INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator. INSULATION TESTER CALIBRATION Overview. The 3200 has five functions for calibrating insulation testers : High value resistance output for insulation testing (0 MOhms to 2,000MOhms) - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator
Page 2: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

INSULATION TESTER CALIBRATIONUsing the 3200 Electrical Test Calibrator

Page 3: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

The 3200 has five functions for calibrating insulation testers :

1. High value resistance output for insulation testing (0 MOhms to 2,000MOhms)

2. Measurement of the insulation test voltage3. Low value resistance for continuity testing

(0.1 Ohm to 20.0 Ohms Variable, 100 Ohms & 1kOhm Fixed Ranges)4. Insulation test current measurement5. A.C. voltage output at 100V, 200V, 240V, 300V and 400V

Dedicated inputs and test buttons for Insulation Testers provides safe, convenient connection & operation

INSULATION TESTER CALIBRATIONOverview

Page 4: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

1. CONNECT TESTER TO INSULATION TESTER TERMINALS

2. SET CONTINUITY RANGE ON TESTER

3. SET CONTINUITY RESISTANCE ON 3200, EG. 1 OHM

4. MEASURE RESISTANCE USING TESTER

INSULATION TESTER CALIBRATIONContinuity Resistance

Page 5: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

1. CONNECT TESTER TO INSULATION TESTER TERMINALS2. SET CONTINUITY RANGE ON TESTER3. MEASURE CONTINUITY CURRENT FROM TESTER

NOTE : 1 OHM RANGE MUST BE SET ON 3200 TO ENABLECURRENT MEASUREMENT

CURRENT MEASUREMENT MODE ENABLED BY USING 1 OHM CONTINUITY RANGE

INSULATION TESTER CALIBRATIONContinuity Current

Page 6: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

1. CONNECT TESTER TO INSULATION TESTER TERMINALS

2. SET INSULATION RANGE ON TESTER

3. SET INSULATION RESISTANCE ON 3200, EG. 10 MOHM

4. MEASURE RESISTANCE USING TESTER

INSULATION TESTER CALIBRATIONInsulation Resistance

Page 7: INSULATION TESTER CALIBRATION Using the 3200 Electrical Test Calibrator

1. CONNECT TESTER TO INSULATION TESTER TERMINALS2. SET INSULATION TEST VOLTAGE RANGE ON TESTER3. MEASURE INSULATION TEST VOLTAGE & CURRENT FROM TESTER

BY SELECTING THE VOLTAGE MEASUREMENT FUNCTION USING THE SOFT KEY (SHOWN BELOW)

VOLTAGE / CURRENT MEASUREMENT ENABLED BY USING THE SOFT KEY

INSULATION TESTER CALIBRATIONInsulation Test Voltage