instrumented impact testing: influence of machine variables and
TRANSCRIPT
OPEN REPORT
SCK•CEN-BLG-1058
Instrumented Impact Testing:
Influence of Machine Variables and
Specimen Position
Work performed during a 3-month secondment at NIST,
Boulder CO (USA), June-August 2008
Enrico Lucon
Institute for Nuclear Material Science, SCK•CEN, Mol
(Belgium)
Chris N. McCowan and Ray A. Santoyo
National Institute for Standards and Technology,
Material Reliability Division, Boulder CO (USA)
September, 2008
SCK•CEN Boeretang 200
BE-2400 Mol
Belgium
OPEN REPORT OF THE BELGIAN NUCLEAR RESEARCH CENTRE
SCK•CEN-BLG-1058
Instrumented Impact Testing:
Influence of Machine Variables and
Specimen Position
Work performed during a 3-month secondment at NIST,
Boulder CO (USA), June-August 2008
Enrico Lucon
Institute for Nuclear Material Science, SCK•CEN, Mol
(Belgium)
Chris N. McCowan and Ray A. Santoyo
National Institute for Standards and Technology, Material
Reliability Division, Boulder CO (USA)
September, 2008
Status: Unclassified
ISSN 1379-2407
SCK•CEN Boeretang 200
BE-2400 Mol
Belgium
Distribution List
R. Chaouadi
E. Lucon
J.-L. Puzzolante
J. Schuurmans
M. Scibetta
S. Van Dyck
R. Gérard
NMS, SCK•CEN
NMS, SCK•CEN
NMS, SCK•CEN
NMS, SCK•CEN
NMS, SCK•CEN
NMS, SCK•CEN
SUEZ-Tractebel
Secretariaat NMS
1
1
1
1
1
1
1
3
© SCK•CEN
Studiecentrum voor Kernenergie
Centre d’étude de l’énergie Nucléaire
Boeretang 200
BE-2400 Mol
Belgium
Phone +32 14 33 21 11
Fax +32 14 31 50 21
http://www.sckcen.be
Contact:
Knowledge Centre
RESTRICTED
All property rights and copyright are reserved. Any communication or reproduction of this document, and
any communication or use of its content without explicit authorization is prohibited. Any infringement to
this rule is illegal and entitles to claim damages from the infringer, without prejudice to any other right in
case of granting a patent or registration in the field of intellectual property.
SCK•CEN, Studiecentrum voor Kernenergie/Centre d'Etude de l'Energie Nucléaire
Stichting van Openbaar Nut – Fondation d'Utilité Publique - Foundation of Public Utility
Registered Office: Avenue Herrmann Debroux 40 – BE-1160 BRUSSEL
Operational Office: Boeretang 200 – BE-2400 MOL
SCK•CEN Open Report BLG-1058 – Page 1 of 45
Preface
This work was performed during my secondment at NIST (National Institute of
Standards and Technology) in Boulder, Colorado (USA), between June and August 2008.
An abridged version of this report was submitted to Journal of Testing and Evaluation for
publication on 25 August 2008. The report is also published as a NIST Internal Report (NISTIR).
Abstract
An investigation has been conducted on the influence of impact machine variables and specimen
positioning on characteristic forces and absorbed energies from instrumented Charpy tests.
Brittle and ductile fracture behavior has been investigated by testing NIST reference samples of
low, high and super-high energy levels. Test machine variables included tightness of foundation,
anvil and striker bolts, and the position of the center of percussion with respect to the center of
strike. For specimen positioning, we tested samples which had been moved away or sideways
with respect to the anvils. In order to assess the influence of the various factors, we compared
mean values in the reference ("unaltered") and "altered" conditions; for machine variables, t-test
analyses were also performed in order to evaluate the statistical significance of the observed
differences. Our results indicate that the only circumstance which resulted in variations larger
than 5% for both brittle and ductile specimens is when the sample is not in contact with the
anvils. These findings should be taken into account in future revisions of instrumented Charpy
test standards.
Keywords
Instrumented Charpy tests, impact machine variables, specimen positioning, t-test analysis.
SCK•CEN Open Report BLG-1058 – Page 2 of 45
Table of Contents
Preface ................................................................................................................... 1
Abstract.................................................................................................................. 1
Keywords............................................................................................................... 1
1 Introduction..................................................................................................... 3
2 Material and experimental ............................................................................... 4
2.1 Loosening machine bolts .......................................................................... 5
2.1.1 Foundation bolts................................................................................. 5
2.1.2 Anvil bolts ....................................................................................... 13
2.1.3 Striker bolts...................................................................................... 18
2.2 Changing the position of the Center of Percussion (COP)....................... 23
2.3 Specimen positioning.............................................................................. 28
2.3.1 Specimen not in contact with the anvils ........................................... 28
2.4 Lateral specimen offset ........................................................................... 34
2.5 Discussion .............................................................................................. 39
3 Conclusions................................................................................................... 43
References............................................................................................................ 44
SCK•CEN Open Report BLG-1058 – Page 3 of 45
1 Introduction
Impact testing has a history extending back to the 1850's [1], when engineers realized that
loading rate can significantly affect material properties. For structural steels, the main result of
increasing loading rate is to raise the yield and tensile strength, which in turn causes a decrease
in impact energy required for cleavage fracture, an increase in energy required for ductile
fracture and a shift in the ductile-to-brittle transition temperature. Loading rate must therefore be
considered when designing structures subject to dynamic loading, such as buildings in
earthquakes, ships and overland vehicles in collisions, aircraft landing gear and, at the highest
rates, structures subjected to ballistic projectiles and explosive shocks.
The conventional, non-instrumented Charpy test, which has recently celebrated its
official centennial [2,3], can provide useful comparative information, but generally cannot offer
any detailed insight into the failure mechanisms, or yield quantifiable material properties. This is
also true for some other impact test methods, such as the Pellini drop-weight and the Izod
pendulum test.
Instrumenting the machine striker in order to measure and record the force applied to the
notched sample during impact provides a much clearer picture of the events occurring, providing
additional insight on the behavior of material under impact loading. Instrumented Charpy tests
are currently regulated by test standard ISO 14556 [4], while standardization is also in progress
within ASTM [5].
For fracture mechanics-based structural integrity evaluations, using fatigue-precracked
Charpy (PCC) specimens is particularly useful. At the time of writing, ISO and ASTM
standardization of fracture toughness tests at impact loading rates is ongoing [6,7].
In the nuclear field, the effects of neutron exposure on the fracture properties of the
reactor pressure vessel can be assessed more reliably by using the results of instrumented Charpy
tests than by applying the current regulatory practice, which relies on an empirical correlation
between fracture toughness and the temperature corresponding to a fixed Charpy energy level
(41 J). This is done in the framework of the so-called Enhanced Surveillance Approach [8-11],
which includes the determination of characteristic force values from instrumented Charpy tests
and the derivation of alternative index temperatures [12,13].
Numerous studies have been conducted in the past [14-17] in order to investigate the
experimental factors, related to both the machine and the specimen, which can affect the results
of a Charpy test in terms of absorbed energy, as provided by the machine dial or the optical
encoder (dial energy, KV). In this study, we have examined the influence of some of these
machine/specimen variables on the results of instrumented Charpy tests, namely in terms of
characteristic forces at general yield (Fgy), maximum forces (Fm) and absorbed energies
calculated from the instrumented test record (Wt). KV values have been also included in the
analyses, as well as the ratio between Wt and KV.1
From the test machine point of view, we focused our attention on the bolts that connect
the pendulum to its foundation and those that hold the anvils and the instrumented striker in
place. Tests were run with the bolts partially loose (anvil, striker) or completely loose
(foundation), and the results were compared to data obtained with all the bolts tightened to the
manufacturer’s specifications. Additionally, the center of percussion (COP) was moved away
from the center of strike (COS) by adding weights to the pendulum hammer.
As far as specimen positioning is concerned, we investigated the effects of a lateral offset
with respect to the centered position of the notch between the anvils and of a gap between the
1
Note that Fgy values were not considered in the case of low-energy specimens, since general yield does not occur or
cannot be unambiguously defined in case of fully brittle behavior.
SCK•CEN Open Report BLG-1058 – Page 4 of 45
sample and the anvils. Again, results were compared to those obtained under ideal conditions
(i.e. specimen centered and flush against the anvils).
The ultimate goal of this investigation was to find out whether, with respect to the
"conventional" dial energy KV, instrumented Charpy forces and energies are less, equally or
more sensitive to experimental variations that could eventually occur during Charpy tests.
2 Material and experimental
NIST-certified impact-verification specimens of different energy levels were used in this
study. For all investigated variables, low (~ 15 J) and high (~ 100 J) energy specimens were
used; super-high energy (~ 240 J) samples were tested only when looking at the effect of
foundation bolts.
Tests were performed on an instrumented Tinius-Olsen Model 84 pendulum with 407.6 J
capacity, located at NIST in Boulder, Colorado. The instrumented striker conforms to the
requirements of ASTM E 23 (i.e. 8 mm radius of the striking edge).
Tests were conducted at room temperature (RT, ~ 21°C), in order to avoid issues related
to temperature control. However, when investigating the effect of loosening the foundation bolts,
an additional set of low-energy specimens was tested at -40°C. This is the prescribed temperature
for both low- and high-energy specimens, when tested for the verification of impact test
machines.
The overall test matrix is given in Table 1. In the reference condition (i.e. with the
machine fully compliant to ASTM E 23), sets of 5 specimens were tested for each energy level,
except for the low-energy samples, where two sets were tested (one at RT and one at -40°C).
Five to seven specimens per energy level and per condition were tested when investigating bolt
loosening and center of percussion. Eight or nine tests per energy level were conducted to
evaluate specimen positioning effects.
Table 1 - Overall test matrix.
Low energy Variable considered
RT -40°C
High
energy
Super-high
energy
Reference tests 5 5 5 5
Loose foundation bolts 5 5 5 5
Loose anvil bolts 5 - 6 -
Loose striker bolts 5 - 5 -
COP position changed 7 - 5 -
Lateral specimen offset 9 - 9 -
Specimen/anvils distance 8 - 8 -
TOTAL 44 10 43 10
To provide additional benchmarking, results are compared to the outcome of a recent
instrumented Charpy round-robin exercise conducted by NIST among several European and US
laboratories (6 participants) using the same low- and high-energy specimens tested here (10 tests
per material and laboratory) [18]. For the comparisons at room temperature, we used grand mean
values for force from [18] and absorbed energy values. For tests at -40°C, and with super-high
energy specimens, no force data were available, so absorbed energy values were considered.
SCK•CEN Open Report BLG-1058 – Page 5 of 45
2.1 Loosening machine bolts
ASTM E 23 requires the impact machine to be securely bolted to a concrete floor at least
150 mm thick or to a foundation having a mass at least 40 times that of the pendulum. The direct
verification for parts requiring annual inspection includes ensuring that the foundation bolts and
the bolts that attach the anvils and striker to the machine are tightened to the manufacturer's
specifications.
2.1.1 Foundation bolts
The machine used in this study is bolted to a concrete foundation having a mass of
approximately 1500 kg, over 50 times that of the pendulum. The T-bolts securing the machine
to the foundation are tightened to a torque value of 515 J (380 ft-lb), which is higher than the
manufacture specification.
Charpy tests on low, high, and super-high-energy specimens were performed with the
foundation bolts completely loosened. This is obviously an extreme situation which is unlikely to
occur in real life.
The results (instrumented forces, instrumented absorbed energy, dial energy and Wt/KV
ratio) and the comparison with reference data are shown in Table 2 and Figure 1 through Figure
7. In the Figures, solid lines connect average values and error bars (often hardly visible)
correspond to ± 2σ.
Note that, in the case of super-high-energy specimens, the acquisition time used was too
short and the tails of the instrumented curves were truncated. As a consequence, Wt and Wt/KV
values cannot be considered reliable and are, therefore, reported in italics; however, the
comparison with the reference condition remains meaningful.
Table 2 – Effect of loosening the foundation bolts. NOTE: only NIST test results are given.
Baseline Foundation unbolted
Fgy Fm Wt KV Fgy Fm Wt KV Material
(kN) (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
- 32.15 18.36 19.65 0.934 - 32.38 18.78 19.48 0.964
- 32.38 18.64 19.68 0.947 - 31.90 18.54 19.48 0.952
- 32.74 19.21 20.25 0.949 - 32.64 18.80 19.78 0.950
- 31.00 18.62 19.08 0.976 - 31.90 19.35 20.21 0.957
Low Energy - RT
- 32.16 18.25 19.05 0.958 - 32.33 18.82 19.51 0.965
Average - 32.09 18.62 19.54 0.953 - 32.23 18.86 19.69 0.958
Standard dev. - 0.653 0.372 0.497 0.015 - 0.323 0.298 0.316 0.007
- 31.49 16.64 16.70 0.996 - 31.87 15.72 15.78 0.996
- 31.55 16.84 16.89 0.997 - 30.70 16.33 16.37 0.998
- 31.35 16.75 16.70 1.003 - 31.05 16.45 16.50 0.997
- 31.60 16.98 17.03 0.997 - 31.98 17.01 17.29 0.984
Low Energy -40°C
- 31.14 16.66 16.76 0.994 - 31.37 17.02 16.99 1.002
Average - 31.43 16.77 16.82 0.998 - 31.39 16.51 16.59 0.995
Standard dev. - 0.185 0.140 0.143 0.003 - 0.541 0.541 0.584 0.007
SCK•CEN Open Report BLG-1058 – Page 6 of 45
21.19 24.56 104.81 107.54 0.975 20.94 24.79 105.76 110.37 0.958
20.81 24.59 107.86 113.01 0.954 20.96 24.78 105.69 110.49 0.957
20.91 24.55 106.04 109.08 0.972 20.98 24.78 105.61 110.61 0.955
20.97 24.55 105.78 111.53 0.948 20.99 24.77 105.54 110.73 0.953
High Energy - RT
20.95 24.56 105.47 108.88 0.969 21.01 24.77 105.46 110.85 0.951
Average 20.97 24.56 105.99 110.01 0.964 20.98 24.78 105.61 110.61 0.955
Standard dev. 0.140 0.016 1.141 2.211 0.012 0.027 0.009 0.120 0.190 0.003
20.20 25.63 221.51 237.97 0.931 20.52 25.65 225.13 243.21 0.926
20.75 25.55 231.83 252.49 0.918 20.52 25.64 228.33 246.61 0.926
20.43 25.55 226.32 244.32 0.926 20.64 25.60 221.77 239.44 0.926
20.70 25.61 222.74 239.96 0.928 20.33 25.57 221.21 235.61 0.939
Super High
Energy - RT
20.76 25.66 227.11 244.83 0.928 20.76 26.12 241.24 260.65 0.926
Average 20.57 25.60 225.90 243.91 0.926 20.55 25.72 227.54 245.10 0.928
Standard dev. 0.246 0.049 4.064 5.602 0.005 0.160 0.228 8.177 9.614 0.006
10
13
16
19
22
25
Fo
rce a
t g
en
era
l y
ield
(k
N)
High energy RT
Super high energy RT
Reference Foundation
unbolted
Fgy grand mean
from NIST
round-robin
± 95% repr. limit
(high energy)
Figure 1 - Effect of loosening the foundation bolts on forces at general yield. NIST round-robin
information only refers to high-energy specimens.
SCK•CEN Open Report BLG-1058 – Page 7 of 45
20
23
26
29
32
35
Ma
xim
um
fo
rce (
kN
)
Low energy RT
Low energy -40°C
High energy RT
Super high energy RT
Reference Foundation
unbolted
Fm reference
value from NIST
round-robin
± exp. uncert.
Figure 2 - Effect of loosening the foundation bolts on maximum forces. NIST round-robin
information refers to low and high-energy specimens tested at RT.
10
13
16
19
22
25
Wt (J
)
Low energy RT
Low energy -40°C
Reference Foundation
unbolted
Wt data from
NIST round
robin ± 1.4 J
Certified
value ± 1.4 J
(ASTM E23)
Figure 3 - Effect of loosening the foundation bolts on instrumented absorbed energies for low-
energy specimens. NOTES: tests at -40°C were not performed in the NIST round robin; certified
values only refer to tests at -40°C.
SCK•CEN Open Report BLG-1058 – Page 8 of 45
100
125
150
175
200
225
250
Wt (J
)
High energy RT
Super high energy RT
Reference Foundation
unbolted
Wt data from
NIST round
robin ± 5%
Figure 4 - Effect of loosening the foundation bolts on instrumented absorbed energies for high
and super-high-energy specimens.
10
13
16
19
22
25
KV
(J
)
Low energy RT
Low energy -40°C
Reference Foundation
unbolted
KV data from
NIST round robin
± 1.4 J
Certified value
± 1.4 J (ASTM E23)
Figure 5 - Effect of loosening the foundation bolts on dial energies for low-energy specimens.
SCK•CEN Open Report BLG-1058 – Page 9 of 45
100
120
140
160
180
200
220
240
260
KV
(J
)
Super high energy RT
High energy RT
Reference Foundation
unbolted
KV data from
NIST round robin
± 5%
Certified value
± 5% (ASTM E23)
Figure 6 - Effect of loosening the foundation bolts on dial energies for high and super-high-
energy specimens.
0.75
1
1.25
Wt/K
V
Low energy -40°C
Low energy RT
High energy RT
Reference Foundation
unbolted
Figure 7 - Effect of loosening the foundation bolts on the ratio between instrumented and dial
energy. NOTE: super-high energy specimens aren't shown since test records are incomplete.
SCK•CEN Open Report BLG-1058 – Page 10 of 45
None of observed variations appears particularly substantial. However, we used a simple
statistical test (unpaired t-test) to determine whether the differences between mean values are
statistically significant or not at the 95% confidence level. The degree of statistical significance
depends on the value of the two-tailed probability P2 using a threshold value 0.05, as follows:
• P > 0.05 ⇒ not significant
• < P < 0.05 ⇒ significant
• < P < 0.01 ⇒ very significant
• P < 0.001 ⇒ extremely significant
The results of the t-test for the loose foundation bolts are shown in Table 3. No influence
of the loosening of the foundation bolts is detected, except for the maximum forces of the high-
energy specimens.
Table 3 - Results of the statistical t-test for the loose foundation bolts.
Specimens / Energy level Parameter P Result (difference is…)
Low energy (RT)
Fm
Wt
KV
0.6701
0.2887
0.5845
Not significant
Not significant
Not significant
Low energy (-40°C)
Fm
Wt
KV
0.9035
0.3147
0.4169
Not significant
Not significant
Not significant
High energy
Fgy
Fm
Wt
KV
0.8789
< 0.0001
0.4800
0.5609
Not significant
Extremely significant
Not significant
Not significant
Super-high energy
Fgy
Fm
Wt
KV
0.9176
0.2985
0.6995
0.8170
Not significant
Not significant
Not significant
Not significant
It is also interesting to observe that all test results (reference and unbolted) are in
excellent agreement with the NIST round-robin results. Note that, for the super-high-energy
specimens, the test machine would pass the indirect verification even with the foundation
completely unbolted (Figure 6). Coincidentally, the low energy level at -40°C (Figure 5) would
fail the indirect verification in the reference condition but remarkably would pass with the
foundation bolts completely loose! In this particular case, the absorbed energy is slightly
decreased for the unbolted condition. This is contrary to what might be expected, but the scatter
for the unbolted result is much higher than for the bolted condition.
The ratio between dial and instrumented energy (Figure 7) is fairly consistent between
low-and high energy samples, and it is very close to unity for the tests conducted at -40°C.
A direct comparison of the instrumented test records (Figure 8 through Figure 11, using
only one curve per condition (for the sake of clarity) does not offer any additional insight.
2 In statistical hypothesis testing, P is the probability of obtaining a value of the test statistic at least as extreme as the one that
was actually observed, given that the null hypothesis (i.e. no difference between the means) is true.
SCK•CEN Open Report BLG-1058 – Page 11 of 45
0
5
10
15
20
25
30
35
0 0.0005 0.001 0.0015 0.002 0.0025 0.003 0.0035 0.004 0.0045 0.005
Displacement (m)
Fo
rce
(kN
)
Baseline
Unbolted
Figure 8 - Effect of loosening the foundation bolts on the instrumented test records of the low-
energy specimens (RT).
0
5
10
15
20
25
30
35
0 0.001 0.002 0.003 0.004 0.005
Displacement (m)
Fo
rce
(k
N)
Baseline
Unbolted
Figure 9 - Effect of loosening the foundation bolts on the instrumented test records of the low-
energy specimens (-40°C).
SCK•CEN Open Report BLG-1058 – Page 12 of 45
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014 0.016
Displacement (m)
Fo
rce (
kN
)
Baseline
Unbolted
Figure 10 - Effect of loosening the foundation bolts on the instrumented test records of the high-
energy specimens.
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014 0.016
Displacement (m)
Fo
rce
(k
N)
Baseline
Unbolted
Figure 11 - Effect of loosening the foundation bolts on the instrumented test records of the super-
high energy specimens. Note the curves truncated at 0.014 m.
SCK•CEN Open Report BLG-1058 – Page 13 of 45
2.1.2 Anvil bolts
The torque applied to the anvil bolts for the baseline condition is 36.6 J (27 ft-lb), as
specified by the manufacturer. The applied torque on the bolts was reduced to 2.8 J (25 in-lb), a
condition which might be described as "finger-tight".
Five tests on low-energy specimens and five tests on high-energy specimens were
performed at RT with loosened anvils bolts and the results compared to the baseline data (Table
4 and Figure 12 through Figure 16).
Table 4 – Effect of loosening the anvil bolts.
Baseline Anvil bolts loosened
Fgy Fm Wt KV Fgy Fm Wt KV Material
(kN) (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
- 32.15 18.36 19.65 0.934 - 33.55 18.80 19.94 0.943
- 32.38 18.64 19.68 0.947 - 33.78 18.64 19.31 0.965
- 32.74 19.21 20.25 0.949 - 32.58 18.57 19.61 0.947
- 31.00 18.62 19.08 0.976 - 33.63 19.48 19.78 0.985 Low Energy
- 32.16 18.25 19.05 0.958 - 32.79 18.67 19.61 0.952
Average - 32.09 18.62 19.54 0.953 - 33.27 18.83 19.65 0.958
Standard dev. - 0.653 0.372 0.497 0.015 - 0.542 0.372 0.234 0.017
21.19 24.56 104.81 107.54 0.975 20.66 24.35 106.90 110.58 0.967
20.81 24.59 107.86 113.01 0.954 20.66 24.36 100.57 104.19 0.965
20.91 24.55 106.04 109.08 0.972 21.02 24.47 107.57 111.36 0.966
20.97 24.55 105.78 111.53 0.948 20.92 24.36 104.69 108.34 0.966
20.95 24.56 105.47 108.88 0.969 20.75 24.36 107.43 112.58 0.954
High Energy
20.81 24.47 104.69 107.85 0.971
Average 20.97 24.56 105.99 110.01 0.964 20.80 24.40 105.31 109.15 0.965
Standard dev. 0.140 0.016 1.141 2.211 0.012 0.145 0.058 2.660 3.021 0.006
SCK•CEN Open Report BLG-1058 – Page 14 of 45
0
5
10
15
20
25
Fg
y (
kN
)
Reference Anvil bolts
loosened
Fgy grand mean
from NIST
round-robin
± 95% repr. limit
Figure 12 - Effect of loosening the anvil bolts on forces at general yield for the high-energy
specimens.
20
23
26
29
32
35
Fm
(k
N)
Low energy
High energy
Reference
Fm reference
value from NIST
round-robin
± exp. uncert.
Anvil bolts
loosened
Figure 13 - Effect of loosening the anvil bolts on maximum forces.
SCK•CEN Open Report BLG-1058 – Page 15 of 45
0
20
40
60
80
100
120
Wt (J
)
High energy
Low energy
Reference
Wt data from
NIST round
robin
± 1.4 J or 5%
Anvil bolts
loosened
Figure 14 - Effect of loosening the anvil bolts on instrumented absorbed energies.
0
20
40
60
80
100
120
KV
(J
) High energy
Low energy
Reference
KV data from
NIST round robin
± 1.4 J or 5%
Anvil bolts
loosened
Figure 15 - Effect of loosening the anvil bolts on dial energies.
SCK•CEN Open Report BLG-1058 – Page 16 of 45
0.75
1
1.25
Wt/K
VLow energy
High energy
Reference Anvil bolts
loosened
Figure 16 - Effect of loosening the anvil bolts on the ratio between instrumented and dial energy.
The results of the t-test to assess the statistical significance of the observed differences
between mean values are provided in Table 5.
Table 5 - Results of the statistical t-test for the loosened anvil bolts.
Specimens / Energy level Parameter P Result (difference is…)
Low energy
Fm
Wt
KV
0.0144
0.3851
0.6719
Significant
Not significant
Not significant
High energy
Fgy
Fm
Wt
KV
0.0918
0.0002
0.6078
0.6113
Not significant
Extremely significant
Not significant
Not significant
The effect of the loosened anvil bolts appears significant only on maximum force data.
However, it is noted that Fm increases for the low-energy specimens and decreases for the high-
energy specimens (Figure 13).
Wt/KV ratios remain substantially constant and the difference between dial and
instrumented energies does not exceed 5% (Figure 16).
The direct comparison of instrumented curves (Figure 17 and Figure 18) does not offer
additional insight.
SCK•CEN Open Report BLG-1058 – Page 17 of 45
0
5
10
15
20
25
30
35
0 0.0005 0.001 0.0015 0.002 0.0025 0.003 0.0035 0.004 0.0045 0.005
Displacement (m)
Fo
rce
(k
N)
Baseline
Anvil bolts loosened
Figure 17 - Effect of loosening the anvil bolts on the instrumented test records of the low-energy
specimens.
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014
Displacement (m)
Fo
rce (
kN
)
Baseline
Anvil bolts loosened
Figure 18 - Effect of loosening the anvil bolts on the instrumented test records of the high-energy
specimens.
SCK•CEN Open Report BLG-1058 – Page 18 of 45
2.1.3 Striker bolts
The bolts holding the instrumented striker in place were loosened from manufacture
specification of 74.6 J (55 ft-lb) to 4.5 J (40 in-lb), which again might be described as "finger-
tight".
Five tests on low-energy specimens and five tests on high-energy specimens were
performed at RT with loosened striker bolts and the results compared to the baseline data (Table
6 and Figure 19 through Figure 23).
Table 6 – Effect of loosening the striker bolts.
Baseline Striker bolts loosened
Fgy Fm Wt KV Fgy Fm Wt KV Material
(kN) (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
- 32.15 18.36 19.65 0.934 - 32.95 19.12 19.51 0.980
- 32.38 18.64 19.68 0.947 - 33.82 19.32 19.31 1.001
- 32.74 19.21 20.25 0.949 - 33.51 19.69 19.71 0.999
- 31.00 18.62 19.08 0.976 - 33.54 18.93 20.14 0.940
Low Energy
- 32.16 18.25 19.05 0.958 - 33.82 19.59 19.78 0.990
Average - 32.09 18.62 19.54 0.953 - 33.53 19.33 19.69 0.982
Standard dev. - 0.653 0.372 0.497 0.015 - 0.355 0.317 0.311 0.025
21.19 24.56 104.81 107.54 0.975 21.19 24.64 104.30 106.83 0.976
20.81 24.59 107.86 113.01 0.954 21.42 24.58 104.29 108.50 0.961
20.91 24.55 106.04 109.08 0.972 21.36 24.57 106.86 111.76 0.956
20.97 24.55 105.78 111.53 0.948 21.33 24.80 107.29 111.32 0.964
High Energy
20.95 24.56 105.47 108.88 0.969 21.20 24.59 104.58 107.07 0.977
Average 20.97 24.56 105.99 110.01 0.964 21.30 24.64 105.46 109.10 0.967
Standard dev. 0.140 0.016 1.141 2.211 0.012 0.101 0.096 1.483 2.326 0.009
SCK•CEN Open Report BLG-1058 – Page 19 of 45
0
5
10
15
20
25
Fg
y (
kN
)
Reference Striker bolts
loosened
Fgy grand mean
from NIST
round-robin
± 95% repr. limit
Figure 19 - Effect of loosening the striker bolts on forces at general yield for the high-energy
specimens.
20
23
26
29
32
35
Fm
(k
N)
Low energy
High energy
Reference
Fm reference
value from NIST
round-robin
± exp. uncert.
Striker bolts
loosened
Figure 20 - Effect of loosening the striker bolts on maximum forces.
SCK•CEN Open Report BLG-1058 – Page 20 of 45
0
20
40
60
80
100
120
Wt (J
)
High energy
Low energy
Reference
Wt data from
NIST round
robin
± 1.4 J or 5%
Striker bolts
loosened
Figure 21 - Effect of loosening the striker bolts on instrumented absorbed energies.
0
20
40
60
80
100
120
KV
(J
) High energy
Low energy
Reference
KV data from
NIST round robin
± 1.4 J or 5%
Striker bolts
loosened
Figure 22 - Effect of loosening the striker bolts on dial energies.
SCK•CEN Open Report BLG-1058 – Page 21 of 45
0.75
1
1.25
Wt/K
VLow energy
High energy
Reference Striker bolts
loosened
Figure 23 - Effect of loosening the striker bolts on the ratio between instrumented and dial
energy.
The statistical significance of the observed differences between mean values was
assessed by means of the t-test at a confidence level of 95% (Table 7).
Table 7 - Results of the statistical t-test for the loosened striker bolts.
Specimens / Energy level Parameter P Result (difference is…)
Low energy
Fm
Wt
KV
0.0025
0.0114
0.5880
Very significant
Significant
Not significant
High energy
Fgy
Fm
Wt
KV
0.0025
0.1263
0.5456
0.5429
Very significant
Not significant
Not significant
Not significant
In comparison with foundation and anvil bolts, loosening the striker bolts seems to have
somewhat more detectable effects, particularly for the low-energy specimens.
The difference between dial and instrumented energies remains within 5% regardless of
the torque applied to the striker bolts (Figure 23).
Once again, no additional insight is provided by comparing instrumented test records for
the two conditions (Figure 24 and Figure 25).
SCK•CEN Open Report BLG-1058 – Page 22 of 45
0
5
10
15
20
25
30
35
0 0.0005 0.001 0.0015 0.002 0.0025 0.003 0.0035 0.004 0.0045 0.005
Displacement (m)
Fo
rce (
kN
)
Baseline
Striker bolts loosened
Figure 24 - Effect of loosening the striker bolts on the instrumented test records of the low-
energy specimens.
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014
Displacement (m)
Fo
rce (
kN
)
Baseline
Striker bolts loosened
Figure 25 - Effect of loosening the striker bolts on the instrumented test records of the high-
energy specimens.
SCK•CEN Open Report BLG-1058 – Page 23 of 45
2.2 Changing the position of the Center of Percussion (COP)
The center of percussion (COP) of an object is defined as the point where a perpendicular
impact will produce translational and rotational forces which perfectly cancel each other out at
some given pivot point so that the pivot will not be moving momentarily after the impulse.
Centers of percussion are often discussed in the context of a bat, racquet, sword or other long
thin objects. For such objects, the COP may or may not be the "sweet spot" depending on the
pivot point chosen.
ASTM E 23 requires the COP of the pendulum hammer to be within 1% of the distance
between the axis of rotation and the center of strike (COS) on the specimen (corresponding to the
length of the pendulum). This requirement ensures that minimum force is transmitted to the point
of rotation. Moving COP and COS away from each other is expected to cause additional
vibrations and therefore increase vibrational energy losses, particularly when testing brittle
materials.
The location of the COP is determined from the equation:
2
2
4π
gpL = (1)
where g is the local gravitational acceleration in m/s², p is the period of a complete oscillation
measured with a suitable time-measuring device in s, and L is the distance between the axis of
rotation and the COP in m.
The length of the pendulum used in this study is 900.65 mm, and the original position of
the COP is 898.79 mm from the axis of rotation. The COP-COS distance is therefore 1.86 mm,
or 0.2% of the pendulum length.
By adding 2 kg to the pendulum mass, we reduced the oscillation period and therefore
shortened the distance between axis of rotation and COP to 893.73 mm and increased the
COP-COS distance to 6.95 mm (0.77% of the pendulum length3). Subsequently, tests were run
on low and high-energy specimens and results were compared to data from reference tests and
the NIST round robin (Table 8 and Figure 26 through Figure 30).
Table 8 – Effect of moving the COP away from the COS4.
Baseline COP moved
Fgy Fm Wt KV Fgy Fm Wt KV Material
(kN) (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
- 32.15 18.36 19.65 0.934 - - - 18.94 -
- 32.38 18.64 19.68 0.947 - 31.98 17.88 19.40 0.922
- 32.74 19.21 20.25 0.949 - 31.43 17.89 19.47 0.919
- 31.00 18.62 19.08 0.976 - 31.62 17.77 19.30 0.921
- 32.16 18.25 19.05 0.958 - 31.56 17.67 18.98 0.931
- 31.94 17.64 - -
Low Energy
- 31.08 17.90 - -
Average - 32.09 18.62 19.54 0.953 - 31.65 17.80 19.22 0.923
Standard dev. - 0.653 0.372 0.497 0.015 - 0.235 0.104 0.244 0.005
3 It was not possible to add more weight to the pendulum mass, and therefore increase further the COP-COS distance.
4 For one of the low energy tests, instrumented data were lost. For two additional low energy tests, an incorrect value of pendulum length was
entered and therefore KV is not reliable.
SCK•CEN Open Report BLG-1058 – Page 24 of 45
Baseline COP moved
Fgy Fm Wt KV Fgy Fm Wt KV Material
(kN) (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
21.19 24.56 104.81 107.54 0.975 20.84 24.55 109.39 115.61 0.946
20.81 24.59 107.86 113.01 0.954 20.98 24.45 108.40 113.87 0.952
20.91 24.55 106.04 109.08 0.972 20.77 24.20 105.58 111.37 0.948
20.97 24.55 105.78 111.53 0.948 21.00 24.39 107.95 113.83 0.948 High Energy
20.95 24.56 105.47 108.88 0.969 20.96 24.35 102.91 107.06 0.961
Average 20.97 24.56 105.99 110.01 0.964 20.91 24.39 106.85 112.35 0.951
Standard dev. 0.140 0.016 1.141 2.211 0.012 0.100 0.129 2.608 3.319 0.006
0
5
10
15
20
25
Fg
y (
kN
)
Reference COP moved
away
Fgy grand mean
from NIST
round-robin
± 95% repr. limit
Figure 26 - Effect of changing the COP position on forces at general yield for the high-energy
specimens.
SCK•CEN Open Report BLG-1058 – Page 25 of 45
20
23
26
29
32
35
Fm
(k
N)
Low energy
High energy
Reference
Fm reference
value from NIST
round-robin
± exp. uncert.
COP moved
away
Figure 27 - Effect of changing the COP position on maximum forces.
0
20
40
60
80
100
120
Wt (J
)
High energy
Low energy
Reference
Wt data from
NIST round
robin
± 1.4 J or 5%
COP moved
away
Figure 28 - Effect of changing the COP position on instrumented absorbed energies.
SCK•CEN Open Report BLG-1058 – Page 26 of 45
0
20
40
60
80
100
120
KV
(J
) High energy
Low energy
Reference
KV data from
NIST round robin
± 1.4 J or 5%
COP moved
away
Figure 29 - Effect of changing the COP position on dial energies.
0.75
1
1.25
Wt/K
V
Low energy
High energy
Reference COP moved
away
Figure 30 - Effect of changing the COP position on the ratio between instrumented and dial
energy.
SCK•CEN Open Report BLG-1058 – Page 27 of 45
Table 9 shows the results of the statistical analyses conducted using the t-test.
Table 9 - Results of the statistical t-test for the modified COP location tests.
Specimens / Energy level Parameter P Result (difference is…)
Low energy
Fm
Wt
KV
0.1453
0.0006
0.2268
Not significant
Extremely significant
Not significant
High energy
Fgy
Fm
Wt
KV
0.4866
0.0175
0.5212
0.2259
Not significant
Significant
Not significant
Not significant
The most significant effect is observed on instrumented energies for the low-energy
specimens which, contrary to expectations, decreased when the COP was moved away from the
COS. As a consequence, the ratio between instrumented and dial energies falls below 0.95
(Figure 30).
Figure 31 and Figure 32 compare instrumented test records for low- and high energy
samples.
0
5
10
15
20
25
30
35
0 0.0005 0.001 0.0015 0.002 0.0025 0.003 0.0035 0.004 0.0045 0.005
Displacement (m)
Fo
rce
(k
N)
Baseline
COP moved away
Figure 31 - Effect of changing the COP position on the instrumented test records of the low-
energy specimens.
SCK•CEN Open Report BLG-1058 – Page 28 of 45
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014
Displacement (m)
Fo
rce
(k
N)
COP moved away
Baseline
Figure 32 - Effect of changing the COP position on the instrumented test records of the high-
energy specimens.
2.3 Specimen positioning
The influence of specimen impact position was investigated by moving the samples away
from the correct location (i.e. flush against the anvils and with the notch centered between the
anvils). Conditions with specimens moved away from flush, and specimen notch positions
moved off center were evaluated.
For every selected distance, multiple impact tests on low- and high-energy specimens
were performed. Due to the limited number of data available for each condition (from 2 to 4), we
decided not to perform any statistical analysis (t-test).
2.3.1 Specimen not in contact with the anvils
In everyday laboratory practice, it is not uncommon that a Charpy specimen is not fully
in contact with the anvils, particularly in the case of manual positioning and when the tests are
performed at low or high temperatures, since the operator must fulfill the < 5 s transfer time
requirement (ASTM E 23).
A sample which was not against the anvils can be easily detected from its instrumented
test record, which shows significantly amplified dynamic oscillations up to general yield and a
characteristic delay in the rise of the force signal with respect to the onset of data acquisition.
For our study, we selected three values of the distance between specimen and anvils: 0.5
mm, 1 mm and 2 mm. Up to 5 tests per condition were performed, although instrumented data
were lost for several tests as will be discussed later. The results are summarized and compared
with the reference data and NIST round-robin results in Table 10 and Figure 33 through Figure
37.
SCK•CEN Open Report BLG-1058 – Page 29 of 45
Table 10 – Effect of the distance between specimen and anvils.
Low Energy High Energy
Fm Wt KV Fgy Fm Wt KV Distance
from anvils (kN) (J) (J)
Wt/KV (kN) (kN) (J) (J)
Wt/KV
32.15 18.36 19.65 0.934 21.19 24.56 104.81 107.54 0.975
32.38 18.64 19.68 0.947 20.81 24.59 107.86 113.01 0.954
32.74 19.21 20.25 0.949 20.91 24.55 106.04 109.08 0.972
31.00 18.62 19.08 0.976 20.97 24.55 105.78 111.53 0.948
0 mm
(reference)
32.16 18.25 19.05 0.958 20.95 24.56 105.47 108.88 0.969
Average 32.09 18.62 19.54 0.953 20.97 24.56 105.99 110.01 0.964
Standard dev. 0.653 0.372 0.497 0.015 0.140 0.016 1.141 2.211 0.012
Fm Wt KV Fgy Fm Wt KV Distance from
anvils (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
33.98 19.73 20.61 0.957 19.54 24.92 108.95 114.19 0.954
34.78 21.22 22.70 0.935 19.22 24.52 107.48 112.30 0.957
34.83 20.22 21.38 0.946 19.49 24.79 107.84 110.50 0.976 0.5 mm
34.86 20.45 21.52 0.950
Average 34.61 20.41 21.55 0.947 19.42 24.74 108.09 112.33 0.962
Standard dev. 0.423 0.621 0.863 0.009 0.172 0.204 0.766 1.845 0.012
Fm Wt KV Fgy Fm Wt KV Distance from
anvils (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
33.70 22.10 22.76 0.971 19.94 24.82 114.48 118.62 0.965 1 mm
35.62 21.04 22.53 0.934 19.66 24.84 110.03 113.53 0.969
Average 34.66 21.57 22.65 0.952 19.80 24.83 112.26 116.08 0.967
Standard dev. 1.358 0.750 0.163 0.026 0.198 0.014 3.147 3.599 0.003
Fm Wt KV Fgy Fm Wt KV Distance from
anvils (kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
35.62 24.54 25.48 0.963 20.45 24.84 114.76 117.02 0.981
35.02 25.10 26.23 0.957 22.00 24.87 112.22 113.29 0.991 2 mm
20.66 25.14 111.75 112.88 0.990
Average 35.32 24.82 25.86 0.960 21.04 24.95 112.91 114.40 0.987
Standard dev. 0.424 0.396 0.530 0.004 0.841 0.165 1.619 2.281 0.006
SCK•CEN Open Report BLG-1058 – Page 30 of 45
0
5
10
15
20
25
0 0.5 1 1.5 2
Specimen distance from anvils (mm)
Fg
y (
kN
)
Fgy grand mean
from NIST
round-robin
± 95% repr. limit
Figure 33 - Effect of the distance between specimen and anvils on forces at general yield for the
high-energy specimens.
20
25
30
35
40
0 0.5 1 1.5 2
Specimen distance from anvils (mm)
Fm
(k
N)
Low
energy
High
energy
Fm reference
value from NIST
round-robin
± exp. uncert.
Figure 34 - Effect of the distance between specimen and anvils on maximum forces.
SCK•CEN Open Report BLG-1058 – Page 31 of 45
0
20
40
60
80
100
120
140
0 0.5 1 1.5 2
Specimen distance from anvils (mm)
Wt (J
)
Low
energy
High
energy
Wt data from
NIST round robin
± 1.4 J or 5%
Figure 35 - Effect of the distance between specimen and anvils on instrumented absorbed
energies.
0
20
40
60
80
100
120
140
0 0.5 1 1.5 2
Specimen distance from anvils (mm)
KV
(J)
Low
energy
High
energy
KV data from
NIST round robin
± 1.4 J or 5%
Figure 36 - Effect of the distance between specimen and anvils on dial energies.
SCK•CEN Open Report BLG-1058 – Page 32 of 45
0.75
1
1.25
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Specimen distance from the anvils (mm)
Wt/K
V
Low
energy
High
energy
Figure 37 - Effect of the distance between specimen and anvils on the ratio between
instrumented and dial energies.
Increasing the distance between specimen and anvils results in a generalized increase of
the characteristic values of force and energy, as well as the Wt/KV ratio. The increment is
particularly significant for maximum force values measured from low-energy specimens (Figure
34), where the enhanced dynamic oscillations lead to a significant overestimation of Fm. For the
same reason, the uncertainty associated with Fgy values from high-energy specimens increases,
while ringing effects are sufficiently dampened once maximum force is reached.
In general, brittle specimens are more affected than ductile specimens.
As previously mentioned, the occurrence of instrumented data loss was problematic for
the low energy samples at 1 mm and 2 mm distance: data were not acquired in 5 out of 9 tests,
whereas in only one instance (out of 6 tests performed) high energy data were lost. No
acquisition failure was recorded for specimens placed 0.5 mm away from the anvils. It is
therefore clear that distances of 1 mm or more increase the likelihood of acquisition failure,
probably due to false triggering.
Two interesting features emerge from the comparison of instrumented test records
(Figure 38 and Figure 39):
• the traces of the offset low-energy specimens are clipped at a force level of 35.6 kN, which
corresponds to the threshold of the acquisition system (Figure 38);
• curves for the reference condition do not reach this force level;
• on the displacement axis, the rise of the force signal roughly corresponds to the initial
distance between specimen and anvils.
In all cases, large inertia peaks are observed at the beginning of the instrumented traces,
caused by the specimen rebounding between anvils and striker before full contact is achieved.
SCK•CEN Open Report BLG-1058 – Page 33 of 45
0
5
10
15
20
25
30
35
0 0.0005 0.001 0.0015 0.002 0.0025 0.003 0.0035 0.004 0.0045 0.005
Displacement (m)
Fo
rce (
kN
)
Reference (0 mm)
0.5 mm
1 mm
2 mm
Figure 38 - Effect of the distance between specimen and anvils on the instrumented test records
of low-energy specimens. Note the clipped force signals above 35.6 kN.
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014
Displacement (m)
Fo
rce (
kN
)
Reference (0 mm)
0.5 mm
1 mm
2 mm
Figure 39 - Effect of the distance between specimen and anvils on the instrumented test records
of high-energy specimens.
SCK•CEN Open Report BLG-1058 – Page 34 of 45
2.3.2 Lateral specimen offset
Annex A1 of ASTM E 23 prescribes that the specimen be positioned against the anvils
such that the center of the notch is located within 0.25 mm of the midpoint between the anvils.
The use of self-centering tongs is recommended to achieve this.
In our study, we selected three values for the specimen lateral offset: 0.5 mm, 1 mm and
2 mm. All distances were outside the ASTM E 23 tolerance.
For each condition, three tests on low-energy and three test on high-energy specimens
were conducted. No failed data acquisition was experienced. Results and comparison with
reference values and NIST round-robin data are provided in Table 11 and Figure 40 through
Figure 44.
Table 11 – Effect of lateral specimen offset.
Low Energy High Energy
Fm Wt KV Fgy Fm Wt KV Lateral offset
(kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
32.15 18.36 19.65 0.934 21.19 24.56 104.81 107.54 0.975
32.38 18.64 19.68 0.947 20.81 24.59 107.86 113.01 0.954
32.74 19.21 20.25 0.949 20.91 24.55 106.04 109.08 0.972
31.00 18.62 19.08 0.976 20.97 24.55 105.78 111.53 0.948
0 mm
(reference)
32.16 18.25 19.05 0.958 20.95 24.56 105.47 108.88 0.969
Average 32.09 18.62 19.54 0.953 20.97 24.56 105.99 110.01 0.964
Standard dev. 0.653 0.372 0.497 0.015 0.140 0.016 1.141 2.211 0.012
Fm Wt KV Fgy Fm Wt KV Lateral offset
(kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
32.32 18.46 18.94 0.975 21.06 24.56 108.62 113.61 0.956
33.14 18.93 19.21 0.985 20.98 24.57 108.30 113.08 0.958 0.5 mm
32.60 18.95 19.91 0.952 21.05 24.37 106.40 111.28 0.956
Average 32.69 18.78 19.35 0.971 21.03 24.50 107.77 112.66 0.957
Standard dev. 0.417 0.277 0.501 0.017 0.044 0.113 1.200 1.221 0.001
Fm Wt KV Fgy Fm Wt KV Lateral offset
(kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
33.50 19.36 20.11 0.963 21.17 24.60 105.67 111.73 0.946
32.51 18.53 18.98 0.976 21.10 24.59 108.45 112.87 0.961 1 mm
33.30 18.80 19.04 0.987 20.97 24.62 110.12 114.68 0.960
Average 33.10 18.90 19.38 0.975 21.08 24.60 108.08 113.09 0.956
Standard dev. 0.523 0.423 0.636 0.012 0.101 0.015 2.248 1.488 0.009
Fm Wt KV Fgy Fm Wt KV Lateral offset
(kN) (J) (J) Wt/KV
(kN) (kN) (J) (J) Wt/KV
33.64 19.50 20.68 0.943 21.24 24.72 103.45 113.77 0.909
34.49 20.67 21.01 0.984 21.24 24.93 103.39 109.52 0.944 2 mm
32.79 18.88 19.44 0.971 21.54 24.92 107.92 114.10 0.946
Average 33.64 19.68 20.38 0.966 21.34 24.86 104.92 112.46 0.933
Standard dev. 0.850 0.909 0.828 0.021 0.173 0.118 2.598 2.554 0.021
SCK•CEN Open Report BLG-1058 – Page 35 of 45
0
5
10
15
20
25
0 0.5 1 1.5 2
Lateral specimen offset (mm)
Fg
y (
kN
)
Fgy grand mean
from NIST
round-robin
± 95% repr. limit
Figure 40 - Effect of lateral specimen offset on forces at general yield for the high-energy
specimens.
20
25
30
35
40
0 0.5 1 1.5 2
Lateral specimen offset (mm)
Fm
(kN
)
Low
energy
High
energy
Fm reference
value from NIST
round-robin
± exp. uncert.
Figure 41 - Effect of lateral specimen offset on maximum forces.
SCK•CEN Open Report BLG-1058 – Page 36 of 45
0
20
40
60
80
100
120
140
0 0.5 1 1.5 2
Lateral specimen offset (mm)
Wt (J
)
Low
energy
High
energy
Wt data from
NIST round robin
± 1.4 J or 5%
Figure 42 - Effect of lateral specimen offset on instrumented absorbed energies.
0
20
40
60
80
100
120
140
0 0.5 1 1.5 2
Lateral specimen offset (mm)
KV
(J
)
Low
energy
High
energy
KV data from
NIST round robin
± 1.4 J or 5%
Figure 43 - Effect of lateral specimen offset on dial energies.
SCK•CEN Open Report BLG-1058 – Page 37 of 45
0.75
1
1.25
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Lateral specimen offset (mm)
Wt/K
V Low
energy
High
energy
Figure 44 - Effect of lateral specimen offset on the ratio between instrumented and dial energies.
Moving the specimen off-center appears to have a systematic, though moderate effect on
instrumented forces. The force values steadily increase with the offset (Figure 40 and Figure 41).
The effect is less pronounced and not systematic for energy values and Wt/KV ratio (Figure 42 to
Figure 44), with both Wt and KV being slightly lower for 2 mm than for 1 mm for the high-
energy specimens.
No substantial features emerge from the comparison of instrumented test records (Figure
45 and Figure 46). For low energy tests, the amplitude of dynamic oscillations seems to increase
with respect to the baseline condition, but it is not directly correlated to the magnitude of the
specimen offset.
SCK•CEN Open Report BLG-1058 – Page 38 of 45
0
5
10
15
20
25
30
35
0 0.0005 0.001 0.0015 0.002 0.0025 0.003 0.0035 0.004 0.0045 0.005
Displacement (m)
Fo
rce
(k
N)
Reference (0 mm)
0.5 mm
1 mm
2 mm
Figure 45 - Effect of lateral specimen offset on the instrumented test records of low-energy
specimens.
0
5
10
15
20
25
30
0 0.002 0.004 0.006 0.008 0.01 0.012 0.014
Displacement (m)
Fo
rce
(k
N)
Reference (0 mm)
0.5 mm
1 mm
2 mm
Figure 46 - Effect of lateral specimen offset on the instrumented test records of high-energy
specimens.
SCK•CEN Open Report BLG-1058 – Page 39 of 45
3 Discussion
Table 12 summarizes the results of the statistical t-test applied to the results of the tests
conducted on low, high and super-high-energy specimens in order to investigate the effect of test
machine variables. In the Table, the following convention is used to indicate the statistical
significance (at the 95% confidence level) of the observed differences between mean values:
Symbol Meaning
NS Not significant
* Significant
** Very significant
*** Extremely significant
Table 12 - Statistical significance of the various machine parameters based on the t-test results.
Test machine
variable
Test
parameter
Low energy
specimens (RT)
Low energy
specimens (-40°C)
High energy
specimens
Super-high energy
specimens
Foundation bolts
Fgy
Fm
Wt
KV
-
NS
NS
NS
-
NS
NS
NS
NS
***
NS
NS
NS
NS
NS
NS
Anvil bolts
Fgy
Fm
Wt
KV
-
*
NS
NS
-
-
-
-
NS
***
NS
NS
-
-
-
-
Striker bolts
Fgy
Fm
Wt
KV
-
**
*
NS
-
-
-
-
**
NS
NS
NS
-
-
-
-
COP position
Fgy
Fm
Wt
KV
-
NS
***
NS
-
-
-
-
NS
*
NS
NS
-
-
-
-
From the examination of Table 12, it is generally clear that the investigated effects are
neither particularly substantial nor systematic, nor is the consistency between the different
parameters and energy levels. Therefore, it is unlikely that a user could detect a given effect from
a simple examination of an individual test record, without resorting to a more detailed statistical
analysis. Force values (namely maximum forces) seem to be slightly more sensitive than energy
values.
As far as specimen position is concerned, our results show that moving the samples away
from the anvils is more influential than displacing it sideways. The effects are clearly visible on
the instrumented test record and are particularly significant when the distance between specimen
and anvils exceeds 0.5 mm. Brittle specimens are more affected than ductile specimens, and for
our test system clipping of the force signal due to saturation of the strain gage signal was
observed. Moreover, the likelihood of losing the instrumented data due to mistriggering is
greatly increased. A lateral offset of the sample affects forces more than energies, but the
consequences are not very pronounced.
Table 13 and Figure 47 through Figure 49 summarize the outcome of our investigation
for brittle and ductile Charpy tests. Data are presented in terms of average forces and energies
measured after "altering" one of the experimental parameters, as a function of the average values
for the reference ("unaltered") condition.
SCK•CEN Open Report BLG-1058 – Page 40 of 45
Table 13 - Average values of characteristic forces and absorbed energies obtained in this study.
Fracture
behavior Material/Tests Variable considered
Fgy
(kN)
Fm
(kN)
Wt
(J)
KV
(J)
Reference condition 32.09 18.62 19.54
Foundation bolts 32.23 18.86 19.69
Anvil bolts 33.27 18.83 19.65
Striker bolts 33.53 19.33 19.69
Moving COP 31.60 17.79 19.22
Not against anvils (0.5 mm) 34.61 20.41 21.55
Not against anvils (1 mm) 34.66 21.57 22.65
Not against anvils (2 mm) 35.32 24.82 25.86
Lateral offset (0.5 mm) 32.69 18.78 19.35
Lateral offset (1 mm) 33.10 18.90 19.38
Low energy (RT)
Lateral offset (2 mm) 33.64 19.68 20.38
Reference condition 31.43 16.77 16.82
Brittle
Low energy (-40°C) Foundation bolts
31.39 16.51 16.59
Reference condition 20.97 24.56 105.99 110.01
Foundation bolts 20.98 24.78 105.61 110.61
Anvil bolts 20.80 24.40 105.31 109.15
Striker bolts 21.30 24.64 105.46 109.10
Moving COP 20.91 24.39 106.85 112.35
Not against anvils (0.5 mm) 19.42 24.74 108.09 112.33
Not against anvils (1 mm) 19.80 24.83 112.26 116.08
Not against anvils (2 mm) 21.04 24.95 112.91 114.40
Lateral offset (0.5 mm) 21.03 24.50 107.77 112.66
Lateral offset (1 mm) 21.08 24.60 108.08 113.09
High energy
Lateral offset (2 mm) 21.34 24.86 104.92 112.46
Reference condition 20.57 25.60 225.90 243.91
Ductile
Super-high energy Foundation bolts 20.55 25.72 227.54 245.10
SCK•CEN Open Report BLG-1058 – Page 41 of 45
16
20
24
28
32
36
16 20 24 28 32 36
Reference values (Fm - kN; Wt,KV - J)
Alt
ered
va
lues
(F
m -
kN
; W
t,K
V -
J)
Foundation unbolted
Anvil bolts loosened
Striker bolts loosened
Moving COP away from COS
Sample 0.5 mm from anvils
Sample 1 mm from anvils
Sample 2 mm from anvils
Lateral offset 0.5 mm
Lateral offset 1 mm
Lateral offset 2 mm
Wt, KV
Fm
−10%
+10%
−5%
+5% 1:1
Figure 47 - Effect of test-machine variables and specimen positioning for brittle specimens (low-
energy level at RT and -40°C).
16
20
24
28
16 20 24 28
Reference force values (kN)
Alt
ered
fo
rce
valu
es
(kN
)
Foundation unbolted
Anvil bolts loosened
Striker bolts loosened
Moving COP away from COS
Sample 0.5 mm from anvils
Sample 1 mm from anvils
Sample 2 mm from anvils
Lateral offset 0.5 mm
Lateral offset 1 mm
Lateral offset 2 mm
−10%
+10%
−5%
+5% 1:1
Fgy
Fm
Figure 48 - Effect of test-machine variables and specimen positioning on characteristic forces for
ductile specimens (high and super-high energy).
SCK•CEN Open Report BLG-1058 – Page 42 of 45
100
125
150
175
200
225
250
100 125 150 175 200 225 250
Reference energy values (J)
Alt
ere
d e
ne
rgy
valu
es (
J) Foundation unbolted
Anvil bolts loosened
Striker bolts loosened
Moving COP away from COS
Sample 0.5 mm from anvils
Sample 1 mm from anvils
Sample 2 mm from anvils
Lateral offset 0.5 mm
Lateral offset 1 mm
Lateral offset 2 mm
−10%
+10%
−5%
+5% 1:1
Low
energy
Super-high
energy
Figure 49 - Effect of test-machine variables and specimen positioning on absorbed energies for
ductile specimens (high and super-high energy).
In the case of brittle behavior (Figure 47), most of the investigated variables tend to
increase maximum forces and absorbed energies as a result of additional vibrational energy
losses, except when the COP is moved away from the COS. In general, altering the specimen
position before impact is more influential than loosening the test machine bolts.
The variations observed with respect to the reference condition are relatively moderate,
i.e. less than 5%, except when the specimen is moved away from the anvil by more than 0.5 mm.
For specimen/anvil distances of 1 and 2 mm, absorbed energies increase by much more than
10%.
Fully ductile specimens (Figure 48 and Figure 49) are generally less sensitive than brittle
specimens to the investigated variables. In particular (Figure 48), maximum forces are
essentially unaffected (variations well below 5%), whereas for general yield forces only the
distance between specimen and anvil causes variations between 5% and 10%. In terms of
absorbed energies (Figure 49), from both the instrumented test record and from the machine
encoder, an increase is generally observed, but no larger than 5-6% of the reference value.
SCK•CEN Open Report BLG-1058 – Page 43 of 45
4 Conclusions
Our investigation on the effect of test-machine variables (bolting of foundation, anvils
and striker; position of the center of percussion) on instrumented Charpy test results leads to the
general conclusion that most of the observed variations are lower than 10% and generally neither
consistent nor systematic, particularly in case of fully ductile behavior. As a consequence, they
are unlikely to be detected from the instrumented test record or from the value of absorbed
energy yielded by the machine encoder.
The most notable effect is when the specimen is not in contact with the anvils;
particularly for brittle tests (such as the low-energy NIST-reference samples) and when the
sample is more than 0.5 mm from the anvils, absorbed energies can be overestimated by
considerably more than 10%. It would be advisable that instrumented Charpy test standards
require data from Charpy tests in which the gap between specimen and anvils was more than 0.5
mm to be discarded. This requirement would be relatively easy to implement, since (a) the test
record of such a test can be easily recognized (delay between test initiation and force signal rise;
very pronounced dynamic oscillations prior to general yield), and (b) the displacement
corresponding to the rise of the force signal provides a rough estimate of the initial distance
between specimen and anvils.
One the most significant outcomes of our study is that a "good" machine has a high
probability of still delivering high quality results, including passing the indirect verification of
ASTM E 23, even when it is unbolted from its foundation, the striker and anvils are just "finger
tightened" and the specimens are offset from the midpoint between the anvils. However, we
expect that these results are strongly influenced by machine design. So, the magnitude of the
effects reported here may differ substantially from results obtained on a machine having a
different design.
Finally, we can say that instrumented forces are more sensitive than absorbed energies to
the variables considered, whereas calculated (Wt) and dial (KV) energies are equally
(in)sensitive. This implies that, if Charpy tests are performed using an instrumented striker,
effects can be detected which would not be noticed if one looked at only energy values returned
by the machine encoder.
SCK•CEN Open Report BLG-1058 – Page 44 of 45
References
[1] M. P. Manahan, Sr., and T. A. Siewert, "The History of Instrumented Impact Testing," in
Pendulum Impact Machines: Procedures and Specimens, ASTM STP 1476, T. Siewert,
M. Manahan, and C. McCowan, eds., ASTM International, West Conshohocken, PA,
2006, pp. 3-11.
[2] Pendulum Impact Testing – A Century of Progress, ASTM STP 1380, T. A. Siewert, and
M. P. Manahan, Sr., eds., ASTM International, West Conshohocken, PA, 2000.
[3] From Charpy to Present Impact Testing, D. François, and A. Pineau, eds., Peer-reviewed
Proceedings of the Charpy Centenary Conference, 2-5 October 2001, Poitiers, France,
ESIS 30, 2002.
[4] International Standard ISO 14556:2000(E), "Steel — Charpy V-notch pendulum impact
test — Instrumented test method".
[5] ASTM Subcommittee E28.07, Work Item WK20198, "Instrumented Impact Testing of
Metallic Materials," initiated 06/12/2008.
[6] ISO/CD 26843, "Metallic materials - Measurement of fracture toughness of steels at
impact loading rates using precracked Charpy specimens," 2008-07-07.
[7] ASTM Subcommittee E08.07, Work Item WK20198, New Annex (A17) to ASTM E
1820, "Standard Test Method for Measurement of Fracture Toughness," initiated
08/05/2008.
[8] A. Fabry, et al., “Enhanced Surveillance of Nuclear Pressure Vessel Steels,” SCK•CEN
Open Report BLG-668.
[9] A. Fabry, et al., “Enhancing the Surveillance of LWR Ferritic Steel Components,” IAEA
Specialists’ Meeting on ‘Technology for Lifetime Management of Nuclear Power Plants’,
Tokyo, Japan, 15-17 November 1994.
[10] Fabry, A., et al., “BR2/CHIVAS Irradiations in Support of Enhanced Surveillance of
Nuclear Reactor Pressure Vessels,” IAEA Specialists’ Meeting on ‘Irradiation
Embrittlement and Mitigation’, Espoo, Finland, 23-26 October 1995.
[11] A. Fabry, et al., “On the Use of Instrumented Charpy ‘V’ Impact Signal for Assessment
of RPVS Embrittlement,” in Evaluating Material Properties by Dynamic Testing, ESIS
20, E. van Walle, ed., MEP, London, pp.59-78.
[12] R. Chaouadi, “Status Report on the Load Diagram Approach for Characterization Reactor
Pressure Vessel Steels,” SCK•CEN Report R-3560, August 2001.
[13] R. Gérard, et al., “In-Service Embrittlement of the Pressure Vessel Welds at the Doel I
and II Nuclear Power Plants,” in Effects of Irradiation on Materials: 17th
International
Symposium, ASTM STP 1270, D.S. Gelles, R.K. Nanstad, A.S. Kumar and E.A. Little,
eds., ASTM, Philadelphia, 1995, pp. 294-319.
[14] N. H. Fahey, "The Charpy Impact Test - Its Accuracy and Factors Affecting Test
Results," in Impact Testing of Metals, ASTM STP 466, D. E. Driscoll, ed., ASTM,
Philadelphia, 1970, pp. 76-92.
[15] G. Revise, "Influence of Dimensional Parameter of an Impact Test Machine on the
Results of a Test, in Charpy Impact Test: Factors and Variables, ASTM STP 1072, J. M.
Holt, ed., ASTM, Philadelphia, 1990, pp. 35-53.
[16] E. A. Ruth, D. P. Vigliotti, and T. A. Siewert, "Effect of Surface Finish of Charpy Anvils
and Striking Bits on Absorbed Energy," in Pendulum Impact Machines: Procedures and
Specimens for Verification, ASTM STP 1248, T. A. Siewert, and A. K. Schmieder, eds.,
ASTM, Philadelphia, 1995, pp. 91-100.
[17] Y. Yamaguchi, S. Takagi, and H. Nakano, "Effects of Anvil Configurations on Absorbed
Energy," in Pendulum Impact Testing – A Century of Progress, ASTM STP 1380, T. A.
SCK•CEN Open Report BLG-1058 – Page 45 of 45
Siewert, and M. P. Manahan, Sr., eds., ASTM International, West Conshohocken, PA,
2000, pp 164-180.
[18] C. M. McCowan, J. D. Splett, and E. Lucon, "Dynamic Force Measurement:
Instrumented Charpy Impact Testing", NISTIR 6652, August 2008.