intra-arterial blood pressure monitoring
Embed Size (px)
TRANSCRIPT

INVASIVE ARTERIAL PRESSURE MONITORING
Dr. T. Vikram Kumar Naidu MD(Anaesthesia),
(DM) (Cardiac anaesthesia) (UNMICRC)

01/05/2023 Dr. Vikram Naidu 2
SUBTITLES• Introduction
• Indications
• Basic priciples
• Percutaneous radial artery cannulation
• Complications
• Components
• Levelling and zeroing
• Normal arterial pressure waveforms
• Arterial blood pressure gradients
• Abnormal arterial pressure
waveforms
• Waveform analysis for prediction of
intravascular volume responsivenesss

01/05/2023 Dr. Vikram Naidu 3
INTRODUCTION
• Intra-arterial blood pressure (IBP) measurement is often considered to
be the gold standard of blood pressure measurement.
• Despite its increased risk, cost, and need for technical expertise for
placement and management, its utility in providing crucial and timely
information outweighs its risks in many cases

01/05/2023 Dr. Vikram Naidu 4
Indications for Arterial CannulationContinuous, real-time blood pressure monitoring
Planned pharmacologic or mechanical cardiovascular manipulation
Repeated blood sampling
Failure of indirect arterial blood pressure measurement, e.g. burns or
obesity
Supplementary diagnostic information from the arterial waveform

01/05/2023 Dr. Vikram Naidu 5
BASIC PRICIPLES
• The pressure waveform of the arterial pulse is transmitted via the
column of fluid, to a pressure transducer where it is converted into an
electrical signal.
• This electrical signal is then processed, amplified and converted into a
visual display by a microprocessor.

01/05/2023 Dr. Vikram Naidu 6
Percutaneous Radial Artery Cannulation
• The radial artery is the most common site for invasive blood pressure
monitoring because it is technically easy to cannulate and
complications are uncommon
• Modified Allen’s test

01/05/2023 Dr. Vikram Naidu 7

01/05/2023 Dr. Vikram Naidu 8
• “Transfixation” technique

01/05/2023 Dr. Vikram Naidu 9
ULTRASOUND IMAGING

01/05/2023 Dr. Vikram Naidu 10
Alternative Arterial Pressure Monitoring Sites
• Ulnar
• Brachial
• Axillary
• Femoral – seldinger technique
• Dorsalis pedis

01/05/2023 Dr. Vikram Naidu 11
Complications of Direct Arterial Pressure Monitoring• Hemorrhage• Misinterpretation of data• Distal ischemia• Pseudoaneurysm• Arteriovenous fistula• Arterial embolization• Infection• Peripheral neuropathy

01/05/2023 Dr. Vikram Naidu 12

01/05/2023 Dr. Vikram Naidu 13

01/05/2023 Dr. Vikram Naidu 14

01/05/2023 Dr. Vikram Naidu 15
PHYSICAL PRINCIPLES
• A wave is a disturbance that travels through a medium, transferring
energy but not matter.
• One of the simplest waveforms is the sine wave

01/05/2023 Dr. Vikram Naidu 16
• Fourier Analysis• The arterial waveform is clearly not a simple sine wave, but it can be
broken down into a series of many component sine waves• The process of analysing a complex waveform in terms of its
constituent sine waves is called Fourier Analysis.

01/05/2023 Dr. Vikram Naidu 17
Properties
• Natural frequency
• Damping coefficient

01/05/2023 Dr. Vikram Naidu 18
• The natural frequency of a system determines how rapidly the system
oscillates after a stimulus
• The damping coefficient reflects frictional forces acting on the system
and determines how rapidly it returns to rest after a stimulus

01/05/2023 Dr. Vikram Naidu 19
Natural Frequency
• It is important that the IBP system has a very high natural frequency –
at least eight times the fundamental frequency of the arterial waveform
(the pulse rate).
• Therefore, for a system to remain accurate at heart rates of up to
180bpm, its natural frequency must be at least: (180bpm x 8) / 60secs
= 24Hz.

01/05/2023 Dr. Vikram Naidu 20
Natural Frequency
• The natural frequency of a system may be increased by:
Reducing the length of the cannula or tubing
Reducing the compliance of the cannula or diaphragm
Reducing the density of the fluid used in the tubing
Increasing the diameter of the cannula or tubing
• Commercially available systems -200Hz

01/05/2023 Dr. Vikram Naidu 21
Damping
• Anything that reduces energy in an oscillating system will reduce the
amplitude of the oscillations. This is termed damping.
• Some degree of damping is required in all systems (critical damping),
but if excessive (overdamping) or insufficient (underdamping) the
output will be adversely effected.

01/05/2023 Dr. Vikram Naidu 22
Underdamped arterial pressure waveform Overdamped arterial pressure waveform

01/05/2023 Dr. Vikram Naidu 23
Factors that cause overdamping include:
Friction in the fluid pathway
Three way taps
Bubbles and clots
Vasospasm
Narrow, long or compliant tubing
Kinks in the cannula or tubing

01/05/2023 Dr. Vikram Naidu 24
FAST-FLUSH TEST
• Provides a convenient bedside method for determining dynamic
response of the system
• Natural frequency is inversely proportional to the time between
adjacent oscillation peaks
• The damping coefficient can be calculated mathematically, but it is
usually determined graphically from the amplitude ratio

01/05/2023 Dr. Vikram Naidu 25

01/05/2023 Dr. Vikram Naidu 26
COMPONENTS OF AN IBP MEASURING SYSTEM

01/05/2023 Dr. Vikram Naidu 27
COMPONENTS OF AN IABP MEASURING SYSTEM• Intra-arterial cannula

01/05/2023 Dr. Vikram Naidu 28
COMPONENTS OF AN IABP MEASURING SYSTEM
• Intra-arterial cannula
• Fluid filled tubing

01/05/2023 Dr. Vikram Naidu 29
COMPONENTS OF AN IABP MEASURING SYSTEM
• Intra-arterial cannula
• Fluid filled tubing
• Transducer

01/05/2023 Dr. Vikram Naidu 30
COMPONENTS OF AN IBP MEASURING SYSTEM
• Intra-arterial cannula
• Fluid filled tubing
• Transducer
• Infusion/flushing system

01/05/2023 Dr. Vikram Naidu 31
COMPONENTS OF AN IBP MEASURING SYSTEM
• Intra-arterial cannula
• Fluid filled tubing
• Transducer
• Infusion/flushing system
• Signal processor, amplifier and display

01/05/2023 Dr. Vikram Naidu 32
Levelling and zeroingZeroing :• For a pressure transducer to read accurately, atmospheric pressure
must be discounted from the pressure measurement.
• This is done by exposing the transducer to atmospheric pressure and
calibrating the pressure reading to zero.
• The level of the transducer is not important.

01/05/2023 Dr. Vikram Naidu 33

01/05/2023 Dr. Vikram Naidu 34
•Levelling :• The pressure transducer must be set at the appropriate level in relation
to the patient in order to measure blood pressure correctly.
• This is usually taken to be level with the patient’s heart, at the 4th
intercostal space, in the mid-axillary line.
• A transducer too low over reads, a transducer too high under reads.

01/05/2023 Dr. Vikram Naidu 35

01/05/2023 Dr. Vikram Naidu 36
Normal Arterial Pressure Waveforms
• The systolic waveform components consist of a steep pressure
upstroke, peak, and ensuing decline, and immediately follow the ECG
R wave.
• The downslope of the arterial pressure waveform is interrupted by the
dicrotic notch, continues its decline during diastole after the ECG T
wave, and reaches its nadir at end-diastole

01/05/2023 Dr. Vikram Naidu 37

01/05/2023 Dr. Vikram Naidu 38
• As the pressure wave travels from the central aorta to the periphery,
the arterial upstroke becomes steeper, the systolic peak increases, the
dicrotic notch appears later, the diastolic wave becomes more
prominent, and end-diastolic pressure decreases.

01/05/2023 Dr. Vikram Naidu 39

01/05/2023 Dr. Vikram Naidu 40
Arterial Blood Pressure Gradients
• The nature of the operative procedure is important when choosing the
appropriate site
Ex:
• Coarctation of aorta
• Thoracic and abdominal aortic surgeries
• Cardiopulmonary bypass

01/05/2023 Dr. Vikram Naidu 41
Cardiopulmonary bypass :• The mean radial artery pressure decreases on initiation of bypass and
remains less than mean femoral artery pressure throughout the bypass
period.
• Persists in the first few minutes following separation from bypass,
often by more than 20 mm Hg.

01/05/2023 Dr. Vikram Naidu 42

01/05/2023 Dr. Vikram Naidu 43
Abnormal Arterial Pressure Waveforms• Morphologic features of individual arterial pressure waveforms can
provide important diagnostic information

01/05/2023 Dr. Vikram Naidu 44
Condition CharacteristicsAortic stenosis Pulsus parvus (narrow pulse pressure)
Pulsus tardus (delayed upstroke)
Aortic regurgitation Bisferiens pulse (double peak)Wide pulse pressure
Hypertrophic cardiomyopathy Spike and dome (mid-systolic obstruction)
Systolic left ventricular failure Pulsus alternans (alternating pulse pressure amplitude)
Cardiac tamponade Pulsus paradoxus (exaggerated decrease in systolic blood pressure during spontaneous inspiration)

01/05/2023 Dr. Vikram Naidu 45

01/05/2023 Dr. Vikram Naidu 46
Waveform analysis for prediction of intravascular volume responsiveness
• Variations in arterial blood pressure observed during positive pressure
ventilation, as well as a variety of derived indices, are the most widely
studied of these dynamic indicators.
• They result from changes in intrathoracic pressure and lung volume
that occur during the respiratory cycle.

01/05/2023 Dr. Vikram Naidu 47

01/05/2023 Dr. Vikram Naidu 48
REFERENCESMiller's 8th edition
Physical principles of intra-arterial blood pressure measurement anaesthesia
Kaplan's cardiac anesthesia the echo era 6edition