Ni Sepharose excel (25, 100 and500 ml)HisTrap excel (1 and 5 ml)

Instructions for Use

cytiva.com 29013867 AF

Page 2: Instructions for Use - Cytiva

Table of Contents

1 Abstract ..................................................................................................... 3

2 Product description .............................................................................. 4

3 Buffer preparation ................................................................................. 8

4 Sample preparation .............................................................................. 10

5 Ni Sepharose excel protocols ........................................................... 11

6 HisTrap excel protocol ........................................................................ 15

7 Optimization ............................................................................................ 16

8 Cleaning-in-place .................................................................................. 17

9 Storage ...................................................................................................... 18

10 Troubleshooting ..................................................................................... 18

11 Ordering information ........................................................................... 22

Read these instructions carefully before using the products.

Intended use

The products are intended for research use only, and shall notbe used in any clinical or in vitro procedures for diagnosticpurposes.

Safety

For use and handling of the products in a safe way, refer to theSafety Data Sheets.

2 29013867 AF

Page 3: Instructions for Use - Cytiva

1 AbstractNi Sepharose™ excel is an immobilized metal ion affinitychromatography (IMAC) medium (resin) precharged withnickel ions. The medium is designed mainly for capture andpurification of histidine-tagged proteins secreted intoeukaryotic cell culture supernatants. The very strong bindingof nickel ions to Ni Sepharose excel enables direct loading oflarge sample volumes without having to remove agents thatcause stripping of nickel ions from conventional IMAC media.The strong nickel ion binding also provides very highresistance to EDTA and reducing agents like DTT.NiSepharose excel is also suitable for purification of histidine-tagged proteins from other samples, including E. coli lysates.

The medium is available in the following formats:

• 25, 100 and 500 ml medium packs for packing inchromatography columns, such as Tricorn™, XK orHiScale™ columns.

• HisTrap™ excel , 1 ml and 5 ml columns prepacked with NiSepharose excel .

The special type of filters in the top and bottom of HisTrapexcel columns makes it possible to directly load cell-free,unclarified samples on the columns without causing backpressure problems.

The HisTrap excel columns can be operated with a peristalticpump or liquid chromatography systems such as ÄKTA™.

29013867 AF 3

Page 4: Instructions for Use - Cytiva

2 Product descriptionBackground

Traditionally, IMAC purification of histidine-tagged proteinssecreted into eukaryotic cell culture supernatants is oftenproblematic due to incompatibility with the cell culture media,leading to stripping of the immobilized metal ions duringsample loading. This results in low or no binding of the targetprotein.

The purification is further complicated by the fact that thetarget protein concentration often is low. This requires largesample volumes to be used, which in turn may lead toincreased metal ion stripping.

To overcome these problems, considerable samplepretreatment has been required, such as buffer exchange byultrafiltration in combination with concentration procedures.Such pretreatment is time consuming and potentially harmfulto sensitive proteins.

General description

Ni Sepharose excel and HisTrap excel (columns prepackedwith Ni Sepharose excel) are products for IMAC proteinpurification. Ni Sepharose excel consists of 90 μm highlycross-linked agarose beads, to which a chelating ligand hasbeen coupled. The ligand is precharged with exceptionallystrongly bound nickel ions. The nickel ions remain bound evenafter 24 hours incubation in 10 mM EDTA. No nickel ion re-charging is required after use.

4 29013867 AF

Page 5: Instructions for Use - Cytiva

Samples usually causing stripping of metal ions can be loadedonto the medium without extensive and time-consumingpretreatment procedures. Examples of samples that oftencause stripping problems are cell culture supernatantscontaining secreted histidine-tagged proteins fromeukaryotic cells, such as insect cells or CHO cells. NiSepharose excel and HisTrap excel are also suitable forpurification of histidine-tagged proteins from other samplescausing extensive metal ion stripping from conventional IMACmedia.

For samples without extensive stripping effects, it isrecommended to use Ni Sepharose 6 Fast Flow or NiSepharose High Performance, which normally show higheraffinity for histidine-tagged proteins.

Ni Sepharose excel

Characteristics

The flow properties of Ni Sepharose excel make it excellent forpurifications in all scales and allow loading of large samplevolumes. Due to low target protein concentrations, severalliters often need to be processed when purifying secretedproteins.

Tricorn, XK and HiScale columns are recommended forpacking of the medium.

The key characteristics of Ni Sepharose excel are listed in thetable below.

Table 1. Ni Sepharose excel characteristics

Matrix Highly cross-linked spherical agarose, 6%

Average particle size 90 μm

Metal ion capacity 54 to 70 μmol Ni2+/ml medium

29013867 AF 5

Page 6: Instructions for Use - Cytiva

Dynamic binding capacity 1 , 2 At least 10 mg (histidine)6-tagged protein/mlmedium

Recommended flow

velocity 3 , 4

150 to 600 cm/h

Maximum flow velocity 3 600 cm/h

pH stability 5

Working range

Cleaning-in-place

2 to 12

2 to 14

Compatibility during use Stable in all buffers commonly used in IMAC

Chemical stability6

- 0.01 M HCl and 0.01 M NaOH

- 10 mM EDTA, 5 mM DTT, 5 mM TCEP,

20 mM β-mercaptoethanol, 1 M NaOH,

and 6 M guanidine-HCl

- 500 mM imidazole and 100 mM EDTA

- 30% 2-propanol

Tested for one week

Tested for 24 hours

Tested for 2 hours

Tested for 20 minutes

Storage 20% ethanol at 4°C to 30°C1 Binding capacity is sample and protein dependent. The binding capacity for most

proteins is considerably higher than 10 mg/ml.2 Dynamic binding capacity was tested with 0.5 mg/ml pure (histidine)6-tagged

protein (Mr 43 000) or (histidine)6-tagged protein (Mr 28 000) spiked in EX-CELL™ 420Insect serum-free medium (capacity at 10% breakthrough). Column volume was 1ml and flow rate 1 ml/min.

3 H2O at room temperature. For viscous buffers and samples the flow velocity must beoptimized. Starting with a low flow velocity is recommended.

4 Optimal flow velocity during binding is sample-dependent. During column wash andelution a flow velocity of 150 cm/h is recommended. Do not exceed 150 cm/h duringcleaning-in-place (CIP).

5 Working range: pH interval where the medium can be handled without significantchange in function.

Cleaning-in-place: pH interval where the medium can be subjected to cleaning-in-place without significant change in function.

6 Chemical stability was tested by incubating the medium in the listed solutions atroom temperature, and thereafter measuring either the nickel leakage or theprotein binding capacity.

6 29013867 AF

Page 7: Instructions for Use - Cytiva

HisTrap excel

Characteristics

HisTrap excel, 1 ml and 5 ml , are ready-to-use IMAC columnsprepacked with Ni Sepharose excel . The design of thecolumns in combination with the specific properties of themedium enables fast and convenient purifications.

The special type of filters in the top and bottom of the columnsmakes it possible to directly load cell-free, unclarified sampleson the columns without causing back pressure problems. Thisis time-saving, which helps to prevent degradation and thusloss of sensitive target proteins.

HisTrap excel columns are made of biocompatiblepolypropylene that does not interact with biomolecules. Thecolumns are delivered with a stopper on the inlet and a snap-off end on the outlet. HisTrap excel columns cannot beopened or refilled.

The columns can be operated with either a peristaltic pump orchromatography systems such as ÄKTA systems. For easyscaling-up, two HisTrap excel columns can simply beconnected in series.

The key characteristics of the HisTrap excel columns arelisted in the following table. For characteristics of the NiSepharose excel medium used in the columns, see Table 1, onpage 5.

Table 2. HisTrap excel characteristics

Column volume (CV) 1 ml 5 ml

Column dimensions 0.7 × 2.5 cm 1.6 x 2.5 cm

Recommended flow rate 1 , 2 1 to 4 ml/min 5 to 20 ml/min

Maximum flow rate 1 4 ml/min 20 ml/min

29013867 AF 7

Page 8: Instructions for Use - Cytiva

Column hardware pressurelimit

5 bar (0.5 MPa) 5 bar (0.5 MPa)

1 H2O at room temperature. Maximum flow rate will be lower when using buffers orsamples with high viscosity or performing purification at low temperature.

2 Optimal flow rate during binding is sample-dependent. During column wash andelution, a flow rate of 1 ml/min and 5 ml/min is recommended for 1 ml and 5 mlcolumns, respectively. During CIP, do not exceed a flow rate of 1 ml/min and 5 ml/minfor 1 ml and 5 ml columns, respectively.

Note: The maximum pressure the packed bed can withstanddepends on the chromatography mediumcharacteristics and sample/liquid viscosity. The valuemeasured on the chromatography system used alsodepends on the tubing used to connect the column.

Supplied Connector kit with HisTrap excel columns

Connectors supplied Usage No. supplied

Union 1/16” male/luerfemale

For connection of syringe to HiTrapcolumn

1

Stop plug female, 1/16” For sealing bottom of HiTrap column 2, 5 or 7

3 Buffer preparationWater and chemicals used for buffer preparation should be ofhigh purity. Filter buffers through a 0.45 μm filter before use.Use high purity imidazole, which gives essentially noabsorbance at 280 nm.

Recommended buffers

Equilibration buffer 20 mM sodium phosphate, 0.5 M NaCl,pH 7.4

Wash buffer 20 mM sodium phosphate, 0.5 M NaCl, 0 to 30mM imidazole, pH 7.4

8 29013867 AF

Page 9: Instructions for Use - Cytiva

Elution buffer 20 mM sodium phosphate, 0.5 M NaCl, 500mM imidazole, pH 7.4

Note: It is not recommended to include imidazole in sampleand equilibration buffers.

Note: Optimal imidazole concentration during wash issample-dependent. See below for further information.

Buffer optimization

Imidazole concentration in wash buffers

To minimize host cell proteins in the eluate, it is recommendedto include imidazole at low concentrations in the wash buffer.However, for some target proteins, even a small increase ofthe imidazole concentration in the wash buffer may lead topartial elution.

The imidazole concentration must therefore be optimized toensure the best balance of high purity (low binding of host cellproteins) and high yield (strong binding of histidine-taggedtarget protein). For high purity, a higher imidazoleconcentration is required in the wash buffer, and insect cellculture supernatants generally require higher imidazoleconcentration than CHO cell culture supernatants.

The concentration of imidazole that will give optimalpurification results is sample-dependent, and is usually lowerfor Ni Sepharose excel than for other Ni Sepharose mediaused for IMAC purification. Finding the optimal imidazoleconcentration for a specific histidine-tagged protein is a trial-and-error effort, but 10 to 20 mM in the wash buffer is a goodstarting range for many proteins.

29013867 AF 9

Page 10: Instructions for Use - Cytiva

Eliminating ion-exchange effects

Addition of salt, for example 0.5 to 1.0 M NaCl in bufferseliminates ion-exchange effects.

Elution

In general, imidazole is used for elution of histidine-taggedproteins. Alternatively, the proteins may be eluted by othermethods or combinations of methods, for example bylowering pH within the range 2.5 to 5.0.

With Ni Sepharose excel no recharging with nickel ionsbetween each run is required when elution is performed usingeither low pH or imidazole. This is in contrast to many otherIMAC media, which often have to be recharged with nickel ionsafter elution using low pH.

Note: If the proteins are sensitive to low pH, it isrecommended to collect the pH-eluted fractions intubes containing a suitable volume of 1 M Tris-HCl, pH9.0 to restore a neutral pH.

4 Sample preparationBefore sample loading, whole cells must be removed by, forexample, centrifugation, otherwise clogging of the columnmay occur. When using HisTrap excel columns, no furtherclarification is needed. When using Ni Sepharose excelpacked in other columns, it is recommended to also filter thesample through a 0.45 μm filter to remove cell debris and/orother particulate material.

10 29013867 AF

Page 11: Instructions for Use - Cytiva

For optimal binding, it is not recommended to includeimidazole in sample and equilibration buffer. It isrecommended to perform binding at neutral pH. However,successful purification has routinely been observed withbinding performed at a pH as low as 6.0. Due to theprecipitation risk, avoid using strong bases or acids for pH-adjustments.

5 Ni Sepharose excel protocolsColumn packing

Ni Sepharose excel is supplied pre-swollen in 20% ethanol.

Suitable columns for packing are the Tricorn, XK and HiScalecolumns, see Chapter 11 Ordering information, on page 22.

Preparation

Step Action

1 Allow the medium slurry to sediment for 3 hours.

2 Add enough 20% ethanol to achieve a 1:1 ratio ofsettled medium and overlayed 20% ethanol.

3 Assemble the column (and packing reservoir ifnecessary).1

1 If the purification is to be performed at a high flow rate, HiScale columns arerecommended.

29013867 AF 11

Page 12: Instructions for Use - Cytiva

Step Action

4 Remove air from the end-piece and adapter by flushingwith 20% ethanol.

Make sure no air has been trapped under the columnbed support.

5 Close the column outlet leaving the bed supportcovered with 20% ethanol.

Packing

Step Action

1 Resuspend the medium and pour the slurry into thecolumn in a single continuous motion.

Pouring the slurry down a glass rod held against thecolumn wall will minimize the introduction of airbubbles.

2 Allow the medium slurry to sediment for at least 3hours.

3 If using a packing reservoir, disconnect the reservoirand fill the remainder of the column with 20% ethanol.

4 Mount the adapter and connect the column to a pump.

Avoid trapping air bubbles under the adapter or in theinlet tubing.

5 Immediately open the bottom outlet of the columnand set the pump to run at the desired flow rate.2

12 29013867 AF

Page 13: Instructions for Use - Cytiva

Step Action

6 Maintain packing flow rate for at least 3 columnvolumes (CV) after a constant bed height is reached.

7 Stop the pump and close the column outlet.

8 With the adapter inlet disconnected, quickly push theadapter down into the column until it reaches the bedsurface and then a further 3 to 4 mm into the mediumbed. Lock the adapter at this level.

9 Connect the column to a pump or a chromatographysystem and start equilibration. Re-adjust the adapter ifnecessary.

Purification protocol for Ni Sepharose excel

Note: For purification protocol for HisTrap excel , see nextSection.

Step Action

1 If the column contains 20% ethanol, wash it with 5column volumes (CV) distilled water.

Recommended flow velocity: 50 to 100 cm/h.

2 Equilibrate the column with at least 5 CV equilibrationbuffer.

Recommended flow velocity: 150 to 600 cm/h.

2 Should be at least 133% of the flow rate to be used during subsequentchromatographic procedures. If this cannot be obtained, use the maximumflow rate your pump can deliver. This should also give a well-packed bed.

29013867 AF 13

Page 14: Instructions for Use - Cytiva

Step Action

3 Load the sample.

Recommended flow velocity: 150 to 600 cm/h.

4 Wash with 20 CV wash buffer.

Recommended flow velocity: 150 cm/h.

5 Elute with elution buffer using a one-step procedure. 5CV elution buffer is usually sufficient. Alternatively, alinear elution gradient (10 to 20 CV) may give higherpurity, at the expense of lower target proteinconcentration in eluted fractions.

Recommended flow velocity: 150 cm/h.

Note: A lower flow rate during sample loading might result inhigher yield of the target protein.

Note: Purification at low temperatures increases the sampleand buffer viscosity, leading to increased backpressure. If necessary, decrease flow rate.

Note: If imidazole needs to be removed from the proteinafter the purification, prepacked desalting columnssuch as HiTrap Desalting, HiPrep™ 26/10 Desalting orPD-10 Desalting columns can be used, see Chapter 11Ordering information, on page 22.

14 29013867 AF

Page 15: Instructions for Use - Cytiva

6 HisTrap excel protocolPurification protocol

Step Action

1 Fill the pump tubing with distilled water. Remove thestopper and connect the column to thechromatography system or the laboratory pump"drop-to-drop" to avoid introducing air into thecolumn. Make sure that the connector is tight toprevent leakage.

2 Remove the snap-off end at the column outlet.

3 Wash out the ethanol with 5 column volumes (CV)distilled water.

Recommended flow rate: 1 ml/min.

4 Equilibrate the column with at least 5 CV equilibrationbuffer.

Recommended flow rates: 1 to 4 ml/min and 5 to 20ml/min for the 1 ml and 5 ml columns, respectively.

5 Load the sample.

Recommended flow rates: 1 to 4 ml/min and 5 to 20ml/min for the 1 ml and 5 ml columns, respectively.

6 Wash with 20 CV wash buffer.

Recommended flow rates: 1 ml/min and 5 ml/min forthe 1 ml and 5 ml columns, respectively.

29013867 AF 15

Page 16: Instructions for Use - Cytiva

Step Action

7 Elute with elution buffer using a one-step procedure. 5CV elution buffer is usually sufficient. Alternatively, alinear elution gradient (10 to 20 CV) may give higherpurity, at the expense of lower target proteinconcentration.

Recommended flow rates: 1 ml/min and 5 ml/min forthe 1 ml and 5 ml columns, respectively.

Note: A lower flow rate during sample loading might result inhigher yield of the target protein.

Note: Purification at low temperatures increases the sampleand buffer viscosity, leading to increased backpressure. If necessary, decrease flow rate.

Note: If imidazole needs to be removed from the proteinafter the purification, prepacked desalting columnssuch as HiTrap Desalting, HiPrep 26/10 Desalting orPD-10 Desalting columns can be used, see Chapter 11Ordering information, on page 22.

7 OptimizationThe protocols recommended in this instruction should beregarded as starting points for purification of histidine-taggedproteins. Some conditions may require optimization.

Examples of conditions which may require optimization are:

• sample loading flow rate

• sample volume

• imidazole concentration during wash

16 29013867 AF

Page 17: Instructions for Use - Cytiva

• wash flow rate

• wash volume

• temperature

• buffer composition, pH etc.

For more information, see Chapter 3 Buffer preparation, onpage 8.

8 Cleaning-in-placeWhen an increase in back pressure is seen, the column shouldbe cleaned. Different methods for cleaning-in-place (CIP) aredescribed in the table below. Do not exceed a flow velocity of150 cm/h during CIP.

To remove... Then...

Ionically boundproteins

1. Wash with several column volumes (CV) of 1.5 MNaCl.

2. Wash with approximately 10 CV distilled water orequilibration buffer.

Precipitated proteins,hydrophobically boundproteins, andlipoproteins

1. Wash the column with 1 M NaOH, contact timeusually 1 to 2 h (12 to 24 h for endotoxin removal).

2. Wash with approximately 10 CV equilibrationbuffer.

29013867 AF 17

Page 18: Instructions for Use - Cytiva

To remove... Then...

Hydrophobically boundproteins, lipoproteins,and lipids.

1. Wash with 5 to 10 CV 30% isopropanol for about 15to 20 min.

2. Wash with approximately 10 CV distilled water orequilibration buffer.

OR

1. Wash with 2 CV detergent in a basic or acidicsolution. Use, for example, 0.1 to 0.5% nonionicdetergent in 0.1 M acetic acid, contact time 1 to 2 h.

2. Remove residual detergent by washing with atleast 5 CV 70% ethanol.

3. Wash with approximately 10 CV equilibrationbuffer.

9 StorageStore Ni Sepharose excel and HisTrap excel in 20% ethanol at4°C to 30°C. Do not freeze.

Make sure that the HisTrap excel column is tightly sealed toavoid drying out.

10 TroubleshootingThe following tips may be of assistance.

For further questions about Ni Sepharose excel or HisTrapexcel , please visit cytiva.com/protein-purification, refer to theData file (29016849), or contact our technical support or yourlocal representative.

18 29013867 AF

Page 19: Instructions for Use - Cytiva

Issue Possible causeSuggested action

Increased backpressure

• Freeze-thawing of the sample may increaseprecipitation and aggregation.Avoid freeze-thawing.Remove precipitate.

• Increased sample viscosity.If the purification has been performed in a coldroom, move to room temperature if possible.Alternatively, decrease flow rate during sampleloading.

Column has clogged • High content of particulate material in the sample.Clean the column according to Chapter 8 Cleaning-in-place, on page 17. If column cleaning isunsuccessful,replace the column.Optimize sample pre-treatment before the nextsample loading.

Compression of the gelin manually packedcolumns

• High viscosity, too high flow rate.Re-adjust the top adapter.

No histidine-taggedprotein in the purifiedfractions

• No histidine-tagged protein present in the startingmaterial.Verify presence of histidine-tagged protein in thestarting material, for example by Western blotting.

• Protein found in the flowthrough: The histidine-tagis not sufficiently exposed.Consider using another position for the histidine-tag.

• Protein found in the flowthrough: Buffer/samplecomposition is not optimal.Check pH and composition of sample and equilib-ration buffer. Make sure that the concentration ofimidazole is not too high. Note that the use ofimidazole in equilibration buffer and sample is notrecommended.

29013867 AF 19

Page 20: Instructions for Use - Cytiva

Issue Possible causeSuggested action

No histidine-taggedprotein in the purifiedfractions (cont.)

• Protein eluted during wash.Ensure that the concentration of imidazole is nottoo high.

• Elution conditions are too mild (histidine-taggedprotein still bound).Elute with increased imidazole concentration.

• Protein has precipitated in the column.Decrease amount of sample, or decrease proteinconcentration by eluting with linear imidazolegradient instead of imidazole steps. Try detergentsor change NaCl concentration.

• Nonspecific hydrophobic, ionic or other interactionprevents elution.Add for example a nonionic detergent to the elutionbuffer or change the NaCl concentration.

The eluted protein isnot sufficiently pure(multiple bands on SDSpolyacrylamide gel)

• Partial degradation of tagged protein by proteases.Add protease inhibitors.

• Contaminants have high affinity for nickel ions.Elute with a stepwise or linear imidazole gradient todetermine optimal imidazole concentrations to usefor wash. Increase the wash length before elutionand/or the imidazole concentration during wash. Ashallow imidazole gradient (20 column volumes ormore) may separate proteins with similar bindingstrengths. If optimized conditions do not removecontaminants, adding a second chromatographystep such as gel filtration may be necessary.

• Contaminants are associated with the taggedprotein.Add detergent or glycerol to the wash buffer todisrupt non-specific interactions.

• Electrostatic interactions of contaminants withthe tagged protein or the chromatographymedium.Increase the sodium chloride concentration to1 to 2 M in buffers.

20 29013867 AF

Page 21: Instructions for Use - Cytiva

Issue Possible causeSuggested action

Histidine-taggedprotein is eluted tooearly during elution

• When eluting using imidazole, an initial, transientpH drop may be observed. The lower pH may affectthe affinity for bound proteins negatively, and apartially premature elution may be observed forloosely bound proteins.Increase the buffering capacity of the wash andelution buffers, which mitigates imidazole’s effecton pH.

Inadequate binding ofhistidine-taggedprotein during sampleloading/wash

• No histidine-tagged protein present in the startingmaterial.Verify presence of histidine-tagged protein in thestarting material, for example by Western blotting.

• The histidine-tag is not sufficiently exposed.Consider using another position for the histidine-tag.

• Buffer/sample composition is not optimal.Check pH and composition of sample, equilibrationbuffer and wash buffer. Make sure that theconcentration of imidazole is not too high.

• Column capacity is exceeded.Change to a larger column. For example, if HisTrapexcel 1 ml column has been used, change toHisTrap excel 5 ml.Note:Ni Sepharose excel is mainly intended for IMACpurification of histidine-tagged proteins secretedinto cell culture supernatants from eukaryotic cells,such as insect cells or CHO cells. It is also suitablefor purification of histidine-tagged proteinsoriginating from other samples causing nickelstripping. When purifying proteins from other typesof samples, use Ni Sepharose 6 Fast Flow or NiSepharose High Performance.

Temperature-relateddifferences betweenoptimization andactual purification

• Temperature affects column properties.If optimization is performed, always do this at thetemperature intended for the actual purification.

29013867 AF 21

Page 22: Instructions for Use - Cytiva

11 Ordering informationProduct Quantity Code No.

Ni Sepharose excel 25 ml

100 ml

500 ml

17371201

17371202

17371203

HisTrap excel 1 × 1 ml

5 × 1 ml

5 × 5 ml

29048586

17371205

17371206

Related products Quantity Code No.

His Mag Sepharose excel 2 × 1 ml

5 × 1 ml

10 × 1 ml

17371220

17371221

17371222

Ni Sepharose 6 Fast Flow 5 ml

25 ml

100 ml

500 ml

17531806

17531801

17531802

17531803

HisTrap FF crude 1 × 1 ml

5 × 1 ml

100 × 1 ml1

5 × 5 ml

100 × 5 ml1

29048631

11000458

11000459

17528601

17528602

HiTrap Desalting 1 × 5 ml

5 × 5 ml

100 × 5 ml1

29048684

17140801

11000329

HiPrep 26/10 Desalting 1 × 53 ml

4 × 53 ml

17508701

17508702

PD-10 Desalting Column 30 17085101

1 Special pack size delivered on specific customer order

22 29013867 AF

Page 23: Instructions for Use - Cytiva

Empty lab-scale columns Code No.

Tricorn 10/20 column, 10 mm i.d. 28406413

Tricorn 10/50 column, 10 mm i.d. 28406414

Tricorn 10/100 column, 10 mm i.d. 28406415

XK 16/20 column, 16 mm i.d. 28988937

XK 16/40 column, 16 mm i.d 28988938

XK 26/20 column, 26 mm i.d. 28988948

XK 26/40 column, 26 mm i.d. 28988949

XK 50/20 column, 50 mm i.d. 28988952

XK 50/30 column, 50 mm i.d. 28988953

HiScale 16/20 column, 16 mm i.d. 28964441

HiScale 16/40 column, 16 mm i.d. 28964424

HiScale 26/20 column, 26 mm i.d. 28964514

HiScale 26/40 column, 26 mm i.d. 28964513

HiScale 50/20 column, 50 mm i.d. 28964445

HiScale 50/40 column, 50 mm i.d. 28964444

HiTrap accessories Quantity Code No.

1/16" male/luer female1 2 18111251

Tubing connector flangeless/M6 female1 2 18100368

Tubing connector flangeless/M6 male1 2 18101798

Union 1/16" female/M6 male1 6 18111257

Union M6 female /1/16" male1 5 18385801

Union luerlock female/M6 female1 2 18102712

HiTrap/HiPrep, 1/16" male connector for ÄKTAdesign

8 28401081

Stop plug female, 1/16" 2 11000464

Fingertight stop plug, 1/16"2 5 11000355

1 One connector included in each HiTrap package.2 One fingertight stop plug is connected to the top of each HiTrap column at delivery.

29013867 AF 23

Page 24: Instructions for Use - Cytiva

Related literature Code No.

Recombinant Protein Purification Handbook, Principles andMethods

18114275

Affinity Chromatography Handbook, Principles and Methods 18102229

Affinity Chromatography Columns and Media Selection Guide 18112186

Prepacked chromatography columns for ÄKTA systems,Selection Guide

28931778

Data file, His Mag Sepharose excel, Ni Sepharose excel, HisTrapexcel

29016849

24 29013867 AF

Page 25: Instructions for Use - Cytiva

Page intentionally left blank

Page 26: Instructions for Use - Cytiva

cytiva.com

Cytiva and the Drop logo are trademarks of Global Life Sciences IP Holdco LLC or anaffiliate.

ÄKTA, HiPrep, HiScale, HiTrap, HisTrap, Sepharose and Tricorn are trademarks ofGlobal Life Sciences Solutions USA LLC or an affiliate doing business as Cytiva.

EX-CELL is a trademark of Sigma-Aldrich Co.

IMAC Sepharose products, Ni Sepharose products and Fe Sepharose products. Theseproducts are sold under a license from Sigma-Aldrich under patent number EP1276716 (Metal chelating compositions) and equivalent patents and patentapplications in other countries.

All other third-party trademarks are the property of their respective owners.

© 2020 Cytiva

All goods and services are sold subject to the terms and conditions of sale of thesupplying company operating within the Cytiva business. A copy of those terms andconditions is available on request. Contact your local Cytiva representative for themost current information.

For local office contact information, visit cytiva.com/contact

29013867 AF V:5 11/2020