

Specifies the decimal position, display divisions, units, and conversion multiplier used forthe secondary units. See Level 3 submenu parameter descriptions.

DSPRAT 250MS500MS750MS1SEC1500MS2SEC2500MS3SEC4SEC6SEC8SEC

Display rate. Sets the update rate for displayed values. Values are in milliseconds (MS) orseconds (SEC).

Table 3-3. Format Menu Parameters

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

DSPDIV MULTUNITSDECPNT

0.4535925D

2D

1D number

PRIMAR SECNDR

6SEC

4SEC

3SEC

2500MS

2SEC

1500MS

750MS

500MS

250MS

8SEC

DSPRAT

DSPDIV UNITSDECPNT

OZ

KG888880

888888 1D

5D

2D

LB

88888.8

8888.88

888.888

88.8888

8.88888 888880

888888

88888.8

8888.88

888.888

88.8888

8.88888

T

TN

NONE

G

TN

OZ

KG

G

T

LB

NONE

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

MULEXP

decimalposition

1SEC

Configuration 17

Page 22: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Level 3 submenus

Primary Units (PRIMAR Parameter)

DECPNT 8888888888808.8888888.8888888.8888888.8888888.8

Decimal point location. Specifies the location of the decimal point or dummy zeroes in theprimary unit display. Value should be consistent with local legal requirements.

DSPDIV 1D2D5D

Display divisions. Selects the minimum division size for the primary units displayed weight.

UNITS LBKGOZTNTGNONE

Specifies primary units for displayed and printed weight. Values are: LB=pound;KG=kilogram; OZ=ounce; TN=short ton; T=metric ton; G=gram.

NOTE: Indicators sold outside North America are configured with KG for both primary andsecondary units.

Secondary Units (SECNDR Parameter)

DECPNT 88888.88888888888808.8888888.8888888.8888888.88

Decimal point location. Determines the location of the decimal point or dummy zeros inthe secondary unit display.

DSPDIV 5D1D2D

Display divisions. Selects the value of minimum division size of the displayed weight.

UNITS KGOZTNTGLBNONE

Specifies secondary units for displayed and printed weight. Values are: KG=kilogram;OZ=ounce; TN=short ton; T=metric ton; G=gram; LB=pound.

MULT 0.453592Enter other choices via keyboard

Multiplier. Specifies the conversion factor by which the primary units are multiplied toobtain the secondary units. The default is 0.453592, which is the conversion factor forchanging pounds to kilograms. NOTE: This parameter does not display the decimal pointlocation for some values less than 1; use the MULEXP parameter to shift the decimalposition of the multiplier. See Section 7.6 on page 40 for a list of multipliers.

To toggle between primary and secondary units, press the UNITS key.

MULEXP decimal position

Multiplier exponent. Sets the decimal position for multiplier values.

FORMAT Menu

Parameter Choices Description

Table 3-3. Format Menu Parameters (Continued)

18 IQ plus 355 Installation Manual

Page 23: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.3 Calibration MenuSee Section 4.0 on page 24 for calibration procedures.

Figure 3-7. Calibration Menu

3.2.4 Serial MenuSee Section 7.3 on page 36 for information about the IQ plus 355 serial data format.

Figure 3-8. Serial Menu

CALIBR Menu

Parameter Choices Description

Level 2 submenus

WZERO — Display and edit the zero calibration A/D count value.

DO NOT adjust this value after WSPAN has been set!

WVAL — Display and edit the test weight value.

WSPAN — Display and edit the span calibration A/D count value.

REZERO — Press Enter to remove an offset value from the zero and span calibrations.

Use this parameter only after WZERO and WSPAN have been set. See Section 4.1 onpage 24 for more information about using this parameter.

Table 3-4. Calibration Menu Parameters

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

WZERO

*CAL*Display and editzero calibrationA/D count value

Display and edittest weight value

WVAL

*CAL*Display and editspan calibrationA/D count value

WSPAN

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

*CAL*Press Enter to

remove offset fromzero and span

calibrations

REZERO

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

EOLDLY

000000

number

EDP PRINT

PRN

EDP

PRNDES

BITS TERMINBAUD

CR

9600

7ODD

7EVEN

8NONE CR/LF

4800

2400

1200

600

300

Same as EDP

EDP

OFF

STREAM

PRN

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

ECHO

ON

OFF

Configuration 19

Page 24: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.5 Program Menu

Figure 3-9. ProgramMenu

SERIAL Menu

Parameter Choices Description

Level 2 submenus

EDP BAUDBITSTERMINEOLDLY

Specifies settings for baud rate, data bits, termination characters, and end-of-line delay used bythe EDP port.

PRINT BAUDBITSTERMINEOLDLY

Specifies settings for baud rate, data bits, termination characters, and end-of-line delay used bythe printer port.

STREAM OFFEDPPRN

Selects the serial port used for continuous transmission. See Section 7.3 on page 36 forinformation about the IQ plus 355 continuous data format.

PRNDES EDPPRN

Print destination. Selects the port for data transmission when the PRINT key is pressed or theKPRINT EDP command is sent.

Level 3 Submenus EDP Port and Printer Port

BAUD 9600300600120024004800

Baud rate. Selects the transmission speed for the EDP or printer port.

BITS 8NONE7EVEN7ODD

Selects number of data bits and parity of data transmitted from the EDP or printer port.

TERMIN CR/LFCR

Termination character. Selects termination character for data sent from the EDP or printer port.

EOLDLY 0number

End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line isterminated to the beginning of the next formatted serial output. Value specified must be in therange 0-255, in tenths of a second (10 = 1 second).

ECHO OFFON

This command enables or disables echoing of the serial commands sent to the indicator.

Table 3-5. Serial Menu Parameters

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

PWRUPM

GO NTEP

OIML

REGULA

000000

number

CONSNU

DELAY

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

000000

number

CONSTU

CANADA

NONE

20 IQ plus 355 Installation Manual

Page 25: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.6 Print Format MenuSee Section 6.0 on page 32 for information about custom print formatting.

Figure 3-10. Print Format Menu

PROGRM Menu

Parameter Choices Description

Level 2 submenus

PWRUPM GODELAY

Power up mode. In GO mode, the indicator goes into operation immediately after a brief powerup display test.

In DELAY mode, the indicator performs a power up display test, then enters a 30-secondwarm up period. If no motion is detected during the warm up period, the indicator becomesoperational when the warm up period ends; if motion is detected, the delay timer is reset andthe warm up period repeated.

REGULA NTEPOIMLCANADANONE

Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than

zero. NONE allows tares to be acquired at any weight value.• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at

no load. NONE allows tares to be cleared at any weight value.• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present.

In CANADA mode, the previous tare must be cleared before a new tare can be acquired.• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net

mode as long as the current weight is within the specified ZRANGE. In OIML mode, thescale must be in gross mode before it can be zeroed; pressing the ZERO key in net modeclears the tare.

CONSNU 000000number

Consecutive numbering. Allows sequential numbering for print operations. The consecutivenumber value is incremented following each print operation.

The initial value of this parameter is set to the start up value specified on the CONSTUparameter. Changing either CONSTU or CONSNU immediately resets the consecutive numberused for printing.

CONSTU 000000number

Consecutive number start up value. Specifies the initial consecutive number (CONSNU) valueused when the indicator is powered on.

Table 3-6. Program Menu Parameters

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

GFMT

Display first 6characters of format

Same as GFMT

NFMT

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

Display and editactive character and

ASCII value

Delete activecharacter

Scroll right in format stringScroll left in format string

Increment ASCII value of active characterDecrement ASCII value of active character

Press to insert a spacebefore the active character

Configuration 21

Page 26: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.7 Digital Input Menu

Figure 3-11. Digital Input Menu

DIG IN Menu

Parameter Choices Description

Level 2 submenus

DIGIN1DIGIN2

OFFZEROTARENT/GRSUNITSDSPTARPRINTCLRCNKBDLOCHOLD

Specifies the function activated by digital inputs 1 and 2.

• ZERO, NT/GRS (net/gross mode toggle), TARE, UNITS, and PRINT provide the samefunctions as the front panel keys.

• DSPTAR displays the current tare value.• CLRCN resets the consecutive number to the value specified on the CONSTU parameter

(PROGRM menu).• KBDLOC disables the keypad while the digital input is held low.• HOLD holds the current display. Releasing this input clears the running average filter.

Table 3-7. Digital Input Menu Parameters

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

DIGIN1

OFF

DIGIN2

ZERO

TARE

NT/GRS

UNITS

DSPTAR

PRINT

CLRCN

OFF

ZERO

TARE

NT/GRS

UNITS

DSPTAR

PRINT

CLRCN

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

HOLD HOLD

KBDLOC KBDLOC

22 IQ plus 355 Installation Manual

Page 27: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

3.2.8 Analog Output MenuThe ALGOUT menu is used only if the analog output option is installed. If the analog output option is installed,configure all other indicator functions and calibrate the indicator (see Section 4.0 on page 24) before configuringthe analog output. See Section 7.8 on page 42 for analog output calibration procedures.

Figure 3-12. Analog Output Menu

3.2.9 Version MenuThe VERS menu is used to check the software version installed in the indicator. There are no parametersassociated with the Version menu: when selected, the indicator displays the installed software version number.

Figure 3-13. Version Menu

ALG OUT Menu

Parameter Choices Description

Level 2 submenus

SOURCE GROSSNET

Specifies the source tracked by the analog output.

OFFSET 0%20%

Zero offset. Selects whether the analog output supplies voltage (0–10 V) or current (4–20 mA)output. Select 0% for 0–10 V output; select 20% for 4–20 mA output.

ERRACT FULLSCHOLDZEROSC

Error action. Specifies how the analog output responds to system error conditions. Possible values are:

FULLSC Set to full value (10 V or 20 mA)HOLD: Hold current valueZEROSC: Set to zero value (0 V or 4 mA)

MIN 000000number

Specifies the minimum weight value tracked by the analog output. Specify a weight value (inprimary units) in the range 0–999 990.

MAX 010000number

Specifies the maximum weight value tracked by the analog output. Specify a weight value (inprimary units) in the range 0–999 990.

TWZERO — Tweak zero. Adjust the analog output zero calibration. Use a multimeter to monitor the analogoutput value. Press and hold or to adjust the output. Press to save the new value.

TWSPAN — Tweak span. Adjust the analog output span calibration. Use a multimeter to monitor the analogoutput value. Press and hold or to adjust the output. Press to save the new value.

Table 3-8. Analog Output Menu Parameters

XXXXXXX XXXXXXXVERSALGOUTDIG INXXXXXXXPROGRM PFORMTSERIALCALIBRXXXXXXXCONFIG FORMAT

SOURCE

GROSS 0%

20%

OFFSET ERRACT

FULLSC

HOLD

000000

number

MIN

10000

number

MAX TWZERO

NET

TWSPAN

ZEROSC

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

Softwareversion

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

Configuration 23

Page 28: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

4.0 CalibrationThe IQ plus 355 can be calibrated using the front panel, EDP commands, or the Revolution™ configurationutility. Each method consists of the following steps:

• Zero calibration• Entering the test weight value• Span calibration• Optional rezero calibration for test weights using hooks or chains.

The following sections describe the calibration procedure for each of the calibration methods.

Figure 4-1. Calibration (CALIBR) Menu

4.1 Front Panel CalibrationTo calibrate the indicator using the front panel, do thefollowing:

1. Place the indicator in setup mode (display readsCONFIG) and remove all weight from the scaleplatform. If your test weights require hooks orchains, place the hooks or chains on the scale forzero calibration.

2. Press until the display reads CALIBR (seeFigure 4-1). Press to go to zero calibration(WZERO).

3. With WZERO displayed, press to calibratezero. The indicator displays *CAL* whilecalibration is in progress. When complete, theA/D count for the zero calibration is displayed.DO NOT adjust this value after WSPAN hasbeen set! Press again to save the zerocalibration value and go to the next prompt(WVAL).

4. With WVAL displayed, place test weights on thescale and press to show the test weight value.Use the procedure shown in Figure 4-2 to enterthe actual test weight, then press to save thevalue and go to span calibration (WSPAN).

5. With WSPAN displayed, press to calibratespan. The indicator displays *CAL* whilecalibration is in progress. When complete, theA/D count for the span calibration is displayed.Press again to save the span calibration valueand go to the next prompt (REZERO).

6. The rezero function is used to remove acalibration offset when hooks or chains are usedto hang the test weights. • If no other apparatus was used to hang the test

weights during calibration, remove the testweights and press to return to the CALIBRmenu.

• If hooks or chains were used duringcalibration, remove these and the test weightsfrom the scale. With all weight removed,press to rezero the scale. This functionadjusts the zero and span calibration values.The indicator displays *CAL* while the zeroand span calibrations are adjusted. Whencomplete, the adjusted A/D count for the zerocalibration is displayed. Press to save thevalue, then press to return to the CALIBRmenu.

7. Press until the display reads CONFIG, thenpress to exit setup mode.

Figure 4-2. Editing Procedure for Numeric Values

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

WZERO

*CAL*Display and editzero calibrationA/D count value

Display and edittest weight value

WVAL

*CAL*Display and editspan calibrationA/D count value

WSPAN

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

*CAL*Press Enter to

remove offset fromzero and span

calibrations

REZERO

When editing numeric values, press or to change thedigit selected. Press or to increment or decrement thevalue of the selected digit.

Press to save the value entered and return to the level above.

0 0 0 0 0 0

24 IQ plus 355 Installation Manual

Page 29: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

4.2 EDP Command CalibrationTo calibrate the indicator using EDP commands, theindicator EDP port must be connected to a terminal orpersonal computer. See Section 2.3.3 on page 6 forEDP port pin assignments; see Section 5.0 on page 27for more information about using EDP commands.

Once the indicator is connected to the sending device,do the following:

1. Place the indicator in setup mode (display readsCONFIG) and remove all weight from the scaleplatform. If your test weights require hooks orchains, place the hooks or chains on the scale forzero calibration.

2. Send the WZERO EDP command to calibratezero. The indicator displays *CAL* whilecalibration is in progress.

3. Place test weights on the scale and use theWVAL command to enter the test weight valuein the following format:

WVAL=nnnnnn<CR>

4. Send the WSPAN EDP command to calibratespan. The indicator displays *CAL* whilecalibration is in progress.

5. To remove an offset value, clear all weight fromthe scale, including hooks or chains used to hangtest weights, then send the REZERO EDPcommand. The indicator displays *CAL* whilethe zero and span calibrations are adjusted.

6. Send the KUPARROW EDP command to exitsetup mode.

4.3 Revolution™ CalibrationTo calibrate the indicator using Revolution, theindicator EDP port must be connected to a PC runningthe Revolution configuration utility.

Use the following procedure to calibrate the indicator:

1. Select Calibrate Indicator from the Revolutionmain menu.

2. On the Indicator Calibration display, select theindicator model (IQ+355) and communicationsport then click OK.

3. Revolution uploads calibration data from theindicator then presents the information in adisplay like that shown in Figure 4-3.

Figure 4-3. Revolution Calibration Display

4. Enter the Value of Test Weight to be used for spancalibration then click OK.

5. The Zero Calibration dialog box prompts you toremove all weight from the scale. Clear the scaleand click OK to begin zero calibration. NOTE: Ifyour test weights require hooks or chains, placethe hooks or chains on the scale for zerocalibration.

6. When zero calibration is complete, the SpanCalibration dialog box prompts you to place testweights on the scale for span calibration. Placetests weights on the scale then click OK.

7. When calibration is complete, the New Settingsfields of the Indicator Calibration display arefilled in. Click Exit to save the new values andreturn to the Revolution main menu; to restorethe previous calibration values, click RestoreSettings.

Calibration 25

Page 30: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

4.4 More About CalibrationThe following topics provide additional informationabout compensating for environmental factors(Section 4.4.1) and diagnostic information fordetermining expected zero and span coefficients.

4.4.1 Adjusting Final CalibrationCalibration may be affected by environmental factorsincluding wind, vibration, and angular loading. Forexample, if the scale is calibrated with 1000 lb, astrain test may determine that at 2000 lb thecalibration is 3 lb high. In this case, final calibrationcan be adjusted by tweaking WVAL to 998.5 lb. Thisadjustment provides a linear correction of 1.5 lb per1000 lb.

To adjust the final calibration, return to the WVALprompt and press to show the test weight value.Press or to adjust calibration up or down. Press to save the value, then press to return to theCALIBR menu.

4.4.2 Zero Deadload A/D CountsTable 4-1 lists the ideal A/D counts that result frominput signals of 0–45 mV with zero deadload. Actualvalues will typically be higher than the values shownin Table 4-1 but the ideal values can be used whencalibrating the indicator with no attached scale.

4.4.3 Calculating the Span CoefficientThe span coefficient value displayed by the WSPANparameter can be calculated using the followingformula:

(21000 * mV_signal_input) + zero_coefficient

where mV_signal_input is the change in the signalinput when the test weight is applied and thezero_coefficient is the WZERO value. Actual valuestypically vary from the calculated value.

Input Signal (mV) Raw A/D Count

0 105 000

1 126 000

2 147 000

3 168 000

4 189 000

5 210 000

6 231 000

7 252 000

8 273 000

9 294 000

10 315 000

15 420 000

20 525 000

30 735 000

45 1 050 000

Table 4-1. Ideal A/D Raw Counts

26 IQ plus 355 Installation Manual

Page 31: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

5.0 EDP CommandsThe IQ plus 355 indicator can be controlled by apersonal computer or remote keyboard connected tothe indicator EDP port. Control is provided by a set ofEDP commands that can simulate front panel keypress functions, display and change setup parameters,and perform reporting functions. The EDP portprovides the capability to print configuration data or tosave that data to an attached personal computer. Thissect ion describes the EDP command set andprocedures for saving and transferring data using theEDP port.

5.1 The EDP Command SetThe EDP command set can be divided into fivegroups: key press commands, reporting commands,the RESETCONFIGURATION special functioncommand, parameter setting commands, and transmitweight data commands.

When the indicator processes an EDP command, itresponds with the message OK. The OK responseverifies that the command was received and has beenexecuted. If the command is unrecognized or cannotbe executed, the indicator responds with ??.

The following sections list the commands andcommand syntax used for each of these groups.

5.1.1 Key Press CommandsKey press EDP commands (see Table 5-1) simulatepressing the keys on the front panel of the indicator.Most commands can be used in both setup andweighing mode. Several of the commands serve as“pseudo” keys, providing functions that are notrepresented by a key on the front panel.

For example, to enter a 15-pound tare weight usingEDP commands:

1. Type K1 and press ENTER (or RETURN).2. Type K5 and press ENTER.3. Type KTARE and press ENTER. The display

shifts to net mode when the tare is entered.

Command Function

KZERO In weighing mode, press the ZERO key

KGROSSNET In weighing mode, press the GROSS/NET key

KGROSS Go to gross mode (pseudo key)

KNET Go to net mode (pseudo key)

KTARE Press the TARE key

KUNITS In weighing mode, press the UNITS key

KPRIM Go to primary units (pseudo key)

KSEC Go to secondary units (pseudo key)

KNEWID Enter new ID (pseudo key)

KPRINT In weighing mode, press the PRINT key

KLEFTARROW In setup mode, move left in the menu

KRIGHTARROW In setup mode, move right in the menu

KUPARROW In setup mode, move up in the menu

KDOWNARROW In setup mode, move down in the menu

K0 Press number 0 (zero)

K1 Press number 1

K2 Press number 2

K3 Press number 3

K4 Press number 4

K5 Press number 5

K6 Press number 6

K7 Press number 7

K8 Press number 8

K9 Press number 9

KDOT Press the decimal point (.)

KENTER Press the ENTER key

Table 5-1. EDP Key Press Commands

EDP Commands 27

Page 32: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

5.1.2 Reporting CommandsReporting commands (see Table 5-2) send specificinformation to the EDP port. These commands can beused in both setup mode and normal mode.

5.1.3 The RESETCONFIGURATION CommandThe RESETCONFIGURATION command can beused to restore all configuration parameters to theirdefault values. Before issuing this command, theindicator must be placed in test mode (press and holdsetup switch for approximately three seconds to showTEST menu).

This command is equivalent to using the DEFLTfunction on the TEST menu. See Section 7.9 onpage 43 for more information about test mode andusing the TEST menu. NOTE: All load cell calibrationsettings are lost when the RESETCONFIGURATIONcommand is run.

5.1.4 Parameter Setting CommandsParameter setting commands allow you to display orchange the current value for a particular configurationparameter (Tables 5-3 through 5-10).

Current configuration parameter settings can bedisplayed in either setup mode or normal mode usingthe following syntax:

command<ENTER>

Most parameter values can be changed in setup modeonly. Use the following command syntax whenchanging parameter values:

command=value<ENTER>

where value is a number or a parameter value. Use nospaces before or after the equal (=) sign. If you type anincorrect command or value, the display reads ??.Changes to the parameters are saved as they areentered but typically do not take effect until you exitsetup mode.

For example, to set the motion band parameter to 5,type the following:

MOTBAND=5D<ENTER>

Command Function

DUMPALL List all parameter values

VERSION Write IQ plus 355 software version

P Write current displayed weight with units identifier. See Section 7.2 on page 35 for more information.

ZZ Write current weight and annunciator status. See Section 7.2 on page 35 for more information.

S Write one frame of stream format

Table 5-2. EDP Reporting Commands

Command Description Values

GRADS Graduations 1–100 000

ZTRKBND Zero track band OFF, 0.5D, 1D, 3D

ZRANGE Zero range 1.9%, 100%

MOTBAND Motion band 1D, 2D, 3D, 5D, 10D, 20D, OFF

OVRLOAD Overload FS+2%, FS+1D, FS+9D, FS

SMPRAT Sample rate 15HZ, 30HZ, 7.5HZ, 3.75HZ

DIGFLTR1DIGFLTR2DIGFLTR3

Digital filtering 1, 2, 4, 8, 16, 32, 64

DFSENS Digital filter cutout sensitivity 2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT

DFTHRH Digital filter cutout threshold NONE, 2DD, 5DD, 10DD, 20DD, 50DD, 100DD, 200DD, 250DD

TAREFN Tare function BOTH, NOTARE, PBTARE, KEYED

Table 5-3. CONFIG EDP Commands

28 IQ plus 355 Installation Manual

Page 33: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Command Description Values

PRI.DECPNT Primary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880

PRI.DSPDIV Primary units display divisions 1D, 2D, 5D

PRI.UNITS Primary units LB, KG, OZ, TN, T, G, NONE

SEC.DECPNT Secondary units decimal position 8.88888, 88.8888, 888.888, 8888.88, 88888.8, 888888, 888880

SEC.DSPDIV Secondary units display divisions 1D, 2D, 5D

SEC.UNITS Secondary units LB, KG, OZ, TN, T, G, NONE

SEC.MULT Secondary units multiplier 0.00000–9999.99

DSPRATE Display rate 250MS, 500MS, 750MS, 1SEC, 1500MS, 2SEC, 2500MS, 3SEC, 4SEC, 6SEC, 8SEC

Table 5-4. FORMAT EDP Commands

Command Description Values

WZERO Zero calibration —

WVAL Test weight value test_weight_value

WSPAN Span calibration —

REZERO Rezero —

LC.CD Set deadload coefficient value

LC.CW Set span coefficient value

Table 5-5. CALIBR EDP Commands

Command Description Values

EDP.BAUD EDP port baud rate 300, 600, 1200, 2400, 4800, 9600

EDP.BITS EDP port data bits/parity 8NONE, 7EVEN, 7ODD

EDP.TERMIN EDP port termination character CR/LF, CR

EDP.EOLDLY EDP port end-of-line delay 0–255 (0.1-second intervals)

PRN.BAUD Printer port baud rate 300, 600, 1200, 2400, 4800, 9600

PRN.BITS Printer port data bits/parity 8NONE, 7EVEN, 7ODD

PRN.TERMIN Printer port termination character CR/LF, CR

PRN.EOLDLY Printer port end-of-line delay 0–255 (0.1-second intervals)

STREAM Streaming port OFF, EDP, PRN

PRNDEST Print destination EDP, PRN

Table 5-6. SERIAL EDP Commands

Command Description Values

PWRUPMD Power up mode GO, DELAY

REGULAT Regulatory compliance NTEP, OIML, CANADA, NONE

CONSNUM Consecutive number 0–999 999

CONSTUP Consecutive number start-up value 0–999 999

Table 5-7. PROGRM EDP Commands

EDP Commands 29

Page 34: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

5.1.5 Normal Mode CommandsThe serial transmit weight data commands (see Table 5-11) transmit data to the EDP port on demand. The SXand EX commands are valid only in normal operating mode; all other commands are valid in either setup ornormal mode.

Command Description Values

GFMT Gross demand print format string See Section 6.0 on page 32 for detailed information

NFMT Net demand print format string

Table 5-8. PFORMT EDP Commands

Command Description Values

DIGIN1DIGIN2

Digital input function OFF, ZERO, TARE, NT/GRS, UNITS, DSPTAR, PRINT, CLRCN, KBDLOC, HOLD

Table 5-9. DIG IN EDP Commands

Command Description Values

SOURCE1 Analog output source GROSS, NET

OFFSET Zero offset 0%, 20%

ERRACT Error action FULLSC, HOLD, ZEROSC

MIN Minimum value tracked 0–999 990

MAX Maximum value tracked 0–999 990

ZERO1 Zero calibration 0–16 383

SPAN1 Span calibration 0–16 383

Table 5-10. ALGOUT EDP Commands

Command Description Response Format

SX Start EDP streaming OK or ??

EX Stop EDP streaming OK or ??

KNEWID Set or query product ID number nnnnnnn

RS Reset system —

XG Transmit gross weight in displayed units nnnnnn UU

where nnnnnn is the weight value, UU is the units.XN Transmit net weight in displayed units

XT Transmit tare weight in displayed units

XG2 Transmit gross weight in non-displayed units

XN2 Transmit net weight in non-displayed units

XT2 Transmit tare weight in non-displayed units

XE Query system error conditions nnnnn nnnnn

See Section 7.1.2 on page 35 for detailed information about the XE command response format.

Table 5-11. Normal Mode EDP Commands

30 IQ plus 355 Installation Manual

Page 35: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

5.2 Saving and Transferring DataConnecting a personal computer to the IQ plus 355EDP port allows you to save indicator configurationdata to the PC or to download configuration data fromthe PC to an indicator. The following sections describethe procedures for these save and transfer operations.

5.2.1 Saving Indicator Data to a Personal ComputerConfiguration data can be saved to a personalcomputer connected to the EDP port. The PC must berunning a communicat ions program such asPROCOMMPLUS®. See Section 2.3.3 on page 6 forinformation about serial communications wiring andEDP port pin assignments.

When configuring the indicator, ensure that the valuesset for the BAUD and BITS parameters on theSERIAL menu match the baud rate, bits, and paritysettings configured for the serial port on the PC. Setthe PRNDES parameter to EDP.

To save all configuration data, place the indicator insetup mode and send the DUMPALL EDP commandto the indicator. The IQ plus 355 responds by sendinga l l configura t i on pa r ame te r s t o t he PC a sASCII-formatted text.

5.2.2 Downloading Configuration Data from PC toIndicator

Configuration data saved on a PC or floppy disk canbe downloaded from the PC to an indicator. Thisprocedure is useful when a number of indicators withsimilar configurations are set up or when an indicatoris replaced.

To download configuration data, connect the PC to theEDP port as described in Section 5.2.1. Place theind i ca to r i n s e tup mode and u se t he PCcommunicat ions sof tware to send the savedconfiguration data to the indicator. When transfer iscomplete, calibrate the indicator as described inSection 4.0 on page 24.

NOTES:

• Calibration settings are included in theconfigurat ion data downloaded to theindicator. If the receiving indicator is a directreplacement for another IQ plus 355 and theattached scale is not changed, recalibration isnot required.

• When downloading configurations thatinclude changed serial communicationssettings, edit the data file to place the serialcommunications changes at the end of the file.Communication between the PC and indicatorwill be lost once the indicator receivessettings for baud rate (BAUD parameter) ordata bits and parity (BITS parameter) that donot match those configured for the PC.

EDP Commands 31

Page 36: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

6.0 Print FormattingThe IQ plus 355 provides two print formats, GFMTand NFMT, that determine the format of the printedoutput when the PRINT key is pressed or when aKPRINT EDP command is received. If a tare has beenentered or acquired, NFMT is used; otherwise, GFMTis used.

Each print format can be customized to include up to300 characters of information, such as company nameand address, on printed tickets. You can use theindicator front panel (PFORMT menu), EDPcommands, or the Revolution™ configuration utility tocustomize the print formats.

6.1 Print Formatting CommandsTable 6-1 lists commands you can use to format thegross and net print formats. Commands included inthe format strings must be enclosed between < and >delimiters. Any characters outside of the delimitersare printed as text on the ticket. Text characters caninclude any ASCII character that can be printed by theoutput device.

The default GFMT and NFMT print formats use onlythe new line (<NL>) command and the commands forgross, net, and tare weights in displayed units (<G>,<N>, and <T>).

The default IQ plus 355 print formats are shown inTable 6-2:

NOTES:

• The <G2>, <N2>, and <T2> commands listedin Table 6-1 print the gross, net, and tareweights in non-displayed units—that is, in theunits not currently displayed on the indicator.

• ID numbers included in the print format string(<ID> command) must be set using theKNEWID EDP command.

• The 300-character limit of each print formatstring includes the output field length of theprint formatting commands, not the commandlength. For example, if the indicator isconfigured to show a decimal point, the <G>command generates an output field of 13characters: the 10-character weight value(including decimal point), one space, and atwo-digit units identifier.

6.2 Customizing Print FormatsThe following sections describe procedures forcustomizing the GFMT and NFMT formats using theEDP port, the front panel (PFORMT menu), and theRevolution configuration utility.

6.2.1 Using the EDP Port With a personal computer, terminal, or remotekeyboard attached to the IQ plus 355 EDP port, youcan use the EDP command set to customize the printformat strings.

To view the current setting of a format string, type thename of the string (GFMT or NFMT) and pressENTER . Fo r example , t o check the cu r ren tconfiguration of the GFMT format, type GFMT andpress ENTER. The indicator responds by sending thecurrent configuration for the gross format:

GFMT=<G> GROSS<NL>

Command Description

<G> Gross weight in displayed units

<G2> Gross weight in non-displayed units

<N> Net weight in displayed units

<N2> Net weight in non-displayed units

<T> Tare weight in displayed units

<T2> Tare weight in non-displayed units

<ID> ID number

<CN> Consecutive number

<NLnn> New line (nn = number of termination (<CR/LF> or <CR>) characters)*

<SPnn> Space (nn = number of spaces)*

<SU> Toggle weight data format (formatted/unformatted)**

Gross, net, and tare weights are 9 digits in length, including sign (10 digits with decimal point), followed by a space and a two-digit units identifier. Total field length with units identifier is 12 (or 13) characters.

ID and consecutive number (CN) fields are 1–6 characters in length, as required.

* If nn is not specified, 1 is assumed. Value must be in the range 1–99.

** After receiving an SU command, the indicator sends unformatted data until the next SU command is received. Unformatted data omits decimal points, leading and trailing characters.

Table 6-1. Print Format Commands

Format Default Format String Sample Output

GFMT <G> GROSS<NL> 2046.81 LB GROSS

NFMT <G> GROSS<NL><T> TARE<NL><N> NET<NL>

4053.1 LB GROSS15.6 LB TARE4037.5 LB NET

NOTE: In OIML and CANADA modes, the letters PT (presettare) are automatically inserted after the printed tare weight.

Table 6-2. GFMT and NFMT Formats

32 IQ plus 355 Installation Manual

Page 37: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

To change the format, use the GFMT or NFMT EDPcommand followed by an equals sign (=) and themodified print format string. For example, to add thename and address of a company to the gross format,you could send the following EDP command:

GFMT=FINE TRANSFER CO<NL>32400 WEST HIGHWAY ROAD<NL>SMALLTOWN<NL2><G> GROSS<NL>

A ticket printed using this format might look like thefollowing:

FINE TRANSFER CO32400 WEST HIGHWAY ROADSMALLTOWN

1345 LB GROSS

6.2.2 Using the Front PanelI f you have no acce s s t o equ ipmen t f o rcommunication through the EDP port or are workingat a site where such equipment cannot be used, youcan use the PFORMT menu (see Figure 6-1) tocustomize the print formats.

Using the PFORMT menu, you can edit the printformat strings by changing the decimal values of theASCII characters in the format string.

NOTE: Lower-case letters and some special characterscannot be displayed on the IQ plus 355 front panel(see the ASCII character chart on page 37) and areshown as blanks. The IQ plus 355 can send or receiveany ASCII character; the character printed depends onthe particular ASCII character set implemented for thereceiving device.

Figure 6-1. PFORMT Menu, Showing Alphanumeric Character Entry Procedure

6.2.3 Using RevolutionThe Revolution configuration utility provides a printformatting grid with a tool bar. The grid allows you toconstruct the print format without the formattingcommands (<NL> and <SP>) required by the frontpanel or EDP command methods. Using Revolution,you can type text directly into the grid, then selectweight value fields from the tool bar and place themwhere you want them to appear on the printed ticket.

Figure 6-2 shows an example of the Revolution printformatting grid.

Figure 6-2. Revolution Print Format Grid

XXXXXXX XXXXXXXXXXXXXXXXXXXXX

GFMT

Display first 6characters of format

Same as GFMT

NFMT

VERSALGOUTDIG INPROGRM PFORMTSERIALCALIBRCONFIG FORMAT

Display and editactive character and

ASCII value

Delete activecharacter

Scroll right in format stringScroll left in format string

Increment ASCII value of active characterDecrement ASCII value of active character

Press to insert a spacebefore the active character

Print Formatting 33

Page 38: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.0 Appendix

7.1 Error MessagesThe IQ plus 355 indicator provides a number of error messages. When an error occurs, the message is shown onthe indicator LED display. Error conditions can also be checked remotely by using the XE EDP command asdescribed in Section 7.1.2.

7.1.1 Displayed Error MessagesThe IQ plus 355 provides a number of front panel error messages to assist in problem diagnosis. Table 7-1 liststhese messages and their meanings.

Error Message Description Solution

E A/D A/D physical error Call Rice Lake Weighing Systems (RLWS) Service.

EEEROM EEPROM physical error

EVIREE Virgin EEPROM Use TEST menu to perform DEFLT (restore defaults) procedure, then recalibrate load cells.EPCKSM Parameter checksum error

EACKSM A/D calibration checksum error A/D converter requires recalibration. Call RLWS Service.

EFCKSM Printer format checksum error Call RLWS Service.

ELCKSM Load cell calibration checksum error Recalibrate load cells.

EIDATA Internal RAM checksum error Call RLWS Service.

E REF A/D reference error A/D converter requires recalibration. Call RLWS Service.

Error Internal program error Check configuration. Run XE command (see Section 7.1.2 on page 35) to determine error type.

Call RLWS Service if unable to clear error by cycling power or if error recurs.

OVERFL Overflow error Weight value too large to be displayed.

- - - - - - Gross > overload limit Gross value exceeds overload limit. Check configuration or signalinput level. Overload can be caused by input signal > 45mV orcommon mode voltage > 950 mV.

______ A/D underrange A/D reading < –4 mV. Check scale for binding or damage.

Table 7-1. IQ plus 355 Error Messages

34 IQ plus 355 Installation Manual

Page 39: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.1.2 Using the XE EDP CommandThe XE EDP command can be used to remotely querythe IQ plus 355 for the error conditions shown on thefront panel. The XE command returns two 5-digitnumbers in the format:

xxxxx yyyyy

where xxxxx contains a decimal representation of anyexisting error conditions as described in Table 7-2.

If more than one error condition exists, the numberreturned is the sum of the values representing the errorconditions. For example, if the XE command returnsthe number 1040, this value represents the sum of anA/D reference error (1024) and an A/D calibrationchecksum error (16).

The second number returned (yyyyy) uses the same bitassignments as shown in Table 7-2 to indicate whetherthe test for the error condition was run. For example,the value yyyyy = 50815 represents the decimalequivalent of the binary value 1100 0110 0111 1111.Using the bit assignments in Table 7-2, this valueindicates all tests were run.

7.2 Status MessagesTwo EDP commands, P and ZZ, can be used toprovide status about the indicator. These commandsare described in the following sections.

7.2.1 Using the P EDP CommandThe P EDP command returns the current displayedweight value to the EDP port, along with the unitsidentifier. If the indicator is in an underrange oroverload condition, the weight value is replaced with&&&&&& (overload) or :::::: (underrange).

7.2.2 Using the ZZ EDP CommandThe ZZ EDP command can be used to remotely querywhich annunciators are currently displayed on theindicator front panel. The ZZ command returns thecurrently displayed weight and a decimal numberrepresenting the LED annunciators currently lit. Theformat of the returned data is:

wwwwww uu zzz

where wwwwww uu is the current displayed weight andunits, zzz is the annunciator status value (seeTable 7-3). If more than one annunciator is lit, thesecond number returned is the sum of the valuesrepresenting the active annunciators.

For example, if the annunciator status value returnedon the ZZ command is 145, the gross, standstill, andlb annunciators are lit: 145 represents the sum of thevalues for the standstill annunciator (128), gross modeannunciator (16), and the lb/primary units annunciator(1).

Error Code Description Binary Value

0 No error 0000 0000 0000 0000

1 EEPROM physical error 0000 0000 0000 0001

2 Virgin EEPROM 0000 0000 0000 0010

4 Parameter checksum error 0000 0000 0000 0100

8 Load cell calibration checksum error

0000 0000 0000 1000

16 A/D calibration checksum error

0000 0000 0001 0000

32 Print format checksum error 0000 0000 0010 0000

64 Internal RAM checksum error 0000 0000 0100 0000

128 not assigned 0000 0000 1000 0000

256 not assigned 0000 0001 0000 0000

512 A/D physical error 0000 0010 0000 0000

1024 A/D reference error 0000 0100 0000 0000

2048 Non-maskable interrupt (NMI) error

0000 1000 0000 0000

4096 Stack overflow error 0001 0000 0000 0000

8192 Zero division error 0010 0000 0000 0000

16384 A/D underrange 0100 0000 0000 0000

32768 Gross > overload limit 1000 0000 0000 0000

Table 7-2. Error Codes Returned on XE Command

Decimal Value Annunciator

1 lb/primary units

2 kg/secondary units

4 oz

8 g

16 Gross

32 Net

64 Center of zero

128 Standstill

Table 7-3. Status Codes Returned on the ZZ Command

Appendix 35

Page 40: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.3 Continuous Output (Stream) FormatFigure 7-1 shows the continuous output format sent to the IQ plus 355 EDP or printer port when the STREAMparameter (SERIAL menu) is set to either EDP or PRN.

Figure 7-1. Continuous Output Data Format

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02(decimal)

Polarity:<Space> = Positive<–> = Negative<^> = Overload<]> = Underrange

Weight data: 7 digits, right-justified, withdecimal point, leading zero suppression.Overload = ^^^^^^^Underrange = ] ] ] ] ] ] ]Display overflow = OVERFL

L = poundsK = kilogramT = tonG = gramsO = ounces<space> = none

G = GrossN = Net

Status:<space> = validI = InvalidM = In motionO = Over/under range

<CR> <LF>or <CR>

ASCII 13, 10(decimal)

36 IQ plus 355 Installation Manual

Page 41: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.4 ASCII Character ChartUse the decimal values for ASCII characters listed inTables 7-4 and 7-5 when specifying print formatstrings on the IQ plus 355 PFORMT menu. The actualcharacter printed depends on the character mappingused by the output device.

The IQ plus 355 can send or receive any ASCIIcharacter value (decimal 0–255), but the indicatordisplay is limited to numbers, upper-case, unaccentedletters, and a few special characters. See Section 7.5on page 39 for information about the IQ plus 355 LEDdisplay.

Table 7-4. ASCII Character Chart (Part 1)

Control ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex

Ctrl-@ NUL 00 00 space 32 20 @ 64 40 ` 96 60

Ctrl-A SOH 01 01 ! 33 21 A 65 41 a 97 61

Ctrl-B STX 02 02 “ 34 22 B 66 42 b 98 62

Ctrl-C ETX 03 03 # 35 23 C 67 43 c 99 63

Ctrl-D EOT 04 04 $ 36 24 D 68 44 d 100 64

Ctrl-E ENQ 05 05 % 37 25 E 69 45 e 101 65

Ctrl-F ACK 06 06 & 38 26 F 70 46 f 102 66

Ctrl-G BEL 07 07 ’ 39 27 G 71 47 g 103 67

Ctrl-H BS 08 08 ( 40 28 H 72 48 h 104 68

Ctrl-I HT 09 09 ) 41 29 I 73 49 i 105 69

Ctrl-J LF 10 0A * 42 2A J 74 4A j 106 6A

Ctrl-K VT 11 0B + 43 2B K 75 4B k 107 6B

Ctrl-L FF 12 0C , 44 2C L 76 4C l 108 6C

Ctrl-M CR 13 0D - 45 2D M 77 4D m 109 6D

Ctrl-N SO 14 0E . 46 2E N 78 4E n 110 6E

Ctrl-O SI 15 0F / 47 2F O 79 4F o 111 6F

Ctrl-P DLE 16 10 0 48 30 P 80 50 p 112 70

Ctrl-Q DC1 17 11 1 49 31 Q 81 51 q 113 71

Ctrl-R DC2 18 12 2 50 32 R 82 52 r 114 72

Ctrl-S DC3 19 13 3 51 33 S 83 53 s 115 73

Ctrl-T DC4 20 14 4 52 34 T 84 54 t 116 74

Ctrl-U NAK 21 15 5 53 35 U 85 55 u 117 75

Ctrl-V SYN 22 16 6 54 36 V 86 56 v 118 76

Ctrl-W ETB 23 17 7 55 37 W 87 57 w 119 77

Ctrl-X CAN 24 18 8 56 38 X 88 58 x 120 78

Ctrl-Y EM 25 19 9 57 39 Y 89 59 y 121 79

Ctrl-Z SUB 26 1A : 58 3A Z 90 5A z 122 7A

Ctrl-[ ESC 27 1B ; 59 3B [ 91 5B 123 7B

Ctrl-\ FS 28 1C < 60 3C \ 92 5C | 124 7C

Ctrl-] GS 29 1D = 61 3D ] 93 5D 125 7D

Ctrl-^ RS 30 1E > 62 3E ^ 94 5E ~ 126 7E

Ctrl-_ US 31 1F ? 63 3F _ 95 5F DEL 127 7F

Appendix 37

Page 42: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

Table 7-5. ASCII Character Chart (Part 2)

ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex ASCII Dec Hex

Ç 128 80 á 160 A0 192 C0 α 224 E0

ü 129 81 í 161 A1 193 C1 β 225 E1

é 130 82 ó 162 A2 194 C2 Γ 226 E2

â 131 83 ú 163 A3 195 C3 π 227 E3

ä 132 84 ñ 164 A4 196 C4 Σ 228 E4

à 133 85 Ñ 165 A5 197 C5 σ 229 E5

å 134 86 ª 166 A6 198 C6 µ 230 E6

ç 135 87 º 167 A7 199 C7 τ 231 E7

ê 136 88 ¿ 168 A8 200 C8 Φ 232 E8

ë 137 89 169 A9 201 C9 Θ 233 E9

è 138 8A ¬ 170 AA 202 CA Ω 234 EA

ï 139 8B 1/2 171 AB 203 CB δ 235 EB

î 140 8C 1/4 172 AC 204 CC ∞ 236 EC

ì 141 8D ¡ 173 AD 205 CD φ 237 ED

Ä 142 8E « 174 AE 206 CE ∈ 238 EE

Å 143 8F » 175 AF 207 CF ∩ 239 EF

É 144 90 176 B0 208 D0 ≡ 240 F0

æ 145 91 177 B1 209 D1 ± 241 F1

Æ 146 92 178 B2 210 D2 ≥ 242 F2

ô 147 93 179 B3 211 D3 ≤ 243 F3

ö 148 94 180 B4 212 D4 ⌠ 244 F4

ò 149 95 181 B5 213 D5 ⌡ 245 F5

û 150 96 182 B6 214 D6 ÷ 246 F6

ù 151 97 183 B7 215 D7 ≈ 247 F7

ÿ 152 98 184 B8 216 D8 ° 248 F8

Ö 153 99 185 B9 217 D9 249 F9

Ü 154 9A 186 BA 218 DA 250 FA

¢ 155 9B 187 BB 219 DB 251 FB

£ 156 9C 188 BC 220 DC 252 FC

¥ 157 9D 189 BD 221 DD 2 253 FD

Pts 158 9E 190 BE 222 DE 254 FE

ƒ 159 9F 191 BF 223 DF 255 FF

38 IQ plus 355 Installation Manual

Page 43: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.5 Front Panel Display CharactersFigure 7-2 shows the 7-segment LED character set used to display alphanumeric characters on the IQ plus 355front panel.

Figure 7-2. IQ plus 355 Display Characters

%

&

(

)

*

+

,

-

. ./

0

1

2

3

4

5

6

7

8

9

:

;

<

=

>

?

@

A

B

C

D

E

F

G

H

I

J

K

L

M

N

O

P

Q

R

S

T

U

V

W

X

Y

Z

[

\

Appendix 39

Page 44: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.6 Conversion Factors for SecondaryUnits

The IQ plus 355 has the capability to mathematicallyconvert a weight into many different types of units andinstantly display those results with a press of theUNITS key.

Secondary units can be specified on the FORMATmenu using the SECNDR parameter, or by using EDPcommands.

• To configure secondary units using the frontpanel menus, use the Table 7-6 to find theconvers ion mul t ip l i e r fo r the MULTparameter. For example, if the primary unit ispounds and the secondary unit is short tons,set the MULT parameter to 0.000500. You must use the MULEXP parameter to setthe decimal point position. For example, if theprimary unit is ounces and the secondary unitis grams, the conversion factor shown inTable 7-6 is 28.3495. To enter this value usingthe menus, first enter 283495 for the MULTparameter, then use the MULEXP parameterto adjust the decimal point to 28.3495.

• To configure secondary units using EDPcommands, use the Table 7-6 to find theconversion multiplier for the SEC.MULTcommand. For example, if the primary unit ispounds and the secondary unit is short tons,send the EDP command SEC.MULT=0.0005<CR> to set the multiplier for thesecondary units.

• Long tons and grains units listed in Table 7-6cannot be directly specified as primary orsecondary units on the IQ plus 355 indicator.For these or other unlisted units of weight,specify NONE on the UNITS parameter.

NOTE: Ensure that the secondary decimal pointposition is set appropriately for the scale capacity inthe secondary units. If the converted value requiresmore digits than are available, the indicator willdisplay an overflow message (OVERFL).

For example, if the primary units are short tons,secondary units are pounds, and the secondarydecimal point is set to 8888.88, the indicator willoverflow if 5 tons or more are applied to the scale.With 5 tons applied, and a conversion factor of 2000,the secondary units display needs five digits to the leftof the decimal point to display the 10000 lb secondaryunits value.

Primary Unit x Multiplier Secondary Unit

grains 0.064799 grams

0.002286 ounces

0.000143 pounds

0.000065 kilograms

ounces 437.500 grains

28.3495 grams

0.06250 pounds

0.02835 kilograms

pounds 7000.00 grains

453.592 grams

16.0000 ounces

0.453592 kilograms

0.000500 short tons

0.000446 long tons

0.000453 metric tons

short tons 2000.00 pounds

907.185 kilograms

0.892857 long tons

0.907185 metric tons

grams 15.4324 grains

0.035274 ounces

0.002205 pounds

0.001000 kilograms

kilograms 15432.4 grains

35.2740 ounces

1000.00 grams

2.20462 pounds

0.001102 short tons

0.000984 long tons

0.001000 metric tons

metric tons 2204.62 pounds

1000.00 kilograms

1.10231 short tons

0.984207 long tons

long tons 2240.00 pounds

1016.05 kilograms

1.12000 short tons

1.01605 metric tons

Table 7-6. Conversion Factors

40 IQ plus 355 Installation Manual

Page 45: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.7 Digital FilteringThe IQ plus 355 uses averaged digital filtering to reduce the effect of vibration on weight readings. Adjustablethreshold and sensitivity functions allow quick settling by suspending filter averaging, allowing the weightreading to jump to the new value. Figure 7-3 shows the digital filter parameters on the CONFIG menu.

Figure 7-3. Digital Filtering Parameters on the Configuration (CONFIG) Menu

Figure 7-4. Flow Diagram for IQ plus 355 Digital Filters

7.7.1 DIGFLx ParametersThe first three digital filtering parameters, DIGFL1,DIGFL2, and DIGFL3, are configurable filter stagesthat control the effect of a single A/D reading on thedisplayed weight. The value assigned to eachparameter sets the number of readings received fromthe preceding filter stage before averaging (seeFigure 7-4).

The overall filtering effect can be expressed by addingthe values assigned to the three filter stages:

DIGFL1 + DIGFL2 + DIGFL3

For example , i f the fil ters are configured asDIGFL1=4, DIGFL2=8, DIGFL3=8, the overallfi l t e r ing e ffec t i s 20 (4 + 8 + 8 ) . Wi th th i sconfiguration, each A/D reading has a 1-in-20 effecton the displayed weight value. Setting the filters to 1effectively disables digital filtering.

7.7.2 DFSENS and DFTHRH ParametersThe three digital filters can be used by themselves toeliminate vibration effects, but heavy filtering alsoincreases settling time. The DFSENS (digital filtersensitivity) and DFTHRH (digital filter threshold)parameters can be used to temporarily override filteraveraging and improve settling time:

• DFSENS specifies the number of consecutivescale readings that must fall outside the filterthreshold (DFTHRH) before digital filteringis suspended.

• DFTHRH sets a threshold value, in displaydivisions. When a specified number ofconsecutive scale readings (DFSENS) falloutside of this threshold, digital filtering issuspended. Set DFTHRH to NONE to turn offthe filter override.

1

4

2

8

DIGFL1

32

16

64

1

4

2

8

DIGFL2

32

16

64

1

4

2

8

DIGFL3

32

16

64

8OUT

32OUT

16OUT

64OUT

DFSENS DFTHRH

2OUT

128OUT

4OUT

2DD

NONE

5DD

20DD

10DD

50DD

200DD

100DD

250DD

DIGFL1=n1 DIGFL2=n2 DIGFL3=n3

3rd StageFilter Average

2nd StageFilter Average

1st StageFilter Average

DisplayedValue

n32nd Stage

Filter Averages

n21st Stage

Filter Averages

n1A/D Readings

Appendix 41

Page 46: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.7.3 Setting the Digital Filter ParametersFine-tuning the digital filter parameters greatlyimproves indicator performance in heavy-vibrationenvironments. Use the following procedure todetermine vibration effects on the scale and optimizethe digital filtering configuration.

1. In setup mode, set all three digital filters(DIGFL1, DIGFL2, DIGFL3) to 1. SetDFTHRH to NONE. Return indicator tonormal mode.

2. Remove all weight from the scale, then watchthe indicator display to determine themagnitude of vibration effects on the scale.Record the weight below which all but a fewreadings fall. This value is used to calculatethe DFTHRH parameter value in Step 4.For example, if a heavy-capacity scaleproduces vibration-related readings of up to50 lb, with occasional spikes to 75 lb, record50 lb as the threshold weight value.

3. Place the indicator in setup mode and set thedigital filters (DIGFLx) to eliminate the

vibrat ion effects on the scale. (LeaveDFTHRH set to NONE.) Reconfigure asnecessary to find the lowest effective valuesfor the DIGFLx parameters.

4. With optimum values assigned to the DIGFLxparameters, calculate the DFTHRH parametervalue by converting the weight value recordedin Step 2 to display divisions:

threshold_weight_value / DSPDIV

In the example in Step 2, with a thresholdweight value of 50 lb and a display divisionvalue of 5lb: 50 / 5lb = 10DD. DFTHRHshould be set to 10DD for this example.

5. Finally, set the DFSENS parameter highenough to ignore transient peaks. Longertransients (typically caused by lower vibrationfrequencies) will cause more consecutiveout-of-band readings, so DFSENS should beset higher to counter low frequency transients.Reconfigure as necessary to find the lowesteffective value for the DFSENS parameter.

7.8 Analog Output CalibrationThe following calibration procedure requires amultimeter to measure voltage or current output fromthe analog output module. If the option is not alreadyinstalled, see Section 2.4 on page 7.

NOTE: The analog output must be calibrated after theindicator itself has been configured (Section 3.0) andcalibrated (Section 4.0).

Figure 7-5. Analog Output Menu

1. Enter setup mode and go to the ALGOUTmenu (see Figure 7-5).• Set OFFSET to 0% for 0–10 V output,

20% for 4–20 mA output• Set MIN to lowest weight value to be

tracked by the analog output• Set MAX to highest weight value to be

tracked by the analog output2. Connect multimeter to analog output:

• For voltage output, connect voltmeterleads to pins 3 and 4

• For current output, connect ammeterleads to pins 1 and 2

3. Adjust zero calibration: Scroll to theTWZERO parameter. Check voltage orcurrent reading on multimeter. Press and hold

or to adjust the zero value up or down.4. Adjust span calibration: Scroll to the

TWSPAN parameter. Check voltage orcurrent reading on multimeter. Press and hold

or to adjust the span value up or down.5. Final zero calibration: Return to the

TWZERO parameter and verify that the zerocalibration has not drifted. Press and hold or to re-adjust the zero value as required.

6. Return to normal mode. Analog outputfunction can be verified using test weights.

XXXXXXX XXXXXXXVERSALGOUTDIG INXXXXXXXPROGRM PFORMTSERIALCALIBRXXXXXXXCONFIG FORMAT

SOURCE

GROSS 0%

20%

OFFSET ERRACT

FULLSC

HOLD

000000

number

MIN

10000

number

MAX TWZERO

NET

TWSPAN

ZEROSC

42 IQ plus 355 Installation Manual

Page 47: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.9 Test ModeIn addition to normal and setup modes, test modeprovides a number of diagnostic functions for the IQplus 355, including:

• Display raw A/D count• Display digital input states• Reset configuration parameters to default

values• Transmit test character (“U”) from serial port• Display characters received by serial port• Set analog output state to zero or full scale

To enter test mode, press and hold the setup switchuntil the front panel display shows the word TEST.After about three seconds, the test mode displayautomatically shifts to the first test menu function,A/DTST.

Figure 7-6 shows the Test Menu structure; Figure 7-7shows the front panel key functions in test mode. Notethat, because the Test Menu functions are all on asingle menu level, the GROSS/NET ( ) key has nofunction. Press the ZERO ( ) key to exit test mode.

Table 7-7 on page 44 summarizes the test menufunctions.

Figure 7-6. Test Menu

Figure 7-7. Front Panel Key Functions in Test Mode

A/DTSTTEST D/I 1 D/I 2 AOUT=0 AOUTFS

DEFLT XMT U ECHO R

ZERO GROSSNET

TARE UNITS PRINT

B/N Units

Move LEFT

Move RIGHTEXIT

Not used

ENTER

TEST MODE KEY FUNCTIONS

Appendix 43

Page 48: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.10 Software Revision HistoryFor a complete description of recent software revisions, visit our web site at www.rlws.com.

The following list summarizes principal revision software changes:

Version 1.10• Entering setup mode now clears any tare value in the system.• The configured end-of-line delay (EOLDLY) now functions in both demand and continuous mode.• A/D gain (ADGAIN) and A/D offset (ADOFFS) functions removed from TEST menu.

Version 1.12• New SMPRAT parameter on the CONFIG menu provides selectable A/D conversion rate ranging from

3.75–30 Hz.• Default value for REGULAT parameter (on PROGRM menu) changed to NTEP.

Version 1.15• This parameter added in SERIAL menu under EDP port (EDP.ECHO=On/Off) to enable or disable

echoing of serial commands sent to indicator.

Version 1.16• This version corrects an issue introduced in Version 1.15 where the primary unit with a fixed zero was

divided by 10 when displaying secondary units.

TEST Menu

Function Description

A/DTST Display A/D testPress and hold Enter key to display raw count from A/D converter.

D/I 1 Display digital input 1

Press and hold Enter key to show status of DIGIN1 (DI1=HI or DI1=LO).

D/I 2 Display digital input 2

Press and hold Enter key to show status of DIGIN2 (DI2=HI or DI2=LO).

AOUT=0 Set analog output to zero

Press and hold Enter key to set analog output to its zero value.

AOUTFS Set analog output to full scale

Press and hold Enter key to set analog output to its full scale value.

DEFLT Default parameters

Press setup switch and Enter key at the same time to reset configuration and calibration parameters to factorydefault values. Load cells must be recalibrated before using the indicator (see Section 4.0 on page 24).

XMT U Transmit “U”

Press and hold Enter key to send ASCII “U” characters (decimal 85) from the serial port.

ECHO R Echo received characters

Press and hold Enter key to view characters received at serial port.

NOTE: IQ plus 355 display shows lower-case characters as blanks.

Table 7-7. Test Menu Functions

44 IQ plus 355 Installation Manual

Page 49: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

7.11 SpecificationsPowerLine Voltages 115 or 230 VACFrequency 50 or 60 HzPower Consumption 70 mA @ 115 VAC (8W)

35 mA @ 230 VAC (8W)Fusing

115 VAC 2 x 200 mA TR5 subminiature fusesWickmann Time-Lag 19374 SeriesUL Listed, CSA Certified

230 VAC 2 x 100 mA TR5 subminiature fusesWickmann Time-Lag 19372 SeriesUL Recognized, VDE Approved

Analog SpecificationsFull Scale Input Signal Up to 45 mVExcitation Voltage 10 ± 0.5 VDC,

8 x 350Ω or 16 x 700Ω load cellsSense Amplifier Differential amplifier with

4- and 6-wire sensingAnalog Signal

Input Range 0.6 mV/V – 4.5 mV/VAnalog Signal

Sensitivity 0.3 µV/graduation minimum, 1.5 µV/grad recommended

Input Impedance 200 MΩ, typicalNoise (ref to input) 0.3 µV p-p with digital filters at 4-4-4Internal Resolution 308 000 countsDisplay Resolution 100 000 ddMeasurement Rate Up to 30 measurements/secInput Sensitivity 150 nV per internal countSystem Linearity Within 0.01% of full scaleZero Stability 150 nV/°C, maximumSpan Stability 3.5 ppm/°C, maximumCalibration Method Software, constants stored in

EEPROMCommon Mode

Voltage –0.15 to +0.95 V, referred to earthCommon Mode

Rejection 130 dB minimum @ 50 or 60 HzNormal Mode

Rejection 90 dB minimum @ 50 or 60 HzInput Overload ± 12 V continuous, static discharge

protectedRFI Protection Signal, excitation, and sense lines

protected by capacitor bypassAnalog Output Optional: fully isolated, voltage or current

output,14-bit resolution. Voltage output: 0 –10 VDC Load resistance:1KΩ minimumCurrent output: 4–20 mAExternal loop resistance: 500Ω maximum

Digital SpecificationsMicrocomputer Phillips 80C51XA-G49 main processor @

9.8304 MHzDigital Inputs 2 inputs, TTL or switch closure,

active-lowDigital Filters 3 filters, software selectable

Serial CommunicationsEDP Port Full duplex RS-232Printer Port Output-only RS-232 or active 20 mA

current loop Both Ports 9600, 4800, 2400, 1200, 600, 300 bps;

7 or 8 data bits; even, odd, or no parity

Operator InterfaceDisplay 6-digit LED display. 7-segment, 0.8 in

(20 mm) digitsLED annunciators Gross, net, center of zero, standstill,

lb/primary units, kg/secondary units, oz, g

Keyboard 5-key flat membrane panel

EnvironmentalOperating Temperature–10 to +40°C (legal);

–10 to +50°C (industrial)Storage Temperature –25 to +70°CHumidity 0–95% relative humidityAltitude 2000 m (6500 ft) maximum

EnclosureEnclosure Dimensions 9.5 in x 6 in x 2.75 in

24 cm x 15 cm x 7 cmWeight 2.8 Kg (6.1 lb)Rating/Material UL Type 4X

Certifications and Approvals

NTEPCoC Number 97-130A1Accuracy Class III/III L

nmax

: 10 000

Measurement CanadaApproval AM-5213Accuracy Class III/III HD

nmax

: 10 000

OIMLR76-2 Test Certificate TC6008Accuracy Class III n

max : 3 000

•N

ATIO

NAL CONFERENC

E•

ON

WE

IGH T S A N D M E A

SU

RE

S

Appendix 45

Page 50: Installation Manual - Cisco- · PDF file3 1.4.4 Acquire Tare ... IQ plus 355 Installation Manual

IQ plus 355 Limited WarrantyRice Lake Weighing Systems (RLWS) warrants that all RLWS equipment and systems properly installed by aDistributor or Original Equipment Manufacturer (OEM) will operate per written specifications as confirmed bythe Distributor/OEM and accepted by RLWS. All systems and components are warranted against defects inmaterials and workmanship for two years.

RLWS warrants that the equipment sold hereunder will conform to the current written specifications authorizedby RLWS. RLWS warrants the equipment against faulty workmanship and defective materials. If any equipmentfails to conform to these warranties, RLWS will, at its option, repair or replace such goods returned within thewarranty period subject to the following conditions:

• Upon discovery by Buyer of such nonconformity, RLWS will be given prompt written notice with adetailed explanation of the alleged deficiencies.

• Individual electronic components returned to RLWS for warranty purposes must be packaged toprevent electrostatic discharge (ESD) damage in shipment. Packaging requirements are listed in apublication, Protecting Your Components From Static Damage in Shipment, available from RLWSEquipment Return Department.

• Examination of such equipment by RLWS confirms that the nonconformity actually exists, and wasnot caused by accident, misuse, neglect, alteration, improper installation, improper repair orimproper testing; RLWS shall be the sole judge of all alleged non-conformities.

• Such equipment has not been modified, altered, or changed by any person other than RLWS or itsduly authorized repair agents.

• RLWS will have a reasonable time to repair or replace the defective equipment. Buyer is responsiblefor shipping charges both ways.

• In no event will RLWS be responsible for travel time or on-location repairs, including assembly ordisassembly of equipment, nor will RLWS be liable for the cost of any repairs made by others.

THESE WARRANTIES EXCLUDE ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUTLIMITATION WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. NEITHERRLWS NOR DISTRIBUTOR WILL, IN ANY EVENT, BE LIABLE FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES.

RLWS AND BUYER AGREE THAT RLWS’S SOLE AND EXCLUSIVE LIABILITY HEREUNDER IS LIMITED TOREPAIR OR REPLACEMENT OF SUCH GOODS. IN ACCEPTING THIS WARRANTY, THE BUYER WAIVES ANY ANDALL OTHER CLAIMS TO WARRANTY.

SHOULD THE SELLER BE OTHER THAN RLWS, THE BUYER AGREES TO LOOK ONLY TO THE SELLER FORWARRANTY CLAIMS.

NO TERMS, CONDITIONS, UNDERSTANDING, OR AGREEMENTS PURPORTING TO MODIFY THE TERMS OF THISWARRANTY SHALL HAVE ANY LEGAL EFFECT UNLESS MADE IN WRITING AND SIGNED BY A CORPORATEOFFICER OF RLWS AND THE BUYER.

© 2002 Rice Lake Weighing Systems, Inc. Rice Lake, WI USA. All Rights Reserved.

RICE LAKE WEIGHING SYSTEMS • 230 WEST COLEMAN STREET • RICE LAKE, WISCONSIN 54868 • USA

46 IQ plus 355 Installation Manual